EFFECTS OF PERİPHERİC CATHATER INSERTİON SUPPORTED BY ACCUVEİN VASCULAR IMAGİNG DEVİCES
IN PEDİATRİC PATİENTS TO THE SUCCESS OF THE PROCESS
Duygu Demir1
Sevil İnal2
1Nurse, Pediatric Hematology and Oncology Service,Okmeydanı Educational and Research Hospital, İstanbul
2Istanbul University, Health Science Faculty, Midwifery [email protected]
INTRODUCTION
Intravenous cathater application is the most encountered medical procedure for hospitalized
pediatric patients
Mostly, IV cathater application is made in order to achieve the cure succesfully and early recovery of the
patients
Vessel diameter is
small
Difficult to Place The IV Catheter in
Children Than Adults
Difficult to palpate
PainfulHard to tolerate
Thick subcutaneous
tissue
Moving Difficult to stabilize
• Nurse IV catheter should be aimed at atraumatic care• For this reason, it is important to use technologies that
reduces the pain and also shorten the process time by the nurse.
İn the long-termIV catheter applications
Repeated IV applications
Difficult IV Catheter Applications
The purpose of the nurse;• with minimal pain• As soon as• the minimum number of intravenous venture
with processes must perform
• IV catheter placement in pediatric patients in the limited research demonstrating the efficacy of different vascular imaging device
• There is no research Accuvein AV 400 demonstrating the effectiveness of the device
Literature review;
OBJECTİVES
İt is aimed to determine effects of peripherical
intravenous cathater insertion supported by Accuvein
vascular imaging device for pediatric patients
THE SUCCESS OF THE PROCESS
Practice Number Per Patient
Cathater İnsertion Period
Pain Threshold Of The Children
Materyal-Metod
• Randomized, controlled experimental study
• Sample, 3-18 years old 129 pediatric patients
• Interview and Observation Form
• Wong-Baker FACES Pain Scale (WBFPS-R)
• AccuVein AV 400 vascular ımagıng devıces
Vein Viewing System(AccuVein AV 400)
The patient has not affected by movement
Under the skin can scan up to 8 mm
Real-time scanning and showing
Weighing only (280g)
Cordless and rechargeable
It is non-contact and portable
Ability to work in light and dark
(accuvein.com)
Use Manually or With Hanger
Two Color Options
White Black
HİPOTEZLER
H0. Peripheral intravenous catheter implementation supported by Accuvein vascular imaging devices in pediatric patients has no effect on the success of the procedure in children.H1. Peripheral intravenous catheter implementation supported by Accuvein vascular imaging devices in pediatric patients reduces the pain during the procedure in children. H2. Peripheral intravenous catheter implementation supported by Accuvein vascular imaging devices in pediatric patients reduces the processing time in children.H3. Peripheral intravenous catheter implementation supported by Accuvein vascular imaging devices in pediatric patients increases the success of the procedure in children.
INCLUSİON CRİTERİA
Case and control subjects have been selected children will create the random methodDue to the treatment procedure IV catheter placement needs to beBoth the experimental and control groups at the same hospital where the children research service in designated inpatient treatment to be seen (Pediatric Hematology (n=54), Pediatric İnternal Medicine(n=75) ) Subject must be within 3-18 ages Same nurse practicing intravenous catheter implementation Same needle size for every intravenous catheter implementation (24-G) Not having taken an analgesic within the last 6 hours The stable health status Not having a developmental disorder that may result in not being
able to answer the pain and anxiety scales (mental retardation or hearing impairment)
Experimental Group
AccuVein assisted
Control Group
Conventional (standard) application method
İntervention
• Data were analyzed with SPSS 15.00.• The quantitative data; Student t Test, Mann
Whitney U testi • The qualitative data Pearson chi-square tests ve
Yates Continuity Correction test • Significance is obtained in p<0,01 and p<0,05
level.
Statistical Analysis
RESULTS
Experimental Group(n=72)
Control Group (n=57)
Mean±SD (Med). Mean±SD (Med).
Age (year) 8,51±4,41 (7,0) 9,34±4,54 (9,5) Z:-0,969 b0,333
BMI (kg/m2) 17,27±3,08 (16,6) 17,34±2,53 (17,1) t: -0,140 a0,889
Number of vascular
access establishment before the process
3,14±1,73 (3) 2,76±1,92 (3) Z: -1,355 0,175
Table1. Comparing The Experimental And The Control Group Success İn Terms Of Process Variables That Can
Affect-1
aStudent t Test bMann Whitney U Test
n %n %
χ2 p
SexFemale 32 44,4 27 47,4 0,110 c0,741
Male 40 55,6 30 52,6
Skin tone
White 39 54,2 34 59,6 0,567 c0,753
Brown 7 9,7 6 10,5
Dark 26 36,1 17 29,8
The degree of difficulty of the veins
Easy 24 33,3 26 45,6 2,355 c0,308
Medium 20 27,8 15 26,3
Difficult 28 38,9 16 28,1
Inserting the cannula in last 1 week
Yes 33 64,7 33 71,7 0,274 d0,600
No 18 35,3 13 28,3
Table 1. Comparing The Experimental And The Control Group Success İn Terms Of Process Variables That Can Affect -2
Tablo 2. IV Catheter Opening Process in Terms of The Time and Number of Attempts
Experimental Group(n=72)
Control Group (n=57)
Z pMean±SD
(Min-Max)Mean±SD
(Min-Max)
IV catheter opening process in terms of the time (sec)
37,24±20,07 (15-83)172,65±153,21
(40-700)-8,648 b0,001
Number of attempts
1,08±0,28 (1-2) 2,23±1,57 (1-8) -5,859 b0,001
aStudent t Test bMann Whitney U Test
The number of attempts in the experimental group than the control group advanced significantly lower (p <0.01).
The experimental group, a statistically high level of processing time significantly shorter in duration than
were observed (p = 0.001).
Pain Levels
Experimental Group(n=72)
Control Group (n=57)
Z p
Mean±SD(Min-Max)
Mean±SD(Min- Max)
Children’s assesment 2,61±1,37 (0-7) 6,40±2,38 (1-10) -7,915 b0,001
Observer’s assesment 2,14±0,97 (0-5) 5,30±2,52 (1-10) -7,575 b0,001
Children in the experimental group andthe expression of children's pain levels were assessed
according to the reports of both observersand statistically determined to be at an advanced level
was significantlylower (p = 0.001).
Tablo 3. Children’s Pain Levels
bMann Whitney U Test
0102030405060708090
100
Ora
n (%
)
%90.3
%9.7
%45.6%54.4
Accuvein Group Control Group
The success of the process (+) The success of the process (-)
Tablo 4. When Compared İn Terms Of The Success İn The First Attempt
Χ2 =28,429 dYates’ Continuity Correction Test
CONCLUSİON
Peripheral intravenous catheter implementation supported by Accuvein AV-400 vascular imaging
devices in pediatric patientsincreases the success of the procedure,
reduces the processing time andthe pain during the procedure
in children
Available in the Pediatrics Unit IV Catheter Placement
THE SUGGESTİONS
THANK YOU