S6 Service Manual Ver2.0

76
MMED6000DP/SF-CSN801 Service Manual Copyright @ 2005 Choice Electronic Technology Co. Ltd. Edition: MMED6000DP-CSN801 Issued date: Aug.2005

description

service manual for pacient monitor choice med

Transcript of S6 Service Manual Ver2.0

MMED6000DP/SF-CSN801 Service Manual

Copyright @ 2005 Choice Electronic Technology Co. Ltd. Edition: MMED6000DP-CSN801 Issued date: Aug.2005

MMED60000DP (S6) Service Manual

COPYRIGHTBeijing Choice Electronic Technology Co., Ltd. (hereinafter called Beijing Choice) owns all right to this unpublished work and intends to maintain this work as confidential. Beijing Choice may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose of reference, operation, maintenance, or repair of Choice equipment. No part of this can be disseminated for other purposes. In the event of inadvertent or deliberate publication, Beijing Choice intends to enforce its right to this work under copyright laws as a published work. Those having access to this work may not copy, use, or disclose the information in this work unless expressly authorized by Beijing Choice to do so. All information contained in this publication is believed to be correct. Choice shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this material. This publication may refer to information and protected by copyrights or patents and does not convey any license under the patent rights of Beijing Choice, nor the rights of others. Beijing Choice does not assume any liability arising out of any infringements of patents or other rights of third parties. Content of the manual is subject to changes without prior notice. PROPERTY OF BEIJING CHOICE ELECTRONIC TECHNOLOGY CO., LTD. ALL RIGHTS RESERVED

MMED60000DP (S6) Service Manual

CONTENTChapter 1 Introduction General Description of abbreviation Explanation of symbols Special feature Appearance of monitor Specification Chapter 2 Special operation in menu Chapter 3 Monitor function and principle Power supply unit CSN801 board Chapter 4 Component and structure Component Structure Chapter 5 Troubleshooting Diagram of solutions to various troubles Troubleshooting Error Message Chapter 6 Calibration of monitor Required testing tool Power on the monitor Front panel test Audio system test Test of automatic storage Alarm test ECG test Respiration test Temperature test 5 5 5 5 6 7 13 17 19 19 19 21 21 21 30 30 40 43 44 44 44 44 45 45 45 46 47 47

MMED60000DP (S6) Service Manual

NIBP test SPO2 test Chapter 7 Adjustment procedure ECG adjustment NIBP adjustment SpO2 adjustment IBP Calibration ET-CO2 calibration Chapter 8 Disassembly procedure Monitor Disassembly Battery Removal Chapter 9 Cleaning and Disinfection Maintenance check General cleaning Sterilization Precondition and cleaning TEMP sensor cleaning and disinfection SpO2 sensor cleaning and disinfection Chapter 10 Warranty Warranty and repair Exemption and restriction Customer guarantee Non-warranty and non-replacement policy Customer special warranty period Repackaging

47 49 50 50 50 50 52 54 55 55 55 57 57 57 58 59 59 60 61 61 62 62 62 62 62

MMED60000DP (S6) Service Manual

MMED60000DP (S6) Service Manual

CHAPTER 1 INTRODUCTION1.1 GeneralMMED6000DP is a Portable Patient Monitor that has abundant monitoring functions and is used for the clinical monitoring of adult, pediatric and neonate. Besides, the user may select the different parameter configuration according to different requirements. MMED6000DP can be connected to the central working station to form a network monitoring system. MMED6000DP can monitor vital signals as ECG, RR, SpO2, PR, NIBP, TEMP, Dual-channel IBP (optional), CO2 (optional). It integrates parameter measuring modules, display and recorder in one device, featuring in compactness, lightweight and portability. Large true color display provides clear view of 6 waveforms and full monitoring parameters.

1.2 Description of AbbreviationHR: SPO2: NIBP: IBP: S: M: D: PR: RR: TEMP1: TEMP2: CO2: heart rate 2-channel ECG waveform arterial oxygen saturation non-invasive blood pressure invasive blood pressure systolic mean blood pressure diastolic pulse rate respiration rate temperature channel 1 temperature channel 2 end-tidal CO2 concentration

1.3 Explanation of SymbolsAttention Refer to relevant the prompt. Resistant defibrillator BF type equipment Rotated knob Power on/off Counter-clockwise and clockwise the Class II Equipment Heart Beat Detected

Alarm on

Alarm off

MMED60000DP (S6) Service Manual

Sound on

Silence

System setup menu

Trend menu

Equipotential grounding terminal

AC 50/60 HZ

Fuse

This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The unit displaying this symbol contains an F-type isolated (floating) patient applied part providing a high degree of degree of protection against shock, and is suitable for use during defibrillation. This item is compliant with Medical Device Directive 93/42/EEC of 14 June 1993, a directive of the European Economic Community.

1.4 Special Features Portable, compact, AC power and internal rechargeable battery; Automatically and rapid charge, 80-90% power will be recharged within 3 hours; Resistant high-frequency electrotome design, reliable and special module is used in operation room; Optional inner printer with 3 channels wave or external printer; Adjust volume more accurately by digital system; Menu design adopts Huffman decode, operating more effectively; Support Ethernet, wireless LAN and could connect with Central monitoring system.

MMED60000DP (S6) Service Manual

1.5 Appearance of Monitor1.5.1. Screen display 4 channels

Fig.1-1

The display of MMED6000DP patient monitor is TFT LCD, which displays the patient parameters, waveforms, historical data and monitor status. The screen is divided into five areas: (1) monitor status, (2) waveforms, (3) the table of historical data, (4) parameters (5) menu as well as (6) frame of menu status.

MMED60000DP (S6) Service Manual

Fig. 1-2

Monitoring status frame Bed No.: indicates the bed number of the patient being monitored when the patient monitor is connected with the workstation. DEMO: indicates the device is in DEMO monitoring mode now. In clinical application, this function is not recommended because the DEMO will mislead the hospital workers to treat the waveform and parameter as actual data of the patient, which may result in delay of treatment or mistreatment. Start time: indicates the time powered on the monitor, it shows in 16:25. Current time: indicates the current date and time, it shows in 05/16/2004, 10:10

Waveforms displaying area The waveform area can maximally display 6 waveforms. The colour of waveforms matches with the parameters on the right of the screen. For the maximum configuration, the waveforms provided by the system for selection are (from up to bottom): 2-lead ECG waveforms SpO2 waveform RESP waveform There are two means to gain RESP waveform, trans-thoracic impedance and Et-CO2 (optional). If operator selects Et-CO2 (side stream) to monitor the patients respiration, the CO2 waveform will instead of the RESP waveform acquired by Trans-thoracic impedance. 2 IBP waveforms(option) If operator selects IBP operation, the table of historical data will be instead of 2 IBP waveforms.

MMED60000DP (S6) Service Manual

Numeric displaying area The parameter area lies to the right side of the graphic area, which includes: ECG: heart rate (unit: beats/minute) NIBP: from left to right, there are Systolic pressure, Mean pressure and Diastolic pressure (unit: mmHg or kPa) IBP: the BP of channel 1 and 2. From left to right, there are Systolic pressure, Mean pressure and Diastolic pressure (unit: mmHg or kPa) SpO2: SpO2(unit: %) pulse rate (unit: beats/minute) RESP: respiration rate (unit: breaths/minute) TEMP: temperature of channel 1 and 2 (unit: or ) Et CO2: unit: mmHg or kPa

Menu The menu always occupies the fixed position on the bottom of the screen, user can set the system and perform operation. Frame of menu status Silence on/off: showing the status of silence on Alarm on/off: shows the status of alarm on or off

or off ,

Trend: the function is for operator observing the patients latest 72 hours change, including historical data table, HR, NIBP, SpO2, ST segment. System setup: you can configure various aspects of the monitor, including system time, simulation, print setup, color, display wave, language, color, etc.. Menu bar: it shows the different menus of every operation.

Other modes:IBP EXT:

MMED60000DP (S6) Service Manual

GRAPH EXT:

ECG EXT:

1.5.2

Function buttons

MMED60000DP (S6) Service Manual

1 Alarm ALARM button, pressing the button will prohibit all the technically audio alarm and physically audio alarm for 3 minutes. The audio alarm function will be restored automatically after 3 minutes or be activated when new alarm occurs. As pressing the

button, the icon in menu status shows have been shut off. NOTE When

which indicates all the audio alarms

mark appears, the system can not give out the audio alarm

prompt. Therefore, the operator should use this function carefully.2Silence SILENCE button, when pressing the button, this mark appears in the menu status, indicating that all kinds of sound including the audio alarm and heart beep have been manually muted until the button has been pressed again, the system will immediately restore the normal status, and the icon shows as

NOTE When pressing SILENCE, the system cannot give out the audio alarm and heart beep; therefore, the operator should be considerate in using the function.3Freeze FREEZE button, pressing the button can freeze the waveform of ECG. The screen displays two traces of ECG waveform. When one trace is active, pressing the FREEZE button will freeze another trace. Pressing FREEZE button again will restore the normal monitor status. 4Start/Cancel

MMED60000DP (S6) Service Manual

START/CANCEL BP measuring button, pressing the button will inflate the cuff to start a new NIBP measurement. When measuring, press it to cancel the measurement and deflate the cuff, including the CYCLE and STAT (NOT STOP) measuring mode. 5Print PRINT button, pressing this button will motivate the recorder or desktop printer to output the result if the monitor is equipped with them. (6) Menu Main MENU button, pressing the button will exit the submenu and refresh the screen. 7Rotary Knob The operator uses the rotary knob to select the menu item and modify the setup. It can be rotated clockwise or counter-clockwise and pressed like other buttons. The operator uses the knob to realize the operations.

The method of using the knob to execute the operation:

Fig. 1-4

Rotary knob is just like the cursor of computer. When operator rotates the knob on the icon where the operation is wanted, the icon will be automatic highlight. Then pressing the knob, operator will open the setup menu of the corresponding parameter so as to set up the menu. (8) Power Supply Indicator Indicating the power supplying state of the monitor Orange: 220Vac main power supplying Orange and flash: 220Vac main power supplying and battery charging Green and flash: internal battery power supplying and 50% power has been consumed.1.5.3. Side Panel (2) (1)

(3)

(4) (7)

(5)

(6)

Fig.1-5 (1) ECG socketconnect ECG 5-lead wire with 6 PIN connector

MMED60000DP (S6) Service Manual

(2) NIBP socketconnect NIBP cuff with extension tube (3)(4) IBP (IBP1, IBP2) socket: Single, Dual channel (optional) (5)(6) Temperature socket: connecting the TEMP probe (7)SPO2 socketconnect the SpO2 sensor with extension cable (8) Et CO2 socketconnect the nasal cannula with connector.

(8)

Fig.1-6 1.5.4 Recorder

Fig.1-7

Fig. 1-8 A thermal dot matrices recorder with standard 50mm (+1/-1) wide printout paper is used for MMED6000DP portable patient monitor. NOTE The thermal array recorder should be installed by the adequate technician.

1.5.5 Rear panel

MMED60000DP (S6) Service Manual

(1)

(2) (3) (5)

(6) (4)Fig.1-9

(1) Power switch (2) NET: Internet access point: connecting with the central monitoring system through the standard RJ-45. exteriorly connecting screen and standard VGA. (3) CRT: (4) Printer: exteriorly connecting printer, which type is designated by the manufacturer. (5) AC Power supply: 100-230 (VAC), 50/60(Hz); Fuse: standard T 1.6A (6) Equivalent electric ground access for connecting with the hospitals grounding system.

1.6 SpecificationEnvironmentThe operation environment should comply with the following conditions: Operating Temperature: 5 to 45 Relative Humidity: 0 to 80%, non-condensing Height of sea level: -500m ~ 4600m The transport and storage environment should comply with the following conditions: Storage Temperature: -20 to 70 Relative Humidity: 0 to 93%, non-condensing Height of sea level: -500 m~ 13100m

DisplayType: Screen Size: 640*480 pixel color TFT 10.4 inch diagonal

Displayed ParametersTime: Battery-backed quartz crystal clock

MMED60000DP (S6) Service Manual

Alarms: ECG: Heart Rate: NIBP: Pulse Oximeter: Respiration Rate: Trends: Temperatures: Trace Freeze

High and low limits selectable on patient parameters ECG Waveform Scale, displayed lead Derived from ECG OR Spo2 Pressure (systolic, mean and diastolic) Pulse Rate, Pulse waveform, and percent saturation. Respiration rate derived from ECG. Heart rate, respiration rate, NIBP (systolic, mean, diastolic), Temperatures and SpO2 Two channels Traces A or B

ECGProtected against defibrillator and electro surgery potentials Standard Lead IIIIIIaVLaVRaVFV1V2V3V4V5V6 Display Gain Scales 5mm/mV 7mm/mV 10mm/mV 15mm/mV 20mm/mV 25mm/m V Sweep Speed 12mm/s25mm/s50mm/s Input Resistance > 5M Ohm (at 10 Hz, not including patient cable) Frequency Response 0.05Hz-100Hz (3dB) Common Mode >60dB Rejection Ratio Electrode Offset Maximum 0.3V Potential Baseline Recovery