Syringes for gas chromatography
Transcript of Syringes for gas chromatography
-
7/29/2019 Syringes for gas chromatography
1/4
278 LCGC NORTH AMERICA VOLUME 24 N UMBER 3 MARCH 2006 www.chromatographyonline.com
John V. HinshawGC Connections Editor
GC CONNECTIONSGC
Syringes for GasChromatography
A bewildering variety of
syringes is available for
gas chromatography.Volumes range from 1
L to as much as 2 mL,
with many different
needle styles and tips,
plunger arrangements,
and accessories. Syringe
styles include both
manual and autosampler
types. Choosing a suitable
syringe for a specific
application can be
difficult, especially if the
inlet system has special
requirements; choosing
the wrong syringe can
cause significant
problems.
n liquid chromatography (LC), the
syringe functions primarily as a
pipette or liquid-transfer device
that loads a sample loop. The syringe
generally does not take an active role in
injection, which occurs only after samplehas been displaced from the syringe. The
same is true of most gas chromatography
(GC) gas-sampling valves the time at
which the gaseous sample is injected into
the column is separate from the moment
that it is transferred into the injection
system. In GC analysis of liquid samples,
however, the syringe becomes an integral
part of the inlet during injection: sample,
in liquid or gaseous form, starts to enter
the column as soon as the syringe enters
the inlet.In GC inlet systems for liquids, the
injection technique, choice of syringe,
and inlet operating conditions all play a
crucial role in the injection process. Two
principal sample-transfer mechanisms
move sample from the syringe into the
inlet while the syringe is in the inlet.
First, liquid-sample transfer takes place as
the syringe plunger is depressed and liq-
uid is expelled from the syringe tip. In
cold injection, where the inlet tempera-
ture is not high enough to produce sig-nificant solvent vaporization, this
pipette-like action is the only major sam-
ple-transfer mechanism. A competing
process occurs in a hot injector, however.
Within a few tenths of a second after the
needle enters a hot inlet, sample begins
to evaporate inside the needle. Bubble
formation and concomitant increased
internal pressures force some liquid out
along with the vapor, so that part or all
of the sample contained in the syringe
needle volume is introduced into theinlet as a mixture of liquid and vapor. As
the plunger is depressed, additional
room-temperature liquid sample is forced
from the syringe through the needle,
which cools the needle and suppresses
but does not entirely stop in-needle evap-
oration. The needle heats up again oncesyringe plunger motion ceases, which
causes additional sample vaporization
from the needle into the inlet. All of
these processes take place in a matter of
seconds. The total amount of sample that
actually is injected into the inlet depends
strongly upon these two processes, their
timing, the volumes involved, and the
inlet conditions. Along with judicious
injection condition control, a good
understanding of the role of the syringe
in these processes will help gas chro-matographers obtain better injections.
Sample Distortion During
Injection
Syringe-needle effects influence not only
the injected sample volume; they also can
modify the relative amounts of individ-
ual sample components that enter the
inlet. To understand this secondary
effect, consider that not all sample com-
ponents have the same vapor pressures at
a given temperature. Lower molecularweight compounds have higher vapor
pressures, and conversely, heavier mole-
cules have lower vapor pressures. This
differentiation forms the basis for simple
thermal fractionation of a mixture of
compounds in a distillation column,
for example.
Unfortunately, for many gas chro-
matographers, the same kind of thermal
fractionation process can occur in the
syringe needle during injection. When
in-needle vaporization occurs, the lightercomponents vaporize first and leave the
I
-
7/29/2019 Syringes for gas chromatography
2/4
www.chromatographyonline.com280 LCGC NORTH AMERICA VOLUME 24 N UMBER 3 MARCH 2006
syringe needle quickly. Heavier com-
pounds take longer to evaporate and
leave the syringe needle more slowly. The
effect of initial needle heating during
injection is remediated largely by subse-
quent bulk liquid transfer through the
needle: straggling heavy compounds will
be rinsed out. At the end of injection,
however, if the needle is withdrawn fromthe inlet before complete transfer of all
compounds can occur, then the sample
that enters the inlet system will contain
more of the volatile components and less
of the heavier components than were
present in the sample before injection.
This effect is called mass discrimination
because it tilts the sample composition
according to components vapor pres-
sures, which are related largely to their
masses.
Two other thermal side-effects gohand-in-hand with needle fractionation.
Many compounds are sensitive to ther-
mal stress, especially in the presence of a
hot steel syringe needle, and decompose
rapidly when subjected to high inlet tem-
peratures: high molecular weight glyc-
erols are one good example. Polar com-
pounds can become strongly adsorbed on
the needle surface. Many of the polar
polyaromatic hydrocarbons (PAHs) as
well as a host of chlorinated pesticides
suffer from this problem. In both cases,the use of nickel syringe needles, or nee-
dles that have been deactivated by
advanced surface treatments, can help
tremendously, as can deactivation of inlet
liners and injector inner surfaces exposed
to the sample. In these situations, ther-
mally mild injection techniques such as
cold on-column injection or pro-
grammed temperature vaporization usu-
ally produce results superior to what can
be obtained by careful deactivation of
needles and other hot injection materials,but the specialized inlets needed for these
somewhat more complex techniques
might not be available.
Judicious data-system calibration and
the use of multiple internal standards can
compensate for inaccuracies caused by
mass discrimination, decomposition, and
adsorption, as well as for a host of other
injection problems. However, the preci-
sion and repeatability of multiple analy-
ses are affected adversely when the rela-
tive amounts of individual componentsstrongly depend upon injection condi-
tions. Heavier components against
which injection strongly discriminates
or any other components that do not
come through the injection process at
100%, will exhibit degraded lower detec-
tion limits (LDLs) as well. Degradation
of precision, repeatability, and LDL are
significant effects even for samples that
do not span a wide range of volatilitiesand do not suffer from mass discrimina-
tion per se.
When this type of injection problem
does arise, its effects must be either sup-
pressed or controlled by applying special
syringe-handling techniques or by using
cold injection if possible. The sandwich
technique is a good example. The syringe
is first filled with 1 L or so of pure liq-
uid solvent followed by a small air plug.
The sample is pulled into the syringe,
then another plug of air, and finally somemore solvent. The entire liquid array is
pulled up into the syringe so that the
needle is largely empty. Upon injection,
only a little solvent is in the needle as it
enters the inlet. Sample passes through
the needle as the syringe plunger is
depressed, followed by a plug of solvent
that forces any remaining sample into
the inlet. Only pure solvent is left in the
needle at the end of injection, so needle
fractionation does not occur. The draw-
back of the sandwich technique is that itsignificantly increases the amount of sol-
vent that enters the inlet, which might
overload the inlet with solvent vapor or
broaden the solvent peak and interfere
with early-eluted analytes.
A high-speed autosampler also can
remediate needle fractionation. By inject-
ing the sample in a very short time, typi-
cally less than 500 ms and certainly
much faster than a human can achieve,
the syringe needle residence time in the
inlet is insufficient for significant needleheating to occur, and so needle fractiona-
tion is eliminated. Another benefit of a
fast injection is the suppression of sensi-
tive-component thermal decomposition
or adsorption on hot steel needle sur-
faces, again due to reduced contact
times. Most modern GC autosamplers
employ high-speed actions by default.
In-needle fractionation and other ther-
mal effects are some of the syringe-
related processes that interfere with injec-
tion accuracy and repeatability. There aremany other effects that occur before,
during, and after sample has entered the
inlet system: see the book by Grob (1)
for a comprehensive discussion.
Types of Syringes
The choice of syringe type plays an
important role in obtaining the best per-
formance from a gas chromatograph.
There are two basic kinds of manualsyringes: plunger-in-barrel (Figure 1a)
and plunger-in-needle (Figure 1b). In the
first type, sample is drawn up into a cali-
brated glass barrel; in the second, the
entire sample is contained in the needle,
and a thin wire plunger fitted into the
bore of the needle forces liquid out dur-
ing injection.
Plunger-in-Barrel Syringes
In this type of syringe, sample is con-
tained in the glass barrel (see Figure 1a).Because you can see the sample, it is easy
to observe the presence of liquid sample
and determine if the syringe needle is
obstructed. In use, the syringe usually is
overfilled with liquid, and the extra is
then ejected. A final clinging droplet can
be removed from the needle tip with a
clean laboratory wipe or by rinsing with
solvent. To gauge the injected volume,
consider that the amount displaced from
the markings on the syringe barrel does
not include sample vaporized from theneedle and is only an approximation.
The actual amount injected is closer to
the needle volume (about 0.7 L for a
10-L syringe) plus the amount dis-
placed. Another way to determine
injected volume is to load the syringe
with sample, pull the plunger back, and
read the total volume of the liquid plug.
After injection, pull the plunger back
slightly, read the remaining liquid vol-
ume, and subtract it from the original
amount. In this way, sample evaporatedfrom the needle is included in the total
amount. Plunger-in-barrel syringes can
be used for most routine applications.
The common 10-L syringe can reliably
inject liquid volumes ranging from 1.0
L to 10 L.
Plunger-in-barrel syringes can be
cleaned in several ways. Pumping solvent
in and out of the syringe is effective and
can be a good idea between injections to
remove leftover sample before it evapo-
rates and leaves a film of residue behind.Eventually, a layer of nonvolatile residue
-
7/29/2019 Syringes for gas chromatography
3/4
-
7/29/2019 Syringes for gas chromatography
4/4
fashion. These syringes tend to be more
tightly integrated with the autosampler
mechanics and are harder to replace. As
mentioned previously, and worth repeat-
ing, is the fact that autosampler syringes
need better care than manual syringes
because they are subjected to greater
mechanical stresses and more operating
cycles.Another familiar specialty syringe is
the cold on-column syringe. This is gen-
erally a plunger-in-barrel type with a spe-
cial removable needle made of narrow-
bore steel or fused-silica tubing. The
needle is designed to penetrate a narrow-
bore (0.20- or 0.25-mm i.d.) capillary
column to accomplish on-column injec-
tion. Such syringes are not interchange-
able across different injector designs; the
manufacturers specifications must be fol-
lowed. On-column syringes can be moresusceptible to needle blockage because of
their smaller inner diameters. Many on-
column injectors use a precolumn or
retention gap made of uncoated
0.53-mm i.d. tubing, which accepts a
standard 26-gauge syringe needle.
Syringe Repair
Many syringe manufacturers offer syringe
repair kits, needle replacements, and
repair service. Most removable needles
can be replaced in the field if the rest ofthe syringe is in good shape. For
plunger-in-needle syringes, needle dam-
age often is accompanied by plunger
damage, and in such cases, it is necessary
to replace both as a matched pair.
Syringe plungers generally are not
field-replaceable (or interchangeable)
because of the tight matching tolerances
between plunger and barrel. One excep-
tion is microliter gas-tight syringes with
PTFE-tipped plungers, which will accept
replacement plungers if the barrel is nototherwise damaged.
Is syringe repair worth the cost? In the
case of plunger-in-needle syringes, the
answer is a definite yes. Repair kits of
matched plungers and needles are one-
quarter to one-third the cost of a new
syringe. For the less-expensive plunger-
in-barrel types, replacement needles are
as much as half the cost of a new unit,
and it might be worthwhile instead to
attempt to reduce syringe damage and
extend syringe life through better han-dling and cleaning procedures.
John V. HinshawGC Connections edi-tor John V. Hinshaw issenior staff engineerat Serveron Corp.,Hillsboro, Oregon, anda member ofLCGCseditorial advisoryboard. Direct corre-spondence about thiscolumn to GC Con-
nections, LCGC, Woodbridge Corporate Plaza,485 Route 1 South, Building F, First Floor,Iselin, NJ 08830, e-mail [email protected].
For an ongoing discussion of GC issues withJohn Hinshaw and other chromatographers,
visit the Chromatography Forum discussiongroup at http://www.chromforum.com.
Syringe Accessories
A few syringe and sampling accessories
are especially useful. First, a syringe rack
is a good way to keep syringes in one
place, rather than on top of the gas chro-
matograph or the bench, where they can
be pushed off and broken easily. Syringe
labels are also handy to keep track of
which syringe is used for which type ofsample.
Another useful item is one of the vari-
ous repeating adapters that permit the
plunger to be returned to the same posi-
tion each time, which can improve the
repeatability of multiple injections. Also,
repeating adapters act as guides and pre-
vent plunger bending or accidental with-
drawal from the barrel, a feature that is
important on plunger-in-needle syringes.
Several manufacturers also make syringes
with extended or reinforced barrels andplungers that prevent damage from a
heavy hand.
Conclusion
When viewed as an active part of an
inlet, the syringe takes on new impor-
tance. While injection technique is cru-
cial, the proper care and cleaning of the
syringe can contribute significantly to
improved accuracy and precision. Choos-
ing the correct syringe to meet sample
and injector requirements, and thenapplying and maintaining it correctly,
can go a long way toward producing bet-
ter and more reliable analytical results.
References(1) K. Grob, Split and Splitless Injection for Quan-
titative Gas Chromatography: Concepts,Processes, Practical Guidelines, Sources of Error,4th, Completely Revised Edition (Wiley,Hoboken, New Jersey, 2001).
www.chromatographyonline.com284 LCGC NORTH AMERICA VOLUME 24 N UMBER 3 MARCH 2006
Circle 40