Engineering\CADD Systems Office CADD Manager's Series Site Integration.
CADD -Solis Ambulatory Infusion Pump with Wireless …/media/M/Smiths-medical... · 2019. 11....
Transcript of CADD -Solis Ambulatory Infusion Pump with Wireless …/media/M/Smiths-medical... · 2019. 11....
CADD®-Solis Ambulatory Infusion Pump with Wireless Communication
Safe, Simple, Smart and Secure. Now Smarter with Wireless
CADD®-SOLIS AMBULATORY INFUSION PUMP WITH WIRELESS COMMUNICATIONA SMART SYSTEM THAT COMBINES SMART CONNECTIVITY WITH SAFE AMBULATORY INFUSION
Wireless communication transforms the CADD®-Solis system from a standalone pump to an integrated system with PharmGuard® Server software that helps increase efficiency and safety within hospitals, decrease cost, and improve outcomes.
• Provide actionable reports when you need them, where you want them
• Deploy protocol libraries and software updates from the server to the pump
• Helps improve safety and effectiveness of drug protocols with PharmGuard® Device Reports
• Monitor drug dosing
• Enable patient mobility to assist in faster recoveries and reduced length of stay and complications
• Note: The wireless feature can be turned off or on without affecting pump delivery. Wireless connection does not impact pump delivery or data retention.
PHARMGUARD® SERVER SOFTWARE CAN REPORT ACROSS YOUR FACILITY OR EVEN BETWEEN MULTIPLE FACILITIES WITHIN THE SAME WIRELESS NETWORK
The PharmGuard® Server software connects your pumps, libraries and stakeholders to create reports by group, facility, library, profile and date.
Medfusion® Pumps
CADD®-Solis Pumps
PharmGuard® Server Software
• View separate facilities• Review protocol compliance
and utilization• Drill through reports for more
detail• Analyze safety events
• Customize pain management reports
• Track pump location to last known wireless access point
• Automate delivery of customized reports to email at preferred intervals
All Three Facilities
Facility 1 Facility 2 Facility 3
PHARMGUARD® DEVICE REPORTSCONNECTINGINFUSING
A SMART AMBULATORY INFUSION PLATFORM THAT JUST GOT SMARTER
STANDARDIZE ON THE VERSATILITY, FLEXIBLITY AND MOBILITY OF THE CADD®-SOLIS SYSTEM PLATFORM FOR ALL YOUR INFUSIONS
Mobility by Design• Exclusive CADD™ medication
cassette reservoirs promote patient mobility
• Seamless pump information collected and reported throughout the hospital network
• Post-Op• Labor & Delivery• ICU• Trauma
• Pediatrics• Oncology• ER• And more
Flexibility of Care AreasAdvance infusions across in-patient and out-patient care settings, like:
• Epidural • Peripheral Nerve
Blocks• Local Infiltration
Analgesia
• Subcutaneous• Intrathecal• IV PCA• And more
Versatility of TherapiesVisually differentiate routes of administration with color-coded screens
50 mL
1 mg/hr
Standard dose
50 mL
PCA Dose 0.5 mgPCA Lockout 15 Min1 Hour Limit 2.5 mg
IV Acetaminophen
IV PCA
Stopped
Stopped
Continuous 3:21 PM Chemo - 4 Day Infusion5-FU (4-day) in 50mL mLChemo - 4 Day Infusion5-FU (4-day) in 50mL m
Exit Accept Value
'Accept Value', or press 'select' to edit
Continuous Rate 1 mL/hr
KVO Rate 0 mL/hrReservoir Vol. 50 mL
Wireless Off Wireless Off (white indicator)
Wireless On
Wireless On, not associated with wireless access point (orange indicator)
Wireless On, associated with wireless access point (orange indicator with green center)
Wireless On, associated with wireless access point, and communicating with PharmGuard® Server software (green indicator with green center)
Wireless communication status at a glance New with version 4
COMBINES SAFE AMBULATORY INFUSION WITH SMART CONNECTIVITY AND VERSATILITY
Immediate visual indication of pump status with color-coded status bar
Drug protocol always displayed on Protocol Bar with selected Therapy, Qualifier and Drug
Easy navigation with soft key interface
Promotes clinician confidence with pump lock status indicator
Helps reduce medication errors with clearly displayed infusion delivery parameters with unit of measure
Intuitive programming with step-by-step instructions on color-coded screen
Effortless disposable attachment with cassette latch lever
Helps reduce numeric key press errors with scrolling keys
Convenient PCA dosing while ambulatory with PCA dose key or handheld remote dose cord
Dosing graph Delivery chart
ADVANTAGES OF THE CADD®-SOLIS AMBULATORY INFUSION SYSTEM
SMART PUMP FEATURES HELP PROMOTE CLINICIAN CONFIDENCE AND PATIENT SAFETY
• Task-based interface helps improve program efficiency and user confidence
• On-screen color graphs and trending data provide immediate patient assessment and helps support the clinician at bedside
• On-board library provides up to 500 protocols that can be edited within the safety limits for specific patient needs
• Pump programming errors are prevented by:
• Soft and hard maximum and minimum programming limits• Tall man/Short man with no trailing zeroes• Support for full drug names
SAFE. SIMPLE. SMART. NOW SMARTER WITH WIRELESS
The performance of any system comes from the quality of its components. The CADD®-Solis infusion pump platform includes innovative products that are known for delivering a high standard of performance and safety.
CADD®-Solis Ambulatory Infusion Pump and Communication ModuleThe pump is still the heart of the system, now with a wireless Communication Module (CM), 802.11 a/b/g radio and Li-Ion battery.
CADD™ Medication Cassette ReservoirsProvide medication security and free-flow protection of fluid delivery.
PharmGuard® Administrator Medication Safety Software PC software for developing protocol libraries. Version 4 now supports longer drug names and profiles
PharmGuard® Device Reports SoftwareProvides Continuous Quality Improvement (CQI) reports that enable data-driven analysis.
PharmGuard® Server SoftwareLoaded onto hospital servers to store infusion data and communicate to and from CADD®-Solis infusion pumps.
PharmGuard® SureLink® Remote Support SoftwareProvides secure support and distribution of infusion pump firmware and PC/server software.
Clinician Bolus 0 to 50 mL (or mg or mcg equivalent)
Max Fluid Delivery Rate
Intermittent Bolus, Clinician Bolus, and PCA Dose: with standard volume tubing: 40-250 mL/hr with high volume tubing: 40-500 mL/hr
Delivery Limit Method
Delivery limit, max doses per hour or not in use
Delivery Limit Period
1 to 12 hours in increments of 1 hour
Power Sources 4 AA (IEC LR6) alkaline batteries; AC adapter; rechargeable battery pack, or Communication Module
Battery Life (Alkaline)
Approximately 113 hours at 10 mL/hr4 hours with Communication Module
Moisture Protection
Splashproof (IPX4) per IEC 60529
Event Log 5000 events
Air Detector, Integral
Turn on or off
Air Detector, Sensitivity
Low – single bubble greater than 2 mL High – single bubble greater than 150 uL or <1 mL over 15 minutes - 2 mL of air / 15 min and multiple bubbles that accumulate >1mL of air over 15 minutes
Alarms and Messages
Multiple customizable alarms are all color coded, many with options to “acknowledge” or “silence”. Alarms include high priority; medium priority; low priority; informational messages and system fault alarms
Alarm Volume High, medium, low55db (+/-8dB) as measured at high volume
Alarm Sound Theme
Standard, intense, distinctiveDistinctive theme is designed to meet the audible tones and patterns set forth by the Medical Device Safety Standards for Alarm Systems
High Pressure Alarm
18 +/- 9 psi
Graphs PCEA/PCA dose graph; Intermittent Bolus graph, PCEA/PCA and Intermittent graph, Delivery history and pie chart
Reports Intermittent Bolus status, Given and Dose counters, delivery log, event log, protocol library summary, device information, daylight savings time, current profile/protocol, wireless status, wireless setting, PCA dose graph.
OtherComponents
Scroll keys; task-based soft key user interface; cassette latch; cassette/keypad lock; indicator lights; continuous backlight display with AC power; data interface port; remote dose cord; upstream and downstream occlusion sensors; polemount adapter; CADD™ medication cassette reservoir; CADD®-Solis lockboxes; compatible with CADD™-Solis medication safety software
Indications
Intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, and/or an intermittent bolus, and/or with patient-controlled demand doses, and/or clinician boluses.
Pump Size 1.6 in. x 4 in. x 5 in. excluding cassette or other accessories. 1.6 in. x 4 in. x 7 in. with CM attached
Screen Size 2.12 in. x 2.12 in. (320 pixels x 320 pixels)
Weight 21 oz. (595g) including 4 AA alkaline batteries, excluding other accessories.
On-board Protocol Library
Stores up to 500 protocols
Security Cassette/keypad lock and three customizable security access levels: keypad code; clinician code; administrator code
Accuracy +/- 6% (nominal)
Delivery Methods
Continuous rate; PCA dose; Clinician bolus; Intermittent Bolus
Reservoir Volume
0 to 9999 mL; programmable in 1 mL increments, displayed in 0.1 mL increments
Units Milliliters (mL), milligrams (mg), micrograms (mcg)
Concentration mg/mL: 0.1 to 0.5 mg/mL in increments of 0.1 mg/mL 0.5 to 1 mg/mL in increments of 0.5 mg/mL 1 to 15 mg/mL in increments of 1 mg/mL 15 to 100 mg/mL in increments of 5 mg/mLmcg/mL: 1 to 15 mcg/mL in increments of 1 mcg/mL15 to 100 mcg/mL in increments of 5 mcg/mL100 to 500 mcg/mL in increments of 100 mcg/mL
Continuous Rate 0 to 100 mL/hr (or the mg or mcg equivalent)
PCA Dose 0 to 50 mL (or mg or mcg equivalent) Intermittent Bolus, Clinician Bolus, and PCA Dose: with standard volume tubing: 40-250 mL/hr with high volume tubing: 40-500 mL/hr (high volume tubing compatible with Model 2110 pumps only)
PCA Dose Lockout
1 minute to 24 hours in the following increments: 1 minute for values between 1 and 20 minutes 5 minutes for values between 20 minutes and 24 hours
Max Doses Per Hour
1 to 60
Delivery Limit Amount
0.1 to 1000 mL (or the mg or mcg equivalent) in increments of: 0.01 mL from 0.01 to 0.5 mL 0.5 mL from 0.5 to 100 mL 1.0 mL from 100 to 1000 mL
Intermittent Bolus
0 mL to 50 mL (or the mg or mcg equivalent)
Intermittent Bolus Interval
0 to 4 hours
Next Bolus 0 to 4 hours
Given 0 to 99,999.99 in 0.01 unit increments
SPECIFICATIONS & REQUIREMENTS
CADD®-Solis Ambulatory Infusion Pump Specifications
SPECIFICATIONS & REQUIREMENTS
PharmGuard® Medication Safety Software v4.1 System Requirements
The following are the minimum system requirements for the PharmGuard® Administrator or PharmGuard® Point of Care software applications:
Operating systems (32-bit)
Microsoft® Windows® XP Professional (SP3)Microsoft® Windows Vista® Business Edition (SP2)Microsoft® Windows® 7 Professional (SP1)Microsoft® Windows® 8 Professional
Operating systems (64-bit)
Microsoft® Windows Vista® Business Edition (SP2)Microsoft® Windows® 7 Professional (SP1)Microsoft® Windows® 8 Professional
Supported operating system languages
English, Danish, Spanish, French, German, Italian, Japanese, Dutch, Portuguese, Swedish, Norwegian, and Finnish.
Server Microsoft® SQL Server® 2005 or 2008 or 2012 [Required if installing the PharmGuard® software database onto your own Microsoft® SQL Server® instance. Microsoft® SQL Server® may be installed on Microsoft® Windows Server® 2003 or 2008 SP2 (x64) or 2012 which may be optionally hosted using VMware® ESXi virtual server technology.)
RAM 256 MB
Hard disk space 70 MB (application); 350 MB (database)
Ports USB port and USB Type A to Type Mini-B cable (for connection to a CADD®-Solis or CADD®-Solis VIP Ambulatory Infusion Pump)
Display 1024 × 768 display with 16-bit video card
Navigation Mouse or other pointing device
Reports Adobe® Reader® 9.4 or greater (for viewing reports exported in PDF format)
The following items are required if using the barcode reading and label printing features of PharmGuard® software. For information about installing and configuring barcode reading or printing equipment, refer to the manufacturer’s instructions.
A barcode scanner configured as a Human Interface Device (HID) and compatible with Code 128 barcode
A printer capable of printing a 1.5 inch (3.81 centimeters) high by 3 inches (7.62 centimeters) wide label as an image
USB ports for connection to the barcode reader and/or barcode label printer
PharmGuard® Server Software v2.1 System Requirements
PharmGuard® Server Software
Install on hospital-selected vendor independent hardwareMicrosoft® Windows Server® 2008 R2/2012 (native or virtualized)Requires Microsoft®.NET Framework v4.5 Supports VMWare® ESX 5.X, and Microsoft® Windows Server® 2008 Hyper-V®
Transactions: XML/SOAP via secure sockets using SSL
Single Server Configuration
Hardware Combined web, database, and medical device application server with:
Two 64-bit processors designed or certified for Microsoft® Windows Server® 2008 R2 or Windows Server® 201248 GB of RAM with 16 GB reserved for SQL Server®
1TB GB for data storage for 1,000 devices at full utilization for two years
Software Microsoft® Windows Server® 2008 R2/2012Microsoft® Internet Information Server (IIS)Microsoft® SQL Server® 2012Microsoft® .NET Framework v4.5
Multi-Server Configuration
Separate database server
Hardware Two 64-bit processors designed or certified for Microsoft® Windows Server® 2008 R2 or Windows Server® 201216 GB reserved for SQL Server®
2TB for data storage for 2,000 devices at full utilization for two years
Software Microsoft® Windows Server® 2008 R2/2012Microsoft® Internet Information Server (IIS)Microsoft® SQL Server® 2012Microsoft® .Framework v.4.5
Separate webserver
Hardware Two 64-bit processors designed or certified for Microsoft® Windows Server® 2008 R2 or Windows Server® 201216 GB of RAM 40 GB+ of available disk space
Software Microsoft® Windows Server® 2008 R2/2012Microsoft® Internet Information Server (IIS)Microsoft® .NET Framework v4.5
Separate medical device application server (MDS)
Hardware One 64-bit processor designed or certified for Microsoft® Windows Server® 2008 R2 or Windows Server® 201216 GB of RAM 120 GB+ of available disk space
Software Microsoft® Windows Server® 2008 R2/2012Microsoft® .NET Framework v4.5
Separate PharmGuard® Interoperability Platform server software
Hardware Two 64-bit processors designed or certified for Microsoft® Windows Server® 2008 R2 or Windows Server® 201232 GB of RAM450 GB+ available disk space
Software Microsoft® Windows Server® 2008 R2/2012
PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES
Smiths Medical ASD, Inc.6000 Nathan Lane NorthMinneapolis, MN 55442, USATel: 1-614-210-7300Toll-Free USA: 1-800-258-5361www.smiths-medical.com
Find your local contact information at: www.smiths-medical.com/customer-support
Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings and precautions. CADD, PharmGuard and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trademarks or service marks of their respective owners. ©2017 Smiths Medical. All rights reserved. IN193500EN-082017
802.11 a/b/g Network Configurations Supported
Network Mode Select the network mode to match the network:802.11 a/b/g802.11 b/g802.11 a
Authentication Mode
OPENWPA-EnterpriseWPA2-EnterpriseWPA-PersonalWPA2-Personal
EAP Mode NoneTLSTTLSPEAPV0 (PEAP with MS-CHAPv2)PEAPV1 (PEAP with GTC)LEAP
NOTE: EAP mode is not applicable for OPEN, WPA-Personal, WPA2-Personal authentication mode.
Encryption Mode
The encryption mode for the network:TKIPCCMP
Wireless Encryption Key
You can either enter a key consisting of 64 characters or a passphrase consisting of between 8 and 63 printable ASCII characters.
NOTE: This field is not used with Enterprise authentication modes.
Identity Enter your identity (if necessary)
Client Certification
Client Certificate: Browse to select your client certificate file (in .pem format with maximum size of 4000 bytes)Client Key: Browse to select your client key file (in .pem format with maximum size of 2000 bytes)Private Key Password: Enter your private key passwordCertificate of Authority: Browse to select your certification authority certificate file (in .pem format with maximum size of 4000 bytes)
SPECIFICATIONS & REQUIREMENTS
MPAUCA-1631
Contact your sales representative at 1-800-258-5361 orfor more information visit our website at www.smiths-medical.com
CADD®-Solis Ambulatory Infusion Pump with Wireless Communication
Safe, Simple, Smart and Secure. Now Smarter with Wireless
Wireless Communication Module Specifications
Battery type Rechargeable Lithium-ion battery (2 cells)Equivalent Lithium content <1.6 g
Battery capacity 18.7 Wh (when fully charged)
Battery output 3.6V (nominal)
Rechargeable battery weight
<100 g
Product battery life
400 full cycles (charge / discharge) at room temperature, 23°C (73°F)Battery discharge cycle of 4 hours at nominal condition
Maximum charge time (low battery to full charge)
10 hours when plugged into AC Adapter