Report No . 917 Waters Proficiency Testing Round No. 179 ...

74
SD 9.17. C PTA wis Mrs J M Associa distribut .09 W Chrom shes to gra McGuire, Ch ates (USA) ting the sam P R Waters R mium, atefully ackn hemCentre for the sup mples. © Copyrig PO Box 750 Repo s Pro Roun - M Copp Thall Jun Ack nowledge th (Australia). ply of samp ght Profici 07 SILVER ort No oficien nd No Metals per, Ir llium, ne 20 knowledgm he technica . Also our th ples and to iency Tes RWATER N o. 917 ncy T o. 179 s - ron, L Zinc 015 ments al assistance hanks go to Global Pro sting Austr NSW 212 7 Testin 9 ead, N e provided o the Enviro oficiency Pty ralia, 2015 28, Austral ng Nicke for this pro onmental R y Ltd (Austr 5 alia el, ogram by Resource ralia) for

Transcript of Report No . 917 Waters Proficiency Testing Round No. 179 ...

SD 9.17.09

Chromium, Copper, Iron, Lead, Nickel,

PTA wishes to gratefully acknowledge the technical assistance provided for this program by Mrs J McGuire, ChemCentreAssociates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for distributing the samples.

SD 9.17.09

Waters Proficiency Testing

Chromium, Copper, Iron, Lead, Nickel,

PTA wishes to gratefully acknowledge the technical assistance provided for this program by Mrs J McGuire, ChemCentreAssociates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for distributing the samples.

PO Box 7507

Report No

Waters Proficiency Testing

Round

Chromium, Copper, Iron, Lead, Nickel,

PTA wishes to gratefully acknowledge the technical assistance provided for this program by Mrs J McGuire, ChemCentre Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for distributing the samples.

© Copyright

PO Box 7507

Report No

Waters Proficiency Testing

Round

- MetalsChromium, Copper, Iron, Lead, Nickel,

Thallium, Zinc

June 2015

Acknowledgments

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

© Copyright Proficiency Testing Australia

PO Box 7507 SILVERWATER

Report No

Waters Proficiency Testing

Round No.

MetalsChromium, Copper, Iron, Lead, Nickel,

Thallium, Zinc

June 2015

Acknowledgments

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

Proficiency Testing Australia

SILVERWATER NSW

Report No . 917

Waters Proficiency Testing

o. 179

Metals - Chromium, Copper, Iron, Lead, Nickel,

Thallium, Zinc

June 2015

Acknowledgments

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

Proficiency Testing Australia

NSW 2128

917

Waters Proficiency Testing

179

Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

Proficiency Testing Australia , 2015

2128, Australia

Waters Proficiency Testing

Chromium, Copper, Iron, Lead, Nickel,

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

2015

, Australia

Chromium, Copper, Iron, Lead, Nickel,

PTA wishes to gratefully acknowledge the technical assistance provided for this program by (Australia). Also our thanks go to the Environmental Resource

Associates (USA) for the supply of samples and to Global Proficiency Pty Ltd (Australia) for

SD 9.17.09

CONTENTS

1. Foreword ............................................................................................................................ 1

2. Program Features and Design ........................................................................................... 1

3. Statistical Format ............................................................................................................... 2

4. PTA and Technical Adviser's Comments ........................................................................... 4

5. Outlier Results ................................................................................................................. 35

6. Reference ........................................................................................................................ 35

APPENDIX A – Results and Data Analysis

Chromium .............................................................................................................................. A1

Copper ................................................................................................................................... A4

Iron ........................................................................................................................................ A7

Lead .................................................................................................................................... A10

Nickel ................................................................................................................................... A13

Thallium ............................................................................................................................... A16

Zinc ...................................................................................................................................... A18

APPENDIX B – Sample Homogeneity and Stability

Homogeneity and Stability Testing ........................................................................................ B1

APPENDIX C – Documentation

Instructions to Participants .................................................................................................... C1

Method Codes ....................................................................................................................... C3

Digestion Codes .................................................................................................................... C4

Results Sheet ........................................................................................................................ C5

1

SD 9.17.09

1. Foreword

This report summarises the results of a proficiency testing program on the determination of metals (Chromium, Copper, Iron, Lead, Nickel, Thallium and Zinc) in waters. This is round 179 in a planned series of programs involving the analysis of chemical and physical parameters of waters. The exercise was conducted in April 2015 by Proficiency Testing Australia (PTA). The main aim of the program was to assess laboratories’ abilities to competently perform the prescribed analyses. The Program Coordinator was Mrs D Mihaila and the Technical Adviser was Mrs J McGuire, ChemCentre (Australia). This report was authorised by Mrs F Watton, PTA Quality – Business Development Manager.

2. Program Features and Design 2.1 Each laboratory was randomly allocated a unique code number for the program to

ensure confidentiality of results. Reference to each laboratory in this report is by code number only. Please note that a number of laboratories reported more than one set of results and, therefore, their code numbers (with letter) could appear several times in the same data set.

2.2 Laboratories were provided with the "Instructions to Participants" and "Results Sheet"

(see Appendix C). Laboratories were requested to perform the tests according to their routine methods.

2.3 Participants were provided with a sealed vial (labelled R179) containing solutions of

metals (Chromium, Copper, Iron, Lead, Nickel, Thallium and Zinc). 2.4 A total of 47 laboratories received samples, comprising:

- 36 Australian participants; and

- 11 overseas participants, including:

- Brunei Darussalam (1), Indonesia (2), Kenya (1), Kingdom of Bahrain (1), Malaysia (3), New Zealand (1), Russia (1), Tanzania (1).

Of these 47 laboratories, two were unable to submit results by the due date. 2.5 Results (as reported by participants) with corresponding summary statistics (i.e.

number of results, median, normalised interquartile range, uncertainty of the median, robust coefficient of variation, minimum, maximum and range) are presented in Appendix A (for each of the analyses performed).

2.6 A robust statistical approach, using z-scores, was utilised to assess laboratories’

testing performance (see Section 3). Robust z-scores and ordered z-score charts relevant to each test are presented in Appendix A. The document entitled Guide to Proficiency Testing Australia, 2014 (reference [1]) defines the statistical terms and details the statistical procedures referred to in this report.

2

SD 9.17.09

2.7 A tabulated listing of laboratories (by code number) identified as having outlier results can be found on page 35.

2.8 Prior to sample distribution, a number of randomly selected samples were analysed

for homogeneity and stability. Based on the results of this testing (see Appendix B) it was considered that the samples utilised for this program were homogeneous and stable. As such, any results later identified as outliers could not be attributed to any notable sample variability.

3. Statistical Format

For each test, where appropriate, the following information is given:

- a table of results and calculated z-scores;

- a list of summary statistics; and

- ordered z-score charts.

3.1 Outlier Results and Z-scores In order to assess laboratories’ testing performance, a robust statistical approach,

using z-scores, was utilised. Z-scores give a measure of how far a result is from the consensus value (i.e. the median), and gives a "score" to each result relative to the other results in the group.

A z-score with an absolute value less than or equal to 2.0 is considered to be satisfactory, whereas, a z-score with an absolute value greater than or equal to 3.0 is considered to be an outlier and is marked by the symbol “§”. Laboratories are also encouraged to review results which have an absolute z-score value between 2.0 and 3.0 (i.e. 2.0 < |z-score| < 3.0). These results are considered to be questionable results.

Each determination was examined for outliers with all methods pooled. The table on

page 35 summarises the outlier results detected. 3.2 Results Tables and Summary Statistics The tables in Appendix A contain the results returned by each laboratory, including

the code number for the method used and the robust z-score calculated for each result.

Results have been entered exactly as reported by participants. That is, laboratories

which did not report results to the precision (i.e. number of decimal places) requested on the Results Sheet have not been rounded to the requested precision before being included in the statistical analysis.

3

SD 9.17.09

A list of summary statistics appears at the bottom of each of the results tables and consists of:

- No. of Results: the total number of results for that test/sample;

- Median: the middle value of the results;

- Normalised IQR: the normalised interquartile range of the results;

- Uncertainty of the Median: a robust estimate of the standard deviation of the Median;

- Robust CV: the robust coefficient of variation expressed as a percentage, i.e. 100 x Normalised IQR / Median;

- Minimum: the lowest laboratory result;

- Maximum: the highest laboratory result; and

- Range: the difference between the Maximum and Minimum.

The median is a measure of the centre of the data. The normalised IQR is a measure of the spread of the results. It is calculated by multiplying the interquartile range (IQR) by a correction factor, which converts the IQR to an estimate of the standard deviation. The IQR is the difference between the upper and lower quartiles (i.e. the values above and below which a quarter of the results lie, respectively). For normally distributed data, the uncertainty of the median is approximated by:

��2 ������

√�

� = number of results. Please see reference [1] for further details on these robust summary statistics.

3.3 Ordered Z-score Charts The charts in Appendix A indicate each laboratory's robust z-score, in order of

magnitude, marked with its laboratory code number. From these charts, each laboratory can readily compare its performance relative to the other laboratories.

These charts contain solid lines at +3.0 and -3.0, so that outliers are clearly

identifiable as those laboratories whose "bar" extends beyond these "cut-off" lines. The y-axis of these charts has been limited, so very large z-scores appear to extend beyond the chart boundary.

4

SD 9.17.09

4. PTA and Technical Adviser’s Comments 4.1 Metrological Traceability and Measurement Uncertainty of Assigned Values

Consensus values (median) derived from participants’ results are used in this program. These values are not metrologically traceable to an external reference.

Sample preparation was undertaken according to Environmental Resource Associates’ Standard Operating Procedures to ensure samples were fit-for-purpose, homogeneous and stable. Solutions were stable and homogeneous, and medians obtained from this proficiency round were in good agreement with the expected levels (manufacturer’s assigned values), as shown in Table 1. As the assigned value for each analyte in this program is the median of the results submitted by the participants, the uncertainty of the median for each analyte has been calculated and is presented in the table below. Table 1. Comparison of expected levels and proficiency medians. The values of the calculated uncertainty of the median are also presented.

Analyte Expected Levels (µg/L)

Median (µg/L)

Uncertainty of the median

(µg/L)

Chromium 291 291.0 3.0

Copper 712 714.5 4.6

Iron 354 357.5 5.7

Lead 548 554.5 4.4

Nickel 1190 1190.0 10.6

Thallium 289 291.0 5.0

Zinc 1240 1249.0 14.4

Overall, the performance of participants in this round was good, with robust CVs less than 10% for all analytes and comparable to those obtained in previous rounds.

5

SD 9.17.09

4.2 Analysis of Round 179 Results 4.2.1 Chromium Table 2 compares the Chromium median and robust CV from this round to those obtained in previous PTA rounds. Table 2. Comparison of current round variability and proficiency median of Chromium testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 291.0 5.2 39

Report 862 R165 854.0 4.4 38

Report 805 R152 273.0 4.6 41 Bias / Accuracy The Chromium testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 261 – 321 µg/L. Out of 39 participants, three questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 331, 382 and 428). Two outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 369 and 517. Figure 1 exhibits the spread of results and the methods used for Chromium testing in this round.

Figure 1. Spread of results for Chromium testing of sample R179, with a median of 291.0 µg/L.

0

2

4

6

8

10

12

14

16

18

195 210 225 240 255 270 285 300 315 330 345 360 375 390

Fre

quen

cy

Results (µg/L)

Chromium - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

6

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 1), with approximately 43.6% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA methods (20.6%), US EPA methods (23.1%) or other methods (12.8%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 2. Out of 39 participants, 31 (79%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 2. MU for Chromium testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 31.5 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Chromium testing was 291.0 ± 31.5 µg/L for sample R179 (95% CI). Laboratories 144, 272, 369, 382, 466 and 545 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 3 below.

0

1

2

3

4

5

6

7

8

0 10 20 30 40 50 60 70 80 90 100 110 120 130

Fre

quen

cy

Measurement Uncertainty (µg/L)

Chromium ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

7

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Chrom

ium - S

ample R

179

Figure 3. Chromium - Results of sample R179, including MU, compared to the median.

190

215

240

265

290

315

340

365

390

126

128

144

196

210

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

428

464

466

499

545

551

568

596

629

674

Re

sults

(µg/

L)

Laboratory Code

Chromium - Sample R179

Result ± MU Median Uncertainty of the Median

8

SD 9.17.09

4.2.2 Copper Table 3 compares the Copper median and robust CV from this round to those obtained in previous PTA rounds. Table 3. Comparison of current round variability and proficiency median of Copper testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 714.5 3.4 44

Report 862 R165 775.5 4.0 44

Report 805 R152 528.0 5.6 44 Bias / Accuracy The Copper testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 667 – 762 µg/L. Out of 44 participants, seven questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 331, 369, 428, 474, 517, 545 and 551). Three outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 272, 329 and 498. Figure 4 exhibits the spread of results and the methods used for Copper testing in this round.

Figure 4. Spread of results for Copper testing of sample R179, with a median of 714.5 µg/L.

0

4

8

12

16

20

510 540 570 600 630 660 690 720 750 780 810 840 870 900

Fre

quen

cy

Results (µg/L)

Copper - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6010

US EPA 6020 A

Other

<

9

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 4), with approximately 40.9% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA methods (22.7%), US EPA methods (20.5%) or other methods (15.9%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 5. Out of 44 participants, 35 (80%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 5. MU for Copper testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 56.2 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Copper testing was 714.5 ± 56.2 µg/L for sample R179 (95% CI). Laboratories 257a, 272, 329, 363, 382, 466, 498 and 545 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 6 below.

0

2

4

6

8

10

12

14

0 15 30 45 60 75 90 105 120 135 150 165 180 195

Fre

quen

cy

Measurement Uncertainty (µg/L)

Copper ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

10

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Copper - S

ample R

179

Figure 6. Copper - Results of sample R179, including MU, compared to the median. The outlier result ofLaboratory 498 is not shown.

550

600

650

700

750

800

850

900

950

126

128

144

196

210

214

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

418

428

464

466

499

545

551

568

590

596

629

674

Re

sults

(µg/

L)

Laboratory Code

Copper - Sample R179

Result ± MU Median Uncertainty of the Median

11

SD 9.17.09

4.2.3 Iron Table 4 compares the Iron median and robust CV from this round to those obtained in previous PTA rounds. Table 4. Comparison of current round variability and proficiency median of Iron testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 357.5 8.5 44

Report 862 R165 1169.0 6.2 44

Report 805 R152 1400.0 5.4 46 Bias / Accuracy The Iron testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 297 – 418 µg/L. Out of 44 participants, two questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 428 and 474). Two outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 323 and 498. Figure 7 exhibits the spread of results and the methods used for Iron testing in this round.

Figure 7. Spread of results for Iron testing of sample R179, with a median of 357.5 µg/L. The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 7), with approximately 43.2% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA protocols (22.7%), US EPA methods (18.2%) or other methods (15.9%).

0

2

4

6

8

10

12

14

190 215 240 265 290 315 340 365 390 415 440 465 490 515

Fre

quen

cy

Results (µg/L)

Iron - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

12

SD 9.17.09

Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 8. Out of 44 participants, 36 (82%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 8. MU for Iron testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 62.4 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Iron testing was 357.5 ± 62.4 µg/L for sample R179 (95% CI). Laboratories 196, 257a, 272, 323, 382, 428, 466, 498, 545 and 551 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 9 below.

0

1

2

3

4

5

6

7

8

0 10 20 30 40 50 60 70 80 90 100 110 120 130

Fre

quen

cy

Measurement Uncertainty (µg/L)

Iron ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

<

13

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Iron - Sam

ple R179

Figure 9. Iron - Results of sample R179, including MU, compared to the median.

150

200

250

300

350

400

450

500

550

126

128

138

144

196

210

214

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

418

428

464

466

498

499

545

551

568

590

596

629

674

Re

sults

(µg/

L)

Laboratory Code

Iron - Sample R179

Result ± MU Median Uncertainty of the Median

14

SD 9.17.09

4.2.4 Lead Table 5 compares the Lead median and robust CV from this round to those obtained in previous PTA rounds. Table 5. Comparison of current round variability and proficiency median of Lead testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 554.5 4.0 40

Report 862 R165 627.0 4.8 40

Report 805 R152 1470.0 5.5 43 Bias / Accuracy The Lead testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 510 – 599 µg/L. Out of 40 participants, four questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 329, 375, 382 and 418). Three outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 428, 498 and 699. Figure 10 exhibits the spread of results and the methods used for Lead testing in this round.

Figure 10. Spread of results for Lead testing of sample R179, with a median of 554.5 µg/L.

0

4

8

12

16

20

330 360 390 420 450 480 510 540 570 600 630 660 690 720

Fre

quen

cy

Results (µg/L)

Lead - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

<

15

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 10), with approximately 37.5% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA protocols (22.5%), US EPA methods (22.5%) or other methods (17.5%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 11. Out of 40 participants, 32 (80%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 11. MU for Lead testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 53.2 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Lead testing was 554.5 ± 53.2 µg/L for sample R179 (95% CI). Laboratories 257a, 375, 382, 428, 466, 498 and 545 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 12 below.

0

1

2

3

4

5

6

7

8

9

10

0 20 40 60 80 100 120 140 160 180 200 220 240 260

Fre

quen

cy

Measurement Uncertainty (µg/L)

Lead ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

16

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Lead - Sam

ple R179

Figure 12. Lead - Results of sample R179, including MU, compared to the median.

250

350

450

550

650

750

850

126

128

144

196

210

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

418

428

464

466

498

499

545

551

590

596

629

Re

sults

(µg/

L)

Laboratory Code

Lead - Sample R179

Result ± MU Median Uncertainty of the Median

17

SD 9.17.09

4.2.5 Nickel Table 6 compares the Nickel median and robust CV from this round to those obtained in previous PTA rounds. Table 6. Comparison of current round variability and proficiency median of Nickel testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 1190.0 4.5 41

Report 862 R165 1280.0 3.5 40

Report 805 R152 1360.0 6.3 43 Bias / Accuracy The Nickel testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 1082 – 1298 µg/L. Out of 41 participants, four questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 126, 274, 331 and 382). One outlier result (|z-score| ≥ 3.0) was obtained, requiring follow-up action by laboratory 272. Figure 13 exhibits the spread of results and the methods used for Nickel testing in this round.

Figure 13. Spread of results for Nickel testing of sample R179, with a median of 1190.0 µg/L.

0

2

4

6

8

10

12

14

955 1000 1045 1090 1135 1180 1225 1270 1315 1360 1405 1450 1495 1540

Fre

quen

cy

Results (µg/L)

Nickel - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

<

18

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 13), with approximately 41.5% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA protocols (22.0%), US EPA methods (21.9%) or other methods (14.6%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 14. Out of 41 participants, 32 (78%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 14. MU for Nickel testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 128.1 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Nickel testing was 1190.0 ± 128.1 µg/L for sample R179 (95% CI). Laboratories 126, 257a, 272, 274, 382, 428, 466, 545 and 629 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 15 below.

0

1

2

3

4

5

6

7

8

0 25 50 75 100 125 150 175 200 225 250 275 300 325

Fre

quen

cy

Measurement Uncertainty (µg/L)

Nickel ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

19

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Nickel - S

ample R

179

Figure 15. Nickel - Results of sample R179, including MU, compared to the median.

800

900

1000

1100

1200

1300

1400

1500

1600

1700

126

128

144

196

210

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

428

464

466

499

545

551

568

590

596

629

674

Re

sults

(µg/

L)

Laboratory Code

Nickel - Sample R179

Result ± MU Median Uncertainty of the Median

20

SD 9.17.09

4.2.6 Thallium Table 7 compares the Thallium median and robust CV from this round to those obtained in previous PTA rounds. Table 7. Comparison of current round variability and proficiency median of Thallium testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 291.0 6.9 25

Report 862 R165 773.0 6.0 27

Report 805 R152 571.5 5.8 28 Bias / Accuracy The Thallium testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 251 – 331 µg/L. Out of 25 participants, two questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 274 and 331). Three outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 272, 362 and 590. Figure 16 exhibits the spread of results and the methods used for Thallium testing in this round.

Figure 16. Spread of results for Thallium testing of sample R179, with a median of 291.0 µg/L.

0

2

4

6

8

10

12

100 125 150 175 200 225 250 275 300 325 350 375 400 425

Fre

quen

cy

Results (µg/L)

Thallium - Sample R179

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

21

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 16), with approximately 40.0% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry) and 20.0% using APHA Part 3125 B (Inductively Coupled Plasma / Mass Spectrometry). The remaining participants used US EPA methods (32.0%) or other methods (8.0%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 17. Out of 25 participants, 22 (88%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 17. MU for Thallium testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 47.2 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Thallium testing was 291.0 ± 47.2 µg/L for sample R179 (95% CI). Laboratories 196, 272, 274, 382, 545 and 590 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 18 below.

0

1

2

3

4

5

6

0 10 20 30 40 50 60 70 80 90 100 110 120 130

Fre

quen

cy

Measurement Uncertainty (µg/L)

Thallium ± MU - Sample R179

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

22

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Thallium

- Sam

ple R179

Figure 18. Thallium - Results of sample R179, including MU, compared to the median.

150

200

250

300

350

400

126

128

196

210

257

a

272

274

286

310

323

329

363

369

376

382

398

410

464

545

568

590

596

Re

sults

(µg/

L)

Laboratory Code

Thallium - Sample R179

Result ± MU Median Uncertainty of the Median

23

SD 9.17.09

4.2.7 Zinc Table 8 compares the Zinc median and robust CV from this round to those obtained in previous PTA rounds. Table 8. Comparison of current round variability and proficiency median of Zinc testing with the results of the previous two rounds.

Round Sample Median (µg/L) Robust CV (%) Participan ts

This study R179 1249.0 6.1 44

Report 862 R165 830.0 4.5 45

Report 805 R152 1304.0 6.3 47 Bias / Accuracy The Zinc testing was successfully performed, with satisfactory results (|z-score| ≤ 2.0) ranging between 1098 – 1400 µg/L. Out of 44 participants, five questionable results (2.0 < |z-score| < 3.0) were reported (laboratories 274, 369, 474, 517 and 575). Five outlier results (|z-score| ≥ 3.0) were obtained, requiring follow-up action by laboratories 210, 272, 323, 428 and 498. Figure 19 exhibits the spread of results and the methods used for Zinc testing in this round.

Figure 19. Spread of results for Zinc testing of sample R179, with a median of 1249.0 µg/L.

0

4

8

12

16

20

24

500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800

Fre

quen

cy

Results (µg/L)

Zinc - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6010

US EPA 6020 A

Other

24

SD 9.17.09

The data set formed an approximately normal distribution with no significant bias attributable to any one method. There was a large variety of methods used (Figure 19), with approximately 43.2% of participants using the method APHA 3120 B (Inductively Coupled Plasma / Atomic Emission Spectrometry). The remaining participants used other APHA protocols (20.4%), US EPA methods (20.5%) or other methods (15.9%). Measurement Uncertainty (MU) The MU reported by participants can be seen in Figure 20. Out of 44 participants, 35 (80%) submitted MU information. Many of the stated MUs did not accurately reflect the difference between the median and the participant’s result for this proficiency sample.

Figure 20. MU for Zinc testing of sample R179, as reported by participants, compared with 95% confidence interval for overall reproducibility, ± 173.9 µg/L in this round, shown as a dashed line. A t-test of all the results (outliers removed), indicated that the overall reproducibility for Zinc testing was 1249.0 ± 173.9 µg/L for sample R179 (95% CI). Laboratories 196, 210, 257a, 272, 274, 323, 369, 375, 428, 466, 498 and 545 may wish to re-examine their MU calculations, as their result was further from the median than their stated MU, as shown in Figure 21 below.

0

1

2

3

4

5

6

7

0 25 50 75 100 125 150 175 200 225 250 275 300 325

Fre

quen

cy

Measurement Uncertainty (µg/L)

Zinc ± MU - Sample R179

APHA 3111 B

APHA 3120 B

APHA 3125 B

US EPA 0200.7

US EPA 0200.8

US EPA 6020 A

Other

25

SD

9.17.09

Results of sam

ple R179, including M

U, com

pared to the m

edian

Zinc - S

ample R

179

Figure 21. Zinc - Results of sample R179, including MU, compared to the median. The outlier results ofLaboratories 428 and 498 are not shown.

800

900

1000

1100

1200

1300

1400

1500

1600

1700

1800

126

128

144

196

210

214

222

257

a

271

272

274

286

310

323

329

363

369

375

376

382

398

410

418

464

466

499

545

551

568

590

596

629

674

Re

sults

(µg/

L)

Laboratory Code

Zinc - Sample R179

Result ± MU Median Uncertainty of the Median

26

SD 9.17.09

4.3 Analysis of Results by Method Groups

Further analysis of results by method groups is undertaken to provide specific information on individual method performance. In order for methods to be grouped for analysis, PTA requires at least 11 sets of results from the same method group. For methods other than those presented below, there were less than 11 results submitted, therefore reliable conclusions cannot be drawn from analysing them separately on this occasion. The majority of the participant laboratories in this round (72.5%-92.0%) used either ICP/AES (Inductively Coupled Plasma / Atomic Emission Spectrometry) or ICP/MS (Inductively Coupled Plasma / Mass Spectrometry) techniques compared to AAS (Atomic Absorption Spectrometry) technique, which shows the advantage of multi elemental technique over single elemental. The methods used in this round suggest the preference for more recent techniques. Thus, none of the participating laboratories used the APHA Methods 3111 C-E, (method codes 2-4), APHA Electrothermal AAS (APHA 3113 B, method code 5) or US EPA 0200.9 GFAA (method code 10). The method APHA 3120 B (ICP/AES) - Method code 6, was most frequently employed for all analytes and the statistical analyses of the results from this method are outlined in table 9 below. The results for Thallium are not shown, as there were insufficient reported results to conduct an accurate statistical analysis. Table 9. Variability and proficiency medians of results obtained by method 6 in Round 179.

Analysis Method code Participants

Median ± Uncertainty of the Median

(µg/L)

Robust CV (%)

Chromium 6 17 295.0 ± 3.8 4.3

Copper 6 18 718.5 ± 3.8 1.8

Iron 6 19 354.0 ± 4.8 4.7

Lead 6 15 554.0 ± 5.6 3.1

Nickel 6 17 1199.0 ± 7.4 2.0

Zinc 6 19 1258.0 ± 16.3 4.5

27

SD 9.17.09

Seven laboratories indicated the use of other methods than those listed in the Instructions to Participants. Out of these, six laboratories specified their choice method as follows:

• laboratory 274 used the GOST 31870-2012 method for all analytes, • laboratory 418 used an in-house method based on APHA 3111 B for all

elements tested (Copper, Iron, Lead and Zinc), • laboratory 428 used GFAAS (graphite furnace atomic absorption

spectrometry) for all elements tested (Chromium, Copper, Iron, Lead, Nickel and Zinc),

• laboratory 466 used ISO 11885 for all elements tested (Chromium, Copper, Iron, Lead, Nickel and Zinc),

• laboratory 590 used an in-house ICP method for all elements tested (Copper, Iron, Lead, Nickel, Thallium and Zinc), and

• laboratory 699 used FAAS (flame atomic absorption spectrometry) for analysis of Iron and Zinc and GFAAS (graphite furnace atomic absorption spectrometry) for Chromium, Copper, Lead and Nickel.

Most of the digestion methods involved nitric acid or a combination of nitric and hydrochloric acid. One laboratory used a combination of nitric acid, hydrochloric acid and hydrogen peroxide and one laboratory used the microwave in the digestion process. It is worth noting that from the results reported, there is no notable difference between the non-digestion and various digestion methods, most of the laboratories being able to demonstrate competency in controlling the problems of contamination and loss of analyte during the process.

28

SD 9.17.09

4.3.1 Chromium Out of 39 results submitted, 21 (53.8%) were obtained by ICP/AES methods, 10 (25.6%) by ICP/MS methods, 4 (10.3%) by FAAS and 2 (5.1%) by GFAAS. Figure 22 below shows the Chromium results distribution for each technique.

Figure 22. Spread of results for Chromium testing of sample R179, by different analysing techniques. Two outlier results were obtained for Chromium in this round, one by ICP/MS and one by FAAS. There were three questionable results, obtained by ICP/AES (1), ICP/MS (1) and GFAAS (1).

0

2

4

6

8

10

12

14

16

18

195 210 225 240 255 270 285 300 315 330 345 360 375 390

Fre

quen

cy

Results (µg/L)

Chromium - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

29

SD 9.17.09

4.3.2 Copper Out of 44 results submitted, 23 (52.3%) were obtained by ICP/AES methods, 9 (20.5%) by ICP/MS methods, 7 (15.9%) by FAAS and 2 (4.5%) by GFAAS. Figure 23 below shows the Copper results distribution for each technique, suggesting a larger spread for ICP/AES methods compared to ICP/MS.

Figure 23. Spread of results for Copper testing of Sample R179, for different analysing techniques. Three outlier results were obtained for Copper in this round, one by ICP/AES, one by ICP/MS and one by FAAS, however, laboratory 498 (FAAS) reported low results for all analytes tested, indicating more a systematic error than a technique problem. There were seven questionable results, obtained by ICP/AES (3), ICP/MS (1), FAAS (2) and GFAAS (1).

0

4

8

12

16

20

510 540 570 600 630 660 690 720 750 780 810 840 870 900

Fre

quen

cy

Results (µg/L)

Copper - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

<

30

SD 9.17.09

4.3.3 Iron Out of 44 results submitted, 23 (52.3%) were obtained by ICP/AES methods, 8 (18.2%) by ICP/MS methods, 9 (20.5%) by FAAS and 1 (2.3%) by GFAAS. Figure 24 below shows the Iron results distribution for each technique, suggesting a larger spread for ICP/AES methods compared to ICP/MS.

Figure 24. Spread of results for Iron testing of Sample R179, by different analysing techniques. Two outlier results were obtained for Iron in this round, one by ICP/AES and one by FAAS (systematic error). There were two questionable results in this round, obtained by ICP/AES (1) and GFAAS (1), while all results obtained by ICP/MS methods were satisfactory.

0

2

4

6

8

10

12

14

190 215 240 265 290 315 340 365 390 415 440 465 490 515

Fre

quen

cy

Results (µg/L)

Iron - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

31

SD 9.17.09

4.3.4 Lead Out of 40 results submitted, 19 (47.5%) were obtained by ICP/AES methods, 10 (25.0%) by ICP/MS methods, 6 (15.0%) by FAAS and 2 (5.0%) by GFAAS. Figure 25 below shows the Lead results distribution for each technique.

Figure 25. Spread of results for Lead testing of Sample R179, by different analysing techniques. Three outlier results were obtained for Lead in this round, one by FAAS (systematic error) and two by GFAAS. There were four questionable results in this round, obtained by ICP/AES (1), ICP/MS (2) and FAAS (1).

0

4

8

12

16

20

330 360 390 420 450 480 510 540 570 600 630 660 690 720

Fre

quen

cy

Results (µg/L)

Lead - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

<

32

SD 9.17.09

4.3.5 Nickel Out of 41 results submitted, 21 (51.2%) were obtained by ICP/AES methods, 10 (24.4%) by ICP/MS methods, 5 (12.2%) by FAAS and 2 (4.9%) by GFAAS. Figure 26 below shows the Nickel results distribution for each technique, suggesting a larger spread for ICP/AES methods compared to ICP/MS.

Figure 26. Spread of results for Nickel testing of Sample R179, by different analysing techniques. One outlier result was obtained for Nickel in this round, by ICP/AES. There were four questionable results, obtained by ICP/AES (2), ICP/MS (1) and other methods (1).

0

2

4

6

8

10

12

14

955 1000 1045 1090 1135 1180 1225 1270 1315 1360 1405 1450 1495 1540

Fre

quen

cy

Results (µg/L)

Nickel - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

<

33

SD 9.17.09

4.3.6 Thallium Out of 25 results submitted, 12 (48.0%) were obtained by ICP/AES methods and 11 (44.0%) by ICP/MS methods. Figure 27 below shows the Thallium results distribution for each technique.

Figure 27. Spread of results for Thallium testing of Sample R179, for different analysing techniques. Three outlier results were obtained for Thallium in this round, two by ICP/AES and one by other methods. There were two questionable results in this round, obtained by ICP/AES (1) and other methods (1), while all results obtained by ICP/MS methods were satisfactory.

0

2

4

6

8

10

12

100 125 150 175 200 225 250 275 300 325 350 375 400 425

Fre

quen

cy

Results (µg/L)

Thallium - Sample R179

ICP/AES

ICP/MS

Other

34

SD 9.17.09

4.3.7 Zinc Out of 44 results submitted, 24 (54.5%) were obtained by ICP/AES methods, 8 (18.2%) by ICP/MS methods, 8 (18.2%) by FAAS and 1 (2.3%) by GFAAS. Figure 28 below shows the Zinc results distribution for each technique.

Figure 28. Spread of results for Zinc testing of Sample R179, by different analysing techniques. Five outlier results were obtained for Zinc in this round, two by ICP/AES, one by ICP/MS, one by FAAS (systematic error) and one by GFAAS. There were five questionable results reported, obtained by ICP/AES (2), ICP/MS (1), FAAS (1) and other methods (1). Overall, the results of Round 179 suggest a better performance of ICP/MS methods compared to ICP/AES. The ICP/MS methods produced three outlier results (one for Chromium, one for Copper and one for Zinc) and six questionable results (one for Chromium, one for Copper, two for Lead, one for Nickel and one for Zinc). For comparison, the ICP/AES gave 7 outliers and 11 questionable results. Should be noted however that, except for Thallium and Lead, there were double or more results obtained by ICP/AES than by ICP/MS methods.

0

4

8

12

16

20

24

500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800

Fre

quen

cy

Results (µg/L)

Zinc - Sample R179

FAAS

GFAAS

ICP/AES

ICP/MS

Other

35

SD 9.17.09

5. Outlier Results

Laboratories reporting results that have been identified as outliers are listed in Table 10 below. Table 10. Laboratory results identified as outliers for each analysis performed.

Lab Code

Analysis

Chromium Copper Iron Lead Nickel Thallium Zinc

210 § 272 § § § § 323 § § 329 § 362 § 369 § 428 § § 498 § § § § 517 § 590 § 699 §

Note:

A “§” indicates the occurrence of a z-score outlier result (i.e. those results for which |z-score| ≥ 3.0).

6. Reference

[1] Guide to Proficiency Testing Australia, 2014 (This document can be found on the PTA website, www.pta.asn.au)

SD 9.17.09

APPENDIX A

Results and Data Analysis Chromium .............................................................................................................................. A1

Copper ................................................................................................................................... A4

Iron ........................................................................................................................................ A7

Lead .................................................................................................................................... A10

Nickel ................................................................................................................................... A13

Thallium ............................................................................................................................... A16

Zinc ...................................................................................................................................... A18

SD 9.17.09

Chromium Results

Sample R179

A1

SD 9.17.09

Chromium (Cr) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 317 ± 34 1.71 6 14

128 306 ± 34 0.99 6 #

144 265 ± 7 -1.71 1 15

196 286 ± 10 -0.33 8 14

210 301 ± 50 0.66 7 14

222 299 ± 20 0.53 6 14

257a 303 ± 12.0 0.79 6 #

271 277 ± 33 -0.92 1 #

272 314 ± 15 1.51 6 #

274 298 ± 23.5 0.46 13 14

286 270 ± 54 -1.38 12 22

310 291 ± 9 0.00 8 #

323 291 ± 93 0.00 6 #

329 294 ± 29 0.20 12 22

331 322 # 2.04 8 14

342 295 # 0.26 6 15

362 286 # -0.33 6 14

363 295 ± 14 0.26 6 14

369 236 ± 35 -3.62 § 9 #

375 285 ± 14 -0.39 6 14

376 298 ± 30 0.46 7 #

382 258 ± 5 -2.17 7 14

398 290 ± 44 -0.07 12 22

410 284 ± 28 -0.46 7 #

428 325 ± 35 2.24 13 24

464 291 ± 2 0.00 6 #

466 301 ± 3 0.66 13 15

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A2

SD 9.17.09

Chromium (Cr) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

474 268 # -1.51 6 15

499 295 ± 21 0.26 6 24

517 377 # 5.66 § 1 14

545 272 ± 6 -1.25 6 20

551 275 ± 10% -1.05 1 #

552 311 # 1.32 6 #

568 279 ± 45 -0.79 6 14

575 280 # -0.72 12 14

596 280 ± 46 -0.72 12 24

629 287 ± 13 -0.26 13 24

674 293 ± 20 0.13 6 #

699 300 # 0.59 13 # No of Results: 39

Median: 291.0

Normalised IQR: 15.2

Uncertainty of the Median:

3.0

Robust CV: 5.2%

Minimum: 236

Maximum: 377

Range: 141

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Chromium - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Chrom

ium - S

ample R

179

A3

Robust Z-Scores

369

382 14

4

474

286

545

551

271

568

575

596

410

375

196

362

629

398

310

323

464

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Copper Results

Sample R179

A4

SD 9.17.09

Copper (Cu) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 696 ± 64 -0.77 6 14

128 730 ± 66 0.64 6 #

144 730 ± 18 0.64 1 15

196 739 ± 25 1.02 8 14

210 702 ± 80 -0.52 7 14

214 729 ± 73 0.60 6 #

222 721 ± 20 0.27 6 14

257a 734 ± 17.2 0.81 6 #

271 675 ± 68 -1.64 1 #

272 896 ± 30 7.53 § 6 #

274 719 ± 36.3 0.19 13 14

286 696 ± 139 -0.77 12 22

310 701 ± 21 -0.56 8 #

323 722 ± 74 0.31 6 #

329 788 ± 62 3.05 § 12 22

331 767 # 2.18 8 14

342 710 # -0.19 6 15

362 704 # -0.44 6 14

363 730 ± 10 0.64 6 14

369 770 ± 108 2.30 9 #

375 710 ± 36 -0.19 6 14

376 719 ± 72 0.19 7 #

382 671 ± 5 -1.81 7 14

398 706 ± 106 -0.35 12 22

410 698 ± 70 -0.68 7 #

418 721 ± 51 0.27 13 14

428 773 ± 93 2.43 13 24

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A5

SD 9.17.09

Copper (Cu) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

464 714 ± 7 -0.02 6 #

466 751 ± 8 1.52 13 15

474 655 # -2.47 6 15

498 367 ± 46 -14.42 § 1 14

499 723 ± 43 0.35 6 24

517 666 # -2.01 1 14

545 652 ± 8 -2.59 6 20

551 666 ± 11.2% -2.01 1 #

552 716 # 0.06 6 #

568 710 ± 68 -0.19 6 14

575 696 # -0.77 11 14

590 708 ± 20 -0.27 13 24

596 697 ± 130 -0.73 12 24

629 723 ± 68 0.35 13 24

674 722 ± 70 0.31 6 #

688 715 # 0.02 1 14

699 690 # -1.02 13 # No of Results: 44

Median: 714.5

Normalised IQR: 24.1

Uncertainty of the Median:

4.6

Robust CV: 3.4%

Minimum: 367

Maximum: 896

Range: 529

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Copper - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Copper - S

ample R

179

A6

Robust Z-Scores

498

545

474 51

7

551

382

271 69

9

126

286

575

596

410

310

210

362

398

590

342

375

568

464

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Iron Results

Sample R179

A7

SD 9.17.09

Iron (Fe) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 359 ± 38 0.05 6 14

128 367 ± 76 0.31 6 #

138 328 ± 30 -0.97 1 14

144 357 ± 7 -0.02 1 15

196 396 ± 17 1.27 8 14

210 330 ± 50 -0.90 7 14

214 344 ± 34 -0.44 6 #

222 358 ± 20 0.02 6 14

257a 413 ± 16.8 1.83 6 #

271 325 ± 45 -1.07 1 #

272 408 ± 20 1.66 6 #

274 367 ± 49.7 0.31 13 14

286 338 ± 68 -0.64 12 22

310 348 ± 10 -0.31 8 #

323 469 ± 57 3.67 § 6 #

329 366 ± 61 0.28 12 22

331 397 # 1.30 8 14

342 340 # -0.58 6 15

362 369 # 0.38 6 14

363 350 ± 9 -0.25 6 14

369 330 ± 50 -0.90 9 #

375 344 ± 17 -0.44 6 14

376 367 ± 37 0.31 7 #

382 320 ± 5 -1.23 6 14

398 379 ± 57 0.71 12 22

410 372 ± 37 0.48 7 #

418 390 ± 131 1.07 13 14

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A8

SD 9.17.09

Iron (Fe) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

428 420 ± 24 2.06 13 24

464 351 ± 7 -0.21 6 #

466 388 ± 4 1.00 13 15

474 267 # -2.98 6 15

498 211 ± 51 -4.82 § 1 14

499 362 ± 22 0.15 6 24

517 394 # 1.20 1 14

545 327 ± 9 -1.00 6 20

551 297 ± 16.5% -1.99 1 #

552 354 # -0.12 6 #

568 334 ± 51 -0.77 6 14

575 <500 # na 11 14

590 344 ± 20 -0.44 13 24

596 369 ± 75 0.38 12 24

629 345 ± 16 -0.41 13 24

674 354 ± 30 -0.12 6 #

688 396 # 1.27 1 14

699 385 # 0.90 13 # No of Results: 44

Median: 357.5

Normalised IQR: 30.4

Uncertainty of the Median:

5.7

Robust CV: 8.5%

Minimum: 211

Maximum: 469

Range: 258

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Iron - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Iron - Sam

ple R179

A9

Robust Z-Scores

498

474

551 38

2

271

545

138

210

369

568

286

342

214

375

590

629

310

363

464

552

674

144

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Lead Results

Sample R179

A10

SD 9.17.09

Lead (Pb) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 531 ± 78 -1.05 6 14

128 572 ± 95 0.78 6 #

144 562 ± 21 0.33 1 14

196 557 ± 23 0.11 8 14

210 561 ± 75 0.29 7 14

222 559 ± 20 0.20 6 14

257a 591 ± 15.2 1.63 6 #

271 597 ± 67 1.90 1 #

272 545 ± 20 -0.42 6 #

274 581 ± 51.2 1.18 13 14

286 519 ± 130 -1.58 12 22

310 542 ± 16 -0.56 8 #

323 554 ± 60 -0.02 6 #

329 600 ± 89 2.03 12 22

331 557 # 0.11 8 14

342 550 # -0.20 6 15

362 564 # 0.42 6 14

363 548 ± 13 -0.29 6 14

369 518 ± 98 -1.63 9 #

375 505 ± 25 -2.21 6 14

376 565 ± 57 0.47 7 #

382 492 ± 5 -2.79 7 14

398 529 ± 79 -1.14 12 22

410 552 ± 55 -0.11 7 #

418 610 ± 214 2.47 13 14

428 799 ± 45 10.90 § 13 24

464 556 ± 17 0.07 6 #

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A11

SD 9.17.09

Lead (Pb) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

466 591 ± 5 1.63 13 15

474 516 # -1.72 6 15

498 341 ± 39 -9.52 § 1 14

499 559 ± 56 0.20 6 24

517 535 # -0.87 1 14

545 529 ± 9 -1.14 6 20

551 540 ± 4.7% -0.65 1 #

552 566 # 0.51 6 #

575 560 # 0.25 12 14

590 538 ± 30 -0.74 13 24

596 555 ± 92 0.02 12 24

629 537 ± 38 -0.78 13 24

699 375 # -8.00 § 13 # No of Results: 40

Median: 554.5

Normalised IQR: 22.4

Uncertainty of the Median:

4.4

Robust CV: 4.0%

Minimum: 341

Maximum: 799

Range: 458

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Lead - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Lead - Sam

ple R179

A12

Robust Z-Scores

498

699

382 37

5 474

369

286 39

8

545

126

517

629

590

551

310

272

363

342

410

323

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Nickel Results

Sample R179

A13

SD 9.17.09

Nickel (Ni) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 1320 ± 84 2.40 6 14

128 1213 ± 189 0.43 6 #

144 1230 ± 49 0.74 1 15

196 1180 ± 50 -0.18 8 14

210 1120 ± 100 -1.29 7 14

222 1199 ± 50 0.17 6 14

257a 1245 ± 14.8 1.02 6 #

271 1230 ± 280 0.74 1 #

272 1610 ± 30 7.76 § 6 #

274 1342 ± 74.9 2.81 13 14

286 1148 ± 230 -0.78 12 22

310 1172 ± 41 -0.33 8 #

323 1193 ± 54 0.06 6 #

329 1240 ± 157 0.92 12 22

331 1313 # 2.27 8 14

342 1210 # 0.37 6 15

362 1207 # 0.31 6 14

363 1180 ± 11 -0.18 6 14

369 1185 ± 166 -0.09 9 #

375 1190 ± 60 0.00 6 14

376 1160 ± 116 -0.55 7 #

382 1072 ± 10 -2.18 7 14

398 1181 ± 177 -0.17 12 22

410 1190 ± 120 0.00 7 #

428 1096 ± 47 -1.74 13 24

464 1193 ± 20 0.06 6 #

466 1270 ± 14 1.48 13 15

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A14

SD 9.17.09

Nickel (Ni) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

474 1105 # -1.57 6 15

499 1200 ± 96 0.18 6 24

517 1142 # -0.89 1 14

545 1123 ± 23 -1.24 6 20

551 1088 ± 9.8% -1.88 1 #

552 1243 # 0.98 6 #

568 1100 ± 200 -1.66 6 14

575 1200 # 0.18 12 14

590 1159 ± 200 -0.57 13 24

596 1153 ± 205 -0.68 12 24

629 1100 ± 34 -1.66 13 24

674 1159 ± 75 -0.57 6 #

688 1221 # 0.57 1 14

699 1105 # -1.57 13 # No of Results: 41

Median: 1190.0

Normalised IQR: 54.1

Uncertainty of the Median:

10.6

Robust CV: 4.5%

Minimum: 1072

Maximum: 1610

Range: 538

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Nickel - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Nickel - S

ample R

179

A15

Robust Z-Scores

382 55

1

428

568

629

474

699

210

545 51

7

286

596

590

674

376

310

196

363

398

369

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Thallium Results

Sample R179

A16

SD 9.17.09

Thallium (Tl) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 312 ± 43 1.05 6 14 128 296 ± 14 0.25 6 # 196 263 ± 10 -1.40 8 14 210 298 ± 50 0.35 7 14 257a 308 ± 17.2 0.85 6 # 272 352 ± 25 3.05 § 6 # 274 344 ± 44.6 2.65 13 14 286 283 ± 57 -0.40 12 22 310 299 ± 12 0.40 9 # 323 319 ± 63 1.40 6 # 329 301 ± 35 0.50 12 22 331 345 # 2.70 8 14 362 190 # -5.05 § 6 14 363 278 ± 16 -0.65 6 14 369 275 ± 47 -0.80 9 # 376 297 ± 30 0.30 7 # 382 266 ± 5 -1.25 7 14 398 268 ± 40 -1.15 12 22 410 290 ± 29 -0.05 7 # 464 291 ± 5 0.00 6 # 545 265 ± 6 -1.30 6 20 552 286 # -0.25 6 # 568 295 ± 42 0.20 7 14 590 171 ± 10 -6.00 § 13 24 596 274 ± 55 -0.85 12 24 No of Results: 25

Median: 291.0

Normalised IQR: 20.0 Uncertainty of

the Median: 5.0

Robust CV: 6.9%

Minimum: 171

Maximum: 352

Range: 181 1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Thallium - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Chart

Thallium

- Sam

ple R179

A17

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

Zinc Results

Sample R179

A18

SD 9.17.09

Zinc (Zn) Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

126 1220 ± 110 -0.38 6 14

128 1265 ± 177 0.21 6 #

144 1250 ± 53 0.01 1 15

196 1360 ± 72 1.46 8 14

210 1650 ± 100 5.28 § 7 14

214 1347 ± 135 1.29 6 #

222 1248 ± 50 -0.01 6 14

257a 1321 ± 17.2 0.95 6 #

271 1180 ± 118 -0.91 1 #

272 1720 ± 40 6.20 § 6 #

274 1421 ± 86.9 2.26 13 14

286 1120 ± 280 -1.70 12 22

310 1215 ± 49 -0.45 8 #

323 1537 ± 89 3.79 § 6 #

329 1220 ± 129 -0.38 12 22

331 1353 # 1.37 8 14

342 1300 # 0.67 6 15

362 1258 # 0.12 6 14

363 1253 ± 9 0.05 6 14

369 1093 ± 109 -2.05 9 #

375 1390 ± 70 1.86 6 14

376 1277 ± 128 0.37 7 #

382 1253 ± 10 0.05 6 14

398 1228 ± 184 -0.28 12 22

410 1212 ± 120 -0.49 7 #

418 1290 ± 177 0.54 13 14

428 536 ± 26 -9.38 § 13 24

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

A19

SD 9.17.09

Zinc (Zn) cont. Results by Laboratory Code

Sample R179

Lab Code Result ± MU1

(µg/L) Robust z-score2

Method Code3

Digestion Code3

464 1254 ± 12 0.07 6 #

466 1375 ± 26 1.66 13 15

474 1074 # -2.30 6 15

498 644 ± 34 -7.96 § 1 14

499 1265 ± 114 0.21 6 24

517 1090 # -2.09 1 14

545 1160 ± 61 -1.17 6 20

551 1157 ± 8.2% -1.21 1 #

552 1294 # 0.59 6 #

568 1120 ± 210 -1.70 6 14

575 1090 # -2.09 11 14

590 1241 ± 200 -0.11 13 24

596 1207 ± 255 -0.55 12 24

629 1220 ± 113 -0.38 13 24

674 1219 ± 90 -0.39 6 #

688 1232 # -0.22 1 14

699 1311 # 0.82 13 # No of Results: 44

Median: 1249.0

Normalised IQR: 76.0

Uncertainty of the Median:

14.4

Robust CV: 6.1%

Minimum: 536

Maximum: 1720

Range: 1184

1 Where reported, results are shown with their corresponding measurement

uncertainty (MU). 2 "§" denotes an outlier (i.e. those results for which |z-score| ≥ 3.0). Robust z-

scores are calculated as: z = (A - median) ÷ normalised IQR, where A is the participant laboratory's result.

3 Please refer to Appendix C (page C3-C4) for method and digestion code descriptions.

SD 9.17.09

Zinc - Sample R179 - Robust Z-Scores

Ordered R

obust Z-S

core Charts

Zinc - S

ample R

179

A20

Robust Z-Scores

428

498

474

517

575

369 28

6

568 55

1

545

271 59

6

410

310

674

126

329

629

398

688

590

222

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

-5

-4

-3

-2

-1

0

1

2

3

4

5

z-sc

ore

lab code

SD 9.17.09

APPENDIX B

Sample Homogeneity and Stability Homogeneity and Stability Testing ........................................................................................ B1

B1

SD 9.17.09

Homogeneity and Stability Testing Certified reference samples for this program were obtained from Environmental Resource Associates (ERA, USA). As such, all samples are subjected to rigorous stability and homogeneity testing. On the basis of this testing, the samples utilised for this program were considered to be homogeneous and stable. Table B1 below presents the certification data and analytical verification results prior to sample release1. Table B1. Certified values and analytical verification results for samples used in Round 179.

Analytical Verification

Analyte Certified Value 2

Uncertainty 3 (%) Mean Recovery

(%) No. of

samples

Chromium 291 4.90 292 100 41

Copper 712 1.86 715 100 48

Iron 354 6.26 354 100 39

Lead 548 5.72 545 99.5 45

Nickel 1190 5.88 1190 99.6 45

Thallium 289 4.31 286 99.0 18

Zinc 1240 0.462 1230 99.2 44

1 ERA certification and analytical verification data issued 09 October 2014. 2 The Certified Values are the actual “made-to” concentrations confirmed by ERA analytical verification. 3 The stated Uncertainty is the total propagated uncertainty at the 95% confidence interval. The uncertainty is based on the preparation and analytical verification of the product by ERA, multiplied by a coverage factor. The uncertainty applies to the product as supplied and does not take into account any required or optional dilution and/or preparations the laboratory may perform while using this product.

SD 9.17.09

APPENDIX C

Documentation Instructions to Participants .................................................................................................... C1

Method Codes ....................................................................................................................... C3

Digestion Codes .................................................................................................................... C4

Results Sheet ........................................................................................................................ C5

SD 9.17.09

CHEMICAL ANALYSIS ROUND 179

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples. 1. Sample Information

i)

ii) iii) iv) v)

vi)

Please Note:laboratory sample.

2. Sample Preparation

i) ii)

Please read the Caution notes below before this step

iii)

iv) v) vi) vii) viii)ix)

Caution:

� Caution must be taken when analysing corrosive samples.� While technically it is unnecessary to digest the sample prior to analysis,

performed if this is your normal procedure.

� A volume of acid different from the 2order to matrix match calibration standards or meet any other method criteria.

� If analysing usito analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2nitric acid suggested in

SD 9.17.09

CHEMICAL ANALYSIS ROUND 179Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

Sample Information

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The vial contains approximately 14mL of

Please note the temperature of vial on receipt.

The sample has been acidified with approximately 2% (v/v) nitric acid. The sample must be thoroughly mixed prior to analysis.

The vial will require dilution in deionised water below).

The vial may be stored at room temperature.

Please Note: Where possible, the proficiency testing sample should be treated as a routine laboratory sample.

Sample Preparation

Analysis must begin immediately after vial is opened

Prepare sample at 20before opening.

Please read the Caution notes below before this step

Add approximately 100concentrated ni

Mix the vial prior to opening.

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask. Dilute the flask to the final volume with deionised water.

Stopper and mix by inversion.

viii) Immediately Report your results as µg/L for the diluted sample.

Caution: Caution must be taken when analysing corrosive samples.While technically it is unnecessary to digest the sample prior to analysis, performed if this is your normal procedure.

A volume of acid different from the 2order to matrix match calibration standards or meet any other method criteria.If analysing usito analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2nitric acid suggested in

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY

CHEMICAL ANALYSIS ROUND 179Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

INSTRUCTIONS TO PARTICIPANTS

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

Sample Information

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The vial contains approximately 14mL of

Please note the temperature of vial on receipt.

The sample has been acidified with approximately 2% (v/v) nitric acid.The sample must be thoroughly mixed prior to analysis.

The vial will require dilution in deionised water

The vial may be stored at room temperature.

Where possible, the proficiency testing sample should be treated as a routine laboratory sample.

Sample Preparation

Analysis must begin immediately after vial is opened

Prepare sample at 20-22ºC, and allow sufficient time for the vial to reach room temperature before opening.

Please read the Caution notes below before this step

Add approximately 100concentrated nitric acid to a 500mL volumetric flask.

Mix the vial prior to opening.

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.Dilute the flask to the final volume with deionised water.

Stopper and mix by inversion.

Immediately analyse the diluted sample by your normal procedures.Report your results as µg/L for the diluted sample.

Caution must be taken when analysing corrosive samples.While technically it is unnecessary to digest the sample prior to analysis, performed if this is your normal procedure.

A volume of acid different from the 2order to matrix match calibration standards or meet any other method criteria.If analysing using colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2nitric acid suggested in step (iii)

PROFICIENCY TESTING AUSTRALIAWATERS PROFICIENCY

CHEMICAL ANALYSIS ROUND 179 Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

INSTRUCTIONS TO PARTICIPANTS

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The vial contains approximately 14mL of

Please note the temperature of vial on receipt.

The sample has been acidified with approximately 2% (v/v) nitric acid.The sample must be thoroughly mixed prior to analysis.

The vial will require dilution in deionised water

The vial may be stored at room temperature.

Where possible, the proficiency testing sample should be treated as a routine

Analysis must begin immediately after vial is opened

22ºC, and allow sufficient time for the vial to reach room temperature

Please read the Caution notes below before this step

Add approximately 100-200mL deionised water followed by 2tric acid to a 500mL volumetric flask.

Mix the vial prior to opening.

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.Dilute the flask to the final volume with deionised water.

Stopper and mix by inversion.

analyse the diluted sample by your normal procedures.Report your results as µg/L for the diluted sample.

Caution must be taken when analysing corrosive samples.While technically it is unnecessary to digest the sample prior to analysis, performed if this is your normal procedure.

A volume of acid different from the 2-order to matrix match calibration standards or meet any other method criteria.

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2

step (iii) .

C1

PROFICIENCY TESTING AUSTRALIAWATERS PROFICIENCY TESTING PROGRAM

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

INSTRUCTIONS TO PARTICIPANTS

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The vial contains approximately 14mL of artificial waste water concentrate.

Please note the temperature of vial on receipt.

The sample has been acidified with approximately 2% (v/v) nitric acid.The sample must be thoroughly mixed prior to analysis.

The vial will require dilution in deionised water (please follow the

The vial may be stored at room temperature.

Where possible, the proficiency testing sample should be treated as a routine

Analysis must begin immediately after vial is opened

22ºC, and allow sufficient time for the vial to reach room temperature

Please read the Caution notes below before this step

200mL deionised water followed by 2tric acid to a 500mL volumetric flask.

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.Dilute the flask to the final volume with deionised water.

analyse the diluted sample by your normal procedures.Report your results as µg/L for the diluted sample.

Caution must be taken when analysing corrosive samples.While technically it is unnecessary to digest the sample prior to analysis, performed if this is your normal procedure.

-5mL of nitric acidorder to matrix match calibration standards or meet any other method criteria.

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2

PROFICIENCY TESTING AUSTRALIATESTING PROGRAM

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

INSTRUCTIONS TO PARTICIPANTS

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The artificial waste water concentrate.

The sample has been acidified with approximately 2% (v/v) nitric acid.The sample must be thoroughly mixed prior to analysis.

(please follow the

Where possible, the proficiency testing sample should be treated as a routine

Analysis must begin immediately after vial is opened.

22ºC, and allow sufficient time for the vial to reach room temperature

Please read the Caution notes below before this step

200mL deionised water followed by 2tric acid to a 500mL volumetric flask.

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.Dilute the flask to the final volume with deionised water.

analyse the diluted sample by your normal procedures.Report your results as µg/L for the diluted sample.

Caution must be taken when analysing corrosive samples.While technically it is unnecessary to digest the sample prior to analysis,

nitric acid suggested in order to matrix match calibration standards or meet any other method criteria.

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2

PROFICIENCY TESTING AUSTRALIA TESTING PROGRAM

INSTRUCTIONS TO PARTICIPANTS

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

Please note the following before commencing the analysis of the samples.

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The artificial waste water concentrate.

The sample has been acidified with approximately 2% (v/v) nitric acid.

(please follow the Sample Prepar

Where possible, the proficiency testing sample should be treated as a routine

22ºC, and allow sufficient time for the vial to reach room temperature

200mL deionised water followed by 2-5mL of high purity,

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.

analyse the diluted sample by your normal procedures.

Caution must be taken when analysing corrosive samples. While technically it is unnecessary to digest the sample prior to analysis,

suggested in step (iii)order to matrix match calibration standards or meet any other method criteria.

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2

APRIL, 2015

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The

Sample Prepar ation

Where possible, the proficiency testing sample should be treated as a routine

22ºC, and allow sufficient time for the vial to reach room temperature

5mL of high purity,

Using a dry, clean volumetric pipette, transfer 5.0mL from the vial into the flask.

While technically it is unnecessary to digest the sample prior to analysis, digestion should be

step (iii) may be added in order to matrix match calibration standards or meet any other method criteria.

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2

APRIL, 2015

**Please record (on the Results Sheet) the approximat e temperature of the samples upon receipt**

One sealed vial labelled R179 supplied by Environmental Resource Associates (ERA). The

ation steps

Where possible, the proficiency testing sample should be treated as a routine

22ºC, and allow sufficient time for the vial to reach room temperature

5mL of high purity,

digestion should be

may be added in

ng colorimetric techniques, it may be necessary to pH adjust the sample prior to analysis. If using colorimetric techniques, it is acceptable to omit the addition of the 2-5mL

C2

SD 9.17.09

3. Tests Requested

i) Chromium (Cr) ii) Copper (Cu) iii) Iron (Fe) iv) Lead (Pb) v) Nickel (Ni) vi) Thallium (Tl) vii) Zinc (Zn)

(It is recommended that a reagent water blank is analysed by the same method used to analyse the samples.)

If unable to perform the above please note this on your Results Sheet.

4. Safety

i) Samples are for laboratory use only. ii) Participants should have sufficient experience and training to take the necessary precautions

when handling the samples and reagent chemicals and during disposal. iii) Use of safety glasses, gloves, and fume hoods, where appropriate during the

determinations, is recommended. 5. Reporting

i) Report results to the nearest whole number (no decimal places). ii) Report results in micrograms per litre (µg/L ). iii) Do not correct results for recovery. iv) Select the appropriate method code for each test from the Method Code Table and record it

on the Results Sheet. v) Calculate the measurement uncertainty (MU) for each reported result. All estimates of MU

must be given as a 95% confidence interval (coverage factor k ≈ 2) and reported in micrograms per litre (µg/L ). Report to the nearest whole number (no decimal places).

6. Testing should commence as soon as possible after receiving the samples and results reported

NO LATER THAN 1 MAY 2015 to:

Delfina Mihaila Proficiency Testing Australia PO Box 7507 SILVERWATER NSW 2128 AUSTRALIA Phone: +612 9736 8397 Fax: +612 9743 6664 Email: [email protected]

7. For this program your laboratory has been allocated the code number shown on the attached Results Sheet. All reference to your laboratory in reports associated with the program will be through this code number, thus ensuring the confidentiality of your results.

8. The expected concentration ranges of each of tests are given in the following table.

Analyte Range (µg/L) Chromium (Cr) 17 – 1000 Copper (Cu) 40 – 900 Iron (Fe) 200 – 4000 Lead (Pb) 70 – 3000 Nickel (Ni) 80 – 3000 Thallium (Tl) 60 – 900 Zinc (Zn) 100 – 2000

C3

SD 9.17.09

Method Codes to be used for the Results Sheet

ANALYSIS METHOD REFERENCE METHOD DESCRIPTION CODE

Chromium (Cr)

Copper (Cu)

Iron (Fe)

Lead (Pb)

Nickel (Ni)

Thallium (Tl)

Zinc (Zn)

APHA SM

APHA 3111 B. Direct Air-Acetylene Flame Method 1

APHA 3111 C. Extraction/Air-Acetylene Flame Method 2

APHA 3111 D. Direct Nitrous Oxide-Acetylene Flame Method 3

APHA 3111 E. Extraction/Nitrous Oxide-Acetylene Flame Method

4

APHA 3113 B. Electrothermal Atomic Absorption Spectrometric Method

5

APHA 3120 B. METALS BY PLASMA EMISSION SPECTROSCOPY Inductively Coupled Plasma (ICP) Method

6

APHA 3125 B. Inductively-Coupled Plasma/Mass Spectrometry (ICP/MS) Method

7

US EPA

US EPA 0200.7 Metals and Trace Elements - ICP/AES 8

US EPA 0200.8 Trace Elements in Water & Wastes - ICP/MS 9

US EPA 0200.9 Trace Elements - GFAA 10

US EPA 6010 Inductively Coupled Plasma-Atomic Emission Spectrometry 11

US EPA 6020 A Inductively Coupled Plasma/MS 12

Other Other (please specify) 13

C4

SD 9.17.09

Digestion Codes to be used for the Results Sheet

DIGESTION PROCEDURE CODE

APHA

APHA 3030 E (HNO3) 14

APHA 3030 F (HNO3/HCl) 15

APHA 3030 G (HNO3/H2SO4) 16

APHA 3030 H (HNO3/HClO4) 17

APHA 3030 I (HNO3/HClO4/HF) 18

APHA 3030 J (Dry Ashing) 19

APHA 3030 K (Microwave Assisted) 20

US EPA

US EPA 3005 A (HNO3/HCl) 21

US EPA 3010 A (HNO3/HCl) 22

US EPA 3015 A (Microwave Assisted) 23

Other (please specify) 24

Method Reference Key

i) APHA SM APHA “Standard Methods for the Examination of Water and Wastewater” (18,

19, 20, 21 and 22 Edition).

ii) USEPA U.S Environmental Protection Agency, http://www.epa.gov/osa/fem/methcollectns.htm.

SD 9.17.09

*Approximate temperature of samples upon receipt:

Chromium

Copper (Cu)

Iron (Fe)

Lead (Pb)

Nickel (Ni)

Thallium (Tl)

Zinc (Zn)

Please note:

i) For ii) Report results for the diluted sample.

iii) Report results to the nearest whole number

iv) Report results in micrograms per litre (v) Do not correct results for recovery.

vi) MU* Laboratmicrograms per litre (

DATE:

INSTRUCT WATERS PROF TEST PROG 179

SD 9.17.09

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

*Approximate temperature of samples upon receipt:

ANALYSIS

Chromium (Cr)

Copper (Cu)

Iron (Fe)

Lead (Pb)

Nickel (Ni)

Thallium (Tl)

Zinc (Zn)

Please note: Where possible, the proficiency testing sample should be treated as a routine laboratory sample.

For each analyteReport results for the diluted sample.

Report results to the nearest whole number

Report results in micrograms per litre (Do not correct results for recovery.

MU* Laboratmicrograms per litre (

DATE: ______________________

Return results Delfina MihailaProficiency Testing AustraliaPO Box 7507 SILVERWATER NSW 2128 AUSTRALIA

INSTRUCT WATERS PROF TEST PROG 179

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY TESTING PROGRAM

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

*Approximate temperature of samples upon receipt:

ANALYSIS

Where possible, the proficiency testing sample should be treated as a routine laboratory sample.

analyte only a single result is requested.Report results for the diluted sample.

Report results to the nearest whole number

Report results in micrograms per litre (Do not correct results for recovery.

MU* Laboratories Measurement Uncertainty (MU) if known for the result. Please report in micrograms per litre (µg /L

______________________

Return results NO LATER THAN 1 MAY 2015 Delfina Mihaila Proficiency Testing AustraliaPO Box 7507 SILVERWATER NSW 2128 AUSTRALIA

INSTRUCT WATERS PROF TEST PROG 179

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY TESTING PROGRAM

CHEMICAL ANALYSIS ROUND 179

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

RESULTS SHEET

*Approximate temperature of samples upon receipt:

Result

Where possible, the proficiency testing sample should be treated as a routine laboratory sample.

only a single result is requested.Report results for the diluted sample.

Report results to the nearest whole number

Report results in micrograms per litre (Do not correct results for recovery.

ories Measurement Uncertainty (MU) if known for the result. Please report in /L). Report to the nearest whole number

______________________

NO LATER THAN 1 MAY 2015

Proficiency Testing Australia PO Box 7507 SILVERWATER NSW 2128

INSTRUCT WATERS PROF TEST PROG 179

C5

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY TESTING PROGRAM

CHEMICAL ANALYSIS ROUND 179

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

APRIL 2015

RESULTS SHEET(µg/L)

*Approximate temperature of samples upon receipt:

SAMPLE R179

Result

Where possible, the proficiency testing sample should be treated as a routine

only a single result is requested.Report results for the diluted sample.

Report results to the nearest whole number (no decimal places).

Report results in micrograms per litre (µg /L).

ories Measurement Uncertainty (MU) if known for the result. Please report in ). Report to the nearest whole number

SIGNATURE:

NO LATER THAN 1 MAY 2015

PO Box 7507 Phone:SILVERWATER NSW 2128 Fax:

Email:

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY TESTING PROGRAM

CHEMICAL ANALYSIS ROUND 179

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

APRIL 2015

RESULTS SHEET

SAMPLE R179

±MU

Where possible, the proficiency testing sample should be treated as a routine

only a single result is requested.

(no decimal places).

ories Measurement Uncertainty (MU) if known for the result. Please report in ). Report to the nearest whole number

SIGNATURE: _______________________________

NO LATER THAN 1 MAY 2015 to:

Phone: +61 2 9736 8397 +61 2 9743 6664

Email: [email protected]

PROFICIENCY TESTING AUSTRALIA

WATERS PROFICIENCY TESTING PROGRAM

CHEMICAL ANALYSIS ROUND 179

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

Laboratory Code

METHODCODE

Where possible, the proficiency testing sample should be treated as a routine

(no decimal places).

ories Measurement Uncertainty (MU) if known for the result. Please report in ). Report to the nearest whole number (no decimal places).

_______________________________

+61 2 9736 8397 +61 2 9743 6664 [email protected]

Metals (Chromium, Copper, Iron, Lead, Nickel, Thallium, Zinc)

Laboratory Code

METHOD CODE

DIGESTIONCODE

Where possible, the proficiency testing sample should be treated as a routine

ories Measurement Uncertainty (MU) if known for the result. Please report in (no decimal places).

_______________________________

[email protected]

DIGESTION CODE

Where possible, the proficiency testing sample should be treated as a routine

ories Measurement Uncertainty (MU) if known for the result. Please report in

_______________________________

SD 9.17.09

- End of Report -