USP for Validating UV/Visible Spectrophotometersamcham.dk/files/editor/USP_857.pdf · USP •...
Transcript of USP for Validating UV/Visible Spectrophotometersamcham.dk/files/editor/USP_857.pdf · USP •...
HUMAN HEALTH • ENVIRONMENTAL HEALTH
© 2014 PerkinElmer
USP<857> for Validating
UV/Visible Spectrophotometers Steve Upstone/Helge Madsen
15th April 2016
USP<857>
• Introduced on 1st May 2015
• Mandatory from 1st May 2016 (USP numbers <1000 are
mandatory, > 1000, advisory)
• Covers validation of UV/Visible Spectrophotometers
• Major change is the new stray light tests. Other checks similar to
existing Ph Eur (European Pharmacopoeia)
USP <857> Document
What Testing is Covered by USP<857>?
• Control of Wavelengths
◦ D2 or Hg emission lines
◦ Rare earth solutions (eg HoClO2, didymium)
◦ Rare earth glasses (Holmium, didymium)
• Control of Absorbance
◦ Potassium Dichromate (235, 257, 313 and 350nm) – 430nm not mandated
◦ ND glass filters (eg NIST 930e, 1930 or 2930 grey-glass)
• Stray Light
◦ KCl (12g/L aqueous) 190-205nm region
◦ NaI (10g/L aqueous) 210-259nm region
◦ Acetone (pure) 250-320nm region
◦ NaNO2 (50g/L aqueous) 300-385nm region
• Resolution
◦ Toluene in hexane (0.02% v/v solution)
What Testing is Covered by USP<857>?
• Control of Wavelengths
◦ D2 or Hg emission lines
◦ Rare earth solutions (eg HoClO2, didymium)
◦ Rare earth glasses (Holmium, didymium)
• Control of Absorbance
◦ Potassium Dichromate (235, 257, 313 and 350nm) – 430nm not mandated
◦ ND glass filters (eg NIST 930e, 1930 or 2930 grey-glass)
• Stray Light
◦ KCl (12g/L aqueous) 190-205nm region
◦ NaI (10g/L aqueous) 210-259nm region
◦ Acetone (pure) 250-320nm region
◦ NaNO2 (50g/L aqueous) 300-385nm region
• Resolution
◦ Toluene in hexane (0.020% v/v solution)
6
Why Stray Light is Important – It Affects Measurement at
Higher Absorbances
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0
True Absorbance
Me
asu
red
Ab
so
rba
nce
0.005% Stray Radiation
0.01% Stray Radiation
0.05% Stray Radiation
0.1% Stray Radiation
Stray Light – How it is Manifested
How Stray Light Has Been Measured in the Past
• Comparison of trace in transmittance with a blocked beam
(ASTM method using NaI and NaNO2)
• Measurement of absorbance at a specified test point (eg Ph Eur
test using 12g/L aqueous KCl at 200nm or 198nm) – should
exceed a minimum value (eg 2A)
Ph Eur Stray Light Test
9
ASTM Stray Light Methodology
• ASTM methodology measures stray light (in transmittance
compared to a block beam) using NaI, NaNO2 and acetone
USP<857> Methodology for Stray Light
• Based on 1982 paper of Mielenz et al. (Appl Opt. 1982 Sep
15;21(18):3354-6. doi: 10.1364/AO.21.003354.)
• Method developed around PMT/ chopper based instrumentation
(only type that was around in 1982) so some features of the test
are now superfluous and could cause issues with current dual-
beam and PDA instruments
• Requires four standards (KCl, NaI, NaNO2, Acetone) in 10mm
sealed cuvettes – together with 5mm sealed cuvettes for the
reference channel
Methodology Developed on Double Beam Spectrophotometer
• Original 1982 Mielenz paper based on chopper-based
double beam instrument where beam is multiplexed
between sample and reference position 25 times a second
Principle of Reference Beam Attenuation for Double Beam
Instruments
The top picture shows the
system what happens in a
double beam instrument when
a highly absorbing sample is
measured. The energy
difference between the front
and rear beam is too great
and the system becomes
noisy
By attenuating the reference
beam, the energy difference
becomes less and the noise is
eliminated
USP<857> Standards Set
What is Being Measured – Measurement of the difference
between a 10mm and 5mm NaI cuvette
What the Result Looks Like - Stray Light Measurement (NaI)
Different Instruments Give Different “Peak” Locations
Data source: Starna Catalog
Stray Light Calculation
• Maximum absorbance (A) is recorded
• Stray Light Defined by
• 𝑆𝑡𝑟𝑎𝑦 𝐿𝑖𝑔ℎ𝑡 𝑠 = 0.25 ∗ 10(−2∗𝐴)
• A must be >=0.7 and S =< 0.01 to pass
Typical Lambda 25/35 Values (1nm slit)
Test Stray Light
KCl 0.0048
NaI 0.0016
Acetone 0.0014
NaNO2 0.0028
Potential Issues
• Instrument has to have ability and sensitivity to measure close to 4AU – most solid state instruments can measure up to around 3AU
• Using the 5mm cuvette as a reference beam attenuator is irrelevant to non-PMT/chopper instruments. It actually makes life worse for dual beam and photodiode array spectrophotometers
• Customers may mistake the maximum as being a “peak” and attempt to use it as a way of calibrating wavelengths or indicate that there is a “problem” with the instrument if it varies from the position found on the standards certificate.
• Not all instruments can measure above 3AU so the norm recommends attenuation during the autozero phase
• The norm is ambiguous regarding instruments with only one sample position (ie no reference position)
Bag of nails…..?
• Definition taken from ”The 1811 Dictionary of the Vulgar
Tongue”, originally by Francis Grose
• ”He squints like a bag of nails…” ie his eyes are directed as
many ways as the points of a bag of nails
Compatibility with New PerkinElmer UOQ
New PerkinElmer Instruments for Pharmaceutical Analysis
LAMBDA 365 LAMBDA 465 PDA