THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412...

36
THE CATALOG 2019

Transcript of THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412...

Page 1: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

THE CATALOG

2019

Page 2: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

“The role of the infinitely smallin nature is infinitely great.“ - Louis Pasteur -

Page 3: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

...with t .

GRIMM provides you with precision instrumentation for your applications in research, federally regulated monitoring, health & safety studies and many more, you name it.

he products in this catalog

EXPLORE THE WORLD OF AEROSOLS

4

323028

26

22

201816

1210

8

6

Condensation Particle Counters For Nanoparticles

Scanning Mobility Particle Sizer With Condensation Particle Counter

Scanning Mobility Particle Sizer With Faraday Cup Electrometer

Indoor Wide Range Aerosol Spectrometer

Particle Counters For Automotive Emission Measurements

Mini Wide Range Aerosol Spectrometer

Portable Aerosol Spectrometer

Transportable Air Quality Monitoring Station

Environmental Dust Monitor For Automated PM Measurements

Mobile Environmental Ultrafine Particle Counter

Outdoor Wide Range Aerosol Spectrometer

Aerosol Diluters

Emission Sampling System

Aerosol Generators

Aerosol Neutralizers

Accessories

Notes34

14

62

42

table of contents

Page 4: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLESCPC 5410 5412 5416 5420 5421

With the CPC lineup of models 5410 to 5421, GRIMM establishes a new standard for Condensation Particle Counters. The detection head enables single particle counting for concentrations up to 150 000 particles/cm3; moreover, it features improved detection efficiency and response time. These models are optimized for stationary use in any indoor or outdoor application.

All models feature the well-established condensate removal pump and the anti-spill saturator design. In addition, a novel saturator shutter enables transport of the CPC without the need of removing or drying the saturator.

We offer models with or without rugged internal pumps (the pumps fully meet the requirements of continuous long-term measurements), and with or without built-in DMA controller. The photometric mode for high con-centrations (up to 107 p/cm3) is integrated in all models.

The CPCs can be combined with a GRIMM DMA for measurements of particle size distributions (see datasheet for the Scanning Mobility Particle Sizer, SMPS+C). Furthermore, a GRIMM Optical Particle Counter expands the SMPS+C system to a Wide Range Aerosol Spectrometer (WRAS) that measures particle size distributions up to the particle size of 32 µm.

FEATURES

• five models optimized for laboratory andlong-term use

• improved detection limit with D50 = 4.0 nm fortungsten oxide particles

• single particle counting up to 150 000 p/cm3

• tolerant to high ambient temperatures (40°C)• improved response time with t90 < 3 seconds• preconfigured software on a netbook• analog inputs for additional sensors• comprehensive self-tests for high reliability

The CPC line includes also 19“ rack versions. GRIMM offers also mini-containers with additional meteorolo- gical sensors and online data transfer via mobile networks for unattended long-term measurements at remote sites.

APPLICATIONS

• fundamental aerosol research• filter testing• environmental & climatic studies• nanotechnology process monitoring• printer emission studies• inhalation & exposure studies• workplace monitoring

CPC SMPS+C WRAS n - butanol real - time

4

Page 5: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONS

5410 5412 5416 5420 5421

basic CPC standard CPC with sample pump

high-end CPC withSMPS+C capability

high-end 19" CPC with SMPS+C capability

standard 19" CPC with sample pump

max. conc. (p/cm3) single count mode(p/cm3) photometric mode

100 000107

100 000107

150 000107

150 000107

150 000107

sample flow rate [L/min] 0.6 0.6 0.3 0.3 0.3

SMPS option - - yes yes -

internal pump - yes yes yes yes

port for external sensors - - yes yes yes

size (h x w x d) (cm) 23 x 25 x 29 23 x 25 x 29 40 x 25 x 29 19“, 22 x 48 x 41 19“, 22 x 48 x 41

weight (kg) 8.9 8.9 12.4 16.2 16.2

particle detection systemparticle size range 4.0 nm (D50 measured with tungsten oxide particles) to greater than 3 µm;

adjustable to 7.0 nm for compliance with the standard CEN 16967 for determina- tion of the particle number concentration of atmospheric aerosolsconcentration accuracy >95% for single particle counting

rise time t10- t90

air flow systemflow rate sample air

flow rate sheath air flow control

aerosol carrier gas working fluid condensate removal

FUNCTIONCPC control data recording status indication analog input

HANDLINGambient temperature ambient humidity absolute pressure range power supply power consumption

(up to 100 000 p/cm3 at 0.6 L/min, up to 150 000 p/cm3 at 0.3 L/min)10% for optional photometric mode (up to 107 p/cm3 at specified conditions) < 3 s

0.6 L/min for counter models 5410 and 5412 0.3 L/min for counter models 5416, 5420, and 5421 as CPC or SMPS3.0 L/mincritical orifice with stabilized temperature; constant volume flow independent from ambien ct onditionsair and inert gasesn-butanol (n-butyl alcohol)continuous drain with micropump

USB or RS-232directly on PC (GRlMM universal software 5477), optionally on USB stick display and 4 LEDs with 3 colorsport for 3 optional analog climatic or gas sensors, plug and play

10 to 40°C (50 to 104°F)0 to 95% RH, non-condensing500 - 1100 mbar 100 - 240 VAC, 50 - 60 Hz40 W standard operation, 30 W standby, 80 W warm-up

5

Page 6: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

6

SCANNING MOBILITY PARTICLE SIZER WITH CONDENSATION PARTICLE COUNTER SMPS+C 5416 5420 DMA

The GRIMM SMPS+C systems feature the Vienna-type DMA design (Winklmayr et al., 1991; Reischl et al., 1997), well known for highest size resolution and lowest par-ticle diffusional losses – even for the smallest particles.

The SMPS+C systems, based on the GRIMM 5416 and the 19" version 5420 CPC, are operated at an aerosol inlet flow rate of 0.3 L/min and a sheath flow rate of 3 L/min. GRIMM offers a flexible design of the Vienna- type DMA with two electrodes of different length to accommodate a variety of experimental needs.

To expand the size range, the GRIMM SMPS+C can be combined with a GRIMM Optical Particle Counter to build a Wide Range Aerosol Spectrometer (WRAS) that measures particle size distributions up to the size of 32 µm.

FEATURES• particle size distribution from 5 – 1094 nm

• two Vienna-type DMAs

• sample flow rate = 0.3 L/min

• sheath flow rate = 3 L/min

• rugged, compact and reliable

• fully automated use with our software

• analog inputs for additional sensors

• anti-spill CPC saturator design

• comprehensive self-test for highest reliability

APPLICATIONS• fundamental aerosol research

• environmental & climate studies

• nanotechnology process monitoring

• printer emission studies

• inhalation & exposure studies

• studies on atmospheric nucleation

• studies on nanoparticle growth, coagulation &transport

• engine exhaust studies

• mobile aerosol studies

• workplace monitoring

SMPS+CM-DMA

5 - 350 nmL-DMA

10 - 1094 nmAM-241, aDBD

soft X-rayreal - time

Page 7: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSdetector type condensation particle counter (CPC) working fluid n-butanol (n-butyl alcohol)max. concentration single count mode 150 000 p/cm³max. concentration photometric mode 107 p/cm³reproducibility

response time t10- t90 size range size resolution

> 95% for single count mode> 90% or photometric mode< 3 s5 – 350 nm (M-DMA); 10 - 1094 nm (L-DMA)stepping mode: 45 - 255 channelsscanning mode: 64 channels per decade; logarithmicspacing

FUNCTION DMA dimensions output HV module internal pump sample flow rate sheath flow rate port for external sensors

HANDLINGambient temperature ambient humidity absolute pressure range power supply power consumption interfaces dimensions DMA (h x w x d)

weight DMA dimensions CPC (h x w x d) weight CPC

Ri = 13 mm, Ro = 20 mm; L= 88 mm (M) or 350 mm (L)5 – 10 000 V positive polarity; negative polarity on request yes0.3 L/min3 L/minyes

10 – 40°C (50 – 104°F)0 – 95% RH, non-condensing600 - 1100 mbar at full voltage range100 – 240 VAC, 50 – 60 Hz40 W standard operation, 30 W standby, 80 W warm-up USB or RS-232M–DMA: 23.4 x 14 x 15.6 cm (9.2 x 5.5 x 6.1 in) L–DMA: 47.8 x 14 x 15.6 cm (18.8 x 5.5 x 6.1 in)M–DMA: 5.7 kg (12.6 lbs); L–DMA: 7.9 kg (17.3 lbs)40 x 25 x 29 cm (15.7 x 9.8 x 11.4 in)12.4 kg (27.3 lbs)

6 7

Page 8: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

SCANNING MOBILITY PARTICLE SIZER WITH FARADAY CUP ELECTROMETER SMPS+E 5705 5706 DMA

The GRIMM SMPS+E systems feature the Vienna-type DMA design (Winklmayr et al., 1991; Reischl et al., 1997), well known for highest size resolution and lowest par-ticle diffusional losses – even for the smallest particles – with a Faraday Cup Electrometer (FCE) as detector.

The SMPS+E systems include the GRIMM 5706 DMA controller and the GRIMM 5705 fast and low noise FCE and can be operated at aerosol inlet flow rates of 1 - 5 L/min and sheath flow rates of 3 - 20 L/min.

GRIMM offers a flexible design of the Vienna-type DMA with three electrodes of different length to accommo-date a variety of experimental needs. GRIMM’s unique FCE design applies a rinsing air flow around the insu-lator of the Faraday Cup to minimize effects of leakage currents due to internal particle contaminations.

The instrument is optimized to reduce the effects of me-chanical shocks and pressure differences, enabling the SMPS+E as a reference instrument for the calibration of nanoparticle counters.

FEATURES• particle size distribution from 0.8 – 1094 nm

• three Vienna-type DMAs

• sample flow rate = 1 - 5 L/min

• sheath flow rate = 3 - 20 L/min

• rugged, compact, and reliable

• fully automated use with our software

• analog inputs for additional sensors

• data sampling rate up to 16 Hz

• no consumables

• operates with air and inert gases

• comprehensive self-test for highest reliability

APPLICATIONS• fundamental aerosol research

• filter testing

• environmental & climate studies

• nanotechnology process monitoring

• printer emission studies

• inhalation & exposure studies

• studies on atmospheric nucleation

• studies on nanoparticle growth, coagulation &transport

• engine exhaust studies

• mobile aerosol studies

• workplace monitoring

SMPS+E3 DMAsS, M, L

L-DMA10 - 1094 nm

SI traceablereference

16 Hz

8

Page 9: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSdetector type sensitivity noise maximum current maximum particle concentration response time t10– t90 resistor size range

size resolution

FUNCTIONDMA dimensions

output HV module sample flow rate sheath flow rate rinsing air flow rate port for external sensors

HANDLINGambient temperature ambient humidity absolute pressure range power supply power consumption interfaces dimensions DMA (h x w x d)

weight DMA

dimensions FCE (h x w x d) weight FCE dimensions DMA controller (h x w x d) weight DMA controller

Faraday Cup Electrometer (FCE)0.1 fA at 1 Hz0.35 fA ± 4000 fA 108 p/cm³200 ms 1 TΩ 0.8 – 1094 nm (depending on sheath flow rate:0.8 - 53 nm (S–DMA); 5 – 350 nm (M–DMA); 10 – 1094 nm (L–DMA)stepping mode: 45 – 255 channelsscanning mode: 64 channels per decade; logarithmic spacing

Ri = 13 mm, Ro = 20 mm; L = 15 (S) or 88 mm (M) or 350 mm (L) 5 – 10 000 V positive polarity; negative polarity on request 1 – 5 L/min3 – 20 L/min0.6 L/minyes

0 – 40°C (32 – 104°F)0 – 95% RH, non-condensing600 – 1100 mbar100 – 240 VAC, 50 – 60 Hz 80 -130 W USB or RS-232S–DMA: 16.1 x 14 x 15.6 cm (6.3 x 5.5 x 6.1 in) M–DMA: 23.4 x 14 x 15.6 cm (9.2 x 5.5 x 6.1 in) L–DMA: 47.8 x 14 x 15.6 cm (18.8 x 5.5 x 6.1 in)S–DMA: 5.6 kg (12.2 lbs); M–DMA: 5.7 kg (12.6 lbs)L–DMA: 7.9kg (17.3 lbs)19 x 9 x 9 cm (7.5 x 3.5 x 3.5 in)1.36 kg (3.0 lbs)31 x 25.5 x 22 cm (12.2 x 10.0 x 8.7 in)12.2 kg (26.9 lbs)

9

Page 10: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

INDOOR WIDE RANGE AEROSOL SPECTROMETERWRAS SMPS+C & 11-D

The Indoor Wide Range Aerosol Spectrometer combines two technologies for particle counting and classifying: a Scanning Mobility Particle Sizer (SMPS+C) with a butanol condensation particle counter for nanoparticles and the portable optical aerosol spectrometer (11-D) for dust particles.

The system can be used for accurate and high resolution measurements of the entire particle size range from 5 nm to 35 µm in 31 size channels for 11-D and user selectable number of channels forSMPS+C (e. g. 64 per decade).

The system is easy to operate and suitable for all kind of aerosol research.

FEATURES

• real-time monitoring of the entire particle sizerange

• high precision with CPC and OPC at low and high

concentrations

• excellent counting statistics and reproducibility

• low diffusion losses

• self-test of all optical and pneumatic components

for high quality standards

• instrument parameters secured against data loss

APPLICATIONS

• monitoring of ultrafine particles and dust

• aerosol science

• workplace monitoring

SMPS+C 11-D 5 nm - 35 µm5478software real - time

10

Page 11: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSSMPS+Cmeasurement principle

particle size range minimum scan time max concentration single count mode max concentration photometric mode

electrostatic classification with subsequent detection by condensational growth selectable M–DMA (5 – 350 nm) or L–DMA (10 – 1094 nm) 150 s150 000 p/cm3 107 p/cm3

reproducibility > 95% for single particle count modeworking fluid n-butanol (n-butyl alcohol)

optical aerosol spectrometermeasurement principle light scattering at single particles;

detection volume aerodynamically focused, no border zone errorparticle size range 0.253 μm - 35.15 μm concentration range 1 – 3 000 000 p/Lreproducibility > 97% for single particle count mode

FUNCTIONsample air flow rate 0.3 L/min CPC, flow control with critical orifice, temperature

stabilized1.2 L/min aerosol spectrometer, ± 3% constant due to self-regulation

HANDLINGoperation Wide Range Aerosol Spectrometer Software 5473 for online data

presentationinterfaces RS-232, USB, Ethernet, Bluetooth power supply SMPS+C: 85 – 264 VAC, 47 – 440 Hz,

optical aerosol spectrometer: 100 – 240 VAC, 47 – 60 Hz, out: 13 VDC

temperature range 10 to +35°C (50 to 95°F), RH < 95%pressure range SMPS+C: 600 – 1100 mbar

optical aerosol spectrometer: 530 - 1100 mbar;flow rate automatically adjustable to pressure

11

Page 12: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

PARTICLE COUNTERS FOR AUTOMOTIVE EMISSION MEASUREMENTS PMP-CPC 5430 5431

The GRIMM 5430 and 5431 are high accuracy statio- nary particle counters for automotive emission mea-surements with improved response time (t10– t90< 3 s) and an aerosol inlet flow rate of 0.6 L/min.

The instruments comply with the GRPE particle mea-surement program (PMP) for Euro 5 and 6 regulation 83. The new detection head enables single particlecounting up to 50 000 p/cm³ with a coincidence cor-rection <10% for up to 23 000 particles/cm³. The mo- del 5431 is equipped with an internal sample pump.

As all GRIMM CPC models, the 5430 and 5431 feature the well-established condensate removal and anti-spill saturator design with an integrated saturator shutter, enabling the transport of the CPC without removing or drying the saturator.

FEATURES• compact and rugged counter for automotive emis-

sion measurements• aerosol flow rate = 0.6 L/min (with external pump)• single particle counting up to 50 000 p/cm³• coincidence correction <10% up to 23 000 p/cm³• linearity R² ≥ 0.98 for 1 – 23 000 p/cm³• improved response time with t10 – t90 < 3 s• counting efficiency within 50 ± 12% at 23 nm and

≥ 90% at 41 nm• anti-spill saturator design• operates with air and inert gases• continuous removal of condensate• comprehensive self-test for highest reliability

APPLICATIONS• engine emissions for Euro 5 and 6 certifications• testing of Diesel particulate filter• particle counter in dilution systems• high-resolution concentration time series in

driving cycles

CPC 23 nm PMP EURO 5 and 6 real - time

12

Page 13: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSworking fluid concentra rtion ange coincidenc core rection concentration linearity concentra action curacy response time t10 – t90

counting efficiency at 23 nm counting efficiency at 41 nm false background counts readability

FUNCTION internal pump sample flow rate flow control

HANDLINGambient temperature ambient humidity absolute pressure range power supply pow cer onsumption interfaces dimensions (h x w x d) weight

n-butanol (n-butyl alcohol)0 – 23 000 p/cm³ single particle counting<10%R² ≥ 0.98≥ 90%< 3 s50 ± 12%, independent of altitude above sea level≥ 90%, independent of altitude above sea level< 0.001 p/cm³0.1 p/cm³

available in model 54310.6 L/mintemperature stabilized critical orifice

10 – 40°C (50 – 104°F)0 – 95% RH, non-condensing500 – 1100 mbar100 – 240 VAC, 50 – 60 Hz 40 W standard operation, 30 W standby, 80 W warm-up RS-232, USB 22.6 x 25.4 x 22.3 cm (8.9 x 10.0 x 8.8 in)8.9 kg (19.6 lbs)

13

Page 14: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

MINI WIDE RANGE AEROSOL SPECTROMETER MiniWRAS 1371

The compact Wide Range Aerosol Spectrometer (MiniWRAS) is the only portable instrument on the market that allows simultaneous and real-time monitoring of both dust and nanoparticles.

Designed and specifically built for indoor air quality monitoring, the MiniWRAS is a fit for purpose, state-of-the-art system that combines optical and electrical par-ticle detection in one instrument.

The MiniWRAS features the measurement of an ultra-wide particle size range from 10 nm - 35 µm in 41 high resolution particle size channels and the simultaneous measurement of PM10, PM2.5 and PM1 with remote instrument control and wireless data transmission. This portable and ready-to-use instru- ment can be flexibly deployed for various IAQ monitoring projects.

FEATURES

• ultra-wide size range from 10 nm to 35 µm• PM10, PM2.5, PM1 and particle size distribution, par-

ticle surface, and dust mass• high precision over 41 equidistant channels• no consumables• non-radioactive particle charger• versatile data aquisition and communication

interfaces (Bluetooth, USB, RS-232)• easy to use with GRIMM software• optional sensor for temperature and relative

humidity• self-test of all optical and pneumatic components

for high quality standards• rinsing air for protecting laser and detector in

optical cell

APPLICATIONS• nanoparticle and PM monitoring (e. g. PM2.5)

• Indoor Air Quality (IAQ) in buildings

• IAQ in vehicles, airplane cabins, cockpits, busses,

trains

• nanoparticle source identification

• workplace monitoring

• R&D testing in industry

NANOPM10 PM2.5

PM1

10 nm - 35 µm IAQ real - time

14

Page 15: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSmeasured parameters dust fractions acc. to EN 481 (inhalable, thoracic, respirable)

PM10, PM2.5, PM1, number concentration and size distribution

dust mass 0 – 100 000 µg/m3 particle size range 10 nm – 35 µm (10 – 193 nm electrical, 0.253 – 35 µm optical)size channels 41 (10 electrical and 31 optical)particle number 3 000 – 500 000 p/cm3 (electrical)

0 – 3 000 000 p/L (optical)reproducibility > 97% of total measuring range (optical)

FUNCTIONdetection principle optical

optical cell detector

time resolution

detection principle electrical detector sensitivity time resolution

volume flow internal rinsing air flow rate

HANDLINGoperation interfaces analog input power supply battery dimensions (h x w x d) weight operating conditions

light scattering at single particles;detection volume aerodynamically focused, no border zone error diode laser 660 nmfast signal processing with 2 µs pulse length, 2 x 16 raw data channels 6 s, 31 channels (storage interval 1 min)

electrical mobility spectrometer with Faraday Cup Electrometer 0.25 fA 60 s, 10 channels 6 s each (storage interval 1 min)

1.2 L/min, ± 3% constant due to self-regulation0.4 L/min, protects laser optics, reference air for self-test

GRIMM MiniWRAS software (wireless or data cable)Bluetooth, USB RS, -232 external sensor for temperature and relative humidityin: 100 – 240 VAC, 47 – 63 Hz, out: 18 VDC, 2.5 ALi-ion battery, 14.4 VDC, 4.8 Ah for 8 h operation34 x 31 x 12 cm (13.4 x 12.2 x 4.7 in)7.6 kg (16.8 lbs)+4 to +40°C (39 - 104°F), RH < 95%, non-condensing

15

Page 16: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

PORTABLE AEROSOL SPECTROMETER DUST DECODER 11-D

The model 11-D, in its compact and rugged design, combines all advantages of the previous portable GRIMM aerosol spectrometers with the improved optical detection, long-term battery operation, and facilitated handling.

This configuration places the 11-D in the leading position of the portable aerosol spectrometers for monitoring inhalable, thoracic and respirable dust, PM values, and particle number concentration.

The 11-D is the optimal solution for reliable, flexible and real-time measurements for aerosol research and indoor air quality, e.g. at workplaces, interior of vehicles, or for process analysis.

FEATURES• real-time monitoring of particle number, occupa-

tional dust mass fractions, and PM values• additional information on particle number, particle

surface, and dust mass distribution• 31 equidistant size channels, PSL traceable• integrated 47 mm PTFE filter (GRIMM dual

technology)• versatile data aquisition and communication

interfaces (Bluetooth, USB, Ethernet, RS-232)• rinsing air for protecting laser and detector in

optical cell• internal sensor for temperature (T) and relative

humidity (RH) in optical cell• total inlet flow (1.2 L/min) analyzed in optical cell• self-test of all optical and pneumatic components

for high quality standards

APPLICATIONS• aerosol science• PM2.5 in indoor environments according to

VDI 4300, part 11• Indoor Air Quality (IAQ) in buildings and vehicles• process control in industry• workplace monitoring (inhalable, thoracic,

respirable) according to EN 481• monitoring of Permissible Exposure Limit (PEL)

with high time resolution• dust pollution measurements

inhalable thoracic

respirable

TSP PM10 PM4 PM2.5 PM1

PMcoarse

counts & mass 0.25 - 35 µm real - time portable

16

Page 17: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

0 – 100 000 µg/m 3

0.253 – 35.15 µm31, equidistant0 – 3 000 000 p/L , diluter available for higher concentrations

> 97% of total measuring range, according to ISO 21501-1

SPECIFICATIONSmeasured parameters

dust mass particle size range size channelsparticle number

reproducibility

FUNCTIONdetection principle

detector time resolution

volume flow rate

internal rinsing air flow rate gravimetric control

HANDLINGoperation connectivity analog input power supply pow cer consumption battery

operating conditions

storage and transport dimensions (h x w x d) weight

ACCESSORIES

1152 1145A 1158-TRH 1159-10, 1159-100

light scattering at single particles with diode laser;detection volume aerodynamically focused, no border zone error fast signal processing, 2 x 16 raw data channels 6 s, 31 channels (selectable storage intervals)1 s, 16 channels (either 0.253 - 2.982 µm or 2.982 - 35.15 µm) 1.2 L/min, ± 3% constant due to self-regulation, according to ISO 21501-1; automatic altitude correction 0.4 L/min, protects laser optics, reference air for self-test47 mm PTFE filter

keypad or PC with GRIMM software (wireless or data cable) Bluetooth, RS-232, Ethernet1 port (0 – 10 V) for external sensorsin: 100 – 240 VAC, 47 – 60 Hz, out: 13 VDC, 2.5 A5.4 W Li-ion battery, 10.8 V, 6.8 Ah for minimum 10 h operation with desktop smart quick charger+4 to +40°C (39 - 104°F), RH < 95%, non-condensing,non-corrosive or explosive gases-20 to +50°C (-4 – 122°F), RH < 95%27 x 13 x 7 cm (10.5 x 4.8 x 2.6 in)1.9 kg (4.2 lbs)

GRIMM softwar e for 11-DGPS receiverisokinetic sampling probe for 4 - 25 m/scarrying bagexternal sensor for temperature and relative humidity capillary diluter (1:10 or 1:100)

17

TSP, PM10, PM4, PM2.5, PM1, and PM coarse

dust fractions acc. to EN 481 (inhalable, thoracic, respirable),pm10, pm2.5, pm1

number concentration and size distribution

11791146-11-D

Page 18: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TRANSPORTABLE AIR QUALITY MONITORING STATIONEDM 164

The EDM 164, in its compact and mobile weather hous-ing, performs reliable optical detection for counting and classifying dust particles.

The EDM 164 is applicable for short and long-term continuous monitoring of dust pollution and enables a real-time data analysis of dust and meteorological measurement data.

This configuration places the EDM 164 in the leading position of the mobile environmental dust monitoring for such applications as dust mass, PM values, total counts and particle size distribution.

The EDM 164 is the optimal solution for reliable envi-ronmental monitoring, e.g. PM monitoring, source identification, or pollution control at construction and mining sites.

FEATURES• fully automatic monitoring system with remote

access• extremely energy-efficient, low maintenance• no consumables• real-time monitoring of PM values (PM10, PM2.5,

PM1) and particle number• additional information on particle size and mass

distribution• high precision over 31 size channels• versatile data aquisition and communication (data

logger with GSM via internet)• self-test of all optical and pneumatic components

for high quality standards• rinsing air for protecting laser and detector in

optical cell• meteorological sensors• 11 size channels < 1 μm for precise submicron

detection• total inlet flow volume (1.2 L/min) entirely analyzed

in optical cell• excellent counting statistics and reproducibility at

low and high dust concentrations

APPLICATIONS• mobile and simultaneous monitoring of PM10, PM2.5

and PM1

• hot spot monitoring• public site and urban monitoring• source identification• quantification of diffusive emissions according to

VDI 4285 part 3• early warning system for forest fires

PM10PM2.5 PM1 0.25 - 35 µm real - time

18

Page 19: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSmeasured mass fraction PM10, PM2.5, PM1 particle size range 0.25 – 32 µmsize channels 31 particle number 0 – 3 000 000 p/Ldust mass 0 – 100 000 µg/m3

reproducibility > 97% of total measuring rangeoptionally GPS positioning, TSP (Total Suspended Particles), TC (Total Counts),

and particle number for all size channels (size distribution)

FUNCTIONdetection principle light scattering at single particle

detection volume aerodynamically focused, no border zone erroroptical cell diode laser 660 nm detector fast signal processing, 2 x 16 raw data channelstime resolution selectable storage intervals 6 s; 1, 5, 10, 15, 30, 60 minsample flow rate 1.2 L/min, ± 3% constant due to self-regulation rinsing air 0.4 L/min, protection of laser optics, reference air for self-test sampling inlet heated, constant above ambient temperature

HANDLINGoperation keypad, data logger or PC with GRIMM software (wireless or data

cable)interfaces data logger (4 x RS-232, RS-485, Ethernet, μSD, GSM) or RS-232analog input 1 port (0 - 10 V) for auxiliary sensorspower supply in: 110 – 230 VAC, 50 – 60 Hzpower consumption 22 W standard, Imax: 1.4 Atemperature range - 20 to +60°C (-4 – 140°F), RH < 95%, non-condensingabsolute pressure range 900 - 1100 mbardimensions (h x w x d) housing only: 21 x 45 x 44 cm (8.3 x 17.7 x 17.3 in);

with meteorological sensor 157L / 158L / 159L23 x 51 x 73 / 79 / 85 cm (9.1 x 20.0 x 28.7 / 31.1 / 33.5 in)

weight 20 kg (44.1 lbs)

19

Page 20: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

ENVIRONMENTAL DUST MONITOR FOR APPROVED PM MEASUREMENTS EDM 180

The GRIMM EDM 180 is the leading Automated Mea-suring System (AMS) for measuring particulate matter concentration (PM10, PM2.5) in ambient air.

This system offers outstanding features such as simul-taneous PM measurements in 31 particle size channels, 0.1 μg/m3 resolution, and an isothermal inlet with an in-tegrated Nafion dryer. The EDM 180 runs silent, requires low maintenance and can be validated on site using the field test kit together with our system diagnosis soft-ware.

The EDM 180 is the optimal solution for reliable environ-mental monitoring, e. g. automated PM measurements in environmental networks, epidemiological studies, urban and rural PM monitoring. The EDM 180 is in ser-vice of governmental networks and institutes in over 30 countries.

FEATURES• certificates and approvals: EN, US-EPA, UK-MCERTS,

CN-CMA; demonstration of equivalence in over 20countries

• real-time measurement of PM10, PM2.5, PM1, totalcounts (TC), and particle number distribution

• fully automated monitoring system with remoteaccess

• extremely energy-efficient, low maintenance,no consumables

• no loss of semi-volatile compounds• no radioactive source, insensitive to vibrations

(applicable also in vehicles)• versatile data acquisition and communication

(GSM data logger)• self-test of all optical and pneumatic components

for high quality standards• rinsing air for protecting laser and detector in

optical cell• temperature and relative humidity sensors• total inlet flow analyzed in optical cell• excellent counting statistics and reproducibility at

low and high dust concentrations

APPLICATIONS• AMS for PM networks

• PM monitoring

• epidemiological studies

• monitoring of construction and mining sites

EN 12341 PM10

PM10 PM2.5 PM1

US EPAPM2.5

EN 14907 PM2.5

MCERTSPM10 PM2.5

20

Page 21: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSmeasured mass fractions optionally

particle size range size channels particle number

PM10, PM2.5, PM1

TC (Total Counts) and particle number for all size channels (size distribution) 0.25 - 32 μm31 0 - 3 000 000 p/L

reproducibility > 97% of total measuring range

FUNCTIONdetection principle optical light scattering at single particles;

detection volume aerodynamically focused, no border zone error optical cell diode laser 660 nmdetector 2 x 16 raw data channels time resolution selectable storage intervals: 6 s; 1, 5, 10, 15, 30, 60 min sample air flow rate 1.2 L/min, ± 3% constant due to self-regulation internal rinsing air flow rate 0.4 L/min, protection for laser optics, reference air for self-testsampling inlet isothermal humidity extraction via Nafion membrane,

sensor-controlled, without loss of semi-volatile compounds (SVC)

HANDLINGoperation keypad or PC with GRIMM software or Hyper Terminal interfaces RS-232 (GESYTEC) analog input 1 port (0 - 10 V) for auxiliary sensorspower supply in: 230 V/50 Hz; optional 115 V/60 Hzpower consumption 18 W standard, 104 W with Nafion dryer, 116 W maximum,

Imax: 1.4 Atemperature range -20 to +50°C (-4 to 122°F), non-condensingabsolute pressure range 900 - 1100 mbar;

adjustable sample flow rate at high altitudes over 2000 m weather protection housing model 199, stand-alone, fully air-conditioned, providing space

for EDM180 and other 19" rack instruments (see Accessories)dimensions (h x w x d) 26.6 x 48.3 x 36.4 cm (10.5 x 19 x 14.3 in) without sampling

inlet (19" rack, 4 HU, extra 2 HU for rack adapter)weight 18 kg (39.7 lbs) without rack adapter and sampling pipe

21

Page 22: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

MOBILE ENVIRONMENTAL ULTRAFINE PARTICLE COUNTEREDM 465

The EDM 465 combines the reliable technology of our butanol condensation particle counters with easy handling and flexible application for environmen-tal monitoring due to a compact, robust and mobile weather housing.

The EDM 465 is applicable for short and long-term continuous monitoring of ultrafine particles and enables a real-time data analysis of nanoparticles and meteorological measurement data.

This configuration places the EDM 465 in the leading position of the mobile ultrafine particle monitoring. The EDM 465 is a fit for purpose, state-of-the-art system capable of performing accurate and high-resolution measurements.

FEATURES• real-time monitoring of ultrafine particles

according to CEN TS 16976:2016• fully automatic 24/7 monitoring system• low maintenance, 30 days unattended operation,

remote access• energy-efficient sampling with isothermal drying

system• high precision at low and high concentrations• excellent counting statistics and reproducibility• low diffusion losses• versatile data acquisition and communication

(data logger with GSM via internet)• self-test of all optical and pneumatic components

for high quality standards• rinsing air for protecting laser and detector in

optical cell• meteorological sensors• instrument parameters secured against data loss

APPLICATIONS• mobile monitoring of ultrafine particles• traffic emission monitoring• source identification• epidemiological health studies• public site and urban monitoring

CPC CEN/TS16976

24/7 GPS real - time

22

Page 23: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSmeasurement principle condensation particle counter working fluid n-butanol (n-butyl alcohol)particle size range 4 nm to 1 µm (pre-impactor)detection efficiency D50 = 7 nm (verified with silver particles), D90 < 14 nmmax concentration single count mode 150 000 p/cm3

max concentration photometric mode 107 p/cm3

reproducibility > 95% for single particle count moderesponse time trise<5 s, tfall< 5 s

FUNCTIONsampling and conditioning 1 m sampling pipe with sampling head,

isothermal humidity extraction via Nafion membrane, sensor- controlleddiffusion losses < 30% for smallest relevant particle size of 7 nmweather housing stainless steel, powder-coated, thermally isolated, temperature- controlledclimate sensors wind speed and direction, precipitation, pressure, temperature

relative humidity; GPS positioningpumps pulse free carbon vane pumps,

flow rate of sample air 0.3 L/minflow control critical orifice, temperature-stabilizedtotal flow rate 1.5 L/min, ≤ 5% difference to the nominal flow rate

HANDLINGoperation data logger and netbook integrated in housing for online data,

meteorological sensor and GPS positioninterfaces data logger, USB, GSM with SIM card for mobile networkanalog input 1 port (0 – 10 V) for auxiliary sensorspower supply 110 – 220 VAC, 50 – 60 Hzpower consumption 100 - 150 Wtemperature range - 20 to + 40°C (-4 to104°F), RH < 95%,pressure range 500 – 1100 mbardimensions (d x w x h) housing: 49 x 28 x 65 cm (19.3 x 11 x 25.6 in),

total height with sampling pipe and meteorological sensor:140 cm (55.1 in)

weight 38 kg (83 lbs)

23

Page 24: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

WIDE RANGE AEROSOL SPECTROMETEREDM 665

The EDM 665 Environmental Wide Range Aerosol Spec-trometer combines two technologies for particle count-ing and classifying in one device: the Scanning Mobility Particle Sizer (SMPS+C) with the butanol condensation particle counter for nanoparticles and the approved EDM 180 for the larger fractions.

Designed and specifically built for atmospheric mo- nitoring, the EDM 665 is a unique high-tech system for accurate and highly resolved measurements over the entire particle size range from 5 nm to 32 μm in 31 size channels for EDM 180 and user selectable number of channels for SMPS+C (e. g. 64 per decade)

The system requires low maintenance, and can be transported and deployed in the field for short and long-term atmospheric monitoring projects. This con-figuration places the EDM 665 in the leading position of the atmospheric particle monitoring systems.

FEATURES

• real-time monitoring of the entire particle sizerange, fully automatic 24/7 monitoring system

• low maintenance, 30 days unattended operation,remote access

• energy-efficient, sampling with isothermal dryingsystem

• high precision with CPC and OPC at low and high-concentrations

• excellent counting statistics and reproducibility• low diffusion losses• versatile data acquisition and communication

(data logger with GSM via internet)• self-test of all optical and pneumatic components

for high quality standards• meteorological sensors for wind speed and

direction, precipitation, pressure, temperature andrelative humidity

• instrument parameters secured against data loss

APPLICATIONS• atmospheric monitoring of ultrafine particles

and dust• source identification• atmospheric science• traffic emission monitoring

SMPS+C EDM 180 24/7 5 nm - 32 µm real - time

24

Page 25: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

SPECIFICATIONSSMPS+Cmeasurement principle

particle size range minimum scan time max concentration single count mode max concentration photometric mode

electrostatic classification with subsequent detection by condensational growthselectable M–DMA (5 – 350 nm) or L–DMA (10 – 1094 nm) 150 s150 000 p/cm3 107 p/cm3

reproducibility > 95% for single particle count modeworking fluid n-butanol (n-butyl alcohol)

optical aerosol spectrometermeasurement principle light scattering at single particles;

detection volume aerodynamically focused, no border zone errorparticle size range 0.25 μm – 32 μm concentration range 1 to 3 000 000 p/Lreproducibility > 97% of total measuring range

FUNCTIONsampling and conditioning 1 m sampling pipe with sampling head,

isothermal humidity extraction via Nafion membrane, sensor- controlled, without loss of semi-volatile compounds (SVC)weather housing stainless steel, powder-coated, air-conditioned climate sensors wind speed and direction, precipitation, pressure, temperature

relative humidity; GPS positioningtotal flow rate 1.5 L/min, ≤ 5% difference to the nominal flow ratesample air flow rate 0.3 L/min CPC, flow control with critical orifice, temperature

stabilized1.2 L/min aerosol spectrometer, ± 3% constant due to self-regulation

HANDLINGoperation data logger and netbook integrated in housing for online data,

meteorological sensor and GPS positioninterfaces data logger, USB, GSM with SIM card for mobile networkpower supply 230 VAC, 60 Hz power consumption 750 Wtemperature range -20 to +55°C (-4 to 131°F), RH < 95%pressure range SMPS+C: 600 – 1100 mbar,

optical aerosol spectrometer: 900 - 1100 mbar;flow rate adjustable to pressure

dimensions (h x w x d) housing: 107 x 65 x 224 cm (42.1 x 25.6 x 88.2 in)total height with meteorological sensor: 270 cm (106.3 in)

weight 250 kg (551 lbs)

25

Page 26: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

EMISSION SAMPLING SYSTEM ESS 7917

AEROSOL DILUTERS 1159-10 1159-100 7951

CAPILLARY DILUTER (1:10) 1159-10CAPILLARY DILUTER (1:100) 1159-100The 1159 -10 / - 100 is a diluter with a fix dilution ratio for a defined sample flow rate. It can be easily con-nected to any portable GRIMM aerosol spectrometer.

The aerosol flow is split by a precision capillary into a sample and bypass flow. In the bypass flow all particles are removed by the built-in high-efficiency HEPA filter capsule. Downstream, both flows are mixed again. The actual dilution ratio is monitored by measuring the differential pressure over the capillary.

No compressed air is needed.

VARIABLE DILUTER "VKL-MINI" 7951

The VKL-mini is a diluter with two cascadeable, variable dilution modules. The system needs an external com-pressed air supply (8 bar max.). Due to the modular set-up with the connector block and the dilution modules, a wide range of dilution ratios can be covered.

To install and operate the 7951 VKL-mini no additional tools are required. All connections are easy to lock or unlock. This enables an easy installation and fast change of dilution ratios. Due to the operating principle no sampling in under or over pressure conditions is possible. Aerosol in, aerosol out, and exhaust air out have to be operated under the same atmospheric pressure.

No power supply is needed.

EMISSION SAMPLING SYSTEM (ESS) 7917

The Emission Sampling System (ESS) combines a heated sampler and diluter for direct sampling in hot emissions up to 500°C.

The variable two-stage diluter is operated with recircu-lated conditioned air. All process air is generated by the ESS. No compressed air supply is required.

The ESS is suitable for all GRIMM SMPS systems or CPCs.

26

Page 27: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

principle capillary diluterdilution media internal HEPA filtered airsample flow rate 1.2 L/minflow control differential pressure sensor, manually setdilution ratio depending on capillary flow

dilution ratio capillary flow sample flow rate1:10 0.12 L/min 1.2 L/min1:100 0.012 L/min 1.2 L/min

temperature range 0 to +40 °C (32 to 104 °F)pressure range ± 50 mbarpower supply in: 110 – 220 VAC, 50 – 60 Hz, out: 9 – 12 VDC

principle dilution media sample flow rate flow control

inlet nozzles

inject noor zzleexter cnal ompressed air variablecircular orifice, with defined upstream pressure (200 mbar), manually set variable, 1:10 or 1:31.6dilution ratio modules

one (1:10) one (1:31.6)

two (1:10 x 1:10) two (1:10 x 1:31.6)

1:10 1: 31.6 1:100 1: 316 1:1,000 two (1:31.6 x 1:31.6)

temperature range 0 to +40 °C (32 to 104 °F)pressure range Pin = Pout

(for atmospheric pressure only!)

principle injector nozzledilution media internal conditioned airsample flow rate variableflow control heated critical orificesdilution ratio variable with two-stage diluter depending on sample flow rate,

e.g.dilution ratio nozzles sample flow rate1:10 one (1:10) 1 L/min1:100 two (1:10 x 1:10) 1 L/min1: 31 one (1:31) 0.3 L/min1: 961 two (1:31 x 1:31) 0.3 L/min

temperature range up to 500°Cpressure difference ± 100 mbarpower supply 230 VAC, 50 Hz or 115 VAC, 60 Hz

27

Page 28: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

AEROSOL GENERATORS 7811 7860

7811The Grimm 7811 multi-purpose nebulizer is a ready-to-use aerosol generator with two integrated pumps to provide air for atomizing particles and an optional dilution and drying of the raw aerosol.

Both flow rates can be independently controlled, the dilution air flow can be monitored by an integrated flowmeter.

A diffusion dryer column is located at the front of the instrument for easy maintenance and regeneration of the silica gel.

For convenience, up to six nebulizers can be stored inside the removable cover on the instruments backside.

7860The Grimm 7860 WOx generator is a ready-to-use instru-ment to produce well defined aerosol particles in the size range between 1.2 – 20 nm. Its principle of operation is based on the sublimation of tungsten oxide. Heated tungsten reacts in dry, clean air to several oxides and nitrates. A heating cell around a WOx coil provides a temperature of approximately 900°C – the sublimation temperature of WO3. Tungsten oxide sublimates into a controlled fraction of the carrier gas and is immediately diluted when exiting the heated zone by a flow of purified air. (Reischl et al. 1997, Ankilov et al. 2002, Steiner 2006).

By means of three adjustable flows (WOx air, carrier air, dilution air) and a variable heating source, the mean particle diameter and output number concentration of the aerosols can be controlled.

The generator operates with an external compressed air supply (2 – 6 bar). An automatically operating security valve secures the instrument in the case of accidental overpressure.

The unit is designed in accordance with the German VDI Standard 3491.

28

Page 29: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

SPECIFICATIONS

particle generation method nebulizerparticle concentration dependent on material (e.g. DEHS > 107 p/cm³)outlet aerosol flow rate nebulizer: ~ 2.5 - 7.0 L/min; dryer: ~ 7.5 - 17 L/minpower supply 100 – 240 VAC / 50 – 60 Hz / 1.7 Aair supply two integrated pumps (two controllable flows)maximum altitude 2000 m (6500 ft)ambient temperature 0 – 40°C (32 – 104°F)ambient humidity 0 – 90% RH, non-condensingsize (h x w x d) 32.5 x 31 x 28 cm (12.8 x 12.2 x 11 in)weight 11 kg (24.3 lbs)

SPECIFICATIONS

particle size range 1.2 – 20 nmparticle concentration variable up to 107 p/cm³outlet aerosol flow rate ~ 0.03 - 33 L/minmaterial tungsten oxidepower supply 85 – 264 VAC / 47 – 440 Hz

max. 1 A (RMS by 110 VAC)air supply external compressed air (2 – 6 bar; oil and particle free)cleaning system integratedsize (h x w x d) 22 x 17.7 x 27 cm (8.7 x 7 x 10.6 in)weight 7.5 kg (16.5 lbs)

APPLICATIONS

• instrument calibration and testing• filter efficiency testing• inhalation and toxicology studies• multi-purpose test aerosol generation

APPLICATIONS

• fine filter efficiency test• inhalation and toxicology studies• mixing and coating processes (e.g. for ceramic tech- nology)• determination of the detection limits and efficiency of

condensation particle counters

TECHNICAL DATA

FEATURES

• particle size range 1.2 - 20 nm• well-defined aerosol material (WOx)• controllable mean particle diameter• controllable particle number concentration• external compressed air supply (2 – 6 bar)• integrated cleaning system

FEATURES

• generation of aerosols from all kinds of liquids,suspensions and solutions e.g. NaCl, DEHS, PSL

• integrated pumps• no compressed air needed• independently controllable flows (nebulizer flow

and dry air flow)• suitable for small liquid volumes

29

Page 30: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

AEROSOL NEUTRALIZERS5523-Ni 5522-A 5524-X 5520 5520-19“

5523-Ni • Ni-63 radioactive β-source• nominal activity 95 MBq• aerosol flow rate up to 1 L/min• no handling license required1

• low maintenance• easy to operate and safe• compatible with GRIMM SMPS+C and SMPS+E• best suited for ambient aerosol measurements1hazardous goods transportation according to UN 2910; no special measures needed

5522-A• Am-241 radioactive α-source• nominal activity 3.7 MBq• aerosol flow rate up to 5 L/min• low maintenance• easy to operate and safe• compatible with GRIMM SMPS+C and SMPS+E• best suited for various applications

5524-X• non-radioactive source• no transport / storing restrictions• no particle, electromagnetic waves or ozone generat.• low maintenance• easy to operate and safe• embedded controller• compatible with GRIMM SMPS+C and SMPS+E• best suited for various applications

5520, 5520-19“• non-radioactive source• no license needed• low maintenance• easy to operate and safe• available as mobile (5520) and 19“ version (5520-19“)• compatible with GRIMM SMPS+C systems• best suited for mobile applications

30

Page 31: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

TECHNICAL DATA

source type Ni-63; unsealed radioactive source; cleaning not permittedion generation method beta (β) radiation, maximum energy 66 keV, no photonsnominal activity 95 MBq (+0 / -10%)half-life 96 yaerosol flow rate up to 1 L/minaerosol medium air or N2housing stainless steel with additional lead sheathsize (h x w x d) 18 x 6.9 x 6.9 cm (7.1 x 2.7 x 2.7 in)weight 2.5 kg (5.5 lbs)

source type Am-241; sealed radioactive source; covered with gold layerion generation method alpha (α) radiation, maximum energy 5.6 MeVnominal activity 3.7 MBq half-life 433 yaerosol flow rate up to 5 L/minaerosol medium air or N2housing stainless steel size (h x w x d) 4.0 x 5.4 x 5.0 cm (1.8 x 2.1 x 2.0 in)weight 0.6 kg (1.2 lbs)

ion generation method soft X-ray < 9.5 keVtube acceleration voltage 11kVequivalent X-ray dose < 0.13 µSv/h at 10 cm distanceaerosol flow rate 0.3 – 5 L/minmaximum particle concentration 107 p/cm³aerosol medium air or N2power supply in: 100 – 240 VAC, 50/60 Hz , out: 12 VDC, 3.33 Apower consumption 7.2 Wcooling natural cooling with ambient airsize (h x w x d) 19.1 x 7.2 x 27.3 cm (7.5 x 2.8 x 10.8 in)weight 1.1 kg (2.5 lbs)

ion generation method dielectric barrier discharge plasmahigh voltage supply 6 – 8 kV; frequency 20 kHz ± 20%aerosol flow rate 0.3 L/minmaximum particle concentration >104 p/cm³aerosol medium air (not useable with Ar, N2 or corrosive gases)ambient temperature 0 – 40°C (50 – 104°F)ambient humidity 20 – 9% RHambient pressure 700 – 1100 mbarpower supply 100 - 240 VAC, 50/60 Hzpower consumption 15 Wdimensions (h x w x d) 16.0 x 21.0 x 6.5 cm (6.3 x 8.3 x 2.6 in)weight 1.6 kg (3.5 lbs)

31

Page 32: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

ACCESSORIES

WEATHER SENSOR 1158-TRH-0.6 or 1158-TRH-1.5for temperature and relative humidity, with 0.6 or 1.5 meter cable, for operation with our handheld instruments

WEATHER SENSOR 158-TRH Kitfor temperature and relative humidity, with 3.5 meter cable and rain protection housing, for operation with our outdoor instruments

DIGITAL WEATHER SENSOR 157L• for temperature, relative humidity, and barometric pressure• for operation with our outdoor instruments

DIGITAL WEATHER SENSOR 158L • for temperature, relative humidity, barometric pressure, wind speed, and wind

velocity• for operation with our outdoor instruments

DIGITAL WEATHER SENSOR 159L• for temperature, relative humidity, barometric pressure, wind speed, wind velo- city, and precipitation• for operation with our outdoor instruments

RADIAL-SYMMETRICAL SAMPLING HEAD 1111• with a defined inlet slit simulating the average human breathing behavior

47 MM PTFE DISC FILTER 1113A• for our handheld instruments where all optically measured particles are beingdeposited for later analysis (gravimetric, microscopic, or chemical)

ISOKINETIC SAMPLING SET 1152• for applications at higher wind speeds than 2.5 m/s up to 25 m/s and high or lowpressure areas.

DATA LOGGER 164-DL | 180-DL • for our measurement equipment including the capability to transmit all data livevia GSM onto a password-protected platform

FIELD TEST KIT 184 • consisting of a polystyrene latex (PSL) generator nebulizing a 1.0 µm or 2.5 µmPSL suspension to validate the instrument´s performance and some other toolsfor leak checks, Nafion cleaning, and more

WEATHER PROTECTION HOUSING 199• compact and fully air-conditioned, providing space for several 19” rack instru-ments. It measures 65 x 62.5 x 95 cm (25.6 x 24.6 x 37 in) and is due to its weight of125 kg (275 lbs) nearly theft-proof

32

Page 33: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

FLOW SPLITTER 5483• with an integrated mixing chamber assuring equal aerosol concentrations at all outlets• easy to use, install, handle, and clean• simple and effective tool for calibration setups, aerosol research, or comparisons of up to four devices

material stainless steelinlet 1 x hose nozzle for 8 mm inner diameteroutlet 4 x hose nozzle for 6 mm inner diameterflow range Re < 2 300 for up to 3.0 L/mintemperature range 0 – 140 °C (32 – 284 °F)dimensions 9.5 x 2.2 cm (3.7 x 0.9 in)

NANO AEROSOL SAMPLER (PRECIPITATOR) 5561• samples aerosol particles over a wide size range due to electrostatic deflection• particles can be collected for offline analysis on different substrates suitable for a variety of common analyticaltools: transmission & scanning electron microscopes (TEM & SEM) or atomic force microscopes (AFM)

• compatible to all GRIMM DMAs

size range from 0.8 to 1094 nm, user-selectable voltage flow range 1 to 5 L/minsampling efficiency 100% for < 220 nm (for single charged particles)power supply in: 110 – 220 VAC, 50 – 60 Hz, out 12 VDCweight 2.2 kg (4.7 lbs)dimensions (Ø x H) 11.5 x 23 cm (4.5 x 9.1in)

SHEATH AIR DRYER AND ADSORBER 5540• filled with silica gel and active carbon• recommended for continuous measurements with the SMPS+C or +E system in high humidity environments• compatible with GRIMM SMPS systems

sheath air flow rate 3 to 20 L/min weight 2.2 kg (4.7 lbs)dimensions (Ø x H) 15 x 52.5 cm (5.9 x 20.7 in)

SMALL DIFFUSION DRYER 7813 | LARGE DIFFUSION DRYER 7814• filled with silica gel• reduces relative humidity of the aerosol sample• refillable and regenerable silica gel for low maintenace and operational costs• compatible with Grimm SMPS systems

sample flow rate 0.3 to 5 L/minweight small 2.3 kg (5.07 lbs) | large 7.4 kg (16.3 lbs)dimensions (Ø x H) small 29 x 19 cm (11.4 x 7.5 in)

large 50 x 19 cm (19.7 x 7.5 in)

ACCESSORIES

33

Page 34: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

NOTES

34

Page 35: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

NOTES

35

Page 36: THE CATALOG - grimm-aerosol.com · CONDENSATION PARTICLE COUNTERS FOR NANOPARTICLES CPC 5410 5412 5416 5420 5421 With the CPC lineup of models 5410 to 5421, GRIMM establishes a new

GRIMM Aerosol Technik Ainring GmbH & Co. KGDorfstr. 9 | 83404 Ainring, Germany www.GRIMM-aerosol.com