Software Revision 4 12 th December 2005 D Longstaff 1 of 27 REGARD 3900 Config Software User description To install the software on either a Laptop or Desktop PC follow the instructions given below:- Installing the Draeger Regard 3900 configuration program Program requirements: IBM PC/compatible with at least 128MB RAM Windows 2000 or Windows XP Screen resolution 1024 x 768 Mouse or other pointing device If you are installing the Regard 3900 configuration program on a PC that already has the program installed, please uninstall the existing program first. The program is distributed as a single installation file with the filename like Regard 3900_v1.exe or Regard 3900_v2d.exe. (You may not see the .exe extension on your system.) Double–click the installation file (eg, Regard 3900_v1). You will see… When you see this… …click Next >

Transcript of REGARD 3900 Config Software - Draeger Web...

Page 1: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 1 of 27

REGARD 3900 Config Software

User description To install the software on either a Laptop or Desktop PC follow the instructions given below:- Installing the Draeger Regard 3900 configuration program Program requirements: • IBM PC/compatible with at least 128MB RAM • Windows 2000 or Windows XP • Screen resolution 1024 x 768 • Mouse or other pointing device If you are installing the Regard 3900 configuration program on a PC that already has the program installed, please uninstall the existing program first. The program is distributed as a single installation file with the filename like Regard 3900_v1.exe or Regard 3900_v2d.exe. (You may not see the .exe extension on your system.) • Double–click the installation file (eg, Regard 3900_v1).

You will see…

When you see this…

…click Next >

Page 2: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 2 of 27

If you are installing the program on Windows XP, you may see a warning about Windows XP compatibility. The Regard 3900 program works fine on Windows XP, so click “Continue Anyway”:

When this appears…

…click Install. Installation should begin:

Page 3: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 3 of 27

..then:

If installation is successful, you will see a Regard 3900 program icon on the desktop:

Regard 3900.url

Regard 3900 Double–click the icon to start the program. Plug the Regard 3900 RS–232 interface into a serial port on your PC. The program automatically detects which serial port the interface is connected to — you don’t need to set anything. Issue 1 – October 2005

Page 4: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 4 of 27

How to use Config 3900 software

After double clicking the REGARD 3900 icon on the desktop the opening (default) screen is shown below.

Screen shot 1. The left window box shows the default setting of a control panel having four 4..20mA modules. The first module has its channels (CH1 to CH4) expanded. The right window box shows the details for channel one. At this stage the configuration PC is standalone and is not connected to an operational REGARD 3900.

Page 5: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 5 of 27

Along the top of the tool bar we see the four words File, Configure , Language and Help. These are explained as follows File – (see screen shot 2)

• Open – allows a preconfigured file to be opened from a storage device i.e. hard drive

• Save – allows the opened configuration to be saved but with the option to give a new file name, location etc

• Save As – allows the opened configuration file to be saved, but with the option to give a new file name, location etc

• Print – prints the configuration even if ‘cancel’ is selected • Print Setup – allows the printer options to be selected without printing • Exit – closes the program, but first prompts to Save if not already Saved. If no

alterations have been made then ‘Exit’ will not prompt ‘Save’.

Screen shot 2.

Page 6: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 6 of 27

Configure – (see screen shots 3, 4, 5 and 6) Note 1: A RS232 CAN interface cable (4208785) is required to connect the PC with the REGARD 3900

• Send PC configuration to REGARD 3900 – this function uploads the entire configuration from the PC to the REGARD 3900 control unit (see Note 1 above).

• Receive configuration from REGARD 3900 – this function downloads the entire configuration of the REGARD 3900 into the PC (see Note 1 above).

• Compare system configuration to program – this function allows a comparison between the PC configuration file and that of the installed REGARD 3900. The differences will toggle; Blue bold text indicates system configuration whereas Black bold text indicates PC configuration.

• New configuration – this function allows a new configuration to begin. A second pop-up window (see screen shot 4) will allow the choice of modules before configuration commences. No connection to a REGARD 3900 is required.

• Copy Gas Channels – this function allows any configured channel to be copied and pasted into another channel allocation (see screen shot 5). No connection to a REGARD 3900 is required.

• Add New Board – this function allows new modules to be configured and the configuration stored ready for uploading into a REGARD 3900 (see screen shot 6).

Screen shot 3.

Page 7: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 7 of 27

Module 1 and 2 allows either a 4..20mA module or None installed.

Screen shot 4. Module 3 and 4 allows either a 4..20mA module, relay module or None installed.

Page 8: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 8 of 27

From the drop down menu select the channel which is to be copied.

Screen shot 5. Select which channels the copied configuration is to be pasted. More than one channel can be selected at any one time. The RED square indicates where it is not possible to copy channel configuration too.

Page 9: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 9 of 27

Screen shot 6. When this function is chosen there must be a spare capacity within the configuration file or REGARD 3900. Otherwise the drop down menus will be ‘greyed out’ to represent no further configuration possible. If spare capacity exists within the configuration file then refer to screen shot 4 above and follow the instruction given there.

Page 10: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 10 of 27

Language -

• English – shows English text

• German – shows German text

Screen shot 7.

Page 11: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 11 of 27

Help – Modules – this allows the information supporting the module type & revision and interface cable type & revision to be displayed (see Note 1 above). Program – this allows the information supporting the Config 3900 software to be displayed

Screen shot 8.

Page 12: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 12 of 27

The following pages explain the function and configuration of each input parameter.

Screen shot 9.

Unique channel tag. Used in the Config. 3900 software only. Max of 16 characters

Indicates active channels within the system. Check R box to make channel active.

Displays which channel is being configured or viewed

Drop down menu for gas name selection. Or type unique gas name to a max of 6 characters

Explorer menu showing module types

Click to send individual channel configuration to the REGARD 3900

Click to receive individual channel configuration from the REGARD 3900

Page 13: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 13 of 27

Screen shot 10.

Drop down menu for gas measurement units. Or type unique measurement unit to a max of 4 characters

Fault (F1). Enter value for fault threshold between 1mA and +3.8mA. This threshold trips the Fault relay on each input module and any subsequent configured relay (F1) on a relay module.

Fault (F2). Enter value for fault threshold between 1mA and +3.8mA. This threshold will only trip F2 relays configured on a relay module. A value of 0 (zero) switches this function ‘off’.

Page 14: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 14 of 27

Screen shot 11.

Drop down menu for gas (FSD) measurement range. No custom measuring ranges are allowed.

Page 15: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 15 of 27

Screen shot 12.

Option to set whether the alarm thresholds are rising or falling per alarm threshold

Enter A1 threshold between 4% and 100% of measuring range. Up to 2 decimal places are allowed subject to measuring range

Enter A2 threshold between 4% and 100% of measuring range. Up to 2 decimal places are allowed subject to measuring range

Enter A3 threshold between 4% and 100% of measuring range. Up to 2 decimal places are allowed subject to measuring range. A value of 0 (zero) switches this function

Page 16: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 16 of 27

This page shows the common functions of the 4 channel input module.

Screen shot 13.

This screen shows the common functions of the 4 channel input module

Option to set whether the alarm thresholds are Latching or Non Latching per alarm threshold

Click to send entire module configuration to the REGARD 3900

Click to receive entire module configuration from the REGARD 3900

Page 17: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 17 of 27

Adding into the system two relay boards (using the ‘New Configuration’ function) the following pages explain the function and configuration of each relay parameter.

Screen shot 14. The Explorer menu now shows two relay modules. Each relay module is expanded to show relays 1 to 16. Relay 1 has been selected. The first relay of each relay module is not configurable, therefore remains as a fail-safe, energised relay.

No configuration allowed for the first relay of each relay module.

Page 18: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 18 of 27

Relay 2 has been selected and its configuration is displayed as ‘Common A1’. The A1 is selected from the drop down menu.

Screen shot 15. Common alarm means that any alarm in the system at the selected threshold will trip the selected relay.

Drop down menu for threshold selection

Click to send individual relay configuration to the REGARD 3900

Click to receive individual relay configuration to the REGARD 3900

Page 19: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 19 of 27

Relay 3 has been selected and its configuration is displayed as ‘Single A1’. The A1 is selected from the drop down menu. For single alarms to operate it is necessary to choose a single channel number. To do this select the channel number under the heading ‘channel’.

Screen shot 16. It is only possible to select a single channel from those which are not ‘greyed out’. The software automatically recognises the total number of available inputs.

Drop down menu for threshold selection

Select individual channel for single operation

Page 20: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 20 of 27

Relay 4 has been selected and its configuration is displayed as ‘Voted A1’. The A1 is selected from the drop down menu. The following information is required to set a voting configuration

• Threshold. • Total number of channels in the voting group. • The channel numbers of the channels in the voting group. • The number of channels to satisfy the vote from within the voting group.

Screen shot 17.

Drop down menu for threshold selection

Select channel identity for those channels in the total voting group

Total number of channels in the voting group

Number of channels to satisfy the vote from within the voting group

Page 21: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 21 of 27

Relay 6 has been selected and its configuration is displayed as ‘No Function’. No configuration is therefore needed. The relay 6 selection is ‘greyed out’.

Screen shot 18.

No configuration options

Relay 6 ‘greyed out’.

Page 22: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 22 of 27

Relay 8 has been selected and its configuration is displayed as ‘Alarms Inhibited. No configuration is therefore needed.

Screen shot 19. Relay 8 will operate whenever the manual switch on the display board is switched to Inhibit. This gives warning that the control panel cannot generate gas alarms.

No configuration options

Page 23: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 23 of 27

This page shows the ability to configure each individual relay, subject to it’s configuration as being

• Energise on alarm or Normally energised (failsafe) • Latching or Non Latching • Acknowledgeable or Non Acknowledgeable

Screen shot 20.

Select either Energise on alarm or Normally energised

Select Latching or Non Latching

Select Acknowledgeable or Non Acknowledgeable

Page 24: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 24 of 27

This page shows the common functions of the relay output module.

Screen shot 21.

Click to send entire module configuration to the REGARD 3900

Click to receieve entire module configuration from the REGARD 3900

Page 25: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 25 of 27

Warning and Fault messages.

This message is displayed if trying to make a communication between the REGARD 3900 and a PC without a RS232 CAN interface cable connection.

This message is displayed when trying to set a gas alarm threshold outside of the limits.

This message is displayed if no gas alarm threshold limit is entered.

This message is displayed when trying to set the fault threshold outside of the limits.

This message is displayed when a zero is not used before the decimal point.

Page 26: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 26 of 27

This message is displayed when a value has not been entered e.g. F2

This message gives a warning and a reminder of the constraints of BSEN 61779-1.

This message is shown when choosing a single relay function but without a channel selection (relay module).

This message is shown when the voting details are incomplete. The correct number of ‘target’ channels have not been selected.

This message is displayed when trying to make a communication connection between the REGARD 3900 and PC.

Page 27: REGARD 3900 Config Software - Draeger Web Apptoxicgasdetection.co.uk/products/regard3900/documents...Software Revision 4 12th December 2005 D Longstaff 6 of 27 Configure – (see screen

Software Revision 4 12th December 2005

D Longstaff 27 of 27

This message is displayed as a reminder to save the configuration.