Rapid Resolution
High Definition
LC Columns
New Columns for Use with
the 1290 Infinity LC
Edward Kim
Application Engineer
August 11, 2010
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 2
What are Rapid Resolution High Definition (RRHD)
Columns?
RRHD columns are the columns designed to be compatiblewith the 1290 Infinity LC – and be backwards compatible to other Agilent instruments with lower pressure limits (though not the first choice) and fully UHPLC compatible
RRHD columns can be used up to 1000-1200 bar and represent a complete Agilent solution with the 1290 LC.
They are 1.8um ZORBAX columns packed to withstand the higher pressures of the 1290 LC.
Like other ZORBAX columns we will maximize compatibility across instruments and labs by maintaining scalabilityacross particle sizes.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 3
What is Different Between RRHT 1.8um Columns
and RRHD 1.8um Columns
The pressure limit of the column
• RRHD columns are designed for 1000-1200 bar use, RRHT columns are designed for 600 bar use on the Rapid Resolution LC (1200 SL)
• RRHD columns use improved hardware and packing processes
– the packing pressure and process are proprietary
– improved hardware is being transferred to RRHT columns (for consistency)
This impacts the backpressure the column
1. These columns operate at a higher backpressure than a 600 bar column – higher pressure packing = higher packing density, more backpressure
2. A 50mm length RRHD can not be used on a 400 bar instrument but can be used on a 1200 RRLC (600 bar instrument)
3. The 100 and 150mm length columns are only recommended for use on the 1290 Infinity LC (or UHPLC instruments)
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 4
LC Beyond 600 bar Requires New RRHD Columns
Agilent ZORBAX RRHT Eclipse Plus C18, 2.1 x 100 mm, 1.8 um; PN 959764-902,
SN USUXU01664, LN B08101
y = 935.21x + 7.1903
R2 = 0.9993
y = 1025.3x + 13.279
R2 = 0.9973
y = 1817.7x + 21.259
R2 = 0.9946
0
100
200
300
400
500
600
700
0 0.1 0.2 0.3 0.4 0.5 0.6
Flow Rate (mL/min)
Pre
ssu
re (
bar
)
Acetonitrile/Water(60:40)
Acetonitrile/Water(50:50)
Methanol/Water(60:40)
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 5
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 5
Comparison of Pressure Generated on RRHT and
RRHD 1.8um Columns
RRHT 1.8um 600 bar RRHD 1.8um 1200 bar
2.1 x 50mm 292 bar 350 bar
2.1 x 100mm 452 bar 566 bar
2.1 x 150mm 660 bar 750 bar
The RRHD columns are ideal for use on the 1290 Infinity LC or on other
UHPLC instruments.
To withstand maximum pressure requires a denser packed bed that
generates a little more backpressure.
This is necessary for optimum lifetime and use up to 1200 bar.
Change in pressure may vary a little until optimized columns ship!
40:60 water:acetonitrile, Flow Rate: 0.5 mL/min, ambient temperature
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 6
RRHT, 600 bar and RRHD Column Comparison
Blue = RRHT, Red=RRHDColumns: 2.1 x 50mm, 1.8um Flow: 0.6 mL/min Sample: Alkylphenones
Expected Differences due to higher packing density:
RRHD – longer retention, same selectivity
Higher Back Pressure of RRHD = 434 bar
vs. Back Pressure of RRHT = 323 bar
N = 11,041N=11,168
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 7
Gain N and Rs w/ Densely Packed RRHD columns
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 7
Rs3,2: 4.16
N6 : 9200
Rs3,2: 10.71
N6 : 33300
Rs3,2: 8.38
N6 : 18700
RRHT 1.8um 600 bar RRHD 1.8um 1200 bar
Mobile Phase: 40:60 water:acetonitrile, Flow Rate: 0.5 mL/min, Temp: ambient, Columns: RRHD/RRHT SB-C18, 2.1 mm ID,
Lengths as noted Peaks: 1: uracil, 2acetophenone, 3 propiophenone, 4 butyrophenone, 5 valerophenone, 6 hexanophenone
1
2
6
54
3
Rs3,2: 5.64
N6 : 10700
Rs3,2: 11.98
N6 : 34100
Rs3,2: 9.02
N6 : 21100
50 mm
100 mm
150 mm
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
min0 1 2 3 4
min0 1 2 3 4
min0 1 2 3 4
0
The additional packing results in more retention, then Rs and efficiency.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 8
Optimum Flow Rates and Efficiencies-Van Deemters
John W. Henderson JrGroup/Presentation
Title
Agilent Restricted
0
2
4
6
8
10
12
14
0 1 2 3 4 5 6 7 8
h
u (mm/s)
DAZ01002
LFS Column
performance
report results
N=20600
TF 1.02
P=324
F=0.3
RRHD Eclipse Plus C18, 2.1 x 100 mm, 1.8 um;
PN 959758-902, SN DAZ01002, LN B009013
Naphthalene k=4.0
RRHT Eclipse Plus C18, 2.1 x 100 mm, 1.8 um;
PN 959764-902, SN USUXU01664, LN B08101
Heptanophenone k=10.8
N=24900
TF 1.08
P=520
F=0.489
N=24900
TF 1.08
P=550
F=0.520
N=22300
TF 1.01
P=535
F=0.571
2.1 x 100 RRHD Optimum flow rate is about 0.5 mL/min. (60% MeCN, Themostatted 26°C, ~LC Performance Report
Conditions)
Note pressure of more than 500 bar on 100mm length!
With just acetonitrile and water – new 1290 Infinity LC ideal
Van Deemters are very similar, this is expected.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 9
Why Is It Important to Know the Optimal Flow Rate?
Efficiency is Highest at Optimal Flow Rate
John W. Henderson JrGroup/Presentation
Title
Agilent RestrictedMonth ##, 200X
min0.5 1 1.5 2 2.5 3
3.5
min0.25 0.5 0.75 1 1.25 1.5 1.75
2
DAZ01002
LFS Column
performance
report results
N=20600
TF 1.02
P=324
F=0.3
DAZ01002
N=25100
TF 1.08
P=550
F=0.520
Sub-Optimal Optimal
0.3 mL/min 0.5 mL/min
N= 20,600 N= 25,100
P= 324 bar P = 550 bar
At sub optimal flow rates the efficiency is lower and customers may
not see the benefits of a 1.8um column or UHPLC technology.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 10
What is Different Between RRHT 1.8um Columns
and RRHD 1.8um Columns
The column hardware – on the inside
• We redesigned the hardware to improve available frit surface area to reduce the likelihood 2.1 mm ID columns will plug
• We increased the frit pore size to 0.5um to also reduce the likelihood a column will plug
The column label – on the outside
• We now use a colored label so customers can visually tell the difference between a 600 bar RRHT column and new RRHD column
Available column dimensions
1. Initially the columns will only be available in 2.1 mm ID
2. We will test 3.0 mm ID
3. 4.6 mm ID will not be available due to technical limitations in this configuration, other 4.6 mm ID with lower pressure limits can be used on the 1290 Infinity LC with appropriate pressure limits entered
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 11
Examples of the RRHT and RRHD Columns
RRHT columns show this groove in the column between the end of the
fitting and the column body. They also have a white label.
RRHD columns have a light blue label
RRHD columns have no visible groove on the column between
end of the fitting and the column body.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 12
What is Different Between RRHT 1.8um Columns
and RRHD 1.8um Columns
The part numbers!
• These new columns will have new part numbers (table below)
• The column shipping with the 1290 Infinity LC will be the ZORBAX Rapid
Resolution High Definition (RRHD) Eclipse Plus C18, 2.1 x 50mm, 1.8um
PN Description
959759-902 RRHD Eclipse Plus C18, 2.1 x 150mm, 1.8um
959758-902 RRHD Eclipse Plus C18, 2.1 x 100mm, 1.8um
959757-902 RRHD Eclipse Plus C18, 2.1 x 50mm, 1.8um
859700-902 RRHD SB-C18, 2.1 x 150mm, 1.8um
858700-902 RRHD SB-C18, 2.1 x 100mm, 1.8um
857700-902 RRHD SB-C18, 2.1 x 50mm, 1.8um
859700-906 RRHD SB-C8, 2.1 x 150mm, 1.8um
858700-906 RRHD SB-C8, 2.1 x 100mm, 1.8um
857700-906 RRHD SB-C8, 2.1 x 50mm, 1.8um
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 13
What’s Not Different Between the RRHT 1.8um and
RRHD 1.8um Columns
The ZORBAX packing materials
• These are still the same ZORBAX 1.8um packing materials with the same
bonded phases - Eclipse Plus, StableBond etc.
• The packing material still has the same particle size distribution – the
change in pressure is only related to change in packing density, not particle
size distribution
– We could have changed it, but there was no performance advantage
• We will start with our most popular C18 phases – Eclipse Plus C18 and
StableBond C18
– Eclipse Plus delivers better peak shape than the BEH Hybrid material
– StableBond C18 is still the best column for low pH lifetime
– ZORBAX columns can deliver excellent performance
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 14
Updated Peak Shape Comparisons – Why Compare
Chromatograms with Dextromethorphan?
Dextromethorphan is a strongly basic compound that generates
the best results in LC at low pH.
• Eclipse Plus claims superior peak shape so comparisons for peak shape
are most relevant
Peak shape is a critical issue and poor peak shape is common
with this compound.
• Peak shape continues to be a major driver in column choice for customers
Because of this, efficiency can also be low on columns so it is a
critical parameter to monitor.
• High efficiency and good peak shape provide the most quantitative
accuracy
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 15
Peak Shape of Comparison of sub 2 um columns
Eclipse Plus is Superior to Acquity BEH C18
min1 2 3 4 5 6
5.837
min1 2 3 4 5 6
5.503
ACQUITY BEH C18
2.1 x 50 mm, 1.7 um
P = 228 Bar N = 10107,
Peak Width = 0.137 min,
Tf= 1.39
RRHT Eclipse Plus C18
2.1 x 50 mm, 1.8 umP = 202 Bar N = 10881,
Peak Width = 0.124 min,
Tf= 1.23
Mobile Phase = 80% 25 mM phosphate (pH 3): 20% ACN, Flow Rate = 0.3 mL/min Detection: 205nm
Analyte: Dextromethorphan
Tf = 1.4
Tf = 1.2
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 16
Peak Shape Comparison of 2.5 & 1.8 um ColumnsEclipse Plus Delivers Peak Shape and Efficiency
min1 2 3 4 5 6 7 8 9
5.08
min1 2 3 4 5 6 7 8 9
6.73
min1 2 3 4 5 6 7 8 9
6.90
min1 2 3 4 5 6 7 8 9
5.50
XBRIDGE C18 2.1 x 50 mm, 2.5 umP = 134 Bar N = 6094, PW = 0.153 min, TF= 1.92
Luna HST C18 2.0 x 50 mm, 2.5 um P = 128 Bar N = 4953, PW = 0.225 min, TF= 1.09
Shim Pak XR C18 2.0 x 50 mm, 2.5 um P = 159 Bar N = 6940, PW = 0.195 min, TF= 1.42
Eclipse Plus C18 2.1 x 50 mm, 1.8 umP = 202 Bar N = 10881, PW = 0.124 min, Tf = 1.23
Mobile Phase = 80% 25 mM phosphate (pH 3): 20% ACN, Flow = 0.3 mL/min (2.1mm), 0.27 mL/min (2.0mm ID)
Detection: 205nm Sample: Dextromethorphan
Acquity and XBridge are the same packing material, yet we see worse performance on
the XBridge – maybe not as optimized packing process as for Acquity?
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 17
Comparing RRHD to Competition on 1290 Infinity
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 17
min0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8
mAU
0
50
100
150
200
250
300
350
min0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8
mAU
0
50
100
150
200
250
300
350
min0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8
mAU
0
50
100
150
200
250
300
350
α3,2: 1.36
Rs3,2 5.64
N6 : 10700
TF6: 1.04
α3,2: 1.22
Rs3,2 3.69
N6 : 9400
TF6: 1.17
α3,2: 1.29
Rs3,2 4.58
N6 : 11000
TF6: 1.14
RRHD SB-C18 2.1 x 50mm, 1.8um USCES1057
Acquity HSS C18 SB
2.1 x 50 mm 1.8um
Acquity BEH C18
2.1 x 50 mm, 1.7um
1
2
6
54
3
Mobile Phase: 40:60 water:acetonitrile, 0.1% Formic Acid Flow Rate: 0.5 mL/min, Temp: ambient, Det: UV 254 nm Columns:
RRHD SB-C18, 2.1 x 50mm, 1.8um and competitive Peaks: 1: uracil, 2acetophenone, 3 propiophenone, 4 butyrophenone, 5
valerophenone, 6 hexanophenone
StableBond is still the best column for low pH lifetime on the market!
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 18
These Columns Are Now Available!
Customers receiving beta test or pre-release LC’s can receive beta test or pre-release columns
• These customers are a critical part of our beta testing program and we want feedback from all of them
• Feedback to date has been excellent – though mostly verbal
A RRHD column will ship with every 1290 Infinity LC
• Part number 959758-902 – RRHD Eclipse Plus C18, 2.1 x 50mm, 1.8um
• This allows quick evaluation of performance and uses our best performing bonded phase for peak shape – Eclipse Plus
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 19
The Technical Details….
Status Report
Currently we have made more than 100 columns available in a working hardware design
We are finalizing hardware design – these are small details on the internal part of the hardware and impact nothing visible to the customer
Every column configuration is tested multiple times to a minimum of 5000 injections to be considered optimized
We have good packing processes for some of the phases that meet these requirements (one shown earlier)
We will continue to optimize the packing process for the multiple bonded phase and configurations so that they are ready for introduction
We will continue to add part numbers and phases each month after initial intro – but this will be driven by requests/customer needs and sales
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 20
RRHD Columns Tested to Minimum Lifetime of
5000 Injections at 1000 Bar
RRHD 2.1 x 100mm, 1.8um
2157622200
2120521480
0
500
1000
1500
2000
0 1000 2000 3000 4000 5000 6000
Injections
Bar
0
5000
10000
15000
20000
25000
N
Pressure
QC Plates
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 21
Overlay of First and Last Injections of QC Test for
Lifetime of RRHD 2.1 x 100mm, 1.8um Column
min0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
5
10
15
20
25
30
35
40
Inj #1 = N4= 21,480
Inj #5000 N4= 21,205
Column QC Test: Mobile Phase: ACN:water Flow Rate: 0.3 mL/min Sample: Uracil, Phenol, 4-
Chloronitrobenzene, Naphthalene
The QC chromatogram shows consistent performance over the 5000 injections.
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 22
Pressure of RRHT Column with New Hardware
200.0
225.0
250.0
275.0
300.0
325.0
350.0
375.0
400.0
425.0
450.0
475.0
500.0
525.0
550.0
575.0
600.0
0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000
Pressure
Linear(Pressure)
Column Test of RRHT Column with new RRHD
Type Hardware
Consistent pressure for over 5500 injections (test ended).
Trend line is straight!Conditions: Mobile Phase: 70% Methanol/30% water Flow Rate: 2mL/min flow Temp: RT
Detection: 254nm, VWD Injection volume: 2ul Sample: QC sample 1B - Uracil, phenol,
4-Cl nitrobenzene, toluene Run time 1.1 minutes
Column: RRHT Eclipse Plus C18, 1.8um
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 23
How Do They Compare On Performance….
Use and Lifetime at 1000 bar or more…
• Current feedback from beta testers is excellent
• Roche feedback indicated use at over 1000 bar with continued good
performance
• Column was taken above 1000 bar though focus is on 1000 bar for
routine use and higher pressure as the “room” left for pressure
changes, shifts due to gradients etc.
• We are expecting more data to be available very soon in this area
and we will update as it is available
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 24
Gain Efficiency and Resolution with Long Length
150 +100 mm = 250 mm 1.8um @ 1000 bar
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 24
α3,2: 1.15
Rs3,2: 6.75
N6 : 49200Pw2: 0.017 min
Pw6: 0.029 min
α3,2: 1.15
Rs3,2: 4.92
N6 : 29600Pw2: 0.015 min
Pw6: 0.023 min
Mobile Phase 20:80 water:acetonitrile, F=0.5 mL/min, ambient temperature,
ZORBAX RRHD SB-C18, 1.8um columns
150 mm (P:560bar)
150 +100 mm (P: 960 bar)1. Uracil
2. Acetophenone
3. Propiophenone
4. Butyrophenone
5. Valerophenone
6. Hexanophenone
1
2
6
5
43
1
min0 0.5 1 1.5 2 2.5 3 3.5
mAU
0
50
100
150
200
250
300
min0 0.5 1 1.5 2 2.5 3 3.5
mAU
0
50
100
150
200
250
300
Highest Rs in less than 3 minutes, 250mm, 1.8um at 1000 bar
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 25
High Resolution Separation with Long 1.8um
Columns
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 25
min0 2 4 6 8 10 12 14
mAU
0
50
100
150
200
DAD1 A, Sig=254,4 Ref=off (E:\090422JWH\PRESSURE000027.D) PMP1, PMP1B, Pressure
Mobile Phase A: water B:acetonitrile, F=0.25 mL/min, ambient temperature,
Columns: ZORBAX RRHD, SB-C18, 2.1 x 150+100 mm, 1.8um
Gradient 10-100% B/15 min.
Pressure 800- 900 bar
Low point of gradient about 480 bar
Long column or columns in series provide max Rs
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 26
Fast Analysis on ZORBAX RRHD Eclipse Plus C18
min0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8
mAU
0
100
200
300
400
500
600
DAD1 D, DAD1C, DAD: Signal C, 280 nm/Bw:10 nm Ref 400 nm/Bw:50 nm (G:\1290\AOX JASMIN\AOX-AM2-45C-101.D)
0.2
72
0.4
15
0.4
80
0.7
57
0.8
83
1.0
11
1.0
55
1.5
24
1.5
56
DAD1 E, DAD1D, DAD: Signal D, 255 nm/Bw:10 nm Ref 400 nm/Bw:50 nm (G:\1290\AOX JASMIN\AOX-AM2-45C-101.D)
1.6
84
Eclipse Plus C18, 1.8um,
Pmax 1070 bar
PG
THBP
TBHQNDGA
BHA
Ionox-100
OGDG
BHT AP
Data Courtesy of: Gerd Vanhoenacker
Research Institute for Chromatography
Kennedypark 26 8500 Kortrijk Belgium
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 27
Fast Analysis on Agilent Zorbax Rapid
Resolution High Definition Eclipse Plus C18
MethodColumn ZORBAX RRHD Eclipse Plus C18 2.1 x 50 mm, 1.8 µm
(SN USDAY01001)Mobile phase A=0.02% H3PO4, B=acetonitrile/methanol 3/1Flow rate 1.6 mL/minGradient 0-1.85 min 30-100% BInjection volume 1 µLDetector Sig=280/10 nm, Ref=400/50, 80 Hz
Sig=255/10 nm, Ref=400/50, 80 HzTemperature 45°CSample Antioxidant mixture, ca. 100 ppm
Data Courtesy of: Gerd Vanhoenacker Research Institute for Chromatography
Kennedypark 26 8500 Kortrijk Belgium
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 28
3 minute analysis, Zorbax Eclipse + C18 2.1x100mm, 1.8umSeiko TOF, 100pg on-column, 10 Hz Acquisition.162/220 Pesticides Identified using TOF,
Customer Example Generated on 1290 with RRHD
Eclipse Plus C18
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 29
What Applications Are Planned on the RRHD
Columns? – CSD Led Information
Note: A consolidated applications plan is available and covered
in another presentation
Applications Categories for CSD
1. Proof of Performance
1. Lifetime, Efficiency
2. Resolution/Peak Capacity Message
1. Resolution per time based on improving resolution with different
gradient times and column lengths
3. Comparisons to Acquity
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 30
CSD Initial Applications Plan for RRHD Columns
General App Samples Columns
Lifetime at 1000 and 1100 bar Formic acid mp, dirtier sample All
Efficiency -simple tests of each dimension QC and Waters acid base neutral mix
2.1 x 50mm,
2.1 x 100mm
Van Deemters, consideration of frictional
heating, limited to 2.1 mm and 3.0 mm ID? Phenones - RRLC checkout sample
2.1 x 50mm,
2.1 x 100mm
Efficiency - 2 columns in series Neutrals - and complex pharma mix 2.1 x 100mm
Comparisons to Acquity Widest range of sample types, show off selectivity
2.1 x 50mm,
2.1 x 100mm
Demonstrate use on an Acquity system -
1000 bar means any 1000 bar LC (Prove
fittings work appropriately)
Duplicate Acquity app - target big pharma
examples so that we can sell columns where they
already have experience
2.1 x 50mm,
2.1 x 100mm
Efficiency - 2.1 x 150mm in 10/20/30
minute increments - demonstrate a short
and long run with efficiency/Rs per time
Several samples of different complexity - peptide
map (most complex), pharma sample with
impurities, sample with at least 5 components with
at least one closely eluting 2.1 x 150mm
Group/Presentation Title
Agilent Restricted
Month ##, 200XPage 31
Summary
New ZORBAX RRHD – Rapid Resolution High Definition –
LC columns for the 1290 Infinity LC are designed for the
demands of this UHPLC and any other UHPLC.
RRHD columns are made from the same ZORBAX 1.8um
particles and bonded phases in the ZORBAX RRHT columns
for the RRLC.
RRHD columns deliver long lifetime and high resolution over
an expanded performance range compared to the RRHT
columns
Practically speaking they can go to the high pressure limits
with new hardware design and packing processes.