SR-933a A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher M. Gittins, Alan H. Gelb, and B. David Green Physical Sciences Inc. 20 New England Business Center Andover, MA 01810-1077 Applied Optics 38(16), 2594-2604 (2000) Copyright © 2000, Optical Society of America. All rights reserved. Reprinted by permission of the Optical Society of America.

Transcript of SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot...

Page 1: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

SR-933a

A Tunable Fabry-Perot Etalon-BasedLong-Wavelength Infrared Imaging Spectroradiometer

William J. Marinelli, Christopher M. Gittins, Alan H. Gelb, and B. David GreenPhysical Sciences Inc.

20 New England Business CenterAndover, MA 01810-1077

Applied Optics 38(16), 2594-2604 (2000)

Copyright © 2000, Optical Society of America. All rights reserved.

Reprinted by permission of the Optical Society of America.

Page 2: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

1

A Tunable Fabry-Perot Etalon-Based

Long-Wavelength Infrared Imaging Spectroradiometer

William J. Marinelli, Christopher M. Gittins, Alan H. Gelb, and B. David Green

Physical Sciences Inc.

20 New England Business Center

Andover, MA 01810-1077

ABSTRACT

Imaging spectrometry enables passive, stand-off detection and analysis of the chemical

composition of gas plumes and surfaces over wide geographic areas. This paper describes the

use of a long-wavelength infrared imaging spectroradiometer, comprised of a low-order tunable

Fabry-Pérot étalon coupled to an HgCdTe detector array, to perform multispectral detection of

chemical vapor plumes. The tunable Fabry-Perot etalon used in this work provides coverage of

the 9.5 to 14 µm spectral region with a resolution of 7 to 9 cm-1. The etalon-based imaging

system provides the opportunity to image a scene at only those wavelengths needed for chemical

species identification and quantification and thereby minimize the data volume necessary for

selective species detection. We present initial results using a brassboard imaging system for

stand-off detection and quantification of chemical vapor plumes against near ambient

temperature backgrounds. These data shows detections limits of 22 ppmv × m and 0.6 ppmv × m

for DMMP and SF6 respectively for a gas/background ûT of 6 K. The system noise-equivalent

spectral radiance is approximately 2 µW cm-2 sr-1 µm-1. Model calculations are presented

comparing the measured sensitivity of the sensor to the anticipated signal levels for two chemical

release scenarios.

Keywords: remote sensing, hyperspectral, imaging, chemical agent, Fabry-Perot, interferometer

Page 3: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

2

I. INTRODUCTION

Imaging spectrometry enables passive, stand-off detection and analysis of chemical vapor

plumes spread over wide geographic areas. Remote monitoring of chemical vapor releases is a

logical extension of existing visible and near-infrared imaging spectrometer technology;

however, the absorption and emission features of most organic chemical vapors are weak and

poorly structured in the visible and near-infrared. Sensitive, selective detection of chemical

plumes requires spectroscopic measurements in the 8 to 14 µm (long-wavelength infrared,

LWIR) atmospheric transmission window where most organic vapors have their strongest

identifying signatures. Multispectral imaging enables chemical plume visualization by

exploiting the radiant flux difference between the gas plume and the background to reveal

characteristic absorption and emission spectra of target chemicals. Quantitative analysis of

multispectral imagery requires knowledge of both the spectral signatures of the analyte vapors as

well as the effective temperature difference between the vapor plume and the background scene.

The integration of a spectral filter into an IR focal plane array (FPA) based sensor

provides the spectral resolution requisite to identify the spectral signatures of specific chemical

compounds and the radiometric accuracy necessary to obtain quantitatively accurate

measurements of chemical plume column density. Previous efforts to provide imaging capability

for chemical vapors focussed largely on the use of filter wheels in conjunction with forward-

looking IR (FLIR) sensors1-4. In contrast to filter wheel-based imagers, the transmission

wavelength of a tunable etalon is continously variable. Additionally, the number of spectral

resolution elements available to a tunable etalon-based system is defined by the etalon’s finesse,

typically between 25 and 50, rather than by the number of slots available in a filter wheel. In its

Page 4: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

3

simplest configuration, the low-order tunable etalon replaces a filter wheel assembly with a

frequency agile system offering considerable improvement in data acquisition capability.

In contrast to imaging FTIR spectrometers such as the Livermore Imaging Fourier

Transform Infrared Spectrometer (LIFTIRS)5,6 , which covers the 8 to 13 µm atmospheric

transmission window at up to 0.5 cm-1 resolution, the adaptive sampling capability of the tunable

etalon-based imager provides reduced data volume and computational simplicity. In contrast to

dispersive imagers such as the Spectrally Enhanced Broadband Array Spectrograph System

(SEBASS)7, a liquid helium-cooled device which covers the 7.8 to 13.5 µm region at 0.05 µm

resolution (5 cm-1 at 10 µm), the imaging spectroradiometer described in this paper provides

common pixel registry for all wavelengths, ambient temperature operation of the tunable filter,

and higher optical throughput.

II. FABRY-PEROT IMAGING SPECTRORADIOMETER CONCEPT

The LWIR multispectral imaging spectroradiometer system described here is a variation

on a mid-wavelength infrared (MWIR) multispectral imaging spectroradiometer based on the

same tunable etalon technology.8 The optical system concept for the MWIR imager is depicted

in Figure 1: a 128 x 128 pixel InSb IR focal plane array (FPA) views the far field through a

tunable, piezoelectric-actuated Fabry-Pérot interferometer placed in an afocal region of the

optical train. The tunable etalon is in operated in low order (mirror spacing comparable to

wavelength of limited transmitted) and functions as an interference filter which selects the

wavelength viewed by the FPA. This configuration of the interferometer affords both wide a

field-of-view (FOV) and broad wavelength coverage at high spectral resolution.

Page 5: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

4

Optical Configuration

Because no suitable large format FPA (�64 x 64 pixels) was available at the time the

LWIR system was constructed, the optical layout of the brassboard system is somewhat different

than that depicted in Figure 1. The layout of the LWIR system is illustrated in Figure 2. Its four-

key components are: 1) an eight- element (2 x 4 format) photoconductive HgCdTe detector array

housed in a liquid nitrogen-cooled dewar; 2) the computer controlled tunable etalon; 3) a rotating

wheel mechanical chopper used to modulate the far field radiation; and 4) a computer controlled

galvo-driven two-axis scanning mirror system to provide a wide Field-of-Regard. The IR

emission from the far field is sampled with the scan mirrors, brought to an intermediate focus by

an off-axis parabolic reflector (OAP), modulated by the mechanical chopper (2870 Hz),

recollimated by a matched OAP, and directed through the tunable etalon. The radiation

transmitted by the etalon is directed into the detector array through a 10.0 to 11.5 µm

transmitting cold filter chosen to closely match the etalon’s free spectral range.

Each 0.5 mm square detector array element is biased and amplified using an eight-

channel, custom-built amplifier board. The use of the mechanical chopper in conjunction with an

eight-channel lock-in amplifier (LIA) board improves the system signal-to-noise performance by

a factor of ~20 over direct sampling of the DC amplifier output. The system noise equivalent

spectral radiance (NESR) achieved using the LIA is similar to that anticipated using a large

format LWIR FPA to provide imaging capability. Conversion of raw data into radiometrically

calibrated output is accomplished using wavelength dependent gain and offset corrections for

each detector element. The gain and offset coefficients are established from a two temperature

radiance calibration of the system at 25 oC and 40 oC. The measured NESR of the system with

the LIA output time constant set at 10 ms is 2 µW cm-2 sr-1 µm-1 (NEûT� 0.1 K) in the middle of

Page 6: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

5

the spectral tuning range and ~3 µW cm-2 sr-1 µm-1 at the 10.0 and 11.5 µm operating limits of the

system.

A three-dimensional “datacube”, 2D-spatial by 1-D spectral, is produced by positioning

the interferometer to transmit a chosen wavelength, scanning the galvo mirrors through 288

discrete positions to produce a 48 x 48 pixel image of the field, and repeating the process for all

desired wavelengths. Each 48 x 48 pixel spatial element in the datacube comprises a 40 deg by

40 deg FOV (14 mrad/pixel). The tunable etalon is typically scanned from 10.1 µm to 11.4 µm

in increments of 0.1 µm during chemical vapor plume imaging experiments so datacubes consist

of 14 spectral elements. The galvo mirror scan is the rate-limiting step of the data acquisition

and takes ~9 seconds per wavelength to complete. The use of the eight element array in

conjunction with the scanning mirrors is not an ideal arrangement for rapid acquisition of

multispectral imagery. The current brassboard system demonstrates of the utility of the tunable

Fabry-Perot as the basis for a compact, field deployable LWIR multispectral imaging

spectroradiometer incorporating a large format FPA.

Interferometer

The tunable etalon module used in this work is described in detail in a previous

publication.9 A digital capacitance micrometry system is used to measure mirror spacing and

alignment. A closed loop control system is used to position and align the interferometer to the

desired wavelength in times as short as 1.3 ms. The theoretical basis for the development of the

tunable etalon based imaging spectroradiometer, as well as the consequences and advantages of

low-order operation, is evident from the mathematical description of etalon transmission.

Transmitted wavelengths fulfill the resonance condition:

Page 7: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

6

25 × cos� m �0(�)�

× � (1)

I(�)I0(�)

T(�)2

[1R(�)]2× 1 �

2F(�)�

2

sin2 2�5cos��

0(�)1

(2)

where 5 is the mirror spacing, � is the angle of incidence of the light with respect to the optical

axis of the etalon, m is the order of interference, and 0(�) is the phase change upon reflection

from the mirror coating.10

The wavelength dependent filter transmission is described by an Airy function:

where T(�) and R(�) are the wavelength dependent mirror coating transmission and reflectivity

and F(�) is the interferometer finesse. The interferometer is equipped with nominal 94%

reflectivity mirrors spanning the range 9.5 to 14.5 µm. Figure 3 provides an expanded view of

the third order transmission band of the etalon near 10.55 µm. Peak transmission is >60% and

includes slight attenuation by the two anti-reflection coated Ge windows which provide optical

access to the etalon in its hermetically sealed housing. A detailed examination of the lineshape

shows a good fit to the expected Airy function with a FWHM of 6.9 cm-1. The etalon’s spectral

resolution (finesse) is governed not only by mirror reflectivity, but also by surface roughness,

collimation of light incident on the filter, and degree of mirror parallelism.10 For practical

operation in the infrared, the substrate surface roughness is the limiting factor in determining

total finesse. The limiting spectral resolution is seldom greater than û�FSR/40, finesse = 40, due

to this limitation. The observed resolution corresponds to a finesse ~25 and is better than 1% of

� over the entire tuning range of the device.

Page 8: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

7

û�FSR �m �m � 1 (3)

û�FSR

�max

m � 1 (4)

The cos � dependence in Eqs. (1) and (2), see also Figure 1, means that a detector array

with an extremely wide FOV acquires a non-monochromatic image. Detector array elements

view shorter wavelengths the further they lie off axis. The maximum useful system FOV is

defined by the user’s requirement for wavelength uniformity in the image, e.g., a maximum

allowable off-axis wavelength shift of FWHM/10, ~0.7 cm-1, would restrict the full angle FOV to

~9 degrees; if the criterion were FWHM/2, then a 20 degree full angle FOV would be allowable.

The maximum incidence angle associated with off-axis rays in the brassboard LWIR imager is

~1.9 degrees. The use of the scanning galvo mirrors provides a wide field-of-regard while

preserving a monochromatic image.

The free spectral range (FSR) of the etalon is defined as the wavelength spacing between

adjacent transmission orders:

where m is the operating order of the interferometer. For an ideal Fabry-Perot etalon, this

expression simplifies to:

where �max is the longest transmitted wavelength when operating in order m. A wide bandpass

filter, matched to the etalon FSR, is placed in front of the detector array to select a single trans-

mission order. Reflected phase dispersion in the dielectric mirror coating, 0(�) in Eqs. (1) and

(2), contracts the observed FSR relative to that predicted by Eq. (4). The observed FSR of our

LWIR etalon in its typical operating mode (m = 3) is ~2 µm. The effective FSR of the system is

Page 9: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

8

demonstrated in Figure 4 where we show a transmission spectrum of the interferometer recorded

with an open path FTIR spectrometer.

In contrast to an absorbing or scattering filter, e.g., a LWIR AOTF,11 the tunable etalon

has extremely low self-radiance at when operated at ambient temperature. Self-radiance of

components in the optical train not only reduces the effective detector dynamic range but also

increases photon statistical noise, further reducing detection sensitivity. The etalon is reflective

where it is not transmitting and, in the configuration depicted in Figures 1 and 2, detector

emission at wavelengths outside of the filter passband is refocussed on the array. This self-

radiance is insignificant for a liquid nitrogen-cooled detector. The small but finite absorption of

the mirror coating, ~0.5% per mirror surface, is the primary source of system self-radiance.

Mirror absorption causes the etalon to have a finite emissivity. We calculate 0�0.12 at the

transmission wavelength and 0�0.005 outside its passband. Although we have not made precise

measurements of etalon self-radiance, our experience suggests that the system performance is

consistent with our radiometric model for the Fabry-Perot. The LWIR interferometer has an

apparent field temperature of ~200 K when operated at 298 K.

III. CHEMICAL DETECTION EXPERIMENTS

A chemical vapor plume may appear in either emission or absorption depending upon the

temperature and emissivity differences between the plume and the background. We calculate the

plume spectral signature using a simplified three layer radiative transfer model, as shown in

Figure 5. The model has been used previously by Flanigan12 and is an excellent approximation

when the temperature variation within the plume is small in comparison to the temperature

difference between the plume and the background. The total infrared radiance incident upon the

Page 10: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

9

ûN(�) tplume N(�,Tbb) � [1tplume] × N(�,Tplume) N(�,Tchopper) (5)

tplume(�) exp[�ki(�)Ci5] (6)

sensor at a given wavelength (W cm-2 sr-1 µm-1) is the sum of the contributions from each layer.

The brassboard LWIR spectroradiometer measures the difference between the scene radiance and

chopper wheel radiance:

where N(�,T) is the Planck radiance, Tbb is the effective temperature of the background, Tplume is

the plume temperature, Tchopper is the mechanical chopper temperature. The quantity tplume is the

optical transmission of the vapor cloud calculated using Beers’ Law:

where Ci is the average concentration of the chemical compound over the plume length 5 and

ki(�) is its wavelength-dependent absorption coefficient. The sum over index i in Eq. (6) is over

all spectrally relevant chemical species. In Eq. (5) we have made the simplifying assumption that

the atmospheric transmission over our relatively short viewing paths is equal to unity. It is

essential to note that if there is no temperature difference between the background and the plume

then no spectral signature of the chemical plume is observed. The dependence of the minimum

detectable chemical column density, Ci5 in Eq. (6), on temperature difference and molecular

absorption coefficient is discussed below.

Work performed under the auspices of the Army ERDEC has compiled spectral data for a

variety of chemical agents.13,14 We chose dimethyl methyphosphonate (DMMP) as one simulant

in our experimental measurements because of its prior use in this field and because the band

Page 11: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

10

spectrum of DMMP is quite similar to the other G-agents. We have also tested the system using

SF6 because it possess a strong absorption band in the spectral region of interest and has low

toxicity. The chemical imaging system was characterized using both direct absorption and

passive IR emission measurements. Initial experiments were conducted utilizing the system to

view an extended blackbody source through a 10 cm path absorption cell. Gas mixtures could be

produced and flowed through the cell using a manifold of mass flow controllers coupled to vapor

generators as well as pure target and diluent gases. An example of an absorption spectrum taken

with the imaging system is shown in Figure 6 for a mixture of 7 ppmv × m SF6 and 76 ppmv × m

DMMP. The SF6 and DMMP absorption coefficients were obtained from laboratory spectra

recorded at 0.5 cm-1 resolution with an FTIR spectrometer. The SF6 absorption spectra were

recorded in an atmospheric pressure flow cell (10 ppmv SF6 in dry N2). The calculated SF6

absorption coefficient was scaled by ~10% to match the published value of the integrated

absorption coefficent of the �3-fundamental band15. This scaling correction is consistent with the

calibration uncertainty in the SF6 mass flow system. The DMMP absorption spectra were

recorded in a sealed cell containing the saturation vapor pressure of the liquid.

The individual components of the calculated transmission spectrum are shown in

Figure 6. Transmission spectra are simulated by convolving high resolution spectra with the

observed Fabry-Perot resolution function. The data agrees quite well with the calculations,

indicating that the spectral resolution of the system as well as the absorption coefficients of the

target gases are both well understood. It should be noted that the mirror coating used in the

brassboard system was initially developed to detect chlorinated hydrocarbon gases at short range.

Hence, the coating was designed for longer wavelengths than are optimal for the detection of

chemical agents and their simulants, such as DMMP. As a result of this constraint, we are forced

Page 12: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

11

d[ûN]d!

k(�)×[N(�,Tplume)N(�,Tbb)] (7)

/! �/[ûN]

k(�)× 0N0T

×[TplumeTbb](8)

to use a weaker band of DMMP (10.85 µm peak) than is used in most other studies of chemical

agent detection.

Combining Eq. (6) and with the derivative of Eq. (5) with respect to species column

density allows us to calculate detection limits for analyte species from the measured noise

characteristics of the LWIR imaging system. For a single species present in an optically thin

plume, the instrument’s responsivity for that compound is:

where ! is equal to Ci5. The Planck radiance is very nearly a linear function of temperature near

300 K in the LWIR, so the minimum detectable change in column density is well approximated

as inversely proportional to the absorption coefficient of the species and the temperature

difference between the plume and the background:

The quantity /[ûN] is the measured NESR of the system. Using Eq. (8), the calculated detection

limits against a ûT of 6 K is 0.6 ppmv × m for SF6 using the 10.55 µm absorption band and

22 ppmv × m for DMMP using the 10.9 µm absorption band. Use of the stronger DMMP band

at 9.4 µm would improve our detection sensitivity by approximately a factor of three.

Initial imaging experiments involved viewing the absorption cell containing 20 ppmv × m

SF6 held at 298 K against a 308 K reference blackbody. The location of the SF6 cell in the scene

Page 13: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

12

was independently determined by the two-band correlation analysis shown in Figure 7. Figure 7

depicts the measured radiance at 10.55 µm, the peak of the SF6 absorption band, plotted versus

the measured radiance at 10.85 µm, which is outside the SF6 absorption band. This approach

captures differences in apparent emissivity of the scene from pixel to pixel independent of

variations in absolute radiance across the scene. The presence of a structured gas absorption

feature, which is quite different from the more slowly varying background spectrum, is observed

as pixels lying outside the primary vector defined by the majority of the scene pixels. Points

falling above the primary vector would indicate SF6 emission, whereas points lying below the

primary vector indicate absorption. Because the cell is colder than the blackbody background,

only absorption is evident. The pixels corresponding to those points falling >31 below the

primary vector locate the absorption cell in the scene. Figure 8 shows the pixels identified by

two-band correlation analysis highlighted in black. We use the three layer radiative transfer

model to calculate the SF6 column density at the highlighted pixels and find good agreement with

the concentration injected into the cell. Figure 9 depicts calculated column density as a function

of position in the image. The calculated peak column density from the optical measurement,

18 ± 3 ppmv × m, is in excellent agreement with the 20 ppmv × m determined from calibrated

mass flow measurements. Spectral data acquired from these measurements were subsequently

used in spectrally matched filter analysis of SF6 plumes released in an outdoor setting.

We have also tested the system sensitivity against controlled, calibrated plume releases of

DMMP. The DMMP plume releases were performed at the Analytical Chemistry Laboratory at

Argonne National Laboratory (ANL). The laboratory at ANL possesses an apparatus to vaporize

and release low concentration plumes of CW agent simulants.16 The simulant release

concentrations ranged from ~30 to ~220 ppmv × m and a NIST traceable blackbody was placed

Page 14: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

13

in back of the plume at its release point so that it is viewed against a calibrated background. The

plume generating apparatus was contained in a laboratory hood and the LWIR imager was

positioned outside the hood at ~1.5 m standoff distance during data acquisition. In this

configuration the calibrated background occupies only a few pixels in the scene and the imagery

is of little interest.

The spectra of the DMMP releases is analyzed by a least-squares fit of experimental data

to Eq. (5). The background blackbody temperature and plume release temperature were

continuously monitored and controlled during the measurements and therefore could be fixed

during the fit to the data. (The plume release temperature was 307 K and the calibrated back-

ground temperature was 313 K.) The column density of DMMP and the mechanical chopper

wheel temperature were treated as independent variables. The effect of treating chopper wheel

temperature as an independent variable was to provide a modest baseline adjustment. The

chopper wheel temperature was measured periodically between data collections. Derived values

of the chopper temperature differ from measured values by <0.5 K. The 11 uncertainty in the

best derived temperature was ~0.1 K for all spectra analyzed.

Figure 10 depicts the best fit of Eq. (5) to representative spectroradiometer data. The

scene radiance in Figure 10 is plotted in absolute units rather than differential units by adding the

appropriate radiance corresponding to the derived chopper temperature. The data is taken from a

point in the image corresponding to the center of the reference blackbody. The released DMMP

column density is 141 ppmv × m based on liquid flow rate into plume generating apparatus and

the value determined from the optical data is 163 ± 29 ppmv × m. Figure 11 shows the DMMP

column density calculated from spectroradiometer data plotted versus release concentration

calculated from the liquid flow rate for the full range of release concentrations covered in the test

Page 15: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

14

series. If the correlation between the calculated and released column density were perfect, all

points would fall on the dashed line. The error bars shown correspond to ± 11 for a fit of

experimental data between 10.4 and 11.2 µm to Eq. (5). Statistical analysis of the data in

Figure 11 yields a correlation of 1.01 ppmv × m measured per ppmv × m DMMP released (11

uncertainty: ±0.19) and a detection limit of ±22 ppmv × m. The latter figure is in very good

agreement with the detection limit anticipated from the measured system NESR. When the

region of the data spectral comparison is extended to cover the full 10.1 to 11.4 µm spectral

region stored in the data cube, the 11 uncertainties in correlation and detection limits increase,

but neither the slope nor intercept of the data plot changes appreciably. This increase occurs

because the system NESR is greater near the red and blue limits of the operating range than in the

center of the range.

Finally, the system was tested against an outdoor plume release where a known flow rate

of SF6 was diluted in N2 and released from a pipe of known cross section. This arrangement is

depicted in Figure 12. Gas is released from the short section of black pipe slightly below the

center of the scene. The location of the SF6 plume in the scene was identified using both two-

band correlation analysis and a spectrally matched filter (SMF) algorithm17-19 targeted on the SF6

absorption feature at 10.55 µm. The results of the SMF analysis is shown in Figure 13. The

spatial extent of Figure 13 covers approximately the central horizontal two-thirds of Figure 12.

The SMF analysis locates a larger plume than does the two-band correlation analysis and those

pixels are indicted by lighter tones. The SMF analysis appears to yield fewer false alarms, albeit

through the use of data in the entire spectral data cube. The SF6 column density in the plume at

the pixels identified by two-band correlation (>31 below the primary vector as for the absorption

cell image analysis) was calculated using the three-layer radiative transfer model described

Page 16: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

15

previously and is depicted in Figure 14. Unfortunately, the outdoor plume images were recorded

on a windy afternoon and the SF6 plume was rapidly dispersed in the atmosphere. Because of the

rapid dispersion and fact that the opening of the release pipe was less than one pixel wide in the

LWIR image, the effective column density of SF6 at the pipe exit was not well-known. The

building wall, which served as the background, was considerably warmer than ambient

temperature, 313 to 315 K as opposed to 298 K, and the plume is observed most clearly there.

The calculated column density using the radiative transfer model serves as a good cross check for

both the two-band correlation and the SMF analyses. False alarms occurring in each analysis

correspond to column densities very near or below the anticipated detection limits for the species.

The radiative transfer model indicates an SF6 column density of 5 ppmv × m at the apparent ûT

of 15 K. The SNR of the data is approximately 5:1 and is consistent with detection limits

determined using Eq. (8) with the SF6 absorption coefficient appropriate for the system spectral

resolution.

IV. CHEMICAL RELEASE DETECTION

Two scenarios were modeled for comparison with the capabilities of the chemical

imaging system described in this paper: 1) a look down view of a chemical weapons release

from a SCUD-B missile, and 2) observation of dilute emissions from a chemical production

facility stack. In the SCUD-B release scenario, DMMP was used to simulate a volatile G-type

nerve agent such as Sarin. In the stack emissions viewing scenario, DMMP was assumed to

simulate tributyl phosphate, a compound used in the PUREX process for plutonium

reprocessing.20 Two different models were employed to translate chemical releases into infrared

signatures: 1) a Gaussian model of plume dispersion in the atmosphere, and 2) a radiative

Page 17: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

16

c(x,y,z) q

�u1y1z

expy 2

212y

exp(zh)2

212z

� exp(z�h)2

212z

(9)

transfer model to describe transport of radiation from the background which traverses the

chemical plume and the intervening atmosphere to reach the sensor. The radiative transfer model

was discussed earlier in this paper, although we no longer assume the intervening atmosphere to

be transparent. The plume dispersion model is discussed below prior to conducting a sensitivity

assessment.

The dispersion of a chemical species in the atmosphere from the release site is the major

source of uncertainty in modeling the ability of a sensor to locate the chemical cloud later in

time. Diffusion and gross scale motions of the atmosphere will dilute and transport the species

of interest along the prevailing wind direction. In this effort a classical Gaussian dispersion

model was used to estimate the concentration of the chemical of interest downstream of the

release site. The Gaussian model is applicable to a highly idealized set of atmospheric

conditions, but nonetheless is used extensively by the EPA and others as a basis for more

extensive models.21 We have used the model to predict the downstream concentrations for

steady-state (chemical stack) and single puff (chemical agent attack) transient releases. These

expressions are given for steady-state and transient releases below.

Steady-State

where

q = mass release rate

u = local wind velocity

Page 18: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

17

c(x,y,z) S

(2�)3/21x1y1z

exp(xx1u(tt1)2

212x

(yy1)2

212y

x

exp (zz1)2

212z

� exp (z�z1)2

212z

(10)

1y Ryxry (11)

h = release height

x = wind direction

z = distance above ground

y = distance transverse to wind direction and parallel to ground

and

1x, 1y, 1z = dispersion parameters along x, y, z axis.

Transient

where

S = mass released

and

x1, y1, z1 = release points.

The dispersion parameters, 1x, 1y, and 1z, are empirically derived from atmospheric dispersion

measurements under a range of atmospheric conditions. The parameters 1y and 1z are defined for

distances downwind of the release point from the expressions:

and

Page 19: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

18

1z Rzxrz (12)

where the parameters Ry, ry, Rz, and rz are established based on atmospheric conditions referred to

as Pasquill Stability Classes. In the transient emission case 1x = 1y. In our model we have used

values of these parameters defined for a neutral atmosphere and a wind velocity of 10 mph

(4.5 m/s). In the agent dispersion case, the model indicated that the agent cloud grows to

approximately 350 meters in diameter approximately 2 km downstream of the release point,

which is in good agreement with published data on other dispersion models.22

In both modeling scenarios the choice of source strength is critical to the overall

sensitivity estimate. In the chemical agent dispersion case a single release of 100 kg was

assumed to occur. This source strength was thought to be typical of a theater ballistic missile

warhead (SCUD) or of a multiple shell artillery attack. We believe the assumption of a 100 kg

release is conservative. In a chemical delivery mode, the SCUD-B is capable of carrying a

warhead containing 555 kg23 of the persistent agent VX and could carry an equivalent amount of

a G-type agent. Based on data available at the Chemical Weapons Convention web site,24 a

100 kg release would require approximately forty 155 mm howitzer shells, or a similar number of

115 mm rockets. In the stack monitoring case data on anticipated chemical concentrations are

less readily available due to the sensitive nature of the facilities which could be monitored. In

our model, we took an extremely conservative approach to this problem by assuming that the

chemical concentrations at the stack exit were 10 ppmv. From literature available on chemical

process facility design,25 we determined that properly designed exhaust vents have air velocities

which range from 10 to 30 meters/second. This combination of chemical concentration, stack

dimension, and flow velocity sets the mass flow rate of the compound entering the atmosphere.

Page 20: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

19

The detection of emissions from chemical stacks represents the more difficult of the two

detection tasks. Initial calculations, based on a model which assumed emissions occurred from a

chemical hood facility (20 m/s flow at 298K) with a 1 meter diameter stack at a concentration of

10 ppmv × m, showed signal levels well below system detection limits. An example of the

model results are shown in Figures 15 and 16. The column density shown in Figure 15 is quite

small, leading to an optical depth of 1.25 x 10-3 for the strongest band in DMMP. The

corresponding change in scene radiance, shown in Figure 16, reveals a net change in radiance of

only 1.5 x 10-8 W cm-2 sr-1 µm-1. This radiance level is well below the detection limits for any

operational or projected multispectral imaging system known to the authors.

It was determined subsequently that emissions from combustion driven stacks

(Texit�150 oC) were most appropriate for the monitoring scenario. Calculations were performed

to simulate this method of chemical release. The monitoring scenario and a plume model was

described by Barden and Kroutil.26 By increasing the plume temperature, the net radiances from

the scene increased approximately 100 fold. For a system with a spatial resolution comparable to

the plume dimensions, the plume radiance would exceed 4 x 10-4 W cm-2 sr-1 µm, which is well in

excess of the calculated NESR for the systems discussed in this paper. Though the increase in

plume temperature increases the signal dramatically, the increase also has implications for the

spectral signature of the background. This change is primarily due to the presence of

combustion-produced species (primarily H2O and CO2) in the stack effluent which tends to

obscure spectral features in the plume

The results of the chemical agent release model were quite promising in relation to the

anticipated performance of the imaging system. The extent of the chemical cloud is indicated in

Figure 17, where the C × 5 product (column density) is shown for the 45 degree look-down view

Page 21: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

20

at a range of 5 km. The peak column density was approximately 0.5 cm g liter-1 (1000 ppmv ×

m) DMMP. for the modeled release. (Although these calculations were carried out in cm g liter-1,

units common to the stand-off detection community, ppmv × m is in general more useful because

it does not require knowledge of analyte’s molecular weight.) The DMMP absorption cross-

section at 9.4 µm results in a plume optical depth of approximately 3, implying that it attenuates

95% of the radiation originating from the far field, which in this case is the ground below. The

radiative transfer model utilized the MODTRAN mid-latitude summer atmospheric profile27 to

determine atmospheric transmission from the release point to the sensor. From this model the air

temperature is given as 294 K, the effective blackbody temperature of the ground is 290 K, and

the chemical plume appears in emission. The data presented in Figures 18 and 19 show total and

differential radiance for the DMMP feature at 9.4 µm, the strongest feature in the spectrum. The

primary difference between these data and a similar calculation performed for the weaker

10.85 µm DMMP feature used in the experimental measurements is the flattening of the spatial

radiance distribution near the region of maximum radiance due to the plume being optically

thick. The differential radiance reaches a maximum of 2.5 x 10-5 W cm-2 sr-1 µm-1 in this viewing

scenario, approximately a factor of 12 greater than the NESR of the brassboard imager.

V. CONCLUSIONS AND FUTURE WORK

The objective of this effort was the development of a LWIR hyperspectral imaging system

specifically designed for imaging of gaseous chemical species in air. We have demonstrated that

the tunable Fabry-Perot etalon provides sufficient spectral resolution, tuning range, and tuning

rate to detect and identify vapor plumes of the two tested hazardous vapor simulants, SF6 and

DMMP. The current mirror set and capacitance micrometry system may be improved to extend

Page 22: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

21

the spectral tuning range and reduce the tuning time between spectral resolution elements to

nearer the limit imposed by the mechanical tuning limit of the mirror actuators.

The NESR of the LWIR spectroradiometer is sufficient to allow detection of optical thin

chemical vapor plumes against near ambient temperature backgrounds. Our experimentally

determined detection limit of ±22 ppmv × m for DMMP against a temperature drop of 6 K is in

excellent agreement with the value predicted using a simple radiative transfer model and the

measured noise characteristics of the system detection electronics. In future efforts we will be

replacing the scanned 2 x 4 element HgCdTe array with a liquid nitrogen-cooled 64 x 64 pixel

format, 61 µm square pixel staring array. This improvement will increase spatial resolution by

approximately a factor of 10. Furthermore, system models indicate that the sensor NESR can be

improved to approximately 0.5 µW cm-2 sr-1 µm-1. For comparison, the LIFTIRS and SEBASS

imagers report NESR values ~1 µW cm-2 sr-1 µm-1 using liquid helium-cooled large format FPAs

(Refs. 6 and 7).

The brassboard Fabry-Perot imaging spectroradiometer has provided hyperspectral

imagery of sufficient quality to allow quantitative analysis of chemical vapor plume releases.

Data is currently processed off-line, however we continue to upgrade system hardware and data

processing algorithms with the goal of providing a stand-off imaging sensor with real-time

analysis of the scene chemical composition.

ACKNOWLEDGMENTS

Recent development efforts on the LWIR imaging spectroradiometer were funded by the

U.S. Army Night Vision and Electronic Sensors Directorate under Phase I Small Business

Innovation Research contract No. DAAB07-97-C-G006 and by the U.S. Army Edgewood

Page 23: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

22

Research, Development, and Engineering Center (ERDEC) under contracts No. DAAD05-97-P-

2618 and No. DAAD05-98-P-0625. The authors thank Dr. Jack Demirgian for conducting

controlled plume releases at ANL, Dr. James Jensen and Dr. Alan Samuels of U.S. Army

ERDEC for helpful discussions, Will Lawrence for assisting with the outdoor SF6 plume imaging

experiments, and George Dippel for development of the system electronics.

REFERENCES

1. H. Walter, Jr. and D. Flanigan, “Detection of atmospheric pollutants: a correlation

technique,” Appl. Opt. 14, 1423-1428 (1975).

2. M.L.G. Althouse, “Spectral filtering of thermal imagers for gas detection,” in Proc. 3rd

Int. Symp. Protection Against Chemical Agents, p. 143-148, Umea, Sweden (1989).

3. L.G. Carr, L. Fletcher, P.L. Holland, J. Leonelli, D. McPherrin, and M.L. Althouse,

“Characterization of filtered FLIR systems designed for chemical vapor detection and

mapping,” in Infrared Imaging Systems: Design, Analysis, Modeling, and Testing, G.C.

Holst, Ed., Proc. SPIE Vol. 1309, p. 90-103 (1990).

4. M.L.G. Althouse and C.I. Chang, “Chemical vapor detection with a multispectral thermal

imager,” Opt. Eng. 30, 1725-1733 (1991).

5. C. L. Bennett, M. R. Carter, and D. J. Fields, "Infrared hyperspectral imaging results from

vapor plume experiments," in Imaging Spectrometry, M.R. Descour, J.M. Mooney, D.L.

Perry, and L.R. Illing, Eds., Proc. of SPIE Vol. 2480, p. 435-444 (1995).

6. C. L. Bennett, M. R. Carter, and D. J. Fields, "Hyperspectral imaging in the infrared using

LIFTIRS," in Infrared Technology XXI, B.F. Andresen and M.S. Scholl, Eds., Proc. of

SPIE Vol. 2552, p. 274-283 (1995).

Page 24: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

23

7. J.A. Hackwell, D.W. Warren, R.P. Bongiovi, S.J. Hansel, T.L. Hayhurst, D.J.Marby,

M.G. Sivjee, and J.W. Skinner, "LWIR/MWIR imaging hyperspectral sensor for airborne

and ground-based remote sensing," in Imaging Spectrometry II, M.R. Descour and J.M.

Mooney, Eds., Proc. of SPIE Vol. 2819, p. 102-107 (1996).

8. C.M. Gittins, W.J. Marinelli, and A.J. Ratkowski, “AIRIS Hyperspectral Imaging

Technology,” presented at the 6th Annual AIAA/BMDO Technology Readiness

Symposium, Report No. N97-31890, American Institute of Aeronautics and Astronautics,

Reston, VA, 1997.

9. C.M. Gittins, W.G. Lawrence, and W.J. Marinelli, “Frequency Agile Bandpass Filter for

Direct Detection Lidar Receivers,” Appl. Opt. 37, 0000-0000 (1998).

10. P.D. Atherton, N.K. Reay, J. Ring, and T.R. Hicks, “Tunable Fabry-Perot Filters,” Opt.

Eng. 20, p. 806-814 (1981)

11. D.R. Suhre and E. Villa, “Imaging Spectroradiometer for the 8-12 µm region with a

3 cm-1 Passband Acousto-Optic Tunable Filter,” Appl. Opt. 37, 2340-2345 (1998).

12. D.F. Flanigan, “Prediction of the limits of detection of hazardous vapors by passive

infrared with the use of MODTRAN,” Appl. Opt. 35, 6090-6098 (1996).

13. D.F. Flanigan, “The Spectral Signatures of Chemical Agent Vapors and Aerosols,”

CRDC-TR-85002, US Army Armament, Munitions and Chemical Command, Aberdeen

Proving Ground, MD, April 1985 (unclassified).

14. L.D. Hoffland, R.J. Piffath, and J.B. Bouck, “Spectral signatures of chemical agents and

simulants,” Opt. Eng. 24, 982-984, 1985.

Page 25: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

24

15. P. Varanasi, Z. Li, V. Nemtchinov, and A. Cherukuri, “Spectral Absorption-Coefficient

Data on HCFC-22 and SF6 for Remote-Sensing Applications,” J. Quant. Spectrosc.

Radiat. Transfer 52, 323-332 (1994).

16. J. Demirgian, S. Macha, and S.Darby, “The Challenge: Collection of Quantitative

Chemical Release Data,” in Proc. Third Workshop on Stand-Off Detection for Chemical

and Biological Defense, Science and Technology Corp., Hampton VA, p. 299-306 (1994).

17. R.N. Clark, A.J. Gallagher, and G.A. Swayze, “Material absorption band depth mapping

of imaging spectrometer data using the complete band shape least-squares algorithm with

library reference spectra,” in Proc. Third Airborne Visible/Infrared Imaging Spectrometer

(AVIRIS) Workshop, JPL Publication 90-54, p.176-186, Jet Propulsion Laboratory,

Pasadena CA, 1990.

18. R.N. Clark, G.A. Swayze, A. Gallagher, N. Gorelick, and F.A. Kruse, “Mapping with

imaging spectrometer data using the complete band shape least-squares algorithm

simulaneously fit to multiple spectral features from multiple materials,” in Proc. Third

Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) Workshop, JPL Publication 91-

28, p.2-3, Jet Propulsion Laboratory, Pasadena CA, 1991.

19. R.N. Clark, G.A. Swayze, and A. Gallagher, “Mapping the minerology and lithology of

Canyonlands, Utah with imaging spectrometer data and the multiple materials,” in

Summaries of the Third Annual JPL Airborne Geoscience Workshop, JPL Publication

92-14, v. 1, p. 11-12, Jet Propulsion Laboratory, Pasadena CA, 1992.

20. U.S. Department of Energy, Richland Operations Web Site,

http://www.hanford.gov/twrs/char.pub/docs/tcr/t107/t107s2.htm

Page 26: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

25

21. Atmospheric Chemistry and Physics of Air Pollution, J.H. Seinfeld, John Wiley & Sons,

New York, 1986.

22. E.E. Uthe, N.B. Nielson, and R.D. Kaiser, “Airborne LIDAR and Radiometric Detection

and Analysis of Effluent Plumes,”in Proc. Third Workshop on Stand-Off Detection for

Chemical and Biological Defense, Science and Technology Corp., Hampton VA, p.211-

212 (1994).

23. Fact sheet, Chemical and Biological Defense Information Center,

http://www.cbiac.apgea.army.mil/faq.htm.

24. J.E. O’Pray, “Regional Power Ballistic Missiles: An Emerging Threat to Deployed US

Forces?” Analytical study, Air War College, Air University, Maxwell Air Force Base,

AL, May 1990.

25. Design, Construction and Refurbishment of Laboratories, R. Lees and A.F. Smith, Eds.,

Ellis Horwood Limited, Chichester, England, 1984.

26. J.D. Barden and R. Kroutil, “Development and Preliminary Field Evaluation of a Field-

of-View Near Real-Time 3-D Stack Plume Model Developed for the Measurement

Attributes of Remote Optical Sensors,” in Proc. Third Workshop on Stand-Off Detection

for Chemical and Biological Defense, Science and Technology Corp., Hampton VA, p.

457-467 (1994).

27. MODTRAN2 (Ontar Corp., North Andover MA, 1992).

Page 27: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

26

FIGURE CAPTIONS

Figure 1. Basic configuration for the tunable Fabry-Perot etalon based imaging

spectroradiometer. FPA, focal plane array.

Figure 2. LWIR imager optical configuration employing the off-axis parabolic reflectors

(OAPs) as focussing elements

Figure 3. Expanded view of LWIR tunable filter transmission spectrum.

Figure 4. Representative Transmission Spectrum of the LWIR tunable filter module

Figure 5. Schematic diagram of three-layer model for quantitative analysis of scene radiance

levels and plume concentration determination.

Figure 6. Observed and calculated transmission through an absorption cell containing

8 ppmv × m SF6 and 80 ppmv × m DMMP in dry N2. The reference blackbody is

at 308 K and the gas temperature is 298 K. The data represent a six scan co-

average.

Figure 7. Two-band correlation plot of narrowband scene images at 10.55 and 10.85 µm.

See text for details.

Figure 8. IR image of absorption cell at 10.55 µm with the location of SF6 absorption

identified by two-band correlation analysis highlighted in black.

Figure 9. Calculated SF6 column density in the absorption cell containing scene as

determined using the three-layer radiative transfer model.

Figure 10. Best fit LWIR imager data to Eq. (7) for a 141 ppmv × m DMMP plume release.

Figure 11. Measured versus calculated DMMP plume column densities.

Figure 12. Visible image of the outdoor plume imaging configuration. Gas is released from

the black pipe slightly below the center of the scene.

Page 28: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

27

Figure 13. SF6 plume from outdoor release experiment as identified using the spectrally

matched filter algorithm. Light pixels indicate a high degree of correlation with

the SF6 absorption spectrum.

Figure 14. SF6 column density in the outdoor plume release calculated using LWIR imager

data and the three-layer radiative transfer model.

Figure 15. Calculated column density (concentration × path length) for stack emission over

the filed of view of the system.

Figure 16. Estimated differential radiance for chemical species in the stack plume for a non-

combustion scenario.

Figure 17. Column density of agent for a simulated release from a SCUD-B.

Figure 18. Total radiance at 9.4 µm for a simulated chemical agent release using DMMP.

Figure 19. Net radiance at 9.4 µm for a simulated chemical agent release. Flattening of the

profile at the highest radiance levels indicates that the cloud is optically thick in

the center.

Page 29: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 1Marinelli

Applied Optics28

λ

R

θ

F-P

Lens

FPA

IncidenceAngle

MirrorSpacing D-3994z

Figure 1. Basic configuration for the tunable Fabry-Perot etalon-based

imaging spectroradiometer. FPA, focal plane array.

Page 30: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 2Marinelli

Applied Optics29

OAP

Detector

Interferometer9.5-11.5 µm

Chopper3 kHz

Horizontal

OAP

OAP

Vertical

Galvo-Scanner

± 20 degHorizontaland Vertical

D-5263

Figure 2. LWIR imager optical configuration employing the off-axis parabolic

reflectors (OAPs) as focussing elements

Page 31: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 3Marinelli

Applied Optics30

Figure 3. Expanded view of LWIR tunable filter transmission spectrum.

Page 32: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 4Marinelli

Applied Optics31

Figure 4. Representative Transmission Spectrum of the LWIR tunable filter module

Page 33: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 5Marinelli

Applied Optics32

Layer 3

Atmosphere

Layer 2

Sensor Plume

Layer 1

D-7659

BlackbodyBackground

Figure 5. Schematic diagram of three-layer model for quantitative analysis of scene

radiance levels and plume concentration determination.

Page 34: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 6Marinelli

Applied Optics33

Figure 6. Observed and calculated transmission through an absorption cell

containing 8 ppmv × m SF6 and 80 ppmv × m DMMP in dry N2. The

reference blackbody is at 308 K and the gas temperature is 298 K. The data

represent a six scan co-average.

Page 35: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 7Marinelli

Applied Optics34

Figure 7. Two-band correlation plot of narrowband scene images at

10.55 and 10.85 µm. See texts for details.

Page 36: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 8Marinelli

Applied Optics35

Figure 8. IR image of absorption cell at 10.55 µm with the location of SF6 absorption identified

by two-band correlation analysis highlighted in black.

Page 37: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 9Marinelli

Applied Optics36

Figure 9. Calculated SF6 column density in the absorption cell containing scene as

determined using the three-layer radiative transfer model.

Page 38: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 10Marinelli

Applied Optics37

Figure 10. Best fit LWIR imager data to Eq. (7) for a 141 ppmv × m

DMMP plume release.

Page 39: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 11Marinelli

Applied Optics38

Figure 11. Measured versus calculated DMMP

plume column densities.

Page 40: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 12Marinelli

Applied Optics39

Figure 12. Visible image of the outdoor plume imaging configuration. Gas is released

from the black pipe slightly below the center of the scene.

Page 41: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 13Marinelli

Applied Optics40

Figure 13. SF6 plume from outdoor release experiment as identified using

the spectrally matched filter algorithm. Light pixels indicate a high degree

of correlation with the SF6 absorption spectrum.

Page 42: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 14Marinelli

Applied Optics41

Figure 14. SF6 column density in the outdoor plume release calculated

using LWIR imager data and the three layer radiative transfer model.

Page 43: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 15Marinelli

Applied Optics42

Figure 15. Calculated column density (concentration × path length) for stack

emission over the field of view of the system.

Page 44: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 16Marinelli

Applied Optics43

Figure 16. Estimated differential radiance for chemical species in the

stack plume for a non-combustion scenario.

Page 45: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 17Marinelli

Applied Optics44

Figure 17. Column density of agent for a simulated relese from a SCUD-B.

Page 46: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 18Marinelli

Applied Optics45

Figure 18. Total radiance at 9.4 µm for a simulated chemical agent

release using DMMP.

Page 47: SR-933a A Tunable Fabry-Perot Etalon-Based William J ... · PDF file1 A Tunable Fabry-Perot Etalon-Based Long-Wavelength Infrared Imaging Spectroradiometer William J. Marinelli, Christopher

Figure 19Marinelli

Applied Optics46

Figure 19. Net radiance at 9.4 µm for a simulated chemical agent release.

Flattening of the profile at the highest radiance levels indicates that the

cloud is optically thick in the center.