30956_Williams_2-2008_lo_res

download 30956_Williams_2-2008_lo_res

of 32

Transcript of 30956_Williams_2-2008_lo_res

  • 8/8/2019 30956_Williams_2-2008_lo_res

    1/32

    S E R I E S O F M E T E R I N G P U M P S

    DUAL SEAL PLUNGER

    V

    INSTALLATION, OPERATION& MAINTENANCE MANUAL

  • 8/8/2019 30956_Williams_2-2008_lo_res

    2/32

    P125V125P250V225P250V300

    P500V225P500V300P500V400(1)

    P750V400(1)

    P1000V400(1)

    P1000V600(1)

    P1000V800(1)

    NOTE: (1)The 400, 600 and 800 motor cylinders are only available with the CR (controller-relay) or SR (solenoid-relay) control methods.

    CR P1000V800 A PE

    C: ControllerCR: Controller - Relay

    S: Solenoid ValveSR: Solenoid Valve - Relay

    Control

    Method

    BasicModel(2)

    A: 17-4 phB: 316 SS

    CR: Ceramic

    PlungerMaterial

    PE: Polyethylene (UHMW)TC: Teflon Composite

    TG: Teflon GraphiteV: Viton

    BR: Buna NK: Kalrez

    EPR: Ethylene Propylene

    SealMaterial

    PART NUMBER DESIGNATION

    S E R I E S O F M E T E R I N G P U M P SDUAL SEAL PLUNGER

    V

    11339 M ROY LOTA 7/18/08 10:18 AM Page 2

  • 8/8/2019 30956_Williams_2-2008_lo_res

    3/32

    TABLE OF CONTENTS

    Section Description Page

    1.0 FUNCTIONAL DESCRIPTION 1

    1.1 Physical Description 1

    1.2 Capabilities 3

    1.3 General Operating Sequence 9

    2.0 INSTALLATION OF PUMP AND CONTROLLER 10

    2.1 General 10

    2.2 Pump Assembly 10

    2.3 Typical Installation 11

    2.4 Supply Reservoir 11

    2.5 Relief Valve11

    3.0 START UP, OPERATION & SHUTDOWN 11

    3.1 General 11

    3.2 Startup 11

    3.3 Operation 12

    3.4 Shutdown and Storage 12

    4.0 MAINTENANCE 12

    4.1 General 12

    4.2 Disassembly and Assembly 13

    4.3 Preventive Maintenance 23

    4.4 Troubleshooting 23

    5.0 PARTS LIST AND REPAIR KIT ORDERING REFERENCE 26

    5.1 Controller Part Lists and Repair Kits 26

    5.2 Relay Part Lists and Repair Kits 26

    6.0 LIMITED WARRANTY 26

    11339 M ROY 3_32 7/18/08 10:53 AM Page 3

  • 8/8/2019 30956_Williams_2-2008_lo_res

    4/32

    SECTION 1.0: FUNCTIONAL

    DESCRIPTION

    1.1 PHYSICAL DESCRIPTION

    The letters CP at the beginning of a pump

    assembly model number mean that the

    model consists of a controller (C) and

    pump (P) For example: CP125V125,

    CP250V225, CP250V300, CP500V225, and

    Control Methods For The Pump

    1

    1.1.1 Controller

    The controller, consisting of an upper and

    lower chamber separated by a sliding

    spool, uses a capillary tube with a needle

    valve to transfer the supply air/gas from

    the lower to the upper chamber. When the

    sliding spool is in its highest position, a

    pilot plug closes a vent and moves the

    supply air/gas to the pump or relay. In the

    spools lowest position, the reverse is true;

    the pilot plug prevents supply air/gas from

    CP500V300. The letters CRP mean that the

    model also has a relay (R). For example:

    CRP500V400, CRP750V400, CRP1000V400,

    CRP1000V600, and CRP1000V800.

    Some pumps use different relay sizes:

    CRP500V400, CRP750V400,CRP1000V400, and CRP1000V600

    pumps use the PO3-6S relay.

    CRP1000V800 pump uses the PO4-6S

    relay.

    The PNEUMATIC RELAY is a pilot operated valve designed to provide the higher

    air or gas flow rates necessary for PNEUMATIC DRIVE CYLINDER

    diameters greater than 3 inches. The PNEUMATIC RELAY is actuated by the pulses produced by

    the CONTROLLER. A single acting PNEUMATIC RELAY is used with pumps that have return springs as illustrated to the left.

    The air or gas pressure is required to return the PISTON-PLUNGER ASSEMBLY on the CRP1000V800. Therefore a double

    acting PNEUMATIC RELAY is required, illustrated to the right.

    The MK XIIA Controller operates on the same operating principal as the MK X Controller.

    The MK XIIA has the same upper and lower chambers, but are separated with flexible

    diaphragms rather than sliding seals. A capillary tube, controlled by a needle valve,

    transfers the air/gas supply to the pump from the lower to the upper chamber.

    When the spool is in the highest position, a pilot plug closes a vent and opens thesupply air/gas to the pump. When the spool is in its lowest position, the pilot plug

    prevents the supply air/gas from entering the pump, and opens the air/gas vent to let it

    exhaust the pump. The spool then returns to its highest position to repeat the process.

    1. Oscillamatic Controller

    STROKE RATECONTROL KNOB

    NEEDLE VALVE

    CONTROL SPOOL

    AIR/GAS SUPPLYTO CONTROLLER

    SUPPLY TO PUMP1/4 NPT

    EXHAUST VALVE

    OPTIONAL3-15CONTROL PORT(MK XIIB) 1/8 NPT

    CONTROLAIR PASSAGE

    UPPER DIAPHRAGM

    MIDDLE DIAPHRAGM

    LOWER DIAPHRAGM

    SUPPLY VALVE

    EXHAUSTFROMPUMP

    1/4 NPT

    3. Solenoid Valves

    The pumps can be automated by replacing the

    CONTROLLER with a 3-way electro-pneumatic

    SOLENOID VALVE. The SOLENOID VALVE can be

    cycled in order to achieve the desired pump output.

    Flow tracking can be accomplished by having a

    FLOWMETER or PH METER signal interpreted by our

    WPC9001 or a PLC. The typical arrangement for a

    WPC-9001 installation is shown at right.

    CONTROLLER

    PO3-6S RELAY

    CONTROLLER

    P04-6S RELAY

    PILOT PULSES FROMCONTROLLER

    EXHAUST

    SINGLE ACTINGPO3-6S DOUBLE ACTING

    PO4-6S

    PUMP

    SUPPLYPRESSURE(REAR)

    SUPPLY PRESSURE

    EXHAUST NO. 1

    PUMP POWERSTROKE NO. 1

    PUMP RETURNSTROKE NO. 2

    EXHAUST NO. 2

    PUMP INSTALLATIONCRP500V400,CRP750V400

    CRP1000V400CRP1000V600

    PUMP INSTALLATIONCRP1000V800

    PILOT PULSESFROM CONTROLLER

    2. Controller-Pneumatic Relay Combination

    11339 M ROY LOTA 7/18/08 10:18 AM Page 4

  • 8/8/2019 30956_Williams_2-2008_lo_res

    5/32

    At the time the controller or relay exhausts

    the air/gas, if the pump has a motor return

    spring, fluid pressure in the fluid chamber

    and the motor return spring force the

    piston plunger upwards. If the pump has

    no motor return spring, supply pressure

    from a relay enters beneath the piston to

    help force the piston/plunger assembly

    upwards. In either case, the plunger

    assembly stops when the piston contacts

    the end of the stroke adjuster.

    The piston face area where the air/gas

    pressure is applied is much greater than

    the plunger face area, which works against

    the pressure of the process fluid. This area

    ratio, called the amplification ratio,

    allows the pump to work against process

    fluid pressures much greater than the

    air/gas supply pressure.

    Note: process pressure + 200psi /required air supply = amplification ratio

    air/gas pressure. This relay has two poppet

    valves. The upper valve operates in the

    same way as the PO3-6S relay valves,

    while the lower valve operates in the

    opposite order. That is, when the upper

    valve applies air/gas pressure to the pump,

    the lower valve exhausts it, and when the

    upper valve exhausts the air/gas, the lower

    valve applies pressure to the opposite side

    of the piston.

    1.1.3 Pumps

    The pneumatic V Series metering plunger

    pump consists of a body, fluid chamber,

    and a motor cylinder or air chamber,

    separated by a piston/plunger assembly

    and seals. In the fluid chamber, the inlet

    (suction) and discharge (outlet) ports have

    check valves that control the direction of

    the fluid flow. In the motor cylinder, the

    air/gas supplied from the controller or relay

    enters and exhausts through a pipe nipple

    that connects the controller or relay to the

    motor cylinder.

    35 TO

    100 PSIG

    50 TO

    150 PSIG

    CONTROLLER

    RELAY

    DISCHARGE CHECK VALVE

    SUCTION CHECK VALVE

    35 TO

    100 PSIG

    CONTROLLER

    DISCHARGE

    CHECK VALVE

    SUCTION CHECK VALVE

    35 TO

    100 PSIG

    50 TO

    150 PSIG

    CONTROLLER

    RELAY

    SUCTION CHECK VALVE

    DISCHARGE CHECK VALVE

    TYPICAL CRP1000V800(WITH PO4-6S RELAY)

    TYPICAL CRP500V400, CRP750V400, CRP1000V400 & 600(WITH PO3-6S RELAY)

    TYPICAL CP125V, CP250V, CP500V(WITH 125, 225 & 300 CYLINDERS)

    2

    entering pump or relay, and opens the vent

    to let it exhaust. The spool then returns to

    its highest position and repeats the cycle.

    1.1.2 Relays

    A relay is a pilot operated servo valve.

    Air/gas pressure from a controller forces apiston and one or more attached poppet

    valves to move. Then, when the controller

    removes the pressure, a spring forces both

    the piston and poppet valve(s) in the

    reverse direction.

    Relay models differ in operation. The

    PO3-6S relay has one poppet valve that

    connects the supply air/gas to the pump

    and blocks the exhaust when the controller

    applies air/gas pressure to the relay piston.

    When the controller removes pressure

    from the piston, the poppet valve stops the

    supply air/gas and connects the pump to

    the exhaust.

    The PO4-6S relay, used on pumps that do

    not have a motor return spring, drives the

    air piston both up and down with the

    11339 M ROY LOTA 7/18/08 10:18 AM Page 5

  • 8/8/2019 30956_Williams_2-2008_lo_res

    6/32

    3

    1.2.1.1 Controller Supply Pressure

    Controllers will operate with the following

    supply pressures:

    Maximum Minimum

    psi bar psi bar

    MK XII 100 6.9 35 2.4

    To establish the proper air/gas supply

    pressure for the controller or relay, add

    200 psig or 13.8 barg to the process

    pressure the pump is working against.

    Then use the performance graphs located

    in section 1.2.2.2.

    Remember for a controller-relay

    combination the controller supply pressure

    need only be set at the minimum pressure

    of 35 psig or 2.4 barg. The relay supply

    would be established in the above

    procedure or refer to section 1.2.1.2.

    CAUTION: To prevent damage to the

    controller, always use a regulator

    between the supply and the controller

    when the air/gas supply pressure is

    more than the maximum rating of your

    controller.

    1.2.1.2 Relay Supply Pressure

    Relays, which require a separate air

    supply, operate with supply pressuresbetween 35 psig (2.41 bar) and 150 psig

    (10.3 bar). The supply pressure must be 35

    psig minimum or equal to or greater than

    the process pressure plus 200 psi, divided

    by the pump amplification ratio. (EXAMPLE:

    CRP1000V400 @ 1000 PSI + 200 PSI

    divided by 15.38 = 78 air supply required.

    Refer to section 1.1.3 for the amplification

    ratio formula) When using a controller/relay

    combination, hold the air/gas supply

    pressure to the controller to the minimum

    value of 35 psig, and the relay at 35 psig

    minimum.CAUTION: To prevent damage to the

    relay, always use a regulator between

    the supply and the relay when the

    supply pressure is more than 150 psig

    (10.3 bar).

    1.2 CAPABILITIES

    1.2.1 Controllers and Relays

    Controllers and relays will require separate

    supply sources, and fortunately will

    operate with air or any gas, such as carbon

    dioxide, nitrogen or natural gas.

    WARNING: TO PREVENT INJURY, MAKE

    SURE THAT ANY FLAMMABLE GAS

    SUCH AS NATURAL GAS IS PROPERLY

    VENTED FOR SAFETY.

    CAUTION: If the gas could possibly

    damage the standard elastomeric

    material, please contact your

    distributor or Williams Instrument

    Incorporated for advice.

    To increase the process fluid flow rate, two

    or more pumps can be multiplexed: their

    inlets and outlets connected in parallel.

    1.2.2 Pumps

    1.2.2.1 Flow Pressure Performance Table*:

    MAX MAX AIR CONSUMPTIONMODEL MAX VOLUME STROKE STROKES DISCHARGE 100 6.9 150 10.3@ AIR/GAS VOLUME PER LENGTH PER MINUTE PRESSURE PSIG BAR PSIG BARSUPPLY PRESSURE STROKE (RANGE) SCF SCM SCF SCM

    GPH/LPH CC INCH PSIG /BARG PER DAY PER DAY PER DAY PER DAY

    CP125V125

    @ 100 PSI/6.9 BAR .07 /.27 .1 .5 1-45 8650 / 596.4 180 5

    CP250V225

    @ 100 PSI/6.9 BAR .57 /2.16 .8 1 1-45 7200 / 496.4 1163 33

    CP250V300

    @ 100 PSI/6.9 BAR .57 /2.16 .8 1 1-45 13,100 / 903.2 2068 59

    CP500V225

    @ 100 PSI/6.9 BAR 2.26 /8.55 3.2 1 1-45 1750 / 120.7 1163 33

    CP500V300

    @ 100 PSI/6.9 BAR 2.26 /8.55 3.2 1 1-45 3250 / 224.1 2068 59

    CRP500V400

    @ 100 PSI/6.9 BAR 2.26 /8.55 3.2 1 1-45 6300 / 434.4 3676 104@ 150 PSI/10.3 BAR 2.26 /8.55 3.2 1 1-45 9200 / 634.3 5280 150

    CRP750V400

    @ 100 PSI/6.9 BAR 5.00 /18.9 7.0 1 1-45 2600 / 178.3 3676 104 5280 150@ 150 PSI/10.3 BAR 5.00 / 18.9 7.0 1 1-45 4000 / 275.8

    CRP1000V400

    @ 100 PSI/6.9 BAR 9.08 /34.37 12.7 1 1-45 1520 / 104.8 3676 104@ 150 PSI/10.3 BAR 9.08 /34.37 12.7 1 1-45 2300 / 158.6 5280 150

    CRP1000V600

    @ 100 PSI/6.9 BAR 9.04 /34.22 12.6 1 1-45 3400 / 234.4 8273 234@ 150 PSI/10.3 BAR 7.00 /26.50 9.8 1 1-35 4700 / 324.0 9239 262

    CRP1000V800

    @ 100 PSI/6.9 BAR 8.81 /33.35 12.3 1 1-45 6300 / 434.4 14707 416@ 150 PSI/10.3 BAR 6.82 /25.81 9.5 1 1-35 9100 / 627.4 16426 465

    *This data should only be used to provide you with an initial size selection. You must refer to the actual performance graphs in order to verify your pump selection.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 6

  • 8/8/2019 30956_Williams_2-2008_lo_res

    7/32

    4

    1.2.2.2 Using The Graphs*

    Use the following Performance Flow Curves* in order to:

    1) Determine the flow capability of the pump you have selected.If you have sized the pump too close to the upper or lower strokerate limit, you may wish to change to a different pump size. Of

    course you can also change your flow for a given stroke rate byadjusting the stroke length. Example:

    2) Determine the air pressure necessary to provide the desiredpump discharge pressure.

    A) The flow curves show the maximum flow/pressure limit of the pumps.The upper near horizontal line represents the maximum flow capability

    (45 SPM @ 100% stroke). The near vertical lines represent themaximum discharge pressure at the corresponding air/gas pressure.The area under the curve represents the entire flow/pressure range forthe pumps.

    B) THE GRAY CURVE defines the relationship between air/gas supplypressure and discharge pressure. For each discharge pressure thereis a minimum air/gas supply pressure required. Always add 200 PSIto your discharge pressure in order to ensure positive injection.Find the discharge pressure on the horizontal axis and follow it up tothe red curve. At that point, read your air/gas pressure requirementson the right axis in PSIG. The minimum air/gas supply pressure willproduce discharge pressures found to the left of the 35 PSIG limit line.

    The required air/gas supply pressure can be read off the graph by firstadding 200 psi to the discharge pressure. Then locate the psi on thedischarge pressure axis and follow it up until it intersects the red line.Follow this point to the air/gas supply pressure axis on the right and youwill find the appropriate air/gas pressure necessary to operate the pump.

    These Settings Will Produce The Same Pump Flow Rate

    STROKE RATE STROKE LENGTH

    10 1 (100%)20 1/2 (50%)40 1/4 (25%)

    The performance curve graph illustrated here is, in fact, two

    graphs overlayed upon each other. The first, determines theair/gas pressure, illustrated in gray. The second, determinesyour discharge pressure, shown in black. Where these twooverlayed graphs intersect define your optimum flow range.

    0 1000 2000 3000 4000 5000 6000 7000 8000 9000

    0.09

    0.08

    0.07

    0.06

    0.05

    0.04

    0.03

    0.02

    0.01

    0.00

    FLOW

    -GPH

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    DISCHARGE PRESSURE - PSIG

    100

    80

    60

    40

    20

    0

    AIR/GASPRESSUREPSIG

    35

    1 0 0

    P S I G A

    I R / G A S

    35PSI GAI R/G

    AS

    0.054

    50

    MAXFLOW(GPH)

    Reading The Graphs

    11339 M ROY LOTA 7/18/08 10:18 AM Page 7

  • 8/8/2019 30956_Williams_2-2008_lo_res

    8/32

    0 1000 2000 3000 4000 5000 6000 7000 8000 9000

    0.09

    0.08

    0.07

    0.06

    0.05

    0.04

    0.03

    0.02

    0.01

    0.00

    FLOW

    -GPH

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    DISCHARGE PRESSURE - PSIG

    100

    80

    60

    40

    20

    0

    AIR/GASPRESSUREPSIG

    35

    1 0 0 P S I G

    A I R / G

    A S

    35

    PSIG A

    IR/ G

    AS

    0.054

    50

    MAXFLOW(GPH)

    AIR/GAS

    PRE

    SSUR

    E

    0.80

    0.70

    0.60

    0.50

    0.40

    0.30

    0.20

    0.10

    0.00

    FLOW

    -GPH

    0 1000 2000 3000 4000 5000 6000 7000 8000

    100

    80

    60

    40

    20

    0

    AIR/GASPRESSUREPSIG

    35

    DISCHARGE PRESSURE - PSIG

    35PSIG AIR

    100PSIGAIR

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    MAXFLOW(GPH)

    AIR/

    GASPR

    ESSU

    RE

    5

    0 1500 3000 4500 6000 7500 9000 10500 12000 13500 15000

    90

    100

    80

    70

    60

    50

    40

    30

    35

    20

    10

    0

    0.90

    1.00

    0.80

    0.70

    0.60

    0.50

    0.40

    .030

    0.20

    0.10

    0

    FLOW

    -GPH

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    DISCHARGE PRESSURE - PSIG

    AIR/GASPRESSUREPSIG

    100PSIGAIR/G

    AS

    AIR

    /GASPRES

    SURE

    MAXFLOW(GPH)

    35PSIGAIR/GAS

    CP125V125

    Discharge Pressure PSI 0 500 1000 1900 5000 8500Gph .069 .066 .063 .059 .043 .025Air Pressure PSI 35 35 35 35 62 98

    CP250V225

    Discharge Pressure PSI 0 500 1000 1900 5000 6600gph .57 .565 .557 .552 .520 .502Air Pressure PSI 35 35 35 35 70 90

    CP250V300

    Discharge Pressure PSI 0 1000 3000 6000 10000 11000gph .522 .502 .462 .402 .321 .301Air Pressure PSI 35 35 35 47 77 85

    Performance Flow Curves

    11339 M ROY 8 7/18/08 11:01 AM Page 8

  • 8/8/2019 30956_Williams_2-2008_lo_res

    9/32

    6

    160

    140

    120

    100

    80

    60

    40

    20

    150

    0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000

    2.50

    2.25

    2.00

    1.75

    1.50

    1.25

    1.00

    0.75

    0.50

    0.25

    0.00

    DISCHARGE PRESSURE - PSIG

    FLOW

    -GPH

    AIR/GASPRESSURE-PSIG

    35

    3 5

    P S

    I G A I R / G

    A S

    100PSIGAIR/GAS

    150PSIGA

    IR/G

    AS

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    MAXFLOW(GPH)

    AIR/GA

    SPR

    ESSU

    RE

    100

    90

    80

    70

    60

    50

    40

    30

    20

    10

    0 200 400 600 800 1000 1200 1400 1600 1800 2000

    2.50

    2.25

    2.00

    1.75

    1.50

    1.25

    1.00

    0.75

    0.50

    0.25

    0.00

    DISCHARGE PRESSURE - PSIG

    FLOW

    -GPH

    AIR/GASPRESSURE-PSIG

    35

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    100PSIGAIR/G

    AS

    3 5

    P S

    I G A

    I R / G

    A S

    MAXFLOW(GPH)

    AIR

    /GASPRES

    SURE

    0 500 1000 1500 2000 2500 3000 3500

    3.00

    2.75

    2.50

    2.25

    2.00

    1.75

    1.50

    1.25

    1.00

    0.75

    0.50

    0.25

    0.00

    DISCHARGE PRESSURE - PSIG

    100

    90

    80

    70

    60

    50

    40

    30

    20

    10

    0

    AIR/GASPRESSUREPSIG

    35

    FLOW

    -GPH

    PLUNGER SPEED: 45 SPM

    100PSIG

    35PSIG

    MAXFLOW(GPH)

    AIR/GAS

    PRE

    SSURE

    CP500V225

    Discharge Pressure PSI 0 200 450 1000 1400 1500gph 2.26 2.26 2.26 2.23 2.21 2.21Air Pressure PSI 35 35 35 60 82 88

    CRP500V400

    Discharge Pressure PSI 0 1500 3000 5000 7000 8100Gph 2.26 2.20 2.15 2.07 1.99 1.96Air Pressure PSI 35 35 52 80 110 130

    CP500V300

    Discharge Pressure PSI 0 500 800 1500 2000 2500gph 2.26 2.20 2.16 2.08 2.2 1.96Air Pressure PSI 35 35 35 48 62 77

    Performance Flow Curves (cont.)

    11339 M ROY LOTA 7/18/08 10:18 AM Page 9

  • 8/8/2019 30956_Williams_2-2008_lo_res

    10/32

    7

    160

    180

    200

    140

    120

    100

    80

    60

    40

    50

    20

    150

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    10.00

    9.00

    8.00

    7.00

    6.00

    5.00

    4.00

    3.00

    2.00

    1.00

    0.00

    DISCHARGE PRESSURE - PSIG

    FLOW

    -GPH

    AIR/GASPRESSURE-PSIG

    50

    PSIGAIR/ GAS

    100PSIGAIR/GAS

    150PSIGAIR/G

    AS

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    MAXFLOW(GPH)

    AIR/GAS

    PRE

    SSUR

    E

    160

    140

    120

    100

    80

    60

    40

    20

    150

    0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000

    10.00

    9.00

    8.00

    7.00

    6.00

    5.00

    4.00

    3.00

    2.00

    1.00

    0.00

    DISCHARGE PRESSURE - PSIG

    FLOW

    -GPH

    AIR/GASPRESSURE-PSIG

    125

    35

    3 5

    P S I G

    A IR

    /G A S

    150PSIGAIR/GAS

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM & 35 SPM @ 150 PSIG

    1 00P

    SIG

    AIR/G

    AS

    125PSIGAIR/GAS

    MAXFLOW(GPH)

    AIR/GAS

    PRE

    SSUR

    E

    160

    140

    120

    100

    80

    60

    40

    20

    150

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    10.00

    9.00

    8.00

    7.00

    6.00

    5.00

    4.00

    3.00

    2.00

    1.00

    0.00

    DISCHARGE PRESSURE - PSIG

    FLOW

    -GPH

    AIR/GASPRESSURE-PSIG

    125

    35

    35PSIGAIR/ G

    AS

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM & 35 SPM @ 150 PSIG

    125PSIGAIR/GAS

    100PSIGAIR/GAS

    150PSIGAIR/G

    AS

    MAXFLOW(GPH)

    AIR/GA

    SPR

    ESSU

    RE

    0 500 1000 1500 2000 2500

    10.00

    9.00

    8.00

    7.00

    6.00

    5.00

    4.00

    3.00

    2.00

    1.00

    0.00

    160

    140

    120

    100

    80

    60

    40

    20

    150

    AIR/GASPRESSURE-PSIG

    35

    35

    PS I G

    AI R/G

    AS

    STROKE LENGTH: 100% PLUNGER SPEED: 45 SPM

    DISCHARGE PRESSURE - PSIG

    150PSIGAIR/GAS

    100PSIGAIR/G

    AS

    MAXFLOW(GPH)

    AIR/GA

    SPR

    ESSU

    RE

    CRP1000V400

    Discharge Pressure PSI 0 315 1000 1200 1750 2000gph 9.12 9.12 8.99 8.89 8.88 8.85Air Pressure PSI 35 35 66 78 112 130

    CRP750V400

    Discharge Pressure PSI 0 500 1500 2000 3000 3500gph 5.00 4.90 4.80 4.70 4.60 4.50Air Pressure PSI 35 35 60 80 115 130

    CRP1000V600

    Discharge Pressure PSI 0 600 2000 2500 4000 4500Gph 9.05 9.00 8.84 8.79 1 1Air Pressure PSI 35 35 60 75 120 140

    CRP1000V800

    Discharge Pressure PSI 0 1000 1950 5500 7300 8500gph 8.86 8.79 8.72 8.46 6.25 3Air Pressure PSI 35 35 35 88 120 140

    Performance Flow Curves (cont.)

    11339 M ROY LOTA 7/18/08 10:18 AM Page 10

  • 8/8/2019 30956_Williams_2-2008_lo_res

    11/32

    8

    1.2.2.3.2 Seal Material Selection

    The seal material must be chosen to satisfy both the chemical compatibility and the pressures/temperatures at which you are operating

    Below is a general guideline for seal material selection.

    1.2.2.3 Plunger & Seal Material

    1.2.2.3.1 Plunger Material Selection

    The materials available vary in hardness and chemical compatibility. We offer three materials based on our many years of industry experience

    with various chemicals. Hardness is a key property when selecting the proper plunger material. Our experience has shown that the harder

    plunger materials not only provide longer plunger life, they also provide greater seal life. A hard plunger is a must when pumping a chemical

    that is prone to crystallization or if the chemical is contaminated. Of course both of the preceding conditions will affect seal life. Below is atable that compares the chemical compatibility and hardness properties of each material.

    Tough material with excellent wear resistance. Excellentchemical inertness. Good for all types of chemicals, acids,bases or solvents. Recommended for use with the harderceramic plunger and higher pressures.

    Tough material with excellent wear resistance. Excellentchemical inertness. Good for all types of chemicals, acids,bases or solvents.

    Tough material with excellent wear resistance. Good for waterand alcohol based chemicals. Not recommended for solvents.

    Soft material with fair wear resistance. Broad chemicalcompatibility but its not to be used with ethyl or methylalcohols. Suggested only for hard to seal fluids in lowpressure applications when PE or TC will not seal.

    Soft material with fair wear resistance. Limited chemical com-patibility. Used mainly in Methanol pumping at low pressure.

    Soft material with fair wear resistance. Excellent chemical

    compatibility. Used when Viton

    is not compatible and PE or TCwill not seal.

    Material has very good abrasion resistance. Excellent chemi-cal resistance to phosphate esters, good to excellent to mildacids, alkalis, silicone oils and greases, ketones and alcohols.Not recommended for petroleum oils or di-esters.

    SEAL TEMP SUGGESTED

    MATERIAL TYPE RANGE PRESSURE RANGE COMMENTS

    TG Mechanical -30 to 180F 1,000 to 10,000 psiTeflon (Spring Loaded) -34 to 82C 207 to 690 barGraphite (High Pressure)

    TC Mechanical -30 to 180F 750 to 9,000 psiTeflon (Spring Loaded) -34 to 82C 52 to 620 bar

    Composite (Low Pressure)

    PE Mechanical -30 to 180F 100 to 3,000 psiUHMW (Spring Loaded) -34 to 82C 6.9 to 207 bar

    Polyethylene

    V O-ring -10 to 200F 100 to 750 psiViton -23 to 93C 6.9 to 52 bar

    BR O-ring -40 to 200F 100 to 750 psiBuna N -40 to 93C 6.9 to 52 bar

    K O-ring 32 to 200F 100 to 750 psi

    Kalrez

    0 to 93C 6.9 to 52 bar

    EPR O-ring -40 to 200F 100 to 750 psiEthylene -40 to 93C 6.9 to 52 bar

    Propylene

    Selecting the proper seal material for your application is important. We suggest using the harder plastic seals (PE, TC or TG) whenever possiblebecause they provide excellent wear life. The elastomers (V, BR, K or EPR) offer enhanced sealing at low pressure because they are soft and morecompliant than the plastics. However, the elastomers do not provide the same toughness or wear resistance.

    DESIGNATION MATERIAL HARDNESS CHEMICAL COMPATIBILITY

    CR CeramicBetween Sapphire and Diamond Excellent Chemical Inertness inon the Mohs Scale all Acids, Bases, Solvents

    A 17-4 ph 40 Rc

    General Corrosion-resistantStainless SteelLimited Acid Resistance

    B 316 SS 28 Rc

    Excellent Corrosion-resistantStainless SteelLimited Acid Resistance

    We recommend the use of ceramic because of its extreme hardness and excellent chemical inertness.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 11

  • 8/8/2019 30956_Williams_2-2008_lo_res

    12/32

    9

    Physical Specifications

    B

    H

    I

    CONTROLLER

    INLET 1/4" NPT (F)

    BLEEDER

    1/4" BARBED FITTING

    E

    D

    F

    C

    G

    A ACONTROLLER

    E

    F

    C

    B

    P03-6S P500V400

    RELAY INLET3/8" NPT (F)

    RELAY

    P1000V400P750V400

    P1000V600

    P04-6S P1000V800

    DI

    H

    J

    G

    BLEEDER

    1/4" BARBED FITTING

    CONTROLLER

    INLET 1/4" NPT (F)

    Model A B C D E F G H I J WT

    Inch/mm Inch/mm Inch/mm Diameter (IN) Connector Connector Inch/mm Inch/mm Inch/mm Inch/mm LBS/KG

    CP125V125 4.50/114.3 9.25/235 8.12/206.2 1 7/847mm 1/4 NPT (F) 1/4 NPT (M) 13/445mm 13/445mm 6 1/4159mm n/a 7.0/ 3.2

    CP250V225 6.00/152.4 11.68/296.7 11.00/279.4 2 1/263.5mm 1/4 NPT (F) 1/4 NPT (M) 2 9/1665mm 2 11/1668mm 8 7/8214mm n/a 9.0/ 4.1

    CP250V300 6.25/158.8 11.68/296.7 11.00/279.4 3 1/482.5mm 1/4 NPT (F) 1/4 NPT (M) 2 9/1665mm 2 11/1668mm 8 7/8214mm n/a 9.0/ 4.1

    CP500V225 5.50/139.7 12.00/304.8 11.00/279.4 2 1/263.5mm 1/4 NPT (F) 1/2 NPT (M) 2 5/867mm 2 13/1669mm 8 9/16217mm n/a 10.0/ 4.5

    CP500V300 6.00/152.4 12.00/304.8 11.00/279.4 3 1/482.5mm 1/4 NPT (F) 1/2 NPT (M) 2 5/867mm 2 13/1669mm 8 9/16217mm n/a 10.0/ 4.5

    CRP500V400 9.12/232 16.00/406 11.00/279.4 4 1/4108mm 1/4 NPT (F) 1/2 NPT (M) 2 5/867mm 2 13/1669mm 12 3/4324mm 9 7/16240mm 15.0/6.8

    CRP750V400 9.75/247.6 16.25/412.7 11.31/287.2 4 9/16116mm 1/2 NPT (F) 3/4 NPT (M) 35/892mm 376mm 13332mm 7 9/16240mm 16.7/7.5

    CRP1000V400 10.50/266.7 19.00/482.6 14.12/358.6 4 3/8111mm 1/2 NPT (F) 3/4 NPT (M) 4102mm 3 3/886mm 14 5/8365mm 8 7/8225mm 29.0/13.2

    CRP1000V600 12.50/317.5 19.00/482.6 14.12/358.6 6 3/8162mm 1/2 NPT (F) 3/4 NPT (M) 4102mm 3 3/886mm 17 3/4451mm 12 1/4214mm 35.5/16.1

    CRP1000V800 14.50/368.3 19.00/482.6 14.12/358.6 8 3/8213mm 1/2 NPT (F) 3/4 NPT (M) 4102mm 3 3/886mm 16406mm 11279mm 47.6/21.6

    Model Plunger PistonDiameter (In.) Diameter (In.)

    CP125V125 1/8 1 1/4

    CP250V225 1/4 2 1/4

    CP250V300 1/4 3

    CP500V225 1/2 2 1/4

    CP500V300 1/2 3

    CRP500V400 1/2 4

    CRP750V400 3/4 4

    CRP1000V400 1 4

    CRP1000V600 1 6

    CRP1000V800 1 8

    1.3 GENERAL OPERATING SEQUENCE

    1.3.1 Controller

    1.3.1.1 MK XII Controller

    The spool spring forces the spool upward

    to its highest position and unseats the top

    of the pilot plug from the upper seat. The

    exhaust spring forces the pilot plug

    upward and seats it on the lower seat. This

    blocks the air/gas exhaust port.

    When high pressure air/gas enters the

    supply port, it passes around and through

    the spool and past the open upper seat to

    the motor cylinder port.

    High pressure air/gas passes through the

    control passage in the controller, past the

    valve stem, and into the valve body upper

    chamber which causes pressure to build

    up in the chamber. Because the surface

    area of the upper U-cup diaphragm is

    much larger than that of the middle U-cup

    diaphragm, the downward force on the

    spool is greater than the upward force.

    This pressure pushes the spool down until

    the pilot plug seats itself on the upper

    valve seat, shutting off the air/gas supply.

    As the spool continues to move down, it

    pushes the pilot plug until the plug is

    unseated from the lower valve seat and

    allows the air/gas to exhaust through the

    lower valve from both the motor cylinder

    and the valve body volume chamber.

    When the pressure in the chamber is low

    enough, the spool spring starts pushing

    the spool upward. The exhaust spring

    pushes the pilot plug upward, and the

    controller returns to its initial position.

    1.3.2 Relays

    1.3.2.1 PO3-6S Relay

    When there is no pilot pulse (high pressure

    air or gas) entering the top of the relay

    from the controller, the piston spring

    pushes the piston and the poppet

    assembly upwards until the O-ring on top

    of the poppet presses against the upper

    body section, sealing the pump port from

    the exhaust port. The space below the

    poppet provides a path between the supply

    and pump ports.

    When the controller sends a pilot pulse,

    the high pressure gas on top of the piston

    overcomes the piston spring force, and

    pushes the piston and the poppet assembly

    downward until the O-ring on the bottom of

    the poppet presses against the lower body

    section, sealing the pump port from the

    supply pressure. The space above the

    poppet provides a path between the pump

    and exhaust ports.

    1.3.2.2 PO4-6S Relay

    When there is no pilot pulse (high pressure

    air or gas) entering the top of the relay

    from the controller, the piston spring

    pushes the piston and two poppets

    attached to the piston upwards until the O-

    ring on top of the upper poppet presses

    against the upper body section and at the

    same time the O-ring on the lower poppet

    presses against the middle body section.

    The upper poppet O-ring seals the number

    one pump port from the number oneexhaust port; the space below provides a

    path between the supply and the number

    one pump ports. The lower poppet O-ring

    seals the number two pump port from the

    supply port; the space below provides a

    path between the number two exhaust and

    pump ports.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 12

  • 8/8/2019 30956_Williams_2-2008_lo_res

    13/32

    10

    When the controller sends a pilot pulse, the

    high pressure gas on top of the relay piston

    overcomes the piston spring force and

    pushes the piston and two poppets

    assemblies downward. The piston and two

    poppets assemblies move down until the O-

    ring on the bottom of the upper poppetpresses against the middle body section

    and the O-ring on the bottom of the lower

    poppet presses against the lower body

    section. The upper poppet O-ring seals the

    number one pump port from the supply

    port; the space above provides a path

    between the number one pump and

    exhaust ports. The lower poppet O-ring

    seals the number two pump port from the

    number two exhaust port; the space below

    provides a path between the number two

    pump port and the supply port.

    1.3.3 Pump Motor (Air Chamber)

    The motor forces the piston plunger to move

    alternately into and out of the pump

    chamber. When the controller sends the

    supply air/gas into the motor chamber

    through the nipple connector, the pressure

    on the piston and diaphragm overcomes

    the combined force of the process fluid

    pressure on the piston plunger and plunger

    return spring, and pushes the plunger into

    the fluid chamber. When the external

    controller exhausts the air/gas, the piston

    plunger return spring and process fluidpressure push the piston plunger out of the

    fluid chamber.

    1.3.4 Pump (Fluid Chamber)

    The pump operating cycle consists of fluid

    being discharged and suctioned into the

    fluid chamber. During discharge, the piston

    plunger moves into the pump fluid

    chamber, decreasing the volume of the

    chamber and raising the pressure in thechamber fluid. This higher pressure closes

    the suction check valve and opens the

    discharge check valve, sending the fluid

    into the discharge line.

    During the suction part of the cycle, the

    piston plunger moves out of the fluid

    chamber, increasing the volume of the

    chamber and lowering the pressure of the

    chamber fluid. This lower pressure opens

    the suction check valve and a spring closes

    the discharge check valve, sending fluid

    from the suction line into the fluidchamber.

    SECTION 2.0: INSTALLATION OF

    PUMP AND CONTROLLER

    2.1 GENERAL

    Always install separate pressure regulators

    in the air/gas supply lines for the controller

    and the relay. Also, for the most efficient

    performance of your pump assembly, we

    recommend the following:

    A dryer and a dump valve in the air/gas

    supply line to remove any moisture from

    the supply air/gas.

    Isolation valves (ball type) on inlet and

    discharge lines of the pump and in the

    air/gas supply line to simplify

    maintenance.

    A check valve where the pump

    discharge line joins the main process

    line to prevent process fluid back flow. An inlet filter, with filtration

    approximately 25 microns, on pump

    suction line.

    A flow meter or a rate setting gauge in

    the suction line or process fluid

    discharge line, if you need precise flow

    rate adjustment or recording.

    2.2 PUMP ASSEMBLY

    Position the pump assembly with enough

    space around it to allow easy access to all

    components for maintenance. Install the

    assembly with the pump inlet/suction

    check valve pointing straight down. The

    pump will not work as efficiently in any

    other position since the inlet/suction check

    valve has no spring.

    NOTE: The pump assembly can be

    installed directly in the process line

    piping without any additional support

    brackets.

    SUPPLY BODY STROKES ELASTOMER SPOOLMODELS PRESSURE MATERIAL (SPM) STYLE

    MK XIIA 35-100 PSI (2.4-6.9 Bar) 316 ss 1 - 45 Neoprene Diaphragm

    MK X 35-100 PSI (2.4-6.9 Bar) 316 ss 1 - 45 Buna/TFE, Viton/TFE U-Cup

    CONTROLLER SPECIFICATIONS

    11339 M ROY LOTA 7/18/08 10:18 AM Page 13

  • 8/8/2019 30956_Williams_2-2008_lo_res

    14/32

    11

    2.4 SUPPLY RESERVOIR

    Position the supply reservoir so that the

    liquid level will not be less than six inches

    above the inlet check valve (flooded

    suction). While you can locate the reservoir

    at any height above the inlet check valve

    (net positive suction head), the limit is 100

    psig net positive suction head, which is the

    cracking pressure of the discharge check

    valve. We do not recommend using the

    pumps in a suction lift position since theywere not designed for such operation.

    2.5 RELIEF VALVE

    A safety relief valve is not necessary if the

    downstream piping can withstand the

    maximum pressure the pump can generate

    at the available air supply pressure. The

    maximum discharge pressure the pump

    can generate can be taken from the

    performance graphs in section 1.2.2.2.

    When this pressure is reached the pump

    will stop. Example: A CP250V225ATC

    operating at 35 psig is capable of

    generating 2520 psig. The pump will stop

    pumping when the 2520 psig is reached.

    SECTION 3.0: STARTUP, OPERATION,

    SHUTDOWN, AND STORAGE

    3.1 GENERAL

    While these procedures for startup,

    operation, shutdown, and storage are

    simple, following them carefully and

    correctly will improve the performance and

    increase the life of your pump assembly.

    CAUTION: To avoid damaging thecontroller valve stem, do not make a

    habit of turning the pump ON and OFF

    with the stroke rate control. Use the

    recommended ball valve in the air/gas

    supply line.

    3.2 STARTUP

    These startup procedures will produce a

    flow rate close to what you want. For a

    more precise flow rate, monitor with a flow

    meter while making the final adjustment.

    3.2.1 Air/Gas Supply

    Before starting up your pump assembly,

    make sure that the primary air/gas supply,

    compressor, tank of gas, or other source, is

    turned OFF. Also, set the pressure

    regulator(s) to ZERO pressure.

    3.2.1.1 Supply Pressure: The air/gas

    supply pressure must be large enough to

    produce a pump discharge pressure 200

    psi higher than the process pressure.

    Therefore, if the process pressure is 2800

    psig, the air/gas supply should provide a

    pump discharge pressure of 3000 psig. To

    set the supply pressure properly, use the

    performance graphs in 1.2.2.2.

    NOTE: With a controller/relay

    combination, the controller supply

    pressure should be at the minimum

    value, 35 psig (Ref. 1.2.1.1) and therelay supply pressure set per the above

    procedure. Remember, the controller is

    now supplying the relay and the relay is

    supplying the pump. Refer to section

    1.2.1.2 for the relay supply pressures.

    3.2.1.2 Supply Piping: Since supplying

    the proper air volume to the pump is

    critical for its operation, avoid long runs of

    small diameter tubing, as follows:

    Locate the main air/gas supply header

    as close to the pump as possible.

    Use no more than five feet of 1/4tubing to supply air/gas to the controller.

    Always locate the controller or relay on

    the pump.

    Use no more than ten feet of 1/2 tubing

    to supply air/gas to the relay.

    If a solenoid valve controls the pump,

    locate it no more than two feet away and

    use 1/2 tubing to supply air/gas to the

    pump.

    2.3 TYPICAL INSTALLATION

    Below you will find a schematic for a typical V plunger pump installation. The key ingredients for a successful installation are: 1:Clean, dry

    regulated air for the controller 2:A flooded inlet supply 3:An inlet filter 4:A rate setting gauge 5:A line check valve at the point of injection

    6:Isolation valves for maintenance on each component.

    Flow Tracking Controller Configuration Standard Pneumatic Controller Configuration

    3-WAY

    SOLENOID VALVE

    FLOWOR PHMETER

    PROCESS FLUID

    WPC 9001,APU,

    OR PLC

    24VDCPULSE

    4-20mASIGNAL

    CHECK VALVE

    FLOWMETER

    TOTALIZER

    PUMP

    DUMPVALVE

    AIR/GAS

    DRYERPRESSURE

    REGULATOR

    AIR/GASSUPPLY

    RATESETTING

    GAUGE

    INLET

    FILTER

    CHEMICALSUPPLY

    600 STROKESPER MINUTE

    RESETNOFLOWSELECT

    PUMP1

    PUMP

    RATE

    AUTO

    MANUAL

    CONTROL LOSSO F I/PMANUALADJUST

    0 1 00 %

    PUMP2

    T OT AL

    WPC9001

    12345678

    PUMP

    &

    CONTROLLER

    DUMP

    VALVE

    AIR/GASDRYER

    PRESSUREREGULATOR

    AIR/GASSUPPLY

    INLETFILTER

    CHEMICALSUPPLY

    PROCESS FLUID

    CHECK VALVE

    RATE

    SETTINGGAUGE

    FLOWMETER

    TOTALIZER

    11339 M ROY LOTA 7/18/08 10:18 AM Page 14

  • 8/8/2019 30956_Williams_2-2008_lo_res

    15/32

    SECTION 4.0: MAINTENANCE

    4.1 GENERAL

    This section contains procedures for

    disassembly and assembly of the

    controller, pump, and check valves, plus

    procedures for preventive and corrective

    maintenance. To maintain the reliability,

    durability, and performance of your pump

    assembly and related components, it is

    essential to follow these proceduresexactly and carefully.

    For consistent, reliable performance,

    replace any O-rings, U-cups, or other seals

    that you remove. Order replacement seal

    kits with detailed instructions from your

    distributor or Williams Instrument

    Incorporated.

    12

    3.2.2 Controller Stroke Rate

    The controller is preset at the factory to

    provide each pump the maximum number

    of strokes per minute at 35 psig (Ref.

    1.2.2.1 Performance Table). This setting

    falls at 100 on the controller scale. At

    ZERO on the scale, the pump will not cycle

    and there will be no output.

    To calibrate the controller stroke rate,

    follow this procedure:

    1. Rotate the stroke rate knob on the

    controller clockwise (CW) to ZERO on

    the stroke rate reference scale.

    2. Turn the main air/gas supply to the

    regulator(s) to ON, and adjust the

    regulator to the desired pressure. (see

    3.2.1.1 for determining pressure.)

    3. Set the flow rate for your application by

    using the controller's stroke rate knob

    in combination with the pump's stroke

    adjuster. (Ref. 1.2.2.1 Performance

    Table)

    If necessary, adjust the controller stroke

    rate knob as follows:

    1. Loosen the set screw and remove the

    knob.

    2. Adjust the valve stem to the desired

    rate by hand by turning the stem

    clockwise (CW) to decrease the stroke

    rate or counterclockwise (CCW) to

    increase the rate. Use a timer, such as

    a stop watch, to determine the actualstroke rate.

    3. Attach and set the knob at the desired

    position on the scale (100 on scales for

    45 spm, for example).

    3.2.3 Pump Stroke Length

    The adjuster scale for the pump's stroke

    length is factory set so that a ZERO reading

    equals ZERO stroke length. Calibrate the

    scale as follows:

    1. Turn the stroke adjuster tee set screw

    counterclockwise until the stroke

    adjuster knob is unlocked.2. Turn the stroke adjuster knob

    counterclockwise as far as it can go.

    The piston/plunger will be fully

    bottomed in the pump.

    3. Loosen the two screws holding the data

    plate; move the data plate until the

    middle of the pin is at ZERO on the

    scale. Tighten the two screws.

    4. Turn the stroke adjuster knob until the

    middle of the pin is at the desired stroke

    length on the scale.

    5. Tighten the tee knob clockwise until the

    stroke adjuster knob is locked.

    3.3 OPERATION

    3.3.1 Bleeder Plug

    Shortly after the pump assembly begins

    operating, the metered liquid should begin

    flowing through the pump. To bleed air

    trapped in the pump chamber, turn the

    bleeder plug CCW about a quarter turn.

    When the liquid is flowing steadily with

    each pump stroke from the end of the

    bleeder plug, turn it CW until the flow

    stops. It is best to close the bleeder plug

    when the pump is discharging and before

    the suction stroke.

    NOTE: To catch the escaping liquid, slip

    a length of 1/4 plastic or rubber tubing

    over the hose barb of the bleeder plug.

    3.3.2 Stroke Rate

    Set the operating stroke rate, as follows:

    1. Set the stroke rate knob to a mark on

    the scale that will produce a stroke rate

    close to the one you want. Keep in

    mind that the scale reading is only an

    approximate percent indication of the

    actual rate; generally the pump

    maximum stroke rate will be set at 100

    on the scale. NOTE: At the ZEROsetting on the controller stroke rate

    scale, the pump will not stroke, but

    as you rotate the knob toward 100,

    the rate will increase to the

    maximum strokes per minute for

    each pump. (Ref. 1.2.2.1 Performance

    Table) To set the stroke rate correctly,

    you must time the exhausts as they

    leave the bottom of the controller.

    2. Count the number of pump strokes

    during a one minute interval, using a

    timer such as a stop watch to

    determine the actual stroke rate.3. Adjust the knob to correct the stroke

    rate as needed. Confirm by timing the

    stroke rate.

    4. Repeat the above steps until you get

    the correct stroke rate. EXAMPLE: To

    get a stroke rate of 22 strokes per

    minute, set the knob to 50 which

    should produce approximately 25

    strokes per minute. Then reduce the

    rate by resetting the knob to 48. If this

    produces 21 strokes a minute, move

    the knob to 49, which should be very

    close to the 22 strokes per minute you

    want. Confirm the rate by timing it.

    3.4 SHUTDOWN AND STORAGE

    To shut down the pump assembly, set the

    pressure regulator(s) to ZERO, and turn the

    air/gas supply to OFF.

    To store the pump assembly or if it will not

    be used for a long time, do the following:

    1. Remove pump from the system.

    2. Flush out the pump chamber and check

    valves with water or solvent; drain and

    then blow the pump dry with

    compressed air.

    CAUTION: To prevent damage to the

    pump when you clean it, be sure to use

    a solvent compatible with the metered

    fluid that will not damage the pump

    seals. For a recommended solvent,

    contact your distributor or Williams

    Instrument Incorporated.

    3. Cap off the suction and discharge

    check valve ports.

    4. You may leave the pump, controller,

    and relay assembled, but make sure to

    store them in a dry, protected place.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 15

  • 8/8/2019 30956_Williams_2-2008_lo_res

    16/32

    13

    remove. Separate upper valve body

    from the lower section. (Fig. 2 & 3)

    3. Lift off the upper body and diaphragm.

    Lift out the inner diaphragm assembly.

    Set aside. (Fig. 4 & 5)

    4. Lift out the spool spring. (Fig. 6)

    5. Turn lower controller body upside down.

    Use a 3/16 hex wrench to unscrew

    bottom plug. Remove the bottom plug,lower spring and pilot plug. (Fig. 7 & 8)

    6. Return controller body as before and

    unscrew lower seat with a 3/16 hex

    wrench. Remove lower seat.

    (Fig. 9 & 10)

    Whenever you disconnect any air/gas or

    fluid piping, cover all open ports in the

    pump assembly to prevent dirt from

    entering.

    4.2 DISASSEMBLY AND ASSEMBLY

    4.2.1 Required Tools and Materials

    Necessary tools will vary by pump

    assembly model but the following are

    typical:

    Adjustable wrench: 12

    Belt-spanner or web wrench

    Open-end wrenches: various sizes

    Hex wrenches: 7/64, 1/8, 9/64,

    3/16, and 5/32, 1/4, 3/8

    Socket wrenches: various sizes with 2

    extension

    Flat-blade Screwdrivers: 1/8 (2

    required) and 1/4

    MK X Screwdriver in 3/16 hex Socket

    Drive (Drawing 1)

    Brass or plastic O-ring pick (1)

    Torque wrench (15 in-lb to 122 in-lb

    range)

    Bench vise

    Silicone grease, (Williams G321M4), or

    synthetic grease (Williams GS102149)

    Teflon tape 1/4

    Thread sealing compound

    NOTE: See 4.3 Preventive Maintenance

    for inspection and replacement of partsidentified throughout these procedures.

    4.2.2 Controllers

    4.2.2.1 Mark XII Controller

    Refer to the Mark XII Controller Parts List.

    To disassemble, do the following:

    1. Remove red cap. (Fig. 1)

    2. The Mark XII has (4) four socket head

    cap screws holding the controller

    together. Use a 5/32 hex wrench to

    Figure 2

    Figure 3

    Figure 4

    Figure 5

    Figure 1

    Fig. 6

    Fig. 7

    Fig. 8

    Fig. 9

    Fig. 10

    11339 M ROY LOTA 7/18/08 10:18 AM Page 16

  • 8/8/2019 30956_Williams_2-2008_lo_res

    17/32

    14

    7. To disassemble the inner diaphragm

    and spool assembly, first remove outer

    sleeve and mid ring by sliding past the

    diaphragms towards the upper seat.

    (Fig. 11, 12 & 13) Use a small screw

    driver or hex wrench placed through the

    inner spacer holes and with a 9/16

    wrench unscrew the upper seat and

    lower diaphragm. (Fig. 14) Place the top

    diaphragm stop (and inner sleeve with

    screw driver/hex wrench) into a soft jaw

    vice with a vee notch. Lightly clamp top

    reinstall the lower seat and spool

    spring. Make sure the capillary holes inthe upper diaphragm, the mid ring and

    the mid diaphragm are in line with the

    capillary hole of the lower body. Use a

    small awl or hex wrench to thread

    together. Install (1) one of the (4) body

    screws from the under side through the

    loose parts and through the top

    diaphragm. Now remove the awl and

    place on top of the assembly the upper

    controller body. Insure its capillary hole

    is inline with the others. Loosely thread

    together the (1) one body screw. Installremaining (3) three screws and torque

    all to 28 - 32 inch pounds.

    diaphragm stop. Unscrew inner sleeve.

    (Fig. 15) Remove the mid diaphragm

    from the top diaphragm stop. (Fig. 16)

    Remove the lower diaphragm from the

    upper seat. (Fig. 18)

    8. Clean all metal parts. Inspect and or

    replace all three diaphragms. To

    reassemble, push the mid diaphragm

    onto the top diaphragm stop. Push the

    lower diaphragm onto the upper seat.

    Thread the inner spacer onto these (2)

    two diaphragm assemblies. With a

    screw driver and 9/10 wrench, tighten

    securely, but notenough to pucker the

    diaphragms. Install the mid ring,

    counter bore first, onto the diaphragm

    assembly past the lower diaphragm and

    then past the mid diaphragm. Some

    maneuvering of the diaphragms will be

    needed. Install the outer sleeve by

    sliding past the lower diaphragm. Insure

    the narrow seat on the sleeve goes

    against the mid diaphragm and the

    wider seat is against the lower

    diaphragm. Some maneuvering of the

    lower diaphragm will also be needed.

    The inner assembly is now complete.

    (Fig. 17 & 19)

    9. To install the inner assembly into the

    lower controller body, be sure to

    Figure 11

    Figure 12

    Figure 14

    Figure 13

    Figure 15

    Figure 16

    Figure 17

    Figure 18

    Figure 19

    11339 M ROY LOTA 7/18/08 10:18 AM Page 17

  • 8/8/2019 30956_Williams_2-2008_lo_res

    18/32

    15

    4.2.3 Relays: PO3-6S and PO4-6S

    Refer to the appropriate Parts List.

    Disassembly instructions 1 through 9 apply

    to all relays, instructions 10 and 11 apply

    only to the PO3-6S, and instructions 12

    through 17 apply to the PO4-6S.

    1. Use a 9/64 hex wrench to remove the

    top cap screws (Fig. 20 & 21). Turn

    screws CCW.

    Figure 20

    Figure 21

    2. Separate the top cap with data plate

    from the upper body section. (Fig. 22)

    Figure 22

    3. Put a 7/64 hex wrench through the

    side of the relay into the upper body

    exhaust port and into the hole in the

    upper poppet stem. (Fig. 23)

    Figure 23

    4. Use a 9/64 hex wrench to remove the

    piston lock screw (Fig. 23)

    5. Pull out the piston, O-ring, and spring.

    (Fig. 24 & 25)

    Figure 29

    9. Use a flat-blade screwdriver to unscrew

    the lower poppet stem. (Fig. 29)

    10. PO3-6S: Pull out the upper poppet and

    poppet body; the poppet stem

    connector bolt will remain in one of

    the stems.(Fig. 30)

    Figure 30

    11. PO3-6S: Put a screwdriver in the holes

    of the poppet O-ring retainers and pry

    them off. (Fig. 31 & 32)

    Figure 31

    Figure 32

    NOTE: You can now reassemble thePO3-6S relay by coating the poppet

    stems, seals, and mating surfaces with

    silicone grease, and then reversing the

    above procedure.

    Figure 24

    Figure 25

    6. Use a 5/32 hex wrench to remove the

    bottom cap screws. (Fig. 26 & 27)

    Figure 26

    Figure 27

    7. Separate the lower body section from

    the upper body section. (Fig. 26 & 27)

    Figure 28

    8. Again, put the 7/64 hex wrench

    through the hole in the upper poppet

    stem. (Fig. 29)

    11339 M ROY LOTA 7/18/08 11:35 AM Page 18

  • 8/8/2019 30956_Williams_2-2008_lo_res

    19/32

    16

    To continue disassembling the PO4-6S, do

    the following:

    12. PO4-6S: At this point, repeat step 8,

    however, if the lower and middle

    poppet stems should separate, modify

    the rest of this procedure accordingly.

    (Fig. 33 & 34)

    Figure 33

    Figure 34

    13. PO4-6S: Separate the upper poppet

    stem from the middle stem and the

    poppet body. The poppet stem

    connector bolt will remain on one of

    the stems. (Fig. 35)

    Figure 35

    14. PO4-6S: Use the flat-blade

    screwdriver to unscrew the lower

    poppet stem.

    15. PO4-6S: Pull out the middle stem and

    poppet body. The poppet stem

    connector bolt will remain in one of

    the stems. (Fig. 35)16. PO4-6S: Put a 7/64 hex wrench or

    screwdriver in the holes of the poppet

    O-ring retainers and pry them off.

    (Fig. 31) Then use the screwdriver to

    pry out the O-rings.

    Figure 40

    b. Remove the motor cylinder by unfastening

    the 4 to 10 screws (depending on the

    pump model) holding the motor cylinder

    to the faceplate. (Fig. 41)

    Figure 41

    c. Remove the piston plunger guide ring

    and return spring. (Fig. 42 & 43)

    Figure 42

    Figure 43

    d. Now remove the retaining ring and

    spring seat. The secondary seal (and

    associated backup ring if O-rings are

    being used), will be found beneath the

    spring seat. (Fig. 44 & 45)

    Figure 36

    NOTE: Reassemble the PO4-6S relay by

    coating the poppet stems, seals, and

    mating surfaces with silicone grease,

    and then reversing the above

    procedure.

    4.2.4 P125, P250, P500, P750 & P1000

    Pumps

    4.2.4.1 Changing Plunger Seals

    1. Disassembly:

    a. Unscrew the pump body from the fluid

    cylinder. The primary seal is visible in

    the pump body. The plunger will

    protrude from the pump body, through

    the seal. (Fig. 37, 38, 39 & 40)

    Figure 37

    Figure 38

    Figure 39

    11339 M ROY LOTA 7/18/08 10:18 AM Page 19

  • 8/8/2019 30956_Williams_2-2008_lo_res

    20/32

    17

    Such conditions could lead to leakage or

    excessive seal wear if not corrected.

    Replace parts that exhibit wear.

    g. Install the new seals in accordance with

    the illustrations shown. Use illustration I

    if your are installing the spring loaded U

    cup type seals, and illustration II if you

    are installing O-ring type seals. In each

    case the seals and back up rings can be

    easily be pushed in place using only

    your fingers. Make certain the order and

    position in which the components are

    installed matches the illustration shown.

    Also make certain the seal assemblies

    are installed completely, none of the

    seal assembly should protrude. NOTE:

    APPLY A LITTLE GREASE FROM

    GS102149 GREASE TUBE TO THE

    SEALS. BACK UP RINGS AND ALL

    THREADS BEFORE INSTALLING.(Figs. 52, 53, 54, 55, 56 & 57)

    Figure 52

    Figure 53

    Figure 54

    Figure 47

    Figure 48

    Figure 49

    Figure 50

    e. Remove both 1/8 NPT plugs in the

    pump body so that all the old grease

    and any dirt or debris can be

    cleaned/flushed from the pump body.

    (Fig. 51)

    Figure 51

    f. Thoroughly clean all internal pump

    surfaces. Then inspect all metal sealing

    surfaces such the seal gland and

    plunger for scratches, nicks or

    irregularities.

    Figure 44

    Figure 45The primary and secondary seals can

    be removed from the pump body

    assembly using either the piston-

    plunger or a brass (or plastic) pick. Be

    careful not to scratch the metal

    sealing surfaces during the extraction

    process. The piston-plunger can be

    inserted alternately into each end of

    the pump body assembly to push out

    the opposing seal (and backup ring if

    present). The brass or plastic pick can

    be used to hook the seal (and backup

    ring) to pull them out. Caution mustbe used to make certain that the

    sealing surfaces are not scratched

    during the removal of the seal and

    back up ring. (Fig.46, 47, 48, 49 & 50)

    Figure 46

    11339 M ROY LOTA 7/18/08 10:18 AM Page 20

  • 8/8/2019 30956_Williams_2-2008_lo_res

    21/32

    18

    2. REASSEMBLY

    a. Replace spring seat and retaining ring.

    NOTE: The spring seat may be used

    as the secondary seal installation

    tool. Place the seal face down in the

    cavity and push it into place with spring

    seat. (Fig. 58 & 59)

    Figure 58

    Figure 59

    b. Thread the fluid cylinder to the pump

    body. The parts are designed to be

    tightened using a strap type wrench

    with the following torque values:

    Figure 55

    Figure 56

    Figure 57

    PUMP MODEL TORQUE (Ft-lbs)

    P125V (all) 50

    P250V (all) 50

    P500V 225 50

    P500V 300 100

    P500V 400 100

    P750V 400 100

    P1000V (all) 100

    c. Now apply Williams GS102149 synthetic

    grease to the plunger, and the piston

    U-cup.

    d. Install the spring in the spring seat.

    Then simultaneously insert and rotate

    the plunger through the spring into the

    pump body seals. It may be necessary

    to open the bleeder plug during this

    procedure to vent any fluid in fluid

    cylinder.

    e. Now, apply synthetic grease (GS102149)

    to the internal surface of the motor

    cylinder. Wrap the piston guide ring

    around the piston, and while holding it

    in place, slide the motor cylinder over

    the piston, the U-cup and guide ring.

    Align the holes in the motor cylinder

    with the face plate and tighten the 4 to

    10 screws removed during the

    disassembly process. Assembly screw

    torque is 25 in-lbs.

    Retaining Ring

    Spring Seat(Seal Retainer)

    Secondary Seal (Upper)

    Backup Ring (Spacer)

    O-ring Seal

    Retaining Ring Groove

    Faceplate

    Pump Body

    Grease Access Ports

    Body O-ring Seal

    Backup Ring(Plunger Guide)

    Primary Seal (Lower)

    O-ring Seal

    Fluid Cylinder

    Illustration I Illustration II

    11339 M ROY LOTA 7/18/08 10:18 AM Page 21

  • 8/8/2019 30956_Williams_2-2008_lo_res

    22/32

    19

    f. Lubricate the seals as follows:

    1. Both 1/8 NPT plugs should have

    been removed during the disassembly

    of the pump to facilitate cleaning. If

    this was not done, please do it now.

    2. Thread the tube of grease (GS102149)

    that came with the seal kit into one of

    the lubrication access ports. (Fig. 60)

    Figure 60

    3. Now, squeeze the grease into the

    lubrication cavity. Hold your finger over

    the opposite lubrication cavity accesshole, alternately removing it to allow

    air to escape and to feel the grease

    pressurize the lube cavity. (Fig. 61)

    Figure 61

    4. When all the air have been expelledfrom the lubrication cavity and only

    grease is exiting the opposite access

    hole, the cavity is full.

    5. Apply Swagelok SWAK thread sealant

    to both 1/8 NPT plugs and replace in

    pump housing. (Fig. 62)

    Figure 62

    4.2.4.2 Complete Pump Rebuild:

    Although there are some differences

    between the pumps, they are

    disassembled and reassemble in

    approximately the same way. The

    procedures describe the differences.

    Refer to the appropriate parts list, for

    the pump you are rebuilding.(Fig.62)

    Figure 63

    Disassemble the pumps as follows:

    1. Clamp the pump at the fluid cylinder in

    a vice. Use jaw protectors to avoid

    scratching the pump. (Fig. 64)

    Figure 64

    2. Use the belt-spanner wrench to loosen

    the body from fluid cylinder. Clamp on

    the body or the motor cylinder. Continue

    turning the assembly by hand until the

    two parts separate. (Fig. 65)

    Figure 65

    NOTE: The fluid cylinder of the pump

    can remain installed on the plumbing if

    only the pump and not the check valves

    are to be serviced. Also, since the body

    and the fluid cylinder are made as a

    matched set, keep them together. do

    not substitute parts from other pumps

    of the same model.

    3. Use the appropriate hex wrench to

    remove the socket-head cap screws

    and washers fastening the motor

    cylinder to the body. If the pump has a

    piston return spring it will push the two

    pieces apart (exception:CRP1000V800);

    if not, separate the motor cylinder from

    the body and faceplate.

    4. Remove the piston/plunger assembly,

    piston U-cup (piston O-ring on the

    P1000V800), piston guide ring and the

    piston return spring (on all pumps

    except the P1000V800). The piston

    guide ring will fall away from the

    piston. Remove the piston U-cup

    (piston O-ring on the P1000V800).

    (Fig.66 & 67)

    Figure 66

    Figure 67

    NOTE: Before doing the next step, put a

    piece of tape or other mark on the

    outside surface of the cylinder

    faceplate to align it with a body part for

    reassembly.

    5. Use the 5/16 open end wrench tounscrew and remove the vent plug, and

    the 3/16 hex wrench to unscrew and

    remove the two grease chamber plugs.

    Use the appropriate hex wrench to

    unscrew and remove the cap screws

    and split washers fastening the cylinder

    faceplate to the body. Slide the

    faceplate off the body. On the

    P1000V800, remove the two faceplate

    O-rings. (Fig. 68, 69, 70 & 71)

    Figure 68

    11339 M ROY LOTA 7/18/08 10:18 AM Page 22

  • 8/8/2019 30956_Williams_2-2008_lo_res

    23/32

    20

    Figure 69

    Figure 70

    Figure 71

    6. On the P250 and P500, use a small

    screwdriver or punch to push the filter

    plugs from the cylinder faceplate. (Fig.

    72, 73 & 74)

    Figure 72

    Figure 73

    Figure 74

    CAUTION: Do not remove the face plate

    plug from the P1000V800. The filter plug

    retainers are permanently pressed into

    place at the factory. DO NOT REMOVE

    THEM.

    7. Use a pick if necessary to remove the

    body seal O-ring, the primary plunger

    seal, and the plunger seal backup ring

    from the body. The piston-plunger can

    also be used to push out the primary

    seal and backup ring by reinstalling it in

    the body faceplate assembly.(Fig. 75,

    76 & 77)

    Figure 75

    Figure 76

    Figure 77

    8. Clamp the body in the bench vise,

    faceplate side up.

    9. Use a pick or small screwdriver to lift

    the end of the spring seat retainer

    from the groove in the body. While

    holding the end free, slide the other

    pick between the retainer and the body

    until the retainer is loose. (Fig.78 & 79)

    Figure 78

    Figure 79

    10. Remove the spring seat. On the

    P1000V800 remove the two spring

    seat O-rings. (Fig. 80 & 81)

    Figure 80

    Figure 8111. Use a pick if necessary to remove the

    secondary plunger seal located

    beneath the spring seat or remove the

    seal by installing the plunger in the

    opposite end of the body-faceplate

    assembly and pushing the seal out.

    (Fig. 82 & 83)

    11339 M ROY LOTA 7/18/08 10:18 AM Page 23

  • 8/8/2019 30956_Williams_2-2008_lo_res

    24/32

    21

    Figure 82

    Figure 83

    12. Use the appropriate socket wrench

    (3/4 or 7/8) to unscrew the nut that

    secures the piston stop and remove

    the stroke adjuster assembly from the

    motor cylinder. (Fig. 84, 85 & 86)

    Figure 84

    Figure 85

    Figure 86

    13. Clamp the fluid cylinder in the vise.

    Use the adjustable wrench to unscrew

    and remove the inlet check valve,

    discharge check valve, and bleeder

    plug. Before reassembling the pump,

    clean all chambers, motor cylinder,

    and cylinder faceplate with an

    approved solvent. Contact your

    distributor or Williams Instrument

    Incorporated for a recommended

    solvent. Also, lubricate all O-rings and

    U-cups with silicone grease or

    synthetic grease. (Fig. 87, 88, 89, 90

    & 91)

    Figure 87

    Figure 88

    Figure 89

    Figure 90

    Figure 91

    Reassemble the pump as follows:

    1. On all pumps except the P125 and

    P1000V800, install the filter plugs in

    the body.

    2. On all pumps except the P125, slide

    the cylinder faceplate over the body

    and line up the mounting holes. Make

    sure that the marked side of the

    faceplate aligns with the outside of the

    pump. On all pumps except the P125and P1000V800 also line up the vent

    holes in the faceplate with the filter

    plugs.

    3. Install the screws with split

    lockwashers and torque them to the

    values in the table.

    PUMP TORQUE (in-lb)

    P125 15 - 20

    P250 15 - 20

    P500 15 - 20

    P750 15 - 20

    P1000 28 - 32

    4. Clamp the body in the bench vise with

    the faceplate end up.

    5. For the P250, P500, P750, and

    P1000V600, install a new plunger seal

    in the upper area of the body by facing

    the open side of the seal down toward

    the grease chamber and using the

    spring seat to push the seal in. For the

    P125, push the seal in by hand or use a

    special tool made for this purpose

    (part no. WT201192).

    6. On the P1000V800, put the two spring

    seat O-rings on the spring seat and put

    the spring seat in the body. On all

    pumps except the P1000V800, align

    the hole in the side of the spring seat

    with the vent plug and use the 5/16

    open end wrench to install the plug.

    7. Separate one end of the retaining ring

    and put it in the groove inside the body.

    Use a finger to work the rest of the ring

    into the groove until the other end is in

    place.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 24

  • 8/8/2019 30956_Williams_2-2008_lo_res

    25/32

    22

    8. Remove the body from the bench vise.

    At the threaded end of the body install

    the body seal O-ring and the plunger

    seal backup ring followed by the

    plunger seal. Face the open side of the

    seal away from the grease chamber.

    9. Place the fluid cylinder in the vise with

    the body mating end facing up. Use the

    adjustable wrench to screw in the

    suction check valve (ball end or hex

    plug end into the pump), discharge

    check valve, and bleeder plug. Arrows

    on the check valves indicate flow

    direction. Use thread sealant and teflon

    tape on the suction and discharge

    check valves.

    10. Apply synthetic grease to the body

    threads and body seal O-ring. Screw

    the body into the fluid cylinder until it is

    hand tight, then tighten with the web or

    belt-spanner wrench per the following

    torque values.

    PUMP MODEL TORQUE (Ft-lbs)

    P125V (all) 50

    P250V (all) 50

    P500V 225 50

    P500V 300 100

    P500V 400 100

    P750V 400 100

    P1000V (all) 100

    11. On all pumps except the P1000V800,

    put one end of the piston return spring

    in the spring seat. Install the piston U-

    cup (piston O-ring on the P1000V800)

    on the piston. Face the open side of

    the U-cup away from the plunger.

    12. Apply synthetic grease liberally to the

    piston/plunger assembly and inside

    the motor cylinder; install the

    piston/plunger assembly, without the

    piston guide ring, into the body.

    13. Put the piston guide ring on the piston;

    while holding the ends of the guide

    ring together with a finger or thumb,

    slide the motor cylinder over the

    piston.14. Push the motor cylinder down to the

    cylinder faceplate; align the motor

    cylinder mounting holes.

    15. Install all the screws with split

    lockwashers.

    16. Use the 9/64 hex wrench on the

    P125, P250, P500, and P750 and the

    1/4 hex wrench on the P1000, to

    torque the socket-head cap screws

    and split lock washers to the values in

    the table.

    PUMP TORQUE (in-lb)

    P125 15 - 20

    P250 15 - 20

    P500 15 - 20P750 15 - 20

    P1000 118 - 122

    17. To lubricate the seals for running, hold

    a finger over one vent hole and insert

    the end of the synthetic grease tube

    into the other vent hole. Squeeze

    synthetic grease into the grease

    chamber until you feel it pressing

    against your finger. Vent all the air.

    Wipe off excess synthetic grease,

    apply thread sealing compound to the

    vent plugs, and replace them.

    4.2.5 Discharge Check Valves:

    All Pumps

    Although there are several sizes of

    discharge check valves, they are all

    disassembled and reassembled the same

    way. Refer to the appropriate parts list.

    Disassemble the check valve as follows:

    1. Clamp the check valve body in the vise.

    2. Use the appropriate hex wrench to

    unscrew and remove the retainer.

    3. Remove the body from the vise and

    dump out the spring, ball seat, ball,sleeve, and Teflon O-ring (Fig. 92)

    Figure 92

    4. Remove the Teflon O-ring from the

    sleeve.

    5. Inspect all the parts and replace them ifthey are worn.

    Reassemble the check valve as follows:

    1. Put the Teflon O-ring in the sleeve and

    drop the sleeve, O-ring first, into the

    body.

    2. Drop the ball into the body.

    3. Put the small end of the spring in the

    spring cavity on the wide end (not the

    slotted end) of the ball seat. Drop the

    two parts, ball seat first, into the body.

    4. Drop the retainer, spring cavity first, if it

    has one, into the body and use the

    appropriate hex wrench to torque to the

    values in the table.

    PUMP TORQUE (in-lb)

    P125 230 - 240

    P250/500 230 - 240

    P750/P1000 118 - 122

    4.2.6 Suction Check Valves: All Pumps

    Although there are several sizes of check

    valves with minor construction differences,

    they are all disassembled and reassembled

    the same way. The procedures describe

    the differences. Refer to the appropriate

    parts list. Disassemble the check valve as

    follows:

    1. Clamp the check valve body in the vise.

    2. Use the appropriate hex wrench to

    unscrew and remove the retainer.

    (Fig. 93)

    Figure 93

    3. Remove the body from the vise and

    dump out the ball, sleeve, and O-ring.

    NOTE: On the P125 and P250 you

    may need a pick or small tool to

    remove the Teflon O-ring inside the

    body. (Fig. 94)

    Figure 94

    11339 M ROY LOTA 7/18/08 10:18 AM Page 25

  • 8/8/2019 30956_Williams_2-2008_lo_res

    26/32

    23

    4. On the P500, P750 and P1000, remove

    the Teflon O-ring from the sleeve.

    (Fig. 95)

    Figure 95

    Reassemble the check valve as follows:

    1. On the P500, P750 and P1000, put the

    Teflon O-ring in the sleeve and drop

    the sleeve, O-ring first, into the body.

    On the P125 and P250, drop the O-ring

    into the body, followed by the sleeve,

    the end with a shoulder first.2. Drop the ball into the body.

    3. On the P125 and P250, use the

    appropriate hex wrench to screw the

    retainer into the body. On the P125 and

    P250, tighten until you feel the Teflon

    O-ring resistance, then tighten another

    1/4 to 1/3 turn to compress the O-ring.

    On the other pumps, tighten securely.

    No Loctite #277 is required.

    4.3 PREVENTIVE MAINTENANCE

    4.3.1 Periodic Maintenance

    4.3.1.1 Once a week: Perform the

    following procedures:

    1. Check for process fluid leaking from the

    plumbing.

    2. Check for process fluid leaking from the

    pump's vent hole, cylinder faceplate,

    and body O-ring.

    3. Check for air/gas leaks.

    4. Check for loose fittings and screws.

    4.3.1.2 At least once a month: Unscrew

    the plug from the top of the tee fitting

    between the relay and the controller; put a

    few drops of Williams SF96-100 silicone

    oil in the hole.

    4.3.1.3 At least every six months:

    Inspect the piston-plunger assembly and

    seals. Replace the seals and check the

    plunger for wear; replace the piston/

    plunger assembly if it is scored, rough, or

    discolored.

    4.3.1.4 At least every twelve months:

    Perform the following procedures:

    1. Disassemble and inspect the pump inlet

    and outlet check valves. Replace worn

    parts.

    2. Inspect the piston/plunger assembly,

    piston return spring (on all pumps

    except the P1000V800), and all seals.

    Replace the piston/plunger assembly

    and the seals.

    4.3.2 Cleaning and Lubrication

    Whenever the pump assembly is

    disassembled: Clean all inside and

    outside surfaces with an approved solvent,

    and blow them dry with compressed air.

    CAUTION: To prevent damage to the

    pump when you clean it, use a solvent

    that is compatible with the process

    fluid and that will not damage pump

    seals. Contact your distributor or

    Williams Instrument Incorporated for a

    recommended solvent.

    4.4 TROUBLESHOOTING

    4.4.1 Proper Pump Use

    When a pump is either not working or

    working incorrectly, the trouble can be in

    two basic areas: the pneumatic or the fluid

    ends of the pump. However, since factors

    other than the pump can affect its

    operation, first check that the pump isbeing used properly. To help you

    determine this, use the following checklist:

    1. Is the air/gas supply available in

    sufficient volume and at the proper

    pressure?

    2. Is the air/gas supply of clean

    instrument quality, not dirty or wet?

    3. Is a pressure regulator in use to

    maintain a constant air supply?

    4. Is the tube or pipe size of the air/gas

    supply line correct for your pump

    model?

    5. Is the pump correct for the nature and

    characteristics of the material(s) it

    handles: composition, viscosity,

    necessary line pressure, etc.?

    6. Is the process fluid containerdrum, day

    tank or large storage tankclean and

    free of contaminants?

    7. Is the filtration adequate? Disassemble

    and inspect.

    8. Is the size of the process fluid line

    correct for the pump?

    9. Is the distance between the pump and

    supply air/gas correct?

    10. Is the pump operating within

    acceptable minimum and maximum

    temperature limits?

    11. Is the pump being used for more than

    one purpose? Plunger seals are

    affected by this.

    12. Is the proper cleaning fluid being used

    to flush out the pump?

    13. Is there a current and accurate

    service/maintenance/breakdown

    record for the pump?

    4.4.2 Proper Amount of Use

    While answers to the questions on the

    above checklist will provide considerable

    information about how the pump is being

    used, it is equally important to determine if

    it is overworked.

    Fortunately, you can use the amplification

    ratio of the pump (ref. 1.1.3 Pumps) and

    the process pressure (the pressure the

    pump plunger is working against) to check

    this. Use the following example:

    Example: process pressure is 2800 psi.

    Add 200 psi to the example process

    pressure. This additional 200 psi will

    insure that the chemical is positively

    injected: 2800 psi + 200 psi = 3000

    psi.

    To set the air/gas supply pressure use

    the performance graphs in 1.2.2.1.

    From this information you can determine if

    the pump is working properly. In the above

    example, if the supply pressure had been

    100 psi instead of 35, it would have been

    excessive, resulting in premature failure of

    the pump's moving parts and sealing

    capabilities.

    4.4.3 Troubleshooting Guide

    The Troubleshooting Guide on the

    following pages identifies the most

    common problems, their possible causes,and corrective action for each problem.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 26

  • 8/8/2019 30956_Williams_2-2008_lo_res

    27/32

    24

    TROUBLESHOOTING GUIDE

    PROBLEM POSSIBLE CAUSE(S) CORRECTIVE ACTION

    CONTROLLER NOT Foreign material in controller Put finger over exhaust port; alternately seal & vent port toOSCILLATING clear exhaust valve.

    No air/gas supply Connect pressure gauge to port opposite supply line; verifyrequired supply pressure.

    Supply pressure too high or too low Reset regulator to proper pressure.

    Too much pressure drop in a ir/gas l ine Increase connecting tube size or clean air l ines.

    Stroke rate valve open too much Disconnect air/gas supply. Rotate stroke rate knob CCW topeg, wait 5 seconds, & rotate knob CW until it stops.DO NOT FORCE KNOB.

    Reconnect supply & rotate knob CCW until oscillations start.

    Adjust stroke rate to proper strokes per minute according toSpecification Sheet; loosen knob set screw, rotate knob CCWto peg, & tighten screw.

    Leak between valve body & Loosen; then retighten the connection between valve &

    controller body controller body. If dirty, disassemble, wipe clean andreassemble.

    Continuous air flow from controller Inspect & replace damaged lower seat and pilot plug.exhaust port (Pilot plug not seatingproperly)

    Air flowing from equal izer hole on Inspect & replace ruptured or improperly seated seals.side of lower control body

    Put finger over exhaust port; alternately seal & vent port toclear exhaust valve.

    Broken pilot plug, exhaust spring, Replace damaged parts.or spool return spring

    Excessive water in controller Install an air/gas dryer or separator in supply line.

    RELAY NOT OPERATING Broken piston return spring Replace return spring. See page 15

    Poppet stem loosened from connector bolt Tighten poppet stem to connector bolt. See page 15

    Air blow-by caused by poorly sealed O-rings Improve quality of air supply and clean dirt from unit.Replace O-rings if damaged.

    Inadequate air supply Check air regulator for proper pressure.

    PLUNGER NOT STROKING Controller control knob set at ZERO Turn knob to proper setting on dial.

    Air/gas supply turned OFF Open valve to allow air supply to flow to controller.

    Broken motor return spring Replace return spring.

    Plunger stuck due to tight or dry seal If seal is swollen, check its chemical compat ibi li ty withprocess fluid; replace with compatible seal material.

    If seal is dry, lubricate & fill reservoir with grease.

    Plunger bottomed Readjust plunger stroke length. Replace return spring ifbroken.

    Excessive grease between cylinder and Remove excess grease. Piston seal may be leaking; checkfaceplate and replace if necessary.

    Air/gas supply pressure too low to Increase supply pressure to controller or relay.overcome process line pressure

    Discharge or suction line plugged Clean the lines.

    11339 M ROY LOTA 7/18/08 10:18 AM Page 27

  • 8/8/2019 30956_Williams_2-2008_lo_res

    28/32

    PLUNGER NOT STROKING Air/gas flow to controller too low (controller Install a larger capacity regulator or supply line. Vent supply(Cont.) locked up and will not cycle) side of controller and try to start pump at slowest speed;

    increase speed slowly if controller starts to cycle.

    Motor cylinder-air piston blow-by Check piston seal; replace as needed.

    Check motor cylinder surface for damage from dirt or sand;install clean filters on bottom of cylinder faceplate. Replacecylinder if necessary.

    LOW PUMP OUTPUT Viscosity of the chemical being pumped Review and enlarge size of supply and discharge lines totoo high. improve flow of chemical.

    Pump mounted too high to suck adequate Remount pump to create a flooded suction (six inchsupply of chemical to fluid cylinder minimum).

    Pump appears sluggish in stroking. Piston Remount pump as close to controller and relay as possiblenot returning all the way. to allow the controller and relay to exhaust quickly. Check

    for ice in exhaust port.

    Suction lift condition inadequate. Change tank elevation to get flooded suction if change notpossible, add foot valve at end of suction line, and increase

    suction line diameter.

    Blocked suction filter Clean or replace filter element.

    Supply and discharge lines too small. Install correct tubing size.See pump sluggish

    Erratic controller operation Rebuild, clean and lubricate controller; add air inlet filter orair/gas dryer.

    Check valves leaking or contaminated. Rebuild, replace damaged parts.Loss of pump capacity

    Improper chemical supply Make sure top of chemical supply tank is ventedto atmosphere or pressurized.

    PROCESS FLUID IN Premature wear on plunger seals from Calculate proper speed and air supply pressure. (Refer toGREASE CHAMBER OR excessive pump speed amplification ratio principle.) Replace seals and plunger.

    LEAKING FROM BODYOR CYLINDER Foreign material in process fluid Check to see if chemical supply is clean; if not, installFACEPLATE VENT HOLE chemical filter in supply line.

    Seals incorrectly assembled or damaged Refer to instructions for install ing seals.during installation

    Plunger nicked, burred or scratched Replace plunger and seals.

    PROCESS FLUID IN Seal or plunger materials not compatible Refer to compatibility charts; contact distributor orGREASE CHAMBER OR with process fluid Williams Instrument IncorporatedLEAKING FROM BODYOR CYLINDER Crystallized chemical on plunger Maintain lubricant and decrease time between inspections.FACEPLATE VENT HOLE scoring seal(cont.)

    Lubricant incompatible with process fluid Change lubricant; contact distributor or WilliamsInstrument Incorporated

    NO PUMP DISCHARGE Suction check valve or discharge check Clean or replace check valves.valve not seating

    Suction or discharge line clogged Inspect line for closed connections or valves.

    Air entering suction line Tighten fittings; Inspect and replace sealants.

    Pump vapor locked Open bleeder plug and prime pump.

    PROBLEM POSSIBLE CAUSE(S) CORRECTIVE ACTION

    25

    11339 M ROY LOTA 7/18/08 10:18 AM Page 28

  • 8/8/2019 30956_Williams_2-2008_lo_res