Computed Tomography III Reconstruction Image quality Artifacts.
-
Upload
kelly-lindsey -
Category
Documents
-
view
227 -
download
4
Transcript of Computed Tomography III Reconstruction Image quality Artifacts.
Computed Tomography III
Reconstruction
Image quality
Artifacts
Simple backprojection
• Starts with an empty image matrix, and the value from each ray in all views is added to each pixel in a line through the image corresponding to the ray’s path
• A characteristic 1/r blurring is a byproduct
• A filtering step is therefore added to correct this blurring
Filtered backprojection
• The raw view data are mathematically filtered before being backprojected onto the image matrix
• Involves convolving the projection data with a convolution kernel
• Different kernels are used for varying clinical applications such as soft tissue imaging or bone imaging
Convolution filters
• Lak filter increases amplitude linearly as a function of frequency; works well when there is no noise in the data
• Shepp-Logan filter incorporates some roll-off at higher frequencies, reducing high-frequency noise in the final CT image
• Hamming filter has even more pronounced high-frequency roll-off, with better high-frequency noise suppression
Bone kernels and soft tissue kernels
• Bone kernels have less high-frequency roll-off and hence accentuate higher frequencies in the image at the expense of increased noise
• For clinical applications in which high spatial resolution is less important than high contrast resolution – for example, in scanning for metastatic disease in the liver – soft tissue kernels are used– More roll-off at higher frequencies and therefore produce
images with reduced noise but lower spatial resolution
CT numbers or Hounsfield units
• The number CT(x,y) in each pixel, (x,y), of the image is:
• CT numbers range from about –1,000 to +3,000 where –1,000 corresponds to air, soft tissues range from –300 to –100, water is 0, and dense bone and areas filled with contrast agent range up to +3,000
water
wateryxyxCT
),(
000,1),(
CT numbers (cont.)
• CT numbers are quantitative• CT scanners measure bone density with
good accuracy– Can be used to assess fracture risk
• CT is also quantitative in terms of linear dimensions– Can be used to accurately assess tumor volume
or lesion diameter
Digital image display
• Window and level adjustments can be made as with other forms of digital images
• Reformatting of existing image data may allow display of sagittal or coronal slices, albeit with reduced spatial resolution compared with the axial views
• Volume contouring and surface rendering allow sophisticated 3D volume viewing
Image quality
• Compared with x-ray radiography, CT has significantly worse spatial resolution and significantly better contrast resolution
• Limiting spatial resolution for screen-film radiography is about 7 lp/mm; for CT it is about 1 lp/mm
• Contrast resolution of screen-film radiography is about 5%; for CT it is about 0.5%
Image quality (cont.)
• Contrast resolution is tied to the SNR, which is related to the number of x-ray quanta used per pixel in the image
• There is a compromise between spatial resolution and contrast resolution
• Well-established relationship among SNR, pixel dimensions (), slice thickness (T), and radiation dose (D):
T
SNRD
3
2
Factors affecting spatial resolution
• Detector pitch (center-to-center spacing)– For 3rd generation scanners, detector pitch determines
ray spacing; for 4th generation scanners, it determines view sampling
• Detector aperture (width of active element)– Use of smaller detectors improves spatial resolution
• Number of views– Too few views results in view aliasing, most noticeable
toward the periphery of the image
Factors affecting spatial resolution (cont.)
• Number of rays– For a fixed FOV, the number of rays increases as
detector pitch decreases
• Focal spot size– Larger focal spots cause more geometric unsharpness
and reduce spatial resolution
• Object magnification– Increased magnification amplifies the blurring of the
focal spot
Factors affecting spatial resolution (cont.)
• Slice thickness– Large slice thicknesses reduce spatial resolution in the
cranial-caudal axis; they also reduce sharpness of edges of structures in the transaxial image
• Slice sensitivity profile– A more accurate descriptor of slice thickness
• Helical pitch– Greater pitches reduce resolution. A larger pitch
increases the slice sensitivity profile
Factors affecting spatial resolution (cont.)
• Reconstruction kernel– Bone filters have the best spatial resolution, and soft
tissue filters have lower spatial resolution
• Pixel matrix• Patient motion
– Involuntary motion or motion resulting from patient noncompliance will blur the CT image proportional to the distance of motion during scan
• Field of view– Influences the physical dimensions of each pixel
Factors affecting contrast resolution
• mAs– Directly influences the number of x-ray photons used to
produce the CT image, thereby influencing the SNR and the contrast resolution
• Dose– Dose increases linearly with mAs per scan
• Pixel size (FOV)– If patient size and all other scan parameters are fixed,
as FOV increases, pixel dimensions increase, and the number of x-rays passing through each pixel increases
Factors affecting contrast resolution (cont.)
• Slice thickness– Thicker slices uses more photons and have better SNR
• Reconstruction filter– Bone filters produce lower contrast resolution, and soft
tissue filters improve contrast resolution
• Patient size– For the same technique, larger patients attenuate more
x-rays, resulting in detection of fewer x-rays. Reduces SNR and therefore the contrast resolution
Factors affecting contrast resolution (cont.)
• Gantry rotation speed– Most CT systems have an upper limit on mA, and for a
fixed pitch and a fixed mA, faster gantry rotations result in reduced mAs used to produce each CT image, reducing contrast resolution
Beam hardening
• Like all medical x-ray beams, CT uses a polyenergetic x-ray spectrum
• X-ray attenuation coefficients are energy dependent– After passing through a given thickness of patient,
lower-energy x-rays are attenuated to a greater extent than higher-energy x-rays are
• As the x-ray beam propagates through a thickness of tissue and bones, the shape of the spectrum becomes skewed toward higher energies
Beam hardening (cont.)
• The average energy of the x-ray beam becomes greater (“harder”) as it passes through tissue
• Because the attenuation of bone is greater than that of soft tissue, bone causes more beam hardening than an equivalent thickness of soft tissue
Beam hardening (cont.)
• The beam-hardening phenomenon induces artifacts in CT because rays from some projection angles are hardened to a differing extent than rays from other angles, confusing the reconstruction algorithm
• Most scanners include a simple beam-hardening correction algorithm, based on the relative attenuation of each ray
• More sophisticated two-pass algorithms determine the path length that each ray transits through bone and soft tissue, and then compensates each ray for beam hardening for the second pass
Motion artifacts
• Motion artifacts arise when the patient moves during the acquisition
• Small motions cause image blurring
• Larger physical displacements produce artifacts that appear as double images or image ghosting
Partial volume averaging
• Some voxels in the image contain a mixture of different tissue types
• When this occurs, the is not representative of a single tissue but instead is a weighted average of the different values
• Most pronounced for softly rounded structures that are almost parallel to the CT slice
Partial volume averaging (cont.)
• Occasionally a partial volume artifact can mimic pathological conditions
• Several approaches to reducing partial volume artifacts– Obvious approach is to use thinner CT slices– When a suspected partial volume artifact occurs
with a helical study and the raw scan data is still available, additional CT images may be reconstructed at different positions