Seegene HPV28 manual en inglés

download Seegene HPV28 manual en inglés

of 43

Transcript of Seegene HPV28 manual en inglés

  • 7/26/2019 Seegene HPV28 manual en ingls

    1/43

    AnyplexTM II

    HPV28 Detection(Cat. No. HP7S00X)

    AnyplexTM II PCR System for detection of human papillomavirus - 19 high-risk HPV

    types(16, 18, 26, 31, 33, 35, 39, 45, 51, 52, 53, 56, 58, 59, 66, 68, 69, 73, 82) and 9

    low-risk HPV types(6, 11, 40, 42, 43, 44, 54, 61, 70) from cervical swab and liquid

    based cytology specimens.

    For use with the

    1. CFX96TMReal-time PCR System(Bio-Rad)

    European Union : (In Vitro Diagnostic Use)

    This product can be used for IVD purposes in European union.

    Other Countries : (Research Use Only)

    This product should be used for RUO purposes in other countries.

    Not available in the U.S.

  • 7/26/2019 Seegene HPV28 manual en ingls

    2/43

    2 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    TABLE OF CONTENTS

    NOTICES --------------------------------------------------------------------------------------------

    INTENDED USE -----------------------------------------------------------------------------------

    PRINCIPLES AND PROCEDURE OVERVIEW -------------------------------------------

    BACKGROUND INFORMATION --------------------------------------------------------------

    REAGENTS -----------------------------------------------------------------------------------------

    STORAGE AND HANDLING -------------------------------------------------------------------

    MATERIALS REQUIRED BUT NOT PROVIDED -----------------------------------------

    PROTOCOL -----------------------------------------------------------------------------------------

    REAL-TIME PCR INSTRUMENT SET UP AND RESULTS ANALYSIS ------------

    RESULTS --------------------------------------------------------------------------------

    TROUBLESHOOTING ---------------------------------------------------------------------------

    PERFORMANCE ----------------------------------------------------------------------------------

    REFERENCES -------------------------------------------------------------------------------------

    EXPLANATION OF SYMBOLS ----------------------------------------------------------------

    ORDERING INFORMATION --------------------------------------------------------------------

    3

    4

    5

    7

    8

    9

    9

    10

    16

    28

    34

    36

    40

    41

    42

  • 7/26/2019 Seegene HPV28 manual en ingls

    3/43

    3 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    NOTICES

    This product can be used for IVD(In Vitro Diagnostics) purposes in EU as the IVD CE Mark

    is reviewed by EU Directive(98/79/EC), but should be used for RUO(Research Use only)

    purposes in other countries.

    If this product is used with MICROLAB NIMBUS IVD and MICROLAB STARlet, maximum 5

    separate runs.

    This test has been validated for the following specimen types: cervical swab and

    liquid based cytology specimens. This test has not been validated for any other types of

    specimens.

    Store DNA samples at -70

    until use and keep on ice during use.

    The sensitivity of an assay may decrease if samples are repeatedly frozen/thawed or stored

    for a longer period of time.

    Reliability of the results depends on adequate specimen collection, transport, storage and

    processing procedure.

    Workflow in the laboratory should proceed in a unidirectional manner.

    Always wear disposable gloves in each area and change them before entering different

    areas. Change gloves immediately if contaminated or treat them with DNA decontaminating

    reagent.

    Dedicate supplies and equipment to the separate working areas and do not move them

    from one area to another.

    Do not pipette by mouth.

    Do not eat, drink or smoke in laboratory work areas. Wear disposable powder-free gloves,

    laboratory coats and eye protections when handling specimens and reagents. Wash hands

    thoroughly after handling specimens and test reagents.

    Avoid contamination of reagents when removing aliquots from reagent tubes. The use of

    sterile aerosol resistant disposable pipette tips is recommended.

    Do not pool reagents from different lots or from different tubes of the same lot.

    Do not use the product after its expiration date.

    Use screw-capped tubes and prevents any potential splashing or cross-contamination of

    specimens during preparations.

    Please be careful not to contaminate reagents with extracted nucleic acids, PCR products,

    and positive control. To prevent the contamination of reagents, the use of filter tips is

    recommended.

    Use separated and segregated working areas for each experiment.

  • 7/26/2019 Seegene HPV28 manual en ingls

    4/43

    4 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    Prepare and use a different pipette set for each of the following areas: Nucleic acid

    extraction, reagent mixing, nucleic acid template addition, and PCR product addition.

    Only at the designated space, open the reaction tubes or strips after the amplification, to

    avoid contamination with amplicon.

    Store positive materials separated from kits reagents.

    Laboratory safety procedures(refer to Biosafety in Microbiological and Biomedical

    Laboratories & CLSI Documents) must be taken when handling specimens. Thoroughly

    clean and disinfect all work surfaces with 0.5% sodium hypochlorite(in de-ionized or distilled

    water).

    INTENDED USE

    The AnyplexTM

    II HPV28 Detection is a qualitative in vitro test for the detection of human

    papillomaviruses in liquid based cytology and cervical swab specimens.

    The AnyplexTM

    II HPV28 Detectionassay consist of two PCR reaction(A set and B set).

    A setis a multiplex assay that permits the simultaneous amplification of target DNA of 14 high-

    risk human papillomaviruses.

    B setis a multiplex assay that permits the simultaneous amplification of target DNA of 5 high-

    risk and 9 low-risk human papillomaviruses.

    Category Types

    A set14 high-risk HPV types

    (16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, 68)

    B set5 high-risk HPV types(26, 53, 69, 73, 82)

    9 low-risk HPV types(6, 11, 40, 42, 43, 44, 54, 61, 70)

  • 7/26/2019 Seegene HPV28 manual en ingls

    5/43

    5 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    PRINCIPLES AND PROCEDURE OVERVIEW

    1. Principles

    The AnyplexTM

    II HPV28 Detection Assay, represents Seegenes proprietary technologies and is

    based on a newly developed TOCETM

    technology which makes it possible to detect multi-

    pathogens in a single fluorescence channel on real-time PCR instruments.

    In current melting curve analysis, temperature differences are often observed among DNAs that

    have high sequence variation, resulting in issues the field of clinical diagnostic where accurate

    and reproducible test results are critical. However, TOCETM

    technology is designated not to be

    affected by sequence variations; therefore it guaranteeing consistent Tm values.

    The AnyplexTM

    II HPV28 Detection can perform multiplex examination by either End point-CMTA

    (End point-Catcher Melting Temperature Analysis) or Cyclic-CMTA(Cyclic-Catcher Melting

    Temperature Analysis) method. Cyclic-CMTA method which represents a new class of

    molecular tests can discriminate major pathogen in the co-infected samples. The AnyplexTM

    II

    HPV28 Detection is a multiplex real-time PCR assay that permits the simultaneous

    amplification, detection and differentiation of target nucleic acids of 19 high-risk HPV types(16,

    18, 26, 31, 33, 35, 39, 45, 51, 52, 53, 56, 58, 59, 66, 68, 69, 73, 82) and 9 low-risk HPV types(6,

    11, 40, 42, 43, 44, 54, 61, 70) as well as Internal Control(IC).

    In PCR, efficiency can be reduced by inhibitors that may be present in the clinical specimens.

    An Internal Control(IC) is incorporated into the product as an endogenous whole process

    control in order to monitor nucleic acid isolation, and to check for possible PCR inhibition. The

    IC is co-amplified with the target nucleic acids within the clinical specimens. The AnyplexTM

    II

    HPV28 Detection uses Human house-keeping gene as an endogenous IC which can ensure

    purification of DNA, verification of PCR reaction and clarification of cell adequacy from each

    specimen.

    The Uracil-DNA glycosylase(UDG)d-UTP system is employed in the AnyplexTM

    II HPV28

    Detection. The UDG-dUTP system is commonly use when performing PCR to eliminate

    amplicon carry-over using UDG excises uracil residues from DNA by cleaving the N-glycosylic

    bond.

  • 7/26/2019 Seegene HPV28 manual en ingls

    6/43

    6 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    2. Procedure Overview

    < AnyplexTMII HPV28 Detection procedure overview >

  • 7/26/2019 Seegene HPV28 manual en ingls

    7/43

    7 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    BACKGROUD INFORMATION

    Human Papilloma Virus(HPV) infection is linked with cervical cancer. HPV can be divided into

    high-risk(HR) and low-risk(LR) groups on the basis of their association with cervical lesions.

    Therefore, it is very important to know which type of HPV is infected in patients to prevent

    cancer development and transmission of disease. Currently, commercially available major

    products to diagnose HPV are based on probe-hybridization method to detect and/or genotype

    HPV. However, main defect of the probe-hybridization based methods are high false positive

    rate due to cross-reactivity between probes and various kinds of viral DNA or PCR amplicons

    used for hybridization. Here we are introducing an innovative HPV detection/genotyping assay

    system which amplifies only specific targets without any cross reactivity and is automated in

    detection using real-time PCR method. Eventually the AnyplexTM

    II HPV28 Detection only

    specifically detects true HPV and accurately genotypes them. It also contains endogenous

    Internal Control to check any inhibition that might occur during PCR reaction.

    Cervical cancer, which progresses from the precancerous stage to invasive cancer, has 7-20

    years of precancerous stage; consequently early diagnosis is possible when HPV infection is

    suspected. High-risk HPV group may lead to the development of cervical cancer; especially,

    HPV16 and 18 are associated with 70% of cervical cancer case. On the other hands, low-risk

    HPV group including HPV6 and 11 may cause genital warts. AnyplexTM II HPV28 Detection can

    identify 19 high-risk HPV types including HPV16 and 18 and also detect for 9 low-risk HPV

    types such as HPV6 and 11 at the same time.

  • 7/26/2019 Seegene HPV28 manual en ingls

    8/43

    8 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    REAGENTS

    The reagents contained in one set are sufficient for 100 determinations.

    AnyplexTM

    II HPV28 Detection order information( HP7S00X).

    AnyplexTM II HPV28 Detection

    Symbols Contents Volume Description

    4X HPV28 A TOM 500 LTOCE Oligo Mix(TOM):

    - Amplification and detection reagents

    4X HPV28 B TOM 500 LTOCE Oligo Mix(TOM):

    - Amplification and detection reagents

    4X Anyplex

    PCR Master Mix

    (with UDG)

    500 L

    X 2

    - DNA polymerase

    - Uracil-DNA glycosylase(UDG)

    - Buffer containing dNTPs

    HPV28 PC1 100 LPositive Control(PC) :

    - Mixture of pathogen clones

    HPV28 PC2 100 LPositive Control(PC) :

    - Mixture of pathogen clones

    HPV28 PC3 100 LPositive Control(PC) :

    - Mixture of pathogen clones

    RNase-free Water1,000 L

    X 2

    Ultrapure quality, PCR-grade

    Negative Control(NC) :

    - Sterilized water as Negative Control

    User manual

    AnyplexTM II is a trademark of Seegene Inc.

  • 7/26/2019 Seegene HPV28 manual en ingls

    9/43

    9 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    STORAGE AND HANDLING

    The components of the AnyplexTM

    II HPV28 Detection should be stored at -20. All

    components are stable under recommended storage conditions until the expiry date stated on

    the label. Repeated thawing and freezing should be avoided, as this may reduce the sensitivity.

    If the reagents are to be used only intermittently, they should be frozen in aliquots.

    MATERIALS REQUIRED BUT NOT PROVIDED

    Disposable powder free gloves(latex or nitrile)

    Pipettes(adjustable) and Sterile pipette tips

    1.5 ml microcentrifuge tube

    Nucleic acid isolation kit(see Nucleic Acid Isolation)

    Proteinase K(For SEEPREP12TM

    , Cat. No.P4850, SIGMA)

    Ice Maker

    Desktop centrifuge

    Vortex mixer

    CFX96TMReal-time PCR system(Bio-Rad)

    Optical Flat 8-Cap Strips(Cat No. TCS0803, Bio-Rad)

    Low-Profile 0.2 mL 8-Tube Strips without Caps(white color, Cat. No. TLS0851, Bio-Rad)

    96-Well Skirted PCR Plate, white well(Cat. No. HSP-9655, Bio-Rad)

  • 7/26/2019 Seegene HPV28 manual en ingls

    10/43

    10 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    PROTOCOL

    1. Specimen Collection, Storage, and Transport

    Note: All samples have to be treated as potentially infectious materials. Only those sample

    materials are permitted, which are collected, transported and stored attending strictly the

    following rules and instructions :

    Note: To ensure a high sample quality, the specimens should be transported as fast as possible.

    The specimens have to be transported at the indicated temperature conditions.

    A. Specimen Collection

    Liquid based cervical cyto logy s pecimen

    Follow the manufacturers instructions for collecting cervical cell specimens into ThinPrep

    or SurePathTM

    media.

    Cervical swab sp ecimen

    For the collection of cervical swab specimen, please use following materials :

    Cervical swabs can be collected and transported in the following mediums :

    - ENAT(COPAN)

    Cervical specimen collection kit Manufacturer Cat. No.

    ENAT PM 2ML L-SHAPE APPLICATOR COPAN 606CS01L*

    * If you would like to purchase the above products from Seegene, Inc., please use this Cat. No.

    Leave the swab in the culture transport medium. Close and label the sample container.

    Stick closely to the instructions given for storage and transport.

    Please follow a recommended protocol to collect columnar and squamous epithelium cells

    after removal of the cervical mucus.

    B. Specimen Storage

    The sensitivity of an assay may decrease if samples are repeatedly frozen/thawed or stored for a

    longer period of time.

  • 7/26/2019 Seegene HPV28 manual en ingls

    11/43

    11 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    Liquid based cervical cyto logy s pecimen

    Cervical cell specimens collected in ThinPrep

    medium may be stored at 2 ~ 8

    for up to 6weeks. Cervical cell specimens collected in SurePath

    TMmedium may be restored at 2 ~ 8

    for up to 2 weeks.

    Note:The specimens should be extracted to nucleic acid as quickly as possible.

    Cervical swab sp ecimen

    If the swab specimens are not processed directly after their receipt in the laboratory, they have

    to be stored at 2 ~ 8and have to be processed within seven days.

    C. Specimen Transport

    To ensure a high quality of sample, specimens should be transported as soon as possible at

    indicated temperature.

    Liquid based cytology specimen

    Cervical cell specimens collected in ThinPrep medium can be transported at 2 ~ 25.

    Specimens collected in SurePathTM

    medium must be transported at 2 ~ 8

    Cervical swab sp ecimen

    Cervical swab specimens must be transported cooled.

    Cervical swab specimens should be shipped to a laboratory as soon as possible after

    collection, following the laboratory instructions for transports under cooling. The samples

    should be transported following also the local and national instructions for the transport of

    pathogen material.

    2. Nucleic Acid Isolation

    Various manufacturers offer nucleic acid isolation kits. Use right amount of sample according to the

    protocol in use. The following isolation kits have been validated for use with this kit.

    A. Pre-treatment of Liquid based cervical cytology specimen

    Equilibrate samples to room temperature(19 ~ 25).

    Centrifuge 1 mL of liquid based cervical cytology specimen for 15 minutes at 15,000 x g

    (13,000 rpm).

    The supernatant has to be discarded. Afterwards, the recommend volume(200 ~ 450 L, See

    Recommended Vol. of 2-C, D) should be resuspended in 1X PBS by vortexing thoroughly

  • 7/26/2019 Seegene HPV28 manual en ingls

    12/43

    12 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    to redissolve.

    Note: Process pre-treatment step using lysis buffer in extraction kit not 1X PBS if the samples arecollected in SurePath

    TMmedium and would be analyzed with NIMBUS or STARlet.

    Follow the manufacturers protocol.

    B.Cervical swab specimen

    Equilibrate samples to room temperature(19 ~ 25).

    For cervical swab specimens which contain a swab in the culture transport media specimens

    should be mixed by vortexing.

    The caps from specimen tubes have to be removed carefully to avoid contaminations. Any

    excess mucus in the specimen should be removed at this time by collecting it on the swab.

    Any residual liquid from the mucus and the swab should then be expressed by pressing the

    swab against the slide of the tube. Finally the swab and the mucus should be removed and

    discarded.

    ENAT specimens may be processed directly out of their primary container.

    C. Manual Prep Kits

    Isolation Kit Manufacturer Cat. No. Recommended Vol.

    QIAampDNA Mini Kit* QIAGEN 51304

    Specimen: 200 L

    Elution: 50 L

    Ribo_spin vRD**(Viral RNA/DNA Extraction Kit)

    GeneAll302-150

    SG1701***Specimen: 200 L

    Elution: 50 L

    * Process lysis step using 180 L of ATL buffer instead of AL buffer in case of SurePathTM

    media.

    **Ribo_spin vRD kit is not compatible with SurePathTM

    media.

    *** If you would like to purchase the above products from Seegene, Inc., please use this Cat. No.

    D. Automated Purification System

    Note: See MICROLAB NIMBUS IVD operation manual.

    Automated Purification System Manufacturer Cat. No. Recommended Vol.

    MICROLABNimbus IVD Hamilton 65415-02*

    STARMag 96 Tissue Seegene744300.4.

    205875

    Specimen: 450 L

    Elution: 100 L

    STARMag 48 X 8 Tissue Cartridge

    Kit

    Seegene744300.4.

    TC384

    Specimen: 450 L

    Elution: 100 L

  • 7/26/2019 Seegene HPV28 manual en ingls

    13/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    14/43

    14 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    E. Summary

    QIAGEN1 GeneAll

    2 SEEPREP12 NIMBUS/STARlet

    Swab ENAT O O O O

    LBCThinPrep O O O O

    SurePathTM

    O3 X O O4

    1. QIAamp DNA Mini Kit

    2. Robo_spin vRD (Viral RNA/DNA Extraction Kit)

    3. Process lysis step using 180L of ATL buffer instead of AL buffer.

    4. If DNA is extracted from SurePathTM

    specimens with NIMBUS or STARlet, there is a possibility that the

    sensitivity could be reduced compared to other extraction methods.

  • 7/26/2019 Seegene HPV28 manual en ingls

    15/43

    15 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    3. Preparation for Real-time PCR

    Note: The correct tubes and caps must be used(see MATERIALS REQUIRED BUT NOT

    PROVIDED).

    Note:Aerosol resistant filter tips and tight gloves must be used when preparing specimens. Use

    an extreme care to ensure no cross-contamination.

    Note: Completely thaw the reagents on ice.

    Note: Briefly centrifuge the reagent tubes to remove drops from the inner cap.

    5 L 4X HPV28 A TOM or B TOM

    5 L 4X Anyplex PCR Master Mix (with UDG)

    5 L

    5 L

    RNase-free Water

    samples nucleic acid

    20 L Total volume of PCR reaction

    Note: Calculate the necessary amount of each reagent needed based on the number of

    reactions(samples + controls).

    Note: Use a new sterile pipette tip for each sample.

    Note: For Negative Control, use 5 L of RNase-free Water instead of samples nucleic acid.

    Note: For Positive Control, use 5 L of each HPV28 PC1, PC2 and PC3.

    Note: Please be careful not to cross-contaminate the PCR Mastermix and samples with the

    Positive Control.

    Note: Do not label the cap of the reaction tubes as fluorescence is detected through the cap.

    Positive Control

    There are three Positive Control tubes included in the kit; HPV28 PC1, PC2 and PC3.

    Each PC includes clones for 5 targets in A set(14 types of high risk and IC) and 5 targets in B

    set(5 types of high risk, 9 types of low risk and IC).

    Note:To run the Positive Control reaction, prepare three PCR tubes for each set, six PCR tubes

    in total;

    For A set, first tube with PC1, second tube with PC2 and third tube with PC3.

    For B set, first tube wz`ith PC1, second tube with PC2 and third tube with PC3.

    (See RESULTS)

  • 7/26/2019 Seegene HPV28 manual en ingls

    16/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    17/43

    17 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    REAL-TIME PCR INSTRUMENT SET UP AND RESULTS ANALYSIS

    1. CFX96TM

    Real-time PCR System(Bio-Rad)

    1.1. Real-time PCR Instrument set up

    Note: The CFX96TM

    Real-time PCR System(Bio-Rad) experiment setup program for the

    detection of 28 types of HPV and IC can be divided into following three steps:

    - Protocol Setup

    - Plate Setup

    - Start Run

    A. Protocol Setup

    1) In the main menu, click Protocol to open the Experiment Setup.

    Fig. 1. Protocol Setup. Create a new protocol or load an existing protocol for the experiment.

  • 7/26/2019 Seegene HPV28 manual en ingls

    18/43

    18 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    2) In Protocol Editor, define the thermal profile as follows:

    i) Cyclic-CMTA (Melt analysis of three times)

    Segment Temperature Duration No. of cycles

    1 50C 4 min

    2 95C 15 min

    3 95C 30 sec

    304 60C 1 min

    5 72C 30 sec

    6 GOTO 3, 29 more times7 55C 30 sec

    8* Melting curve 55C ~ 85C (5 s / 0.5C)

    9 95C 30 sec

    1010 60C 1 min

    11 72C 30 sec

    12 GOTO 9, 9 more times

    13 55C 30 sec

    14* Melting curve 55C ~ 85C (5 s / 0.5C)

    15 95C 30 sec

    1016 60C 1 min

    17 72C 30 sec

    18 GOTO 15, 9 more times

    19 55C 30 sec

    20* Melting curve 55C ~ 85C (5 s / 0.5C)

    Note: Plate Read on Segment 8, 14 and 20. Fluorescence is detected at Melting.

  • 7/26/2019 Seegene HPV28 manual en ingls

    19/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    20/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    21/43

    21 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    B. Plate Setup

    1) In Experiment Setup Plate, click Create New to open the Plate Editor to create a new

    plate.

    Fig. 6. Plate Editor. Create a new plate or load an existing plate for the experiment.

    2) Click Select Fluorophores to indicate the fluorophores(FAM, HEX, Cal Red 610, Quasar

    670and Quasar 705) that will be used in the experiment.

    Fig. 7. Select Fluorophores(FAM, HEX, Cal Red 610, Quasar 670and Quasar 705).

    3) Choose the appropriate well and then click the Sample Typefrom the drop-down menu.

    - Unknown: Clinical samples

    - Negative Control

    - Positive Control

  • 7/26/2019 Seegene HPV28 manual en ingls

    22/43

    22 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    4) Click the appropriate checkboxes(FAM, HEX, Cal Red 610, Quasar 670and Quasar 705)

    to specify the fluorophores in the selected wells.5) Type the Sample Nameand PC(PC1, PC2, andPC3), and then press enter key.

    6) In Settings of the Plate Editor main menu, choose the Plate Size and Plate Type(BR

    White).

    Fig. 8.Plate Setup.

    7) Click OKand save a new plate set-up file.

    8) Then Experiment Setupwindow will open.

    Fig. 9. Experiment Setup Plate.

  • 7/26/2019 Seegene HPV28 manual en ingls

    23/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    24/43

    24 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    B. Pre-settings for Data Analysis in CFX96

    1) After the test, click the Melt curve field to confirm the Melt Peak results.

    Fig. 11. Result of Melting peak.

    2) Select Step number8and select Export All Data Sheets to Excelfrom Tools menu.

    Note: Select Export All Data Sheets to Exceldirectly in case of End point-CMTA.

    Fig. 12. Export All Data sheets to excel.

  • 7/26/2019 Seegene HPV28 manual en ingls

    25/43

    25 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    3) Save the result to the 1folder, when a new window is opened.

    Note: Save the result to the arbitrary folder in case of End point-CMTA.

    Fig. 13.Export all data from spreadsheets in data analysis to designated folder.

    4) Make sure the result have been saved to the folder.

    Fig. 14. Exported Result files.

    Note: Skip 5)~8) steps and process next analysis stage in case of End point-CMTA.

    5) Return to step 2) and select Step number14. Repeat steps 3) & 4) and save data in 2

    folder.

    6) Return back to step 2) and select Step number 20.

    7) It is necessary to regulate threshold bar before export data from step number 20. Select

    only Quasar 670, the threshold bar of melt peak should be adjusted to zero.

  • 7/26/2019 Seegene HPV28 manual en ingls

    26/43

    26 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    Fig. 15. Melt peak Threshold of Quasar 670.

    8) Repeat steps 3) & 4) and save data in 3folder. Data of each step number is saved as

    shown below.

    Step number Designated folder

    8 1

    14 2

    20 3

  • 7/26/2019 Seegene HPV28 manual en ingls

    27/43

    27 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    C. Settings for Data Analysis in Seegene Viewer

    1) Open the Seegene Viewer program in the screen, and click open to find the saved file in 1

    folder.

    Note: Open to find the saved file in arbitrary folder in case of End point-CMTA.

    Fig. 16. Seegene viewer.

    2) After opening the results file, click PRODUCT column and select test kit in the lists.

    Fig. 17. Settings for Data Analysis in Seegene Viewer.

    Note: Please check the type of the tube at the time of test kit selection (8-strip or 96 plate).

  • 7/26/2019 Seegene HPV28 manual en ingls

    28/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    29/43

    29 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    RESULTS

    1. Analyte Information

  • 7/26/2019 Seegene HPV28 manual en ingls

    30/43

    30 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    2. Interpretation of Results

    A. Cyclic-CMTA

    HPV Result* IC Result

    * Interpretation

    +++ or ++ or + +++ or ++- HPV DNA, detected

    - Target HPV type identification

    +++ or ++ or + + or -

    - HPV DNA, detected

    - Target HPV type identification

    - Additional HPV genotypes that were not detected may be

    present.

    - +++ or ++ - HPV DNA, not detected

    - + or -

    Invalid

    - Weak or negative IC signal was suggestive of inadequate

    specimen collection, processing or the presence of

    inhibitors.

    - Process another aliquot of the original specimen and

    repeat the test.

    Result

    Cyclic-CMTA

    (Cyclic Catcher Melting Temperature Analysis)

    First CMTA point Second CMTA point Third CMTA point

    +++ + + +

    ++ - + +

    + - - +

    - - - -

  • 7/26/2019 Seegene HPV28 manual en ingls

    31/43

    31 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    B. End point-CMTA

    HPV Result* IC Result

    * Interpretation

    + +- HPV DNA, detected

    - Target HPV type identification

    + -

    - HPV DNA, detected

    - Target HPV type identification

    - Additional HPV genotypes that were not detected may be

    present.

    - + - HPV DNA, not detected

    - -

    Invalid

    - Negative IC signal was suggestive of inadequate specimen

    collection, processing or the presence of inhibitors.

    - Process another aliquot of the original specimen and

    repeat the test.

    * Internal Control or any other signals are not observed: see TROUBLESHOOTING.

    Detection of the Internal Control in the Quasar 670 channel is not required for positive results. High

    titer of another analyte can lead to a reduced or absent Internal Control signal.

  • 7/26/2019 Seegene HPV28 manual en ingls

    32/43

    32 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    3. Application to Clinical Samples

    A. Cyclic-CMTA

    Melt Peak-1st

    (First CMTA point)

    Melt Peak-2nd

    (Second CMTA point)

    Melt Peak-3rd

    (Third CMTA point)

  • 7/26/2019 Seegene HPV28 manual en ingls

    33/43

    33 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    Melt Peak-1

    st

    (First CMTA point)

    Melt Peak-2nd

    (Second CMTA point)

    Melt Peak-3rd

    (Third CMTA point)

    FAM HEX Cal Red 610 Quasar 670 Quasar 705 Auto interpretation

    A set 66 45 58 51 59 16 33 39 52 IC 35 18 56 68 31

    66,52,70

    Sample + - - - - - - - ++ +++ - - - - -

    B set 26 69 73 42 82 53 43 54 70 IC 61 6 44 40 11

    Sample - - - - - - - - ++ +++ - - - - -

  • 7/26/2019 Seegene HPV28 manual en ingls

    34/43

    34 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    B. End point CMTA

    A set

    B set

    FAM HEX Cal Red 610 Quasar 670 Quasar 705 Auto interpretation

    A set 66 45 58 51 59 16 33 39 52 IC 35 18 56 68 31

    66,52,70

    Sample + - - - - - - - + + - - - - -

    B set 26 69 73 42 82 53 43 54 70 IC 61 6 44 40 11

    Sample - - - - - - - - + + - - - - -

  • 7/26/2019 Seegene HPV28 manual en ingls

    35/43

    35 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    TROUBLESHOOTING

    AnyplexTMII HPV28 Detection

    OBSERVATION PROBABLE CAUSES SOLUTION

    Internal Control

    or any other

    signals are not

    observed

    The fluorophores for data

    analysis do not comply with

    the protocol

    Select the correct fluorophores for data analysis.

    Incorrect PCR cycle or

    machine temperature

    Please check the PCR conditions and repeat the

    PCR under the correct setting if necessary.

    Leaving reagents at room

    temperature for a long time

    or incorrect storage condition

    Please check the storage conditions(See 9 page)

    and the expiration date(see the kit label) of the

    reagents and use a new kit if necessary.

    Incorrect programming Repeat the detection procedure with a correct

    setting.

    Nucleic acid extraction failure Make sure that you use a recommended isolation

    method.

    Error in specimen collection If the both target and IC signal were not observed

    that means specimen collected inappropriately

    recollect the specimen.

    Internal Control

    signal is not

    observed

    High load of pathogen's

    nucleic acid

    If the target signal is observed, sample is

    presumably detection for target pathogens, although

    IC signal is not observed.

    If you want to check the IC, dilute the specimen in

    PBS(10-100x) and repeat from extraction step with

    the diluted specimen.

    Presence of PCR Inhibitor Dilute the specimen in PBS(10-100x), and repeat

    from extraction step.

    False positive or

    signals

    observed at the

    Negative Control

    Presence of cross

    contamination

    Decontaminate all surfaces and instruments with

    sodium hypochlorite and ethanol. Use only filter tips

    during the extraction procedure. Change tips among

    tubes. Repeat the nucleic acid extraction with the

    new set of reagents.

    Cross-contamination

    between PC 1, 2 and 3

    Restart from extraction step

    or

    Restart from real-time PCR step.

  • 7/26/2019 Seegene HPV28 manual en ingls

    36/43

    36 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    AnyplexTMII HPV28 Detection

    OBSERVATION PROBABLE CAUSES SOLUTION

    False negative

    or no signals

    observed at the

    Positive Control

    Error in specimen collection Recollect the specimen.

    Incorrect storage of the

    specimen

    Recollect the specimen and repeat the whole

    process. Make sure the product is stored in

    recommended conditions.

    Error in nucleic acid

    extraction

    Re-extract the nucleic acid.

    Error in adding nucleic acid

    to corresponding PCR tubes

    Check the sample numbers for nucleic acid

    containing tubes and make sure to add nucleic acidinto correct PCR tubes during detection process.

    Presence of inhibitor Dilute the specimen in PBS(10-100x), and repeat

    from extraction step with the diluted specimen.

    The fluorophores for data

    analysis do not comply with

    the protocol

    Select the correct fluorophores for data analysis.

    Incorrect programming Repeat the PCR with corrected setting.

    Incorrect PCR mixture Check whether all components are added or not(If

    you use to precomposed premix, should be reduce

    sensitivity). Each reagent used for homogenization

    and spin down reagent tube before put the real-time

    PCR.

    Leaving reagents at room

    temperature for a long time

    or incorrect storage condition

    Please check the storage condition and the

    expiration date(see the kit label) of the reagents and

    use a new kit if necessary.

  • 7/26/2019 Seegene HPV28 manual en ingls

    37/43

    37 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    PERFORMANCE

    1. Specificity

    The high specificity of the AnyplexTM

    II HPV28 Detection is ensured by the primers

    specifically designed for the targets in interest and the reaction condition. AnyplexTM

    II

    HPV28 Detection has been tested for cross-reactivity in 85 different pathogens: result

    illustrated PCR amplifications in targets only.

    Organism Strain No.

    Result

    Acinetobacter baumannii ATCC 15150 -

    Bacteroides fragilis ATCC 25285D -

    Chlamydia trachomatis ATCC VR-577 -

    Corynebacterium genitalium ATCC 33030 -

    Enterobacter cloacae KCTC 13047 -

    Enterococcus faecalis ATCC 700802D-5 -

    Escherichia coli ATCC 15489 -

    Fusobacterium nucleatum ATCC 25586D-5 -

    Gardnerella vaginalis ATCC 14019 -

    Haemophilus ducreyi ATCC 33940 -

    Klebsiella pneumoniae ATCC 13883 -

    Lactobacillus acidophilus ATCC 4357D-5 -

    Lactobacillus crispatus ATCC 33820 -

    Lactobacillus gasseri ATCC 33323 -

    Lactobacillus iners ATCC 55195 -

    Lactobacillus jensenii ATCC 25258 -

    Mobiluncus curtisii ATCC 35241 -

    Mobiluncus mulieris ATCC 35243 -

    Neisseria gonorrhoeae ATCC 700825D -

    Neisseria meningitidis ATCC 700532D -

    Neisseria sicca ATCC 29256 -

    Peptostreptococcus anaerobius ATCC 49031D-5 -

    Propionibacterium acnes ATCC 6919 -

    Proteus mirabilis ATCC 12453 -

  • 7/26/2019 Seegene HPV28 manual en ingls

    38/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    39/43

    39 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    Organism Strain No.

    Result

    HPV6 ATCC 45150D + (HPV6)

    HPV11 ATCC 45151D + (HPV11)

    HPV16 ATCC 45113D + (HPV16)

    HPV18 ATCC 45152D + (HPV18)

    HPV26 Korean isolate + (HPV26)

    HPV31 ATCC 65446 + (HPV31)

    HPV33 Korean isolate + (HPV33)

    HPV35 ATCC 40330 + (HPV35)

    HPV39 Korean isolate + (HPV39)

    HPV40 Korean isolate + (HPV40)

    HPV42 Korean isolate + (HPV42)

    HPV43 ATCC 40339 + (HPV43)

    HPV44 Korean isolate + (HPV44)

    HPV45 Korean isolate + (HPV45)

    HPV51 Korean isolate + (HPV51)

    HPV52 Korean isolate + (HPV52)

    HPV53 Korean isolate + (HPV53)HPV54 Korean isolate + (HPV54)

    HPV56 ATCC 40549 + (HPV56)

    HPV58 Korean isolate + (HPV58)

    HPV59 Korean isolate + (HPV59)

    HPV61 Korean isolate + (HPV61)

    HPV66 Korean isolate + (HPV66)

    HPV68 Korean isolate + (HPV68)

    HPV69 Korean isolate + (HPV69)

    HPV70 Korean isolate + (HPV70)

    HPV73 Korean isolate + (HPV73)

    HPV82 Korean isolate + (HPV82)

    SiHa Cell KCLB 30035 + (HPV16)

    HeLa Cell KCLB 10002 + (HPV18)

  • 7/26/2019 Seegene HPV28 manual en ingls

    40/43

  • 7/26/2019 Seegene HPV28 manual en ingls

    41/43

    41 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    REFERENCES

    1

    2

    3

    4

    5

    6

    7

    8

    9

    10

    11

    12

    13

    14

    15

    Burd EM. [Human papillomavirus and cervical cancer.] Clin Microbiol Rev. (2003) 16(1): 1-17

    Castle PE. [The potential utility of HPV genotyping in screening and clinical management.] J Natl Compr Canc

    Netw. (2008) 6(1): 83-95 Review

    Chris JM, Peter JS, Philip EC. [Clinical utility of HPV genotyping.] Gynecol Oncol. (2006) 103: 12-17

    Chun JY, Kim KJ, Hwang IT, Kim YJ, Lee DH, Lee IK, Kim JK. [Dual priming oligonucleotide system for the

    multiplex detection of respiratory viruses and SNP genotyping of CYP2C19 gene.] Nucleic Acids Res. (2007)

    35(6): e40

    Chun JY. [High Multiplex Molecular Diagnostics.] Seegene Bulletin (2012) 1: 1-4

    Giorgi Rossi P, Bisanzi S, Paganini I, Di Iasi A, Angeloni C, Scalisi A, Macis R, Pini MT, Chini F, Carozzi FM. [HPV

    Prevalence Italian Working GroupPrevalence of HPV high and low risk types in cervical samples from the Italian

    general population: a population based study.]BMC Infect Dis. (2010) 20(10): 214

    Hwang IT. [Cyclic-CMTA: An Innovative Concept in Multiplex Quantification.] Seegene Bulletin (2012) 1: 11-15

    Krane JF, Granter SR, Trask CE, Hogan CL, Lee KR. [Papanicolaou smear sensitivity for the detection of

    adenocarcinoma of the cervix: a study of 49 cases.] Cancer. (2001) 93(1): 8-15

    Lee DH. [TOCE: Innovative Technology for High Multiplex Real-time PCR.] Seegene Bulletin (2012) 1: 5-10

    Li J,Mei J,Wang X,Hu L,Lin Y,Yang P.[Human papillomavirus type-specific prevalence in women with cervical

    intraepithelial neoplasm in Western China.]J Clin Microbiol. (2012) 50(3): 1079-1081

    Novaes LC, Novaes MR, Simes-Barbosa A. [Diagnosis of human papillomatosis by polymerase chain reaction in

    cases of divergence between results of hybrid capture and papanicolaou cytology.] Braz J infect Dis. (2006)

    10(3):169-172

    Son S, Noh HT, An S. [Human papillomavirus status in cervical scrapes and biopsy specimens using the HPV

    genotyping DNA microarray.]Int J Gynaecol Obstet. (2006) 93(3): 258-259

    Sun ZR, Ji YH, Zhou WQ, Zhang SL, Jiang WG, Ruan Q. [Characteristics of HPV prevalence among women in

    Liaoning province, China.]Int J Gynaecol Obstet. (2010) 109(2): 105-109

    Wallace J, Woda BA, Pihan G. [Facile, Comprehensive High-Throughput Genotyping of Human Genital

    Papillomaviruses Using Spectrally Addressable Liquid Bead Microarrays.] J Mol Diagn. (2005) 7(1): 72-80

    Ursu RG, Onofriescu M, Nemescu D, Iancu LS. [HPV prevalence and type distribution in women with or without

    cervical lesions in the Northeast region of Romania.]Virol J. (2011) 22(8): 558

    http://www.ncbi.nlm.nih.gov/pubmed/20646310http://www.ncbi.nlm.nih.gov/pubmed/20646310http://www.ncbi.nlm.nih.gov/pubmed?term=Li%20J%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Mei%20J%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Wang%20X%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Hu%20L%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Lin%20Y%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Yang%20P%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Jinke%20Li%2C%20Pei%20Yanghttp://www.ncbi.nlm.nih.gov/pubmed?term=Human%20papillomavirus%20status%20in%20cervical%20scrapes%20and%20biopsy%20specimens%20using%20the%20HPV%20genotyping%20DNA%20microarrayhttp://www.ncbi.nlm.nih.gov/pubmed/20138618http://www.ncbi.nlm.nih.gov/pubmed/20138618http://www.ncbi.nlm.nih.gov/pubmed/22192090http://www.ncbi.nlm.nih.gov/pubmed/22192090http://www.ncbi.nlm.nih.gov/pubmed/22192090http://www.ncbi.nlm.nih.gov/pubmed/22192090http://www.ncbi.nlm.nih.gov/pubmed/20138618http://www.ncbi.nlm.nih.gov/pubmed/20138618http://www.ncbi.nlm.nih.gov/pubmed?term=Human%20papillomavirus%20status%20in%20cervical%20scrapes%20and%20biopsy%20specimens%20using%20the%20HPV%20genotyping%20DNA%20microarrayhttp://www.ncbi.nlm.nih.gov/pubmed?term=Jinke%20Li%2C%20Pei%20Yanghttp://www.ncbi.nlm.nih.gov/pubmed?term=Yang%20P%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Lin%20Y%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Hu%20L%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Wang%20X%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Mei%20J%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed?term=Li%20J%5BAuthor%5D&cauthor=true&cauthor_uid=22170939http://www.ncbi.nlm.nih.gov/pubmed/20646310http://www.ncbi.nlm.nih.gov/pubmed/20646310
  • 7/26/2019 Seegene HPV28 manual en ingls

    42/43

    42 03/2013 V1.04

    AnyplexTM

    II HPV28 Detection

    EXPLANATION OF SYMBOLS

    Explanation of symbols used in label and manual.

    Symbol Explanation

    In vitro diagnostic use

    Research use only

    Batch code

    Catalogue number

    Use by

    Temperature limitation

    Caution

    Oligonucleotide Mix for amplification and detection

    RNase-free Water

    Positive Control

    AnyplexTM

    PCR Master Mix

    Manufacturer

    Date of manufacture

    Consult instructions for use

    Authorized representative in the European community

  • 7/26/2019 Seegene HPV28 manual en ingls

    43/43

    AnyplexTM

    II HPV28 Detection

    ORDERING INFORMATION

    Cat. No. Product Size

    AnyplexTM

    II HPV Series

    HP7S00X AnyplexTM

    II HPV28 Detection 100 rxns

    Seeplex

    HPV Series

    HP6401Y Seeplex

    HPV4A ACE Screening 50 rxns

    Cervical specimen collection kit

    606CS01L ENAT PM 2ML L-SHAPE APPLICATOR 50 tests

    Manual extraction system

    SG1701 Ribo_spin vRD(Viral RNA/DNA Extraction Kit) 50 preps

    Automated extraction system

    SPN1200 SEEPREP12TM

    EA

    SPN1004 SEEPREP12TM

    Viral NA kit 96 preps

    SPN1101 SEEPREP12TM Tip Set 96 tips

    65415-02 MICROLAB Nimbus IVD EA

    173000-075 MICROLAB STARlet EA

    744300.4.205875 STARMag 96 Tissue 384T / 1box

    744300.4.TC384 STARMag 48 X 8 Tissue Cartridge Kit 384T / 1box