1-s2.0-S0011227504001031-main
-
Upload
sureshjeeva -
Category
Documents
-
view
213 -
download
0
Transcript of 1-s2.0-S0011227504001031-main
-
7/27/2019 1-s2.0-S0011227504001031-main
1/4
Small punch testing for determining the cryogenic fractureproperties of 304 and 316 austenitic stainless steels
in a high magnetic field
Yasuhide Shindo *, Yoko Yamaguchi, Katsumi Horiguchi
Department of Materials Processing, Graduate School of Engineering, Tohoku University, Sendai 980-8579, Japan
Received 22 December 2003; received in revised form 12 April 2004; accepted 16 April 2004
Abstract
This paper examines the effect of magnetic field on the fracture properties of austenitic stainless steels at liquid helium tem-
perature (4 K). Small punch tests were performed on cold-rolled 304 and 316 austenitic stainless steels. Previously proposed
correlation for small punch and elasticplastic fracture toughness test methods was applied to predict a small punch test-based
fracture toughness from equivalent fracture strain.
2004 Elsevier Ltd. All rights reserved.
Keywords: Metals (A); Structural materials (A); Liquid helium (B); Mechanical properties (C); Superconducting magnets (F)
1. Introduction
Austenitic stainless steels are the primary structural
materials for the superconducting applications, in which
they are subjected to high magnetic fields at liquid he-
lium temperature (4 K). The presence of a strong mag-
netic field enhances the strain-induced martensitic
transformation in some of these materials at low
temperatures. If the structural materials selected for
superconducting applications undergo martensitic
transformation under service conditions, there may be
unanticipated effects, such as changes in the fracture and
deformation properties that can potentially degrade the
performance of the device [1].
Experimental efforts have been made to examine theeffect of magnetic field on the tensile and fracture
properties of metastable austenitic stainless steels at 4 K.
Fultz and Morris [2] studied the plastic deformation of
AISI 304L and AISI 304LN stainless steels in magnetic
fields as large as 18 T (tesla) at temperatures of 4, 77 and
290 K. They found that the effects of high magnetic
fields on the deformation behavior were probably too
small to be of engineering importance in the design of
large superconducting magnets. Fukushima et al. [3]using compact (CT) specimens precracked at 77 K,
suggested that there may be a significant decrease in the
fracture toughness of 304 stainless steel at 4 K in a 9 T
magnetic field. Murase et al. [4] also observed that an 8
T magnetic field decreased the 4-K fracture toughness of
304 CT specimens precracked at 77 K. However, Chan
et al. [5] using CT specimens precracked at room tem-
perature, found that an increase in the fracture tough-
ness of 304 CT specimens tested at 4 K in an 8 T
magnetic field was observed relative to the fracture
toughness of CT specimens tested in 0 T. They con-
cluded that this improvement is expected as a result of
magnetostatic effects and transformation strain differ-ences due to the excess martensite formed within the
magnetic field, and the increase in strain hardening
rates. The direction of fracture toughness change is
influenced both by the stability of the alloys and by the
specimen preparation conditions, such as precracking
temperature. The stability of martensitic transforma-
tions and thus the mechanical properties of 304 are
sensitive to factors such as its C, N, and Ni content [6]
and grain size. The C and Ni content in the alloy used in
Ref. [5] was a slightly lower than that in the alloys used
by Fukushima et al. [3] and by Murase et al. [4]. Chan
* Corresponding author. Tel./fax: +81-22-217-7341.
E-mail address: [email protected] (Y. Shindo).
0011-2275/$ - see front matter 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.cryogenics.2004.04.008
Cryogenics 44 (2004) 789792
www.elsevier.com/locate/cryogenics
http://mail%20to:%[email protected]/http://mail%20to:%[email protected]/ -
7/27/2019 1-s2.0-S0011227504001031-main
2/4
et al. [1] examined the fracture behavior of CT speci-
mens made from austenitic stainless steels of differing
stability in a 4 K, 8 T magnetic field environment. The
least stable alloy showed a large reduction in the 4-K
fracture toughness with an 8 T magnetic field. The
amount of fracture toughness reduction with an 8 T
magnetic field decreased as the stability of the specimens
increased. They found that this difference in fracture
behavior is attributed to the enhancement of martensitic
transformation about the crack tip during the fracture
process in a magnetic field.
Because of the limited space available in high-field
mechanical property testing facilities, a test specimen
much smaller than the standard CT specimen would be
extremely desirable. Shindo et al. [7] examined the use of
small punch (SP) testing to estimate fracture toughness of
austenitic stainless steels and weld metals at 4 K, and
assessed correlation between equivalent fracture strain
and elasticplastic fracture toughness JIC. Following JIS
Z 2284 [8] standard test method, all JIC data were ob-tained using 25-mm-thick CT specimens. Shindo et al. [9]
using circumferentially notched bar specimens, also
investigated the cryogenic fracture toughness of austen-
itic stainless steels and weld metals. Recently, Yamaguchi
et al. [10] investigated the effect of magnetic field on the
cryogenic fracture toughness of alloy 908, a ferromag-
netic austenite, using SP and notch tensile specimens. The
4-K fracture properties of alloy 908 were not changed
significantly by magnetic field. The theoretical model [11]
predicted a negligible magnetic field effect on the stress
intensity factor for a crack in low-permeability materials.
The purpose of this study is to examine the effect
of magnetic field on the cryogenic fracture properties
of austenitic stainless steels, and establish the suitabil-
ity of SP testing technique for fracture characterization
of cryogenic structural materials at 4 K in magnetic
fields. SP tests were performed with thin plate specimens
at 4 K in magnetic fields of 0 and 6 T. Two austenitic
stainless steels of differing stability, SUS304 and SUS316
were selected for this test series. A method outlined by
Shindo et al. [7] was applied to predict a SP test-based
fracture toughness from equivalent fracture strain.
2. Experimental procedures
2.1. Materials and specimen
The compositions of the commercial SUS304 and
SUS316 plates used in this work are listed in Table 1.
SUS304 is metastable with respect to austenitic-
to-martensitic transformation and undergo a phase
transition from an fcc structure to a more stable bcc
martensite on deformation at low temperatures; SUS316
is stable with respect to austenitic-to-martensitic trans-
formation. Both were obtained as 10 mm thickness plate
in the cold-rolled condition. The initial volume fraction
of a0 martensite was determined using the relationship
between saturation magnetization and amount of a0
martensite obtained by X-ray diffraction [12]. The sat-
uration magnetization was measured using a vibrating
sample magnetometer. The percent martensite is 0.8%
for 304 and 0.4% for 316. Small, thin plate specimens
of 10 10 0.5 mm were sliced by electro discharge
machining and tested in the as-received condition. The
SP specimens were oriented with the thickness direction
parallel to the rolling direction.
2.2. Testing method
The high-field SP testing was done using a load frame
designed and constructed to fit into the bore of a 8 T
superconducting magnet with a 77 mm diameter work-
ing bore. The punch and the specimen holder, designed
for SP tests, are shown in Fig. 1. The SP specimen
holder consisted of an upper and lower die and four
clamping screws. All test fixtures were fabricated using
SUS310 that is completely stable with respect to mar-
tensitic transformation.
SP tests were performed at 4 K in magnetic fields of 0
and 6 T. The specimens were oriented such that the axis
of the solenoid field was parallel to the specimen loaddirection. The stroke rate was 0.2 mm/min. Displace-
Table 1
Chemical compositions of SUS304 and SUS316 stainless steels (wt%)
C Si Mn P S Ni Cr Mo
SUS304 0.06 0.47 0.89 0.028 0.001 8.54 18.28
SUS316 0.05 0.43 1.04 0.023 0.001 13.69 16.46 2.51
Fig. 1. Schematic of small punch test device.
790 Y. Shindo et al. / Cryogenics 44 (2004) 789792
-
7/27/2019 1-s2.0-S0011227504001031-main
3/4
ment was measured by measurement of the motion of
the punch relative to the lower die using a 20-mm-
diameter ring-shaped clip-on gage. The clip-on gage
response with and without a 6 T magnetic field was
measured at 4 K. The effect of magnetic field on the clip-
on gage response was negligible.
3. Results and discussion
All loaddisplacement curves for both SUS304 and
316 SP specimens are represented in Figs. 2 and 3. In the
4-K SP tests of austenitic stainless steels and weld met-
als, an approximate definition of equivalent fracture
strain eqf was adopted [7]:
eqf 0:0756dmax
t0
1:83; 1
where dmax is the displacement at peak load and t0 is theinitial thickness. The SP test-based fracture toughness
JCSP can be estimated from the equivalent fracturestrain as
JCSP 957:7eqf 29:5; 2
where JCSP has units of kJ/m2. Standard deviation is
24 kJ/m2. The JCSP values of SUS304 decrease 30%from approximately 303 to 213 kJ/m2 on going from 0 to
6 T, while those values of SUS316 decrease 5% from 0 T
condition (274 kJ/m2) to 261 kJ/m2 at 6 T. Decreases in
fracture toughness are detected with an applied mag-
netic field, depending on the alloy stability. The lessstable SUS304 has the large percentage reduction while
SUS316 has not changed significantly. The transfor-
mation related factor important to the fracture behavior
is the formation of lower toughness a0 martensite. A
material that transforms easily produces a brittle zone,
reducing its measured fracture toughness.
Our results showing a decrease in fracture toughness
with application of a magnetic field are consistent withsome experimental data obtained from CT specimens
[3,4]. At present, fracture toughness is usually deter-
mined using CT specimens, which are rather expensive
to manufacture and require relatively sophisticated
laboratory equipment to test. The advantages of the SP
test are reduced specimen machining costs compared to
CT specimens and simplified test methodology. The SP
test approach is shown to be a viable option for evalu-
ating the material toughness at 4 K in magnetic fields.
4. Conclusions
The 4-K fracture toughness of austenitic stainless
steels, SUS304 and SUS316, in magnetic fields are
characterized by small punch test. The results are sum-
marized as follows:
1. SUS304 shows a decrease in the measured fracture
toughness at 4 K with the application of magnetic
field.
2. The magnetic field has a measurable influence on the
4-K fracture toughness of SUS304.
3. For SUS316, the magnetic field effect is not large
enough to affect mechanical design.4. The magnitude of change in 4-K fracture toughness
with the application of magnetic field is a function
of the stability of the alloy.
References
[1] Chan JW, Chu D, Sunwoo AJ, Morris Jr JW. Metastable
austenites in cryogenic high magnetic field environments. Adv
Cry Eng 1992;38:559.
[2] Fultz B, Morris Jr JW. Effects of high magnetic fields on the flow
stress of 18-8 stainless steels. Acta Metall 1986;34(3):37984.
0.5 1.0 1.5
0.5
1.0
1.5
0
0 T
6 T
Displacement (mm)
Load(kN)
SUS304
SP test
4 K
Fig. 2. Loaddisplacement curves for SUS304 SP specimens.
0.5 1.0 1.5
0.5
1.0
1.5
0
0 T
6 T
Displacement (mm)
Load
(kN)
SUS316
SP test
4 K
Fig. 3. Loaddisplacement curves for SUS316 SP specimens.
Y. Shindo et al. / Cryogenics 44 (2004) 789792 791
-
7/27/2019 1-s2.0-S0011227504001031-main
4/4
[3] Fukushima E, Kobatake S, Tanaka M, Ogiwara H. Frac-
ture toughness tests on 304 stainless steel in high magnetic
fields at cryogenic temperatures. Adv Cry Eng 1988;34:367
70.
[4] Murase S, Kobatake S, Tanaka M, Tashiro I, Horigami O,
Ogiwara H, Shibata K, Nagai K, Ishikawa K. Effects of a high
magnetic field on fracture toughness at 4.2 K for austenitic
stainless steels. Fus Eng Des 1993;20:4514.
[5] Chan JW, Glazer J, Mei Z, Kramer PA, Morris Jr JW. Fracturetoughness of 304 stainless steel in an 8 Tesla field. Acta Metall
Mater 1990;38(3):47987.
[6] Pickering FB. Physical metallurgy of stainless steel developments.
Int Mat Rev 1976;21:22768.
[7] Shindo Y, Horiguchi K, Sugo T, Mano Y. Finite element analysis
and small punch testing for determining the cryogenic fracture
toughness of austenitic stainless steel welds. ASTM J Test Eval
2000;28(6):4317.
[8] JIS Z 2284, Method of elasticplastic fracture toughnessJIC testing
for metallic materials in liquid helium. Japanese Standards
Association; 1998.
[9] Shindo Y, Mano Y, Horiguchi K, Sugo T. Cryogenic fracture
toughness determination of a structural alloy weldment by notch
tensile measurement and finite element analysis. ASME J Eng
Mater Technol 2001;123(1):4550.
[10] Yamaguchi Y, Horiguchi K, Shindo Y, Sekiya D, Kumagai S.
Fracture and deformation properties of NiFe superalloy in cryo-genic high magnetic field environments. Cryogenics 2003;43(8):46975.
[11] Shindo Y. The linear magnetoelastic problem for a soft ferro-
magnetic elastic solid with a finite crack. ASME J Appl Mech
1977;44(1):4750.
[12] Kurita Y, Emura S, Fujita K, Nagai K, Ishikawa K, Shibata K.
Effects of magnetic fields on martensitic transformation and
serration of austenitic FeNi and FeCrNi steels at 4 K. Fus Eng
Des 1993;20:44550.
792 Y. Shindo et al. / Cryogenics 44 (2004) 789792