Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research...

18
© 2018 Activated Research Company, LLC © 2018 Activated Research Company, LLC Recent Advances in Catalysis for the Measurement of CO, CO 2 , and Formaldehyde Rajvi Mehta October 17, 2018 Gulf Coast Conference

Transcript of Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research...

Page 1: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Recent Advances in Catalysis for the Measurement of CO, CO2, and

Formaldehyde

Rajvi Mehta

October 17, 2018

Gulf Coast Conference

Page 2: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Measuring CO, CO2, & Formaldehyde

• Applications• Ethylene/propylene

production

• Air quality

• Transformer oil gas analysis

• Renewable Fuels

• Beverage

• Catalysis

2

Page 3: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Measuring CO, CO2, & Formaldehyde

• Applications• Ethylene/propylene

production

• Air quality

• Transformer oil gas analysis

• Renewable Fuels

• Beverage

• Catalysis

• Analytical Techniques• GC-TCD

• GC-MS

• GC-Methanizer-FID

3

Page 4: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Packed Capillary

Columns

HP 5890 Agilent Intuvo

GCs

Photo Credit:http://hashahgroup.com/product_agilent_crosslab.php

http://www.dot-red.com/gc-columns/fused-silica-capillary-gc-columns/http://www.aimanalytical.com/5890.html

http://www.businesswire.com/news/home/20160830005184/en/Agilent-Technologies-Introduces-Transformational-Technology-Gas-Chromatography

GC is Evolving to Meet Scientists’ Needs

4

Page 5: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Traditional Methanizers

• 16+ parts• 2-6 hour installation

5

Page 6: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Recent Advances: Jetanizer™

Traditional Methanizer Jetanizer

16+ parts

1 part

6

Agilent

Shimadzu

Page 7: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™: Simple Installation

Traditional Methanizer Jetanizer

2-6 hours 1-5 minutesInstallation time

7

Page 8: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™ vs. Methanizer

8

Page 9: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™: CO2 Analysis

1 mL sample loop

9

Page 10: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™: Hydrocarbon Matrix

1. Methane2. Ethylene3. Acetylene4. Ethane5. Propylene6. Propane

Conventional FID Jet

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

10

Page 11: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™: Hydrocarbon Matrix

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

1. Carbon monoxide2. Methane3. Carbon dioxide4. Ethylene5. Acetylene6. Ethane7. Propylene8. Propane

11

Page 12: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

CO2 in Confined Space

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

Front

Rear

38,000 feet, in flight for 2 h

12

Page 13: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

CO2 in Carbonated Beverages

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

Fresh drink

Day old drink

13

Page 14: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Flue Gas Analysis

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

1. Methane2. CO2

14

Page 15: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Jetanizer™: Formaldehyde

Luong, J., et al., Analytical Methods, “In-situ Methanation with Flame Ionization Detector for the Determination of Carbon Dioxide in Various Matrices.”

1. Air2. Methane3. CO2

4. Ethane5. Formaldehyde6. Propane7. Butane8. Formaldehyde dimer9. Pentane10. Hexane

10

1

2

4

5

3

6 97

8

15

Page 16: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Polyarc® System: Universal Carbon Response

Microreactor Design and

Stainless Steel 3D Printing

Expertise in Catalysis

Design and Synthesis

First and Only Commercial 3D Printed

Microreactor

16

Page 17: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Polyarc® and Jetanizer™

17

Page 18: Recent Advances in Catalysis for the Measurement of CO, CO ... · © 2018 Activated Research Company, LLC Packed Capillary Columns HP 5890 Agilent Intuvo GCs Photo Credit:

© 2018 Activated Research Company, LLC© 2018 Activated Research Company, LLC

Questions?Thank you!

ARC can be found at Booth #522.

18