FPD Calibration Studies

5
Michael Strang FTG Meeting 2-Sep-03 FPD Calibration Studies Michael Strang University of Texas at Arlington

description

FPD Calibration Studies. Michael Strang University of Texas at Arlington. First Appearance of Problem. This is a recent calibration with 0x400 gain on board. Notice the dropping of the mean at the edge of SIFT chips happening on two boards in different cassettes. Old data looking for problem. - PowerPoint PPT Presentation

Transcript of FPD Calibration Studies

Page 1: FPD Calibration Studies

Michael StrangFTG Meeting 2-Sep-03

FPD Calibration StudiesFPD Calibration Studies

Michael Strang

University of Texas at Arlington

Page 2: FPD Calibration Studies

Michael StrangFTG Meeting 2-Sep-03

First Appearance of ProblemFirst Appearance of Problem14A6

0

10

20

30

40

50

60

70

80

90

100

1 13 25 37 49 61 73 85 97 109 121 133 145 157 169 181 193 205 217 229 241 253 265 277 289 301 313 325 337 349 361 373 385 397 409 421 433 445 457 469 481 493 505

channel

AF

E c

ou

nts

rms

mean

14A0

0

10

20

30

40

50

60

70

80

90

100

1 13 25 37 49 61 73 85 97 109 121 133 145 157 169 181 193 205 217 229 241 253 265 277 289 301 313 325 337 349 361 373 385 397 409 421 433 445 457 469 481 493 505

channel

AD

C c

ou

nt

rms

mean

This is a recent calibration with 0x400 gain on board. Notice the dropping of the mean at the edge of SIFT chips happening on two boards in different cassettes

Page 3: FPD Calibration Studies

Michael StrangFTG Meeting 2-Sep-03

Old data looking for problemOld data looking for problem14A6

0

10

20

30

40

50

60

70

80

90

100

1 13 25 37 49 61 73 85 97 109 121 133 145 157 169 181 193 205 217 229 241 253 265 277 289 301 313 325 337 349 361 373 385 397 409 421 433 445 457 469 481 493 505

channel

AF

E c

ou

nts

rms

mean

This is an old calibration (around April) with 0x400 gain on board. The evidence of this structure is not as clear if it is present at all.

Page 4: FPD Calibration Studies

Michael StrangFTG Meeting 2-Sep-03

Trying to isolate problemTrying to isolate problem14A6

0

10

20

30

40

50

60

70

80

90

100

1 13 25 37 49 61 73 85 97 109 121 133 145 157 169 181 193 205 217 229 241 253 265 277 289 301 313 325 337 349 361 373 385 397 409 421 433 445 457 469 481 493 505

channel

AF

E c

ou

nts

rms

mean

During recent access last week, we unplugged the bayonet, unplugged the cables from the tunnel going to the TPP and isolated the TPP from contact with anything else (flex cables still connected) and structure is very clear.

Page 5: FPD Calibration Studies

Michael StrangFTG Meeting 2-Sep-03

Looked at CFT boards as wellLooked at CFT boards as wellCrate 53, Slot 13, HDI 7

0

10

20

30

40

50

60

70

80

90

100

1 13 25 37 49 61 73 85 97 109 121 133 145 157 169 181 193 205 217 229 241 253 265 277 289 301 313 325 337 349 361 373 385 397 409 421 433 445 457 469 481 493 505

channel

AF

E c

ou

nts

rms

mean

For comparison, looked at a random board from a different crate (53). Surprisingly, this same structure is present in parts of the board so while this problem seems enhanced in the FPD boards, it appears to be present in other boards as well.