RAPIDEC® CARBA NP Test Instructions: Rehydration: 1. Remove the strip from its packaging 2. Open an ampule of API ® Suspension medium 3. Pipette 100 uL into wells a, b, c 4. Place an incubation lid on the test strip and leave for 4-10 minutes at room temperature (15-25°C) Lysis: 1. Gently mix the contents of well b using a stirring stick 2. Place the test strip on top of the two-colored (black and white) support card 3. Inoculate well c with colonies to be tested 4. Mix well and add colonies until the turbidity of well c matches well b 5. Place an incubation lid on the test strip and leave for 30 minutes at room temperature (15-25°C) Hydrolysis: 1. Transfer 25 uL from well c to wells d and e 2. Transfer 25 uL from well a to wells d and e 3. Place an incubation lid on the test strip and incubate for 30 minutes at 33-38°C Interpretation of Results: 1. At the end of the 30 minute incubation period, place the test strip on the two-colored (black and white) support card and remove the incubation lid to perform the initial reading. If a color change from red to yellow, light orange, orange or dark orange is observed in well e, the result is positive and the test is complete 2. If no color change is observed, continue the incubation at 33-38°C for up to 2 hours and perform a final reading Contents of the kit: RAPIDEC® CARBA NP test strips (10) API® Suspension medium 2 ml (10) Incubation lids (10) 1 pack of stirring sticks 1 two-colored (black and white) support Note: Package insert and reading guide are downloadable from the bioMérieux technical library. For more information, please visit our website: www.biomerieux-usa.com/rapidec ©2018 bioMérieux, Inc. • BIOMERIEUX, the BIOMERIEUX logo, API, and RAPIDEC CARBA NP are used pending and/or registered trademarks belonging to bioMérieux or one of its subsidiaries, or one of its companies. • Patents: www.biomerieux-usa.com/patents • PRN 17-0175-00 Same-day carbapenemase detection with results in 30 minutes to 2 hours. BIOMÉRIEUX RAPIDEC® CARBA NP CARBAPENEMASE DETECTION A simple and efficient test for carbapenemase detection in Enterobacteriaceae and Pseudomonas aeruginosa. RAPIDEC® CARBA NP Ref. 417824 To place an order, visit www.biomerieuxDIRECT.com bioMérieux, Inc. • 100 Rodolphe Street • Durham, NC 27712 Tel. : (800) 682-2666 • Fax : (800) 432-9682 www.biomerieux-usa.com BIOMÉRIEUX

Transcript of Rehydration - biomerieux-usa.com · BIOMERIEUX, the blue logo and RAPIDEC are used, pending, and/or...

Page 1: Rehydration - biomerieux-usa.com · BIOMERIEUX, the blue logo and RAPIDEC are used, pending, and/or registered trademarks belonging to bioMérieux, or one of its su bsidiaries or

RAPIDEC® CARBA NP Test Instructions:Rehydration:

1. Remove the strip from its packaging

2. Open an ampule of API® Suspension medium

3. Pipette 100 uL into wells a, b, c4. Place an incubation lid on the test strip and leave for

4-10 minutes at room temperature (15-25°C)

Lysis:

1. Gently mix the contents of well b using a stirring stick

2. Place the test strip on top of the two-colored (black and white) support card

3. Inoculate well c with colonies to be tested

4. Mix well and add colonies until the turbidity of well c matches well b

5. Place an incubation lid on the test strip and leave for 30 minutes at room temperature (15-25°C)

Hydrolysis:

1. Transfer 25 uL from well c to wells d and e

2. Transfer 25 uL from well a to wells d and e

3. Place an incubation lid on the test strip and incubate for 30 minutes at 33-38°C

Interpretation of Results:

1. At the end of the 30 minute incubation period, place the test strip on the two-colored (black and white) support card and remove the incubation lid to perform the initial reading. If a color change from red to yellow, light orange, orange or dark orange is observed in well e, the result is positive and the test is complete

2. If no color change is observed, continue the incubation at 33-38°C for up to 2 hours and perform a final reading

Contents of the kit:• RAPIDEC® CARBA NP test strips (10)

• API® Suspension medium 2 ml (10)

• Incubation lids (10)

• 1 pack of stirring sticks

• 1 two-colored (black and white) support

Note: Package insert and reading guide are downloadable from the bioMérieux technical library.

For more information, please visit our website: www.biomerieux-usa.com/rapidec

©20

18 b

ioM

érie

ux, I

nc. •

BIO

MER

IEUX

, the

BIO

MER

IEUX

logo

, API

, and

RAP

IDEC

CAR

BA N

P ar

e us

ed p

endi

ng a

nd/o

r reg

iste

red

trad

emar

ks b

elon

ging

to

bioM

érie

ux o

r one

of i

ts s

ubsi

diar

ies,

or o

ne o

f its

com

pani

es. •

Pat

ents

: ww

w.bi

omer

ieux

-usa

.com

/pat

ents

• P

RN 17

-017

5-00

Same-day carbapenemase detection with results in 30 minutes to 2 hours.

BIOMÉRIEUX

RAPIDEC® CARBA NP CARBAPENEMASE DETECTION

A simple and efficient test for carbapenemase detection in Enterobacteriaceae and

Pseudomonas aeruginosa.RAPIDEC® CARBA NPRef. 417824

To place an order, visit

www.biomerieuxDIRECT.com

bioMérieux, Inc. • 100 Rodolphe Street • Durham, NC 27712 Tel. : (800) 682-2666 • Fax : (800) 432-9682 www.biomerieux-usa.com

BIOMÉRIEUX

Page 2: Rehydration - biomerieux-usa.com · BIOMERIEUX, the blue logo and RAPIDEC are used, pending, and/or registered trademarks belonging to bioMérieux, or one of its su bsidiaries or

Enhance your antimicrobial stewardship program with same- day carbapenemase detection.

FDA 510(k) cleared RAPIDEC® CARBA NP is a phenotypic (in vitro) diagnostic test for the qualitative detection of carbapenemase enzymes in Enterobacteriaceae and Pseudomonas aeruginosa. Everything needed to perform the test is provided in an easy to use kit and interpretation of results is simply detecting a change in color on the test strip.

Well Reagentsa Phenol red solution

b Turbidity control

c Lysis buffer

d Control well without imipenem

e Reaction well containing imipenem

flattened

RAPIDEC® CARBA NP CARBAPENEMASE DETECTION

See Package Insert - 9310301 - A

417824-417825 21872 A - en - 2017/04

RAPIDEC® CARBA NP Reading Guide

Place the strip on the two-colored (black and white) support. Position wells and on the white background to facilitate reading. Remove the incubation lid. Reading is performed by comparing the colors in wells and , ensuring that the strip is firmly flatten against the two-colored (black and white) support. A test is positive when a significant variation in color is observed between the two wells. For correct interpretation, please refer to the following table.

Control well Test well Interpretation red red Negative

(absence of carbapenemase) orange orange red yellow, light orange, orange, dark orange Positive (presence of

carbapenemase) orange yellow any color other than red or orange Not applicable *Uninterpretable orange red

* An Uninterpretable result should be retested. If the retest yields an uninterpretable result, consider testing with an alternate method to determine carbapenemase statut of the isolate.

red yellow

red light orange

red orange

red dark orange

red red

orange red

Positive Negative Uninterpretable

BIOMERIEUX, the blue logo and RAPIDEC are used, pending, and/or registered trademarks belonging to bioMérieux, or one of its subsidiaries or one of its companies. Any other name or trademark is the property of its respective owner.

bioMérieux SA 376 Chemin de l'Orme 69280 Marcy-l'Etoile – France RCS LYON 673 620 399 Tel. 33 (0)4 78 87 20 00 Fax 33 (0)4 78 87 20 90 www.biomerieux.com

Distributed by bioMérieux, Inc. 100 Rodolphe Street Durham, North Carolina 27712 – USA www.biomerieux.com

Ready to use

Standardized protocol with internal control

Easy to perform

All-inclusive test kit

Results when you need them

In just 30 minutes to 2 hours