Ion Pumps – Operation and Applications - DUNIWAY

40
1 Ion Pumps Ion Pumps Operation and Operation and Applications Applications

Transcript of Ion Pumps – Operation and Applications - DUNIWAY

Page 1: Ion Pumps – Operation and Applications - DUNIWAY

1

Ion Pumps Ion Pumps –– Operation and Operation and ApplicationsApplications

Page 2: Ion Pumps – Operation and Applications - DUNIWAY

2

Ion Pumps Ion Pumps -- OutlineOutline• Where they fit in the

vacuum pumping realm

• Basic principles of operation

• Genealogy of ion pump configurations

• Parameters ‘BVD’

• Pumping mechanisms for common gases

• Typical operating cycle

• Common applications

• Applications in microscopy

• Problems and troubleshooting

Page 3: Ion Pumps – Operation and Applications - DUNIWAY

3

Vacuum Pumps Vacuum Pumps Types & RangesTypes & Ranges

Page 4: Ion Pumps – Operation and Applications - DUNIWAY

4

Ion PumpsIon PumpsMain CharacteristicsMain Characteristics

• “Capture” Pumping; No Backing Required• Contamination-Free• Vibration-Free• Ion Current Provides Pressure Indication• Long Life at HV and UHV• Low Maintenance• Simple Operation

Page 5: Ion Pumps – Operation and Applications - DUNIWAY

5

Ion Pump PrincipleIon Pump PrinciplePenning DischargePenning Discharge

Page 6: Ion Pumps – Operation and Applications - DUNIWAY

6

Ion Pump Current ProvidesIon Pump Current ProvidesPressure IndicationPressure Indication

• Ion Current is Proportional to Pressure• Wide Range Pressure Indicator

– UHV – Limited by Field Emission Current– High Pressure (>10-4 Torr) Power Limited

• Approximate Relationship:– I/P = 10 x S, for example:– I/P = 1000 amps/torr for typical 100 l/s Pump– See Table on Next Page

Page 7: Ion Pumps – Operation and Applications - DUNIWAY

7

Current Current vsvs PressurePressureTypical 100 L/S Ion PumpTypical 100 L/S Ion Pump

Pressure - Torr Ion Current - Amps Comment 1 x 10-3 3 x 10-1 Power Supply Limited 1 x 10-4 1 x 10-1 I = 1000 P 1 x 10-5 1 x 10-2 I = 1000 P 1 x 10-6 1 x 10-3 I = 1000 P 1 x 10-7 1 x 10-4 I = 1000 P 1 x 10-8 1 x 10-5 I = 1000 P 1 x 10-9 1 x 10-6 I = 1000 P 1 x 10-10 5 x 10-7 Field Emission Limited

Page 8: Ion Pumps – Operation and Applications - DUNIWAY

8

Current vs. Pressure 100 L/S Ion PumpCurrent vs. Pressure 100 L/S Ion Pump

1.E- 0 7

1.E- 0 6

1.E- 0 5

1.E- 0 4

1.E- 0 3

1.E- 0 2

1.E- 0 1

1.E+0 0

1.E -10 1.E -09 1.E -08 1.E -07 1.E -06 1.E -05 1.E -04 1.E -03

Pressure - Torr

Cur

rent

- A

mps

Page 9: Ion Pumps – Operation and Applications - DUNIWAY

9

Ion PumpsIon PumpsElement/Cathode GenealogyElement/Cathode Genealogy

Page 10: Ion Pumps – Operation and Applications - DUNIWAY

10

Diode ElementDiode Element

Page 11: Ion Pumps – Operation and Applications - DUNIWAY

11

Slotted Diode ElementSlotted Diode Element

Page 12: Ion Pumps – Operation and Applications - DUNIWAY

12

Triode ElementTriode Element

Page 13: Ion Pumps – Operation and Applications - DUNIWAY

13

Differential/DI/Noble Differential/DI/Noble Diode ElementDiode Element

Page 14: Ion Pumps – Operation and Applications - DUNIWAY

14

Hydrogen ElementHydrogen Element

Page 15: Ion Pumps – Operation and Applications - DUNIWAY

15

StarCellStarCell™ Element™ Element

Page 16: Ion Pumps – Operation and Applications - DUNIWAY

16

Galaxy™ Diode ElementGalaxy™ Diode Element

Page 17: Ion Pumps – Operation and Applications - DUNIWAY

17

Galaxy™ Triode ElementGalaxy™ Triode Element

Page 18: Ion Pumps – Operation and Applications - DUNIWAY

18

Ion Pump Parameters ‘BVD’Ion Pump Parameters ‘BVD’

• B = Magnetic Field Strength

• V = Operating Voltage

• D = Anode Cell Diameter

Page 19: Ion Pumps – Operation and Applications - DUNIWAY

19

B = Magnetic FieldB = Magnetic FieldLow Pressure Operation Improves with Higher B

Typical Magnetic Field 1200 to 1500 Gauss

Figure 5: Discharge Intensity vs. Pressure Three Different Magnetic Fields

(36 Anode Cells, 0.5" Diameter, 3000 Volts)

0

50

100

150

200

1E-11 1E-10 1E-09 1E-08 0.0000001 0.000001 0.00001 0.0001 0.001

P ressure ( to rr)

1000gauss

1500 gauss

2000 gauss

Page 20: Ion Pumps – Operation and Applications - DUNIWAY

20

V = Operating VoltageV = Operating VoltagePumping Speed Increases Linearly with Voltage

Typical Operating Voltage 4000 to 7000 Volds DC

Figure 4: Pumping Speed (l/s) vs. Magnetic Field & Voltage

0

0.5

1

1.5

2

2.5

0 1000 2000 3000 4000 5000

Magnetic Field (gauss)

Pum

ping

Spe

ed (l

/s)

4 kv

10 kv

Page 21: Ion Pumps – Operation and Applications - DUNIWAY

21

D = Anode Cell DiameterD = Anode Cell DiameterLow Pressure Operation Improves with Larger D

Typical Anode Cell Diameter 0.5 to 1.0 inch

Fig. 6: Discharge Intensity (amps/torr) vs. Pressure (torr)

Three Different Cell Diameters (1000 gauss, 3000 volts)

020406080

100120140

1E-11 1E-10 1E-09 1E-08 1E-07 0.000001 0.00001 0.0001

Pressure ( t o rr)

6-1" Cells 36-1/2" Cells3-2" Cells

Page 22: Ion Pumps – Operation and Applications - DUNIWAY

22

Pumping Mechanisms for Pumping Mechanisms for Common GasesCommon Gases

• Chemically active: N, O, H

• Common small molecules: H2O, CO, CO2, CH4, NH3, NO

• Noble gases: Ar, He, Ne, Kr, Xe

Page 23: Ion Pumps – Operation and Applications - DUNIWAY

23

Chemically Active GasesChemically Active GasesNN22 –– OO22 –– HH22

• Dissociated & Ionized in Discharge• Accelerated to Cathode• Nitrogen and Oxygen

Sputter Titanium Cathode MaterialNeutralizedForm stable, chemical combination in

regions of sputtered titanium• Hydrogen

Solution/Diffusion into TitaniumMostly in the Cathodes

Page 24: Ion Pumps – Operation and Applications - DUNIWAY

24

Common Small MoleculesCommon Small MoleculesH2O, CO, CO2, CH4, NH3, NO

• Dissociated & Ionized in Discharge• Accelerated to Cathode• Active Ions – O, N, H

– Same pumping mechanism as active gases

• Carbon– Accelerated to cathode, neutralized– Deposited as free carbon

Page 25: Ion Pumps – Operation and Applications - DUNIWAY

25

Noble Gases Noble Gases He, He, NeNe, , ArAr, Kr, , Kr, XeXe

• Ionized in Discharge• Accelerated to Cathode• Sputter/Neutralized• Sputtering rate goes up with mass• Pumped stably in areas of net buildup of sputtered

material (cathode, pump walls, anode• Heavier noble gases require proper element

configuration for stable pumping(Triode, DI/Noble, Starcell™, Galaxy™

• Helium diffuses into titanium

Page 26: Ion Pumps – Operation and Applications - DUNIWAY

26

Argon InstabilityArgon InstabilityPeriod Shorter at Higher Leak Rates

Pmax ~ 2 x 10-4 Torr (Discharge Mode Shift)Argon Displaced into Stably Pumped Areas During Mode Shift

Slow Rise/Rapid Decline Characteristic Wave Shape

Argon Instability

1.E-06

1.E-05

1.E-04

1.E-03

0 100 200 300 400 500

Time - seconds

Pres

sure

-Tor

r

Page 27: Ion Pumps – Operation and Applications - DUNIWAY

27

Ion PumpsIon PumpsTypical Operating CycleTypical Operating Cycle

• Rough Pump with Clean Technology– Turbo, Trapped Mechanical, Sorption– Rough to Pressure Below 10-3 T0rr– Lower Pressure Saves Time, Extends Life

• Turn on Ion Pump Control Unit High Voltage• Cycle Power if Necessary to Avoid

Overheating• Valve Off Roughing Pump• Walk Away

Page 28: Ion Pumps – Operation and Applications - DUNIWAY

28

Ion PumpsIon PumpsCommon ApplicationsCommon Applications

• Electron Microscopes• Accelerators• Microwave Tubes• X-ray Tubes• Medical Equipment• Coating Equipment• Materials Research• Mass Spectrometers

Page 29: Ion Pumps – Operation and Applications - DUNIWAY

29

Ion PumpsIon PumpsApplications in MicroscopyApplications in Microscopy

• Source• Column• Sample Chamber• Sample Metallizing/Coating• Sample Storage

Page 30: Ion Pumps – Operation and Applications - DUNIWAY

30

Ion PumpsIon PumpsProblems and TroubleshootingProblems and Troubleshooting

• Initial Set-Up and Operation• Field emission leakage• ‘Noisy’• Over-Heats• Hard to Start – High Pressure• Hard to Start – Low Pressure• “Sluggish”, Low Pumping Speed• Lifetime

Page 31: Ion Pumps – Operation and Applications - DUNIWAY

31

Initial SetInitial Set--Up and OperationUp and Operation

• Leak-Tight Connection• Proper Conductance: > 5X Pump Speed• Correct Magnet Strength/Orientation• Visible Grounding Wire• Good High Voltage Connection

– Pump HV Feedthrough– HV Connecter – Pump End– HV Connector – Controller End

Page 32: Ion Pumps – Operation and Applications - DUNIWAY

32

Field Emission LeakageField Emission Leakage

• Buildup of Sharp-Pointed Deposits• Field Emission Current from Points• Masks Ion Current at Low Pressures• Doesn’t Influence Pumping Speed• Can Be Removed by “Hi-Potting”

Page 33: Ion Pumps – Operation and Applications - DUNIWAY

33

‘‘Noisy’Noisy’

• Long-Term Operation Builds up Sputtered Deposits

• Deposits Can ‘Flake Off’• Occurs Especially with Vibration, Shock• Generates Current/Pressure Spikes as

Flakes Fall into Discharge • When Such Noise becomes a Problem,

Pump Needs Cleaning/Rebuilding

Page 34: Ion Pumps – Operation and Applications - DUNIWAY

34

OverOver--HeatsHeats

• Occurs at Higher Pressure, ie >10-5 Torr• With Properly Matched Control Unit,

Damage Will Not Occur• Caused by Excessive Gas Load• Reduce Gas Load or Add Auxiliary Pump

– Titanium Sublimation, Cryo, Turbo, NEG• May Require Re-Roughing if Excessive

Page 35: Ion Pumps – Operation and Applications - DUNIWAY

35

Hard to Start Hard to Start –– High PressureHigh Pressure

• May be Caused by Inadequate Roughing• Badly-Matched Control Unit with

Excessive Power Could Cause Over-Heating

• Older Pump May Have Adsorbed Gases on High Surface Area Deposits or Dissolved Gases (H, He) in Cathodes

• Improve Roughing or Clean/Rebuild

Page 36: Ion Pumps – Operation and Applications - DUNIWAY

36

Hard to Start Hard to Start –– Low PressureLow Pressure

• Most Likely to Occur at UHV • Discharge Needs an ‘Event’ to Initiate• ‘Events’ Less Likely at Low Pressures• Smaller Pumps Most Susceptible• A Gentle Tap May Help• Injected Electrons or UV Sometimes Used

Page 37: Ion Pumps – Operation and Applications - DUNIWAY

37

““Sluggish”, Low Pumping SpeedSluggish”, Low Pumping Speed

• Older Pumps May Release Gas from Deposited High Surface Area

• Baking Pump During Start May Help• Clean/Rebuild makes Pump Like New• System Design May Limit Conductance to

Pump, Limiting Speed• Larger Diameter, Shorter Attachment Will

Help

Page 38: Ion Pumps – Operation and Applications - DUNIWAY

38

Ion PumpsIon PumpsLifetimeLifetime

• Normal Lifetime is 30,000 – 40,000 Hours at a Pressure of 1 x 10-6 Torr

• Heavy Noble Gases May Cut Life• Lifetime is Inversely Proportional to

Pressure: • 1/10X at 1 X 10-5 Torr• 10X at 1 X 10-7 Torr• 100X at 1 X 10-8 Torr• ETC.

Page 39: Ion Pumps – Operation and Applications - DUNIWAY

39

Ion Pumps Ion Pumps -- ConclusionConclusion

• When applied properly, ion pumps are an excellent, reliable solution for HV and UHV

Page 40: Ion Pumps – Operation and Applications - DUNIWAY

July 2006 40

Contact InformationContact Information