Scanned-proximity Probe Microscopy (SPM) Background

14
Scanned-proximity Probe Microscopy (SPM) Background Emphasis on Atomic Force Microscopy (AFM) Reading SPM Features AFM Specifics AFM Operation (Conceptual) AFM and Nanotechnology

description

Scanned-proximity Probe Microscopy (SPM) Background. Emphasis on Atomic Force Microscopy (AFM). Reading SPM Features AFM Specifics AFM Operation (Conceptual) AFM and Nanotechnology. Reading. - PowerPoint PPT Presentation

Transcript of Scanned-proximity Probe Microscopy (SPM) Background

Page 1: Scanned-proximity Probe Microscopy (SPM) Background

Scanned-proximity Probe Microscopy (SPM)

Background Emphasis on Atomic Force

Microscopy (AFM)

•Reading

•SPM Features

•AFM Specifics

•AFM Operation (Conceptual)

•AFM and Nanotechnology

Page 2: Scanned-proximity Probe Microscopy (SPM) Background

Reading• “Window on a Small World,” McGuire, Today’s Chemist,

June 2002Vol. 11, No. 5 pp 21–24 http://pubs.acs.org/subscribe/journals/tcaw/11/i06/html/06inst.html

• “AFM Technology Overview” From Veeco (Digital Instruments) the manufacturer of the AFM that will be used.

• http://nano.nd.edu/SC190/scanningprobe.pdf “AFM Nanomanipulation” From Veeco on the use of the AFM in direct nanotechnology applications. http://www.veeco.com/pdfs.php/70

• Scanning Probe Microscopy (SPM) Instructions http://frontpage.okstate.edu/nanotech/Lab/lab3/SPM_Instructions_082003.pdf

Page 3: Scanned-proximity Probe Microscopy (SPM) Background

Scanned Proximity Probe Microscopy Common Features

• Piezoelectric (PZ) positioning– PZ crystals expand/contract under applied voltage

(d11 ≈ 4x10-10 m/V)• Example: A commercial linear actuator produces 15 m

displacement for 100 V applied with sub nm resolution– PZ crystals generate voltages under applied force

(g33 ≈ 1x10-2 V/Nm)• Deflection Feedback Loop

– Height (z) dependent signal • Example: Tunneling current for Scanning Tunneling Microscope

(STM)• Example: Van der Waals forces for Atomic Force Microscope (AFM)

• Probe/Tip– Photolithographic Fabrication Techniques– Resolution determined by probe size (not diffraction)

Page 4: Scanned-proximity Probe Microscopy (SPM) Background

AFM Capabilities (Advantages)

• Able to achieve a resolution of 10 pm with special tips (typical resolution ≈ 10 nm)

• Able to image samples in air and under liquids

• Able to measure in 3 D (within limits)• Able to measure non-conductive surface

(unlike SEM or STM)

Page 5: Scanned-proximity Probe Microscopy (SPM) Background

AFM Capabilities (Disadvantages)

• Incorrect tip choice can lead to measurement artifacts or sample damage

• Depth of field limited by cantilever / z positioning PZ

• Scan area limited by PZ scanners• Slow scan rate compared to SEM

Page 6: Scanned-proximity Probe Microscopy (SPM) Background

Laser

Position Sensitive Photodetector (PSPD)

Cantilever

Probe Tip

X-Y Piezoelectric Positioning/

Scanning

Substrate

Sample

Optical Beam

Z Piezoelectric Positioning/ Force

Sensor

AFM Components

Page 7: Scanned-proximity Probe Microscopy (SPM) Background

• Tip– Modifies measurements– Typically Si or SiN for ease of fabrication– Many variations depending on application

• Cantilever (Tip at the end)– Low spring constant (Hooke’s Law F = -kz)– Low weight for high resonant frequency ( )– Coated for reflectivity

AFM Probe

Images from http://stm2.nrl.navy.mil/how-afm/how-afm.html#General%20concept

12

kfm

Contact Non-Contact

k (N/m) 0.2 0.006 140 2

f (kHz) 70 10 1100 70

Tip on apex of cantilever

Page 8: Scanned-proximity Probe Microscopy (SPM) Background

AFM Modes• Contact mode (Repulsive-Static)

– AFM tip rides on the sample in close contact with the sample surface (low k)

– The force in the feedback loop is friction– May interact with the sample surface

• Non-contact mode (Attractive-Dynamic)– AFM tip hovers 5-15 nm away from the sample surface– The force in the feedback loop is typically van der Waals (VDW)

forces– Applied force (dependent on height z) changes cantilever

oscillation frequency.• Tapping mode (Repulsive-Dynamic)

– AFM tip taps surface as it maps z.– Eliminates lateral forces or hysteresis due to the tip sticking on

the sample.– Less likely to damage the sample

Page 9: Scanned-proximity Probe Microscopy (SPM) Background

Changes of the surface properties along the scan line

Changes of interaction forces between the probe tip and sample surface

Deflection of the cantilever Change of oscillation amplitude and phase of the cantilever

Change in laser position on PSPD Laser beam position oscillates on the PSPD

Electronic signal to control and recording electronics

Signal processing to form image

AFM Measurement

Contact Mode Non-Contact Mode

Page 10: Scanned-proximity Probe Microscopy (SPM) Background

PSPD measures change optical beam position change in cantilever height

Sample is raster scanned

AFM Visualization (Contact Mode)

Page 11: Scanned-proximity Probe Microscopy (SPM) Background

Forces at Work

• Atomic Forces (approx.)

13 7

12 6a

B AF Uz z

RepulsiveAttractive

Detectable forces for an AFM 1 nN in the contact regime and 1 pN in the noncontact regime (Theoretical Limit 10-18 N with heroic measures.) R. Wiesendanger, "Chapter 11. Future Nanosensors." In H. Meixner, R. Jones, eds., Volume 8: Micro- and Nanosensor Technology / Trends in Sensor Markets. In W. Gopel, J. Hesse, J.N. Zemel, eds., Sensors: A Comprehensive Survey, VCH Verlagsgesellschaft mbH, Weinheim Germany, 1995; pp. 337-356.

B and A coefficients depend on the surfaces involved.

U

Page 12: Scanned-proximity Probe Microscopy (SPM) Background

Snap On

Approach

Impress

Snap Off

Retract

Forc

e

(pro

port

iona

l to

cant

ileve

r de

flect

ion)

Approach-Retract Phenomenon

1. Approach—VDW forces pull tip toward surface

2. Snap On—When close enough, VDW grow stronger than restoring spring force in the cantilever.

3. Impress—Tip pressed into sample with positive force after reaching the set point deflection

4. Retract—Tip adheres to surface giving rise to hysteresis on measured force/position curve

5. Snap Off—Spring force dominates the tip-sample adhesion and tip leaves surface

(NOT the tip Z position!)

Page 13: Scanned-proximity Probe Microscopy (SPM) Background

Alternate Scanning Probe Systems• Magnetic Force—Maps surface magnetic field (think

hard drives…)• Electrostatic Force—Maps surface potential (100 nm

resolution)• Lateral Force—Maps friction force experienced by

scanning probe on surface orthogonal to scanning direction

• Magnetic Resonance—Possible 3D imaging of individual molecules via resonant electron spin flipping with an antenna on the cantilever

• Near Field Optical Microscopy – Probe tip has a small aperture (radius << ) for optical

wavelength measurements• And many, many more…

Page 14: Scanned-proximity Probe Microscopy (SPM) Background

Nanotech Applications

• Measurement of nanostructures• Nano indention of surfaces• Dip Pen Nanolithography • Manipulating nanoparticles (building

nanostructures)• Precision electro chemistry—supply

electrons by applying voltage across AFM tip and substrate