    Multi-Parameter TROLL 9000WQP-100

    Operators Manual

    September 2005

  • 7/31/2019 TROLL_9000 Manual 2005

    2/171

    Copyright 2001 2005 by In-Situ Inc. All rights reserved.

    Revision history initial release May 24, 2002

    rev. 001 October 3, 2002

    rev. 002 December 11, 2002

    rev. 003 April 2, 2003

    rev. 004 April 30, 2003

    rev. 005 September 11, 2003

    rev. 006 December 3, 2003

    rev. 007 May 24, 2004

    rev. 008 November 19, 2004

    rev. 009 April 2005

    rev. 010 September 2005

    This document contains proprietary information which is protected by copyright. No part of this document may be photocopied, reproduced, or translated to

    another language without the prior written consent of In-Situ Inc.

    Mailing & Shipping Address: Phone: 970 498 1500

    In-Situ, Inc. Fax: 970 498 1598

    221 E. Lincoln Ave. Internet: www.in-situ.com

    Ft. Collins, CO 80524 Support Line: 800 446 7488

    USA (US & Canada)

    The information in this document is subject to change without notice. In-Situ Inc. has made a reasonable effort to be sure that the information contained herein is

    current and accurate as of the date of publication.

    In-Situ Inc. makes no warranty of any kind with regard to this material, including, but not limited to, its fitness for a particular application. In-Situ will not be liable

    for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

    In no event shall In-Situ Inc. be liable for any claim for direct, incidental, or consequential damages arising out of, or in connection with, the sale, manufacture,

    delivery, or use of any product.

    Images in this manual have been selected for illustration; actual images may vary from those shown.

    In-Situ and the In-Situ logo, Win-Situ, Pocket-Situ, miniTROLL, TROLL, BaroTROLL, HERMIT, Quick-Connect, FlowSense, Profiler Plus, ComIT, RuggedReader

    and RDO are trademarks or registered trademarks of In-Situ Inc. Microsoft, Windows, Excel, ActiveSync, and Windows CE.Net are trademarks or registered

    trademarks of Microsoft Corporation. iPAQ is a trademark of the Hewlett-Packard Development Company, L.P. Teflon and Delrin are registered trademarks of

    E. I. DuPont de Nemours and Company. Viton is a registered trademark of DuPont Dow Elastomers. Other trademarks are the property of their respective

    owners.

  • 7/31/2019 TROLL_9000 Manual 2005

    3/171

    rev. 010 09/05

    Multi-Parameter TROLL 9000

    Operators Manual

    Contents

    1 INTRODUCTION ................................................ 1General Description of the MP TROLL 9000 ...................................... 1

    How to Use This Manual ............................................ ........................ 1Conventions ............................................. .................................. 2

    Unpacking and Inspection ....................................................... ........... 2Serial Number.... .................................................... .................... 2

    To Our Customers . . . ................................................ ........................ 2

    What We Provide .................................................... ........................... 3Warranty Provisions ................................................... ................ 3

    Firmware & Software Upgrades................................................. 3How to Contact Us .................................................. ........................... 3

    To Obtain Repair Service (U.S. and Canada) ............................ 3

    Outside the U.S. and Canada ........................................... 3Certification ....................................................... ................................. 3

    Cleaning Guidelines .................................................. ......................... 4Instrument Return Procedure......................... ............................ 4

    2 COMPONENTS & FEATURES ............................... 5System Description ............................................... ............................. 5

    Standard Features.................................................... .......................... 5

    Optional Features .................................................... ........................... 5Power Options ............................................... ..................................... 6

    Internal Power ......................................................... .................. 6External Power ......................................................... ................. 6

    Network Power Supplies, 12 VDC ............................................. 6Cables .......................................................... ...................................... 7

    Quick-Connect Submersible Cable ............................................ 7

    Surface Connector ............................................................ 7Vented Cap .............................................................. ......... 7

    Communication Cables .................................................... ......... 7ComIT2 Communication Interface ................................. 7

    Computer Connection Cable ("Comm Cable") .................. 7

    Direct-Read Cable ....................................................... ..... 7Versions and Feature Sets ......................................................... ........ 8

    Control Software ..................................................... ............................ 8Wakeup ................................................... ........................................... 8

    Smart Sensors .......................................................... .......................... 9

    System Connectors ...................................................... ...................... 9Triple Moisture Protection......... .................................................. ........ 9

    Components & Accessories List ................................................. ........ 9Product Specifications ............................................. ......................... 10

    3 GETTING STARTED .......................................... 12Power .................................................... ........................................... 12

    Sensors ................................................ ............................................ 12

    Cable ............................................... ................................................. 13Quick-Connect Cable ................................................ .............. 13

    PC Connection ................................................... .............................. 13

    Software Setup ................................................ ................................. 14Install the Software ..................................................... ............. 14

    Establish Communication with the MP TROLL 9000 ............... 14Set the Real-Time Clock ............................................. ............. 14

    Quick-Cal the Basic Water-Quality Sensors ..................................... 15Quick-Cal Tips ............................................ .................... 15

    Calibrate the Extended Sensors .............................................. 18

    Set Up Custom Pressure/Level Measurements ................................ 18Set Up to Log Data (Set Up a Test) .................................................. 19

    Complete the Setup .................................................... ...................... 19DeploymentProfiling ............................................................ .......... 19

    Checklist .................................................. ................................ 19

    DeploymentLong-Term Monitoring ................................................ 19Checklist .................................................. ................................ 19

    4 CONTROL SOFTWARE......................................20System Requirements ................................................... ................... 20

    Win-Situ ................................................... ................................ 20Pocket-Situ ................................................................ .............. 20

    Installation ................................................... ..................................... 20

    Win-Situ (Desktop or Laptop PC) ............................................ 20Pocket-Situ (PDA) ......................................................... .......... 20

    About PocketSync ............................................... .................... 21Setting PocketSync Options .................................................. .. 22

    The User Interface .................................................... ........................ 22

    The Navigation Tree ..................................................... ........... 22Software Functions.................................................... ....................... 23

    Configuring the Port .................................................... ............. 23Establishing Communication with the MP TROLL 9000 .......... 23

    New Connection............................... ............................... 23Existing Connection ............................................. ........... 23

    Displaying Device Information ............................................ ..... 23Upgrading Firmware and Features .......................................... 24

    Firmware Upgrade ....................................................... ... 24

    Features Upgrade ................................................... ........ 24Editing the Device Properties .................................................. 25

    Setting the Real-Time Clock ........................................... 25

    Setting a Device Name ............................................. ...... 25Setting Battery Information ............................................. 25

    Choosing Measurement Units and Other Preferences ............ 26Units ................................................. .............................. 26

  • 7/31/2019 TROLL_9000 Manual 2005

    4/171

    rev. 010 09/05

    MULTI-PARAMETER TROLL 9000 OPERATORS MANUA

    Other Preferences................ ........................................... 26

    Choosing Pressure Display & Conversion Options .................. 27Performing Water-Quality Calibrations .................................... 27

    Logging Data (Adding a Test) ................................................. . 28

    Taking Manual Readings ................................................. ...... 28Extracting and Viewing Data .................................................... 29

    Deleting Data from Instrument Memory ................................... 29Exiting the Software ...................................................... ................... 29

    Win-Situ ................................................ ................................... 29Pocket-Situ ........................................................... ................... 29

    5 PROFILING .................................................... 30Profiler Features .............................................. ................................. 30

    Starting the Profiler ....................................................... .................... 30

    Graphing Profiler Data ............................................... ....................... 31Profiling Dissolved Oxygen .................................................... .......... 31

    Profiling Pressure or Depth ....................................................... ....... 31Profiling Turbidity........................................................... ................... 31

    Customizing the Profiler ............................................ ....................... 33

    Changing the Channels Displayed .................................. 33Changing Measurement Units ........................................ 33

    Changing the Sample Rate ............................................. 33Starting in Profiler Mode ................................................. 33

    Logging Profiler Data ................................................ ........................ 33To Stop Logging ........................................................ ............... 33

    Retrieving Logged Profiler Data ............................................... 33

    Exiting the Profiler ............................................... ............................. 33

    6 LOGGING DATA ............................................. 34Adding a Test to a Device: The Test Wizard ..................................... 34

    Test Setup Parameters ........................................................ .... 35

    1. Test Name (Optional) ................................................. . 35

    2. Parameters to Include ................................................. 353. Measurement Schedule .............................................. 35

    4. Measurement Interval ................................................. 355. Start Mode .......................................................... ........ 35

    Ending Setup ........................................................ ................... 36

    More on Measurement Schedules.................................................... 36Linear...................... ........................................................ ......... 36

    Event ......................................................... .............................. 36Delta .............................................. ................................. 36

    Minimum Delta Values .................................................... 36Default Storage ....................................................... ........ 36

    Linear Average.......... ....................................................... ........ 38

    Other Test Operations ...................................................... ................ 38Starting a Manual Start Test ...................................................... ....... 38

    Stopping a Test .............................................. ................................... 39Retrieving Logged Data ............................................. ....................... 39

    Deleting Tests ................................................. .................................. 39Cloning a Test .................................................... ............................... 39Transferring Files from a PDA to a Desktop PC ............................... 41

    Viewing Logged Data ............................................... ........................ 42Selecting Data in the Data Folder........ ............................................. 42

    Graphing Data ................................................. ................................. 43

    Changing the Data Display .............................................. ........ 44Transferring Data to Excel from Win-Situ ................................ 44

    Printing Data in Win-Situ ................................................... ...... 44Saving Text Files from Win-Situ ............................................... 44

    7 MONITORING PRESSURE (WATER LEVEL) .......... 45Why Monitor Pressure? ...................................................... .............. 45The Pressure Sensor .................................................. ..................... 45

    Factory Calibration ...................................................... ............ 45

    Operating Principle .................................................... .............. 46Non-Vented (Absolute) vs. Vented (Gauged) Sensors.......... ... 46

    Pressure vs. Depth vs. Level .............................................. ..... 46Zeroing a Gauged Pressure Sensor ........................................ 46

    Setting up Pressure Measurements: The Parameter Wizard ........... 47Pressure Setup Choices ............................................. ............. 47

    1. Name (Optional) ..................................................... ............. 472. Display Mode .................................................... ................... 47

    3. Converting Pressure to Depth or Level ................................ 48

    Conversion Using Specific Gravity ................................. 48Conversion Using Density, Latitude, and Elevation ........ 48

    4. Level Reference ...................................................... ............ 48Reference Time .......................................................... .... 48

    Liquid Density ............................................. ...................................... 48

    Quick Summary of Pressure Setup .................................................. 49Correcting Absolute Pressure Readings for Barometric Pressure .... 50

    Manual Barometric Pressure Compensation ........................... 50Using the MP TROLLs Barometric Pressure Sensor ...... 50

    Using a Second Unit and a Spreadsheet ........................ 50

    Automatic Barometric Pressure Compensation ....................... 50Collecting Barometric Pressure Data for Reference ....... 50

    Collecting PSIA Data ............................................... ....... 50Using the Baro Wizard ................................................ .... 50

    Recalibration Recommendations ...................................................... 51

    8 MONITORING TEMPERATURE........................... 52Why Monitor Temperature? ........................................................ ...... 52

    The Temperature Sensor ........................................................ .......... 52Role of the Sensor in Calibration ............................................. 53

    Software Functions.................................................... ....................... 53Units ...................................................... .................................. 53

    References ...................................................... ................................. 53

    9 MONITORING BAROMETRIC PRESSURE ............. 54What is Barometric Pressure? ...................................................... .... 54

    Why Monitor Barometric Pressure?. ................................................. 54The Barometric Pressure Sensor ..................................................... 54

    Units ....................................................... .......................................... 54Collecting Accurate Measurements without Vented Cable ............... 54

    Pressure vs. Elevation ............................................... .............. 55

    10 MONITORING WATER QUALITY: OVERVIEW..... 56Why Monitor Water Quality? ....................................................... ...... 56

    The Sensors ..................................................... ................................ 56Basic Sensor Set ............................................... ...................... 56

    Extended Sensor Set ........................................................ ....... 57

    Installing Sensors .............................................. ...................... 57Removing Sensors .................................................... .............. 58

    Sensor O-Rings ................................................ .............. 58Calibration Overview .................................................... .................... 58

    Traditional Calibration ................................................. ............. 58Quick Calibration ............................................... ...................... 59

    Out of the Box ................................................... ....................... 59

    Default Coefficients ........................................... ...................... 59Custom Sensor Preparation ............................................... ..... 59

  • 7/31/2019 TROLL_9000 Manual 2005

    5/171

    rev. 010 09/05

    MULTI-PARAMETER TROLL 9000 OPERATORS MANUA

    Preparing to Calibrate........... ....................................................... ..... 59

    Calibration Kits .............................................. .......................... 59The Calibration Cup ................................................. ................ 59

    Effect of Temperature on Calibration ....................................... 60

    Rinsing ....................................................... ............................. 60Stirring .......................................................... ........................... 60

    When to Stir? ........................................................ .......... 60Starting the Stirrer ....................................................... .... 60

    Attaching the Stirrer for Calibration ................................. 60Calibration Procedures .................................................... ................. 60

    After Calibration... ..................................................... ........................ 61Sensor Storage .......................................................... .............. 61

    Calibration History ............................................ ....................... 61

    How Often to Calibrate ............................................. ............... 62How to Check if Sensor is Still in Calibration ........................... 63

    When to Replace the Sensor ................................................... 63References .................................................. ..................................... 63

    11 PH .............................................................. 64What Is pH?.................................................... .................................. 64

    Why Measure pH? .................................................... ............... 64

    The pH Sensor ............................................ ..................................... 64Sensor Installation ....................................................... ..................... 65

    Calibration .................................................. ...................................... 65Overview.... ............................................................... ............... 65

    Nominal vs. Stable .................................................. ........ 66

    Calibration Solutions. ....................................................... ........ 66Calibrating with Other Solutions...................................... 66

    Recommended Calibration Frequency .................................... 66

    pH Quick Cal ............................................... ............................ 66Traditional pH Calibration Procedure ....................................... 66

    Resetting Default Coefficients ........................................... ...... 68Sensor Slope and Offset ................................................. ......... 68

    Units and Calculated Measurements. ............................................... 68Deployment Recommendations and Cautions ................................. 68

    Sensor Care and Handling .................................................... ........... 69

    Sensor Removal .............................................. ........................ 69Maintenance/Inspection/Cleaning ............................................ 69

    Storage .................................................. .................................. 69References .................................................. ..................................... 69

    12 CONDUCTIVITY ............................................. 70What Is Conductivity?... .................................................... ................ 70

    Why Measure Conductivity? ............................................ ........ 70

    How is Conductivity Measured? .............................................. 70The Conductivity Sensors ................................................ ................ 70

    Sensor Installation ....................................................... ..................... 71Calibration .................................................. ...................................... 71

    Overview.... ............................................................... ............... 71Nominal vs. Stable .................................................. ........ 71

    Calibration Solutions (Primary Standards) ............................... 71

    Calibrating with Other Solutions...................................... 71Recommended Calibration Frequency .................................... 71

    Conductivity Quick Cal .............................................. .............. 72

    Traditional Conductivity Calibration Procedure ........................ 72Using a Conductivity Meter as a Secondary Standard ............ 74

    Entering a Cell Constant Manually .......................................... 74Units and Calculated Measurements. ............................................... 74

    Basic Unit: AC ................................................... .............. 74Specific Conductance (SC) ............................................. 74

    Total Dissolved Solids (TDS) .......................................... 75Resistivity .................................................. ..................... 75

    Salinity ........................................... ................................. 75Deployment Recommendations and Cautions ................................. 75

    Conductivity and Temperature ................................................. 75

    Sensor Care and Handling ....................................................... ........ 75Sensor Removal .................................................. .................... 75

    Maintenance/Inspection/Cleaning......................... ................... 75

    Storage ...................................................... .............................. 75References ...................................................... ................................. 75

    13 DISSOLVED OXYGEN ..................................... 76What is Dissolved Oxygen? ................................................. ............ 76

    Why Measure Dissolved Oxygen?. .......................................... 76Polarographic Measurement of Dissolved Oxygen .................... 77

    The Polarographic Dissolved Oxygen Sensor .................................. 77

    Theory of Operation.... ........................................................... .. 77

    Sensor Conditioning ..................................................... ........... 77Sensor Installation .................................................. .......................... 77

    Fill the Membrane Module ........................................... ............ 77

    Install the Sensor in the MP TROLL 9000............................. ... 78Condition a Newly Installed Sensor ......................................... 78

    Calibration ...................................................... .................................. 78

    Overview............................................................. ..................... 78Nominal vs. Stable ............................................. ............. 79

    Calibration Solutions & Equipment .......................................... 79Recommended Calibration Frequency .................................... 79

    Dissolved Oxygen Calibration Tips .......................................... 79

    Dissolved Oxygen Quick Cal ................................................ ... 79Traditional Dissolved Oxygen Calibration Procedure .............. 79

    Sensor Slope and Offset ..................................................... ..... 83

    Units and Calculated Measurements.... ............................................ 83Deployment Recommendations and Cautions ................................. 84

    Stirring ................................................... .................................. 84Attaching the Stirrer ............................................ ............ 84

    Starting the Stirrer ........................................................... 84Sensor Care and Handling ....................................................... ........ 85

    Sensor Removal .................................................. .................... 85Maintenance/Inspection ................................................ ........... 85

    Cleaning the Electrodes .................................................... ...... 85

    Replacing the Membrane Module ............................................ 85Sensor Life ......................................................... ..................... 85

    Storage ...................................................... .............................. 85References ...................................................... ................................. 86

    Optical Measurement of Dissolved Oxygen.............................. ... 87

    The Optical Dissolved Oxygen Sensor ........................................... .. 87Theory of Operation.... ........................................................... .. 87

    Comparison to Polarographic D.O. Sensor .............................. 87Sensor Installation .................................................. .......................... 87

    Unpack the RDO Sensor ............................................. ............ 88

    Prepare the MP TROLL 9000 .................................................. 88Attach the RDO Sensor to the TROLL 9000 ............................ 88

    Calibration ...................................................... .................................. 89Overview............................................................. ..................... 89

  • 7/31/2019 TROLL_9000 Manual 2005

    6/171

    rev. 010 09/05

    MULTI-PARAMETER TROLL 9000 OPERATORS MANUA

    Nominal vs. Stable .................................................. ........ 89Calibration Solutions. ....................................................... ........ 89

    Recommended Calibration Frequency .................................... 89Traditional Two-Point Calibration ............................................. 90

    Oxygen Saturation Point ................................................. ......... 90Zero Oxygen Point ................................................... ................ 91Degrees of Phase Angle ................................................. ......... 92

    Units and Calculated Measurements. ............................................... 92Deployment Recommendations and Cautions ................................. 92

    Sensor Care and Handling .................................................... ........... 93Biofouling ....................................................... .......................... 93

    Cleaning ................................................. ................................. 93

    Foil Check .............................................. ................................. 94Replacing the Foil .................................................... ................ 94

    Updating the Foil Coefficients ....................................... ........... 94Specifications ................................................ ................................... 96

    14 OXIDATION-REDUCTION POTENTIAL ................ 98What is ORP?..... ....................................................... ....................... 98

    Why Measure ORP?................................................ ................ 98The ORP Sensor .................................................... .......................... 98Sensor Installation ....................................................... ..................... 99

    Calibration .................................................. ...................................... 99Overview.... ............................................................... ............... 99

    Nominal vs. Stable .................................................. ...... 100

    Calibration Solutions.................................................... .......... 100Calibrating with Other Solutions.................................... 100

    Recommended Calibration Frequency .................................. 100ORP Quick Cal ................................................ ...................... 100

    Traditional ORP Calibration Procedure .................................. 100

    Resetting Default Coefficients ........................................... .... 102Units and Calculated Measurements.............................................. 102

    Normalization to Standard Hydrogen Electrode Values 102

    Deployment Recommendations and Cautions ............................... 102Sensor Care and Handling .................................................... ......... 103

    Sensor Removal ........................................... ......................... 103Maintenance/Inspection/Cleaning.................................... ...... 103

    Storage .................................................. ................................ 103References .................................................. ................................... 103

    15 AMMONIUM .............................................. 104What is Ammonium ? ....................................................... .............. 104

    Why Measure Ammonium?......................... ........................... 104

    The Ammonium Sensor ............................................................ ...... 104Sensor Preparation ............................................................. ... 104

    Sensor Installation ..................................................... ............ 105Calibration .................................................. .................................... 105

    Overview.... ........................................................... ................. 105

    Calibration Solutions.................................................... .......... 105Recommended Calibration Frequency .................................. 106

    Preparing to Calibrate ...................................................... ...... 106Ammonium Calibration Procedure ......................................... 106

    Sensor Slope and Offset ................................................. ....... 108Units and Calculated Measurements.............................................. 108

    Deployment Recommendations and Cautions ............................... 108

    pH ....................................................... ................................... 108Temperature ................................................ ........................... 108

    Conductivity ........................................................... ................ 108

    Potential Interferences................................................. .......... 108Sensor Care and Handling ....................................................... ...... 109

    Sensor Removal .............................................. ...................... 109

    Maintenance/Inspection/Cleaning............................. ............. 109Storage ...................................................... ............................ 109

    References ...................................................... ............................... 109

    16 CHLORIDE...................................................110What Is Chloride? .................................................. ......................... 110

    Why Measure Chloride? ............................................. ........... 110

    The Chloride Sensor .................................................. .................... 110Sensor Preparation. ........................................................... .... 110

    Sensor Installation .................................................... ............. 110

    Calibration ...................................................... ................................ 111Overview......................................................... ....................... 111

    Calibration Solutions... .................................................... ....... 111Recommended Calibration Frequency .................................. 111

    Preparing to Calibrate ................................................. ........... 111

    Chloride Calibration Procedure.. ............................................ 112Sensor Slope and Offset ..................................................... ... 114

    Units and Calculated Measurements.. ............................................ 114Deployment Recommendations and Cautions ............................... 114

    pH Effects ......................................................... ..................... 114Potential Interferences................................................. .......... 114

    Sensor Care and Handling ....................................................... ...... 114

    Sensor Removal .............................................. ...................... 114Maintenance/Inspection/Cleaning................................. ......... 114

    Storage ...................................................... ............................ 114

    17 NITRATE .....................................................115What Is Nitrate ? ......................................................... .................... 115

    Why Measure Nitrate? ............................................... ............ 115The Nitrate Sensor .................................................. ....................... 115

    Sensor Preparation. ........................................................... .... 116Sensor Installation .................................................... ............. 116

    Calibration ...................................................... ................................ 116

    Overview......................................................... ....................... 116Calibration Solutions... .................................................... ....... 117

    Recommended Calibration Frequency .................................. 117Preparing to Calibrate ................................................. ........... 117

    Nitrate Calibration Procedure ................................................ 117Sensor Slope and Offset ..................................................... ... 119

    Units and Calculated Measurements.. ............................................ 119

    Deployment Recommendations and Cautions ............................... 119Temperature .................................................... ....................... 119

    Potential Interferences................................................. .......... 120Sensor Care and Handling ....................................................... ...... 120

    Sensor Removal .............................................. ...................... 120Maintenance/Inspection/Cleaning............................. ............. 120Storage ...................................................... ............................ 120

    References ...................................................... ............................... 120

    18 TURBIDITY ................................................. 121What Is Turbidity?....................... .................................................... 121

    Why Measure Turbidity? ....................................................... . 121

    How Is Turbidity Measured? .............................................. .... 121

  • 7/31/2019 TROLL_9000 Manual 2005

    7/171

    rev. 010 09/05

    MULTI-PARAMETER TROLL 9000 OPERATORS MANUA

    The Turbidity Sensor ................................................ ...................... 122

    The Turbidity Wiper ........................................................ ................ 122Wiper Installation ............................................. ...................... 123

    Wiper Movement ...................................................... ............. 123

    Manual Wipe............................................... ........................... 124Wiper Guidelines and Precautions ........................................ 124

    Calibration .................................................. .................................... 124Factory Calibration ................................................... ............. 124

    Field Calibration............ ................................................. ........ 124Calibration Solutions.................................................... .......... 124

    Calibrating with Other Solutions.................................... 125Recommended Calibration Frequency .................................. 125

    Turbidity Calibration Procedure ............................................. 125

    Resetting Default Coefficients ........................................... .... 128Sensor Slope and Offset ................................................. ....... 128

    Units and Calculated Measurements.............................................. 128Deployment Recommendations and Cautions ............................... 128

    Common Interferences .......................................................... 128

    Profiling Turbidity........................................................... ................. 128Long-Term Monitoring of Turbidity .................................................. 128

    Sensor Care ................................................ ................................... 128Inspection/Maintenance/Cleaning.................................... ...... 128

    Wiper Maintenance ................................................ ........................ 129

    Removing the Wiper Head ............................................ 129Replacing the Wiper Pad .............................................. 129

    Adjusting Wiper Pressure ............................................. 129References .................................................. ................................... 129

    19 INSTRUMENT NETWORKS ............................ 130Network Description .................................................. ..................... 130Network Boxes ................................................ ............................... 130

    Network T-Box, NTB-100 .................................................... ... 130Network Quad Box, NTB-400 ................................................ 130

    Network Mid-Line Tap Box, MLT-100 ..................................... 130Cables ................................................. ........................................... 130

    Network Communication Cable ............................................. 130

    Smart Device Network Communication Cable ....................... 130Network Interconnect Cable .......................................... ........ 130

    Network Extension Cable ............................................. ......... 131Power ................................................. ............................................ 131

    System Requirements .................................................... ................ 131

    Installation Procedure.. .................................................... ............... 131Telemetry .......................................................... .............................. 131

    20 SDI-12 OPERATION..................................... 133SDI-12 Requirements ...................................................... ............... 133

    Wiring ................................................ ............................................. 133MP TROLL Installation ....................................................... ............ 133

    SDI-12 Support ...................................................... ......................... 133Sensor Identification ...................................................... ........ 134

    Making SDI-12 Measurements ................................................ ....... 134

    Redundant Logging (Win-Situ and SDI-12) .................................... 135Networking Considerations............................ ................................. 135

    Reference .................................................. ..................................... 135

    SDI-12 Version 1.3 Commands ............................................. 136

    21 FIELD INSTALLATION .................................. 138Long-Term Monitoring Installations ............................................. .... 138Securing the Cable .................................................. ....................... 138

    Installation Tips for Long-Term Pressure/Level Monitoring ............. 138

    Stabilization Time ................................................... ........................ 139Installation Checklist .......................................... ............................ 139

    Backshell Hanger Installations ............................................. .......... 139Using a Stirrer .................................................. .............................. 140

    Starting the Stirrer ...................................................... ... 140Attaching the Stirrer for Deployment ............................. 140

    22 APPLICATIONS AND TECHNICAL NOTES ..........141Monitoring Background Levels .................................................... ... 141

    Long-Term Water Quality Monitoring .......................................... .... 141

    Pressure Testing ........................................................ ..................... 141Monitoring Stormwater Runoff ............................................ ............ 141

    Artesian (Sealed-Well) Applications ............................................... 141Low-Flow Monitoring with Flow-Sense ........................................... 142

    Preparation .................................................... ........................ 142

    Optional Preparation........... ................................................... 142Start the Software .............................................................. .... 142

    Launch the Flow-Sense Wizard ............................................. 142Verify Unit Preferences ............................................... .. 143

    Stabilization Readings ................................................. .......... 145Saving the Low Flow Data ................................................. .... 145

    Output ..................................................... ............................... 146

    Export to Excel Option ............................................... ............ 146Using a Custom Excel Template ................................... 146

    Hints and Tips........ ................................................. ........................ 146

    Priming the Flow-Sense Wizard in the Office ........................ 146

    23 CARE & MAINTENANCE ............................... 148Replacing Batteries ................................................. ....................... 148O-Ring Seals ............................................. ..................................... 148

    Lubrication ........................................................ ..................... 148Replacement ................................................... ...................... 149

    General Cleaning ................................................... ........................ 149

    Storage.................................................... ....................................... 149Vent Cap Desiccant ................................................. ....................... 149

    Elastomer ...................................................... ................................. 149Consumables .................................................. ............................... 149

    24 TROUBLESHOOTING ................................... 150Troubleshooring Connections..................................................... .... 150

    Troubleshooting Data Collection (Tests) ......................................... 150

    Troubleshooting Sensors ..................................................... ........... 151Troubleshooting Calibration............ ................................................ 151

    APPENDIX ...................................................... 152

    Boxcar Filtering ....................................................... ....................... 152Electronic Drift and Device Recalibration ....................................... 152Determining Density .................................................... ................... 153

    GLOSSARY ...................................................... 154

    Index ............................................................. 160

  • 7/31/2019 TROLL_9000 Manual 2005

    8/171

    1

    Multi-Parameter TROLL 9000

    Operators Manual

    rev. 010 09/05

    GENERAL DESCRIPTION OF THE MP TROLL 9000

    Your new Multi-Parameter TROLL 9000 water quality probe uses the

    latest sensor and electronics technology to provide a robust, durable,

    and user-friendly instrument.

    Designed for both short- and long-term monitoring, the MP

    TROLL 9000 logs data from up to 7 optional pressure and water-

    quality sensors, as well as built-in temperature and barometric

    pressure sensors. Many custom options and versions are

    available, so your instrument may not look exactly like those we

    have chosen to illustrate in this manual.

    The MP TROLL 9000 provides the convenience of a laboratory-

    quality measurement instrument for field use, providing true in-

    situ monitoring of water level and water quality.

    HOW TO USE THIS MANUAL

    This operators manual is designed as both a start-up guide and

    a permanent reference for the features, uses, and applications

    of the Multi-Parameter TROLL 9000.

    Section 1: Introduction to this Operators Manual and to In-Situ

    Inc.; contacting us for warranty and repair issues

    Section 2: Description of components and features of the Multi-

    Parameter TROLL 9000

    Section 3: Getting Started: setting up the software, connecting

    for the first time, Quick-Calibrating the sensors, and a summary

    overview of the setup to start collecting data

    Section 4: Control Software: Win-Situ and Pocket-Situ

    Section 5: Profiling

    Section 6: How to set up tests and start logging data; extracting

    data to your PC; viewing and graphing data

    1 INTRODUCTION

    Section 7: Monitoring Pressure (Water Level): Info on the two basic

    types of pressure sensors; setting up the pressure channel to obtain

    the most accurate readings for your altitude and latitude; how to

    display data as you wish to see it

    Section 8: Temperature

    Section 9: Barometric Pressure

    Section 10: An overview of monitoring water quality with the MP

    TROLL 9000: Sensor installation and calibration, software options,

    calibration reports

    Section 11: pH

    Section 12: Conductivity

    Section 13: Dissolved Oxygen

    PolarographicRDO Optical Dissolved Oxygen

    Section 14: Oxidation-Reduction Potential (ORP)

    Section 15: Ammonium

    Section 16: Chloride

    Section 17: Nitrate

    Section 18: Turbidity

    Section 19: Setting up networks of MP TROLL 9000s

    Section 20: SDI-12 Operations

    Section 21: Field installation of the MP TROLL 9000guidelines and

    precautions for long-term deployment

    Section 22: Applicationsand a place to store your In-Situ Applica-

    tion Notes and Tech Notes

    Section 23: Care and Maintenance advice

    Section 24: Troubleshooting: Look here first if you have problems

    making connections or calibrating . . .

  • 7/31/2019 TROLL_9000 Manual 2005

    9/171

    2 rev. 010 09/05

    CONVENTIONS

    Throughout this operators manual you will see the following symbols.

    UNPACKING AND INSPECTION

    Your MP TROLL 9000 was carefully inspected before shipping. Check

    the instrument for any physical damage sustained during shipment.

    Notify In-Situ and file a claim with the carrier if there is any such

    damage; do not attempt to operate the instrument.

    The instrument will arrive with batteries installed, basic sensors

    installed and kept moist in the Calibration Cup. Sensor port plugs are

    included for all sensor ports.

    Accessories may be shipped separately and should also be inspected

    for physical damage and the fulfillment of your order.

    SERIAL NUMBER

    The serial number is engraved on the body of the Multi-Parameter

    TROLL 9000. It is also programmed into the instrument and displayed

    when the instrument is connected to a PC running Win-Situ or Pocket-

    Situ. We recommend that owners keep a separate record of thisnumber. Should your Multi-Parameter TROLL 9000 be lost or stolen,

    the serial number is often necessary for tracing and recovery, as well

    as any insurance claims. If necessary, In-Situ maintains complete

    records of original owners names and serial numbers.

    Serial numbers of the water-quality sensors are stored in sensor

    memory and displayed in the software when sensors are installed.

    TO OUR CUSTOMERS . . .

    Thank you for your purchase of an In-Situ product. We are glad you

    chose us and our products to help you with your environmental

    monitoring needs. In-Situ Inc. has been designing and manufacturing

    world-class environmental monitoring instrumentation for over 25

    years in the Rocky Mountains of the United States. As it was in the

    beginning, our expectation is that this product will provide you with

    many trouble-free years of use. To that end, we pride ourselves on

    delivering the best customer service and support possible24 hours a

    day, 7 days a week. We believe that this level of commitment to you,

    our customer, is imperative in helping you ensure clean, safe

    groundwater and surface water resources across the globe. We also

    understand the need for accurate, reliable assessments and we

    continue to make significant investments in Research and Develop-

    ment to ensure that we deliver the latest product and technologicalinnovations to support your needs.

    Whether you are gathering information about your body of water for a

    few moments, or over a period of years, you can rely upon us to

    provide you with a quality product and outstanding customer support

    at a fair price and have that product delivered to you when and where

    you need it.

    We want your experience with In-Situ Inc. to be pleasant and

    professional, whether you are renting from us, or purchasing from us.

    We would be pleased to hear from you and learn more about your

    needs, and your experiences with our products. Again, we thank you

    for choosing In-Situ Inc. and we look forward to serving your needsnow, and in the future.

    Please save packing materials for future storage and

    shipping of your MP TROLL 9000. The shipping boxeshave been performance-tested and provide protection for

    the instrument and its accessories.

    Bob Blythe, President and CEO

    In-Situ Inc.

    [email protected]

    The check mark highlights a tip about a convenient featureof the MP TROLL 9000

    The exclamation point calls your attention to a requirementor important action that should not be overlooked

    Q:

    A:

    Typical usage questions and answers that we hope will

    help you gain a better understanding of your new Multi-Parameter TROLL 9000 and simplify its setup and

    operation

    SECTION 1: INTRODUCTION

  • 7/31/2019 TROLL_9000 Manual 2005

    10/171

    3 rev. 010 09/05

    WHAT WE PROVIDE

    WARRANTY PROVISIONS

    In-Situ Inc. warrants all products sold, excluding batteries sold withsuch products, against defects in materials and workmanship under

    normal operating conditions. Consult the separate warranty for

    specific warranties that may apply.

    FIRMWARE & SOFTWARE UPGRADES

    From time to time In-Situ may make available enhanced versions of

    firmware and software to its customers over the Internet.

    Visit our website at www.in-situ.com to download the latest firmware

    and software.

    HOW TO CONTACT USTechnical Support: 800 446 7488

    Toll-free 24 hours a day in the U.S. and Canada

    Address: In-Situ, Inc.

    221 E. Lincoln Ave.

    Ft. Collins, CO 80524

    USA

    Phone: 970 498 1500

    Fax: 970 498 1598

    Internet: www.in-situ.com

    TO OBTAIN REPAIR SERVICE (U.S. AND CANADA)

    If you suspect that your Multi-Parameter TROLL 9000 is malfunction-ing and repair is required, you can help assure efficient servicing by

    following these guidelines:

    1. Call In-Situ Customer Service toll-free at 1-800-446-7488. Have

    the product model and serial number handy.

    2. Be prepared to describe in detail the exact nature (symptoms) of

    the problem, including how the instrument was being used and the

    conditions noted at the time of the malfunction.

    3. If service personnel determine that service is required, they will

    assign an RMA (return material authorization) number.

    4. Write a description of the problem for service personnel, indicating

    whether the malfunction occurs intermittently or constantly. Save

    printouts or other materials that illustrate the problem.

    5. Clean the Multi-Parameter TROLL 9000 and cable. Decontaminate

    thoroughly if it has been used in a toxic or hazardous environment.

    See the cleaning guidelines and form at the end of this section.

    6. Pack your Multi-Parameter TROLL 9000 in its original shipping box

    if possible. Include your write-up of the symptoms, a statement

    certifying that the instrument and cable have been decontaminated

    and any supporting information.

    7. Send the package, shipping prepaid, to

    In-Situ, Inc.

    Customer Service

    ATTN: RMA # (assigned no. here)

    221 E. Lincoln Ave.

    Ft. Collins, CO 80524

    The warranty does not cover damage during transit. In-Situ recom-

    mends the customer insure all shipments. Warranty repairs will be

    shipped back prepaid.

    Outside the U.S. and Canada

    Contact your international In-Situ distributor for repair and service

    information.

    If an instrument returned for servicing shows evidence ofhaving been deployed in a toxic or hazardous environ-

    ment, Customer Service personnel will require written

    proof of decontamination before they can service the unit.

    CERTIFICATION

    The MP TROLL 9000 complies with all applicable directives per the

    CE and FCC and was tested to the EN 55022 / ICES-003 / FCC

    specifications.

    SECTION 1: INTRODUCTION

  • 7/31/2019 TROLL_9000 Manual 2005

    11/171

    4 rev. 010 09/05

    CLEANING GUIDELINES

    Please help us protect the health and safety of our employees by

    cleaning and decontaminating equipment that has been subjected to

    any potential biological or health hazards, and labeling such equip-

    ment. Unfortunately, we cannot service your equipment without such

    notification. Please complete and sign the form below (or a similar

    statement certifying that the equipment has been cleaned and

    decontaminated) and send it along to us with each downhole

    instrument.

    A good cleaning solution, such as Alconox (a glassware cleaning

    product available from laboratory supply houses) is recommended.

    Clean all cabling. (Dont forget connectors and protective caps.)

    Remove all foreign matter.

    Clean the probe bodyincluding the nosecone, restrictor, cable

    head, and protective caps. Remove all foreign matter.

    Remove and clean all removable sensors. Rinse with deionized or

    distilled water after cleaning.

    INSTRUMENT RETURN PROCEDURE

    1. Contact In-Situ Inc. Customer Service at 1-800-446-7488 and

    obtain an RMA (Return Materials Authorization) number from

    service personnel.

    If an instrument is returned to our Service Center

    for repair or recalibration without a statement that ithas been cleaned and decontaminated, or in the

    opinion of our Service Representatives presents a

    potential health or biological hazard, we reserve the

    right to withhold service until proper certification has

    been obtained.

    Alconox is a registered trademark of Alconox Inc.

    2. Complete and sign a Decontamination & Cleaning Statement

    (below) for each instrument.

    3. Carefully pack the instrument in its original packaging, if possible.

    Include your write-up of symptoms displayed by the instrument, the

    Decontamination & Cleaning form, and any supporting information.

    4. Send the package, shipping prepaid, to:

    In-Situ, Inc.

    Customer Service

    ATTN: RMA# (assigned number here)

    221 E. Lincoln Ave.

    Ft. Collins, CO 80524

    Company Name __________________________________________ Phone ________________________

    Address ________________________________________________________________________________

    City __________________________________ State ________________ Zip _______________________

    Instrument Type __________________________________Serial Number___________________________

    Contaminant(s) (if known) __________________________________________________________________

    ______________________________________________________________________________________

    Decontamination procedure(s) used __________________________________________________________

    ______________________________________________________________________________________

    Cleaning verified by________________________________________ Title __________________________

    Date _____________________________

    Decontamination & Cleaning Statement

    SECTION 1: INTRODUCTION

  • 7/31/2019 TROLL_9000 Manual 2005

    12/171

    5 rev. 010 09/05

    Multi-Parameter TROLL 9000

    Operators Manual

    SYSTEM DESCRIPTION

    The Multi-Parameter TROLL 9000 is a compact, modular system with

    a range of components to customize the instrument to various

    applications and operational modes. Components include theinstrument body with a host of options, sensors, a suite of cables for

    single instrument operation, network boxes and cables for network

    operation, external power accessories, and consumable items.

    STANDARD FEATURES

    4.5 cm (1.79 in) diameter

    316L stainless steel and acetal body construction

    Viton seals

    Fast data sampling

    Continuous real-time readings (profiling)

    User-replaceable off-the-shelf D-cell alkaline batteries

    Two A/D converters (24-bit and 10-bit)

    Triple protection against moisture

    Fully submersible*

    RS485 communications

    Low-power microprocessor

    Permanently installed, factory-calibrated temperature sensor

    Permanently installed, factory-calibrated barometric pressure

    sensor, for use on vented cable

    Temperature-compensated real-time clock

    Error-proof connector system

    Easy-to-use Win-Situ control software for setup, downloading,

    text and graphical data display

    Optically isolated communication signals

    Cable available in standard and custom lengths of vented

    polyurethane or FEP**, or non-vented HDPE

    External power capable

    OPTIONAL FEATURES

    Profiler and Professional feature sets, both available in XPwith

    extended parameter capability

    Smart Sensors for water quality measurementsremovable,

    replaceable, field-calibrated

    Basic Sensors include Dissolved Oxygen (polarographic),

    ORP, pH, Conductivity, Combination pH/ORP

    The Extended Sensor set includes Ammonium, Chloride,

    Nitrate, Turbidity, RDO Optical Dissolved Oxygen

    Pocket-Situ full-featured control software for field use on a

    PDAtake continuous real-time readings, calibrate, program,

    and download logged data from multiple MP TROLLs

    RuggedReader handheld PDA

    Temperature-compensated vented or non-vented pressure

    sensor, available in several ranges, permanently installed and

    factory-calibrated

    Permanently installed, factory-calibrated turbidity sensor

    Integral data logging capabilityup to 4 megabytes flash data-

    storage memory (about a million individual readings)

    Fast data downloads

    Up to 16-test capacity; linear, log, event, and linear average

    measurement schedules

    SDI-12 or RS485 interfaceor both FlowSense low-flow sampling software for automated

    sampling and report creation

    Optional accessories include low-flow flowcell, user-serviceable

    turbidity wiper, battery-powered magnetic stirrer for use in

    stagnant water

    Instrument networking and telemetry

    2 COMPONENTS & FEATURES

    * within the ratings of the installed sensor options.

    ** FEP = fluorinated ethylene propylene, the generic equivalent of DuPont Teflon.

  • 7/31/2019 TROLL_9000 Manual 2005

    13/171

    6 rev. 010 09/05

    POWER OPTIONS

    The MP TROLL 9000 operates in 3 power modes:

    internal power external battery pack

    external AC line power

    These may be used separately or together in any combination.

    INTERNAL POWER

    MP TROLL 9000:

    two standard 1.5V alkaline D cells, OR

    two 3.6V lithium D cellsrecommended for use with an RDO

    optical dissolved oxygen sensor, and with a turbidity wiper

    MP TROLL 9000E:

    4 standard 1.5V alkaline D cells

    External power is recommended for long-term monitoring and for

    networks.

    EXTERNAL POWER

    A single stand-alone MP TROLL 9000 can run exclusively on power

    supplied from a 9-12 VDC line power supply connected to a 90-264

    VAC input. When line power is enabled, the MP TROLL shuts down

    the battery regulator, thus preserving the internal or external batteries.

    Both the comm cable and the direct-read cable include external power

    jacks.

    NETWORK POWER SUPPLIES, 12 VDC

    Three regulated DC network power supplies are available:

    North American (120 VAC, 0 to 50 C)

    Universal (95-250 VAC, 0 to 40 C)

    Universal Environmental (100-120, 200-240 VAC, -10 to 60 C)

    Alternatively, a 12 VDC automotive or marine battery may be used.

    Stainless steelBody

    Back EndRemoves for accessto batteries

    Backshell

    Stainless steelRestrictorprovides generousopenings for circulation of environmentalwater, unscrews for installation/removal ofsensors

    AcetalNose Coneminimizes sedimentbuildup and protects the front end frombottoming out on a hard surface;removable for installation of optionalaccessories

    MP TROLL 9000 MP TROLL 9000E

    Quick-Connect SubmersibleCableTIP:Battery life is dependent upon temperature, cablelength, and how often the device is recording measure-

    ments.

    SECTION 2: COMPONENTS & FEATURES

    Use only replacement lithium batteries supplied by

    In-Situ. Use of any other lithium battery will void

    the product warranty.

  • 7/31/2019 TROLL_9000 Manual 2005

    14/171

    7 rev. 010 09/05

    CABLES

    Several basic cable types are used in the MP TROLL 9000 system:

    Submersible cable: Vented: yellow polyethylene or blue FEP*

    Non-vented: black high-density polyethylene (HDPE)

    Stainless steel suspension cable: non-vented, no communica-

    tion capabilities

    Network cable: non-vented non-submersible gray PVC

    Communication cables

    QUICK-CONNECT SUBMERSIBLE CABLE

    Vented submersible cable includes the instrument's backshell and

    provides conductors for power and communication signals, a vent

    tube, strain relief, and a service grip to anchor the instrument securely.

    Vented cable is designed for use with gauged

    pressure/level sensors, atmospheric pressure

    reference for DO sensors, and to measure

    barometric pressure.

    Non-vented submersible cable is designed for

    use with absolute pressure/level sensors.

    Surface Connector

    The Surface Connector remains above the water

    surface and mates with a variety of cables and

    accessories:

    Communication Interface

    Comm cable with AC line power input

    External battery packs

    Network boxes

    SDI-12 adapter

    High-volume desiccant pack

    Vented Cap

    The surface connector cap is vented to allow the

    cables vent tube to carry atmospheric pressure to the back of the

    pressure sensor diaphragm. A hydrophobic membrane (moisture bar-

    rier) of high-density polyethylene (HDPE) under the vent keeps water

    droplets out. A canister of indicating desiccant (replaceable) protects

    the cable and electronics from condensation of water vapor.

    COMMUNICATION CABLES

    ComIT2 Communication Interface

    A null modem cable that connects the MP TROLL's

    quick-connect submersible cable to a desktop or laptop

    PC or to a handheld PDA. Includes a port-powered

    RS485 to RS232 converter. Weatherproof, withstands

    temporary immersion. Cable vents into ComIT body,

    protected by a hydrophobic membrane.

    Computer Connection Cable ("Comm Cable")

    Vented polyurethane cable (1.8 m, 6 ft),

    connects the MP TROLL's submersible cable to

    a PC or PDA. Converts the instrument's RS485

    signal to RS232. Includes a 2.5 mm external

    power jack to connect the instrument to 9-12

    VDC external power.

    Recommended for applications where a

    permanent desktop PC connection is required.

    Direct-Read Cable

    Combines the functions of the quick-connect

    submersible cable and comm cable; connects

    the MP TROLL 9000 directly to a PC or PDA;

    includes RS485/RS232 converter and external

    power jack.

    * FEP = fluorinated ethylene propylene, the generic equivalent of DuPont Teflon.

    Vented cable is recommended for applications whereaccurate barometric pressure measurement is required

    for example when calibrating dissolved oxygen, or formeasuring DO in percent saturation.

    SECTION 2: COMPONENTS & FEATURES

    DB-9 connector

    to SurfaceConnector

    RS485/232Converter

    ExternalPower Input

    DB-9 Connector

    to SurfaceConnector

    Dust Cap

    RS485/232Converter

    ExternalPower Input

    InstrumentBackshell

    Dust Cap

    DB-9 Connector

    InstrumentBackshell

    Dust Cap

    Service

    Grip

    SurfaceConnector

    Vented capwith desiccant

  • 7/31/2019 TROLL_9000 Manual 2005

    15/171

    8 rev. 010 09/05

    VERSIONS AND FEATURE SETS

    The features available in different versions of the Multi-Parameter

    TROLL 9000 are shown in the table above.

    CONTROL SOFTWAREWin-Situ enables communication between the MP TROLL 9000 and a

    desktop or laptop PC. Pocket-Situ is Win-Situ for PDA.

    Both versions of the software provide instrument control for calibra-

    tion, profiling, long-term data logging, direct readings, data extraction,

    data viewing (text and graphical interface), choice of units and other

    custom display options, battery/memory usage tracking, interface to

    networks and telemetry.

    WAKEUP

    The MP TROLL uses a technique we call sleeping to minimize poweconsumption and extend battery life. When the instrument is asleep

    whenever there is no immediate task to performonly its internal

    clock remains active. The MP TROLL will wake when its clock

    indicates it is time to take a data point, or when you are directly

    communicating with it via Win-Situ or Pocket-Situ.

    SECTION 2: COMPONENTS & FEATURES

    Version

    Profes- Profes- Profes- Profes-

    Feature Profiler Profiler XP sional sional XP LTS sional sional XP LTS

    Available WQ sensor ports 4 4 4 4 1 4 4 1

    Sensors

    Dissolved Oxygen, (mg/L, %)

    Conductivity (SC, TDS, Resistivity, Salinity)

    Temperature

    pH

    ORP

    pH/ORP Combination

    Level, Depth, Pressure Barometric Pressure

    Nitrate

    Chloride

    Ammonium

    Turbidity

    Replaceable Sensors

    Data Logging

    Memory, 4MB

    Quick-Connect Cable

    SDI-12 communications

    Upgradeable Calibration

    Quick (single-point)

    Traditional (multi-point)

    Versions and Feature Sets

    MP TROLL 9000 MP TROLL 9000E

  • 7/31/2019 TROLL_9000 Manual 2005

    16/171

    9 rev. 010 09/05

    SMART SENSORS

    The MP TROLLs smart sensors retain manufacturing and calibration

    information for plug-and-play ease and flexibility in the front end option

    ports. Sensor insertion and removal tools simplify insertion and

    removal.

    SYSTEM CONNECTORS

    The lightweight, weather-resistant nylon connectors used throughout

    the Multi-Parameter TROLL 9000 system are designed for straightfor-

    ward connection. Connector keying prevents misconnections. For

    example, submersible cable cannot mate with network cable.

    Each connector is protected by a dust cap. Replacements for worn or

    damaged dust caps are available from In-Situ Inc. or your local

    distributor.

    TRIPLE MOISTURE PROTECTION

    A vented connector cap at the top of the Multi-Parameter TROLL

    9000's submersible cable is fitted with a hydrophobic membrane that

    will not allow water droplets to pass into the cable. This replaceable

    cap contains indicating desiccant to protect the cable and electronics

    from condensation of water vapor. Additional desiccant in the Multi-

    Parameter TROLL 9000 body provides further protection against

    moisture damage, but is not user-replaceable.

    COMPONENTS & ACCESSORIES LIST

    Cables:

    Quick-Connect Submersible Cable with instrument backshell &

    surface connector

    ComIT Communication Interface with RS485/232 converter

    Communication Cable with RS485/232 converter & external power

    input

    Direct Read Cable with instrument backshell, RS485/232 converter

    & external power input

    Stainless steel suspension cable: non-vented, no communication

    Power Accessories:

    External battery packs

    External power supply

    Alkaline batteries

    Lithium batteries

    Water Quality Accessories:

    Basic Sensors

    Polarographic Dissolved Oxygen

    ORPpH

    Conductivity (Low Range or High Range)

    Combination pH/ORP

    Extended Sensor Set

    RDO Optical Dissolved Oxygen

    Ammonia

    Chloride

    Nitrate

    Turbidity

    Calibration Cup: clear PVC vessel that attaches to the front end of

    the MP TROLL 9000 in place of the restrictor and holds the

    recommended amount of calibration fluid.

    Nose Cone Stirrer: battery-powered magnetic stirrer for sample

    agitation during calibration and/or deployment of the MP TROLL

    9000 for best performance in stagnant water.

    Calibration Kits

    Quick Cal (for Basic Sensors)

    pH

    ORP

    Dissolved Oxygen

    Conductivity (Normal range, Low range, High range)

    Nitrate (Normal range, Low range, High range)

    Ammonium (Normal range, Low range, High range)

    Chloride (Normal range, Low range, High range)

    Network Accessories:

    T-Box, Quad Box, Mid-Line Tap Box

    12 VDC Network Power Supplies

    Cables: Network Comm Cable (2 varieties)

    Network Interconnect Cable

    Network Extension Cable

    Installation Accessories:

    Non-Vented Backshell: stainless steel combination backshell cap

    and hanger to suspend the MP TROLL 9000 while taking data. No

    venting, no communication capabilities.

    Well Dock: top-of-well support for 2 or 4 well

    Flow Cells for low-flow sampling

    SECTION 2: COMPONENTS & FEATURES

  • 7/31/2019 TROLL_9000 Manual 2005

    17/171

    10 rev. 010 09/05

    PRODUCT SPECIFICATIONS

    Wetted Materials 316L stainless steel, acetal, Viton,

    nylon, PVC, FEP or poly cable

    Operating Temperature -5C to 50C (23F to 122F)

    Storage Temperature

    Base unit* -40C to 80C (-40F to 176F)

    Basic W.Q. sensors -5C to 50C (23F to 122F)

    Pressure rating (Base unit) 350 psi (246 m, 807 ft)

    Dimensions

    MP TROLL 9000 45 mm (1.79 in) diameter, 47.3 cm (18.6

    in) long

    with RDO sensor 88.4 mm (3.48 in) diameter, 48.3 cm (19

    in) long

    MP TROLL 9000E 45 mm (1.79 in) diameter, 59.5 cm

    (23.45 in) long

    with RDO sensor 88.4 mm (3.48 in) diameter, 61 cm (24in) long

    Weight (without cable)

    MP TROLL 9000 1.9 Kg (4.2 lbs)

    MP TROLL 9000E 2.7 Kg (5.8 lbs)

    Power

    Battery type

    TROLL 9000 2 standard alkaline D-cells (1.5 V), or

    2 lithium D cells (approx. 3.6 V)

    TROLL 9000E 4 standard alkaline D-cells (1.5 V)

    External power input 9-12 VDC (optional)

    Battery life estimates see table below

    Data Sampling

    Memory type/size (memory 4 megabytes flash data storage, about

    -equipped models) 1,000,000 individual readings

    Fastest linear logging rate 5 seconds (10 seconds with RDO)

    Profiling speed 2 seconds

    Communications RS485

    Computer interface RS232

    Software

    Win-Situ 4 Runs on desktop/laptop PC with Micro-

    soft Windows 95 or later, serial com

    port, CD-ROM drive, 16-64 Mb RAM

    Pocket-Situ 4 In-Situ RuggedReader or other supportedhandheld PDA running Microsoft Pocket

    PC, Pocket PC 2002, or Pocket PC 2003,

    also known as Windows Mobile 2003

    Internal Sensors

    Pressure/level

    Type Media-isolated piezoresistive silicon

    strain gauge

    Range 15 psig (0-11 m, 0-35 ft)

    30 psia/psig (0-21 m, 0-69 ft)

    100 psia/psig (0-70 m, 0-231 ft)

    300 psia/psig (0-211 m, 0-692 ft)

    Accuracy better than 0.1% FS

    Resolution 1 mm

    Temperature

    Type Platinum resistance thermometer

    Range -5C to 50C (23F to 122F)

    Accuracy 0.1C

    Resolution 0.01C

    Barometric pressure

    Type Piezoresistive silicon pressure sensor

    Range 0-16.5 psia (854 mm Hg, 33.6 in Hg)

    Accuracy 0.3% FS (2.54 mm Hg, 0.1 in Hg)

    Resolution 0.1 mm Hg, 0.01 in HgTurbidity

    Type Nephelometer, 90 light scattering, 870

    nm LED, solid-state

    Range 0-2000 NTU

    Pressure Rating 150 psi (105 m, 346 ft)

    Accuracy 5% or 2 NTU (whichever is greater)

    Resolution 0.1 NTU

    SECTION 2: COMPONENTS & FEATURES

    Estimated Battery Life

    Instrument At 60 min. At 15 min. At continuoussample intervals sample intervals sample intervals

    MP TROLL 9000 with Turbidity & Wiper 63 Days 52 Days 10 Days

    MP TROLL 9000 no Wiper 75 Days 70 Days 11 DaysMP TROLL 9000 no Turbidity 76 Days 73 Days

    MP TROLL 9000E with Turbidity & Wiper 137 Days 117 Days 24 Days

    MP TROLL 9000E no Wiper 204 Days 193 Days 35 DaysMP TROLL 9000E no Turbidity 206 Days 198 Days

    * Base unit = pressure, temperature, barometric pressure

  • 7/31/2019 TROLL_9000 Manual 2005

    18/171

    11 rev. 010 09/05

    Basic Sensors

    Conductivity

    Type 4-cell conductivity, AC drive

    Operating RangeLow Range sensor 3 S/cm to 50,000 S/cm

    High Range sensor 70 S/cm to 200,000 S/cm

    Accuracy

    Low Range sensor 0.5% or 2 S/cm (whichever is greater)

    between 5 and 20,000 S/cm when

    calibrated in region of interest

    High Range sensor 0.8% + 2 S/cm between 150 and

    112,000 S/cm when calibrated in region

    of interest

    Resolution Range-dependent

    Pressure Rating 350 psi (246 m, 807 ft)

    pH

    Type Glass sensing bulb with Ag/AgCl refer-

    ence half-cell, double junction, reference

    electrolyte gel

    Range 0 to 12 pH units

    Pressure Rating 350 psi (246 m, 807 ft)

    Accuracy 0.09 pH units

    Resolution 0.01 pH unit

    ORP

    Type Platinum wire with Ag/AgCl reference

    half-cell, double junction, reference

    electrolyte gelRange 1400 mV

    Pressure Rating 350 psi (246 m, 807 ft)

    Accuracy 4.0 mV

    Resolution 1 mV

    Dissolved Oxygen (polarographic)

    Type Clark polarographic

    Range 0 to 20 mg/L, 0 to 200% saturation

    Pressure Rating 350 psi (246 m, 807 ft); deployment &

    retrieval at up to 4 ft per second

    Accuracy 0.2 mg/L

    Resolution 0.01 mg/L

    SECTION 2: COMPONENTS & FEATURES

    Extended Sensors

    Ammonium (NH4+)

    Type PVC membrane sensing element, doublejunction Ag/AgCl reference half-cell,

    reference electrolyte gel

    Range 0.14 to 14,000 ppm N (0.1 to 18,000 ppm

    NH4+)

    Pressure Rating 20 psi (14 m, 46 ft)

    Accuracy 10%

    Resolution 0.01 ppm

    Chloride (Cl)

    Type Solid-state sensing electrode, double

    junction Ag/AgCl reference half-cell,

    reference electrolyte gelRange 0.35 to 35,500 ppm Cl (2 to 35,000 ppm

    Cl)

    Pressure Rating 100 psi (70 m, 231 ft)

    Accuracy 15%

    Resolution 0.01 ppm

    Nitrate (NO3)

    Type PVC membrane sensing element, double

    junction Ag/AgCl reference half-cell,

    reference electrolyte gel

    Range 0.14 to 14,000 ppm N (0.4 to 62,000 ppm

    NO3)

    Pressure Rating 20 psi (14 m, 46 ft)Accuracy 10%

    Resolution 0.01 ppm

    Dissolved Oxygen (RDO)

    Type Optical, fluorescence quenching

    Range 0 to 20 mg/L, 0 to 200% saturation

    Pressure Rating exceeds that of MP TROLL 9000

    Accuracy 0.1 mg/L @ 0-10 mg/L

    0.2 mg/L @ 10-20 mg/L

    Resolution 0.01 mg/L

  • 7/31/2019 TROLL_9000 Manual 2005

    19/171

    12 rev. 010 09/05

    Multi-Parameter TROLL 9000

    Operators Manual

    This section provides a quick overview of the initial steps necessary to

    get the Multi-Parameter TROLL 9000 ready to take measurements.

    You will need

    Power Sensors

    Cable

    Software (Win-Situ or Pocket-Situ) and a desktop or

    laptop PC, or a supported PDA

    Calibration Kitone for each sensor to be calibrated, or

    the QuickCal Kit for the Basic Sensor set (pH, ORP,

    polarographic D.O., Conductivity)

    POWER

    The Multi-Parameter TROLL 9000 and TROLL 9000E ship with

    internal alkaline batteriesstandard off-the-shelf 1.5V alkaline D-cells.

    For use with an RDO optical dissolved oxygen sensor, and with aturbidity wiper, the alkaline batteries in the TROLL 9000 (but not the

    TROLL 9000E)may be replaced with In-Situ-supplied 3.6V lithium D

    cells (shipped separately, customer installation required).

    SENSORS

    Basic water-quality sensorspH, ORP, polarographic D.O., Conduc-

    tivitymay arrive installed in the instruments sensor ports. Proceed to

    Cable if there are no additional sensors to be installed.

    Install sensors in the Extended Sensor Set and any sensors shipped

    separately as follows:

    1. Remove the restrictor from the front end of the MP TROLL 9000.

    This allows access to the sensor block depicted in the drawing

    below. There are four sensor ports, P1 through P4, plus a perma-

    nently installed pressure and/or turbidity sensor or plug.

    2. Remove the sensors protective cap or storage bottle and set asidefor future storage of the sensor. If the connector end

    is covered with a cap, remove it also.

    3. Use the sensor removal tool to remove the plug from

    the port where you will install the sensor. Retain the

    plug for use when deploying the instrument with

    fewer than 4 removable sensors installed.

    4. Check lubrication of the sensor o-rings.

    TIP:When unscrewing the restrictor, it is best to grasp the

    body (not the back end) with one hand and the smooth

    area of the restrictor with the other.

    sensorremovaltool

    The sensor o-rings require generous lubrication beforeinstallation. New sensors will be lubricated at the factory.

    If the o-rings appear dry, apply a silicone lubricant beforeinstallation.

    P1

    P2 P3 P4 Conductivity

    1

    2 3 4

    Pressure/Turbidity(or plug)

    Temperature

    D.O. (polarographic)or pHor ORPor ammoniumor chlorideor nitrate

    pH/ORPor Optical D.O.or pHor ORPor ammoniumor chlorideor nitrate

    pHor ORPor ammoniumor chlorideor nitrateor Optical D.O.or turbidity wiper

    End view of sensor block

    P1

    P2 P3 P4

    Alignment mark

    3 GETTING STARTED

    Back end Body Restrictor

  • 7/31/2019 TROLL_9000 Manual 2005

    20/171

    13 rev. 010 09/05

    5. Align the mark on the side of the sensor with the mark on the port.

    6. Press the sensor firmly into the port until you feel it

    dock with the connector at the bottom. When

    properly inserted a small gap (width of the sensor

    removal tool) remains between the instrument body

    and the widest part of the sensor, for ease of

    removal.

    CABLE

    The cable port at the back end of the MP TROLL 9000 accommodates

    different quick-connect cables, depending on the application.

    Quick-Connect Submersible Cabledesigned to be capped off

    and left in the field during deployment. To connect to a PC,

    simply add a communication interface or comm cable.

    Quick-Connect Direct-Read Cableconnects directly to a PC

    for calibrating, profiling, programming, and downloading.

    QUICK-CONNECT CABLE

    1. Remove the soft dust cap from the instrument backshell at the end

    of the quick-connect cable.

    2. Remove the protective plug from the cable port at the back end of

    the MP TROLL 9000 body.

    3. Attach the backshell end of the quick-connect cable to the cable

    port. The threads will

    screw smoothly

    together. Hand-

    tighten.

    PC CONNECTION

    1. Attach a ComIT2 Communication Interface or comm cable to the

    surface connector of the MP TROLL 9000s quick-connectsubmersible cable.

    2. Attach the DB9 connector on the ComIT or comm cable to a PCs

    standard RS232 serial port (a serial cable or card may be needed

    with some PDAs).

    If the sensing element is at the end of the sensor, becareful to handle the sensor by the sides. Use the

    insertion tool to press the sensor into the port at step 6.

    sensorinsertiontool

    After installing a polarographic D.O. sensor, we recom-mend conditioning it for 2-4 hours, preferably overnight,

    before calibrating. For more on D.O. sensor conditioning,see Section 13.

    TIP:Retain the dust cap to protect the backshell's contact

    board from damage when not attached to the instrument.

    Q:

    A:

    Why do I need to connect the MP TROLL 9000 to a PC?

    A PC running Win-Situ or Pocket-Situ control software is aneasy-to-use interface that enables you to

    calibrate the water-quality sensors

    use the MP TROLL 9000 for profiling

    program the instrument for data collection

    set up a pressure channel to display as desired

    monitor battery and memory usage

    download and view the collected data