Tools for Shape Analysis of Vascular Response using Two Photon Laser Scanning Microscopy

43
Tools for Shape Analysis of Vascular Response using Two Photon Laser Scanning Microscopy By Han van Triest Committee: Prof. Dr. Ir. B.M. ter Haar Romeny Dr. M. A. M. J. van Zandvoort Dr. Ir. H. C. van Assen A. Vilanova i Bartrolí, PhD R.T.A. Megens, MSc

description

Tools for Shape Analysis of Vascular Response using Two Photon Laser Scanning Microscopy. By Han van Triest. Committee: Prof. Dr. Ir. B.M. ter Haar Romeny D r. M. A. M. J. van Zandvoort D r. Ir. H. C. van Assen A. Vilanova i Bartrolí, PhD R.T.A. Megens, MSc. Overview. - PowerPoint PPT Presentation

Transcript of Tools for Shape Analysis of Vascular Response using Two Photon Laser Scanning Microscopy

Page 1: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

Tools for Shape Analysis of Vascular Response using

Two Photon Laser Scanning Microscopy

ByHan van Triest

Committee:

Prof. Dr. Ir. B.M. ter Haar RomenyDr. M. A. M. J. van ZandvoortDr. Ir. H. C. van AssenA. Vilanova i Bartrolí, PhDR.T.A. Megens, MSc

Page 2: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

2/43

Overview

1. Biological Introduction2. Technical Introduction3. Vessel Radius Estimation4. Cell counting5. Conclusion and Recommendations

Page 3: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

3/43

Biological Introduction

Vascular diseases are a big problem in the western world.

It is estimated that arteriosclerosis is the underlying cause of 50 % of all deaths in the western world

To unravel the underlying mechanisms more research is required

Page 4: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

4/43

Biological Introduction – Vessel Anatomy

Page 5: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

5/43

Biological Introduction – Remodelling

Page 6: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

6/43

1. Excitation2. Energy loss processes3. Emission

Energy of Photon:

hhE

Technical Introduction – Fluorescence

1

2

3

Page 7: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

7/43

Technical Introduction – Confocal Laser Scanning Microscopy

Page 8: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

8/43

Technical Introduction – Confocal Laser Scanning Microscopy

Advantages:• Optical sectioning

Disadvantages:• Excitation of out-of-focus regions• High energy of excitation photons• Low penetration depth

Page 9: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

9/43

Technical Introduction – Two Photon Laser Scanning Microscopy

Page 10: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

10/43

Technical Introduction – Two Photon Laser Scanning Microscopy

Page 11: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

11/43

Technical Introduction – Two Photon Laser Scanning Microscopy

Advantages:• No pinhole to block out-of-focus light required• Increased penetration depth• Excitation photons of lower energy• Imaging of viable tissue• Multiple dyes usable for targeting of different structures

Disadvantage:• Higher wavelength limits maximal achievable resolution

Page 12: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

12/43

Technical Introduction – Imaging

Page 13: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

13/43

Processing – Description of Vessels

Features:• Radius of the vessel• Ratio vessel wall thickness – vessel radius• Cell volume fraction

Needed:• Vessel Radius• Number of Cells

Page 14: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

14/43

Radius Estimation

Page 15: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

15/43

Radius Estimation – Methods

1. Statistical methods: Least squares estimators2. Robust statistics: Reduction of the influence of outliers3. Hough Transform

Page 16: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

16/43

Radius Estimation – Hough Transform

xayb

bxay

ˆˆ

ˆˆ

Line through a point in image space

Set of parameters that describe the point

Page 17: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

17/43

Radius Estimation – Hough Transform

Page 18: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

18/43

Radius Estimation – Circular Hough Transform

Circle can be described by: 222 )()( ryyxx cc

Page 19: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

19/43

Radius Estimation – Hough Transform

Advantages:• Robust against noise• Able to find partly occluded objects

Disadvantages:• Expensive, both computational and memory cost

Page 20: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

20/43

Radius Estimation – Proposed method

A circle is defined by three non co-linear points.

• Store only center coordinates• Weight vote by average distance

between p1, p2 and p3

Find r by voting for most likely value of the radius

Page 21: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

21/43

Radius Estimation – Finding Edge Points

A global threshold is infeasible due to differences in optical paths for emitted photons

Page 22: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

22/43

Radius Estimation – Finding Edge Points

Modified Full Width at Half Maximum:

maxI InsideOutside

Page 23: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

23/43

Radius Estimation – Experiments

• 20 images, 10 single slices, 10 taken from three dimensional stacks

• Test images have both sides of the wall vissible• Groundtruth given by the average estimate of 12

volunteers• Results compared with common least squares

estimator• Tests are performed for values of α between 0.2 and

0.8 in steps of 0.05, and using 20 to 250 points in steps of 10

• In total 24960 estimates are made

Page 24: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

24/43

Radius Estimation – Influence of α

xz-scan:

z-stack slice:

Blue line: LSE

Red line: MHT

Page 25: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

25/43

Radius Estimation – Influence of number of points

xz-scan:

z-stack slice:

Blue line: LSE

Red line: MHT

Page 26: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

26/43

Radius Estimation – Conclusion

• Proposed method outperforms least squares fitting method for xz-scans

• Proposed method performs equally compared to least squares fitting method for z-stack slices

• The best value for α used in the proposed method is α = 0.4

• At least 100 points is required for a stable result using the proposed method

Page 27: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

27/43

Cell counting

Page 28: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

28/43

Cell Counting – Algorithm

Noise Reduction

Potential Center Extraction

Potential Edgepoint Extraction

Edgepoint Selection

Ellipsoid Fitting

Oversegmentation Reduction

Page 29: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

29/43

Cell Counting – Noise Reduction

Edge-preserving filtering: Median

Filtering

Each pixel is replaced by the median of its surrounding

Purple line: Original objectBlue line: Degraded objectRed line: Median filter, kernel

width 5 pixelsBlack line: Median filter, kernel

width 25 pixels

Page 30: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

30/43

Cell Counting – Potential Center Detection

Assumption: Blob-like structures

Center is maximum of the blob

Local maxima within a region are potential centers.

Page 31: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

31/43

Cell Counting – Potential Edgepoint Extraction

• Sample rays from each potential center• Rays intersect points along a generalized spiral

Page 32: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

32/43

Cell Counting – Potential Edgepoint Extraction

Constraint:Points on a downward flank

These points can be found at points in which the second order derivative switches from negative to positive.

Blue line: Image intensity along ray

Purple line: First order derivative

Sienna line: Second order derivative

Page 33: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

33/43

Cell Counting – Dynamic Programming

B

c

d

a

b

A

2

4

5 2

3

6

2

6

7 4

25

9

Shortest Route: AbcB

Page 34: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

34/43

Cell Counting – Edgepoint Selection

Find set of most likely edge points

Cost function: 2

max

2max

111

21

,

I

pw

I

ppwppwppC

Ii

l

Ii

Ii

ici

cisii

Page 35: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

35/43

Cell Counting – Ellipsoid Fitting

0222 dzryqxpyxhzxgzyfzcybxaQ

Ellipsoid can be described by a quadric, a general polynomial in three dimensions of order two:

Axes proportions

Orientation Position Size

Fitted on the data using a least squares fitting procedure

Page 36: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

36/43

Cell Counting – Oversegmentation Reduction

1. Find overlapping nuclei2. Check wether nuclei are parallel3. Merge the sets of edgepoints of parallel overlapping nuclei4. Perform Ellipsoid fitting on the combined data sets

Page 37: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

37/43

Cell Counting – Results Before Merging

Page 38: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

38/43

Cell Counting – Results After Merging

Page 39: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

39/43

Cell Counting – Discussion

Three types of frequent mistakes:A Incorrect merging of two blunt nucleiB Center of cell not foundC No distinct directions

Page 40: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

40/43

Cell Counting – Discussion

A another problem is due to leakage of light from other colors

Page 41: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

41/43

Cell Counting – Conclusion

Although the method only has been tested on a single dataset, the results show to be promising.

Most of the cells are found while there is a relatively small amount of false negatives and false positives

Page 42: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

42/43

Recommendations

• Test the algorithm on more datasets• Investigate the influence of parameters• For the calculation of the cost during the dynamic programming

step, take into account more points on the surface• Remove outliers in the selected set, as outliers have great effect

on the least squares algorithm• Optimize the imaging parameters to get as litle non cellular

structures as possible • Classify the cells into subclasses

Page 43: Tools for Shape Analysis of Vascular Response using  Two Photon Laser Scanning Microscopy

Questions?