Gas Chromatography (GC)

• This method depends upon the solubility and boiling points of organic liquids in order to separate them from a mixture. It is both a qualitative (identity) and quantitative (how much of each) tool.

Page 3: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Gas Chromatography (GC)Introduction – Overview of Topics

• Applications– Most common for volatile compounds– More common for non-polar to

moderately polar compounds

• Columns (packed vs. open tubular)• Sample Injection• Detectors

Page 4: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Process Flow Schematic

Carrier gas (nitrogen or helium)

Sample injection

Long Column (30 m)

Detector (flame ionization detector or FID) Hydrogen

Air

Page 5: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Gas Chromatograph Components

Flame Ionization Detector

Column

Oven

Injection Port

top view

front view

Page 6: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Theory• An inert gas such as helium is passed through the column as a carrier gas and is the

moving phase. A sample is injected into a port which is much hotter than the column and is vaporized. The gaseous sample mixes with the helium gas and begins to travel with the carrier gas through the column. As the different compounds in the sample have varying solubility in the column liquid and as these compounds cool a bit, they are deposited on the column support. However, the column is still hot enough to vaporize the compounds and they will do so but at different rates since they have different boiling points. The process is repeated many, many times along the column. Eventually the components of the injected sample are separated and come off of the column at different times (called "retention times").

• There is a detector at the end of the column which signals the change in the nature of the gas flowing out of the column. Recall that helium is the carrier gas and will have a specific thermal conductivity, for example. Other compounds have their own thermal conductivities. The elution of a compound other than helium will cause a change in conductivity and that change is converted to an electrical signal. The detector, in turn, sends a signal to a strip chart recorder or to a computer. Detectors come in several varieties, for example, thermal detectors, flame-ionization and electron capture detectors.

Page 7: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Theory

To detector

Analyte

Column + packing

Time

Page 8: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Separations

To detector

Time 1 Time 2 Time 3

Page 9: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Process

Page 10: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC General• The chromatogram shows the order of elution (order

of components coming off the column), the time of elution (retention time), and the relative amounts of the components in the mixture. The order of elution is related to the boiling points and polarities of the substances in the mixture. In general, they elute in order of increasing boiling point but occasionally the relative polarity of a compound will cause it to elute "out of order". This is analyzing your sample.

Page 11: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Elution Order

• Compound Boiling Point (0C)• pentane 36• hexane 69• cyclohexane 80• isooctane 99• toluene 110• 4-methyl-2-pentanone 117• octane 126

Page 12: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Example Chromatogram

• The observed elution pattern appears below. Notice the reversed elution of toluene and 4-methyl-2-pentanone.

Page 13: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Chromatogram

Page 14: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Components of GC:

• Column, oven, injector, and detector. These parameters (HETP, etc) are affected by the various components of the instrumentation. Perhaps the column is the most important component of the GC. With it, different separations can be accomplished.

Page 15: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Instrumentation• Carrier gas

The carrier gas must be chemically inert. Commonly used gases include nitrogen, helium, argon, and carbon dioxide. The choice of carrier gas is often dependant upon the type of detector which is used. The carrier gas system also contains a molecular sieve to remove water and other impurities.

• Sample injection port • For optimum column efficiency, the sample should not be too large, and should be

introduced onto the column as a "plug" of vapour - slow injection of large samples causes band broadening and loss of resolution. The most common injection method is where a microsyringe is used to inject sample through a rubber septum into a flash vapouriser port at the head of the column. The temperature of the sample port is usually about 50°C higher than the boiling point of the least volatile component of the sample. For packed columns, sample size ranges from tenths of a microliter up to 20 microliters. Capillary columns, on the other hand, need much less sample, typically around 10-3 mL. For capillary GC, split/splitless injection is used. Have a look at this diagram of a split/splitless injector;

Page 16: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Sample Injection and Pretreatment• Gas Sample: if present at moderate to high

concentration, may be directly put into column via gas-tight syringe.

• Another technique is to use a gas-tight valve:

For trace levels like volatile organic compounds (VOC), the sample may need to be preconcentrated by passing it through a solid-absorbing cartridge or using a cryogenic trap.

Page 17: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Sample Injection and Pretreatment• Liquid Samples: direct injection of a relatively large sample (6-10

mL). A calibrated microsyringe may be injected through a gas-tight septum into a heated chamber that volatilizes the sample.

• Note that there are direct, split, and splitless column styles. • Sometimes pretreatment requires dervitatization or moving the

sample into a different solvent (remove from H2O).• Static or Dynamic Headspace analysis of vapor phase above a

sample avoids water in the sample.

Page 18: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Sample Injection and Pretreatment• Solid Samples: First extract compounds of interest

from a solid matrix by liquid-liquid extraction or supercritical fluid. Samples are placed in solvent, then treated as liquid samples.

• Thermal desorption may be used on some solids as analytes may be collected during heating of a solid.

• Pyrolysis GC is employed for substances that are not volatile and cannot be easily derivatized to volatile forms. The solid is heated in a controlled way to break it into smaller, more volatile pieces separated by the column to form a pyrogram, which can be matched to known standards.

Page 19: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Sample Injections• Next, the sample injection system. Here it is important that

the sample be injected onto the column as a plug and of a suitable size. Also, the injector should provide consistent and reproducible injections. See Figure 27-3, Pg 704. The micro-syringe is used to load the sample onto the column. The syringe should be clean and accurate and gas tight. The syringe is injected through a rubber septum. The septum should be replaced after many injections to insure gas tightness onto the column. An auto sampler can be used to inject the samples. Typical volumes range from 0.2 to 20 Ls. With capillary columns it is necessary to use a splitter. (See Figure.) A suitable solvent is also necessary for the proper separations and injections.

Page 20: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Injector

Page 21: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCInjection

• Gas Samples– various methods including injectors used for

liquid samples (below), fixed loop injectors (described later with HPLC), and solid phase microextraction (SPME)

• Liquid Samples – on-column injection (needle goes into column)– direct injection– above two methods mainly for packed columns– split/splitless injector (next page)

Page 22: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCSample Injection – Liquid Samples

• Split/Splitless Injectors– Injectors capable of running

in two modes: split and splitless

– Split injections used to avoid overloading columns

• Injection Process– Syringe pierces septum and

depressing plunger deposits liquid

– Analyte volatilizes– Part injected (usually

smaller fraction)– Part passed to vent– Fraction vented depends on

split valve

outside

Inside oven

Septum

Syringe port

He in

To Column

Split vent

Split valve

liner

Page 23: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCSample Injection – Liquid Samples

• Split injection is used for:– Higher concentrations– Smaller diameter (OT) columns– Greater need for high resolution than

high accuracy• In split injection, solvent overload is

less problematic• Splitless injection is used for trace

analysis (~50% of injected sample put on column)

Page 24: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Carrier Gas

• This is the mobile phase and should be pure gas so as not to react with the column or analyte. Gas is usually He, Ar, N2, or H2. Choice will depend on the type of detector used. He and H2 give better resolution (smaller plate height) than N2. Pressure is also important and as expected the system comes with regulators. Can you find where in GC equations that are dependent on pressure?

Page 25: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Columns

• The column is the most important component of GC. Here is where the separations take place. All the various equations we discussed above are dependent on properties of the column. There are four types of columns: wall-coated open tubular (WCOT), support coated open tubular (SCOT), micropacked, fused silica open tubular (FSOT), and packed column. The FSOT column is the most flexible. Open tubular is also capillary. Particle size is important because the efficiency of GC column increases rapidly with decreasing particle size of the packing material.

Page 26: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Column

• The column sits in a temperature controlled environment that is 0.50. Temperature is very important in GC. Can you remember what equations are affected by temperature? See page 706 Fig. 27-5 for temperature effects on separations. Normally, one does a temperature program to get the various analytes off the column for better separations (resolutions).

Page 27: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Columns

• There are two general types of column, packed and capillary (also known as open tubular). Packed columns contain a finely divided, inert, solid support material (commonly based on diatomaceous earth) coated with liquid stationary phase. Most packed columns are 1.5 - 10m in length and have an internal diameter of 2 - 4mm.

• Capillary columns have an internal diameter of a few tenths of a millimeter. They can be one of two types; wall-coated open tubular (WCOT) or support-coated open tubular (SCOT). Wall-coated columns consist of a capillary tube whose walls are coated with liquid stationary phase. In support-coated columns, the inner wall of the capillary is lined with a thin layer of support material such as diatomaceous earth, onto which the stationary phase has been adsorbed. SCOT columns are generally less efficient than WCOT columns. Both types of capillary column are more efficient than packed columns.

• In 1979, a new type of WCOT column was devised - the Fused Silica Open Tubular (FSOT) column;

Page 28: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

FSOT column

Page 29: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Capillary column

Page 30: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Columns

• Two Common Formats– Packed columns (most

common with bonded liquid coating)

– Open tubular (typically long columns with small diameters)

• Advantages of packed columns– Greater capacity (can

handle larger masses injected on the column)

– Lower cost/somewhat more variety of stationary phases (common as specialty column)

Bonded liquid in green

Packed Columns

Open Tubular(end on, cross section view)

Column Wall (fused silica)

Mobile phase

Stationary phase (wall coating)

Page 31: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Columns

• Advantages of Open Tubular Columns– Best resolution (negligible A term, small C term

in van Deemter Equation)– More robust– Better sensitivity with many detectors (due to

less band broadening vs. lower mass through column)

• Column Selection– High resolution (thin film, 0.25 mm diameter,

60 m) vs. higher capacity (thick film, 0.53 mm diameter)

– Stationary phase based on polarity

Page 32: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Stationary Phase

• Selection of stationary phase affects k and α values• Main concerns of stationary phase are: polarity, functional

groups, maximum operating temperature, and column bleed (loss of stationary phase)

• More polar columns suffer from lower maximum temperatures and greater column bleed

Type Functional Groups Polarity

OV-1 methyl Non-polar

OV-17 50% methyl/50% phenyl

Somewhat polar

OV-225 Cyanopropyl, methyl, and phenyl

More polar

carbowax Ether groups polar

Page 33: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Adjustments

• k is adjusted by changing temperature (higher T means smaller k)

• the a values are adjusted by changing column (will work if there is a difference in solute polarity)– example: separation of saturated and unsaturated fatty acid

methyl esters (FAMEs).

– Retention of C18:0 and C18:1 FAMEs on OV-5 columns is very similar (due to similar boiling points)

– Retention on more polar columns (e.g. OV-17) is greater for the more polar unsaturated FAMEs

• Changing carrier gas has no effect on retention

O

CH3

O CH3

O

CH3

O CH3

C18:0 FAME

C18:1 FAME

bp = 369°C

bp = 352°C

Page 34: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Gradients

• In GC, a “gradient” elution can be performed by increasing the oven temperature during the run

• This is useful when sample solutes have a wide variety of volatilities (or at least retention times)

• This results in a continuation of narrow peaks and less time between peaks, less total time and better late peak S/N ratios

Note: Data from HPLC, but looks very similar to GC chromatograms for a series of compounds with extra –CH2- groups

Isocratic Run

Gradient Run

Page 35: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCDetectors

• Performance criteria– limit of detection (low is better)– Linear range (or useful range for non-

linear detectors)– Destructive vs. non-destructive– Other (cost, speed, size, precision, etc.)

• Most GC detectors perform well for cost vs. HPLC

Page 36: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Instrumentation• Detectors• There are many detectors which can be used in gas chromatography.

Different detectors will give different types of selectivity. A non-selective detector responds to all compounds except the carrier gas, a selective detector responds to a range of compounds with a common physical or chemical property and a specific detector responds to a single chemical compound. Detectors can also be grouped into concentration dependant detectors and mass flow dependant detectors. The signal from a concentration dependant detector is related to the concentration of solute in the detector, and does not usually destroy the sample Dilution of with make-up gas will lower the detectors response. Mass flow dependant detectors usually destroy the sample, and the signal is related to the rate at which solute molecules enter the detector. The response of a mass flow dependant detector is unaffected by make-up gas.

Page 37: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCUniversal Detectors

• Thermal Conductivity Detector (TCD)– Principle: heat conduction by gas (dependent

on gas molecular weight)– All gases detected, but variable sensitivity.– Not very sensitive

• Flame Ionization Detector (FID)– Principle: GC effluent burned in hydrogen flame

producing ions measured by electrode– Good sensitivity– Hydrocarbons are detected with reduced

response for other functional groups– Quantitation is good and possible without

standards

Page 38: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCDetectors - Classes

• Universal Detectors– Should detect wide range of compounds– Response should be based on mass or moles of

analyte (so can quantitate with surrogate standards)

– Used for comprehensive understanding of samples

• Selective Detectors– Should detect only specific types of compounds– Best for selected analytes in complex samples

• Multi-dimensional Detectors (selective and universal – MS is most common)

Page 39: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GCSelective Detectors

• Based on Gas Ionization– Electron Capture Detector (ECD)

• Uses β emitter to produce electrons that cause current

• Very sensitive detector• Selective for compounds with electronegative groups

(e.g. halogens, nitrogroups, etc.)

– Photoionization Detector (PID)• Uses UV light to photoionize compounds (M + hν →

M+ + e-)• Sensitive to unsaturated compounds

• Element Specific Detectors (common for S, P, and N)

Page 40: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Instrumentation

Page 41: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Column temperature

• For precise work, column temperature must be controlled to within tenths of a degree. The optimum column temperature is dependent upon the boiling point of the sample. As a rule of thumb, a temperature slightly above the average boiling point of the sample results in an elution time of 2 - 30 minutes. Minimal temperatures give good resolution, but increase elution times. If a sample has a wide boiling range, then temperature programming can be useful. The column temperature is increased (either continuously or in steps) as separation proceeds.

Page 42: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Mobile Phase (gas)

Page 43: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Under the hood

Page 44: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Column and Oven

Page 45: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Typical GC (dual column)

Page 46: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detection• The effluent from the column is mixed with hydrogen and air, and ignited.

Organic compounds burning in the flame produce ions and electrons which can conduct electricity through the flame. A large electrical potential is applied at the burner tip, and a collector electrode is located above the flame. The current resulting from the pyrolysis of any organic compounds is measured. FIDs are mass sensitive rather than concentration sensitive; this gives the advantage that changes in mobile phase flow rate do not affect the detector's response. The FID is a useful general detector for the analysis of organic compounds; it has high sensitivity, a large linear response range, and low noise. It is also robust and easy to use, but unfortunately, it destroys the sample.

Page 47: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detector types are:

1) FID ( flame ionization detector )

2) TCD ( thermal conductivity detector ) 3) ECD ( electron capture detector )4) FPD ( flame photometric detector )

5) HID ( helium ionization detector ) 6) NPD ( nitrogen-phosphorus detector ) 7) PID ( photo ionization detector ) 8) TID ( thermionic ionization detector ) 9) CCD ( catalytic combustion detector )

10) NPD/DELCD ( combination NPD and dry electrolytic conductivity detector ) 11) FID/DELCD ( combination FID and dry electrolytic conductivity detector ) 12) FID/FPD ( combination FID and FPD ) 13) Dual FPD ( dual wavelength FPD for both sulfur and phosphorus ) 14) FID dual FPD ( dual FPD plus FID combination )

Page 48: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.
Page 49: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors continued• ECD (electron capture detector). Uses Ni-63 as a radiation

source to cause ionization of the substance using N as the carrier gas. Detector is sensitive to functional groups containing electronegative species such as halogens, quinones, peroxides, and nitro groups. Hence, very good detector for environmental analysis where pesticides need to be measured.

• AED (atomic emission detector). It is a AA unit using MIP that accepts the output from the GC.

• Mass spectrometer

Page 50: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Different GC detectors

Page 51: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Selective Detectors• Specific for specific types of

chemicals (i.e., nitrogen-phosphorus detector, NPD)• Measures ions produced from

eluting N or P compounds, but generates e– from a heated surface (not flame) that combine with electronegative elements to form negative ions

• Good selectivity, good LOD, but must periodically change heated material

Page 52: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Electron Capture• Detects electronegative atoms or

groups (Cl, -NO2) and also polynuclear aromatics and conjugated carbonyl compounds

• These groups capture e– that are produced from nuclear radiation from 3H

• Carrier gas hit by these e– can release secondary e–, which will be absorbed by analyte

• Has good LOD, but a narrow linear range and requires radioactive source

Page 53: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Mass Spectrometry• Detects and measures by converting eluting analytes into

gas phase ions (forming a molecular ion or fragmenting analyte [and ionizing]. Compare patterns of ions and fragments to known values. Intensity relates to amounts.

• A portion of analytes is converted to anions via electron impact or chemical impact (softer, less fragments).

• Gas phase ions are separated by mass/charge ratio using a quadrupole mass analyzer. (Uses four parallel rods with well-defined potentials so that only certain mass/charge species may pass [sorted])

Page 54: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Mass Spectrometry

• A Mass spectrum plots intensity of mass/charge vs. time—could be set a certain retention time in a GC.

• Mass chromatogram

Page 55: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Thermal Conductivity Detector (TCD)

IntroductionA TCD detector consists of an electrically-heated wire or thermistor. The temperature of the sensing element depends on the thermal conductivity of the gas flowing around it. Changes in thermal conductivity, such as when organic molecules displace some of the carrier gas, cause a temperature rise in the element which is sensed as a change in resistance. The TCD is not as sensitive as other detectors but it is non-specific and non-destructive.

InstrumentationTwo pairs of TCDs are used in gas chromatographs. One pair is placed in the column effluent to detect the separated components as they leave the column, and another pair is placed before the injector or in a separate reference column. The resistances of the two sets of pairs are then arranged in a bridge circuit.The bridge circuit allows amplification of resistance changes due to analytes passing over the sample thermoconductors and does not amplify changes in resistance that both sets of detectors produce due to flow rate fluctuations, etc.

Page 56: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Schematic of a bridge circuit for TCD detectionTwo filament in one cell ( reference side ) --- carrier gas onlyThe other cell ( sample side ) --- carrier plus sample flowing

1. Universal2. Used primarily for gas analysis3. Sensitive few ppm

Page 57: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Thermal Conductivity

• TCD can be used for organic and inorganic analytes.• The key aspect is the ability of the carrier gas and the

analytes to change the conductivity of a wire filament, which will vary with different analytes.

• The carrier gas should have different thermal conductivity of analytes.

• TCD is a non-destructive type of detection that uses a Wheatstone bridge style. Downsides are the response to impurities, leakage in air, and poor response to LOD.

Page 58: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Wheatstone Bridge TCD

Most common carrier

Page 59: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detectors: Flame Ionization

• FID uses fuel mixed with carrier and organic analyte. Analyte forms ions in the flame.

• Cations from the flame are gathered by the negative electrode—produces a current.

• Advantage:inorganics do not respond (He carrier gas), so the low background signal allows for LOD 100- to 1000-fold lower then TCD.

• Disadvantage: destructive

Page 60: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Flame Ionization Detector

Introduction

The flame ionization detector (FID) is the most sensitive gas chromatographic detector for hydrocarbons such as butane or hexane. With a linear range for 6 or 7 orders of magnitude (106 to 107) and limits of detection in the low picogram or femtogram range, the FID is the gas chromatographic detector for volatile hydrocarbons and many carbon containing

compounds.

FID

Responds to all organic compounds except for formic acid. Response greatest with hydrocarbons and decreases with substitution. Except for vapor of elements in Groups I and II, does not respond to inorganic compounds. Sensitivity high due to low noise level. Insensitivity to water, the permanent gases, and inorganic compounds simplifies the resolution of components in analysis of aqueous extracts and in air pollution studies.

Page 61: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Flame Ionization Detector

Consists of a stainless steel burner assembly installed in the detector compartment and a electrometer system in a separate unit adjacent to the gas chromatograph

Often it is installed in the tandem with the thermal conductivity cell

Effluent form the column enters burner base through millipore filters which remove contaminates

Hydrogen mixed with gas stream at bottom of jet and air or oxygen is supplied axially around the jet.

Hydrogen flame burns at the tip, which also functions as the cathode and is insulated form the body by a ceramic seal

Collector electrode is above the burner tip and is made of platinum

Page 62: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

An FID consists of a hydrogen/air flame and a collector plate. The effluent

from the GC column passes through the flame, which breaks down organic

molecules and produces ions. The ions are collected on a biased electrode

and produce an electrical signal. The FID is extremely sensitive with a large

dynamic range, its only disadvantage is that it destroys the sample.

FIDs are normally heated independently of the chromatographic oven.

Heating is necessary in order to prevent condensation of water generated by

the flame and also to prevent any hold-up of solutes as they pass from the

column to the flame. With the flame extinguished, the column end should be

passed up through the jet and then lightly held in position by slightly

tightening the coupling. Gradually draw the column end back into detector

jet until it is approximately 1 - 2 mm below the jet tip. Then tighten the

coupling to retain it in position. Do not over tighten couplings on capillary

columns.

Page 63: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Mechanism

The effluent from the column is mixed with hydrogen and air and then ignited

electrically at a small metal jet. Most organic compounds produce ions and electrons

that can conduct electricity through the flame. There is an electrode above the flame to

collect the ions formed at a hydrogen/air flame. The number of ions hitting the

collector is measured and a signal is generated.In series with flame gases is a selection of resistors 107 to 1010 ohms. Vibrating reed electrometer used to provide sensitivities up to 5 × 1013 Amps. Carbon counting device that produces a current proportional to number of ions or electrons formed in the flamed gases.

The organic molecules undergo a series of reactions including thermal fragmentation,

chemi-ionization, ion molecule and free radical reactions to produce charged-species.

The amount of ions produced is roughly proportional to the number of reduced carbon

atoms present in the flame and hence the number of molecules. Because the flame

ionization detector responds to the number of carbon atoms entering the detector per

unit of time, it is a mass-sensitive, rather than a concentration-sensitive device. As a

consequence, this detector has the advantage that changes in flow rate of the mobile

phase has little effects on detector response.

Page 64: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Normal Combustion: i.e. burn methane in air and get carbon dioxide and water vapor...

CH4 + O2 CO2 + H2O

or:

CH4 + 3O2 CO2 + 2H2O

Flame Ionization: during combustion, a uniform proportion (about 0.0002%) of the molecules in this reaction do this instead: (simplified for clarity)

CH4 + O2 C++ H2O + e- CO2 + H2O

or:

CH4 + 3O2 C++ O2 + 2H2O + e- CO2 + 2H2O

These oppositely-charged, intermediate products can then be detected by the FID:

Page 65: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Limitations

Molecules that contained only carbon and hydrogen respond best in this

detector but the presence of "heteroatoms" in a molecule, such as

oxygen, decreases the detector's response. For instance, the FID's

methane response (CH4) is fabulous but formaldehyde's (CH2O) is quite

poor. Therefore, highly oxygenated molecules or sulfides might best be

detected using another detector instead of the FID. Sulfides

determination by the flame photometric detector and aldehydes and

ketones analyzed with the photoionization detector are alternatives to

the use of the FID for those molecules.

Page 66: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Functional group, such as carbonyl, alcohol, halogen, and amine, yield fewer ions or none at all in a flame. In addition, the detector is insensitive toward noncombustible gases such as H2O, CO2, SO2 and NOx.

Selectivity: Compounds with C-H bonds. A poor response for some non-hydrogen containing organics (e.g., hexachlorobenzene).

Sensitivity: 0.1 ~ 10 ng

Linear range: 105 ~ 107

Gases: Combustion - hydrogen and air; Makeup - helium or nitrogen

Temperature: 250-300 °C; 400-450 °C for high temperature analyses

Page 67: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Detector Construction

FID is constructed of a small volume chamber into which the gas chromatograph's capillary column in directly plumbed.

Usually the small diameter capillary is fitted directly into the bottom of the detector's flame jet. The gaseous eluents from the column are mixed with separately plumbed in hydrogen and air and all are burned on the jet's tip. After the fuel (H2) and oxidant (O2 in air) are begun, the flame is lit using a electronic ignite, actually an electrically heated filament that is turned on only to light the flame.

The charged particles created in that combustion process create a current between the detector's electrodes. One electrode is actually the metallic jet itself, another is close by and above the jet. The gaseous products leave the detector chamber via the exhaust. The detector housing is heated so that gases produced by the combustion (mainly water) do not condense in the detector before leaving the detector chimney.

Page 68: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Flame Ionization Detector

FIDTCD

View of TCD and FID of HP5890 GC

Page 69: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Flame Ionization Detector

Page 70: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Makeup Gases

The total volume of gas in the FID that yields the most sensitive and widest linear response is not the same volume of gas when the column effluent flow (~ 1 mL/min) and hydrogen and air flows are flowing; these gases' total flow into the detector is too small. Another way to say this is that the optimum column flow to maintain the best chromatography and the best fuel and oxidant flows for the best flame conditions--all added together--don't create the best gas flow for the FID detector's design. This means that to maintain the best analytical conditions, additional gas must be constantly flowed into the detector. This gas makes up the additional needed gas flow and so is termed makeup gas. Since the gas needs to be inert so that its addition doesn't upset the fuel and oxidant balance and since it needs to be added in relatively large amounts (~30+ ml/min in some detector designs) nitrogen is usually the gas of choice. Helium would work also but is a nonrenewable resource and more expensive. All gas flows are controlled by adjustable gas regulators.

Page 71: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Electron Capture Detector (ECD)

The ECD uses a radioactive emitter (electrons) to ionize some of the carrier gas and produce a current between a biased pair of electrodes. When organic molecules that contain electronegative functional groups, such as halogens, phosphorous, and nitro groups pass by the detector, they capture some of the electrons and reduce the current measured between the electrodes. The ECD is as sensitive as the FID but has a limited dynamic range and finds its greatest application in analysis of halogenated compounds.

Schematic of an ECD

Page 72: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

ECD

Selective in its response and highly sensitive

Hewlett Packard makes one with a detection limit of less than 8 fg/sec for lindane

Sensitive toward molecules with electronegative functional groups (halogens, peroxides, quinones, nitro groups)

Insensitive towards amines, alcohols and hydrocarbons

A leak test of an ECD containing nickel-63 (63Ni) must be performed at intervals not to exceed six months. The test must be performed in accordance with the manufacturer's instructions, or by wiping the gas intake and outlet surfaces.

NOTE: Never attempt to directly wipe the inner surface of the component containing the radioactive material. This might cause the ECD to fail and will contaminate the ECD, the gas chromatograph and the surrounding area.

Never open the detector cell for any reason.

Page 73: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Nitrogen Phosphorous DetectorSpecific: sample must contain nitrogen or phosphorous

Destructive

LOD : 0.4 pg N / sec 0.2 pg P / sec

Linear range : ~ 104

Mode of operation: essentially a modified FID

Active element acts to block undesired species

Page 74: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Flame Photometric Detector

The determination of sulfur or phosphorus containing compounds is the job of the flame photometric detector (FPD). This device uses the chemiluminescent reactions of these compounds in a hydrogen/air flame as a source of analytical information that is relatively specific for substances containing these two kinds of atoms. The emitting species for sulfur compounds is excited S2. The lambda max for emission of excited S2 is approximately 394 nm. The emitter for phosphorus compounds in the flame is excited HPO (lambda max = doublet 510-526 nm). In order to selectively detect one or the other family of compounds as it elutes from the GC column, an interference filter is used between the flame and the photomultiplier tube (PMT) to isolate the appropriate emission band. The drawback here being that the filter must be exchanged between chromatographic runs if the other family of compounds is to be detected.

InstrumentationIn addition to the instrumental requirements for 1) a combustion chamber to house the flame, 2) gas lines for hydrogen (fuel) and air (oxidant), and 3) an exhaust chimney to remove combustion products, the final component necessary for this instrument is a thermal (bandpass) filter to isolate only the visible and UV radiation emitted by the flame. Without this the large amounts of infrared radiation emitted by the flame's combustion reaction would heat up the PMT and increase its background signal. The PMT is also physically insulated from the combustion chamber by using poorly (thermally) conducting metals to attach the PMT housing, filters, etc.The physical arrangement of these components is as follows: flame (combustion) chamber with exhaust, permanent thermal filter (two IR filters in some commercial designs), a removable phosphorus or sulfur selective filter, and finally the PMT.

Page 75: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Schematic of a gas chromatographic flame photometric detector

Specific: phosphorous or sulfurDestructiveLOD: 20 pg S /sec, 0.9 pg P / secLinbear range: ~104 P, ~103 S

Page 76: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Photoionization Detector

IntroductionThe reason to use more than one kind of detector for gas chromatography is to achieve selective and/or highly sensitive detection of specific compounds encountered in particular chromatographic analyses. The selective determination of aromatic hydrocarbons or organo-heteroatom species is the job of the photoionization detector (PID). This device uses ultraviolet light as a means of ionizing an analyte exiting from a GC column. The ions produced by this process are collected by electrodes. The current generated is therefore a measure of the analyte concentration. TheoryIf the energy of an incoming photon is high enough (and the molecule is quantum mechanically "allowed" to absorb the photon) photo-excitation can occur to such an extent that an electron is completely removed from its molecular orbital, i.e. ionization.A Photoionization Reaction

Page 77: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

If the amount of ionization is reproducible for a given compound, pressure, and light source then the current collected at the PID's reaction cell electrodes is reproducibly proportional to the amount of that compound entering the cell. The reason why the compounds that are routinely analyzed are either aromatic hydrocarbons or heteroatom containing compounds (like organosulfur or organophosphorus species) is because these species have ionization potentials (IP) that are within reach of commercially available UV lamps. The available lamp energies range from 8.3 to 11.7 ev, that is, lambda max ranging from 150 nm to 106 nm. Although most PIDs have only one lamp, lamps in the PID are exchanged depending on the compound selectivity required in the analysis.

Page 78: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Selective detection using a PIDHere is an example of selective PID detection: Benzene's boiling point is 80.1 degrees C and its IP is 9.24 ev. (Check the CRC Handbook 56th ed. page E-74 for IPs of common molecules.) This compound would respond in a PID with a UV lamp of 9.5 ev (commercially available) because this energy is higher than benzene's IP (9.24). Isopropyl alcohol has a similar boiling point (82.5 degrees C) and these two compounds might elute relatively close together in normal temperature programmed gas chromatography, especially if a fast temperature ramp were used. However, since isopropyl alcohol's IP is 10.15 ev this compound would be invisible or show very poor response in that PID, and therefore the detector would respond to one compound but not the other.

Page 79: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

InstrumentationSince only a small (but basically unknown) fraction of the analyte molecules are actually ionized in the PID chamber, this is considered to be a nondestructive GC detector. Therefore, the exhaust port of the PID can be connected to another detector in series with the PID. In this way data from two different detectors can be taken simultaneously, and selective detection of PID responsive compounds augmented by response from, say, an FID or ECD. The major challenge here is to make the design of the ionization chamber and the downstream connections to the second detector as low volume as possible (read small diameter) so that peaks that have been separated by the GC column do not broaden out before detection.

Specific: compounds ionized by UVLOD: ~ 2 pg Carbon / secLinear range : 107

Page 80: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Atomic-Emission Detector (AED)

This detector, while quite expensive compared to other commercially available GC detectors, is an extremely powerful alternative. For instance, Instead of measuring simple gas phase (carbon containing) ions created in a flame as with the flame ionization detector, or the change in background current because of electronegative element capture of thermal electrons as with the electron capture detector, the AED has a much wider applicability because it is based on the detection of atomic emissions.The strength of the AED lies in the detector's ability to simultaneously determine the atomic emissions of many of the elements in analytes that elute from a GC capillary column (called eluants or solutes in some books). As eluants come off the capillary column they are fed into a microwave powered plasma (or discharge) cavity where the compounds are destroyed and their atoms are excited by the energy of the plasma. The light that is emitted by the excited particles is separated into individual lines via a photodiode array. The associated computer then sorts out the individual emission lines and can produce chromatograms made up of peaks from eluants that contain only a specific element.

Page 81: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Instrumentation

The components of the AED include 1) an interface for the incoming capillary GC column to the microwave induced plasma chamber, 2) the microwave chamber itself, 3) a cooling system for that chamber, 4) a diffraction grating and associated optics to focus then disperse the spectral atomic lines, and 5) a position adjustable photodiode array interfaced to a computer. The microwave cavity cooling is required because much of the energy focused into the cavity is converted to heat.

Schematic of a gas chromatographic atomic emission detector

Page 82: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

PDD (pulsed discharge detector)•

The VICI PDD (pulsed discharge detector) utilizes a stable, low powered, pulsed DC discharge in helium as an ionization source. Performance is equal to or better than detectors with conventional radioactive sources. In the electron capture mode, the PDD is a selective detector for monitoring high electron affinity compounds such as freons, chlorinated pesticides, and other halogen compounds. For this type of compound, the minimum detectable quantity (MDQ) is at the femtogram (10-15) or picogram (10-12) level. The PDD is similar in sensitivity and response characteristics to a conventional radioactive ECD, and can be operated at temperatures up to 400°C. For operation in this mode, He and CH4 are introduced just upstream from the column exit. In the helium photoionization mode, the PDD is a universal, non-destructive, high sensitivity detector. The response to both inorganic and organic compounds is linear over a wide range. Response to fixed gases is positive (increase in standing current), with an MDQ in the low ppb range. The PDD in helium photoionization mode is an excellent replacement for flame ionization detectors in petrochemical or refinery environments, where the flame and use of hydrogen can be problematic. In addition, when the helium discharge gas is doped with a suitable noble gas, such as argon, krypton, or xenon (depending on the desired cutoff point), the PDD can function as a specific photoionization detector for selective determination of aliphatics, aromatics, amines, as well as other species. (Click here for an ionization potential chart in .pdf format.)

Page 83: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

PDD detector

Page 84: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Various configurations:

• Column: capillary• Autosampler• Computer controlled for data acquisition and analysis

of peaks. Some come with their own compressor for the air supply. Sensitivity is very important because of detection limits. More expensive system will use a MS for the detector to do GC-MS. Can do headspace analysis for volatiles in water etc.

Page 85: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Companies

• ThermoQuest• Agilent (Hewlett Packard)• Perkin-Elmer• Buck Scientific• Thermo Nicolet

Page 86: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Gas Chromatography Instrumentation

3.) Packed Columns The major advantage and use is for large-scale or

preparative purification

Industrial scale purification maybe in the kilogram or greater range

500 L chromatography column

Oil refinery – separates fractions of oil for petroleum products

Page 87: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Qualitative analysis : tR

Standard ---- methanol, MEK(tR ), toluene Unknown ----- same tR (X) Conclude (X) = MEK

X

methanol

MEK

toluene

tR

Retention time limitations

tR changes with flow rate, column temperature, liquid phase, column history, sample size

*** WARNING identical retention times do not confirm peak identity

Page 88: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Spiking

Step 1 Peak X --- toluene ? Step 2 Toluene added to sample Step 3 Peak X identified as toluene

X

methanol

MEK

toluene

tR

Toluene added to sample

Page 89: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Pesticides• Analysis of pesticide residues in soil, water, and food is crucial for

maintaining safe levels in the environment. The PDD in the ECD mode is highly selective for monitoring electron capturing compounds such as chlorinated pesticides and other halogens. This chromatogram illustrates the sensitivity of the non-radioactive PDECD for such compounds. Sample: Pesticide calibration mix Detector mode: Electron capture Detector temp: 330°C Column: 25 m x 0.32 mm x 25 µm, HP-5 Column temp: 150°C to 300°C at 10°C/min Sample volume: 1 µL, 10:1 split Discharge gas: Helium, 30 mL/min Dopant gas: 5% methane in helium, 2.4 mL/min Attenuation: 1

Page 90: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Pesticide separations

Retention time (sec)

Page 91: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Headspace gas chromatography analysis

• Headspace GC (HSGC) analysis employs a specialized sampling and sample introduction technique, making use of the equilibrium established between the volatile components of a liquid or solid phase and the gaseous / vapor phase in a sealed sample container. Aliquots of the gaseous phase are sampled for analysis.

Page 92: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Headspace sampling

Headspace sampling

Page 93: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

HSGC

• Examples of HSGC are the forensic analysis of blood and urine alcohol levels, quality and production control of diesel fuel and beer constituents. Aromatic flavors and trace volatiles in foods and soft-drinks are also readily analyzed. and HSGC analysis of volatile free fatty acids produced by bacteria, particularly anaerobic bacteria, enables a fingerprint of the particular microorganisms to be obtained, which assists in the identification of the bacteria.

Page 94: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Food analysis

• Analysis of foods is concerned with the assay of lipids, proteins, carbohydrates, preservatives, flavours, colorants and texture modifiers, and also vitamins, steroids, drugs and pesticide residues and trace elements. Most of the components are non-volatile and although HPLC is now used routinely for much food analysis, GC is still frequently used. For examples, derivatization of lipids and fatty acid to their methyl esters(FAMEs), of proteins by acid hydrolysis followed by esterification (N-propyl esters) and of carbohydrates by silylation to produce volatile samples suitable for GC analysis.

Page 95: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC Food• GC quality control analysis of food products can confirm the presence and

quantities of the analytes For example, fruits, fruit derived foodstuffs, vegetables and soft drinks, tea and coffee, were analyzed for their polybasic and hydroxy acid contents (citric, maleic acids) as TMS derivatives.

• All food and beverage products on sale today must be carefully assayed for contamination with pesticides, herbicides and many other materials that are considered a health risk. The analysis of food involves separating and identifying very complex mixtures, the components of which are present at very low concentrations. GC is the ideal technique for use in food and beverage assays and tests. Furthermore, the origin of many herbs and spices can often be identified from the peak pattern of the chromatograms from their head space analysis.

Page 96: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Food and Cancer

• Chemicals that can cause cancer have a wide variety of molecular structures and include hydrocarbons, amines, certain drugs, some metals and even some substances occurring naturally in plants and molds. In this way, many nitrosamines have carcinogenic properties and these are produced in a number of ways such as cigarette smoke. GC can be used to identify these nitro-compounds in trace quantities.

Page 97: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Drugs• There are still numerous GC applications involving both quantitative and

qualitative identification of the active components and possible contaminants, adulterants or characteristic features which may indicate the source of the particular sample. Forensic analysis frequently users GC to characterize drugs of abuse, in some cases the characteristic chromatographic fingerprint gives an indication of the source of manufacture of the sample or worldwide source of a vegetable material such as cannabis.

• Analytical procedures, chromatographic methods and retention data are published for over 600 drugs, poisons and metabolites. These data are extremely useful for forensic work and in hospital pathology laboratories to assist the identification of drugs.

Page 98: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Pyrolysis gas chromatography

• Pyrolysis GC (PGC) is used principally for the identification of non-volatile materials, such as plastics, natural and synthetic polymers, drugs and some microbiological materials. The thermal dissociation and fragmentation of the sample produces a chromatogram which is a fingerprint for that sample. The small molecules produced in the pyrolysis reaction are frequently identified using a GC-MS system and information on molecular structure for identification is also obtained.

Page 99: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Metal chelates and inorganic materials

• Although inorganic compounds are generally non-volatile, GC analysis can be achieved by converting the metal species into volatile derivatives. Only some metal hydrides and chlorides have sufficient volatility for GC. Organometallics other than chelates, which can be analyzed directly, include boranes, silanes, germanes, organotin and lead compounds.

Page 100: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

Environmental analysis

• Environmental pollution is an age-old trademark of man and in recent years as technology has progressed, populations have increased and standards of living have improved. So the demands on the environment have increased, with all the attendant problems for the ecosystems. Combustion of fossil fuel, disposal of waste materials and products, treatment of crops with pesticides and herbicides have all contributed to the problems. Technological developments have enabled man to study these problems and realize that even trace quantities of pollutants can gave detrimental effects on health and on the stability of the environment. There is a vast amount of literature on the use of GC for studying a wide variety of these problems.

Page 101: Gas Chromatography. Gas Chromatography (GC) This method depends upon the solubility and boiling points of organic liquids in order to separate them from.

GC application• Every year many new substances are synthesized that differ radically from

the natural products that exist in biosystems. The Environmental Protection Agency is empowered to control water pollution and the production, use and disposal of toxic chemicals.It follows that detailed studies must be made of their effect on the environment and their method of movement through the ecosystem. Many of the compounds are not biodegradable and will thus progressively pollute the environment. There are a number of tragic examples of which DDT (dichlorodiphenyltrichloroethane) and the PCBs (polychlorinated biphenyls) are well known instances. The materials of interest are present in environmental samples at very low concentrations and are often to be found among a myriad of other compounds from which they must be separated and identified. It follows that GC, with its inherent high sensitivity and high separating power, is one of the more commonly used techniques in the analysis of environmental samples.