PI_SHC500_en_8007741

4
SHC500 GRAVIMAT Mobile measurement system for gravimetric dust concentration measurement PRODUCT INFORMATION

description

Dispozitiv portabil de masurare a concentratiilor de praf

Transcript of PI_SHC500_en_8007741

Page 1: PI_SHC500_en_8007741

SHC500GRAVIMAT

Mobile measurement system for gravimetric dust concentration measurement

P R o d u C T I n f o R M AT I o n

Page 2: PI_SHC500_en_8007741

2 G R AV I M AT S H C 5 0 0 | S I C k

SHC500 Gravimat –Simple construction, precise measurement, fast results

Version 1: •LC collectors for low (0.1 ... 200 mg/m3)

Version 2: •HC collectors for high (50 ... 50,000 mg/m3) dust concentrationsThedustcollectorsarefittedwith•suitableroundfiltersbeforemea-surements and form a sealed unit during the complete measuring process including weighing proce-dures.

Probe head with integrated support •for the dust collector, pressure measuring openings for isokinetic control of the extraction process, and a PT100 sensor for precise measurement of the exhaust gas temperature Stainless steel probe shaft• Multi-channel hose with integrated •impulse lines for transmission of the pressure signals to the pres-sure sensors in the automatic unit

Flight case with:

Electronic unit•

6 Pressure sensors•

Power supply•

Suction pump•

Control valve, stop valve•

Condensate separator•

d u S T C o l l e C T o R f I lT e R H e A d P R o b e G S 5 A u T o M AT I C u n I T

A R e A S o f A P P l I C AT I o n

Calibration of continuous dust measurement systems•Performancemeasurementsatfilterplants• Compliance measurements•

Flowandtemperatureprofilemeasurements,determi-•nationoftheflowdirection(swirl)

P e R f o R M A n C e f e AT u R e S

Nodustlossesduringfilterhandling,patentedsampling•system

Highest measuring accuracy especially at low level dust •concentrations

Automatic data recording and system control•

Isokinetic control in real time•

Automatic storage and evaluation of the measuring val-•ues

Measurement results immediately available after sam-•pling

Aut• omaticmeasurementoftheflowangle

Detec• tion of turbulence effects

Compact design, low number of components•

Only one person for transport and handling•

Page 3: PI_SHC500_en_8007741

3S I C k | G R AV I M AT S H C 5 0 0

M e A S u R I n G P R I n C I P l e

G R AV I M AT S Y S T e M o V e R V I e W

Apartialvolumeflow,regulatedwith a control valve by adding false air isokinetically based on the re-corded differential pressure, is ex-tracted from the duct by using a dustcollectorwithfilter,afilterhead probe and a condensate se-perator. Gas velocity as well as gas flowangletotheprobecanbecal-culated from these values. Tempe-rature is recorded by an integrated PT100 sensor within the probe head.Acut-offvalvepreventsabackflowto the dust collector before and af-ter sampling. The extracted partial volumeflowismeasuredusingflowmeasurement with restriction.

Simple connection offilterheadprobeandautomatic unit by screw lock gasket mechanism

Special bracketTemperature sensor

HC dust collector

LC dust collector

Automatic unit

Filter head probe GS5

Page 4: PI_SHC500_en_8007741

SICK MAIHAK GmbH | GermanyNimburger Str. 11 | 79276 Reute | www.sick.comPhone +49 7641 469-0 | Fax +49 7641 469-1149 | [email protected]

8007

741/2009-06∙DIV03/KE∙PrintedinGermany(2009-10)∙Subjecttochangewithoutpriornotice

Technical data SHC500 GRAVIMAT

Measuring parameters

Measuring components Dust concentration with gravimetrical comparison measurement

Measuring ranges 0.1 ... 200 mg/m• 3 with dust collector LC50 ... 50,000 g/m• 3 with dust collector HC

Accuracy ± 1 % of maximum throughput

Measuring conditions

Medium temperature Standard: 0 ... 250 °C (32 … 480 °F) •With air cooling: 0 ... 400 °C (32 … 700 °F) •Special version: 0 ... 600 °C (32 … 1,110 °F) •

Internal duct pressure –70 ... +70 hPa (–1.0 … 1.0 psi)

Gas velocity 2 ... 48 m/s (6.6 … 157.5 ft/sec)

Ambient conditions

Umgebungstemperatur –10 ... +50 °C

Interfaces

Interface RS232: 9600, N, 8, 1

General information

System components Automatic unit SHC-AE501/502•Filter head probe GS5•Dust collector LC or•Dust collector HC•

Operation Via LC-Display and Function key pad; Version SHC502 with operational software•

Approvals

Compliances EN 13284-1• US-EPA Method 17•