HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

39
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY INSTRUMENTATION AND WORKING BINDU KSHTRIYA Dept. of Pharmaceutical sciences, Maharishi Dayanand University, Rohtak

Transcript of HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Page 1: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

INSTRUMENTATION AND WORKING BINDU KSHTRIYA

Dept. of Pharmaceutical sciences, Maharishi Dayanand University, Rohtak

Page 2: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

LIQUID CHROMATOGRAPHY

· 1906 - Mikhail Semenovich Tswett used Calcium Carbonate and Petroleum Ether to separate plant chlorophyllsThe separation was named Chromatography from the Greek word chromos for color and to write (graphy)

In the 1940’s, Martin and Synge introduced the concept of Partition Chromatography.In the1950’s Gradient elution was introduced by Tiselius and theory of separation was described by VanDeemter.

Finally ,in the 1960’s the first commercial liquid chromatography was introduced .

Page 3: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SEPARATION MECHANISMChromatographic processes can be defined as separation techniques involving mass-transfer between stationary and

mobile phases.

Separation in based upon differential migration between the stationary (packing material) and mobile phases (solvent). Mobile phase elutes components while Packing materials retain components in the column

Page 4: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

COMPONENTS OF A LIQUID

CHROMATOGRAPH SYSTEM

Mobile Phase / Solvent ReservoirDegasser

Solvent Delivery System (Pump)Injector

Pre-columnColumn

Temperature ControlDetectors

Recorder (Data Collection)

Page 5: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

HPLC SOLVENT RESERVOIR SYSTEMS

Degasser is needed to remove dissolved air by subjecting the mobile phase under vacuum, distillation, spurging with fine spray of an inert gas at lower solubility such as Argon and

Helium or by heating and ultrasonic stirring.

■ Filtration is needed to eliminate suspended particles and organic impurities.

Page 6: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
Page 7: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DEGASSERProblems caused by dissolved air(O2, N2)in mobile phase

Unstable delivery in pumpBigger noise and large baseline-drift in detector cell

▪ In order to avoid causing the problems,

mobile phase should be degassed.vacuum pumping systems

distillation systema system for heating and stirring the solvents

sparging system - bubbles an inert gas of low solubility through the solvent

Page 8: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PUMPS■ Pass mobile phase through column at high pressure and at controlled flow rate.

■ Performance of pump directly affects the Rt, reproducibility, detector sensitivity

Page 9: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

IDEAL CHARCETRISTIC OF A PUMP■ Non corrosive and compatible with solvent.

■ Provide High pressure to push mobile phase

■ Provide constant flow rate to mobile phase.

■ Easy to change for one mobile phase to another.

■ Should have reproducible flow rate and independent of column back pressure

■ Should not leak and should be easy to dismantle and repair

Page 10: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

TYPE OF PUMP USED IN HPLC

■ Reciprocating pump

■ Displacement pump

■ Pneumatic pump

Page 11: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

RECIPROCATING PUMP

Page 12: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

RECIPROCATING PUMP

WORKING

■ Contains reciprocating piston that moves back and forth in hydraulic chamber.

■ By the movement of piston solvent flow into the column under high pressure.

■ When piston moves backward inlet valve open while exit valve closes. This result in mobile phase being drawn into the main chamber (cylinder).

Page 13: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

RECIPROCATING PUMP■ When the piston moves to the front the inlet valve closes and the exit valve opens.

■ The reduction in volume in main chamber due to forward motion of piston result in mobile phase moving out of the exit valve under high pressure.

DISADVENTAGE

■ Pulsed flow which must be damped as they produce a base line noise on the chromatogram

Page 14: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

RECIPROCATING PUMPADVANTAGES

■ Generate high output pressure (upto10000 psi).

■ Ready adaptability to gradient elusion.

■ Provide constant flow rate.

■ Pressure generated is so high that any back pressure generated in the column due to higher viscosity of stationary phase can be easily overcome.

Page 15: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DISPLACEMENT PUMP / SYRINGE PUMP

Page 16: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DISPLACEMENT PUMP / SYRINGE PUMPWORKING

■ Works on the principle of positive solvent pressure.

■ Consist of screw or plunger which revolves continuously driven by motor.

■ Rotatory motion provides continuous movement of the mobile phase which is propelled by the revolving screw at greater speed and pushes solvent through small needle like outlet.

■ Consist of large syringe like chamber of capacity 250 – 500 ml.

Page 17: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DISPLACEMENT PUMP / SYRINGE PUMPADVANTAGES

■ Flow is pulse free.■ Provide high pressure upto 200 – 475 atm.

■ Independent of column back pressure and viscosity of solvent.■ Simple operation.

DISADVENTAGE

■ Limited solvent capacity ■ Gradient elution is not easy.

Page 18: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PNEUMATIC PUMP

WORKING•Pressure from a gas cylinder delivered through a large

piston drivers the mobile phase. •Pressure on the solvent is proportional to the ratio of

piston usually 50: 1.

Page 19: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PNEUMATIC PUMP■ The driving air is applied, piston moves, inlet closes & outlet open pushing mobile phase to the column.

■ A lower pressure gas source of 1- 10 atm can be used to generate high liquid pressure .( 1 – 400 atm )

■ About 70 ml of the mobile phase is pumped from every stroke.

Page 20: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PNEUMATIC PUMPADVENTAGES

■ Pulse free flow.

■ Generates high pressure.

DISADVANTAGES

■ Limited volume capacity (70 ml )

■ Pressure output and flow rate depends on the viscosity and column back pressure.

■ Gradient elusion is not possible.

Page 21: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PULSE DAMPERImportant damping methods include

■ A triple headed pump: - two heads in different stages of filling as the third is pumping.

■ A tube with an air space or a flexible bellows or tube: - where a gas (air space) or a flexible metal vessel takes up some of the solution energy. When pump refills, this energy is released and a smooth pressure pulsation result.

■ A restrictor: - I this method, a 25 cm length of 4 mm bore .stainless steel tubing. Packed with 20µm glass beds, is placed between the pump and the column.

Page 22: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SAMPLE INJECTION SYSTEM■ Septum injectors

■ Stop flow septumless injection.

■ Rheodyne injector / loop valve type.

Page 23: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SAMPLE INJECTION SYSTEM

Septum injection port.

■ Syringe is used to inject the sample through a self sealing inert septum directly into the mobile phase.■ Drawback: - leaching effect of the mobile phase with the septum resulting in the formation of ghost peaks.

Stop flow septumless injection.

■ Flow of mobile phase through the column is stopped for a while.■ Syringe is used to inject the sample.■ Drawback: formation of ghost peak.

Page 24: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SAMPLE INJECTION SYSTEM

Rheodyne injector / loop valve type.

■ Sophisticated modern method with good precision.

■ Sample is introduced in the column without causing interruption to mobile phase flow.

■ Volume of sample ranges between 2 µl to over 100 µl.

Page 25: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SAMPLE INJECTION SYSTEM■ Operation of sample loop.

• sampling mode• Injection mode.

■ Sample is loaded at atmospheric pressure into an external loop in

the micro volume sampling valve, and subsequently injected into the mobile phase by suitable rotation

of the valve.

Micro volume sampling valve operation of a Sampling loop.

Page 26: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

COLUMN■ Made up of stainless steel or heavy glass to withstand the pressure.

■ The columns are usually long (10 – 30 cm) narrow tubes. Contains stationary phase at particle diameters of 25 µm or less.

■ The interior of column should be smooth and uniform.

■ Column end fitting are designed to have a zero void volume.

Page 27: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

CLASSIFACTION OF CLOUMN BASED ON APPLICATION

columnMain column Guard column

Analytical column Preparative column

Standard columnNarrow bore

Short fast column

Micro preparative Preparative columnMacro preparative

Page 28: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

CLASSIFACTION OF COLUMN ON THE BASES OF COMPONENTS

■ BONDED PHASE COLUMN

■ COLUMN WHERE LIQUID IS INPERMAGNETED ON SOLID INERT SUPPORT

Page 29: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

STANDARD COLUMN

• Internal diameter 4 – 5 mm and length 10 – 30 cm. • Size of stationary phase is 3 – 5 µm in diameter.

• Used for the estimation of drugs, metabolites, pharmaceutical preparation and body fluids like plasma.

NARROW BORE COLUMN

■ Internal diameter is 2 – 4 mm. ( signal is increased 4 times )■ Require high pressure to propel mobile phase.

■ Used for the high resolution analytical work of compounds with very high Rt.

Page 30: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SHORT FAST COLUMN

■ Length of column is 3 – 6 cm.■ Used for the substances which have good affinity towards the stationery phase.

■ Analysis time is also less (1- 4 min for gradient elusion & 15 – 120 sec for isocratic elusion).

PREPARATIVE COLUMN

■ Used for analytical separation i.e. to isolate or purify sample in the range of 10-100 mg form complex mixture.• Length – 25- 100 cm

• Internal diameter – 6 mm or more.

Page 31: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Preparative column are of three type : • Micro preparative or semi preparative column

■ Modified version of analytical column■ Uses same packaging and meant for purifying sample less then 100 mg.

• Preparative column ■ Inner diameter – 25 mm .

■ Stationary phase diameter – 15- 100 µm

• Macro Preparative Column

■ Column length – 20 – 30 cm■ Inner diameter – 600 mm

Page 32: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

GUARD COLUMN • They are placed anterior to the separating column.

• Serve as a protective factor that prolongs the life and usefulness of the column.

• They are dependable column designed to filter or remove

■ Particles that clog the separation column.

■ Compounds and ions that could ultimately cause baseline drift, decrease resolution , decrease sensitivity and create false peaks.

Page 33: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

BONDED PHASE COLUMN■ Here the molecules, comprising the stationary phase i.e. the surface of the silica particles, are covalently bonded to

a silica based support particles.

■ The most popular bonded phase ,siloxanes , are formed by heating the silica particles in dilute acid for the day so as to generate the reactive Silonal group.

- OH OH OH ו ו ו

- Si – O – Si - O - Si - ו ו ו ו

Page 34: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

■ Silonal group is the treated with organochlorosilane .

CH3 CH3Si – OH + Cl – Si – R →- Si – O – Si – R + HCl ו ו

│ │ CH3 CH3

• These bonded phases are stable between the pH range 2 – 9 and upto temperature of 80º C.

■ Bonded phase is made with a linear C 18 hydrocarbon, also know as ODS (octadecyl silane) bonded phase. Used in pharmaceutical analysis or separation of less polar components .

Page 35: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

■ An alkyl nitrile column or cyano column which has 12 carbon atoms with the last atom appearing as a nitrile group (CN) .moderately polar column.

■ Amino alkyl bonded phase column which is normally C 8, last C atom bearing NH2 group. Polar column. Use full in separation of CHO, peptides, amino acids.

Advantages • Can withstand high pressure exerted by mobile phase.

• Life of column is more.• No bleeding effect

Disadvantages

■ Very expensive ■ Manually can not be fabricated

Page 36: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

TYPE OF DETECTORSBulk property detectors

Solute property detectors

TYPE OF DETECTORS

Page 37: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

■ BULK PROPERTY DETECTORS

• Compare an over all change in physical property of mobile phase with or without an eluting solute.• These types of detectors tend to be relatively insensitive and require temperature control.

e.g. Refractive index detector.

SOLUTE PROPERTY DETECTORS

• They respond to a physical property of the solute that is not exhibited by the pure mobile phase. • These detectors are more sensitive, detect the sample in nanograms quantity.

e.g. uv visible detector , electrochemical detector, fluorescence detector.

Page 38: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

TYPES OF DETECTORUV/Visible

- Fixed wavelength- variable wavelength

Photo Diode ArrayRefractive index

FluorescenceEvaporative light scattering

ConductivityElectrochemical

Page 39: HIGH PERFORMANCE LIQUID CHROMATOGRAPHY