July 2001 Incubators, Infant Purpose An infant incubator provides a closed, controlled environment that warms an infant by circulating heated air over the skin. The heat is then absorbed into the body by tissue conduction and blood convection. Ideally, both the skin and core temperatures should be maintained with only minor variations. Incubators feature better temperature control than radiant warm- ers (overhead heating units), which can overheat or underheat the infant. However, an infant in an incu- bator is less accessible to the caregiver. At birth, an infant’s core and skin temperatures tend to drop significantly because of heat loss from conduction (heat loss to cooler surfaces in direct con- tact with the infant), convection (heat loss to air mov- ing past the infant), radiation (heat loss to cooler objects not in direct contact with the infant), and water evaporation (heat loss from the infant’s lungs and skin surface). Whereas term neonates naturally regulate 211887 424-008 Scope of this Product Comparison This report covers infant incubators used for in- tensive, intermediate, and primary care. It also covers transport infant incubators and one unit that can be used as either an incubator or a radiant warmer. Dedicated radiant warmers are covered in the report titled WARMERS, RADIANT, INFANT. UMDNS information This Product Comparison covers the following device terms and product codes as listed in ECRI’s Universal Medical Device Nomenclature System™ (UMDNS™): Incubator/Radiant Warming Units, Infant, Mobile [18-857] Incubators, Infant, Mobile [17-432] Incubators, Infant, Transport [12-114] 5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA Telephone +1 (610) 825-6000 Fax +1 (610) 834-1275 E-mail [email protected] Mobile incubator

description

Descripcion del principio de funcionamieto de las icubadoras Neonatales

Transcript of incubadoras neonatales

Page 1: incubadoras neonatales

July 2001

Incubators, Infant

PurposeAn infant incubator provides a closed, controlled

environment that warms an infant by circulatingheated air over the skin. The heat is then absorbed intothe body by tissue conduction and blood convection.Ideally, both the skin and core temperatures should bemaintained with only minor variations. Incubatorsfeature better temperature control than radiant warm-ers (overhead heating units), which can overheat orunderheat the infant. However, an infant in an incu-bator is less accessible to the caregiver.

At birth, an infant’s core and skin temperaturestend to drop significantly because of heat loss from

conduction (heat loss to cooler surfaces in direct con-tact with the infant), convection (heat loss to air mov-ing past the infant), radiation (heat loss to coolerobjects not in direct contact with the infant), and waterevaporation (heat loss from the infant’s lungs and skinsurface). Whereas term neonates naturally regulate

211887424-008

Scope of this Product ComparisonThis report covers infant incubators used for in-tensive, intermediate, and primary care. It alsocovers transport infant incubators and one unitthat can be used as either an incubator or aradiant warmer. Dedicated radiant warmers arecovered in the report titled WARMERS, RADIANT,INFANT.

UMDNS informationThis Product Comparison covers the followingdevice terms and product codes as listed inECRI’s Universal Medical Device NomenclatureSystem™ (UMDNS™):

• Incubator/Radiant Warming Units, Infant,Mobile [18-857]

• Incubators, Infant, Mobile [17-432]

• Incubators, Infant, Transport [12-114]

5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USATelephone +1 (610) 825-6000 ● Fax +1 (610) 834-1275 ● E-mail [email protected]

Mobile incubator

Page 2: incubadoras neonatales

their body temperature to some extent, prematureinfants have thinner skin, which allows surface bloodvessels to more readily lose heat to the environment; alarge ratio of surface area to volume, resulting ingreater heat losses from radiation and convection; andalmost no subcutaneous fat to either metabolize intoheat or act as an insulator. Prolonged cold stress inneonates can cause oxygen deprivation, hypoglycemia,metabolic acidosis, and rapid depletion of glycogenstores; therefore, energy conservation provided bythermal support is critical.

Principles of operationThe neonate lies on a mattress in the infant com-

partment, which is enclosed by a clear plastic hood.Most incubators have hand access ports with doorsthat permit the infant to be handled while limiting theintroduction of cool room air. The clinician can raise orremove the plastic hood or open a panel to gain greateraccess to the infant. Some units feature an air curtainthat causes warm air to sweep past the opening.

Most incubators warm the infant by a forced ornatural flow of heated air. At least one unit supple-ments air convection by actively warming the incuba-tor walls to reduce radiant heat loss. Another unit usesa mattress of warm water, rather than a convectiveairflow, to warm the infant.

Heating and humidification systems are located be-neath the infant compartment. A fan or natural flowcirculates air past the heater and the temperaturemeasuring device, over a water reservoir used to hu-midify the air (if desired), and up into the infantcompartment. Most incubators are equipped with pro-portional heating controls that provide electricalpower to the heating coil in response to the differencebetween the actual temperature and the desired tem-perature. Most units have two modes of operation:air-temperature control and skin-temperature control.With the air-temperature (manual) control, the opera-tor sets the temperature of the air in the incubator;changes in infant body temperature are usually meas-ured periodically with a thermometer, and adjust-ments in air temperature are made accordingly. In theskin-temperature control mode, also called the servo(automatic) mode, a sensor is taped to the infant’s skin,and the heater responds to changes in the sensor tokeep the skin temperature at the preset level.

Most units allow the user to vary relative humidityfrom either a built-in reservoir or an outside source(e.g., a humidifier that attaches to one of the inletports). Although increasing the relative humidity in anincubator can reduce evaporative heat loss, many cli-nicians avoid supplemental humidification because of

concern that infectious organisms may proliferate inthe water reservoir.

Many incubators have one or two oxygen inlet portsand can be equipped with optional oxygen controllers.The incubators can also provide support and protec-tion for oxygen cylinders when oxygen must be deliv-ered to the infant in the incubator.

Because the room temperature of the nursery isnearly always lower than that inside the incubator,radiant heat loss through the incubator walls accountsfor as much as half of the infant’s total heat loss. Insome nurseries, a plastic heat shield is placed over theinfant inside the incubator to minimize radiant heatloss. In addition, some incubators have double wallsseparated by an air space to prevent excessive heatloss. However, in a study comparing heat loss fromservo-regulated single- and double-walled incubators,Bell and Rios (1983) reported that, although the dou-ble-walled incubator decreased radiant heat loss, itincreased convective heat loss; total heat loss andmetabolic heat production were the same as in single-walled incubators.

Incubator/radiant warming units

One manufacturer’s device combines the thermalcapabilities of an infant incubator with the accessadvantages of an infant radiant warming unit. Thishybrid unit provides a centralized control and moni-toring system for both the incubator and the warmingunit; it can be converted automatically from a closedincubator to an open bed radiant warming unit andvice versa with little effort in a few seconds. This deviceeliminates the need to transfer an infant between anincubator and a radiant warming unit.

Transport incubator

Healthcare Product Comparison System

2 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 3: incubadoras neonatales

Transport incubators

During transport within a hospital or to anotherfacility, infants require thermal support. Transportincubators can be operated on a variety of powersources found in hospitals or transport vehicles (e.g.,120 VAC, 12 VDC, 24 VDC). Most units also carry theirown backup power supplies (e.g., 12 VDC rechargeablenickel-cadmium [Ni-Cd], gel cell, or lead-acid batter-ies) that power the unit for short periods in the absenceor failure of other power sources. In addition to havinga reliable portable power source, transport incubatorsalso need to meet strength requirements for air andground travel, provide isolation from external noiseand vibration, and have limited electromagnetic emis-sions to allow for aircraft use. Transport incubators aretypically smaller and lighter than mobile or stationaryincubators, facilitating easy maneuvering in and outof emergency vehicles.

Reported problemsDeaths and injuries to neonates in incubators have

been linked to thermostat failure that caused incuba-tor overheating and infant hyperthermia and to mal-functions or design defects that produced fires andelectric shock hazards.

If a surface temperature sensor does not make ade-quate thermal contact with the infant’s skin, incorrectlow temperature readings may result, and the incuba-tor could overheat the infant in response to the falsereadings. However, using a strongly adherent tape tofasten the sensor can produce skin irritation or dam-age. Periodic checks of sensor placement are stronglyrecommended.

Inadequate control over the amount of oxygen deliv-ered in an incubator can cause hyperoxia or hypoxia(excessive or inadequate oxygen, respectively, in bodytissues). Excessively high concentrations of oxygen cancontribute to retrolental fibroplasia (i.e., formation offibrous tissue behind the lens, leading to detachmentof the retina and arrested growth of the eye) in prema-ture infants.

Incubators can produce excessive internal noise lev-els that might adversely affect infant hearing. Slam-ming a porthole, using a nebulizer, activating analarm, and tapping on the incubator hood furtherincrease the noise level to the infant; levels as high as90 to 100 dB have been reported. The 1991 Associationfor the Advancement of Medical Instrumentationstandard for incubators specifies that the maximumnoise level inside the infant compartment should notexceed 60 dB during normal use.

In a case reported to ECRI, an infant was able topush open the rear isolette door of a mobile infantincubator due to improper door gasket placement. Themattress in the isolette placed the infant at access-doorlevel, and the infant fell out of the incubator to thefloor.

Some older incubators still in use may have hoodthermometers or high-temperature thermostats (orboth) that are filled with mercury. Mercury contami-nation is possible if the thermometer or thermostatbreaks, which is most likely during routine cleaning ormaintenance. Several investigators have expressedconcern about possible adverse effects of an infant’sexposure to mercury vapor. It is not known to whatextent safe limits for chronic adult exposure apply tosick or premature infants who remain in incubators forextended periods of time. Some investigators recom-mend replacing all mercury-containing components ininfant incubators, and most hospitals have followedthis course of action.

ECRI has published an infant incubator checklist(see the citation from the Health Devices Inspectionand Preventive Maintenance System below) that canhelp detect many potential problems with incubatorequipment before they occur.

Purchase considerationsInfant incubators typically cost between $4,000 and

$14,000, depending on accessories purchased. Higher-priced incubators may include such features as amechanism that minimizes the temperature dropwhen the front-access panel is opened, a rail system tomount accessories, a humidification system, and anelectrical height adjustment capability.

Hospitals can purchase service contracts or serviceon a time-and-materials basis from the supplier. Serv-ice may also be available from a third-party organiza-tion. The decision to purchase a service contract shouldbe carefully considered. Purchasing a service contractensures that preventive maintenance will be per-formed at regular intervals, eliminating the possibilityof unexpected maintenance costs. Also, many suppli-ers do not extend system performance and uptimeguarantees beyond the length of the warranty unlessthe system is covered by a service contract.

ECRI recommends that, to maximize bargainingleverage, hospitals negotiate pricing for service con-tracts before the system is purchased.

Stage of developmentManufacturers continue to improve the design of

incubators, focusing on thermal protection. Many

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 3

Page 4: incubadoras neonatales

infant incubators may be upgraded to an intensive careworkstation with several accessories.

BibliographyBell EF. Infant incubators and radiant warmers. Early

Hum Dev 1983;8:351-75.

Bell EF, Rios GR. A double-walled incubator alters thepartition of body heat loss of premature infants.Pediatr Res 1983;17:135-40.

Bell EF, Rios GR. Air versus skin temperature servo-control of infant incubators. J Pediatr 1983 Dec;103(6):954-9.

Hill DW, Dolan AM. Intensive care instrumentation.2nd ed. London (England): Academic Press; 1982.

Korones SB, Fitch CW. Thermal management in thecare of sick neonates: a clinical appraisal. Med In-strum 1982 May-Jun;16(3):163-5.

Marks KH, Lee CA, Bolan CD Jr, et al. Oxygen con-sumption and temperature control of prematureinfants in a double-wall incubator. Pediatrics 1981Jul;68(1):93-8.

Neuman MR. Neonatology, obstetrics, and gynecology.In: Cook AM, Webster JG, eds. Therapeutic medicaldevices: application and design. Englewood Cliffs(NJ): Prentice-Hall; 1982:522-50.

Perlstein PH. Future directions for device design andinfant management. Med Instrum 1987 Feb;21(1):36-41.

Poole DR. Challenges in the design of transport incu-bators. Biomed Instrum Technol 1997 Mar-Apr;31(2):137-42.

Sjörs G, Hammarlund K, Öberg P, et al. An evaluationof environment and climate control in seven infantincubators. Biomed Instrum Technol 1992 Jul-Aug;26(4):294-301.

Standards and guidelinesNote: Although every effort is made to ensure that thefollowing list is comprehensive, please note that otherapplicable standards may exist.

American National Standards Institute/Associationfor the Advancement of Medical Instrumentation.Infant incubators [standard]. ANSI/AAMI II36.1984 (revised 1998).

New standard offers guidelines for infant incubatorsafety [standard]. ANSI/AAMI II36-1997. 1991 (re-vised 1998).

Safe current limits for electromedical apparatus[standard]. 3rd ed. ANSI/AAMI ES1-1993. 1985 (re-vised 1993).

Transport infant incubators [standard].ANSI/AAMI II51-1996. 1996.

Association of Women’s Health, Obstetric, and Neona-tal Nurses. Transition to an open crib for the verylow birth weight infant [miscellaneous]. Committeeon Practice. CC1.

British Standards Institution. Specification for nurs-ing incubators [standard]. BS 5724:Section 2.20.1982.

Specification for transport incubators [standard].BS 5724:Section 2.21. 1983.

International Electrotechnical Commission. Medicalelectrical equipment — part 1: general require-ments for safety [standard]. IEC 60601-1 (1988-12).1988.

Medical electrical equipment — part 1: general re-quirements for safety. Amendment 1 [standard].IEC 60601-1-am1 (1991-11). 1991.

Medical electrical equipment — part 1: general re-quirements for safety. Amendment 2 [standard].IEC 60601-1-am2 (1995-03). 1995.

Medical electrical equipment — part 1: general re-quirements for safety. Section 2. Collateral stand-ard: electromagnetic compatibility — requirementsand tests. IEC 60601-1-2 (1993-04). 1993.

Medical electrical equipment — part 2: particularrequirements for the safety of baby incubators[standard]. IEC 60601-2-19. (1990-12). 1990.

Medical electrical equipment — part 2: particularrequirements for the safety of baby incubators.Amendment 1 [standard]. IEC 60601-2-19-am1(1996-10). 1996.

Medical electrical equipment — part 2: particularrequirements for the safety of transport incubators[standard]. IEC 60601-2-20 (1990-12). 1990.

Medical electrical equipment — part 2: particularrequirements for the safety of transport incubators.Amendment 1 [standard]. IEC 60601-2-20-am1(1996-10). 1996.

Medical Devices Agency. Nursing incubators.Evaluation 1998 May: No. 362.

Citations from other ECRI publicationsHealth Devices

Infant incubators [evaluation]. 1982 May;11:191-9.

Healthcare Product Comparison System

4 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 5: incubadoras neonatales

High-risk equipment: outdated incubators. 1988 Feb;17(2):50.

Reversal of access port gaskets on Hill-Rom Air-Shields Isolette infant incubators [hazard report].Health Devices 1998 Sept-Oct;27(9-10):370-1.

Cable failure causes frequent alarms — and loss ofcontrol — on Hill-Rom Air-Shields C2HS Isoletteinfant incubators [User Experience Network™].Health Devices 1999 Aug;28(8):326-7.

Deposits may form in humidifier of Hill-Rom Air-Shields C2HS Isolette infant incubators [User Ex-perience Network™]. Health Devices 2001 Mar;30(3):94.

Health Devices Alerts

This Product Comparison lists Health Devices Alerts(HDA) citations published since the last update of thisreport. Each HDA abstract is identified by an Acces-sion Number. Recalls and hazard reports include de-scriptions of the problem involved; abstracts of otherpublished articles are referenced by bibliographic in-formation. HPCS subscribers can call the Hotline foradditional information on any of these citations or torequest more extensive searches of the HDA database.

A4070 FDA has designated Class II Recall Nos. Z-234/237-0 certain Hill-Rom Air-Shields K-1 infanttransport incubators, which have a faulty design of thehand ventilator that does not indicate the true pres-sure inside the incubator. The manufacturer initiateda field correction by site visits beginning October 15,1999, and by letters dated March 3 and April 6, 2000.Verify that you have received the letters, and identifyand isolate any affected product in your inventory.Hill-Rom Air-Shields will dispatch service personnelto your facility to address the subject of product im-provement by offering your facility two corrective op-tions. Source: FDA Enforcement Rep 1999 Dec 5;Manufacturer.

A4363 O2 concentration variations can occur inOhmeda Medical Giraffe OmniBed infant incubator/radiant warming units, resulting in too much or toolittle O2 inside the infant compartment when the unitsare operated in incubator mode. The variations occuras a result of changes in the fan speed when the unitsare operated in incubator mode and free-flow O2 isadministered. Changes in the fan speed may occur atany time in response to the temperature of the bedheater or as the result of intentional user activation.The manufacturer states that it has not received anyreports of injuries arising from O2 concentration vari-ations. Injury is possible only if (1) free-flow O2 is beingadministered, (2) no servo-controlled O2 delivery sys-tem is being used, and (3) the O2 saturation of the

patient is not being monitored. The manufacturer in-itiated a safety alert by letter dated February 28, 2001.Verify that you have received the letter, prepaid re-sponse form, and operation and maintenance manualreplacement pages. Identify any affected product inyour inventory. Inform all caregivers and maintenancepersonnel who use or service these units of the safetyalert. If you are a dealer or distributor, or if your officesupplies medical equipment to other locations withinyour institution, send a copy of the notification letterand response form to your customers or other locationswithin your facility. To obtain additional copies of theletter and response form, contact the Ohmeda Medicalsafety alert coordinator by telephone at (800) 249-5026within the U.S. or at (410) 888-5213 outside the U.S.For further information, contact the Ohmeda Medicalsafety alert coordinator by telephone at (800) 249-5026within the U.S. or at +1 (410) 888-5213 outside the U.S.Source: Manufacturer.

Health Devices Inspection and Preventive MaintenanceSystem

Infant incubators. 415.

Supplier information

Airborne Life Support

Airborne Life Support SystemsDiv International Biomedical Inc [101024]8508 Cross Park DrAustin TX 78754-4557Phone: (512) 873-0033, (800) 433-5615Fax: (512) 873-9090E-mail: [email protected]: http://www.int-bio.com

Air-Shields

Hill-Rom Air-ShieldsA Hillenbrand Co [339679]330 Jacksonville RdHatboro PA 19040-2211Phone: (215) 675-5200, (800) 523-5756Fax: (215) 674-3497Internet: http://www.hill-rom.com

Hill-Rom Air-Shields (Australia)A Hillenbrand Co [339681]5-9 Devlin St Suite 402 Ground FloorRyde NSW 2112AustraliaPhone: 61 (2) 98096622

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 5

Page 6: incubadoras neonatales

Hill-Rom Air-Shields (UK)A Hillenbrand Co [339680]Unit 9 Manor WayOld Woking, Surrey GU22 9JYEnglandPhone: 44 (1483) 246270E-mail: [email protected]: http://www.hill-rom.co.uk

Atom Medical

Atom Medical Corp [154422]3-18-15 HongoBunkyo-kuTokyo 113JapanPhone: 81 (3) 38152311Fax: 81 (3) 38152997Internet: http://www.atomed.co.jp

Parker Healthcare Pty Ltd [354556]2 Monomeeth DrMitcham VIC 3132AustraliaPhone: 61 (3) 98735822Fax: 61 (3) 98735911E-mail: [email protected]: http://www.parkerhealth.com.au

Bertocchi

Giacomo Bertocchi Co [187952]Pokrovka Strasse 41/8 Bldg 2103062 MuskvaRussiaPhone: 7 (095) 9613527E-mail: [email protected]

Giacomo Bertocchi SNC [151077]via de Berenzani 8 a/b/cNuova Zona Industriale Casa Postale 115I-26100 Cremona CRItalyPhone: 39 (0372) 471000Fax: 39 (0372) 471533E-mail: [email protected]: http://www.giacomobertocchi.it

Bio MS

AIROX [382727]Parc d’ActivitesPau Pyrenees l’EchangeurF-64008 Pau CedexFrancePhone: 33 (559) 140202Fax: 33 (559) 140200E-mail: [email protected]: http://www.airox.fr

Draeger

Draeger Ltd [157747]The WillowsMark RdHemel Hempstead, Hertfordshire HP2 7BWEnglandPhone: 44 (1442) 213542Fax: 44 (1442) 240327Internet: http://www.draeger.co.uk

Draeger Medical Inc [371341]3135 Quarry RdTelford PA 18969Phone: (215) 721-5400, (800) 462-7566Fax: (215) 721-9561Internet: http://www.draegermedical.com

DraegerMedizintechnik GmbH [374044]Moislinger Allee 53-55Postfach 1339D-23542 LuebeckGermanyPhone: 49 (451) 8820Fax: 49 (451) 8822080E-mail: [email protected]: http://www.draeger.com

Draeger Southeast Asia Pte LtdMedical Div [354511]73 Science Park Dr #02-01/04Cintech 1SingaporeRepublic of SingaporePhone: 65 8729278Fax: 65 7792165

Fanem

Fanem Fabr de Apars Nac de Eletomedicina Ltda[174321]Avenida General Ataliba Leonel 1790Caixa Postal 283602033-020 Sao Paulo-SPBrazilPhone: 55 (11) 69791700Fax: 55 (11) 64122199E-mail: [email protected]: http://www.fanem.com

Healthcare Product Comparison System

6 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 7: incubadoras neonatales

Gigante Recem-Nascido

Gigante Recem-NascidoIndustria e Comercio e Representacoes Ltda[174377]Avenida Marechal Costa e Silva 354114075-610 Ribeirao Preto-SPBrazilPhone: 55 (16) 39691000Fax: 55 (16) 39691001E-mail: [email protected]: http://www.gigante.com

Ginevri

Ginevri srl [306412]via Cancelliera 25/DI-00040 Cecchina RMItalyPhone: 39 (06) 934591Fax: 39 (06) 93459393E-mail: [email protected]: http://www.ginevri.com

Guido Rayos

Guido Rayos X SA [309753]Calle Salcedo 5E-28034 MadridSpainPhone: 34 (91) 3581625Fax: 34 (91) 3580869E-mail: [email protected]

Hero Medical

Hero Medical Pty Ltd [177031]40 St David StFitzroy VIC 3065AustraliaPhone: 61 (3) 94195122

Magna

Magna Medical Systems Inc [377697]7200 NW 7th StMiami FL 33126Phone: (305) 261-2211Fax: (305) 261-4492E-mail: [email protected]: http://www.magnamedical.com

Medap

Heraeus Medical GmbH [254159]Heraeusstrasse 12-14D-63450 HanauGermanyPhone: 49 (6181) 355100Fax: 49 (6181) 35809E-mail: [email protected]: http://www.heraeus.de

Medicor

Medicor Elektronika Rt [277820]Illatos utca 9H-1097 BudapestHungaryPhone: 36 (1) 2806342Fax: 36 (1) 2806336E-mail: [email protected]: http://www.datanet.hu/medelektronika

Mediprema

Mediprema SA [151024]Technopole 4 avenue Marcel Merieuxboite postale 3F-37913 Tours Cedex 9FrancePhone: 33 (47) 284787E-mail: [email protected]: http://www.mediprema.com

Medix

Medix icsa [157008]Jose Arias 2931672 Villa Lynch San MartinPcia de Buenos AiresArgentinaPhone: 54 (11) 47545555Fax: 54 (11) 47541713E-mail: [email protected]: http://www.medix.com.ar

Ohmeda

Datex-Ohmeda (Finland)An Instrumentarium Co [351977]Kuortaneenkat 2Postilokero 900FIN-00031 HelsinkiFinlandPhone: 358 (10) 39411Fax: 358 (9) 1463310E-mail: [email protected]: http://www.datex-ohmeda.com

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 7

Page 8: incubadoras neonatales

Datex-Ohmeda (India) Pvt Ltd [354295]Block EP & GP Sector VPlot XI-16 Salt LakeCalcutta 700 091IndiaPhone: 91 (33) 3574002Fax: 91 (33) 3574001E-mail: [email protected]: http://www.datex-ohmeda.com

Datex-Ohmeda Ltd (UK)An Instrumentarium Co [354403]Ohmeda House71 Great North RdHatfield, Hertfordshire AL9 5ENEnglandPhone: 44 (1707) 263570Fax: 44 (1707) 260065Internet: http://www.datex-ohmeda.com

Ohmeda MedicalAn Instrumentarium Co [347601]8880 Gorman RdLaurel MD 20723Phone: (410) 888-5200, (800) 527-9209Fax: (410) 888-0545Internet: http://www.ohmedamedical.com

Parker

Parker Healthcare Pty Ltd [354556]2 Monomeeth DrMitcham VIC 3132AustraliaPhone: 61 (3) 98735822Fax: 61 (3) 98735911E-mail: [email protected]: http://www.parkerhealth.com.au

Phoenix Medical

Phoenix Medical Systems (P) Ltd [346929]32/4 Jawaharlal Nehru SalaiEkkaduthaugalChennai 600 097IndiaPhone: 91 (44) 2317867Fax: 91 (44) 2317894E-mail: [email protected]: http://www.phoenixmedicalsystems.com

Seward

Seward Ltd [178349]98 Great North RoadLondon N2 OGNEnglandPhone: 44 (20) 83654100Fax: 44 (20) 83653999E-mail: [email protected]: http://www.seward.co.uk

TSE

TSE spol s r o [236960]Manesova 390/74CZ-371 52 Ceske BudejoviceCzech RepublicPhone: 420 (38) 6721114Fax: 420 (38) 6721102E-mail: [email protected]: http://www.tse.cz

About the chart specificationsThe following terms are used in the chart:

Backup thermostat: Backup device that deactivates theheater if the incubator’s internal air temperatureexceeds a limit when the primary controller fails.

Tubing ports: Number and type of openings in the incu-bator that allow therapeutic access to the infant.

O2 inlet ports/controllers: Number of openings in theincubator for tubing (oxygen delivery) and con-trol/monitoring of the oxygen delivered.

Internal noise, dB: The sound level inside the incuba-tor, measured in decibels. The values listed includeonly the noise generated by the incubator equip-ment itself.

Abbreviations:

AAMI — Association for the Advancement of Medi-cal Instrumentation

ANSI — American National Standards Institute

ARO — After receipt of order

CE mark — Conformite Europeene mark

CEI — Comitato Elettrotecnico Italiano

CPAP — Continuous positive airway pressure

CSA — Canadian Standards Association

EN — European Norm

ETL — ETL Testing Laboratories

FAA — Federal Aviation Administration

Healthcare Product Comparison System

8 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 9: incubadoras neonatales

IEC — International Electrotechnical Commission

IMV — Intermittent mandatory ventilation

ISO — International Organization for Standardiza-tion

IV — Intravenous

LCD — Liquid crystal display

LED — Light-emitting diode

SIMV — Synchronized intermittent mandatorybreathing

UL — Underwriters Laboratories

VLBW — Very low birth weight

Note: The data in the charts derive from suppli-ers’ specifications and have not been verified throughindependent testing by ECRI or any other agency.Because test methods vary, different products’ specifi-cations are not always comparable. Moreover,

products and specifications are subject to frequentchanges. ECRI is not responsible for the quality orvalidity of the information presented or for any adverseconsequences of acting on such information.

When reading the charts, keep in mind that, unlessotherwise noted, the list price does not reflect supplierdiscounts. And although we try to indicate whichfeatures and characteristics are standard and whichare not, some may be optional, at additional cost.

For those models whose prices were supplied to usin currencies other than U.S. dollars, we have alsolisted the conversion to U.S. dollars to facilitate com-parison among models. However, keep in mind thatexchange rates change often.

Need to know more?For further information about the contents of this

Product Comparison, contact the HPCS Hotline at +1(610) 825-6000, ext. 5265; +1 (610) 834-1275 (fax); [email protected] (e-mail).

About ECRI . . .ECRI is a nonprofit health services research agency and a Collaborating Center of the World HealthOrganization, providing information and technical assistance to the healthcare community to supportsafe and cost-effective patient care for more than 25 years. The results of ECRI’s research andexperience are available through its publications, information systems, databases, technical assistanceprogram, laboratory services, seminars, and fellowships.

Our full-time staff includes a wide range of specialists in healthcare technology, hospital admini-stration, financial analysis, risk management, and information and computer science, as well ashospital planners, attorneys, physicists; biomedical, electrical, electronic, chemical, mechanical, andregistered engineers; physicians; basic medical scientists; epidemiologists and biostatisticians; andwriters, editors, and communications specialists.

Underlying ECRI’s knowledge base in healthcare technology are its integrity and objectivity. ECRIaccepts no financial support from medical product manufacturers, and no employee may own stockin or consult for a medical equipment or pharmaceutical company.

The scope of ECRI’s resources extends far beyond technology. ECRI keeps healthcare professionals,manufacturers, legal professionals, information specialists, and others aware of the changing trendsin healthcare, healthcare standards and regulations, and the best ways to handle environmental andoccupational health and safety issues. ECRI also advises on management issues related to healthcarecost containment, accreditation, risk management, human resources, quality of care, and othercomplex topics.

ECRI has more than 35 publications, databases, software, and services to fulfill the growing needfor healthcare information and decision support. They focus on three primary areas: healthcaretechnology, healthcare risk and quality management, and healthcare environmental management.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 9

Page 10: incubadoras neonatales

Product Comparison Chart

MODEL AIRBORNE LIFE AIRBORNE LIFE AIR-SHIELDS AIR-SHIELDSSUPPORT SUPPORTIn-house Infant Multipurpose Infant C300 ISOLETTE C400 QT ISOLETTETransport Sys 185A Transport Sys 20H Infant Incubator Infant Incubator

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Yes Yes Yes Yes

CE MARK Yes Yes No Yes

MOBILE OR TRANSPORT Ground transport Air and ground Mobile/transport Mobiletransport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 30-38.9 (0.1) 30-38.9 (0.1) 22-38 (0.1) 20-38.5 (0.1)

Skin (increments) No No NA NA

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 20-45 20-45 17-45 17-45

Skin, display type Digital Digital NA NARange, °C 20-45 20-45 NA NA

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes No YesHigh skin temp No No NA NALow skin temp No No NA NAFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Low battery Low battery Vertical lamp bar, Horizontal lamp bar,

indicator indicator 25% increments 20% increments

HAND PORTS 2, with 1 head door 2, with 1 head door 3 6port port

TUBING PORTS 1 respiratory, 1 respiratory, 1 respiratory, 2 respiratory,4 accessory 4 accessory 2 accessory 4 accessory

O2 INLET PORTS/CONTROLLERS Optional/optional 2/1-way check 1 standard/no 1 standard/no

valves

SUPPLEMENTALHUMIDITY Optional Optional Yes Yes

Adjustable Not specified Not specified No Yes *

PHOTOTHERAPY Optional Optional No No

Colons separate data on similar models of a device. This is the first of* The optional Dew-ette 2 incubator humidifier provides ambient to 85% relative humidity. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

10 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 11: incubadoras neonatales

Product Comparison Chart

MODEL AIRBORNE LIFE AIRBORNE LIFE AIR-SHIELDS AIR-SHIELDSSUPPORT SUPPORTIn-house Infant Multipurpose Infant C300 ISOLETTE C400 QT ISOLETTETransport Sys 185A Transport Sys 20H Infant Incubator Infant Incubator

INTERNAL NOISE, dB <60 <60 ≤60 Typically 49

LINE POWER, VAC 105-125/210-250 105-125/210-250 100/120/240 100/120/240

BATTERY, TYPE 12 V sealed lead- 12 V sealed lead- 12 V sealed lead- NAacid gel acid gel acid gel

DOUBLE WALL Yes Yes No Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 22.9 (9) 22.9 (9) 30.5 (12) 40.6 (16)

SIZEIncubator w/stand,

H x W x D, cm (in) 116.8 x 119.4 x 71.1 109.6 x 94.2 x 53.3 129.5 x 83.8 x 55.9 140 x 116 x 55.9(46 x 47 x 28) (43 x 37 x 21) (51 x 33 x 22) (55.1 x 45.7 x 22)

Mattress, W x D,cm (in) 61 x 30.5 61 x 30.5 63.5 x 33 62.9 x 34.9

(24 x 12) (24 x 12) (25 x 13) (24.75 x 13.75)Hood access door,

H x W, cm (in) 22.9 x 51.2 22.9 x 51.2 22.9 x 58.4 32.5 x 85.1(9 x 20.3) (9 x 20.3) (9 x 23) (12.8 x 33.5)

Caster, cm (in) 12.7 (5) 12.7 (5) 10.2 (4) 12.7 (5)

WEIGHT, kg (lb) Variable, depends on Variable, depends on 54.4 (120) 76 (167.6)configuration configuration

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO 30-45 days 30-45 days 30 days 30 daysList price Not specified Not specified $7,500-11,400 $10,000-14,900

Year first sold Not specified Not specified 1982 1994Number sold Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

OTHER SPECIFICATIONS Available with Available with Optional intubation Air curtain withpatient monitor, patient monitor, access door, battery access door open;ventilator, and ventilator, and pack for transport quiet-touchventilator monitor; ventilator monitor; within a hospital, features; in-bedmobile cart; storage detachable accessory IV pole. scale; rail system,shelves available; and air/oxygen cart; shelf, IV pole,optional ambulance capacity for M-, E-, remote alarm module,cart. Meets or Kevlar cylinders; Dew-ette 2 servo-requirements of CSA, optional controlledETL, and UL. noncollapsible cart. humidification

Meets requirements system, variable-of CSA, ETL, FAA, height stand.UL, and UnitedStates Air Force.

Colons separate data on similar models of a device.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 11

Page 12: incubadoras neonatales

Product Comparison Chart

MODEL AIR-SHIELDS AIR-SHIELDS AIR-SHIELDS AIR-SHIELDS

C450 QT ISOLETTE ISOLETTE Infant ti 500 Globetrotter ti 500 InfantInfant Incubator Incubator Transport Systems Transport Incubator

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Yes Yes Yes Yes

CE MARK Yes Yes Yes Yes

MOBILE OR TRANSPORT Mobile Mobile Transport Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-38.5 (0.1) 20-39 (0.1) 22-38 (0.1) 22-38 (0.1)

Skin (increments) 34-37.9 (0.1) 34-38 (0.1) No No

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 17-45 17-45 17-45 17-45

Skin, display type Digital Digital Digital DigitalRange, °C 17-45 17-45 17-45 17-45

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes No NoHigh skin temp Yes Yes No NoLow skin temp Yes Yes No NoFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Horizontal lamp bar, Trended display * Vertical lamp bar, Vertical lamp bar,

20% increments 25% increments 25% increments

HAND PORTS 6 6 3 3

TUBING PORTS 2 respiratory, 2 respiratory, 1 respiratory, 1 respiratory,4 accessory 4 accessory 2 accessory 2 accessory

O2 INLET PORTS/CONTROLLERS 1 standard/no 1 standard/optional 1 standard/NA 1 standard/NA

servo-controlledoxygen

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable Yes ** Optional servo No Nohumidity

PHOTOTHERAPY No No No No

Colons separate data on similar models of a device. This is the first of* 2, 4, 8, 12, or 24 hr time frame. two pages covering** The optional Dew-ette 2 incubator humidifier provides ambient to 85% relative humidity. the above model(s).

These specificationscontinue onto thenext page.

Healthcare Product Comparison System

12 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 13: incubadoras neonatales

Product Comparison Chart

MODEL AIR-SHIELDS AIR-SHIELDS AIR-SHIELDS AIR-SHIELDS

C450 QT ISOLETTE ISOLETTE Infant ti 500 Globetrotter ti 500 InfantInfant Incubator Incubator Transport Systems Transport Incubator

INTERNAL NOISE, dB Typically 49 <47 <56 <56

LINE POWER, VAC 100/120/240 100/120/220/240 120/220/240 110/220/220/240

BATTERY, TYPE NA NA Sealed lead-acid Sealed lead-acidgel, 12 V/24 A-h gel, 12 V/24 A-hvented vented

DOUBLE WALL Yes Yes Yes Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 40.6 (16) 43.2 (17) 31 (12.2); 33 (13) 31 (12.2); 33 (13)high hood high hood

SIZEIncubator w/stand,

H x W x D, cm (in) 140 x 116 x 56 142.2 x 96.5 x 58.4 [54.1-56.1] x 56.6 [54.1-56.1] x 56.6(55.1 x 45.7 x 22) (56 x 38 x 23) * x 95.9 ([21.3-22.1] x 95.9 ([21.3-22.1]

22.3 x 37.8) ** 22.3 x 37.8) **Mattress, W x D,

cm (in) 62.9 x 34.9 40.6 x 80 63.5 x 33 63.5 x 33(24.75 x 13.75) (16 x 31.5) (25 x 13) (25 x 13)

Hood access door,H x W, cm (in) 32.5 x 85.1 35.6 x 85.1 [21-26] x 66.5 [21-26] x 66.5

(12.8 x 33.5) (14 x 33.5) ([8.3-10.2] x 26.2) ([8.3-10.2] x 26.2)Caster, cm (in) 12.7 (5) 12.7 (5) 12.7 (5) 12.7 (5)

WEIGHT, kg (lb) 76 (167.6) 88.4 (195) *** 75.3 (166) † 49.2 (108.5) †

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year, 90 days for 1 year, 90 days for

battery batteryDelivery time, ARO 30 days 30 days 30 days 30 daysList price $10,000-14,900 $10,500-16,900 $32,995-36,675 $10,650-14,970

Year first sold 1994 1996 1995 1994Number sold Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

OTHER SPECIFICATIONS Air curtain with Servo-controlled O2 Integrated ventila- Power on/standbyaccess door open; and humidity; twin tory and gas capa- mode; power sourcequiet-touch capable; dual access bilities; oxygen indicator; batteryfeatures; in-bed doors with dual air monitoring system; charge status; skinscale; rail system, curtain; micropro- hand and mechanical temperature display;shelf, IV pole, cessor controller ventilation; power 2 E- or D-size tankremote alarm module, with message center; on/standby mode; mounts; IV pole;Dew-ette 2 servo- trended data display power source indi- accessory shelf;controlled for air, skin 1/2, cator; battery optional secondhumidification O2, humidity; in-bed charge status; skin battery.system, variable- scale; IV pole; temperature display;height stand. drawer module; re- IV pole; optional

mote alarm module; second battery andvariable-height air compressor.stand.

Colons separate data on similar models of a device.* On cabinet stand.** Minimum and maximum height on adjustable stand.*** On a fixed-height stand.† Without accessories and stand.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 13

Page 14: incubadoras neonatales

Product Comparison Chart

MODEL ATOM MEDICAL ATOM MEDICAL ATOM MEDICAL ATOM MEDICALFAILED TO RESPOND * FAILED TO RESPOND * FAILED TO RESPOND * FAILED TO RESPOND *TC-500 Transport V-85 Infant V-850 Infant V-850W InfantIncubator Incubator Incubator Incubator

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Not specified Not specified Yes Yes

MOBILE OR TRANSPORT Both Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 30-37 (not specifi- 25-38 (0.1) 25-38 (0.1) 25-38 (0.1)ed), override to 39

Skin (increments) NA 34-38 (0.1) 34-38 (0.1) 34-38 (0.1)

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 30-45 20-42 20-42 20-42

Skin, display type NA Digital Digital DigitalRange, °C NA 30-42 30-42 30-42

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes YesHigh skin temp NA Yes Yes YesLow skin temp NA Yes Yes YesFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR 1-point 12-point 10-point 10-point

HAND PORTS 2, additional hand 5 6 6ports optional

TUBING PORTS 2 2 2 2

O2 INLET PORTS/CONTROLLERS 1/optional 1/optional 1/optional 1/optional

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable Ambient to 70% Ambient to 80% Ambient to 80% Ambient to 70%relative humidity relative humidity relative humidity relative humidity

PHOTOTHERAPY Optional Optional Optional Optional

Colons separate data on similar models of a device. This is the first of* Specifications current as of October 1999. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

14 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 15: incubadoras neonatales

Product Comparison Chart

MODEL ATOM MEDICAL ATOM MEDICAL ATOM MEDICAL ATOM MEDICALFAILED TO RESPOND * FAILED TO RESPOND * FAILED TO RESPOND * FAILED TO RESPOND *TC-500 Transport V-85 Infant V-850 Infant V-850W InfantIncubator Incubator Incubator Incubator

INTERNAL NOISE, dB <50 <48 <48 <48

LINE POWER, VAC 110/240 100/120/220/240 100/120/220/240 100/120/220/240

BATTERY, TYPE PP3 alarm Not specified Not specified Not specified

DOUBLE WALL Optional No No Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 23.5 (9.25) 34.5 (13.4) 34.5 (13.4) 31.5 (12.4)

SIZEIncubator w/stand,

H x W x D, cm (in) 42.5 x 83.5 x 43 132 x 98 x 56 132 x 98 x 56 132 x 98 x 56(16.7 x 32.9 x 16.9) (52 x 38.6 x 22) (52 x 38.6 x 22) (52 x 38.6 x 22)

Mattress, W x D,cm (in) 57.5 x 30.5 64 x 34 64 x 34 64 x 34

(22.6 x 12) (25.2 x 13.4) (25.2 x 13.4) (25.2 x 13.4)Hood access door,

H x W, cm (in) 64 x 19.5 28.5 x 65.9 28.5 x 65.9 28.5 x 65.9(25.2 x 7.7) ** (11.2 x 25.9) (11.2 x 25.9) (11.2 x 25.9)

Caster, cm (in) Not specified Not specified Not specified Not specified

WEIGHT, kg (lb) 21 (46.3) 66.5 (146.7) with 72.5 (160) with 76 (168) withcabinet stand cabinet stand cabinet stand

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO Not specified Not specified Not specified Not specifiedList price Y941,000 (US$8,208) Y880,000 (US$7,676) Y1,260,000 Y1,300,000

(US$10,990) (US$11,339)Year first sold 1996 1994 1986 1986Number sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified

OTHER SPECIFICATIONS Mobile power pack; Optional IV pole, MF rail, in-bed Air curtain withtransport trolley; MF rail, in-bed scale; optional IV access door;optional IV pole scale, phototherapy. pole, variable- optional IV pole,stand. height stand, MF rail, in-bed

phototherapy. scale, variable-height stand,phototherapy.

Colons separate data on similar models of a device.* Specifications current as of October 1999.** Hood access door is optional.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 15

Page 16: incubadoras neonatales

Product Comparison Chart

MODEL ATOM MEDICAL BERTOCCHI BERTOCCHI BERTOCCHIFAILED TO RESPOND * FAILED TO RESPOND ** FAILED TO RESPOND ** FAILED TO RESPOND **V-2100 G Advanced GB 35 BR GB 35 RO GB 58 ROInfant Incubator

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Yes Not specified Not specified Not specified

MOBILE OR TRANSPORT Mobile Mobile Mobile Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 23-37 (not spec), 30-37 (0.1) 30-37 (0.1) 30-37 (0.1)override to 39

Skin (increments) 35-37.5 (not spec), 30-37 (0.1) 30-37 (0.1) 30-37 (0.1)override to 39

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 20-42 30-37 30-37 30-37

Skin, display type Digital Digital Digital DigitalRange, °C 30-42 30-37 30-37 30-37

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes No No NoHigh skin temp Yes Yes Yes NoLow skin temp Yes No No NoFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Low battery

BACKUP THERMOSTAT Yes Not specified Not specified Not specified

HEATER POWERINDICATOR 10-point LED LED LED

HAND PORTS 6 4 4 4

TUBING PORTS 6 2 6 1

O2 INLET PORTS/CONTROLLERS 1/optional 1/not specified 1/not specified 1/not specified

SUPPLEMENTALHUMIDITY Yes Yes Yes No

Adjustable Yes Yes Yes NA

PHOTOTHERAPY Optional Optional Optional Optional

Colons separate data on similar models of a device. This is the first of* Specifications current as of October 1999. two pages covering** Specifications current as of March 1998. the above model(s).

These specificationscontinue onto thenext page.

Healthcare Product Comparison System

16 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 17: incubadoras neonatales

Product Comparison Chart

MODEL ATOM MEDICAL BERTOCCHI BERTOCCHI BERTOCCHIFAILED TO RESPOND * FAILED TO RESPOND ** FAILED TO RESPOND ** FAILED TO RESPOND **V-2100 G Advanced GB 35 BR GB 35 RO GB 58 ROInfant Incubator

INTERNAL NOISE, dB <45 <60 <60 <60

LINE POWER, VAC 100/120/220/240 220 220 220

BATTERY, TYPE NA NA NA 12 V

DOUBLE WALL Yes Optional Optional Optional

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 40.3 (16) 34 (13.4) 34 (13.4) 24 (9.4)

SIZEIncubator w/stand,

H x W x D, cm (in) 135 x 100 x 58 99 x 193 x 104 Not specified 63.5 x 96.5 x 152.4(53.1 x 40 x 22.8) (39 x 76 x 41) (25 x 38 x 60)

Mattress, W x D,cm (in) 74 x 36 63 x 34.5 Not specified 24.5 x 56.5

(30.2 x 14.2) (24.8 x 13.6) (9.6 x 22.2)Hood access door,

H x W, cm (in) 40.5 x 85.1 24 x 52 24 x 52 20 x 54.5(15.9 x 33.5) (9.4 x 20.5) (9.4 x 20.5) (8 x 21.5)

Caster, cm (in) 12.7 (5) 8 (3) Not specified 10 (4)

WEIGHT, kg (lb) 94.5 (186) with 110 (242.6) 115 (253.6) 50 (110.3)cabinet stand

PURCHASE INFORMATIONWarranty Not specified 1 year 1 year 1 year

Delivery time, ARO Not specified 6 weeks 6 weeks 6 weeksList price Not specified Not specified Not specified Not specified

Year first sold Not specified 1989 1989 1990Number sold Not specified Not specified Not specified Not specifiedFiscal year Not specified January to December January to December January to December

OTHER SPECIFICATIONS Servo-controlled Meets requirements Meets requirements Meets requirementsO2 and humidity; of IEC. of IEC. of IEC.dual access doorswith dual aircurtain; in-bedscale; IV pole;drawer module;variable-heightstand.

Colons separate data on similar models of a device.* Specifications current as of October 1999.** Specifications current as of March 1998.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 17

Page 18: incubadoras neonatales

Product Comparison Chart

MODEL BERTOCCHI BIO MS BIO MS BIO MSFAILED TO RESPOND *GB 58 RVO NATISSE AR 500 NATISSE AR 500 NATISSE AR 500

Intensive Care Intermediate Care Primary Care

WHERE MARKETED Worldwide Africa, Asia, Africa, Asia, Africa, Asia,Europe, Middle East, Europe, Middle East, Europe, Middle East,Russia Russia Russia

FDA CLEARANCE Not specified No No No

CE MARK Not specified Yes Yes Yes

MOBILE OR TRANSPORT Transport Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 30-37 (0.1) 28-39 (0.1) 28-39 (0.1) 28-39 (0.1)

Skin (increments) 30-37 (0.1) 35-38 (0.1) 35-38 (0.1) 35-38 (0.1)

TEMPERATURE DISPLAYAir, display type Digital LED 8 segments LED 8 segments LED 8 segments

Range, °C 30-37 10-45 10-45 10-45

Skin, display type Digital LED 8 segments LED 8 segments LED 8 segmentsRange, °C 30-37 10-45 10-45 10-45

ALARMSHigh air temp Yes Yes Yes YesLow air temp No Yes Yes YesHigh skin temp No Yes Yes YesLow skin temp No Yes Yes YesFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Low battery Yes Yes Yes

BACKUP THERMOSTAT Not specified No No No

HEATER POWERINDICATOR LED LED LED LED

HAND PORTS 4 6 6 5

TUBING PORTS 1 7 7 3

O2 INLET PORTS/CONTROLLERS 1/not specified Yes/fixed at 40% Yes/fixed at 40% Yes/fixed at 40%

SUPPLEMENTALHUMIDITY No Yes Yes Yes

Adjustable NA 40-70%, ±10% 40-70%, ±10% 40-70%, ±10%

PHOTOTHERAPY Optional Optional Futura Optional Futura Futura

Colons separate data on similar models of a device. This is the first of* Specifications current as of March 1998. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

18 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 19: incubadoras neonatales

Product Comparison Chart

MODEL BERTOCCHI BIO MS BIO MS BIO MSFAILED TO RESPOND *GB 58 RVO NATISSE AR 500 NATISSE AR 500 NATISSE AR 500

Intensive Care Intermediate Care Primary Care

INTERNAL NOISE, dB <60 <45 <45 <45

LINE POWER, VAC 220 230 230 230

BATTERY, TYPE 12 V NA NA NA

DOUBLE WALL Optional No (optional thermal No (optional thermal No (optional thermalshield) shield) shield)

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 24 (9.4) 40 (15.7) 40 (15.7) 40 (15.7)

SIZEIncubator w/stand,

H x W x D, cm (in) 63.5 x 96.5 x 152.4 140 x 105 x 57 140 x 105 x 57 140 x 105 x 57(25 x 38 x 60) (55 x 41 x 22.4) (55 x 41 x 22.4) (55 x 41 x 22.4)

Mattress, W x D,cm (in) 24.5 x 56.5 60 x 30 60 x 30 60 x 30

(9.6 x 22.2) (23.6 x 11.8) (23.6 x 11.8) (23.6 x 11.8)Hood access door,

H x W, cm (in) 20 x 54.5 16 x 16 16 x 16 16 x 16(8 x 21.5) (6.3 x 6.3) (6.3 x 6.3) (6.3 x 6.3)

Caster, cm (in) 10 (4) 12.5 (4.9) 12.5 (4.9) 12.5 (4.9)

WEIGHT, kg (lb) 56 (123.5) 87 (191.8) 84 (185.2) 82 (180.8)

PURCHASE INFORMATIONWarranty 1 year 15 months 15 months 15 months

Delivery time, ARO 6 weeks 3 weeks 3 weeks 3 weeksList price Not specified Not specified Not specified Not specified

Year first sold 1990 1996 1996 1996Number sold Not specified Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified

OTHER SPECIFICATIONS Ventilator is Optional variable Optional Optional storageincorporated into height, accessories. accessories. (right and left),the incubator. IV pole.Meets requirementsof IEC.

Colons separate data on similar models of a device.* Specifications current as of March 1998.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 19

Page 20: incubadoras neonatales

Product Comparison Chart

MODEL DRAEGER FANEM GIGANTE RECEM- GIGANTE RECEM-NASCIDO NASCIDO

Caleo C-186/TS Infant ALG LUXIncubator

WHERE MARKETED Worldwide Africa, Asia, Cen- Not specified Not specifiedtral America, Eur-ope, North and SouthAmerica, Oceania

FDA CLEARANCE Yes Not specified Not specified Not specified

CE MARK Yes Not specified Not specified Not specified

MOBILE OR TRANSPORT Mobile Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-39 (0.1) 25-37/39 (0.1) 28-37 (0.2) 28-37 (0.1)

Skin (increments) 34-38 (0.1) 34-37/38 (0.1) 35-37 (0.2) NA

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 13-42 21-45 28-39 28-37

Skin, display type Not specified Digital Digital NARange, °C 13-43 21-45 28-37 NA

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes NoHigh skin temp Yes Yes Yes NALow skin temp Yes Yes Yes NAFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes No

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR No Horizontal LED bar Vertical bar Vertical bar

HAND PORTS 6 5 oval and 1 iris 6 4entry port

TUBING PORTS 10 4 2 respiratory, 1 respiratory,2 accessory 2 accessory

O2 INLET PORTS/CONTROLLERS Servo-controlled/ 1/not specified Yes/not specified 1/not specified

optional

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable 30-99%, increments Yes Yes Noof 1%

PHOTOTHERAPY Optional PT4000 Optional Optional Optional

Colons separate data on similar models of a device. This is the first oftwo pages coveringthe above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

20 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 21: incubadoras neonatales

Product Comparison Chart

MODEL DRAEGER FANEM GIGANTE RECEM- GIGANTE RECEM-NASCIDO NASCIDO

Caleo C-186/TS Infant ALG LUXIncubator

INTERNAL NOISE, dB Not specified 50-55 40-50 40-50

LINE POWER, VAC 110/127/230/240 110 or 220 +10% 110-120, 220-240 110/120/220/240

BATTERY, TYPE Lithium-ion For alarm NA NA

DOUBLE WALL Optional, removable Yes Optional Optionalinner walls

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 38 (15) 38 (15) 36 (14.2) 36 (14.2)

SIZEIncubator w/stand,

H x W x D, cm (in) [122-152] x 109 x 68 142.5 x 105 x 57.5 136 x 59 x 100 136 x 59 x 100([48-59.8] x 42.9 x (56.1 x 41.3 x 22.6) (53.5 x 23.2 x 39.4) (53.5 x 23.2 x 39.4)26.8)

Mattress, W x D,cm (in) 64.5 x 50 63 x 34 36 x 80 36 x 80

(25.4 x 19.7) (24.8 x 13.4) (14.2 x 31.5) (14.2 x 31.5)Hood access door,

H x W, cm (in) 22 x 64 25 x 84 15 x 15 15 x 15(8.7 x 25.2) (9.8 x 33.1) (6 x 6) (6 x 6)

Caster, cm (in) 10 (3.9) 10.2 (4) Not specified Not specified

WEIGHT, kg (lb) 137 (302.1) 65 (143.3) 69 (152.1) 67 (147.7)

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO Not specified Not specified 2 weeks Not specifiedList price Not specified $4,800-8,500 $5,500 $2,847

Year first sold 2000 1992 (September) 1997 Not specifiedNumber sold Not specified Several thousand 83 Not specifiedFiscal year Not specified Not specified Not specified Not specified

OTHER SPECIFICATIONS EL display; trend Access door; Optional double Optional head accessanalysis; integrated electronic scale; hood, phototherapy door forx-ray tray; weighing IV pole; guard rail; unit, body stand, resuscitation.scale; thermo- emergency oxygen monitor stand,monitoring; system; humidifier; sensor adhesive box,adjustable height; oxygen hood; 4 microfilter box, andKangaroo Mode; (3-wire) outlets; oxygen tent.1-year calibration- optional shelves,free O2 sensor; cabinet standself-cleaning mode; (3 kinds), andTwincubator for co- phototherapy unit.bedding; electronic Meets requirementsbed tilt; all-round of IEC and ISO 9001.access inemergencies.

Colons separate data on similar models of a device.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 21

Page 22: incubadoras neonatales

Product Comparison Chart

MODEL GIGANTE RECEM- GIGANTE RECEM- GINEVRI GINEVRINASCIDO NASCIDOSMJ TI “BABY SHUTTLE” OGB POLY CARE 1

TRANSPORT INCUBATOR Primary Care

WHERE MARKETED Not specified Not specified Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Not specified Not specified Yes Yes

MOBILE OR TRANSPORT Mobile Mobile Transport Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 28-37 (0.1) 28-37 (0.1) 20-38 (0.1) 20-38 (0.1)

Skin (increments) 28-37 (0.1) NA 25-38 (0.1), NAoptional

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 28-39 28-37 10-45 10-45

Skin, display type Digital NA Digital NARange, °C 28-37 NA 10-45 NA

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes No Yes YesHigh skin temp No NA Optional NALow skin temp No NA Optional NAFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure No No Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Vertical bar Vertical bar Vertical LED bar, Vertical LED bar,

20% increments 20% increments

HAND PORTS 6 6 4, includes main and 8, includes 2 open-lateral doors ing main doors *

TUBING PORTS 2 respiratory, 2 respiratory, 2 respiratory, 2 respiratory,2 accessory 2 accessory 2 accessory 7 accessory

O2 INLET PORTS/CONTROLLERS 1/not specified 1/not specified 2/digital control 2/maximum 31% and

system 52%, optionaldigital controlsystem

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable No No Yes 40-85%

PHOTOTHERAPY Optional Optional Optional Optional

Colons separate data on similar models of a device. This is the first of* Front and rear. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

22 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 23: incubadoras neonatales

Product Comparison Chart

MODEL GIGANTE RECEM- GIGANTE RECEM- GINEVRI GINEVRINASCIDO NASCIDOSMJ TI “BABY SHUTTLE” OGB POLY CARE 1

TRANSPORT INCUBATOR Primary Care

INTERNAL NOISE, dB 40-50 40-50 <53 <45

LINE POWER, VAC 110/120/220/240 110/120/220/240 230; 110-240 opt 230; 110-240 opt

BATTERY, TYPE NA NA 12 V, 40 A-h 4.8 V backup forpower failure

DOUBLE WALL Optional Optional Optional Optional

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 36 (14.2) 36 (14.2) 26 (10.2) 45 (17.7)

SIZEIncubator w/stand,

H x W x D, cm (in) 136 x 59 x 100 136 x 59 x 100 60 x 118 x 105 127 x 80 x 60(53.5 x 23.2 x 39.4) (53.5 x 23.2 x 39.4) (23.6 x 46.5 x 41.3) (50 x 31.5 x 23.6)

Mattress, W x D,cm (in) 36 x 80 36 x 80 55 x 25 60 x 35

(14.2 x 31.5) (14.2 x 31.5) (21.7 x 9.8) (23.6 x 13.8)Hood access door,

H x W, cm (in) 15 x 15 15 x 15 30 x 19 37 x 65(6 x 6) (6 x 6) (11.8 x 7.5) (14.6 x 25.6)

Caster, cm (in) Not specified Not specified 12.5 (4.9) 10 (3.9)

WEIGHT, kg (lb) 67 (147.7) 67 (147.7) 40 (88.2); 32 (70.6); 37-5160 (132.3) with (81.6-112.5) withtrolley trolley

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO Not specified Not specified 4-6 weeks 4-6 weeksList price $4,350 $3,278 Not specified Not specified

Year first sold Not specified Not specified 1997 1995Number sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified January to December January to December

OTHER SPECIFICATIONS Air curtain with Air curtain with Upgradable skin tem- Hood, base, and maindoor. door. perature and O2 mod- components are in-

ules; infant venti- jection-molded poly-lator module with carbonate; tilt bedLCD for CPAP-IMV- positioning; noSIMV and heated hu- glued parts,midifier; O2 cylin- rough edges;der; compressor; easily assembled andVenturi suction disassembled. Meetsunit; resuscitator; requirements ofhead immobilizer; IV CEI 62.22 andpole; shelf; infu- IEC 601-2-19. *sion pump; hood;optional doublewall. Meets require-ments of CEI 62.41,IEC 601-2-20, andRAI n.94-6892/TRM. *

Colons separate data on similar models of a device.* Also meets the requirements of CEI 62.5; EN 55011, 61000-4-2, -4-3, -4-4, -4-5, -4-11; and IEC 601-1, -2.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 23

Page 24: incubadoras neonatales

Product Comparison Chart

MODEL GINEVRI GINEVRI GINEVRI GUIDO RAYOS

OGB POLY CARE 2 OGB POLY CARE 3 OGB POLY CARE 4 NESTORET 5050Intermediate Care Intensive Care Super Intensive Care

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Yes Yes Yes Not specified

MOBILE OR TRANSPORT Mobile Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-38 (0.1) 20-38 (0.1) 20-38 (0.1) 20-37 (0.1),override to 39

Skin (increments) 25-38 (0.1) 25-38 (0.1) 25-38 (0.1) 33-37 (0.1),override to 38

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 10-45 10-45 10-45 10-50

Skin, display type Digital Digital Digital DigitalRange, °C 10-45 10-45 10-45 10-50

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes YesHigh skin temp Yes Yes Yes YesLow skin temp Yes Yes Yes YesFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes 2

HEATER POWERINDICATOR Vertical LED bar, Vertical LED bar, Vertical LED bar, Horizontal LED bar,

20% increments 20% increments 20% increments 4 levels

HAND PORTS 8, includes 2 open- 8, includes 2 open- 8, includes 2 open- 6ing main doors * ing main doors * ing main doors *

TUBING PORTS 2 respiratory, 2 respiratory, 2 respiratory, 2 respiratory,7 accessory 7 accessory 7 accessory 2 accessory

O2 INLET PORTS/CONTROLLERS 2/maximum 31% and 2/maximum 31% and 3/maximum 31%, 52%, 2; 1 limited O2

52%, optional 52% with digital and a digital concentration, 1 updigital control control system servo-controlled to 100% O2 concen-system system 21-100% tration/integrated

O2 monitoring withalarms; optional O2servo-regulator

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes (monitored)

Adjustable 40-85% 40-85% with digital 40-90% with servo- Yescontrol system control system

PHOTOTHERAPY Optional Optional Optional Optional

Colons separate data on similar models of a device. This is the first of* Front and rear. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

24 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 25: incubadoras neonatales

Product Comparison Chart

MODEL GINEVRI GINEVRI GINEVRI GUIDO RAYOS

OGB POLY CARE 2 OGB POLY CARE 3 OGB POLY CARE 4 NESTORET 5050Intermediate Care Intensive Care Super Intensive Care

INTERNAL NOISE, dB <45 <45 <45 <60

LINE POWER, VAC 230; 110-240 opt 230; 110-240 opt 230; 110-240 opt Any upon request

BATTERY, TYPE 4.8 V backup for 4.8 V backup for 4.8 V backup for See footnote *power failure power failure power failure

DOUBLE WALL Optional Optional Optional Optional

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 45 (17.7) 45 (17.7) 45 (17.7) 35.5 (14)

SIZEIncubator w/stand,

H x W x D, cm (in) 127 x 80 x 60 127 x 80 x 60 127 x 80 x 60 131 x 90 x 59(50 x 31.5 x 23.6) (50 x 31.5 x 23.6) (50 x 31.5 x 23.6) (51.6 x 35.4 x 23.2)

Mattress, W x D,cm (in) 60 x 35 60 x 35 60 x 35 66 x 35

(23.6 x 13.8) (23.6 x 13.8) (23.6 x 13.8) (26 x 13.8)Hood access door,

H x W, cm (in) 37 x 65 37 x 65 37 x 65 25.5 x 78.5(14.6 x 25.6) (14.6 x 25.6) (14.6 x 25.6) (10 x 30.9)

Caster, cm (in) 10 (3.9) 10 (3.9) 10 (3.9) Not specified

WEIGHT, kg (lb) 32 (70.6); 37-51 32 (70.6); 37-51 32 (70.6); 37-51 77 (169.8)(81.6-112.5) with (81.6-112.5) with (81.6-112.5) withtrolley trolley trolley

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO 4-6 weeks 4-6 weeks 4-6 weeks 30-45 daysList price Not specified Not specified Not specified Not specified

Year first sold 1995 1996 1996 Not specifiedNumber sold Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

OTHER SPECIFICATIONS Hood, base, and main Hood, base, and main Hood, base, and main IV pole; large LCDcomponents are in- components are in- components are in- display with twojection-molded poly- jection-molded poly- jection molded poly- 20-character rowscarbonate; tilt bed carbonate; tilt bed carbonate; tilt bed for operation para-positioning; no positioning; no positioning; meters and instruc-glued parts, glued parts, “sterile RH” servo tions; rails;rough edges; rough edges; steam; servo O2; no shelves; variableeasily assembled and “Hot Spot” infrared glued parts or height trolley;disassembled; “Hot hood heating system rough edges; scale with digitalSpot" infrared hood with VLBW infants or “Hot Spot” infrared display; 2 sensorsheating system with when doors are open. hood heating system for O2; heater errorVLBW infants or when Meets requirements w/ VLBW infants or alarm; CPU errordoors are opened. of CEI 62.22 and when doors are open. alarm; manual testMeets requirements IEC 601-2-19. ** Meets requirements for all alarms;of CEI 62.22 and of CEI 62.22 and optional pulseIEC 601-2-19. ** IEC 601-2-19. ** oximetry. Meets

IEC standards.

Colons separate data on similar models of a device.* Uninterruptible power supply upon request.** Also meets the requirements of CEI 62.5; EN 55011, 61000-4-2, -4-3, -4-4, -4-5, -4-11; and IEC 601-1, -2.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 25

Page 26: incubadoras neonatales

Product Comparison Chart

MODEL GUIDO RAYOS HERO MEDICAL MAGNA MAGNAFAILED TO RESPOND *

NESTOTRANS DMH 2-Stationary ICT-2000 ICT-2001

WHERE MARKETED Worldwide Asia, Europe Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Yes Yes

CE MARK Not specified Not specified Not specified Not specified

MOBILE OR TRANSPORT Transport Not specified Mobile Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-37 (0.1), 21-38 (0.1) 20-39 (0.1) 34-39 (0.1)override to 39

Skin (increments) 33-37 (0.1), 21-38 (0.1) 20-39 (0.1) 34-39 (0.1)override to 38

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 10-50 21-38 20-42 20-42

Skin, display type Digital Digital Digital DigitalRange, °C 10-50 21-38 20-42 20-42

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes YesHigh skin temp Yes Yes Yes YesLow skin temp Yes Yes Yes YesFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes No No

HEATER POWERINDICATOR Horizontal LED bar, LED Horizontal lamp bar, Horizontal lamp bar,

4 levels 25% increments 25% increments

HAND PORTS 3 5 4 4

TUBING PORTS 1 respiratory, To customer 2 respiratory, 2 respiratory,2 accessory specifications 2 accessory 2 accessory

O2 INLET PORTS/CONTROLLERS 2; 1 limited O2 Yes/yes 1 standard/control 1 standard/control

concentration, 1 up system systemto 100% O2 concen-tration/integratedO2 monitoring withalarms

SUPPLEMENTALHUMIDITY Yes Not specified Yes Yes

Adjustable No Not specified Yes Yes

PHOTOTHERAPY Not specified Not specified Optional Optional

Colons separate data on similar models of a device. This is the first of* Specifications current as of March 1998. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

26 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 27: incubadoras neonatales

Product Comparison Chart

MODEL GUIDO RAYOS HERO MEDICAL MAGNA MAGNAFAILED TO RESPOND *

NESTOTRANS DMH 2-Stationary ICT-2000 ICT-2001

INTERNAL NOISE, dB <60 55 ≤65 ≤50

LINE POWER, VAC Any upon request 110/240 110/220 110/220

BATTERY, TYPE 12 V 4 AA 12 V 12 V

DOUBLE WALL Optional Not specified Optional Standard

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 28.5 (11.2) 38 (15) 37 (14.6) 26.5 (10.4)

SIZEIncubator w/stand,

H x W x D, cm (in) 191 x 57 x 115 137 x 93 x 52 See footnote ** 99 x 61 x 180(75.2 x 22.4 x 45.3) (54 x 36.6 x 20.5) (39 x 24 x 70.9)

Mattress, W x D,cm (in) 53 x 26.5 69.5 x 37.5 64 x 35.5 64 x 35.5

(20.9 x 10.4) (27.4 x 14.8) (25.2 x 14) (25.2 x 14)Hood access door,

H x W, cm (in) Full hood opening 34 x 70 30 x 94 (11.8 x 37) 26 x 94 (10.2 x 37)(13.4 x 27.6)

Caster, cm (in) Not specified Not specified 10 (3.9) 10 (3.9)

WEIGHT, kg (lb) 28 (11) 73 (161) 78 (172) 115 (253.6)

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO 30-45 days Not specified Not specified Not specifiedList price Not specified $4,050-6,000 $7,017 $12,849

Year first sold Not specified 1992 Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified

OTHER SPECIFICATIONS IV pole; large LCD Optional shelves, None specified. None specified.display with two cabinet, IV pole,20-character rows integral oxygenfor operation para- controller, doublemeters and instruc- microprocessor, andtions; built-in multi-incubatoroxygen cylinder; monitoring.2 sensors for O2;heater error alarm;CPU error alarm;manual test for allalarms; optionalobservation lamp.Meets IEC standards.

Colons separate data on similar models of a device.* Specifications current as of March 1998.** 139 × [91, 106.5] × [51.5, 61.3] (54.7 × [35.8, 41.9] × [20.3, 24.1]).

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 27

Page 28: incubadoras neonatales

Product Comparison Chart

MODEL MEDAP MEDICOR MEDICOR MEDICORFAILED TO RESPOND *T1500 BLF-2001 ICB-145 ICB-145i

WHERE MARKETED Not specified Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Not specified Yes Yes Yes

MOBILE OR TRANSPORT Mobile Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 32-40 (2) 27-37 (not specifi- 27-37 (not speci- 27-37 (not specifi-ed), override to 39 fied) ed), override to 39

Skin (increments) NA 34-36 (not specifi- NA 34-36 (not specifi-ed), override to 38 ed), override to 38

TEMPERATURE DISPLAYAir, display type Fluid thermometer Digital Digital Digital

Range, °C Not specified 40-50 40-50 40-50

Skin, display type NA Digital NA DigitalRange, °C NA 40-50 NA 40-50

ALARMSHigh air temp Yes Yes Yes YesLow air temp No Yes No NoHigh skin temp NA Yes NA NoLow skin temp NA Yes NA YesFan failure NA Yes Yes YesSensor failure NA Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT No Yes Yes Yes

HEATER POWERINDICATOR Yes LED LED LED

HAND PORTS 2 6 6 6

TUBING PORTS 2 8 6 8

O2 INLET PORTS/CONTROLLERS Flowmeter O2 Yes Yes Yes

cylinder; space forspare cylinder

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable Not specified No No No

PHOTOTHERAPY Optional Optional Optional Optional

Colons separate data on similar models of a device. This is the first of* Specifications current as of March 1998. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

28 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 29: incubadoras neonatales

Product Comparison Chart

MODEL MEDAP MEDICOR MEDICOR MEDICORFAILED TO RESPOND *T1500 BLF-2001 ICB-145 ICB-145i

INTERNAL NOISE, dB Not specified ≤50 ≤60 ≤60

LINE POWER, VAC 220/240 230 ** 230 ** 230 **

BATTERY, TYPE 12-24 V 12 V 9 V 12 V

DOUBLE WALL Yes Yes No Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 30 (11.8) 31.5 (12.4) 35.6 (14) 31.5 (12.4)

SIZEIncubator w/stand,

H x W x D, cm (in) 90.5 x 80 x 50.5 130 x 85 x 66.5 132 x 105 x 51 132 x 105 x 51(35.6 x 31.5 x 19.9) (51.2 x 33.5 x 26.2) (52 x 41.3 x 20.1) (52 x 41.3 x 20.1)

Mattress, W x D,cm (in) 60 x 30 58 x 33 58 x 33 58 x 33

(23.6 x 11.8) (22.8 x 13) (22.8 x 13) (22.8 x 13)Hood access door,

H x W, cm (in) Not specified 36 x 72 38 x 70 36 x 72(14.2 x 28.3) (15 x 27.6) (14.2 x 28.3)

Caster, cm (in) 12 (4.7) 9.5 (3.7) 9.5 (3.7) 9.5 (3.7)

WEIGHT, kg (lb) 25 (55.1), incubator 63 (138.9) 60 (132.3) 63 (138.9)only

PURCHASE INFORMATIONWarranty Available 1 year 1 year 1 year

Delivery time, ARO 6-8 weeks 45 days 45 days 45 daysList price $7,000-7,500 $6,990 $3,980 $6,250

Year first sold Not specified Not specified Not specified Not specifiedNumber sold ~1,000 Not specified Not specified Not specifiedFiscal year Not specified January to December January to December January to December

OTHER SPECIFICATIONS Flat heater is Built-in scale; Built-in scale; up Built-in scale;vulcanized on lower built-in x-ray bank; to 90% humidity; built-in x-ray bank;side of patient- air curtain; up to optional head hood, up to 90% humidity;supporting tub; no 90% humidity; storage cabinet, optional head hood,fan. optional head hood, infusion stand, storage cabinet,

servo-controlled apnea alarm. Meets infusion stand,O2 unit, storage requirements of apnea alarm. Meetscabinet, infusion ISO 9001. requirements of ISOstand, apnea alarm. 9001.Meets requirementsof ISO 9001.

Colons separate data on similar models of a device.* Specifications current as of March 1998.** Other line power configurations available upon request.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 29

Page 30: incubadoras neonatales

Product Comparison Chart

MODEL MEDIPREMA MEDIPREMA MEDIPREMA MEDIPREMA

MP5 ISIS 3552 MP5 ISIS 3553 MP5 ISIS 3555 NITE

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Not specified Not specified Not specified

CE MARK Yes Yes Yes Yes

MOBILE OR TRANSPORT Mobile Mobile Mobile Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 28-37 (0.1), 28-37 (0.1), 28-37 (0.1), 28-37 (0.1),override to 39 override to 39 override to 39 override to 39

Skin (increments) 35-37 (0.1), 35-37 (0.1), 35-37 (0.1), 35-37 (0.1),override to 38 override to 38 override to 38 override to 38

TEMPERATURE DISPLAYAir, display type Digital Digital Digital Digital

Range, °C 5-45 5-45 5-45 5-45

Skin, display type Digital Digital Digital DigitalRange, °C 5-45 5-45 5-45 5-45

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes YesHigh skin temp Yes Yes Yes YesLow skin temp Yes Yes Yes YesFan failure Yes Yes Yes NASensor failure Yes Yes Yes YesPower failure Yes (and system Yes (and system Yes (and system Yes (and low battery

failure) failure) failure) and system failure)

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR LED or micro- LED or micro- LED or micro- Horizontal bar or

processor processor processor microprocessor

HAND PORTS 4 5 6 6

TUBING PORTS 4 9 13 4 respiratory,6 accessory

O2 INLET PORTS/CONTROLLERS Yes/not specified Yes/not specified Yes/optional Oxygen analyzer with

audible and visualalarms/not specified

SUPPLEMENTALHUMIDITY Yes Yes Yes No

Adjustable 40/60/80% 40/60/80% 40 to 90% optional NA

PHOTOTHERAPY Optional Optional Optional No

Colons separate data on similar models of a device. This is the first oftwo pages coveringthe above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

30 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 31: incubadoras neonatales

Product Comparison Chart

MODEL MEDIPREMA MEDIPREMA MEDIPREMA MEDIPREMA

MP5 ISIS 3552 MP5 ISIS 3553 MP5 ISIS 3555 NITE

INTERNAL NOISE, dB <45 <45 <45 Noiseless (no fan)

LINE POWER, VAC 220/240; 110 opt 220/240; 110 opt 220/240; 110 opt 100-260;50/60/400 Hz

BATTERY, TYPE NA NA NA 12-38 V

DOUBLE WALL Optional Optional Optional No

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 35 (13.8) 35 (13.8) 35 (13.8) 29 (11.4)

SIZEIncubator w/stand,

H x W x D, cm (in) 130 x 104 x 58 134 x 104 x 60 134 x 104 x 60 54 x 96 x 58(51 x 41 x 22.8) (53 x 41 x 24) (53 x 41 x 24) (21.3 x 37.8 x 22.8)

Mattress, W x D,cm (in) 61 x 38.5 61 x 38.5 61 x 38.5 59 x 35

(24 x 15.2) (24 x 15.2) (24 x 15.2) (23.2 x 13.8)Hood access door,

H x W, cm (in) 67 x 32 67 x 32 67 x 32 67 x 24(26.4 x 12.6) (26.4 x 12.6) (26.4 x 12.6) (26.4 x 9.4)

Caster, cm (in) 12.5 (5) 12.5 (5) 12.5 (5) NA

WEIGHT, kg (lb) 80 (176.4) 85 (187) 87 (192) 36.5 (80)

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO 8 weeks 8 weeks 8 weeks 8 weeksList price Not specified Not specified Not specified Not specified

Year first sold 1997 1997 1997 1995Number sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified

OTHER SPECIFICATIONS Convection heated; Convection heated; Convection heated; Same as ISIS models,temp adjust control; temp adjust control; temp adjust control; plus radiationelectronic electronic electronic heating integratedsurveillance system; surveillance system; surveillance system; in the incubatoralarm/communications alarm/communications alarm/communications compartment backeddisplay on 68000 display on 68000 display on 68000 by convection andSeries micro- Series micro- Series micro- conduction system;processor; permanent processor; permanent processor; permanent touchscreen withauto verification of auto verification of auto verification of temperatureprobes and control probes and control probes and control monitoring.devices; micro- devices; micro- devices; micro-processor control; processor control; processor control;RS232 and RS485 RS232 and RS485 RS232 and RS485serial links; 1-week serial links; 1-week serial links; 1-weekmemory of defaults/ memory of defaults/ memory of defaults/temp; x-ray facili- temp; x-ray facili- temp; x-ray facili-ties. ties; optional ties; optional

scale and height. scale and height,autoclavable watertank; servo humid.

Colons separate data on similar models of a device.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 31

Page 32: incubadoras neonatales

Product Comparison Chart

MODEL MEDIX MEDIX MEDIX MEDIX

PC-301 : PC-305 TR-1 TR-200 TR-306

WHERE MARKETED Africa, Latin Africa, Latin Africa, Latin Africa, LatinAmerica, Middle America, Middle America, Middle America, MiddleEast, Western East, Western East, Western East, WesternEurope Europe Europe Europe

FDA CLEARANCE Yes Not specified Not specified Not specified

CE MARK Not specified Not specified Not specified Not specified

MOBILE OR TRANSPORT Mobile Transport Transport Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-39 (0.1) 20-38 (0.25) 20-39 (0.1) 20-39 (0.1)

Skin (increments) NA : 34-38 (0.1) NA 34-38 (0.1) 34-38 (0.1)

TEMPERATURE DISPLAYAir, display type Digital Thermometer Digital Digital

Range, °C 20-42 20-42 20-42 20-42

Skin, display type NA : Digital NA No DigitalRange, °C NA : 20-42 NA NA 20-42

ALARMSHigh air temp Yes NA Yes YesLow air temp Yes NA Yes YesHigh skin temp NA : Yes NA No YesLow skin temp NA : Yes NA No YesFan failure Yes NA Yes YesSensor failure Yes NA Yes YesPower failure Yes NA Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Horizontal LED bar, Vertical LED bar, LED bar, 25% Horizontal LED bar,

25% increments 25% increments increments 25% increments

HAND PORTS 6 2 2 6

TUBING PORTS 4 2 2 4

O2 INLET PORTS/CONTROLLERS 1; 2 positions 1; 2 positions 1; 2 positions 1; 2 positions

(maximum 40% and (maximum 40% and (maximum 40% and (maximum 40% andmaximum 75%)/not maximum 75%)/not maximum 75%)/not maximum 75%)/notspecified specified specified specified

SUPPLEMENTALHUMIDITY Yes Yes Yes Yes

Adjustable Optional No Not specified Optional

PHOTOTHERAPY Optional No No No

Colons separate data on similar models of a device. This is the first oftwo pages coveringthe above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

32 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 33: incubadoras neonatales

Product Comparison Chart

MODEL MEDIX MEDIX MEDIX MEDIX

PC-301 : PC-305 TR-1 TR-200 TR-306

INTERNAL NOISE, dB <60 <60 <60 <60

LINE POWER, VAC 110/220 110/220 110/220 110/220

BATTERY, TYPE NA 12 V 12 V 12 V

DOUBLE WALL Optional Yes Yes Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 34.9 (13.75) 25 (9.8) 30 (11.8) 25 (9.8)

SIZEIncubator w/stand,

H x W x D, cm (in) 132 x 94 x 53.3 110 x 54 x 130 115 x 54 x 130 132 x 94 x 53.3(52 x 37 x 21) (43.3 x 21.3 x 51.2) (45.3 x 21.3 x 51.2) (52 x 37 x 21)

Mattress, W x D,cm (in) 63.5 x 35.1 56 x 25 56 x 25 63.5 x 35.1

(25 x 13.8) (22 x 9.8) (22 x 9.8) (25 x 13.8)Hood access door,

H x W, cm (in) 26.7 x 83.8 23 x 64 30 x 70 26.7 x 83.8(10.5 x 33) (9 x 25.2) (11.8 x 27.6) (10.5 x 33)

Caster, cm (in) 10 (4) Not specified 10.2 (4) 10.2 (4)

WEIGHT, kg (lb) 75.3 (166) 19-70 (41.9-154.4) 21-72 (46.3-158.8) 40-59 (90-130)

PURCHASE INFORMATIONWarranty 1 year 1 year 1 year 1 year

Delivery time, ARO 4 weeks 6 weeks 6 weeks 8 weeksList price $6,200 : $6,500 $3,300-6,400 Not specified Not specified

Year first sold 1988 1978 2000 1997Number sold 450 : 1,300 1,700 120 150Fiscal year November to October November to October November to October November to October

OTHER SPECIFICATIONS Meets requirements Optional folding Optional folding Optional foldingof AAMI/ANSI II36, cart, fixed cart, cart and 2 cylinder/ cart and oxygenIEC 601-2-19, and battery and charger, regulators. Meets cylinder/regulator.ISO 9000. and oxygen cylinder/ requirements of IEC Meets requirements

regulator; ISO 9000. 601-2-20, ISO 9000. of IEC 601-2-20, ISO9000.

Colons separate data on similar models of a device.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 33

Page 34: incubadoras neonatales

Product Comparison Chart

MODEL OHMEDA OHMEDA OHMEDA OHMEDA

Care-Plus Care-Plus Access Giraffe OmniBed Ohio Intensive Care

WHERE MARKETED Worldwide Worldwide Worldwide USA

FDA CLEARANCE Yes Yes Yes Yes

CE MARK Yes Yes Yes No

MOBILE OR TRANSPORT Mobile Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 20-37 (0.1), 20-37 (0.1), 20-37 (0.1), 28-37.5 (0.1),override to 39 override to 39 override to 39 optional 28-39

Skin (increments) 35-37 (0.1), 35-37 (0.1), 35-37.5 (0.1) 35-37.5 (0.1)optional 35-37.5 optional 35-37.5

TEMPERATURE DISPLAYAir, display type Digital Digital Not specified Digital

Range, °C 20-42 20-42 Not specified 15-50

Skin, display type Digital Digital Digital DigitalRange, °C 20-42 20-42 15-50 15-50

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes NoHigh skin temp Yes Yes Yes YesLow skin temp Yes Yes Yes YesFan failure Yes Yes Yes NoSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Vertical lamp bar Vertical lamp bar No * None

HAND PORTS 6 6 4, gumdrop shaped Sliding access hood

TUBING PORTS 4 6 9 2

O2 INLET PORTS/CONTROLLERS Yes/separate O2 Yes/separate O2 Servo-controlled Not specified

control system control system option only;available August2001

SUPPLEMENTALHUMIDITY Yes (variable) Yes (variable) See footnote ** Yes (variable)

Adjustable Optional servo Optional servo 30-95%, increments Not specifiedcontrolled controlled of 5%

PHOTOTHERAPY Optional Optional See footnote *** Optional

Colons separate data on similar models of a device. This is the first of* Trending of set point, skin temp, and air temp. two pages covering** Servo-controlled humidifier, immersed heater type. the above model(s).*** Floor model, rail-mounted spot PT lamps available. These specifications

continue onto thenext page.

Healthcare Product Comparison System

34 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 35: incubadoras neonatales

Product Comparison Chart

MODEL OHMEDA OHMEDA OHMEDA OHMEDA

Care-Plus Care-Plus Access Giraffe OmniBed Ohio Intensive Care

INTERNAL NOISE, dB <60 <60 ≤40 dBA <60

LINE POWER, VAC 100/120/220/240 100/120/220/240 115, 50/60 Hz * 100/120/220/240

BATTERY, TYPE NA NA 9 V Ni-MH NA

DOUBLE WALL Optional Optional Both walls Yes

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 38.1 (15) 43.2 (17) 40 (15.7) 34.3 (13.5)

SIZEIncubator w/stand,

H x W x D, cm (in) 135.9 x 88.9 x 64.8 141 x 88.9 x 64.8 [147-174] x 69 x 112 142.2 x 88.9 x 71.1(53.5 x 35 x 25.5) (55.5 x 35 x 25.5) ([57.9-68.5] x 27.2 (56 x 35 x 28)

x 44.1)Mattress, W x D,

cm (in) 34.3 x 64.8 34.3 x 64.8 48 x 66 38.1 x 62.2(13.5 x 25.5) (13.5 x 25.5) (18.9 x 26) (15 x 24.5)

Hood access door,H x W, cm (in) 25.4 x 76.2 30.5 x 76.2 27 x 78 33 x 76.2

(10 x 30) (12 x 30) (10.6 x 30.7) (13 x 30)Caster, cm (in) 12.7 (5) 12.7 (5) 12.7 (5) 12.7 (5)

WEIGHT, kg (lb) 83.9 (185) 83.9 (185) 129 (284.4) 99.8 (220)

PURCHASE INFORMATIONWarranty 3 years, parts; 3 years, parts; 1 year 1 year

1 year, labor 1 year, laborDelivery time, ARO Not specified Not specified 90 days Not specifiedList price $8,475-13,150 $8,475-13,150 $34,000, varies with $10,125

optionsYear first sold 1988 1996 2000 1981Number sold Not specified Not specified 300 Not specifiedFiscal year Not specified Not specified Not specified Not specified

OTHER SPECIFICATIONS Optional rail and Optional rail and Converts from Optional rail,shelf; optional shelf; optional incubator to radiant shelf, radiantBiliBlanket photo- BiliBlanket photo- warmer; elevating warmer, andtherapy light; full therapy light; full hood; closed bed BiliBlanket photo-line of accessories; line of accessories; 10 cm above center therapy light.continuous tilt continuous tilt of mattress.system on 4000 system on 4000version of version ofCare-Plus. Care-Plus Access.

Colons separate data on similar models of a device.* Also 220/230/240 VAC, 50/60 Hz.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 35

Page 36: incubadoras neonatales

Product Comparison Chart

MODEL PARKER PHOENIX MEDICAL SEWARD TSEFAILED TO RESPOND *TC500-1 : TC500-2 : 100 A INCUBATOR 4000 INKA SI-302TC500-3

WHERE MARKETED Worldwide Worldwide Worldwide Africa, Europe,Middle East

FDA CLEARANCE No No Not specified Not specified

CE MARK No No Yes Yes

MOBILE OR TRANSPORT Transport Mobile Mobile Mobile

TEMPERATURE CONTROLRANGE, °C

Air (increments) 30-37 (0.1) 32-38 (0.1) 25-39 (0.1) 30-37.9 (0.1)

Skin (increments) None 32-38 (0.1) 32-37 (0.1) 34-37.9 (0.1)

TEMPERATURE DISPLAYAir, display type Digital Not specified Backlit graphic LCD Digital

Range, °C 15-40 15-45 25-39 0-99.9

Skin, display type NA Not specified Backlit graphic LCD DigitalRange, °C NA 15-40 32-38 0-99.9

ALARMSHigh air temp Yes Yes Yes YesLow air temp Yes Yes Yes YesHigh skin temp No Yes Yes YesLow skin temp No Yes Yes YesFan failure Yes Yes Yes YesSensor failure Yes Yes Yes YesPower failure Yes Yes Yes Yes

BACKUP THERMOSTAT Yes Yes Yes Yes

HEATER POWERINDICATOR Single LED Vertical lamp bar Not specified On/off

10% increments

HAND PORTS 2 or 4 4 or 6 5 6

TUBING PORTS 2 4 respiratory 4 2 respiratory,2 accessory

O2 INLET PORTS/CONTROLLERS 1/not specified 1 oxygen port with 1 standard/monitor 1/Oxycare

flowmeter/not optional Control Systemspecified

SUPPLEMENTALHUMIDITY Yes Yes Monitoring only Yes

Adjustable Not specified No No Yes

PHOTOTHERAPY Optional Optional No Optional

Colons separate data on similar models of a device. This is the first of* Specifications current as of October 1999. two pages covering

the above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

36 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 37: incubadoras neonatales

Product Comparison Chart

MODEL PARKER PHOENIX MEDICAL SEWARD TSEFAILED TO RESPOND *TC500-1 : TC500-2 : 100 A INCUBATOR 4000 INKA SI-302TC500-3

INTERNAL NOISE, dB <520 A ≤60 <54 <52

LINE POWER, VAC 115/240 : 12/24 VDC 110/220 230, 50/60 Hz, 875 W 230

BATTERY, TYPE TC515 12 V pack NA NA NA

DOUBLE WALL Optional No Yes Optional

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 23 (9) 31 (12.2) 33 (13) 33.5 (13.2)

SIZEIncubator w/stand,

H x W x D, cm (in) 44 x 43 x 83 135 x 96 x 55 105 x 71 x 125 [134-156] x 56 x 105(17.3 x 17 x 32.6) (53 x 38 x 21) (41.3 x 28 x 49.2) ([52.8-61.4] x 22 x

41.3)Mattress, W x D,

cm (in) 57.5 x 30.5 61 x 30.5 61 x 36 36 x 67(22.6 x 12) (24 x 12) (24 x 14.2) (14.2 x 26.4)

Hood access door,H x W, cm (in) Not specified 30.5 x 81.2 Not specified 36.2 x 84.5

(12 x 32) (14.3 x 33.3)Caster, cm (in) Not specified 10 (4) Not specified Not specified

WEIGHT, kg (lb) 23 (50.7) 85 (187.4) 110 (242.6) 106 (233.7)

PURCHASE INFORMATIONWarranty 1 year Not specified 3 years 1 year

Delivery time, ARO 6 weeks Not specified 12 weeks 6-8 weeksList price $4,000-14,000 $3,000 $14,500 Not specified

Year first sold 1993 Not specified 1997 1996Number sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified October to September January to December

OTHER SPECIFICATIONS TGA L49729 and open Crevice-free None specified. Oxygen analyzer;canopy for control interior and message 20-40°Cof temperature on display alarm. Meets thermometer; oxygenTC500-2 and TC500-3. requirements of flowmeter;

IEC 601-1. bacteriologicalfilters; 2-position,bidirectional bedinclination;x-ray cassette; 4wheels with 2brakes; relativehumidity gauge;optional storagecase. CE mark. Meetsrequirements ofCSN, EN, IEC andISO 9001.

Colons separate data on similar models of a device.* Specifications current as of October 1999.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 37

Page 38: incubadoras neonatales

Product Comparison Chart

MODEL TSE TSE

INKA SI-303 TI-400 Transport

WHERE MARKETED Africa, Europe, Africa, Europe,Middle East Middle East

FDA CLEARANCE Not specified Not specified

CE MARK Yes Yes

MOBILE OR TRANSPORT Mobile Transport

TEMPERATURE CONTROLRANGE, °C

Air (increments) 30-37.9 (0.1) 30-38 (0.1)

Skin (increments) 30-37.9 (0.1) NA

TEMPERATURE DISPLAYAir, display type Graphic display Digital

Range, °C 0-99.9 0-99.9

Skin, display type Graphic display NARange, °C 0-99.9 NA

ALARMSHigh air temp Yes YesLow air temp Yes YesHigh skin temp Yes NALow skin temp Yes NAFan failure Yes YesSensor failure Yes YesPower failure Yes Yes

BACKUP THERMOSTAT Yes Yes

HEATER POWERINDICATOR On/off Vertical lamp bar,

25% increments

HAND PORTS 6 3

TUBING PORTS 2 respiratory, 2 respiratory,2 accessory 2 accessory

O2 INLET PORTS/CONTROLLERS 1/Oxycare 1/Oxycare

Control System Control System

SUPPLEMENTALHUMIDITY Yes No

Adjustable Yes NA

PHOTOTHERAPY Optional Optional

Colons separate data on similar models of a device. This is the first oftwo pages coveringthe above model(s).These specificationscontinue onto thenext page.

Healthcare Product Comparison System

38 ©2001 ECRI. Duplication of this page by any means for any purpose is prohibited.

Page 39: incubadoras neonatales

Product Comparison Chart

MODEL TSE TSE

INKA SI-303 TI-400 Transport

INTERNAL NOISE, dB <52 <60

LINE POWER, VAC 230 230, 380 VA, 195 W

BATTERY, TYPE NA 12 V

DOUBLE WALL Yes Optional

VERTICAL HOOD-TO-MATTRESS DISTANCE,

cm (in) 33.5 (13.2) 23.5 (9.3)

SIZEIncubator w/stand,

H x W x D, cm (in) [134-156] x 56 x 105 136.5 x 56 x 86.5([52.8-61.4] x 22.1 (53.7 x 22.1 xx 41.3) 34.1) w/carriage *

Mattress, W x D,cm (in) 36 x 67 30.9 x 56.3

(14.2 x 26.4) (12.2 x 22.2)Hood access door,

H x W, cm (in) 36.2 x 84.5 29.1 x 74.2(14.3 x 33.3) (11.5 x 29.2)

Caster, cm (in) Not specified Not specified

WEIGHT, kg (lb) 108 (238.1) 58 (128); 78 (172)with carriage

PURCHASE INFORMATIONWarranty 1 year 1 year

Delivery time, ARO 10-12 weeks 6-8 weeksList price Not specified Not specified

Year first sold 1997 1995Number sold Not specified Not specifiedFiscal year January to December January to December

OTHER SPECIFICATIONS Microcomputer Temp regulator;controlled; tempera- 20-40°C thermo-ture and oxygen meter; built-in O2measurement and con- flowmeter; addi-trol; double cover; tional external O2elastic sleeve at and air supplies;entry port; adjust- O2 and air pressureable bed; pull-out bottles; polymetha-board for x-ray crylate cover; frontcassettes; breath side can be lowered;and pulse control includes heat unitand weight measuring fan, electronic con-optional; IV pole; trol circuits, de-shelves. CE mark. tachable internalMeets requirements energy source; re-of CSN, EN, IEC and chargeable. MeetsISO 9001. requirements of CSN,

EN, IEC and ISO9001.

Colons separate data on similar models of a device.* The incubator measures 66 x 56 x 86.5 cm (26 x 22.1 x 34.1") without a carriage.

Incubators, Infant

©2001 ECRI. Duplication of this page by any means for any purpose is prohibited. 39