Qaqc Course Total

32
Quality is a Lousy Quality is a Lousy Idea- Idea- If it’s Only an Idea If it’s Only an Idea

Transcript of Qaqc Course Total

Page 1: Qaqc Course Total

Quality is a Lousy Idea-Quality is a Lousy Idea-

If it’s Only an IdeaIf it’s Only an Idea

Page 2: Qaqc Course Total

Quality Assurance vs. Quality Assurance vs. Quality ControlQuality Control

Quality AssuranceQuality AssuranceAn overallmanagement plan to guarantee theintegrity of data(The “system”)

Quality ControlQuality ControlA series of analytical measurements usedto assess thequality of the analytical data(The “tools”)

Page 3: Qaqc Course Total

True Value vs. Measured True Value vs. Measured ValueValue

True ValueTrue Value

The known, accepted value of a quantifiable property

Measured ValueMeasured Value

The result of an individual’s measurement of a quantifiable property

Page 4: Qaqc Course Total

Accuracy vs. PrecisionAccuracy vs. Precision

AccuracyAccuracyHow well a measurement agrees with an accepted value

PrecisionPrecisionHow well a series of measurements agree with each other

Page 5: Qaqc Course Total

Accuracy vs. PrecisionAccuracy vs. Precision

Page 6: Qaqc Course Total

Systematic vs. Systematic vs. Random ErrorsRandom Errors

Systematic ErrorSystematic ErrorAvoidable error due to controllable variables in a measurement.

Random ErrorsRandom ErrorsUnavoidable errors that are always present in any measurement. Impossible to eliminate

Page 7: Qaqc Course Total

Quality Control MeasuresQuality Control Measures

• Standards and Calibration

• Blanks

• Recovery Studies

• Precision and Accuracy Studies

• Method Detection Limits

• NJQLs

Page 8: Qaqc Course Total

Standards and CalibrationStandards and Calibration

• Prepared vs. Purchased Standard

• Signals: Peak Area, Beer’s Law

• Calibration Curves

• Continuing Calibration Checks

• Internal Standards• Performance Testing.

Page 9: Qaqc Course Total

Calibration CurvesCalibration Curves

Graphical representation of the Graphical representation of the relationship between:relationship between:

• The analytical signal

• The concentration of the analyteand

Page 10: Qaqc Course Total

Calibration Curve for DDT

y = 9.3005x + 4.3313

0

100

200

300

400

500

0 10 20 30 40 50 60

Parts per trillion DDT

Pea

k ar

ea x

10

6

R2 = 0.9989

Page 11: Qaqc Course Total

Continuing Calibration Continuing Calibration VerificationVerification

• Many methods don’t require that daily calibration curves are prepared

• A “calibration verification” is analyzed with each batch of samples

Page 12: Qaqc Course Total

Sample BatchSample Batch• 10 - 20 samples (method defined)

or less

• Same matrix

• Same sample prep and analysis

• Contains a full set of QC samples

Page 13: Qaqc Course Total

Internal StandardsInternal Standards• A compound chemically similar to the

analyte

• Not expected to be present in the sample

• Cannot interfere in the analysis• Added to the calibration standards and

to the samples in identical amounts.

Page 14: Qaqc Course Total

Internal StandardsInternal Standards• Refines the calibration process• Analytical signals for calibration

standards are compared to those for internal standards

• Eliminates differences in random and systematic errors between samples and standards

Page 15: Qaqc Course Total

Performance TestingPerformance TestingBlind samples submitted to laboratories

??

?Labs must periodically analyze with acceptable results in order to maintain accreditation

Page 16: Qaqc Course Total

Blanks, Blanks, BlanksBlanks, Blanks, Blanks• Laboratory Reagent Blanks

• Instrument Blanks

• Field Reagent Blanks

• Trip Blanks

Page 17: Qaqc Course Total

Laboratory Reagent BlanksLaboratory Reagent Blanks• Contains every reagent used in the

analysis• Is subjected to all analytical

procedures• Must give signal below detection limit• Most methods require one with every

batch

Page 18: Qaqc Course Total

Instrument BlankInstrument Blank

• A clean sample (e.g., distilled water) processed through the instrumental steps of the measurement process; used to determine instrument contamination.

Page 19: Qaqc Course Total

Field Reagent BlanksField Reagent Blanks

• Prepared in the lab, taken to the field

• Opened at the sampling site, exposed to sampling equipment, returned to the lab.

Page 20: Qaqc Course Total

Trip BlanksTrip Blanks• Prepared in the lab, taken to the

field

• Not opened

• Returned to the lab

• Not always required in EPA methods

Page 21: Qaqc Course Total

Recovery StudiesRecovery Studies

• Matrix Spikes

• Laboratory Control Samples

• Surrogates .

Page 22: Qaqc Course Total

Matrix SpikesMatrix Spikes

• Sample spiked with a known amount of analyte

• Subjected to all sample prep and analytical procedures

• Determines the effect of the matrix on analyte recovery

• Normally one per batch

Page 23: Qaqc Course Total

Laboratory Control SampleLaboratory Control Sample

• Subjected to all sample prep and analytical procedures

• Analyte spiked into reagent water

Page 24: Qaqc Course Total

Laboratory Control SampleLaboratory Control Sample

Also known as:Also known as:

• Laboratory Fortified Blank (LFB)

• Quality Control Sample (QCS)

Page 25: Qaqc Course Total

SurrogatesSurrogates

• Similar to an internal standard

• Added to all analytical samples, and to all QC samples to monitor method performance, usually during sample prep

• Methods often have specific surrogate recovery criteria

• Most common in Organic methods

Page 26: Qaqc Course Total

Quality Control MeasuresQuality Control Measures

• Standards and Calibration

• Blanks

• Recovery Studies

• Precision and Accuracy Studies

• Method Detection Limits

• NJQLs

Page 27: Qaqc Course Total

Precision and AccuracyPrecision and Accuracy

• Required for initial certification and annually thereafter

• A series of four laboratory control samples

• Must meet accuracy (recovery) and precision (standard deviation) requirements, often in method

Page 28: Qaqc Course Total

Precision and AccuracyPrecision and Accuracy

• Required with a change in instrumentation or personnel

• Specific to the analyst

• Other names include:

P&A, DOC, IDOC

Page 29: Qaqc Course Total

Method Detection LimMethod Detection Limitit “The minimum concentration of a

substance that can be measured and reported with 99% confidence that the analyte concentration is greater than zero”

N.J.A.C 7:18 - 1.7N.J.A.C 7:18 - 1.7

Page 30: Qaqc Course Total

Method Detection LimitMethod Detection Limit

• MDLs are determined according to 40 CFR, part 136, Appendix B

• Seven replicate laboratory control samples, analyzed for precision

• Multiply standard deviation by 3.14 (Student’s t- value)

Page 31: Qaqc Course Total

Method Detection LimitMethod Detection Limit• Must be performed initially for

certification• Must meet criteria specified in method• Must be performed with change in

instrumentation or test method• Annually with ELCP

Page 32: Qaqc Course Total

New Jersey New Jersey Quantitation LimitsQuantitation Limits

(NJQLs)(NJQLs)• The minimum concentration of an

analyte that can be quantified with statistical confidence

• 5 x MDL, for the NJ Lab Certification Program