PRIMING DRAWING THE CLEARING VOLUME VAMP Jr. System Inservice Bulletin Venous Arterial blood Management Protection system Safe and Accurate Closed Blood Sampling System Featuring the Edwards TruWave Disposable Pressure Transducer 3 Provide flow by pulling Snap-Tab of the Edwards TruWave disposable pressure transducer. Slowly deliver priming solution to remove air. 4 Close plunger and remove all air from the system. Ensure that VAMP Jr. and catheter connection is a fluid-to-fluid connection. 1 Open reservoir plunger to approximately 1/2 cc to facilitate flow of priming solution. 2 With the shut-off valve in the open position (parallel to the tubing), hold sampling site above the Edwards VAMP Jr. reservoir at 45° angle. 6 Close the proximal shut-off valve by turning handle perpendicular to tubing. 5 Close the distal shut off valve by turning the handle perpendicular to the tubing. Smoothly and evenly pull up on the reservoir plunger to draw the required amount of clearing volume consistent with the patient’s clinical condition, patient’s size, or your hospital protocol. (Recommended pull rate is 1 cc every 10-15 seconds.)

Transcript of VamP Jr. System inservice Bulletin - Edwards...

Page 1: VamP Jr. System inservice Bulletin - Edwards Lifesciencesht.edwards.com/scin/edwards/sitecollectionimages/edwards/... · 2012-07-31 · Edwards Lifesciences Irvine, USA I Nyon, Switzerland

P r i m i n g

D r a w i n g t h e C l e a r i n g V o l u m e

VamP Jr. System inservice BulletinVenous arterial blood management Protection systemSafe and Accurate Closed Blood Sampling SystemFeaturing the Edwards TruWave Disposable Pressure Transducer

3 ProvideflowbypullingSnap-Tab

oftheEdwardsTruWave

disposable pressuretransducer.

Slowlydeliverprimingsolutiontoremoveair.

4 Closeplungerandremoveallairfromthesystem.EnsurethatVAMPJr.andcatheterconnectionisafluid-to-fluidconnection.

1 Openreservoirplungertoapproximately

1/2cctofacilitateflowofprimingsolution.

2 Withtheshut-offvalveintheopenposition(paralleltothetubing),holdsamplingsiteabove

theEdwards VAMPJr.

reservoirat45°angle.

6 Closetheproximalshut-offvalveby

turninghandleperpendicular

totubing.

5 Closethedistalshutoffvalvebyturningthehandleperpendiculartothetubing.Smoothlyandevenlypulluponthereservoirplungertodrawtherequiredamount

ofclearingvolumeconsistentwiththepatient’sclinicalcondition,patient’ssize,oryourhospital

protocol.(Recommendedpullrateis1ccevery 10-15seconds.)

Page 2: VamP Jr. System inservice Bulletin - Edwards Lifesciencesht.edwards.com/scin/edwards/sitecollectionimages/edwards/... · 2012-07-31 · Edwards Lifesciences Irvine, USA I Nyon, Switzerland

Edwards Lifesciences Irvine,USAINyon,SwitzerlandITokyo,JapanISingapore,SingaporeISãoPaulo,Braziledwards.com

Drawing BlooD SamPleS from the VamP Jr. neeDleleSS SamPling Site

For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events.EdwardsisatrademarkofEdwardsLifesciencesCorporation.EdwardsLifesciences,thestylizedElogo,TruWave,VAMPandVAMPJr.aretrademarksofEdwardsLifesciencesCorporationandareregisteredintheUnitedStatesPatentandTrademarkOffice.©2011EdwardsLifesciencesCorporation.Allrightsreserved.AR05992

No components of this package or the product it contains are made from natural rubber latex or dry natural rubber.

10 Opentheproximalshut-offvalvebyturninghandleparalleltotubing.

11 Slowly,smoothly,andevenly,pushdownontheplungeruntilitisfullyclosed.(Recommendedrateis1ccevery10-15seconds.)

12 Fillsyringewith flushsolution. AttachInterlink NeedleLess cannulatothe syringe.Ensure syringeandcannula arefreeofairbubbles. Swabdistalaccessport. Insertsyringeand cannulaintothe distalaccessport.Slowlyflushlineperhospital

procedure.Openthedistalone-wayvalve.

7 Swabproximalsamplesite.Ensureplungerisdepressedtothebottomofthesyringe.Attachcannula.Pushcannulaintoproximalsamplesite.

9 Whileholdingthecannula,removethesyringe/

cannulafromthesamplingsitebypullingstraightout.

8 Slowlydrawthebloodsample.