Service ManualC-4201
Asepsis Design Cuspidor, Vacuum Accessory Group and Water QD.
(used with 145-002 side box only)
Rev 5 - 1-14-20 900-085M
800.237.2303 ◆ 503.538.8756 ◆ 503.538.2845 Fax ◆ www.beaverstatedental.com 1
C-4201Installation 1. Place the shipping carton in the desired location. Remove the cuspidor from the
shipping carton.
2. Run umbilical through cuspidor hole in top of side box. Place cuspidor drain shaft into hole.
3. Placethebowl,cupfillerandbowlrinsespoutsintheirappropriatelocations.
4. Connect the supply and vacuum tubings and gravity drain to the appropriate connections.
5. Connect the low voltage wiring for the vacuum switch to the electrical source.
1. Remove bowl and screw from package. Tuck bowl under the drain spout of the cover andpressbowlfirmlyintodraincup.
2. With the screw provided, fasten cover to the metal frame using a 5/64” hex wrench.
Test the bowl rinse.Testthecupfiller.Check for proper drain.Test HVE and Saliva Ejector valve for proper function.Adjust bowl rinse timer, if desired.
Final Adjustments
Bowl Installation
Ceramic Bowl
DrainSpout
DrainCup
001-1596-32 x 5/8” Screw
800.237.2303 ◆ 503.538.8756 ◆ 503.538.2845 Fax ◆ www.beaverstatedental.com2
C-42011. Thecupfillerisactivatedbypushingthecupfillerbutton.Thecyclestopsautomatically.
2. The timed bowl rinse is activated by pushing the rinse button. The timing for the rinsecyclemaybeadjustedwiththeflowcontrolvalvelocatedunderthecuspidorhousing. To adjust the rinse cycle:
a. Turn adjustment knob counter-clockwise to decrease the rinse cycle time. b. Turn adjustment knob clockwise to increase the rinse cycle time.
3. Flip the electric switch to activate the vacuum for the HVE and Saliva Ejector valve.
Operation
Cup Filler
Bowl Rinse
Bowl RinsePush Button
Cup FillerPush Button
Controls for Cuspidors
Timed Bowl RinseAdjustment Valve
under cover
NOTEThe bowl rinse timer is set at the factory for 15 seconds. This adjustment is very sensitive, only 1/8 of a turn will increaseordecreasethetimesignificantly.
1/4" Female water QDlocated on side box
800.237.2303 ◆ 503.538.8756 ◆ 503.538.2845 Fax ◆ www.beaverstatedental.com 3
Autoclavable HVE & S/E ValvesMaintenance Autoclavable vacuum valves may be autoclaved by steam autoclave or chemical vapor. Due
to the tight clearances and hardness of the anodized surfaces, the valve spool can become jammed or seized in the valve if it is sterilized while contaminated with debris.
Valves can be quickly disassembled for cleaning.
1. Pryonesideoftheleverandflipitoffofthevalve.Forsalivaejectorvalves,removethetip also.
2. Push out the valve spool and clean all the disassembled parts in an ultrasonic cleaner daily. Rinse thoroughly before putting into autoclave.
3. Lubricate the O-Rings on the spool with a silicone based lubricant before reassembly.
AUTOCLAVABLE S/E LEVER VALVE112-280
ITEM 1 2 3 4 5 6 7
PART DESCRIPTION Tip......................................................... Body..................................................... Spool................................................... O-Ring #10........................................ O-Ring, #12...................................... Swivel................................................. Lever....................................................
PART NUMBER 112-174 112-276 112-278 017-110 017-012 112-177 112-181
QTY 1 1 1 2 1 1 1 AUTOCLAVABLE UNIVERSAL HVE VALVE
112-263
ITEM 123456
PART DESCRIPTIONO-Ring, #13........................................Body, Chrome...................................Spool, Chrome..................................O-Ring, #13........................................Swivel.................................................Lever....................................................
PART NUMBER017-113112-272112-268017-013112-170112-169
QTY311111
44ASEPSIS CUSPIDORC-4201
ITEM 12345678
PART DESCRIPTIONCup Fill Spout.......................................Screw, 6-32 x 5/8”................................Cover.......................................................Washer, 7/16”........................................Receptical Spout Kit .........................Screw, 10-32 x 1/4”.............................Air Reservoir.........................................Frame......................................................
PART NUMBER132-068001-159145-027003-047132-069001-087132-044145-030
ITEM9
10111213141516
ITEM171819202122232425
PART DESCRIPTIONDrain Assembly...................................Drain Shaft ............................................Bushing..................................................Syringe ...................................................Autoclavable S/E.................................Autoclavable HVE...............................Bowl Rinse Spout................................Ceramic Bowl .......................................
PART DESCRIPTIONCeramic Bowl O-Ring ........................Cup Fill Valve ........................................Bowl Rinse Valve .................................Signal Valve...........................................Bowl Rinse Adjustment Valve.........Holder Bar .............................................Screw, 10-32 x .75 ...............................Universal HP Holder ..........................Drain Tubing.........................................
PART NUMBER132-041132-103013-017113-400112-280112-262132-067132-110
PART NUMBER017-150132-018132-020132-035132-122132-111001-034141-023115-747
QTY14162411
QTY11111111
QTY21111143
7ft.
55 SIDE BOX145-000
ITEM 123456789
10
PART DESCRIPTIONVacuum Canister.. ................................Canister bracket ................................. Top cover side box..............................Pelton - crane adapter........................Riser post side box .............................Screw, 10-32 x 5/16”...........................Washer #10 ...........................................Water bottle bracket........................... Pressure head ......................................Set screw 5/16”-18 x 1/2” ................
PART NUMBER131-038131-021145-025146-208144-064001-077003-024145-013110-007005-010
ITEM11121314151617181920
ITEM21222324
PART DESCRIPTIONSet Screw, 10-32 x 3/16“ ....................Aluminum block ..................................1 Liter bottle..........................................Side Cover..............................................Screw 10-32 x 50”.................................Washer.....................................................Base housing cover.............................Plug 3/4“ .................................................Plug 1/2” .................................................1/4“ Feemale QD..................................
PART DESCRIPTIONNeedle flow valve........................Screw, 8-32 .375............................40” side box post..........................Screw, 8-32 x 1/4”.........................
PART NUMBER005-037145-026110-041145-041001-171003-063145-040010-005010-013024-003
PART NUMBER121-000001-161146-204001-003
QTY11121
142114
QTY1112441131
QTY1112
663-WAY IC SYRINGE113-400
ITEM 123456789101112
PART DESCRIPTIONSyringe Tip ..................................Syringe Tip Nut ............................O-Ring, 006 .................................O-Ring, 228-040..........................Syringe Tip Adapter .....................O-Ring, 003 .................................Syringe Head...............................Syringe Button Repair Kit ............Pin, Syringe Cartridge .................Plastic Washer, 8-32....................Barb, 8-32 x 1/16’’ .......................Syringe Handle...........................
PART NUMBER113-030113-402017-006017-023113-227017-003113-405113-403113-008003-030022-040113-236
QTY112111121221
Quick-disconnect Syringe Tip Assembly 113-230
1. Remove the existing hex nut. A hex nut wrench has been provided.
2. Remove the syringe tip adapter using the hex wrench.
3. Insert the new syringe tip adapter into the syringe head and tighten using the hex wrench.
4. Place the two (2) O-Rings into the syringe tip nut.
5. Screw the syringe tip nut onto the syringe tip adapter and tighten using the hex wrench.
6. Push the syringe tip into the syringe tip nut until it snaps into place.
Replace Syringe Tip
1. Pull the dirty syringe tip out and insert a clean syringe tip into the syringe tip nut until it snaps into place.
77 BLEED VALVE (CUSPIDOR)132-122
ITEM 1 2 3 4 5 6 7
PART DESCRIPTIONStem ...................................................................O-Ring, 003.......................................................Body...................................................................Set Screw, 4-40 x 1/8" ...................................Plastic Waher, 1/4" .........................................Barb,m 10-32 x 1/16"....................................Plastic Elbow, 10-32 x 1/16" .......................
PART NUMBER132-124017-003132-123005-030003-001022-009022-039
QTY1111111
88PUSH BUTTON VALVE, AIR (CUSPIDOR)132-020
PUSH BUTTON VALVE, WATER (CUSPIDOR)132-018
ITEM 1 2 3 4 5 6 7 8 9
10 11
PART DESCRIPTIONSyringe Button......................................O-Ring, 007.............................................Dress Nut ................................................Body..........................................................Plastic Washer, 1/4"..............................Barb, 10-32 x 1/8".................................O-Ring, 001.............................................Stem .........................................................O-Ring, 002.............................................Spring ......................................................Plug ..........................................................
PART NUMBER113-401017-007002-517132-028003-001022-010017-501132-021017-502008-023020-026
QTY 1 1 1 1 2 2 2 1 1 1 1
ITEM 1 2 3 4 5 6 7 8 9
10 11 12
PART DESCRIPTIONSyringe Button.....................................O-Ring, 008............................................Dress Nut ...............................................Air Signal Adapter...............................Set Screw, 4-40 x 3/16"......................Valve Body.............................................O-Ring, 001............................................Stem ........................................................O-Ring, 002............................................Spring .....................................................Plastic Washer, 1/4" ............................Barb, 10-32 x 1/16" .............................
PART NUMBER113-401017-008002-517132-014005-001123-017017-501123-015017-502008-011003-001022-009
QTY111111211111
99 WATER RELAY VALVE (CUSPIDOR)132-035
ITEM 123456789
10
PART DESCRIPTIONHex Barb, 10-32 x 1/16" ...............................Plastic Washer, 1/4" .......................................Valve, Air Pilot .................................................Cup Seal ............................................................Piston, Air Pilot................................................Stem, 2-Way Toggle Valve...........................Spring, Toggle Valve......................................#4 Split Ring ....................................................Body, Toggle Valve.........................................Hex Barb, 10-32 x 1/8" ..................................
PART NUMBER022-009003-001014-002017-203014-008120-013008-002017-011120-010022-010
QTY1311111112
Cuspidor Bowl
Vacuum Cannister131-030
Air Reservoir132-044
Cup Fill Drain
Relay Valve132-035
S/E Valve112-280
HVE Valve112-262
1/8" Clear
1/4" Blue
Cup Fill Spout132-062
Water Valve, Push Button132-018
Bowl Rinse Spout132-062
Bleed Valve132-122
Air Valve, Push Button132-020
3/8" Sterling
5/8" Sterling
1/2" Spiral Sterling
1/8" Clear - Tee into Foot Control tubing, ribbed.1/4" Blue - Connect each to 1/4" barb on Water Master Valve (may require inserting an additional barb).
Circled letters indicate connection at Junction BoxJunction Box
Tee
1/4” Water QD016-020
BEAVERSTATE DENTAL, INC.115 S. ELLIOTT RD.
NEWBERG, OR 97132 USA800-237-2303 503-538-8756
503-538-2845 FAX
Asepsis CuspidorC-4201
Air Release do not connect
Vent
WARRANTY BDS warranties its product to be free of defects in material and workmanship for SEVEN FULL YEARS from the date of purchase on all units. Auto-blocks and standard foot controls are covered by a LIFETIME warranty when factory installed as part of a complete dental system. Upholstery has a TWO YEAR warrany and does not cover normal wear, stains, cuts or tears. Our LED lights have a FIVE year Warranty and our halogen light has a ONE year Warranty. Operatory chair motors feature a TEN year warranty. All warranty is limited to repair or replacement by BDS. Products returned to the factory and determined to be defective will be repaired or replaced free of charge at discretion of BDS. The warranty does not cover light bulbs, electrical components and trimmed or installed tubing. BDS reserves the right to void all warranties if any product is installed by other than an authorized technician. BDS' warranty does not cover damage to any surface finish, including but not limited to, discoloration or abrasion from cleaners, disinfectants or light sources. No claim for labor or consequential damages will be allowed. Return items must include a copy of the invoice or packing slip as proof of purchase.
Beaverstate Dental Inc.115 S. Elliott Rd.
Newberg, OR 97132800-237-2303503-538-8756
503-538-2845 Faxwww.beaverstatedental.com