Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer...

45
INTRODUCTION Page 1 CHAPTER-1 INTRODUCTION

Transcript of Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer...

Page 1: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 1

CHAPTER-1

INTRODUCTION

Page 2: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 2

Section (i): A brief account on analytical chemistry and its applications

to pharmaceutical analysis

Analytical chemistry seeks ever improved means of measuring the chemical

composition of natural and artificially occurring materials in our surroundings and

environment as well as within ourselves, in our tissues and body fluids and also in

pharmaceutical dosage forms. The techniques of this science involve separation,

identification and determination of the relative amounts of the components in simple or

complex systems. The applications of analytical chemistry are extended to all the

branches of science. Recent advances in analytical chemistry are the significant

developments in the field of chemical analysis, which are versatile in nature adopting

newer techniques. The present trend includes the employment of spectrophotometry,

chromatography, polarography, neutron activation analysis, atomic absorption

spectrophotometry, ICP-atomic emission spectrometry etc. Among these,

spectrophotometric methods occupy special position due to their ease in application,

simplicity and accuracy in the determination of trace amounts of chemical compounds,

i.e. drugs, metals and non-metals.

Pharmaceutical chemistry [1-5]

is the science that makes use of general law of

chemistry to study drugs and their preparation, chemical nature, composition, structure,

influence on an organism, the method of quality control and the conditions of their

storage .etc. All around the world, a large number of varieties of organic compounds are

being synthesized daily and some of them are very useful to cure diseases, pain to get

relief from abnormalities of health in humans or animals and these chemical compounds

Page 3: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 3

are known as pharmaceutical compounds or drugs. A pharmaceutical drug also referred

as medicine, medication or medicament, can be loosely defined as any chemical

substance intended for use in the medical diagnosis, cure treatment or prevention of

disease. [6-7]

Drug is a substance that affects the function of living cells and used in medicine

to diagnose, cure, prevent the occurrence of diseases and disorders and prolong the life of

patients with incurable conditions. Since 1900, the availability of new and more effective

drugs such as antibiotics, which fight bacterial infections, antiviral, which fight viral

infections and vaccines, which prevent diseases caused by bacterial and viruses has

increased the average humans’ life span from about 60 years to about 75 years.

Pharmaceutical drug analysis provide information on the identity, purity, contents

and stability of starting materials, excipients, active pharmaceutical ingredients(APIs)

and also the secondary pharmaceutical products i.e. the dosage forms having either

single or multi-component formulated product.[8]

The quality of the drug product may

deviate from the standard required, but in carrying out an analysis one can confirm that

whether the quality of the product is of the required standard or not [9]

. The fundamental

reasons for the “analysis of drug substances” are perhaps due to the tremendous growth in

the progress of medicinal chemistry towards achieving one ultimate objective which is to

obtain “better drugs for a better world”[10]

.Drug analysis is important in several phases of

drug development, such as formulation, stability studies, and dissolution studies and in

quality control. The importance of reliable analytical methods for the drug determination

in a fast, inexpensive, sensitive and selective way is thus evident. In this context, the

objective of this thesis is to present the main advances in the development and validation

Page 4: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 4

of analytical methods for the determination of drugs using spectrophotometric

techniques.

QUALITY CONTROL AND ASSURANCE [11]

Quality is important in every product or service but it is vital in medicine as it

involves life. Unlike ordinary consumer goods, there can be no second quality in drugs.

Quality control is a concept, which strives to produce a perfect product by series of

measures designed to prevent and eliminate errors at different stages of production. The

word quality is derived from the Latin “Qualitas”, which means, incidentally, only the

‘nature’ and ‘inherent characteristics’ of a thing. Quality Control (QC) and Quality

Assurance (QA) are integral part of Analytical Research. "The primary aim of the new

analytical method is to produce correct results, not by chance, but at all times and the use

of quality control methods to achieve and demonstrate this." Two yardsticks are used in

QA viz. Accuracy and Precision. QA is thus by choosing an appropriate specific,

sensitive and accurate procedure with precision which involves consideration of several

other practical aspects such as the speed, economy and the skill required. Such

procedures assure both accuracy and precision and should be used by the laboratories

aiming QA and QC. In research laboratories, no compromise can be made on QA, as

today’s novel analytical method is the trendsetter for tomorrow’s routine analysis.

Internal and external quality control measures should be adopted and recorded by each

laboratory. Internal quality control measures by using controls, replications and random

sample check determine the drifts occurring in the daily tests. External quality control

methods measure periodic analysis of unknown samples from reputed QC programmes.

The variance index scores are the measure of the performance.

Page 5: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 5

QA & QC methods developed follow standard internal operating procedures directed

toward assuring the quality, safety, purity and effectiveness of the drug supply.

In pharmaceutical analysis, a variety of methods have been used for the

estimation of the drugs in bulk and in their formulations. The methods for the estimation

of drugs are broadly divided in to physical, chemical, physicochemical and biological.

Among these four methods, physicochemical and physical methods are the most used

methods for the analysis of drugs. Physical methods of analysis consider the study of

physical properties of substance, i.e. melting, freezing and boiling points, solubility,

transparency or degree of turbidity and specific gravity. Physicochemical methods are

used to study the physical phenomena that occur as a result of chemical reactions. The

most important physicochemical methods are optical and chromatographic methods. The

optical methods are refractometry, polarimetry, photometry (colorimetry),

spectrophotometry (UV, Visible and IR regions), Flow Injection-spectrofluorimetry and

nephelometry. The chromatographic methods are column, paper, thin layer, gas-liquid,

HPLC, UPLC and RP-HPLC. Recently hyphenated techniques like HPLC with different

detectors, LC-MS, GC-MS, and LC-MS/MS are introduced for the analysis of the drugs

for achieving high sensitivity and selectivity with good precision, accuracy and

reproducibility.

Analytical method development and validation play important roles in the

discovery, development and manufacture of pharmaceuticals. The official test methods

that result from these processes are used by quality control laboratories to ensure the

identity, purity, potency and performance of the drug products. The selection of

analytical method may be based on one or more of the following design criteria,

Page 6: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 6

accuracy, precision, sensitivity, selectivity, robustness, ruggedness, scale of operation,

analysis time, availability of equipment and cost.

Many organic compounds absorb in the ultra-violet region of the spectrum and

pre-treatment involves only separation of interferences. Some elements in the periodic

table absorb strongly in the visible or UVregions, at least in certain oxidation states and

preliminary steps may involve redox reactions as well as separations [12]

. Development of

absorption methods by means of inorganic reagents is occasionally possible. Many of the

colored complexes formed by metal ions with organic reagents are metal-chelates,

offering most impressive variety of spectrophotometric methods, and they are especially

useful in the field of trace analysis in biological fluids and in pharmaceutical dosage

forms. In some regards, the low aqueous solubility of many of the metal chelate

compounds is disadvantageous, but on the other hand, extraction of metals into non

aqueous solvents by means of chelating agents may lead to very powerful analytical

methods. Reasonable absorbance values are generally obtained with chloroform solutions

whose metal concentrations are in the order of a few µg mL-1

. The solvents used in

spectrophotometric methods poses a problem in some regions of spectrum. There is no

solvent which is transparent throughout the infrared region. However, in UV or Visible

region, most of the organic solvents have UV-cutoff points below 210 nm and water is an

excellent solvent in that region, due to transparency throughout the spectrum.

Analytical methods used in analysis

Analytical chemistry is divided into two types such as Classical and Instrumental.

Modern analytical chemistry is dominated by sophisticated instruments. The roots of

analytical chemistry and some of the principles used in modern instruments are from

Page 7: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 7

traditional techniques many of which are still used today. These techniques also tend to

form the back bone of most undergraduate chemistry educational labs.

1. Classical methods

a) Qualitative analysis

A qualitative analysis determines the presence or absorbance of particular

compounds, but not the mass or concentration.

b) Chemical tests

There are numerous qualitative tests, for example, the acid test for the gold and

the Kastle-Meyer test for the presence of blood.

C) Flame test

It is a test where the certain metal ions especially alkali and alkaline earth metals

are identified based on the characteristic colors they exhibit in flame.

d) Gravimetric analysis

Chemical analysis is based on weighing a final product is called gravimetric

analysis. It involves determining the amount of material present by weighing the sample

before and / or after some transformations.

e) Volumetric analysis

It is refers to any procedure in which the volume of reagent needed to react with

analyte is measured. In titration, increments of reagent solution, the titrant, are added to

analyte until their reaction is complete. Form the quantity of titrant required, we can

calculate the quantity of analyte that must have been present. The most common type

Page 8: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 8

titrations are based on acid-base, oxidation-reduction, complex formation and

precipitation reactions. In these titrations different types of indicators are used to

determine the equivalence point.

2. Instrumental methods

a) Chromatographic analysis

Chromatography is the collective term for a family of laboratory techniques for

the separation of mixtures. It involves passing a mixture dissolved in a mobile phase

through a stationary phase, which separate the analyte to be measured from other

components in the mixture and allows it to be isolated. There are several different types

of chromatographic techquines currently in use such as Paper Chromatography, Thin

Layer Chromatography (TLC), Gas Chromatography (GC), Liquid Chromatography

(LC), High performance liquid chromatography (HPLC) with different detectors, Ion

exchange chromatography (IEC) and Gel permeation or Gel filtration chromatography.

b) Electrochemical analysis

Electro analytical methods measure the potential (volts) and/or current (amps) in

an electrochemical cell containing the analyte. Electrochemistry measures the interaction

of the material with electric field. These methods can be categorized according to which

aspects of the cell are controlled and which are measured. The three main categories are

potentiometry (the difference in electrode potentials is measured), coulometry (the cell’s

current is measured over time) and voltammetry (the cell’s current is measured while

actively altering the cell’ potential).

Page 9: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 9

C) Microscopy

The visualization of single molecules, single cells, biological tissues and

nonmaterials is an important activity in analytical science. Microscopy can be categorized

into three different fields, optical microscopy, electron microscopy and scanning probe

microscopy. Recently, this field is rapidly progressing because of the rapid development

of the computer and camera industries.

d) Spectrophotometry and colorimetry

Spectrophotometry involves the use of a spectrophotometer. A spectrophotometer

is a photometer which can measure light intensity as a function of color or more

specifically, the wavelength of light. There are many kinds of spectrophotometers, among

the most important distinctions they used to classify them are the wavelengths they work

with, the measurement technique they use, how they acquire a spectrum and the sources

of intensity variation they are designed to measure. The use of spectrophotometers is not

limited to studies in physics but they are also commonly used in other scientific fields

such as chemistry, biochemistry and molecular biology.

Direct spectrophotometric determinations such as colorimetric analysis or UV

determination are widely used in pharmaceutical analysis. The estimation of an analyte

concentration based on its absorption of ultraviolet or visible radiation is one of the most

frequently encountered quantitative analytical methods. One reason for its popularity is

that many organic and inorganic compounds have strong absorption bands in the UV or

visible radiation. An additional advantage to UV or visible absorption is that in most

cases it is relatively easy to adjust experimental and instrumental conditions so that

Beer’s law is obeyed. The applications of Beer’s law for the quantitative analysis of

Page 10: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 10

samples in environmental chemistry, clinical chemistry, industrial chemistry, forensic and

in pharmaceutical chemistry are numerous. The scale of operation for molecular UV or

visible absorption is commonly used for the analysis and is generally better than that of

IR absorption. It is routinely used for the analysis of trace analytes in macro and meso

samples. The analysis of a sample by molecular absorption spectroscopy is relatively

rapid, although additional time may be required when it is necessary to use a chemical

reaction to transform a non absorbing analyte into an absorbing form. The cost of UV-

Visible instrumentation is relatively less. The selectivity and sensitivity of analyte

towards the absorption of light can be increased by converting it into a chromogenic

derivative, by adopting a suitable chemical reaction, which also prevents the

interferences.

Classification of functional groups in drugs [13-16]

A feature of organic drug is the presence of some specific functional groups in

their molecules. Knowing the reactions of functional groups, one can easily analyze any

organic molecule with a complicated structure.

Category Functional group Examples

1. Acidic nature

- COOH (Carboxylic) Aspirin, Ibuprofen, Cetirizine

-SH (Sulpha hydryl) Captopril, Penicillamine

OH

(Phenolic Hydroxyl)

Nelfinavir

Page 11: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 11

(Sulphone)

Dapsone

(Amide)

Mosapride

2. Basic nature

R1 = R

2 = H;

R1 = H; R

2 = alkyl;

R1 = R

2 = alkyl groups

(Primary, Secondary and

Tertiary amino)

Repagladine, Ampicilline, and

Mesalamine,

3. Neither acidic nor

basic nature

RCR1

O

Ketone

Acuvail

(Hydroxy methyl)

Emtricitabine

NO

O Nitro

Metronidazole, Azathioprine and

Nimesulide

(Methoxy)

Albendazole

R-O-R1

(Ether)

Metoprolol

NH

O (Lactam)

Amoxicillin, Penicillin

(Acetylenic)

Norethindrone, Ethynulesteroid

Page 12: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 12

Selection of reagents for the analysis of solutes [17]

Several papers are being published every year on the reactions and possible

applications of new and old reagents for organic analysis. The selection of an appropriate

reagent for a particular analytical situation is still a challenging problem. The choice of a

particular reagent depends on careful consideration of such factors as the scale and

economy of the reaction, the presence of other functional groups that might be adversely

affected by the reagent, the deactivation of the reaction centre by steric and electronic

effects, the instability or high reactivity of the desired product, the rate of reaction,

position of equilibrium (in the case of a reversible reaction) and other related factors.

The selection of a reagent for the determination of a particular compound is made

after a literature search for methods that have been used in parallel situations elsewhere

or that show reasonable promise for the compound under consideration. If adequate

information is not available in this way, then the reagents that act most rapidly and

stoichiometrically can be chosen after investigation of the performance of several

plausibly selected ones on a pure sample of the compound. Reagent selectivity for a

particular functional group is normally the minimum requirement. Specificity of the

reagent for a single compound containing that functional group is often desirable, not

only to isolate it from other compounds containing the same functional group, but also to

eliminate the effects of interfering compounds.

The general objective of treating a compound with a reagent (for purpose of

determination of the compound) is to get one or more derivatives having a measurable

physical or chemical property that is as completely different as possible from that of any

Page 13: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 13

of the reactants. Some examples of properties in which a specific change may be brought

about by the reagent are as follows:

1. The reagent may bring about a change in or destroy an acidic, alkaline, oxidizing or

reducing property of the functional group, the amount of change being determined

titrimetrically or spectrophotometrically.

2. The reagent forms a product with a solubility product different from that of original

sample and this property serves as basis for the gravimetric determination or for the

isolation, concentration and purification of the compound for examination by other

analytical techniques.

3. The reagent forms a chromophore or reduces the concentration of an already existing

chromophore, the change being measured by one of the spectrophotometric techniques-

ultraviolet, infrared or visual. Many spot tests for functional groups or spray reagents in

TLC depend on the formation of a colored derivative. Similarly a fluorophore may be

generated or quenched and the change is measured fluorimetrically.

4. The reagent act on the sample to produce a gas measurable nanometrically or one that

can be collected and determined by titrimetric, gravimetric or other types of

techniques.

5. The reaction of the reagent with the sample produces a derivative that is less polar

and/or more volatile than the original sample and is thus more amenable for

determination by high performance liquid or gas liquid chromatography. Many

compounds containing polar functional groups show unfavorable properties such

as low volatility, tailing, irreversible adsorption on to column packing and thermal

Page 14: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 14

instability. Vast improvements in these respects are easily realized because the polar

nature of the compound promotes derivatization with suitable reagents to replace the

polar group with a less polar one, giving some times a more sensitive detection

response.

6. The reagent forms derivatives with the sample suitable for structural investigation or

estimation by the application of NMR and Mass spectroscopic techniques.

7. Enzymatic treatment of the sample may catalyze a specific reaction and structural

change which may be monitored by one or more of the several analytical techniques.

Page 15: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 15

Section (ii): SPECTROSCOPY

Spectroscopy is the branch of science that deals with the measurement and

interpretation of electromagnetic radiation (EMR) absorbed or emitted when the

molecules or the atoms or ions of the sample move from one energy state to another. At

ground state, the energy of the molecule is the sum of rotational, vibrational, and

electronic energies. EMR is made up of discrete particles or photons. EMR possess both

wave and particle characteristics. i.e. it can travel in vacuum also. The different regions

of EMR are Visible, Infra-Red, Microwaves, Radio waves, X-rays, Gamma rays, or

Cosmic rays. The energy of the radiation depends upon the frequency and the

wavelength of the radiation.

Energy of EMR:

E = h υ …………… 1.ii.1

Where; E = Energy of radiation

h = Plank’s constant (6.625 × 10 -34

J.sec-1

)

υ = Frequency of radiation

Types of spectroscopy:

1. Based on the study of the spectra obtained that may be at atomic or molecular

level:

(i) Atomic spectroscopy: where the changes in the energy takes place at atomic level.

E.g. Atomic Absorption Spectroscopy, Flame Photometry.

Page 16: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 16

(ii) Molecular Spectroscopy: where the changes in the energy takes place at molecular

level.

E.g. UV Spectroscopy, Colorimetry, Fluorimetry, I.R. Spectroscopy etc.

2. Based on the study of the spectra obtained that may be due to absorption or

emission of the EMR:

(i) Absorption Spectroscopy: Spectroscopy involving the spectra due to the absorption

of the EMR.

E.g. UV- Visible Spectroscopy, I R-Spectroscopy etc.

(ii) Emission Spectroscopy: Spectroscopy involving the spectra due to the emission

(following the absorption) of the EMR.

E.g. Flame Photometry, Fluorimetry

3. Based on the study carried out at electronic or magnetic levels:

(i) Electronic Spectroscopy:

E.g. U.V Spectroscopy, Colorimetry, Fluorimetry

(ii) Magnetic Spectroscopy:

E.g. NMR Spectroscopy, ESR Spectroscopy.

Absorption of EMR by organic molecules:

If we pass light from a ultra-violet lamp through an organic sample, some of the

light is absorbed. In particular, some of the wavelengths (frequencies) are absorbed and

others are virtually unaffected. We can plot the changes in absorption against wavelength

Page 17: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 17

and produce an absorption spectrum. A molecule can only absorb a particular frequency,

if there exists within the molecule an energy transition of magnitude (E = h υ).

Although almost all parts of the electromagnetic spectrum are used for studying

the matter, in organic chemistry we are mainly concerned with energy absorption from

three or four regions -Visible, Infra-Red, and Microwave and Radio frequency absorption

which is shown in Fig 1.ii.1.

Fig 1.ii.1 Electro magnetic spectrum:

Table: 1.ii.1- Spectroscopic techniques in organic chemistry:

Radiation absorbed Effect on the molecule

Ultraviolet-Visible levels

λ: 200 – 380 nm and 380 – 760 nm

Changes in the electronic energy with in the

molecules.

Infra-Red (Mid Infra-Red)

λ : 2.5 – 25 µm

Changes in the vibrational and rotational

movements of the molecule

Microwave

λ : 1 mm – 10 cm

Electronic spin resonance and electronic

paramagnetic resonance; induces changes in

the magnetic properties of the unpaired

electrons.

Page 18: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 18

Principles of UV-Visible spectrophotometry

Basic principles of spectrophotometry [18]

Spectrophotometry is one of the advanced analytical methods used in qualitative

and quantitative analysis. It is based on the absorption of electromagnetic radiation.

Every compound has a unique characteristic relationship with electromagnetic radiation.

This is a direct process which does not change composition and properties of the

experimental solutions. Spectrophotometry is called colorimetry in the visible region of

electromagnetic spectrum. This technique is mainly used for the determination of trace

amounts of metals, non-metals and organic compounds and also used in the

determination of composition of metal complexes and stability of coordination complexes

in solution state.

The measurement of radiation frequency is done experimentally which gives a

value for the change of energy involved. From this, we can conclude the set of possible

discrete energy levels of the matter. The curve which is obtained by plotting the light

transmitted, the height of the ordinate in the plot at a wavelength is directly proportional

to the concentration of the absorbing medium.

Fig 1.ii.2 Energy Transition:

LUMO

HOMO

Excited stateGround state

E

V

h v

= *

=

Page 19: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

The lowest energy transition is that between the highest occupied molecular

orbital (HOMO) and the lowest unoccupied molecular orbital (LUMO) in the ground

state. The absorption of the EMR excites an electron to the LUMO and creates an excited

state. The more highly conjugated the system, the smaller the HOMO

and therefore, the lower the frequ

radiation. The colors that our eyes see are typically due to highly conjugated organic

molecules. The unit of molecule that is responsible for the absorption is called the

chromophore, of which the most common are C=C (

By the interaction of UV

the energy. Thereby the molecule undergoes transition from ground state to excited state.

The intensity of absorption depends upon the concentration and path length as given by

the Beer-Lambert’s Law.

Fig 1.ii

Laws of photometry:

When the light passes through a homogeneous transparent medium, a part of

incident light is reflected, a part is absorbed by the medium, and the rest is allowed to

INTRODUCTION

The lowest energy transition is that between the highest occupied molecular

orbital (HOMO) and the lowest unoccupied molecular orbital (LUMO) in the ground

state. The absorption of the EMR excites an electron to the LUMO and creates an excited

more highly conjugated the system, the smaller the HOMO-

and therefore, the lower the frequency, the longer the wavelength (λ) of the absorbed

The colors that our eyes see are typically due to highly conjugated organic

The unit of molecule that is responsible for the absorption is called the

chromophore, of which the most common are C=C (π → π*) and C=O (n →

By the interaction of UV-visible radiation with matter, only the valence electron absorbs

the molecule undergoes transition from ground state to excited state.

The intensity of absorption depends upon the concentration and path length as given by

Lambert’s Law.

1.ii.3 Spectrophotometer common features:

When the light passes through a homogeneous transparent medium, a part of

incident light is reflected, a part is absorbed by the medium, and the rest is allowed to

INTRODUCTION

Page 19

The lowest energy transition is that between the highest occupied molecular

orbital (HOMO) and the lowest unoccupied molecular orbital (LUMO) in the ground

state. The absorption of the EMR excites an electron to the LUMO and creates an excited

-LUMO gap, ∆E,

) of the absorbed

The colors that our eyes see are typically due to highly conjugated organic

The unit of molecule that is responsible for the absorption is called the

*) and C=O (n → π*) systems.

nce electron absorbs

the molecule undergoes transition from ground state to excited state.

The intensity of absorption depends upon the concentration and path length as given by

When the light passes through a homogeneous transparent medium, a part of

incident light is reflected, a part is absorbed by the medium, and the rest is allowed to

Page 20: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 20

transmit as such. If Io denotes the intensity of incident light, Ir, that of the reflected light,

Ia , that of the absorbed light and It, that of the transmitted light,

We can write

Io = Ia + It + Ir ……..... (1.ii.2)

In spectrophotometric measurements, comparison cells are used for sample and

reference so the Ir is eliminated. Then

Io = Ia + It ………... (1.ii.3)

The fraction of incident radiation passing through a medium is governed by two

laws, known as Lambert’s law and Beer’s law.

Lambert’s law

According to this law, “when a beam of monochromatic light is allowed to pass

through a transparent medium, the rate of decrease of intensity with the thickness of

medium is directly proportional to the intensity of the incident light.”

This law may be expressed mathematically as follows.

���� �� ……………………(1. ii. 4)

or

���� = �� ……………………(1. ii. 5)

Where I denote the intensity of incident light of wavelength λ, ‘l’ denotes the thickness of

the medium and ‘k’ denotes the proportionality factor.

Page 21: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 21

The negative sign indicates that ‘I’ decreases as ‘l’ increases. On integrating

equation (1.ii.5) and putting I = Io when l = 0, we get

�� ���� = ��⋯⋯⋯⋯⋯⋯(1. ii. 6)

or

�� = ������⋯⋯⋯⋯⋯⋯(1. ii. 7) Changing from natural to common logarithms, we get

�� = ������.������ = ��������

Where

�� = intensity of the incident light

�� = intensity of the transmitted light

�= thickness of the medium

�� = absorption coefficient.

The ratio ���� is termed as the transmittance T, and its reciprocal

���� is termed as

opacity. The ratio � ! " "# is termed as absorbance, A.

A = � ! ���� ⋯ ⋯⋯⋯⋯⋯⋯⋯(1. ii. 8)

Page 22: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 22

Beer’s law

This law states that “the intensity of a beam of monochromatic light decreases

exponentially with the increase in concentration of the absorbing medium arithmetically.”

This is expressed as

−���& = ���� ⋯⋯⋯⋯⋯⋯(1. ii. 9)

�� = ������.�������& = ���������& ⋯⋯⋯⋯(1. ii. 10)

Where

c = concentration of the absorbing medium

����= constant

By adding Lambert’s and Beer’s laws, we get

)* = )+10�,-. ⋯⋯⋯⋯⋯⋯⋯(1. ii. 11)

log )+)* = acl ⋯⋯⋯⋯⋯⋯⋯(1. ii. 12)

Where

‘a’ is new proportionality constant.

Since, log ���� = 5,

We can rewrite the above expression as

A = acl ⋯⋯⋯⋯⋯⋯⋯(1. ii. 13)

Page 23: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 23

In the above equation ‘a’ is called absorptivity when the concentration ‘c’ is

expressed in g L-1

. If ‘c’ is expressed in moles L-1

and the thickness of the medium ‘l’ in

cm, we write ‘ε’ in the place of’a’.

Thus

A = ε;< ⋯⋯⋯⋯⋯⋯⋯(1. ii. 14)

Where, ε is molar absorptivity.

The absorbance of a solution is an additive property. If there are more than one absorbing

species present in the solution, the total absorbance is equal to the sum of all the

absorbance values of the components present in the solution.

This can be expressed as follows,

A = =εc> + εc@ + εcA + εcB⋯⋯⋯+ εcCD = EεFcFC

FGH⋯⋯⋯⋯(1. ii. 15)

Where, the path length is kept constant.

Ringbom’s plot [19]

Ringbom introduced a new method of plotting spectrophotometric data in the year

1939. He plotted the percent absorbance (P) against logarithm of concentration (log C).

Absorbance = I - Transmittance (T) and

% Absorbance = 100 - % transmittance.

An S̕ ̕ shaped sigmoid curve would be obtained. Ringbom proposed that the

accuracy of an analytical determination based on the spectrophotometric method is

Page 24: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 24

greatest when ΔP/(ΔC/C) reaches a maximum value, where ΔP is photometric error and

ΔC/C is relative analysis error.

ΔP/(ΔC/C) Can be equated to ΔP/2.303.

The Ringbom plot has two outstanding features.

1. The analysis error is least in the concentration range of determination. This is the

concentration range corresponding to the nearly linear portion or the steep slope

of the plot.

2. At any concentration, the accuracy of the analysis can be easily evaluated. The

relative analysis error is obtained on dividing 230 by the slope of the curve at that

point.

Sensitivity of spectrophotometric methods

The term sensitivity implies the smallest determinable concentration or the

amount of species concerned. In photometry it refers to the slope of the calibration curve

plotted between the concentration and absorbance value. Numerically, it is expressed as

molar absorptivity at the wavelength of maximum absorbance.

ε = Abc⋯⋯⋯⋯⋯⋯(1. ii. 16)

Where

A= absorbance

b = path length in centimeters,

c = concentration in moles L-1

.

Page 25: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 25

The molar absorptivity is expressed in L mole-1

cm-1

in CGS units and in

mole-1

nm-1

in SI units.

The methods [20]

for which ε > 1 x 104 are considered to be sensitive while others

with ε < l x 103 is termed as less sensitive methods. As per the quantum theory, εvalues

never exceed 1.5 x 105. In some indirect methods such as amplification methods higher

value of εcan be obtained.

Sensitivity can also be expressed in terms of specific absorptivity (a) which is

related to molar absorptivity (ε) as

a = εAt. wt × 1000 ⋯⋯⋯⋯⋯⋯(1. ii. 17)

Sandell’s [21]

expressed the sensitivity by a quantity S known as Sandell’s

photometric sensitivity.

S = 10�A

a ⋯⋯⋯⋯⋯⋯⋯(1. ii. 18)

The units of Sare µg cm-2

.

INSTRUMENTATION:

1. Source of Radiation:

UV-Visible spectrophotometer uses two light sources, a deuterium (D2 ) lamp for

ultraviolet light and a tungsten (W) lamp for visible light.

2. Collimating System:

It makes the light beam parallel or focused. It includes lenses, mirrors and slits.

Page 26: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 26

3. Monochromator (Filter):

It converts the polychromatic radiation to monochromatic radiation. It consists of

filters, prisms and gratings.

4. Sample Compartment:

It is to permit radiation to pass through the sample. These are made of quartz,

fused silica glass.

5. Frequency Analyzer:

Which separates out all the individual frequencies generated by the source (the

most familiar being the triangular glass prism as used by Isaac Newton for visible light)

6. Detector:

For measuring the intensity of the radiation at each frequency, allowing the

measurement of how much energy has been absorbed at each of these frequencies by the

sample. e.g. Barrier layer cells, Photo emissive tubes, Photomultiplier tubes etc.

7. Recorder:

A pen recorder or a computerized data station, with a VDU for initial viewing of

the spectrum

Page 27: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 27

Fig 1.ii.4- UV/Vis – spectrophotometer optical design

8. Applications:

1. Quantitative determination of chromophore concentrations in the solution.

2. Impurity determination by spectrum subtraction.

3. Structural elucidation of organic compounds (saturation and unsaturation, and the

extent of aromaticity).

4. Determination of pKa values of indicators.

5. Determination of molecular weights of amines.

6. Identification of cis-trans isomers: Trans isomer absorbs less energy than cis isomer.

7. Chemical kinetics: The rate of reaction is calculated as the change in absorbance per

unit time.

8. Qualitative and quantitative analysis of materials used in dye and pharmaceutical

industries.

Page 28: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 28

Analytical Method Validation:

Accuracy [23]:

The accuracy of an analytical method is the degree of agreement of results

generated by the method to the true value or a conventional true value. Accuracy can be

assessed by applying the analytical method to samples or mixtures of sample matrix

components to which known amounts of the analyte have been added, above and below

the normal levels expected in the samples. Method accuracy is the agreement between the

differences in the measured analyte concentrations of the fortified (spiked) and

unfortified samples and the known amount of analyte added to the fortified sample.

Comparison of the method’s results can be performed by using an established reference

method, assuming that the latter is free from systematic errors. Second, accuracy can be

measured by analyzing a certified reference material, and comparing the measured value

with the true value as supplied with the material. If such reference material is not

available, a blank sample matrix can be spiked with a known concentration that should

cover the range of concern, including one concentration close to the quantization limit.

The expected recovery depends on the sample matrix, on the sample processing

procedure, and on the analyte concentration.

Linearity:

The linearity of an analytical method is its ability to elicit test results that directly,

or by means of well-defined mathematical transformation, is proportional to the

concentration of analyte in samples within a given range. Frequently, the linearity is

evaluated graphically in addition or alternatively to mathematical evaluation. The

deviations from linearity are minimized by Regression analysis; a statistical technique

Page 29: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 29

which provides the means for objectively obtaining a linear line with minimum deviation

between the plot and data. For linear ranges, the deviations should be equally distributed

between the positive and negative values. The minimum concentration for the working

range is known as the Limit of Detection (LOD) where as, the maximum concentration is

called the Limit of Linearity. It is not always possible to obtain a linear graph due to

interferences, signal noise, sample matrix or deviations from Beer’s Law. However,

several programs are commercially available for computing the best fit curve for the data

set.

Significant Figures:

All measurements carry some degree of uncertainty. The degree of uncertainty

depends upon both accuracy of the measuring device and skill of its operator. In order to

determine the number of significant digits in a measurement or number, the following

rules may be applied:

1. If there is no decimal point, the right-most non-zero digit is the least significant digits.

2. In case of numbers that include a decimal point, the right-most non-zero digit is the

least significant digit regardless of its value.

Significant digits should not be considered the same as uncertainty in

measurement systems. They are used for manipulating experimental data, but not for

expressing uncertainty. Significant figures have economic significance simply because

the more the significance numbers expected from the measurement, the more expensive

will be the equipment and higher the quality of reagents required. Care should therefore,

be taken to properly define the optimum number of significant figures when attempting

an analysis.

Page 30: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 30

Other Performance Parameters:

Sensitivity:

The sensitivity of an instrument is the change in output signal while a change in

the physical parameter is being measured. It is determined with the help of the calibration

curves. The sensitivity is constant over the entire range in a linear response system and is

found to be changing with respect to the concentration in case of non-linear systems. In

case of analytical instruments, the sensitivity is usually expressed as the concentration of

analyte required to cause a given instrument response.

Selectivity:

The selectivity of an analytical method is defined as its ability to accurately

measure an analyte in the presence of interference that may be expected to be present in

the sample matrix.

Range:

The range of an analytical method is the interval between the upper and lower

levels (including these levels) that have been demonstrated to be determined with

precision, accuracy and linearity by using the specified method. The range is normally

expressed in the same units as the test results (e.g. percentage, parts per million etc.)

obtained by the analytical method.

Limit of Detection (LOD):

The LOD is the point at which a measured value is larger than the uncertainty

associated with it. It is the lowest concentration of analyte in a sample that can be

detected but not necessarily quantified.

Page 31: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 31

LOD = 3.3σS ⋯⋯⋯⋯⋯(1. ii. 19)

Units for limit of detection is µg mL-1

Where,

σ = Standard deviation of the response.

S = Slope of the calibration curve.

The slope ‘S’ may be estimated from the calibration curve of the analyte.

The estimate of σ may be carried out in a variety of ways.

Limit of Quantification (LOQ):

LOQ =10σS ⋯⋯⋯⋯⋯(1. ii. 20)

Units for limit of Quantification is µg mL-1

Page 32: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 32

Section (iii): TYPES OF REACTIONS AND REAGENTS INVOLVED

IN THE PROPOSED SPECTROPHOTOMETRIC

METHODS

Different types of reactions are applied in the present proposed methods for the

determination of the following selected pharmaceutical drugs. A brief discussion of these

reactions and the reagents used to bring about these reactions are presented here.

Table 1.iii.1 Different types of the proposed methods.

S.No Name of the drug

determined

Method Reagent(s) used

1

Atazanavir(AZV)

AZV-BTB Bromothymol blue (BTB); Phthalate buffer

2 AZV-BPB Bromophenol blue(BPB) ; Phthalate buffer

3 AZV-BCG Bromocresol green(BCG) ;Phthalate buffer

4

Emtricitabine

(EMT)

EMT-PDAB P-Dimethylaminobenzaldehyde(PDAB);

H2SO4

5 EMT-MBTH 3-Methyl 2-benzothiazolinone hydrazone

(MBTH) ; FeCl3

6

EMT-HMBA 4-Hydroxy-3-methoxybenzaldehyde

(HMBA) ; H2SO4

7

Eplerenone(EPL)

EPL-BTB Bromothymol blue(BTB); Phthalate buffer

8 EPL- BPB Bromophenol blue(BPB); Phthalate buffer

9 EPL-BCG Bromocresol green(BCG); Phthalate buffer

10

Olopatadine(OLO)

OLO-BTB Bromothymol blue(BTB);Phthalate buffer

11 OLO- BPB Bromophenol blue(BPB);Phthalate buffer

12 OLO-BCG Bromocresol green(BCG); Phthalate buffer

13

Pemetrexed

disodium (PMD)

PMD-PDAC P-Dimethylaminocinnamaldehyde(PDAC)

; H2SO4

14 PMD-HMBA 4-Hydroxy-3-methoxybenzaldehyde

(HMBA); H2SO4

15 PMD-MBTH 3-Methyl2-benzothiazolinone hydrazone

(MBTH) ; cerric ammonium sulphate

Page 33: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 33

16

Zolmitriptan(ZOL)

ZOL-PDAC P-Dimethylaminocinnamaldehyde(PDAC);

H2SO4

17 ZOL-PDAB P-Dimethylaminobenzaldehyde(PDAB);

H2SO4

18 ZOL-NQS 1,2-Napthoquinone-4Sulphonic acid

(NQS); NaOH

Oxidative coupling reactions:

Oxidative coupling procedure involving the use of 3-methy1-2-benzothiazolinone

hydrazone (MBTH) in the presence of an appropriate oxidant under acidic, neutral or

alkaline conditions to form highly colored products were explored for the determination

of some of the mentioned active pharmaceutical ingredients possessing amenable

structural features.

MBTH – Oxidant: (EMT and PEM)

The reported applications of 3-Methyl-2-benzothiazolinone hydrazone (MBTH)

include the detection and determination of phenols[24-26],

formaldehyde[27]

and other

aldehydes[28,29]

, polyhydroxy compounds[30,31],

aromatic amines including heterocyclic

amines such as antipyrine and amidopyrine,[32]

iminohetero aromatic compounds, indoles,

carbazoles and phenothiazines[33].

MBTH was also used for the indirect determination of

compounds such as glycerides [34]

and glycosamine glycans [35]

etc., which yield one of

the above mentioned groups on preliminary treatment with certain reagent.

The reaction of MBTH with phenols or amines was found to be based mainly on

the nature of the oxidant and acidic, alkaline or neutral conditions employed. Friested.et

al[36]

initially carried out, the coupling under alkaline conditions with ferricyanide, but a

modification, to use cerric ammonium sulphate in acidic medium was successful and led

to an increased sensitivity.[25]

Later, Umeda [37]

carried out the reaction using a basic

Page 34: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 34

solvent (triethanolamine) with phenols. El-Kommos [25]

suggested cerric ammonium

sulphate as an oxidant under acidic conditions for the determination of 17 pharmaceutical

phenols. Ferric chloride has been mostly used for the determination of aromatic and

heteroaromatic amines by Sawicki.et al [33]

(in neutral conditions) and by Pays.et. al [32]

(in acidic conditions). Other oxidants such as periodate (acidic conditions), ammonium

persulphate (alkaline conditions) and potassium dichromate (acidic conditions) were used

for the determination of ethylenic compounds and primary alcohols [38]

(after oxidation

with ruthenium textraoxide) and phenidone [39].

Recently, Sastry and Sastry reviewed

various aspects of MBTH chemistry in pharmaceutical analysis [40].

Under the reaction conditions, on oxidation MBTH loses two electrons and one

proton, forming an electrophilic intermediate which has been postulated to be the active

coupling species.[41-45]

This intermediate undergoes electrophilic substitution with the

drug to form the colored product. It also reacts with phenol or amine by electrophilic

attack on the most nucleophilic site on the aromatic ring of the phenol or amine (i.e.

Para-position to phenolic hydroxyl or amino group, if it is either free or substituted with

electron withdrawing substituent like -Cl, -COOH etc.; where Para-position is substituted

with electron releasing groups, coupling proceeds through ortho-position to phenolic

hydroxyl or amino group) and the oxidant to form the colored species.

The absorption maximum and molar absorptivity values of the colored species are

influenced by the nature (inductive, mesomeric or steric effects) of other substituents

present in the compound besides the desired substitutes, the oxidizing agent used, the

solvent employed and the pH of the medium. Some of the antiviral and anticancer agents

under investigation (Emtricitabine and Pemetrexed disodium) possess structural features

Page 35: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 35

that are likely to enable them to react with MBTH. Attempts were made by the present

author to study the suitability of MBTH reagent for the estimation of these two drugs.

The effect of the concentration of the regent (MBTH with different oxidants) pH,

temperature, time and order of addition of reagents with respect to maximum sensitivity,

minimum blank and obedience to Beer’s law have been investigated and the results are

recorded in chapter 4 of EMT and chapter 7 of PEM in the thesis.

Ion association complex formation reactions: (AZV, EPL and OLO)

Ion-association is a chemical reaction where by ions of opposite electrical charge

come together in solution to form a distinct chemical entity. Ions of opposite charge are

naturally attracted to each other by the electrostatic force. This is described by coulomb’s

law.

F=

q1q2

ε r2

F is the force of attraction, q1 and q2 are the magnitudes of the electrical charges,

and ‘ε’ is the dielectric constant of the medium and ‘r’ is the distance between the

ions[46].

Sulphonaphthalein dyes are commonly used as anionic dyes to form ion-pair

complexes with the nitrogenous compound present in positively charged protonated

forms. Ion-pair complexation reaction has been applied for the spectrophotometric

determination of many organic compounds [47-52]

.

The term molecular complex is used to describe a variety of types of association

products of two or more molecules. In recent years, extensive attention has been given to

a large group of complexes formed by weak interaction of certain classes of organic

compounds functioning as electron donors (bases), with others which act as electron

Page 36: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 36

acceptors (acids) [53-55]

. The forces which lead to the formation of molecular complex

include physical forces such as dipole and induced dipole interactions, London dispersion

forces, hydrogen bonding and dative bonding interactions. The donor-acceptor complexes

(whose composition can be represented by integral ratio of the components) are, in many

instances, so unstable that they cannot be isolated in the pure state at ordinary

temperatures but exist only in solutions in equilibrium with their components. They can,

however, usually be detected readily because of differences in physical properties (e.g.

Absorption spectra, solubility in organic solvents) from those of the pure components.

The ion-association complex or adduct (commonly known as ion pair, if two ions

are involved, colored, neutral and extractable into organic solvents) is a special form of

molecular complex resulting from two components extractable into organic solvents from

aqueous phase at suitable pH. One component is a chromogen (dye or metal complex)

possessing charge (cationic or anionic in nature) and so it is insoluble in organic solvents.

The second component is colorless, possessing opposite charge to that of chromogen.

Acid dyes such as bromothymol blue, bromophenol blue, bromocresol green,

mordant black-III, solochrome black-T, metanil yellow dye, bromocresol red, eosin,

methyl orange, phenol red, alizarin violet 3B, alizarin brilliaint violet R, and fast green

FCF have been used arbitrarily by various workers for the determination of components

exhibiting basic properties e.g. amines, quaternary ammonium compounds and

heterocyclic compounds[56-62]

. Some basic dyes (parent moiety - xanthene, triphyenyl

methane, azine, oxazine and thiazine) and metal complexes have been tried by some

workers for the determination of components exhibiting acidic properties [63-65].

Page 37: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 37

In the present investigations, it was found that atazanavir (AZV), eplerenone

(EPL) and olopatadine (OLO) form ion-association complexes with the bromothymol

blue, bromophenol blue, and bromocresol green, in the presence of phthalate buffer

(pH 2.4).The formed respective colored ion-association complexes are extracted into

chloroform and the analytical results obtained in these methods are reported in

chapters-3, 5 and 6.

Condensation reactions using aromatic aldehydes like PDAC, PDAB and HMBA in

acidic media (EMT, PEM and ZOL)

PDAB, PDAC and HMBA are used as analytical reagents [66-68]

for the detection

and spectrophotometric determination of aromatic and aliphatic primary and secondary

amines. PDAC and PDAB has demonstrated a wide usefulness as a chromogenic reagent

for spectrophotometric analysis of several compounds such as urea, thiourea and their

N-alkyl/aryl derivatives[69]

, aceclofenac[70]

, sodium diclofenac and oxyphenbutazone

[71],

glafenine and metoclopramide[72]

. PDAC and PDAB have also been applied for micro

plate-based assay of p-aminohippuric acid [73]

in plasma and urine samples, and for rapid

testing of benzodiazepines [74]

, nimesulide [75]

, satranidazole [76]

and sympathomimitic

amines in pharmaceuticals. The use of modified dimethylaminocinnamaldehyde has also

been described for the analysis of flavanols in wines. [77]

The majority of the methods

mentioned above involve a reaction in acid medium with heating to produce coloured

compounds (ranging from orange to red or pink).

In the present study, PDAB, PDAC and HMBA form orange red or red colored

chromogens with zolmitriptan, emtricitabine and pemetrexed Disodium. The reaction

between PDAB or PDAC with secondary amino group of zolmitriptan is assumed to take

Page 38: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 38

place via condensation of amino group with the carbonyl group of the reagent to produce

Schiff’s bases or enamines [78-82].

The primary amino group of emtricitabine and

pemetrexed disodium is involved in the condensation process with PDAB, HMBA and

PDAC to form Schiff’s base complexes between these drugs in acidic medium.

The attempts were made by the author to study the suitability of the reagents for

the estimation of these three drugs. The effect of the concentration of the regents,

temperature, time and order of addition of reagents with respect to maximum sensitivity,

minimum blank and obedience to Beer’s law have been investigated and the results are

recorded in Chapter-4, Chapter-7 and Chapter-8 which are to a large extent based on

reactions suggested in the literature.

Reaction of 1, 2-naphthoquinone-4-sulfonic acid (NQS) with zolmitriptan (ZOL):

1, 2-Napthaquinone-4-sulphonic acid (NQS) is a reagent widely applied for

analytical determinations of amines and amino acids by using UV-Visible

spectrophotometric method. NQS is readily able to react in basic medium and at

moderate temperatures with both primary and secondary amino groups to form colored

derivatives, which were determined by spectrophotometrically. The reagent (NQS) was

first used as a colorimetric reagent by Folin [83]

for the determination of amino acids in an

alkaline solution to form highly colored complexes. Later on, this reagent was employed

for the spectrophotometric determination of many amino drugs [84-110].

In the present investigation, the author utilized the reagent, NQS to estimate the

drug zolmitriptan in bulk and in its pharmaceutical formulations and the results are

reported in Chapter-8.

Page 39: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 39

REFERENCES:

[1]. Kvetoslav spurny, Talanta, 51(5), 921-933.

[2]. Tr A.C. Trends in analytical chemistry, 21(9-10), 547-557.

[3]. H. A. Laitinen, History of analytical chemistry in the U.S.A, Talanta, 1989,36 (1-2),

1-9.

[4]. Bard, A.J, Faulkner, L.R, Electrochemical methods: Fundamentals and applications.

Newyork: John Wiley &sons, 2000, 2.

[5]. Skoog, D.A, West, D.M and holler, F.J, Fundamentals of analytical chemistry,

1988, 5.

[6]. US Federal Food, Drug and Cosmetics Act, August 2008, SEC. 210, Accessed 17.

[7]. Directive 2004/27/EC of the European Parliament and the council of 31 March 2004

amending directive 2001/83/ EC on the community code relating to medicinal

products for human use, Article 1 published 2004, march 31.

[8]. Metrohm USA Inc, Pharmaceutical analysis, 2010.

[9]. G.D. Watson, Pharmaceutical analysis, Churchill Livingstone, U.K, 1999.

[10]. Ashutoshkar, Pharmaceutical drug analysis, New Age International Publishers,

Revised 2nd

Edition, India, 2001.

[11]. B.M.Wenclawaik, M. Koch, Quality assurance in analytical chemistry Springer

International edition, 1st Edition, 2004, PP 215 and 216.

[12]. R.A. Day, Jr and A.L. Underwood, Quantitative analysis, 6th

Edition, 1998,

pp. 420-421.

[13]. Robert Thornton Morrison and Robert Neilson Boyd, Organic Chemistry,

Pearson Education, 6th edition, 2007, pp. 354.

[14]. Cheronic, N.D and Ma, T.S., Organic functional group analysis by micro and

semi micro methods, 1964, Inter science, New York.

Page 40: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 40

[15]. Higuchi, T. and Brochman-Hansen, E., Eds., Pharmaceutical Analysis, 1961, Inter

science, London.

[16]. Fiegl, F., Spot tests in organic analysis, 1961, 6th edition, Elsevier, London.

[17]. Koltoff, I.M and Elving, P.J., Edt. “Treatise on Analytical Chemistry”, 1983, Part I,

Vol.3, John Wiley and Sons, New York.

[18]. Skoog, D.A., Principles of instrumental methods of analysis, 1985, 3rd

Edition

(Holt Saunders International Editions).

[19]. Ringbom, A., Z. Anal. Chem., 939,115, 137.

[20]. Marcozenko, Z., Spectrophotometric determination of elements (Ellis Horward,

CTD, Eng), 1976, pp. 19.

[21]. Sand ell, E.B., Colorimetric determination of traces of metals (lnter science, New

York), 2nd Ed., 1950, pp. 367.

[22]. R.A. Day, Jr. and A.L. Underwood, Quantitative analysis 6th

Edn, 1998, pp 400.

[23]. R.S. Khandpur, Handbook of Analytical Instruments, 2nd

edition, Tata McGraw-Hill

Pub., 2006, pp 24-31.

[24]. Kamata, E., Bull. Chem. Society, Japan.1964, 37, 1674.

[25]. Kammos, E. and Michael, E., Arch. Pharm. Chem. Sci. Ed., 1982, 10, 146.

[26]. Pays, M. and Bourdon, R., Ann. Pharm, Franc, 1968, 26, 681.

[27]. Kamata, E., Bull. Chem. Society. Japan,1965, 38, 2005.

[28]. Pays. M., Malergean, P and Bourdon, R., Ann. Pharm Franc, 1966, 24, 763.

[29]. George, JN., Anal. Chem.,1981, 53, 1708.

[30]. Pays, M and Beligean, M., Ann. Pharm, Franc, 1970, 28, 153.

[31]. Sawicki, E., Schumachar, R. and Engel, C.R., Mikrochem.J.1967, 12, 377.

Page 41: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 41

[32]. Pays. M., Baligean M., and Bourdon, R , Anal.Chim.Acta, 1969, 47, pp.101.

[33]. Sawicki, E., Stanley, T.W., Hauser. And Noe, J.L., Anal. chem.1961, 33, 722.

[34]. Pay.M., Ann.Pharm.Franc., 1967, 25, 29.

[35]. Murst. E.R. and Seltine. M.J., Anal. Biochem., 1981,115, pp. 88.

[36]. Friestad. H.O., Ott. D.E. and Gunther, F.A., Anal Chem., 1969, 41, pp.1750.

[37]. Umenda, M., Yakugaku Zasshi, 1963, 83, page 951.

[38]. Passey, M. and Baltos, J., Ann. Pharm. Franc.1970, 28, pp153.

[39]. Sebbort, W.S., Photogr. Sci.,1969, 17, pp13.

[40]. Sastry, C.S.P., and Sastry, B.S., “The Eastern Pharmacist”, 1986, 29[345], pp 31.

[41]. Siddappa, K., Mllikarjun, M., Tukaram, R., Mahesh, T and Mllikarjun, G,

World J Chem, 2009, 4, pp15.

[42]. Prakash, S S., Gowrisankar, D., Abhay, S., and Abdul, F, Int j Chem Tech Res,

2010, 2, pp 282.

[43]. Anindita, B Swapan Kumar, M., Sudam Chandra, S and Dannana Gowri sankar,

Quim Nova, 2011, 34, pp 1349.

[44]. Revanasiddappa, H D, Manju, B and Ramappa, P G, Analytical Sciences, 1999, 15,

661-664.

[45]. Rahul reddy, M B, Gurupadayya, BM and Anilkumar, T, Indian journal of chemical

technology, Nov-2011,18, 431-438.

[46]. http:// en.wikipedia.org / wiki / Ion-association.

[47]. Y.M. Issa, F.M. Abdel-gawad, M.A. Abou, Anal. Letter, 1997, 30, 2071.

[48]. Sastry C.S.P., Ramarao K., Sivaprasad D., Talanta, 1995, 42, 311.

Page 42: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 42

[49]. IUPAC Compendium of chemical terminology, 2nd

Edition, 1997, available at

http://www.IUPAC.org/gold book/I03231.pdf.

[50]. Connors K.A., A text book of pharmaceutical analysis, 2007, 3rd

edition Wiley India

(p) Ltd, New Delhi.

[51]. Remington; The science and practice of pharmacy, 2006, 21st edition Vol.I,

Lippincott Williams and Wilkins, USA.

[52]. Higuchi T., Roubal E., Division of analytical chemistry, ACS Abstract papers,

April-1965, 149 Th meeting, 28B.

[53]. Foster, R. Molecular Complexes, Vol. I (1973) and Vol.II (1974) edt., Eleck

Science, London.

[54]. Foster, R., J. Phys.Chem., 1980, 84, pp. 2135.

[55]. Yarwood, J. (Ed.), Spectroscopy and structure of molecular complexes,1973

Plerum Press, London,.

[56]. Sastry, C.S.P., Reddy, B.S. and Mangala, D.S., Indian Drugs, 1984, 21, pp.526.

[57]. Das Gupta,V., Indian J.Pharm Sci., 1973, 35, pp 77.

[58]. Shingbai, D.M., Ibid., 1973, 35, pp 160.

[59]. Kyoji, T., Shoh, M. and Toru, U., Talanta, 1982, 29, pp 103.

[60]. Reddy, B.S., Sastry, C.S.P., J. Inst. of Chemists, 1983,5, pp 69.

[61]. Sastry, B.S., E.V.,Tummruru, M.K. and Sastry,C.S.P., Indian Drugs 1986, 24(2),

pp. 105.

[62].Vishwanadham, N., Reddy, M.N. and Sastry, C.S.P., IndianJ.Pharm.Sci.1983, 45,

pp. 81.

[63]. Fogg, A.G., Burgess, C. and Thorburn, B.D., Talanta, 1971, 18, pp 1175.

[64]. Eegriwe, E., Z. Anal Chem., 1927, 70, pp. 400.

Page 43: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 43

[65]. Nayak, A.N., Ramappa, P.G., Yathirajan, H.S. and Manjappa, S., Anal. Chem Acta.

1982, 134, pp. 411.

[66]. Saeed A., Haque S., Quereshi S.Z., Talanta, 1993, 40(12), pp. 1867.

[67]. Sastry, BS, Gananadhamu, S, Prasad, SVSGB and Venugopal Raju, K,

International journal of Pharm Tech Research, 2009, 1(3), 416-419.

[68]. Prashanth, KN, Basavaiah, K and Raghu, MS, ISRN Analytical chemistry, 2013, 7.

[69]. Hussain I, and Shaukat U., J. Chem. Soc, Pakistan, 2002, 24(4), pp. 282.

[70]. Zawilla N.H., Mohammed M.A.A., and El Kousy N.M., J.Pharm.Biomed.Anal,

2002, 27(1), pp. 243.

[71]. El Sheriff Z.A., Walash M.I., El-Tarras M.F., Osman A.O., Anal.Lett, 1997, 30(10)

pp. 1881.

[72]. Moussa B.A., J.Pharm.Biomed.Anal, 2000, 23(6), pp.1045.

[73]. Agrawal R., Am. J. Physiolo-Renal , 2002, 283(2), pp. 236.

[74]. Landszun M., Kovanr K., “Pharm. Acta. Helve”, 1991, 66(9-10), pp. 268.

[75]. Chowdary K.P.R., Kumar K.G., Rao G.D., “Ind.J.Pharm.Sci.” 1999, 61(2), pp.86.

[76]. Mruthunjaya swamy B.H.M., Patil S.M.M., Raju S.A., Ind.J.Pharm.Sci, 2001,

63(5), pp. 433.

[77]. Nagel C.W., Glories Y., “Am. J. Enol. Viticult”, 1991, 42(4), pp. 364.

[78]. Narayana, B, Krishnamurthy, A, Narayana shetty, D and Veena, K., Eurasian

journal of analytical chemistry. 2010; 5(1):63-72.

[79]. Siddappa, K, Mallikarjuna, M, Reddy, PT and Tambe, M., Ecletica Quimica.2008;

33(4): 41-46.

[80]. Gowri sankar, D, Sastry, CSP, and Narayana Reddy, M., Indian drugs,1991, 28(6),

269-273.

Page 44: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 44

[81]. Joseph, BL and Wallace, RB. Journal of American chemical society, 1930; 52(4),

1655-1659.

[82]. Youngman, MA, and Dax, SL., Journal of combinatorial chemistry. 2001; 3(5):

469-472.

[83]. Folin, O. Journal of Biological chemistry.1922, 51,377-391.

[84]. Kumar, CHA, Kumar, TA, Gurupadayya, BM, Sloka, SN, Reddy, BMR., Archives

of Applied Science Research.2010; 2 (4): 278-287.

[85]. Mahmoud, AM, Khalil, NY, Darwish, IA, Aboul-Fadl, Y., International journal of

analytical Chemistry.2009; 2009(810104)

[86]. Al-Momani, IF. , J Pharm Biomed Anal 2001, 25(5-6): 751-757.

[87]. Darwish, IA., Analytica Chemica Acta. 2005; 551(1-2): 222-231

[88]. Darwish, IA, Abdine, HH, Amer, SM, Al-Rayes, LI., International journal of

analytical Chemistry.2009, 2009(237601)

[89]. Darwish, IA, Abdine, HH, Amer, SM, Al-Rayes, LI., Spectrochimica Acta A. 2009;

72(4):897-902.

[90]. Smith, J R L., Smart, A U., Hancock, F E and Twigg, M V , J .Chromatogr, 1989,

483, 341.

[91]. Kobayashi, Y., Kubo, H., Kinoshita, T, Anal. Bio Chem., 1987, 160,392.

[92]. Saurina, j and Hern’andez-Cassou, S, J .Chromatogr.A, 1994, 676,311.

[93]. Edder, P., Cominoli, A and Corvi, C, J .Chromatogr.A, 1999, 830,345.

[94]. Nakahara, Y and Takeda, Y, Chromatographia, 1988, 26,363.

[95]. Ulu, T S, Chromatographia, 2006, 64,169.

[96]. Ulu, T S, J. Pharm. Biomed.Anal, 2007,43,1444.

Page 45: Chapter-1shodhganga.inflibnet.ac.in/bitstream/10603/22702/7/07... ·  · 2014-08-19newer techniques. The present trend includes the employment of spectrophotometry, chromatography,

INTRODUCTION

Page 45

[97]. Anilkumar, T, Gurupadayya, BM, Rahul Reddy, MB, Indian journal of chemical

technology, Jan-2012, 19, 56-62.

[98]. Cetin, G and Sungur, S, Sci. Pharm, 1995, 63, 93-98.

[99]. Meyyanathan, SN, Philip, M, and Suresh, B, Indian drugs, 1998, 35, 183-188.

[100]. Meyyanathan, SN, Sebastian,M and Suresh, B, Indian drugs, 1998, 35, 344-347.

[101]. Gallo-Martinez, L, Sevillano-Cabeza, A, Campíns-Falcó, P and Bosch-Reig, F,

J. Chrom.B Biomed. Sci. Appl, 1998, 718, 143-151.

[102]. Wang, H Y, Xu, L X, Xiao,Y and Han, J, Spectrochim. Acta A Mol .Biomol.

Spectros, 2004, 60, 2933-2939.

[103]. Xu, L , Wang, H, and Xiao, Y , Spectrochim. Acta A. Mol .Biomol. Spectros, 2004,

60, 3007-3012.

[104]. Mohamoud, AA, Khalil, NY, Darwish, IA and Aboul-Fadl, T, International

journal of analytical chemistry, 2009, doi: 10.1155/2009/810104.

[105]. Prasad, Ch. H, Chandy Sushma, P, Devala Rao, G and Sudhakara sai Babu, G,

Orient. J. Chem., 2010, 26, 1211-1213.

[106]. Aswani, CH, Kumar, T, Anilkumar, BM, Gurupadayya, S Navya, S and Rahul

Reddy, MB, Arch. Appl. Sci. Res., 2010, 2, 278-287.

[107]. Walashs, M I, Belal, FF, El-Enany, NM, and El-Maghrabey, MH, Chem. Cent. J,

2011, 5, 36-45.

[108]. Sowjanya, K, Thejaswini, JC, Gurupadayya, BM and Indupriya, M.

DerPharmacia Lettre, 2011, 3, 47-56.

[109]. Gouda, AA, Hashem, H, and Hassan, W, Drug Test. Anal, 2011.

[110]. Chavala, AK, Bannimath, G, Sama, NS and Prasanth, KVR, Int. J. Pharm. Tech,

2011, 3, 1750-1767.