What does a mass spectrometer do? - UAB does a mass spectrometer do? 1. It measures mass better than...
Transcript of What does a mass spectrometer do? - UAB does a mass spectrometer do? 1. It measures mass better than...
What does a mass spectrometer do? 1. It measures mass better than any other technique.
2. It can give information about chemical structures.
What are mass measurements good for? To identify, verify, and quantitate: metabolites, recombinant proteins, proteins isolated from natural sources, oligonucleotides, drug candidates, peptides, synthetic organic chemicals, polymers
Experimental investigations on the conduction of electricity by gases led to the discovery of the electron in 1897, for which Thomson was awarded the and to the1906 Nobel Prize in Physics. Thomson J.J. On the Masses of the Ions in Gases at Low Pressures Philosophical Magazine, 1899, 48:295, p.547-567. Thomson observed that the new technique could be a used profitably by chemists to analyze chemicals. Thomson J.J. On Rays of Positive Electricity. The London, Edinburgh, and Dublin Philosophical Magazine and Journal of Science, 1907, XLVII.
The technique of mass spectrometry had its beginnings in J.J. Thomson's vacuum tube.
F = qE = ma y = ½ at2 L = vt
0
y
x -
Consider a Charged Particle in an Electric Field
_ _ _ _ _ _ _ _ _ _ _ _ _
+ + + + + + + + + + + + +
How does a mass spectrometer work?
Ion source: makes ions
Mass analyzer: separates ions
Mass spectrum: presents
information
Sample
Inlet Ion source
Mass Analyzer Detector
Data System
High Vacuum System
Mass Spectrometer Block Diagram
Inlet Ion Source
Mass Analyzer Detector
Data System
High Vacuum System
HPLC Flow injection Sample plate
Sample Introduction
Inlet Ion Source
Mass Analyzer
Detector Data System
High Vacuum System
MALDI ESI FAB LSIMS EI CI
Ion Source
High voltage applied to metal sheath (~4 kV)
Sample Inlet Nozzle (Lower Voltage)
Charged droplets
+ + + + +
+
+ + + + +
+
+ + + + +
+ + + +
+ + + + + +
+ + + + +
+ +
+ +
+
+ + + + + +
+ + +
MH+
MH3+
MH2+
Pressure = 1 atm Inner tube diam. = 100 um
Sample in solution
N2
N2 gas
Partial vacuum
Electrospray ionization:
Ion Sources make ions from sample molecules (Ions are easier to detect than neutral molecules.)
Electrospray Ionization (ESI)
Vacuum Atmospheric pressure
Mass Analyzer
N2 curtain gas
+HV
nebulizing gas
sample solution
+ + +
+ +
+ + + +
+ + +
+ +
+ +
[M + nH]n+ + +
1. Solvent evaporation 2. Coulombic repulsion
Electrospray Ionization (ESI) Process(Positive Mode)
Ions
1-5 kVpower supply
+
Spraycurrent ( ) i
+
Taylor coneNeedle
Site ofoxidation Site of
reduction
+
++
+ +
+
+
++
++ + + +
++++++
+
++
++ + ++
++
++ ++
+ +++
++++
++++ +
++++ +
++++ +
+++
+++
+++
+++
++
+
++
++
+
++
+
Massspectrometer
+
++
+
++
+ +
+
+
+
+
+
Solvent andneutralized ions
++++ +
++++ +
++++ +
+++
+++
+++
+++
++++ +
+++
+++
+++
+++
+++
+++
+++
+
++
+
+
+++++ +
+++
++
++ ++
+ ++ +
ESI droplets
Adapted from Kebarle, P and Tang, L, , , , 972AAnal. Chem. 65(22)1993
hν Laser
1. Sample is mixed with matrix (X) and dried on plate.
2. Laser flash ionizes matrix molecules.
3. Sample molecules (M) are ionized by proton transfer: XH+ + M MH+ + X.
MH+
MALDI: Matrix Assisted Laser Desorption Ionization
+/- 20 kV Grid (0 V)
Sample plate
MALDI generation of ions (Matrix-assisted laser desorption ionization)
Sample mixed with a UV-absorbing matrix and is allowed to co-crystallize on the metal target.
Laser pulse (337 nm)
Peptide/protein deposited on crystal surface
[M+H]+
[M-H]-
matrix ions
Matrices for MALDI analysis
Peptides/proteins - 3,5-dimethoxy-4-hydroxycinnamic acid (sinapinic acid) - α-cyano-4-hydroxycinnamic acid (CHCA) - 2,5-dihydroxybenzoic acid (DHB) - 2-(4-hydroxyphenylazo)-benzoic acid (HABA) Oligonucleotides - 2-aminobenzoic acid - 3-hydroxypicolinic acid (3-HPA) - 2,4,6-trihydroxyacetophenone (THAP) The choice of matrix depends greatly on the solute to be analyzed.
Inlet Ion source
Mass Analyzer Detector
Data System
High Vacuum System
Time of flight (TOF) Quadrupole Ion Trap Magnetic Sector FT-ICR MS Orbitrap MS
Mass Analyzer
•Operate under high vacuum (keeps ions from bumping into gas molecules)
•Actually measure mass-to-charge ratio of ions (m/z) The importance of the mass-to-charge ratio is that according to classical electrodynamics two particles with the same mass-to-charge ratio move in the same path in a vacuum when subjected to the same electric and magnetic fields.
•F=ma (Newton’s second law of motion)
•F=q(E + v × B) (Lorentz force Law)
•(m/q)a = E + v × B
•Key specifications are resolution, mass measurement accuracy, and sensitivity.
Mass analyzers separate ions based on their mass-to-charge ratio (m/z)
Time-of-flight (TOF) Mass Analyzer
+
+
+
+
Source Drift region (flight tube)
dete
ctor
V
• Ions are formed in pulses.
• The drift region is field free.
• Measures the time for ions to reach the detector.
• Small ions reach the detector before large ones.
Particles on which the same amount of work was done will have the same kinetic energy. Work is Force time distance Force = qE force is proportional to charge and field strength The distance is the same for all particles so the kinetic energy of all particles is the same and proportional to: ½ m/q v2
So a particle with twice the m/z will have a reduction in velocity of by a factor of the square root of 2. In general, the reduction in velocity is proportional to the square root of m/z. The consequence of this is that at higher m/z the smaller the difference in arrival times.
Time-of-flight (Tof) analyzer
Ions are accelerated so that they have equal kinetic energy. The ions “drift” down a 1 - 1.5 meter tube before striking a photomultiplier detector. “time of flight” (t) depends on the mass of the ion (m), where t = (m/2eV)1/2*D V is the applied potential and D is the flight tube distance. For a given instrument, the flight time varies as the square root of the mass of the ion.
Resolution 2 x 104
No upper limit of mass
Scan times ~ 1 µsec, good for LC-MSMS
Source
Accelerating pulse
Tof tube (1.0 - 1.5 m)
Effective flight tube (3.0 m)
Detector
Detector Reflector
Quadrupole Mass Analyzer Uses a combination of RF and DC voltages to operate as a mass filter.
• Has four parallel metal rods.
• Lets one mass pass through at a time.
• Can scan through all masses or sit at one fixed mass.
Quadrupole analyzer
Ions are accelerated electrically (5-15V) and passed along the long central axis of four rods arranged symmetrically. By applying combined DC and oscillating RF potentials, the ions drift along irregular flight paths along the rod axis. The DC/RF ratio is held constant and the absolute values of DC and RF are varied. Only ions with a particular m/z value have stable trajectories for a given value of DC and RF. If DC is set to 0, then all ions have stable trajectories. A scan can be accomplished over a period of 10-1000 msec.
Mass resolution 2 x 103
Tolerant of relatively high pressure (10-4 torr)
Upper limit for m/z is 3,000-4,000
Ion Traps
The ion trap is an energy well - ions with sufficient energy to enter the trap are retained by an energy barrier on the exit side of the trap. The advantage of the ion trap is that it accumulates selected ions prior to their analysis giving it high initial sensitivity (detection limit of approx. 20 fmol). Ions are fragmented by collision with helium gas and their daughter ions analyzed within the trap. Selected daughter ions can undergo further fragmentation, thus allowing MSn. The ion trap has a high efficiency of transfer of fragment ions to the next stage of fragmentation (unlike the triple quadrupole instrument).
Inlet Ion source
Mass Analyzer Detector
Data System
High Vacuum System
Microchannel Plate Electron Multiplier Mass analyzer/ion trap AC image
Detector
Ion Mobility Mass Spec (IMS)
• Provides information about the size and shape of a molecule in the gas phase
• Imagine putting a size exclusion column in front of the mass spectrometer
Ion Mobility Mass Spec (IMS)
Provides information about the Size and shape of a molecule In the gas phase
The mass spectrum shows the results R
elat
ive
Abun
danc
e
Mass (m/z)
0
10000
20000
30000
40000
50000 100000 150000 200000
MH+
(M+2H)2+
(M+3H)3+
MALDI TOF spectrum of IgG
Assigning numerical value to the intrinsic property of “mass” is based on using carbon-12, 12C, as a reference point. One unit of mass is defined as a Dalton (Da). One Dalton is defined as 1/12 the mass of a single carbon-12 atom. Thus, one 12C atom has a mass of 12.0000 Da.
How is mass defined?
Isotopes
+Most elements have more than one stable isotope. For example, most carbon atoms have a mass of 12 Da, but in nature, 1.1% of C atoms have an extra neutron, making their mass 13 Da.
+Why do we care? Mass spectrometers can “see” isotope peaks if their resolution is high enough.
If an MS instrument has resolution high enough to resolve these isotopes, better mass accuracy is achieved.
Element Mass AbundanceH 1.0078
2.014199.985%0.015
C 12.000013.0034
98.891.11
N 14.003115.0001
99.640.36
O 15.994916.999117.9992
99.760.040.20
Stable isotopes of most abundant elements of peptides
Calculating Molecular Weight Monoisotopic mass (exact mass) Mass of an ion calculated using the exact mass of the most abundant isotope. Average mass Mass of an ion calculated using average atomic weight of each element. Nominal mass Mass of an ion calculated using the integer mass of the most abundant isotope.
1981.84
1982.84
1983.84
Mass spectrum of peptide with 94 C-atoms (19 amino acid residues)
No 13C atoms (all 12C)
One 13C atom
Two 13C atoms
“Monoisotopic mass”
Mass spectrum of insulin
12C : 5730.61
13C
2 x 13C
Insulin has 257 C-atoms. Above this mass, the monoisotopic peak is too small to be very useful, and the average mass is usually used.
Monoisotopic mass
Monoisotopic masscorresponds tolowest mass peak
When the isotopes are clearly resolved the monoisotopic mass is used as it is the most accurate measurement.
Average mass Average mass corresponds to the centroid of the unresolved peak cluster
When the isotopes are not resolved, the centroid of the envelope corresponds to the weighted average of all the the isotope peaks in the cluster, which is the same as the average or chemical mass.
6130 6140 6150 6160 6170
Poorer resolution
Better resolution
What if the resolution is not so good? At lower resolution, the mass measured is the average mass.
Mass
Two peptides - same nominal mass - simulation
Peptide mixture: [Val5]-Angiotensin II Lys-des-Arg9-Bradykinin Sequence: DRVYVHPF KRPPGFSPF Formula: C49H69N13O12 C50H73N13O11
Exact mass: [M+2H]2+ = 516.76671 [M+2H]2+ = 516.78490 ∆m (mmu): 18.2 mmu
RP = 18,000
Inte
nsity
(%)
0
20
40
60
80
100
m/z 516.65 516.70 516.75 516.80 516.85 516.90
516.77581 (observed)
516.76671 (correct)
516.78490 (correct)
RP = 56,700
Inte
nsity
(%)
0
20
40
60
80
100
m/z 516.65 516.70 516.75 516.80 516.85 516.90
516.78490 516.76671
15.01500 15.01820 15.02140 15.02460 15.02780 15.03100Mass (m/z)
100
0
10
20
30
40
50
60
70
80
90
100
% In
ten
sit
yISO:CH3
15.0229M
FWHM = ∆M
R = M/∆M
How is mass resolution calculated?
Mass measurement accuracy depends on resolution
0
2000
4000
6000
8000
Cou
nts
2840 2845 2850 2855
Mass (m/z)
Resolution = 14200
Resolution = 4500
Resolution =18100 15 ppm error
24 ppm error
55 ppm error
High resolution means better mass accuracy
489 490 491 492 493 494 495 496m/z
489 490 491 492 493 494 495 496m/z
489 490 491 492 493 494 495 496m/z
FT-ICR MS
RP = 491.2594 / 0.0055 amu = 89,319
Mass Resolution = m / Δm 50%
489 490 491 492 493 494 495 496m/z
LTQ MS
RP = 491.45 / 0.69 amu = 712
How Do We Determine Charge State of a Protein?
1599.8
1499.9 1714.1
1845.9 1411.9
1999.6 2181.6
m/z
m/z = (MW + nH+)/n
040308
200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z0
100
%
Mix 406 (7.446) Cm (399:413) TOF MS ES+ 1.06e3985.7959949.3958
949.2087
915.3472883.8630
854.4720
826.8801
826.7056
1025.1672
1025.30191070.7888
1114.34631164.8209
1220.1954
040308
910 920 930 940 950 960 970 980 990 1000 1010 1020 1030 1040 1050 1060 1070 1080 1090m/z0
100
%
Mix 406 (7.446) Cm (399:413) TOF MS ES+ 1.06e3985.7959949.3958
949.2087915.4742
915.3472
914.5703
913.5538
917.9626918.0758
918.3589
918.5995948.7340
919.1232
951.9443952.0451 985.5759
952.2180
952.4198
952.7657 985.2242953.0972
1025.1672988.4808
988.7010
988.7597
988.90661024.7339989.0682
989.4500 1024.6140
1025.30191067.8568
1028.08501067.6586
1028.2646
1028.30961028.4144
1028.6390 1067.3838
1070.7888
1070.9875
1071.0946
1071.3544
1071.5532
Mass Spectrum of Denatured Intact Protein
Q1
Ion Mirror (reflector)
Effective Flight Path = 2.5 m
Q2
Q0
Sample
Quadrupole-Time of Flight (Q-ToF)
Ion Mobility Mass Spec (IMS)
• Provides information about the size and shape of a molecule in the gas phase
• Imagine putting a size exclusion column in front of the mass spectrometer