Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System...
-
Upload
kaley-lippincott -
Category
Documents
-
view
337 -
download
4
Transcript of Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System...
www.anton-paar.com
Product Presentation C-PTD 180/AIR/QC
Air-cooled Peltier Temperature System
28.02.2013
www.anton-paar.com
Content
▸Pictures▸Features and Benefits▸Specification▸Performance▸Measurements▸Special Features▸Article Number and Price▸Configuration Example▸Applications and Markets▸Competition▸Summary
28.02.2013
www.anton-paar.com
Pictures P-CTD 180/AIR/QC
28.02.2013
www.anton-paar.com
Features and Benefits C-PTD 180/AIR/QC
▸ Temperature Range Complete range: 0 °C to 180 °C
▸ Measuring Systems Concentric Cylinder Measuring Systems
Disposable Measuring Systems
Stirrers and Spindles
▸ Air-cooled Peltier system No counter cooling necessary
▸ Typical Measurements Isothermal temperature tests
Temperature ramps
Temperature swing tests
28.02.2013
www.anton-paar.com
C-PTD 180/AIR/QC - Specification
• Temperature range 0 °C to 180 °C• Measuring systems Standard systems, DG42• Beaker Standard beaker, max. CC39 or DG42• General Good cleanability
Radiation system for counter cooling
Surrounding conditions• Temperature range 5 °C to 35 °C• Relative humidity, not condensing < 80 %• Altitude Up to 2000 m• Pollution Pollution degree 2 according to EN 61010• Surrounding Laboratory, indoor us only
Supply• Supply voltage 85 to 265 V• Frequency 50 … 60 Hz• Overvoltage Overvoltage category II according to
EN 6100
28.02.2013
www.anton-paar.com
Technical Specification – Heating Rates
28.02.2013
0
10
20
30
40
50
60
70
80
90
100
°C
T
10
100
1,000
10,000
mPa·s
0 5 10 15 20 25 30 35 40minTime t
C-LTD 0 - 80 °C (max) 1
T Temperature
Viscosity
C-PTD 0 - 80 °C (max) 2
T Temperature
Viscosity
Comparison Peltier / Liquid Temperature Control
Heating rate (Peltier): 8.0 °C/min
Heating rate (Liquid): 5.0 °C/min
www.anton-paar.com
Technical Specification – Cooling Rates
28.02.2013
-10
0
10
20
30
40
50
60
70
80
90
100
°C
T
10
100
1,000
10,000
mPa·s
0 20 40 60 80 100 120 140 160 180 220minTime t
C-LTD 80 - 0 °C (max) 1
T Temperature
Viscosity
C-PTD 80 - 0 °C (max) 3
T Temperature
Viscosity
Comparison Peltier / Liquid Temperature Control
Cooling rate (Peltier): 4.0 °C/min
Cooling rate (Liquid): 0.5 °C/min
www.anton-paar.com
Examples - Heating
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
200
°C
T
10-3
10-2
10-1
100
101
Pa·s
0 2 4 6 8 10 12 14 16 18 20minTime t
Anton Paar GmbH
Jump to 40 °C 2
CC39-SN5666; d=0 mm
T
Jump to 120 °C 2
CC39-SN5666; d=0 mm
T
Jump to 180 °C 2
CC39-SN5666; d=0 mm
T
Stable Conditions after approx. 15 min
28.02.2013
www.anton-paar.com
Examples - Cooling
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
200
°C
T
10-3
10-2
10-1
100
Pa·s
0 2 4 6 8 10 12 14 16 18 20minTime t
Anton Paar GmbH
Jump to 160 °C 2
CC39-SN5666; d=0 mm
T
Jump to 120 °C 6
CC39-SN5666; d=0 mm
T
Jump to 40 °C 3
CC39-SN5666; d=0 mm
T
Stable Conditions after approx. 15 min
28.02.2013
www.anton-paar.com
Special Features
• Wide temperature range: 0 °C to 180 °C
• Air cooled Peltier temperature system
• Rapid changes in temperature
• No external controller necessary
• No counter cooling by an external circulator
• Easy to install
• Space saving design
• Noise reduced
28.02.2013
www.anton-paar.com
Article Number
Item-no.108642
C-PTD 180/AIR/QCPeltier Temp. Control SystemTemp. range: 0 to 180 °C
28.02.2013
www.anton-paar.com
Standard Configuration
18318 Basic Instrument RheolabQC
108642 C-PTD 180/AIR/QC
20654 Disposable Measuring System CC39/D/QC-LTD
20655 Disposable Measuring System CC27/D/QC-LTD
Option:
18385 Software Rheoplus
28.02.2013
www.anton-paar.com
Applications & Markets
Coatings:Flow Curve, Yield Point,
3ITT (Thixotropy)
Food:Flow Curve, Yield Point,
Temperature Test,
3ITT (Thixotropy)BuildingMaterials:Flow Curve, Yield Point,
3ITT (Thixotropy)
28.02.2013
www.anton-paar.com
Applications & Markets
Adhesives: Flow Curve, 3ITT (Thixotropy),
Temperature Test
Cosmetics: Flow Curve, Yield Point,
Temperature Swing Test
Pharma: Flow Curve, Yield Point,
Temperature Swing Test
Soft Gels: Yield Point, 3ITT (Thixotropy)
Lubricants,Oils:
Flow Curve, Temperature Test
Asphalt: Flow Curve, Temperature Test
28.02.2013
www.anton-paar.com
Summary
C-PTD 180/AIR/QC is the new Peltier temperature system for the RheolabQC with:
▸ Wide temperature range: 0 °C to 180 °C▸ Air cooled Peltier temperature system▸ Cheaper configuration compared with the liquid temp. control systems▸ Compact design, no counter cooling by an external circulator
▸ Easy handling▸ Small footprint▸ Noise reduced compared to a fluid circulator ▸ Low power consumption▸ Fast and accurate temperature setting▸ no thermal fluids/oil necessary
▸ Firmware V1.26 for the RheolabQC required
28.02.2013