Quality Control
![Page 2: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/2.jpg)
HIGH QUALITY FILMS
REPRODUCES ANATOMICAL IMAGE EXACTLY
FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY
![Page 3: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/3.jpg)
RESOLUTION
ABLE TO VISUALLY SEPARATE OBJECTS
BONE VS. TISSUE-- IE: SPATIAL RESOLUTION CR VS CONVENTIONAL FILM
DIFFERENCES BETWEEN DIFFERENT TISSUES-- IE: CONTRAST RESOLUTION CR VS. CONVENTIONAL FILM
![Page 4: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/4.jpg)
NOISE
FILM GRAININESS (INHERENT) SIZE OF FILM CRYSTALS
STRUCTURE MOTTLE (INHERENT), PHOSPHOR IN SCREEN
QUANTUM MOTTLE
![Page 5: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/5.jpg)
QUANTUM MOTTLE
LOW NUMBER OF X-RAY PHOTONS INTERACTING WITH SCREEN. INSUFFICIENT AMOUNT OF PHOTONS STRIKING INTENSIFYING SCREEN CAUSING AN INSUFFICIENT LIGHT TO EMIT FROM PHOSPHOR CRYSTALS.
How do we adjust technique for this? CR and Quantum Mottle
![Page 6: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/6.jpg)
SPEED
FILM’S ABILITY TO RESPOND TO X-RAY EXPOSURE
IF FILM “A” RECORDS MORE DENSITY SOONER THAN FILM “B”, WHICH FILM HAS A HIGHER SPEED?
A
![Page 7: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/7.jpg)
SENSITOMETRY
MEASUREMENT OF SENSITIVITY TO LIGHT AND THE FILM’S RESPONSE TO CHANGES IN TECHNIQUE AND PROCESSING.
![Page 8: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/8.jpg)
DENSITOMETRY
MEASUREMENT OF THE DENSITY OF THE FILM BASED ON THE PERCENTAGE OF LIGHT TRANSMITTED THROUGH THE FILM.
![Page 9: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/9.jpg)
ANATOMY OF A CHARACTERISTIC(H & D CURVE)
![Page 10: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/10.jpg)
![Page 11: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/11.jpg)
OVERHEADS!!!!!!!
![Page 12: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/12.jpg)
![Page 13: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/13.jpg)
RECIPROCITY LAW
OD REMAINS THE SAME NO MATTER THE LENGTH OF EXPOSURE
100 mA X 1 sec = 100 mAs ↓↓
200 mA X .5 sec= 100mAs ↓↓
400 mA X .25 sec =100mAs →→same OD 500 mA X .2 sec = 100mAs ↑↑ Applicable to direct screen exposures
![Page 14: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/14.jpg)
RECIPROCITY LAW FAILURE
FILM/SCREEN RADIOGRAPHY EXPOSURE TIMES LESS THAT 10
MILLISECONDS EXPOSURE TIMES GREATER THAN
5 SECONDS INTERVENTIONAL STUDIES MAMMOGRAPHY
![Page 15: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/15.jpg)
RADIOGRAPHIC CONTRAST
IMAGE RECEPTOR CONTRAST
PLUS
SUBJECT CONTRAST
![Page 16: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/16.jpg)
IMAGE RECEPTOR CONTRAST
Formerly known as film contrast. Why the change in name?
Inherent to film Affected by processing
![Page 17: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/17.jpg)
SUBJECT CONTRAST
SIZE SHAPE X-RAY ATTENUATION OF THE
ANATOMY AND ENERGY OF BEAM
![Page 18: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/18.jpg)
REVIEW
L ATITUDE (REMEMBER THIS IN REFERENCE TO THE FILM)
W I D E LATITUDE = LOOOOONNNNG GRAY AKA LOW OR LONG SCALE CONTRAST
NARROW LATITUDE = SHORT SCALE =HIGH CONTRAST
![Page 19: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/19.jpg)
WIDE LATITUDE FILM, BECAUSE OF LONG GRAY SCALE ALLOWS AS MUCH AT 15% ERROR BY RADIOLGRAPHER
![Page 20: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/20.jpg)
MAGNIFICATION FACTORS
MOST IMAGES ARE LARGER ON RADIOGRAPH THEN IN ACTUALITY DUE TO SID AND SOD!
FOR EVERY 1” INCREASE IN OID, SID MUST BE INCREASED BY 7” TO COMPENSATE FOR MAGNIFICATION
![Page 21: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/21.jpg)
SID AND SOD?
SID: SOURCE TO IMAGE DISTANCE
SOD: SOURCE TO OBJECT DISTANCE
MAG. FACTOR IMAGE SIZE OR SID OBJECT SIZE SOD
TO FIND SOD SUBTRACT THE OID FROM THE SID
![Page 22: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/22.jpg)
IF YOU COULD MEASURE A HEART AND IT MEASURED 12.5 CM BUT ON THE RADIOGRAPH IT MEASURED 14.7 CM, THE MAGNIFICATION FACTOR WOULD BE CALCULATED AS 14.7 DIVIDED BY 12.5 WHICH EQUALS A MAGNIFICATION FACTOR OF 1.17.
![Page 23: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/23.jpg)
WAYS TO DECREASE MAGNIFICATION
IF OID IS INCREASED? INCREASE SID! CHEST LATERAL AND OBLIQUE C-SPINE
![Page 24: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/24.jpg)
DISTORTION
Foreshortening Object at an angle and CR is perpendicular
to the part that is angled-trauma elbow Spatial
Tube is angled to separate body parts Clavicle
Elongation object is parallel but tube is angled sigmoid
![Page 25: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/25.jpg)
FOG LEVEL/CONTRAST AND THE PROCESSOR
As time extends in the developing process the fog level____(toe of the characteristic curve) and the contrast____
increases decreases
![Page 26: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/26.jpg)
90 seconds is the optimal developer time
A variation of 5 seconds either way can affect quality of film.
![Page 27: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/27.jpg)
Developer temperature
As the temperature increases the fog level______
increases optimum temperature is usually 90
degrees F.
![Page 28: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/28.jpg)
SUBJECT CONTRAST
Review: Rad. Contrast=film contrast plus
subject contrast film contrast is controlled latitude of
film, and processing factors
![Page 29: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/29.jpg)
Subject contrast
Patient (Size) Tissue (KUB) Atomic # ( higher
produces high contrast)
object shape
![Page 30: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/30.jpg)
MOTION
PATIENT TUBE MOVEMENT
TOMOGRAPHY
![Page 31: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/31.jpg)
MOTION
LARGE SID = LESS FOCAL SPOT BLUR
SMALL SID = MORE FOCAL SPOT BLUR
![Page 32: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/32.jpg)
FOCAL SPOT BLUR
![Page 33: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/33.jpg)
KVP
GIVES ELECTRONS ENERGY TO HIT THE ANODE
HIGH AMOUNT OF KINECTIC ENERGY IN AN ELECTRON PRODUCES A HIGHLY ENERGIZED X-RAY PHOTON
BETTER PENETRATION=LOTS OF GRAYS
![Page 34: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/34.jpg)
KVP
LOW KVP PRODUCES A BLACK AND WHITE IMAGE.
TO PRODUCE AN ADEQUATE OD MUST USE HIGHER MAS
HIGHER MAS MEANS MORE EXPOSURE TO THE PATIENT
![Page 35: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/35.jpg)
HIGH KVP HIGH MAS
+Greater margin of error
+Reduction in patient dose
+Wider latitude -increase scatter -increase fog
+Higher OD +Lower Rad noise -increased pt.
Exposure -less margin of
error
![Page 36: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/36.jpg)
The Radiographer controls
kVp mAs focal spot size distance added filtration screen speed
![Page 37: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/37.jpg)
REMEMBER
200 mA increased to 300mA is a 50% increase in electrons being produced
As mA increases, time must decrease to maintain same density
Distance Maintenance formula--New Sid over Old Sid Squared
![Page 38: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY.](https://reader036.fdocuments.net/reader036/viewer/2022062314/56649e4b5503460f94b40427/html5/thumbnails/38.jpg)
REMEMBER
Small mA station uses smaller focal spot
smaller focal spot = more detail Inherent = .5 mm additional 2 mm of aluminum between
housing and collimator collimator adds 1 mm