BA058D/06/en/10.03 50098512 Valid as of software version: V 1.06.XX (amplifier) V 1.03.XX (communication) PROline promass 80 Coriolis Mass Flow Mesuring System Description of Device Functions MASS FLOW VOLUME FLOW DENSITY TEMPERATURE UNIT VOLUME UNIT MASS FLOW UNIT MASS COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 OVERFLOW CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SW-REV.-NO. SENSOR SENSOR TYPE I/O-MODUL TYPE SW-REV. I/O-MODUL UNIT VOLUME FLOW DEFINE PRIVATE CODE 100% VALUE LINE 1 100% VALUE LINE 2 UNIT SUM VALUE 0_4 mA VALUE 20 mA END VALUE FREQUENCY OFF-VALUE MIN. PULSE WIDTH BUS ADDRESSE OFF-VALUE POSITVE ZERO RETURN NOMINAL DIAMETER ASSIGN SYSTEM ERROR VALUE SIMULATION STATUS ACCESS VALUE F LOW TIME CONSTANT SIMULATION STATUS HART PROTOCOL EPD SYSTEM DAMPING SENSOR TYPE ERROR CATEGORY MEASURING VALUES SYSTEM UNITS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION MASS FLOW VOLUME FLOW DENSITY TEMPERATURE UNIT VOLUME UNIT MASS FLOW UNIT MASS COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 OVERFLOW CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SW-REV.-NO. SENSOR SENSOR TYPE I/O-MODUL TYPE SW-REV. I/O-MODUL UNIT VOLUME FLOW DEFINE PRIVATE CODE 100% VALUE LINE 1 100% VALUE LINE 2 UNIT SUM VALUE 0_4 mA VALUE 20 mA END VALUE FREQUENCY OFF-VALUE MIN. PULSE WIDTH BUS ADDRESSE OFF-VALUE POSITVE ZERO RETURN NOMINAL DIAMETER ASSIGN SYSTEM ERROR VALUE SIMULATION STATUS ACCESS VALUE F LOW TIME CONSTANT SIMULATION STATUS HART PROTOCOL EPD SYSTEM DAMPING SENSOR TYPE ERROR CATEGORY MEASURING VALUES SYSTEM UNITS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION COMMISSIONING LANGUAGE ASSIGN LINE 1 SUM ASSIGN TOTALIZER ASSIGN CURRENT OPERATION MODE ASSIGN STATUS OUTPUT ASSIGN STATUS INPUT TAG NAME ASSIGN LF-CUTOFF INSTALLATION DIRECTION K-FACTOR ACTUAL SYSTEM CONDITION FAILSAFE MODE SERIAL NUMBER SW-REV. AMP. ACCESS CODE ASSIGN LINE 2 CURRENT SPAN ASSIGN FREQUENCY ON-VALUE AKTIVE LEVEL TAG DESCRIPTION ON-VALUE MEASURING MODE ZERO POINT PREVIOUS SYST. CONDITION SIMULATION MEASURAND SENSOR TYPE I/O-MODUL TYPE DE V F MIN BU P AS QUICK SETUP OPERATION USER INTERFACE TOTALIZER CURRENT OUTPUT PULSE/FREQ.- OUTPUT STATUS OUTPUT STATUS INPUT COMMUNICATION PROCESS PARAMETER SYSTEM PARAMETER SENSOR DATA SUPERVISION SIMULATION SYST. SENSOR VERSION AMP.VERSION

Transcript of PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass...

Page 1: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

BA058D/06/en/10.0350098512

Valid as of software version:V 1.06.XX (amplifier)V 1.03.XX (communication)

PROline promass 80Coriolis Mass Flow Mesuring System

Description of Device Functions

MASS FLOW VOLUME FLOW DENSITY TEMPERATURE

UNIT VOLUMEUNIT MASS FLOW UNIT MASS

COMMISSIONING

LANGUAGE

ASSIGN LINE 1

SUMASSIGN

TOTALIZER

ASSIGN CURRENT

OPERATIONMODE

ASSIGN STATUSOUTPUT

ASSIGN STATUSINPUT

TAGNAME

ASSIGNLF-CUTOFF

INSTALLATIONDIRECTION

K-FACTOR

ACTUAL SYSTEMCONDITION

FAILSAFE MODE

SERIALNUMBER

SW-REV.AMP.

ACCESS CODE

ASSIGN LINE 2

OVERFLOW

CURRENT SPAN

ASSIGNFREQUENCY

ON-VALUE

AKTIVE LEVEL

TAGDESCRIPTION

ON-VALUE

MEASURINGMODE

ZEROPOINT

PREVIOUS SYST.CONDITION

SIMULATIONMEASURAND

SW-REV.-NO.SENSOR

SENSORTYPE

I/O-MODULTYPE

SW-REV.I/O-MODUL

UNITVOLUME FLOW

DEFINE PRIVATECODE

100% VALUELINE 1

100% VALUELINE 2

UNIT SUM

VALUE 0_4 mA VALUE 20 mA

END VALUEFREQUENCY

OFF-VALUE

MIN. PULSE WIDTH

BUS ADDRESSE

OFF-VALUE

POSITVE ZERORETURN

NOMINALDIAMETER

ASSIGN SYSTEMERROR

VALUESIMULATION

STATUS ACCESS

VALUE F LOW

TIME CONSTANT

SIMULATIONSTATUS

HART PROTOCOL

EPD

SYSTEMDAMPING

SENSORTYPE

ERRORCATEGORY

MEASURINGVALUES

SYSTEM UNITS

QUICK SETUP

OPERATION

USER INTERFACE

TOTALIZER

CURRENTOUTPUT

PULSE/FREQ.-OUTPUT

STATUSOUTPUT

STATUSINPUT

COMMUNICATION

PROCESSPARAMETER

SYSTEMPARAMETER

SENSOR DATA

SUPERVISION

SIMULATION SYST.

SENSOR VERSION

AMP. VERSION

MASS FLOW VOLUME FLOW DENSITY TEMPERATURE

UNIT VOLUMEUNIT MASS FLOW UNIT MASS

COMMISSIONING

LANGUAGE

ASSIGN LINE 1

SUMASSIGN

TOTALIZER

ASSIGN CURRENT

OPERATIONMODE

ASSIGN STATUSOUTPUT

ASSIGN STATUSINPUT

TAGNAME

ASSIGNLF-CUTOFF

INSTALLATIONDIRECTION

K-FACTOR

ACTUAL SYSTEMCONDITION

FAILSAFE MODE

SERIALNUMBER

SW-REV.AMP.

ACCESS CODE

ASSIGN LINE 2

OVERFLOW

CURRENT SPAN

ASSIGNFREQUENCY

ON-VALUE

AKTIVE LEVEL

TAGDESCRIPTION

ON-VALUE

MEASURINGMODE

ZEROPOINT

PREVIOUS SYST.CONDITION

SIMULATIONMEASURAND

SW-REV.-NO.SENSOR

SENSORTYPE

I/O-MODULTYPE

SW-REV.I/O-MODUL

UNITVOLUME FLOW

DEFINE PRIVATECODE

100% VALUELINE 1

100% VALUELINE 2

UNIT SUM

VALUE 0_4 mA VALUE 20 mA

END VALUEFREQUENCY

OFF-VALUE

MIN. PULSE WIDTH

BUS ADDRESSE

OFF-VALUE

POSITVE ZERORETURN

NOMINALDIAMETER

ASSIGN SYSTEMERROR

VALUESIMULATION

STATUS ACCESS

VALUE F LOW

TIME CONSTANT

SIMULATIONSTATUS

HART PROTOCOL

EPD

SYSTEMDAMPING

SENSORTYPE

ERRORCATEGORY

MEASURINGVALUES

SYSTEM UNITS

QUICK SETUP

OPERATION

USER INTERFACE

TOTALIZER

CURRENTOUTPUT

PULSE/FREQ.-OUTPUT

STATUSOUTPUT

STATUSINPUT

COMMUNICATION

PROCESSPARAMETER

SYSTEMPARAMETER

SENSOR DATA

SUPERVISION

SIMULATION SYST.

SENSOR VERSION

AMP. VERSION

COMMISSIONING

LANGUAGE

ASSIGN LINE 1

SUMASSIGN

TOTALIZER

ASSIGN CURRENT

OPERATIONMODE

ASSIGN STATUSOUTPUT

ASSIGN STATUSINPUT

TAGNAME

ASSIGNLF-CUTOFF

INSTALLATIONDIRECTION

K-FACTOR

ACTUAL SYSTEMCONDITION

FAILSAFE MODE

SERIALNUMBER

SW-REV.AMP.

ACCESS CODE

ASSIGN LINE 2

O

CURRENT SPAN

ASSIGNFREQUENCY

ON-VALUE

AKTIVE LEVEL

TAGDESCRIPTION

ON-VALUE

MEASURINGMODE

ZEROPOINT

PREVIOUS SYST.CONDITION

SIMULATIONMEASURAND

SSENSORTYPE

I/O-MODULTYPE I

DEF

1

VA

EF

O

MIN.

BUS

O

PO

D

ASS

S

QUICK SETUP

OPERATION

USER INTERFACE

TOTALIZER

CURRENTOUTPUT

PULSE/FREQ.-OUTPUT

STATUSOUTPUT

STATUSINPUT

COMMUNICATION

PROCESSPARAMETER

SYSTEMPARAMETER

SENSOR DATA

SUPERVISION

SIMULATION SYST.

SENSOR VERSION

AMP. VERSION

Page 2: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80

2 Endress+Hauser

Page 3: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 Table of Contents

Endress+Hauser 3

Table of Contents

1 Function matrix PROline Promass 80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 The function matrix: layout and use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.2 Graphical illustration of the function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 Group MEASURING VALUES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3 Group SYSTEM UNITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

4 Group QUICK SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

5 Group OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

6 Group USER INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

7 Group TOTALIZER 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

8 Group HANDLING TOTALIZER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

9 Group CURRENT OUTPUT 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

10 Group PULSE/FREQUENCY OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

11 Group STATUS OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3811.1 Information on the response of the status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4011.2 Switching action of the status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

12 Group STATUS INPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

13 Group COMMUNICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

14 Group PROCESS PARAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

15 Group SYSTEM PARAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

16 Group SENSOR DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

17 Group SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

18 Group SIMULATION SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

19 Group SENSOR VERSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

20 Group AMPLIFIER VERSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

21 Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5821.1 SI units (not for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5821.2 US units (only for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

22 Index of key words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Page 4: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Table of Contents Device Functions PROline Promass 80

4 Endress+Hauser

Registered trademarks

HART�

Registered trademark of HART Communication Foundation, Austin, USA

S-DAT™, T-DAT™, F-CHIP™Registered trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Page 5: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 1 Function matrix PROline Promass 80

Endress+Hauser 5

1 Function matrix PROline Promass 80

1.1 The function matrix: layout and use

The function matrix is a two-level construct: the groups form one level, functions the other.The groups are the highest-level grouping of the control options for the measuring device.Each group comprises a number of functions.You select a group in order to access the individual functions for controlling or parameterizing the measuring device.

You will find an overview of the groups in the table of contents on Page 3 and in the graphical representation of the function matrix on Page 6.You will also find an overview of the functions on Page 6, complete with the page references of the detailed function descriptions.The descriptions of the individual functions start on Page 7.

Example of how to parameterize a function (in this case changing the language for the user interface):➀ Entry into the function matrix (F-key).➁ Select the OPERATION group.➂ Select the LANGUAGE function, change the setting from ENGLISH to DEUTSCH

with OS and save F (all texts displayed on the user interface appear in German).➃ Exit the function matrix (ESC > 3 seconds).

F-x

0xxx

xxx-

19-x

x-xx

-xx-

000

>3s

- + E

Esc

E

E

E

E

E E E E E

–+

➀+

Esc

–+Esc

+Esc

E

Page 6: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

1 Function matrix PROline Promass 80 Device Functions PROline Promass 80

6 Endress+Hauser

1.2 Graphical illustration of the function matrixU

NIT

LE

NG

TH

(p

. 12)

AC

TU

AL

FR

EQ

.(p

. 33)

DE

NS

ITY

SE

T V

ALU

E(p

. 47)

DE

NS

ITY

CO

EF.

C 1

(p. 5

3)

UN

IT T

EM

PE

RAT

UR

E(p

. 11)

TE

ST

DIS

PLA

Y(p

. 19)

VALU

E S

IM.

CU

RR

. (p

. 28)

FAIL

SA

FE

VA

LUE

(p

. 33)

VA

LUE

SIM

ULA

TIO

N

PU

LSE

(p.

37)

ZE

RO

PO

INT

AD

JUS

T

(p. 4

7)

DE

NS

ITY

CO

EF.

C 0

(p. 5

3)

OP

ER

ATIO

N H

OU

RS

(p. 5

5)

UN

IT R

EF.

DE

NS

ITY

(p

. 11)

BA

CK

LIG

HT

(p. 1

9)

SIM

ULA

TIO

N C

UR

R.

(p. 2

7)

FAIL

SA

FE

MO

DE

(p. 3

3)

SIM

ULA

TIO

N P

ULS

E(p

. 37)

FIX

. RE

F. D

EN

SIT

Y(p

. 47)

CA

L. C

OE

F. K

D 2

(p. 5

2)

MA

X. T

EM

P. C

AR

R.

(p. 5

3)

SY

ST

EM

RE

SE

T(p

. 55)

UN

IT D

EN

SIT

Y(p

. 11)

CO

NT

RA

ST

LC

D(p

. 19)

AC

TU

AL

CU

RR

EN

T(p

. 27)

TIM

E C

ON

STA

NT

(p. 3

2)

FAIL

SA

FE

MO

DE

(p. 3

6)

VA

L. S

IM. S

WIT

. PN

T(p

. 39)

EP

D R

ES

PO

NS

ET

IME

(p

. 46)

CA

L. C

OE

F. K

D 1

(p. 5

2)

MIN

. TE

MP.

CA

RR

.(p

. 53)

ALA

RM

DE

LAY

(p. 5

5)

TE

MP

ER

ATU

RE

(p. 7

)

UN

IT C

OR

R. V

OLU

ME

(p

. 10)

DIS

PLA

Y D

AM

PIN

G(p

. 19)

RE

SE

T T

OTA

L.(p

. 21)

FAIL

SA

FE

MO

DE

(p. 2

7)

OU

TP

UT

SIG

NA

L (p

. 32)

OU

TP

UT

SIG

NA

L(p

. 36)

SIM

. SW

ITC

H P

OIN

T

(p. 3

9)

DE

VIC

E I

D

(p. 4

4)

EP

D V

ALU

E H

IGH

(p. 4

6)

TE

MP.

CO

EF.

KT

(p. 5

2)

MA

X. T

EM

P. M

EA

S.

(p. 5

3)

ER

RO

R C

ATE

GO

RY

(p. 5

5)

RE

FE

RE

NC

E D

EN

SIT

Y

(p. 7

)

UN

IT C

OR

R. V

OL.

F

LOW

(p.

10)

AC

CE

SS

CO

DE

C

OU

NT

ER

(p.

16)

FO

RM

AT

(p. 1

8)

TOTA

LIZ

ER

MO

DE

(p. 2

1)

TIM

E C

ON

STA

NT

(p. 2

7)

VA

LUE

F H

IGH

(p

. 30)

PU

LSE

WID

TH

(p. 3

5)

AC

TU

AL

STA

TU

S(p

. 39)

VA

L. S

IM. S

TAT.

IN(p

. 43)

MA

NU

FAC

TU

RE

R ID

(p. 4

4)

EP

D V

ALU

E L

OW

(p. 4

6)

PR

ES

SU

RE

(p. 4

8)

FLO

W D

AM

PIN

G

(p. 5

1)

TE

MP.

CO

EF.

KM

2(p

. 52)

MIN

. TE

MP.

ME

AS

.(p

. 53)

AS

SIG

N P

RO

C. E

RR

. (p

. 54)

DE

NS

ITY

(p. 7

)

UN

IT V

OLU

ME

(p. 9

)

STA

TU

S A

CC

ES

S(p

. 16)

100%

VA

LUE

(p. 1

8)

UN

IT T

OTA

LIZ

ER

(p. 2

1)

VA

LUE

20

mA

(p. 2

4)

VALU

E F

LO

W

(p. 3

0)

PU

LSE

VA

LUE

(p. 3

5)

TIM

E C

ON

STA

NT

(p. 3

9)

SIM

. STA

TU

S IN

(p. 4

2)

HA

RT

PR

OTO

CO

L (p

. 44)

EM

PT

Y P

IPE

DE

T.(p

. 46)

PR

ES

SU

RE

MO

DE

(p. 4

8)

DE

NS

ITY

DA

MP

ING

(p. 5

1)

TE

MP.

CO

EF.

KM

(p. 5

2)

DE

NS

ITY

CO

EF.

C 5

(p. 5

3)

ER

RO

R C

ATE

GO

RY

(p

. 54)

SW

-RE

V. I/

O(p

. 57)

CO

RR

.VO

L.-F

LOW

(p

. 7)

UN

IT V

OLU

ME

FLO

W(p

. 9)

DE

F. P

RIV

ATE

CO

DE

(p. 1

6)

100%

VA

LUE

(p. 1

8)

OV

ER

FLO

W(p

. 20)

VA

LUE

0_4

mA

(p. 2

4)

EN

D V

ALU

E F

RE

Q.

(p. 2

9)

AS

SIG

N P

ULS

E(p

. 34)

OF

F-V

ALU

E

(p. 3

8)

MIN

. P

ULS

E W

IDT

H(p

. 42)

BU

S A

DD

RE

SS

(p. 4

4)

OF

F-V

AL.

LF

-CU

T O

FF

(p. 4

5)

RE

STO

RE

OR

IGIN

AL

(p. 4

8)

PO

S. Z

ER

O R

ET

UR

N

(p. 5

0)

NO

MIN

AL

DIA

ME

TE

R

(p. 5

2)

DE

NS

ITY

CO

EF.

C 4

(p. 5

3)

AS

SIG

N S

YS

. ER

RO

R

(p. 5

4)

VALU

E S

IM. M

EA

S.

(p. 5

6)

SW

-RE

V. S

-DAT

(p. 5

7)

I/O M

OD

UL

TY

PE

(p. 5

7)

VO

LUM

E F

LOW

(p. 7

)

UN

IT M

AS

S(p

. 8)

AC

CE

SS

CO

DE

(p. 1

5)

AS

SIG

N L

INE

2(p

. 17)

SU

M(p

. 20)

FAIL

SA

FE

MO

DE

(p

. 22)

CU

RR

EN

T S

PAN

(p. 2

3)

AS

SIG

N F

RE

QU

EN

CY

(p. 2

9)

VA

LUE

SIM

. FR

EQ

. (p

. 34)

ON

-VA

LUE

(p

. 38)

AC

TIV

E L

EV

EL

(p. 4

2)

TAG

DE

SC

RIP

TIO

N(p

. 44)

ON

-VA

L. L

F-C

UT

OF

F(p

. 45)

DE

NS

ITY

AD

JUS

T(p

. 48)

ME

AS

UR

ING

MO

DE

(p

. 49)

ZE

RO

PO

INT

(p. 5

2)

DE

NS

ITY

CO

EF.

C 3

(p. 5

3)

PR

EV.

SY

S. C

ON

D

(p. 5

4)

SIM

. M

EA

SU

RA

ND

(p

. 56)

SE

NS

OR

TY

PE

(p. 5

7)

LAN

GU

AG

E G

RO

UP

(p. 5

7)

MA

SS

FLO

W(p

. 7)

UN

IT M

AS

S F

LOW

(p. 8

)

UN

IT P

RE

SS

UR

E

(p. 1

2)

QS

CO

MM

ISS

ION

(p

. 13)

LAN

GU

AG

E(p

. 15)

AS

SIG

N L

INE

1(p

. 17)

AS

SIG

N T

OTA

LIZ

ER

(p. 2

0)

RE

SE

T A

LL T

OTA

LIZ

-E

RS

(p.

22)

AS

SIG

N C

UR

RE

NT

(p

. 23)

OP

ER

ATIO

N M

OD

E

(p. 2

9)

SIM

ULA

TIO

N F

RE

Q.

(p. 3

4)

AS

SIG

N S

TAT

US

(p. 3

8)

AS

SIG

N S

TAT

US

IN(p

. 42)

TAG

NA

ME

(p. 4

4)

AS

SIG

N L

F-C

UT

OF

F

(p. 4

5)

ME

AS

UR

E F

LUID

(p

. 47)

INS

TL.

DIR

. SE

NS

OR

(p

. 49)

K-F

AC

TOR

(p. 5

2)

DE

NS

ITY

CO

EF.

C 2

(p. 5

3)

AC

TU

AL

SY

S. C

ON

D.

(p. 5

4)

SIM

. FA

ILS

AF

E M

OD

E

(p. 5

6)

SE

RIA

L N

UM

BE

R

(p. 5

7)

SW

-RE

V. A

MP.

(p

. 57)

▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼ ▼

ME

AS

UR

ING

VA

LU

ES

(p

. 7)

SY

ST

EM

UN

ITS

(p. 8

)

QU

ICK

SE

TU

P(p

. 13)

OP

ER

AT

ION

(p. 1

5)

US

ER

INT

ER

FAC

E(p

. 17)

TOTA

LIZ

ER

1/2

(p. 2

0)

HA

ND

LIN

G T

OTA

LIZ

ER

(p. 2

2)

CU

RR

EN

T O

UT

PU

T 1

/2(p

. 23)

PU

LS

E/F

RE

Q. O

UT.

(p

. 29)

STA

TU

S O

UT

PU

T

(p. 3

8)

STA

TU

S IN

PU

T(p

. 42)

CO

MM

UN

ICA

TIO

N

(p. 4

4)

PR

OC

ES

S P

AR

AM

. (p

. 45)

SY

ST

EM

PA

RA

ME

TE

R

(p. 4

9)

SE

NS

OR

DA

TA(p

. 52)

SU

PE

RV

ISIO

N(p

. 54)

SIM

UL

AT.

SY

ST

EM

(p. 5

6)

SE

NS

OR

VE

RS

ION

(p. 5

7)

AM

P. H

W V

ER

SIO

N(p

. 57)

Page 7: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 2 Group MEASURING VALUES

Endress+Hauser 7

2 Group MEASURING VALUES

Function description MEASURING VALUES

! Note!• The engineering unit of the measured variable shown here can be set in the “SYSTEM UNITS” group

(see Page 8).• If the fluid in the pipe flows backwards, a negative sign prefixes the flow reading on the display.

MASS FLOW In this function, the currently measured mass flow appears on the display.

Display shows:5-digit floating-point number, including unit and sign(e.g. 462.87 kg/h; –731.63 lb/min; etc.)

VOLUME FLOW In this function, the currently measured volumetric flow appears on the display. The volumetric flow is derived from the measured mass flowrate and the den-sity of the fluid measured.

Display shows:5-digit floating-point number, including unit and sign(e.g. 5.5445 dm3/min; 1.4359 m3/h; –731.63 gal/d; etc.)

CORRECTED VOLUME FLOW

The calculated corrected volume flow appears on the display. The calculated corrected volume flow is derived from the measured mass flow and the refer-ence density of the fluid (density at reference temperature, measured or fixed entry).

User Interface:5-digit floating-point number, including unit and sign (e.g. 1.3549 Nm3/h; 7.9846 scm/day; etc.)

DENSITY In this function, the currently measured density or the specific gravity appears on the display.

Display shows:5-digit fixed-point number, including unit(e.g. 1.2345 kg/dm3; 993.5 kg/dm3; 1.0015 SG_20 °C; etc.)

REFERENCE DENSITY The density of the fluid, at reference temperature, appears on the display. The reference density can be measured or also specified via the function FIXED REFERENCE DENSITY.

User Interface:5-digit floating-point number, including unit, corresponding to 0.1000...6.0000 kg/dm3

(e.g. 1.2345 kg/dm3; 993.5 kg/m3; 1.0015 SG_20 °C; etc.)

TEMPERATURE In this function, the currently measured temperature appears on the display.

Display shows:5-digit floating-point number, including unit and sign(e.g. –23.4 °C; 160.0 °F; 295.4 K; etc.)

Page 8: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

3 Group SYSTEM UNITS Device Functions PROline Promass 80

8 Endress+Hauser

3 Group SYSTEM UNITS

Function description SYSTEM UNITS

You can select the unit for the measured variable in this function group.

UNIT MASS FLOW Use this function to select the unit for displaying the mass flow (mass/time).

The unit you select here is also valid for:• Current output• Frequency output• Switching points (limit value for mass flow, flow direction)• Low flow cut off

Options:Metric:gram → g/s; g/min; g/h; g/daykilogram → kg/s; kg/min; kg/h; kg/dayMetric ton → t/s; t/min; t/h; t/day

US:ounce → oz/s; oz/min; oz/h; oz/daypound → lb/s; lb/min; lb/h; lb/dayton → ton/s; ton/min; ton/h; ton/day

Factory setting:Country dependent (kg/h or US lb/day)

UNIT MASS Use this function to select the unit for displaying the mass.

The unit you select here is also valid for:• Pulse value (e.g. kg/p)

Options:Metric → g; kg; t

US → oz; lb; ton

Factory setting:Country dependent (kg or US lb)

! Note!The unit for the totalizer is independent of your choice here, it is selected separately in the TOTALIZER group (see Page 20).

Page 9: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 3 Group SYSTEM UNITS

Endress+Hauser 9

UNIT VOLUME FLOW Use this function to select the unit for the volume flow.

The unit you select here is also valid for:• Current output• Frequency output• Switching points (limit value for volume flow, flow direction)• Low flow cut off

Options:Metric:Cubic centimeter → cm3/s; cm3/min; cm3/h; cm3/dayCubic decimeter → dm3/s; dm3/min; dm3/h; dm3/dayCubic meter → m3/s; m3/min; m3/h; m3/dayMilliliter → ml/s; ml/min; ml/h; ml/dayLiter → l/s; l/min; l/h; l/dayHectoliter → hl/s; hl/min; hl/h; hl/dayMegaliter → Ml/s; Ml/min; Ml/h; Ml/day

US:Cubic centimeter → cc/s; cc/min; cc/h; cc/dayAcre foot → af/s; af/min; af/h; af/dayCubic foot → ft3/s; ft3/min; ft3/h; ft3/dayFluid ounce → oz f/s; oz f/min; oz f/h; oz f/dayGallon → gal/s; gal/min; gal/h; gal/dayMillion gallon → Mgal/s; Mgal/min; Mgal/h; Mgal/dayBarrel (normal fluids: 31.5 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/dayBarrel (beer: 31.0 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/dayBarrel (petrochemicals: 42.0 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/dayBarrel (filling tanks: 55.0 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/day

Imperial:Gallon → gal/s; gal/min; gal/h; gal/dayMega gallon → Mgal/s; Mgal/min; Mgal/h; Mgal/dayBarrel (beer: 36.0 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/dayBarrel (petrochemicals: 42.0 gal/bbl) → bbl/s; bbl/min; bbl/h; bbl/day

Factory setting:Country dependent (m3/h or US Mgal/day)

UNIT VOLUME Use this function to select the unit for the volume.

The unit you select here is also valid for:• Pulse value (e.g. m3/p)

Options:Metric → cm3; dm3; m3; ml; l; hl; Ml

US → cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petrochemicals); bbl (filling tanks)

Imperial → gal; Mgal; bbl (beer); bbl (petrochemicals)

Factory setting:Country dependent (m3 or US Mgal)

! Note!The unit for the totalizer is independent of your choice here, it is selected separately in the TOTALIZER 1/2 group (see Page 20).

Function description SYSTEM UNITS

Page 10: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

3 Group SYSTEM UNITS Device Functions PROline Promass 80

10 Endress+Hauser

UNIT CORRECTED VOL-UME FLOW

Use this function to select the unit for displaying the corrected volume flow (corrected volume/time).

The unit you select here is also valid for:• Current outputs• Frequency outputs• Relay switch points (limit value for corrected volume flow, flow direction)• Low flow cut off

Options:Metric:Nl/sNl/minNl/hNl/dayNm3/sNm3/minNm3/hNm3/day

US:Sm3/s; Sm3/min; Sm3/h; Sm3/dayScf/s; Scf/min; Scf/h; Scf/day

Factory setting:Nm3/h

UNIT CORRECTED VOL-UME

Use this function to select the unit for displaying the corrected volume.

The unit you select here is also valid for:• Pulse value (e.g. Nm3/p)

Options:Metric:Nm3

Nl

US:Sm3

Scf

Factory setting:Nm3

! Note!The unit of the totalizers is independent of your choice here. The unit for each totalizer is selected separately for the totalizer in question.

Function description SYSTEM UNITS

Page 11: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 3 Group SYSTEM UNITS

Endress+Hauser 11

UNIT DENSITY Use this function to select the unit for displaying the fluid density.

The unit you select here is also valid for:• Current output• Frequency output• Switching points• Density adjustment value• Density responce value for EPD

Options:Metric → g/cm3; g/cc; kg/dm3; kg/l; kg/m3; SD 4 °C, SD 15 °C, SD 20 °C; SG 4 °C, SG 15 °C, SG 20 °C

US → lb/ft3; lb/gal; lb/bbl (normal fluids); lb/bbl (beer); lb/bbl (petrochemi-cals); lb/bbl (filling tanks)

Imperial → lb/gal; lb/bbl (beer); lb/bbl (petrochemicals)

Factory setting:kg/l

SD = Specific Density, SG = Specific GravityThe specific density is the ratio of fluid density to water (at water temperature = 4, 15, 20 °C)

UNIT REFERENCE DEN-SITY

Use this function to select the unit for displaying the reference density.

The unit you select here is also valid for:• Current outputs• Frequency outputs• Relay switch points (limit value for density)• Fixed reference density (for calculation of corrected volume flow)

Options:Metric:kg/Nm3

kg/Nl

US:g/Scckg/Sm3

lb/Scf

Factory setting:kg/Nl

UNIT TEMPERATURE Use this function to select the unit for displaying the temperature.

The unit you select here is also valid for:• Current output• Frequency output• Switching points

Options:°C (CELSIUS)

K (KELVIN)°F (FAHRENHEIT)°R (RANKINE)

Factory setting:°C (CELSIUS)

Function description SYSTEM UNITS

Page 12: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

3 Group SYSTEM UNITS Device Functions PROline Promass 80

12 Endress+Hauser

UNIT LENGTH Use this function to select the unit for the unit of length for nominal diameter.

The unit you select here is also valid for:• nominal diameter of the sensor (see the NOMINAL DIAMETER function on

Page 52).

Options:MILLIMETERINCH

Factory setting:Country dependent (MILLIMETER or INCH)

UNIT PRESSURE Use this function to select the unit for pressure.

The unit you select here is also valid for:• Specified pressure (see function DRUCK auf Page 48)

Option:BAR GPSI GBAR APSI A

Factory setting:BAR G

Function description SYSTEM UNITS

Page 13: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 4 Group QUICK SETUP

Endress+Hauser 13

4 Group QUICK SETUP

! Note!• The display returns to the QUICK SETUP COMMISSION cell if you press the ESC key

combination during programming of a parameter.

• ➀ Only units not yet configured in the current Quick Setup are offered for selectionin each cycle. The unit for mass and volume is derived from the corresponding flow unit.

• ➁ The “YES” option remains visible until all the units have been parameterised.“NO” is the only option displayed when no further units are available.

• ➂ Only the outputs not yet configured in the current Quick Setup are offered for selection in each cycle.

• ➃ The “YES” option remains visible until all the outputs have been parameterised.“NO” is the only option displayed when no further outputs are available.

• ➄ The “automatic parameterization of the display” option contains the following basicsettings/factory settings:YES: Main line = Mass flow; Additional line = Totalizer 1.NO: The existing (selected) settings remain.

➀ – ➄: see next page

Function description QUICK SETUP

QUICK SETUP COMMISSION

Use this function to start the Setup menu for commissioning.

Options:NOYES

Factory setting:NO

Page 14: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

4 Group QUICK SETUP Device Functions PROline Promass 80

14 Endress+Hauser

++ +E EEsc

E- +

Yes

Yes

Yes

No

No

No

Mass flow

Current output 1

Volume flow Density

Frequency/Pulse output

Selection System Units

Selection output type

Configuration another unit ?

Configuration another output ?

Automatically configuration display ?

Temperature Quit

Quit

UnitMass flow

AssignCurrent

Current span

Value0/4 mA

Value20 mA

Time constant

Time constantFailsafe mode

Failsafe mode

Failsafe mode

AssignFrequency

End valuefrequency

ValueF Low

ValueF High

Output signal

AssignPulse

Pulse value

Pulse width

Output signal

UnitflowVolume

UnitDensity

Operation mode

UnitTemperature

MeasuringMode

UnitTotalizer

UnitTotalizer

Frequency Pulse

Automaticallyparameterization

of the display

Language

Pre-setting

Quick Setup

HOME-POSITION

Quick SetupCommission

AssignCurrent

Current span

Value0/4 mA

Value20 mA

Time constant

Failsafe mode

Current output 2

Corr. Vol. flow

UnitCorr. Vol. flow

UnitRef. Density

FixRef. Density

F06-80xxxxx-19-xx-xx-en-000

Page 15: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 5 Group OPERATION

Endress+Hauser 15

5 Group OPERATION

Function description OPERATION

LANGUAGE Use this function to select the language for all texts, parameters and mes-sages shown on the local display.

! Note!The displayed options depend on the available language group shown in the LANGUAGE GROUP (8226) function.

Options: Language group WEST EU / USA:ENGLISHDEUTSCHFRANCAISESPANOLITALIANONEDERLANDSPORTUGUESE

Language group EAST EU / SCAND:ENGLISHNORSKSVENSKASUOMIPOLISHRUSSIANCZECH

Language group ASIA:ENGLISHBAHASA INDONESIAJAPANESE (syllabary)

Factory setting:Country-dependent (see Page 58 ff.)

! Note!• If you press the OS keys at startup, the language defaults to “ENGLISH”.

ACCESS CODE All data of the measuring system are protected against inadvertent change. Programming is disabled and the settings cannot be changed until a code is entered in this function.If you press the OS key in any function the measuring system automatically goes to this function and the prompt to enter the code appears on the display (programming disabled).

You can enable programming by entering the personal code (Factory setting = 80, see the DEFINE PRIVATE CODE function)

User input:max. 4-digit number: 0...9999

! Note!• The programming levels are disabled if you do not press a key within 60

seconds following return to the HOME position.• You can also disable programming in this function by entering any number

(other than the defined private code).• The Endress+Hauser service organization can be of assistance if you lose

your personal code.

Page 16: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

5 Group OPERATION Device Functions PROline Promass 80

16 Endress+Hauser

DEFINE PRIVATE CODE Use this function to define a personal code number for enabling programming.

User input:max. 4-digit number: 0...9999

Factory setting:80

! Note!• If you define the personal code number = 0, the programming is always

enabled.• Programming has to be enabled before this code can be changed.

When programming is disabled the function can’t be changed, this precaution prevents others from changing your personal code without your knowledge and consent.

STATUS ACCESS Use this function to check the access status for the function matrix.

Display shows:ACCESS CUSTOMER (Parameterization enabled)LOCKED (Parameterization disabled)

ACCESS CODE COUNTER

Displays how often the customer code, service code or the digit “0” (code-free) has been entered to gain access to the function matrix.

Display:max. 7-digit number: 0...9999999

Factory setting:0

Function description OPERATION

Page 17: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 6 Group USER INTERFACE

Endress+Hauser 17

6 Group USER INTERFACE

Function description USER INTERFACE

ASSIGN LINE 1 Use this function to define the display value assigned to the main line (the upper line of the local display) for display during normal measuring operation.

Options:OFFMASS FLOWMASS FLOW IN %VOLUME FLOWVOLUME FLOW IN %DENSITYTEMPERATURETOTALIZER 1TOTALIZER 2CORRECTED VOLUME FLOWCORRECTED VOLUME FLOW IN %REFERENCE DENSITY

Factory setting:MASS FLOW

ASSIGN LINE 2 Use this function to define the display value assigned to the additional line (the bottom line of the local display) for display during normal measuring operation.

Options:OFFMASS FLOWMASS FLOW IN %VOLUME FLOWVOLUME FLOW IN %DENSITYTEMPERATURETOTALIZER 1TAG NAMEOPERATION/SYSTEM CONDITIONDISPLAY FLOW DIRECTIONMASS FLOW BARGRAPH IN %VOLUME FLOW BARGRAPH IN %TOTALIZER 2CORRECTED VOLUME FLOWCORRECTED VOLUME FLOW IN %CORRECTED VOLUME FLOW BARGRAPH IN %REFERENCE DENSITY

Factory setting:TOTALIZER

Page 18: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

6 Group USER INTERFACE Device Functions PROline Promass 80

18 Endress+Hauser

100% VALUE(Line 1) ! Note!

This function is not available unless one of the following was selected in the ASSIGN LINE 1 function:• MASS FLOW IN %• VOLUME FLOW IN %• CORRECTED VOLUME FLOW IN %

Use this function to define the flow value to be shown on the display as the 100% value of the variable assigned to line 1.

User input:5-digit floating-point number

Factory setting:• 10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)

• 10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is selected)

100% VALUE(Line 2) ! Note!

This function is not available unless one of the following was selected in the ASSIGN LINE 2 function:• MASS FLOW IN %• VOLUME FLOW IN %• MASS FLOW BARGRAPH IN %• VOLUME FLOW BARGRAPH IN %• CORRECTED VOLUME FLOW BARGRAPH IN %

Use this function to define the flow value to be shown on the display as the 100% value of the variable assigned to line 2.

User input:5-digit floating-point number

Factory setting:• 10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)

• 10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is selected)

FORMAT Use this function to define the maximum number of places after the decimal point displayed for the reading in the main line.

Options:XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX

Factory setting: X.XXXX

! Note!• Note that this setting only affects the reading as it appears on the display, it

has no influence on the accuracy of the system's calculations.• The places after the decimal point as computed by the measuring device

cannot always be displayed, depending on this setting and the engineering unit. In these instances an arrow appears on the display between the meas-ured value and the engineering unit (e.g. 1.2 → kg/h), indicating that the measuring system is computing with more decimal places than can be shown on the display.

Function description USER INTERFACE

Page 19: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 6 Group USER INTERFACE

Endress+Hauser 19

DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely fluctuating flow variables, either very quickly (enter a low time con-stant) or with damping (enter a high time constant).

User input:0...100 s

Factory setting:1 s

! Note!Setting the time constant to zero seconds switches off damping.

CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions.

User input:10...100%

Factory setting:50%

BACKLIGHT Use this function to optimize the backlight to suit local operating conditions.

User input:0...100%

! Note!Entering the value “0” means that the backlight is “switched off”. The display then no longer emits any light, i.e. the display texts can no longer be read in the dark.

Factory setting:50%

TEST DISPLAY Use this function to test the operability of the local display and its pixels.

Options:OFFON

Factory setting:OFF

Test sequence:1. Start the test by selecting ON.

2. All pixels of the main line and additional line are darkened for minimum 0.75 second.

3. Main line and additional line show an “8” in each field for minimum 0.75 second.

4. Main line and additional line show a “0” in each field for minimum 0.75 second.

5. Main line and additional line show nothing (blank display) for minimum 0.75 second.

6. When the test completes the local display returns to its initial state and the setting changes to OFF.

Function description USER INTERFACE

Page 20: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

7 Group TOTALIZER 1/2 Device Functions PROline Promass 80

20 Endress+Hauser

7 Group TOTALIZER 1/2

Function description TOTALIZER 1/2

ASSIGN TOTALIZER Use this function to assign a measured variable to the totalizer.

Options:MASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

Factory setting:MASS FLOW

! Note!The totalizer is reset to “0” as soon as the selection is changed.

SUM Use this function to view the total for the totalizer measured variable aggregated since measuring commenced. The value can be positive or negative.

Display shows:max. 7-digit floating-point number, including sign and unit (e.g. 15467.04 kg)

! Note!The totalizer response to faults is defined in the FAILSAFE MODE function (see Page 22).

OVERFLOW Use this function to view the overflow for the totalizer aggregated since measuring commenced.

Total flow quantity is represented by a floating decimal point number consisting of max. 7 digits. You can use this function to view higher

numerical values (>9,999,999) as overflows. The effective quantity is thus the total of OVERFLOW plus the value returned by the SUM function.

Example:Reading for 2 overflows: 2 E7 kg (= 20,000,000 kg)The value returned by the SUM function = 196,845.7 kgEffective total quantity = 20,196,845.7 kg

Display shows: Integer with exponent, including sign and unit, e.g. 2 E7 kg

Page 21: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 7 Group TOTALIZER 1/2

Endress+Hauser 21

UNIT TOTALIZER Use this function to define the unit for the totalizer measured variable, as selected beforehand.

Options (for the MASS FLOW assignment):Metric → g; kg; tUS → oz; lb; ton

Factory setting:Depends on nominal diameter and country,[value] / [g ...kg or US oz...US ton]corresponding to the totalizer unit factory setting (see Page 58 ff.)

Options (for the VOLUME FLOW assignment):Metric → cm3; dm3; m3; ml; l; hl; Ml

US → cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petro-chemicals); bbl (filling tanks)

Imperial → gal; Mgal; bbl (beer); bbl (petrochemicals)

Options (for the CORRECTED VOLUME FLOW assignment):Metric → Nl; Nm3

US → Sm3; Scf

Factory setting:Depends on nominal diameter and country,[value] / [dm3 ...m3 or US gal...US Mgal]corresponding to the totalizer unit factory setting (see Page 58 ff.)

TOTALIZER MODE Use this function to define how the flow components are to be totalized.

Options:BALANCEPositive and negative flow components. The positive and negative flow com-ponents are balanced. In other words, net flow in the flow direction is regis-tered.

FORWARD Only positive flow components are totalized.

REVERSE Only negative flow components are totalized.

Factory setting:Totalizer 1 = BALANCETotalizer 2 = FORWARD

RESET TOTALIZER Use this function to reset the sum and the overflow of the totalizer to zero (= RESET).

Options:NOYES

Factory setting:NO

! Note!If the device has a status input and is appropriately configured, a reset for the totalizer can also be triggered by a pulse.

Function description TOTALIZER 1/2

Page 22: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

8 Group HANDLING TOTALIZER Device Functions PROline Promass 80

22 Endress+Hauser

8 Group HANDLING TOTALIZER

Function description HANDLING TOTALIZER

RESET ALL TOTALIZ-ERS

Use this function to reset the totals (including all overflows) of the totalizers (1...2) to “zero” (= RESET).

Options:NOYES

Factory setting:NO

! Note!If the device has a status input and if it is appropriately configured, a reset for the totalizer (1...2) can also be triggered by a pulse (see the ASSIGN STATUS INPUT function ).

FAILSAFE MODE Use this function to define the response of all totalizers (1...2) to a fault.

Options:STOPThe totalizer is paused until the fault is rectified. The totalizer stops at the last value prior to the occurrence of the error.

ACTUAL VALUEThe totalizer continues to count is based on the current flow measured value. The fault is ignored.

HOLD VALUEThe totalizer continues to count the flow is based on the last valid flow value (before the fault occurred).

Factory setting:STOP

Page 23: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

Endress+Hauser 23

9 Group CURRENT OUTPUT 1/2

Function description CURRENT OUTPUT 1/2

ASSIGN CURRENT Use this function to assign a measured variable to the current output.

Options:OFF MASS FLOWVOLUME FLOWCORRECTED VOLUME FLOWDENSITYREFERENCE DENSITYTEMPERATURE

Factory setting:MASS FLOW

! Note!If you select OFF, the only function shown in this group is these (ASSIGN CURRENT) function.

CURRENT SPAN Use this function to define the current span. The selection specifies the opera-tional range and the lower and upper signal on alarm. For the current output 1 the option HART can be defined additionally.

Options:0–20 mA4–20 mA4–20 mA HART (only current output1)4–20 mA NAMUR4–20 mA HART NAMUR (only current output1)4–20 mA US4–20 mA HART US (only current output1)0–20 mA (25 mA)4–20 mA (25 mA)4–20 mA (25 mA) HART (only current output1)

Factory setting:4–20 mA HART NAMUR (für current output 1)4–20 mA NAMUR (für current output 2)

! Note!• The option HART is only supported by the current output designated as cur-

rent output 1 in the device software, (terminals 26 and 27).• When switching the hardware from an active (factory setting) to a passive

output signal select a current span of 4–20 mA (please refer to the Operat-ing Instructions PROline promass 80, BA 057D/06/en)

Page 24: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

24 Endress+Hauser

CURRENT SPAN(contd)

Current span, operational range and signal on alarm level

A = Current span1 = Operational range (measuring information)2 = Lower signal on alarm level3 = Upper signal on alarm level

! Note!• If the measured value exceeds the measuring range (as defined in the func-

tions VALUE 0_4 mA and VALUE 20 mA a notice message is generated (#351...352, current span).

• In case of a fault the behaviour of the current output is according to the selected option in the function FAILSAFE MODE. Change the error category in the function ASSIGN SYSTEM ERROR to generate a fault message instead of a notice message.

VALUE 0_4 mA! Note!This function is not available unless the DENSITY, REFERENCE DENSITY or TEMPERATURE option was selected in the ASSIGN CURRENT function.

Use this function to assign a value to the 0/4 mA current, (see “Setting the span by means of the 0_4 mA and 20 mA value” on Page 25).

Option:5-digit floating-point number (with sign for the TEMPERATURE measured variable)

Factory setting:0.5 [kg/l] or –50 [°C]

VALUE 20 mA Use this function to assign a value to the 20 mA current, (see “Setting the span by means of the 0_4 mA and 20 mA value” on Page 25).

Option:5-digit floating-point number (with sign for the MASS FLOW, VOLUME FLOW, CORRECTED VOLUME FLOW and TEMPERATURE measured variables)

Factory setting:Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [°C]

Function description CURRENT OUTPUT 1/2

12 3

I[mA]

A 1 2 3

0-20 mA

4-20 mA

4-20 mA NAMUR

4-20 mA US

0-20 mA (25 mA)

4-20 mA (25 mA)

0 - 20.5 mA 0 22

4 - 20.5 mA 2 22

3.8 - 20.5 mA 3.5 22.6

3.9 - 20.8 mA 3.75 22.6

0 - 24 mA 0 25

4 - 24 mA 2 25

4-20 mA HART 4 - 20.5 mA 2 22

4-20 mA HART NAMUR 3.8 - 20.5 mA 3.5 22.6

4-20 mA HART US 3.9 - 20.8 mA 3.75 22.6

4-20 mA (25 mA) HART 4 - 24 mA 2 25 F06-

x3xx

xxxx

-05-

xx-x

x-xx

-017

Page 25: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

Endress+Hauser 25

Setting the span by means of the 0_4 mA and 20 mA value

The span for the measured variable selected in the ASSIGN CURRENT function is specified via the VALUE 0_4 mA and VALUE 20 mA functions.

The span is defined differently, depending on the measured variable selected:

MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW• The VALUE 0_4 mA function is not available; the value for the zero flow (0

kg/h or 0m3/h) is assigned to the 0/4 mA current.• The flow value for the 20 mA current is defined in the VALUE 20 mA function,

(input range –99999 to +99999). The appropriate unit is taken from the UNIT MASS FLOW, UNIT VOLUME FLOW or UNIT CORR. VOL. FLOW function.

Example (for standard measuring mode):

➀ = Initial value (0...20 mA)

➁ = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN

➂ = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN

➃ = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN

➄ = Maximum current value: depends on the setting in the function CURRENT SPAN

➅ = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE,

A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

(Continued on next page)

Function description CURRENT OUTPUT 1/2

0

3

45

6

I [mA]

2

A

1

Q

F06

-80x

xxxx

-05-

xx-x

x-xx

-00

8

Page 26: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

26 Endress+Hauser

Setting the span by means of the 0_4 mA and 20 mA value(contd)

DENSITY and REFERENCE DENSITY• The density value for the 0/4 mA current is defined in the

VALUE 0_4 mA function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY function.

• The density value for the 20 mA current is defined in the VALUE 20 mA function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

Example (for standard measuring mode):

➀ = Initial value (0...20 mA)

➁ = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN

➂ = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN

➃ = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN

➄ = Maximum current value: depends on the setting in the function CURRENT SPAN

➅ = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE

A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

TEMPERATURE• The temperature value for the 0/4 mA current is defined in the

VALUE 0_4 mA function, (input range –99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

• The temperature value for the 20 mA current is defined in the VALUE 20 mA function, (input range –99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

! Note!Values with different signs cannot be entered for the 0_4 mA and 20 mA val-ues if SYMMETRY is the option selected in the MEASURING MODE function, (see Page 49), The message “INPUT RANGE EXCEEDED” appears on the dis-play.

Example (for standard measuring mode)

➀ = Initial value (0...20 mA)

➁ = Lower signal on alarm level:: depends on the setting in the function CURRENT SPAN

➂ = Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN

➃ = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN

➄ = Maximum current value: depends on the setting in the function CURRENT SPAN

➅ = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT SPAN and FAILSAFE MODE

A = Measuring range (the minimum measuring range has to exceed the value that corre-lates with a flow velocity of 0.3 m/s)

Function description CURRENT OUTPUT 1/2

0

3

45

6

I [mA]

2

A

1

Q

F06

-80x

xxxx

-05-

xx-x

x-xx

-009

0

3

45

6

I [mA]

2

A

1

Q

F06

-80x

xxxx

-05-

xx-x

x-xx

-010

Page 27: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 9 Group CURRENT OUTPUT 1/2

Endress+Hauser 27

TIME CONSTANT Use this function to enter a time constant defining how the current output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).

User input:Fixed-point number 0.01...100.00 s

Factory setting:1.00 s

FAILSAFE MODE For safety reasons it is advisable to ensure that the current output assumes a predefined state in the event of a fault. The setting you select here affects only the current output. It has no effect on other outputs and the display (e.g. total-izers).

Options: MIN. CURRENTThe current output adopts the value of the lower signal on alarm level (as defined in the function CURRENT SPAN.

MAX. CURRENTThe current output adopts the value of the upper signal on alarm level (as defined in the function CURRENT SPAN.

HOLD VALUE (not recommended)Measuring value output is based on the last measuring value saved before the error occurred .

ACTUAL VALUE Measured value output is based on the current flow measurement.The fault is ignored .

Factory setting:MIN. CURRENT

ACTUAL CURRENT Use this function to view the computed value of the output current.

Display shows:0.00...25.00 mA

SIMULATION CURRENT Use this function to activate simulation of the current output.

Options:OFFON

Factory setting:OFF

! Note!• The “SIMULATION CURRENT OUTPUT” message indicates that simulation

is active. • The measuring device continues to measure while simulation is in progress,

i.e. the actual measured values are output correctly via the other outputs.

" Caution!The setting is not saved if the power supply fails.

Function description CURRENT OUTPUT 1/2

Page 28: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

9 Group CURRENT OUTPUT 1/2 Device Functions PROline Promass 80

28 Endress+Hauser

VALUE SIMULATION CURRENT ! Note!

This function is not available unless the SIMULATION CURRENT function is active (= ON).Use this function to define a selectable value (e.g. 12 mA) to be output at the current output. This value is used to test downstream devices and the flow-meter itself.

User input:floating-point number: 0.00...25.00 mA

Factory setting:0.00 mA

" Caution!The setting is not saved if the power supply fails.

Function description CURRENT OUTPUT 1/2

Page 29: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

Endress+Hauser 29

10 Group PULSE/FREQUENCY OUTPUT

Function description PULSE/FREQUENCY OUTPUT

OPERATION MODE Use this function to configure the output as a pulse or frequency output.The functions available in this function group vary, depending on which option you select here.

Options:PULSE FREQUENCY

Factory setting:PULSE

ASSIGN FREQUENCY! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to assign a measured variable to the frequency output.

Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOWDENSITYREFERENCE DENSITYTEMPERATURE

Factory setting:MASS FLOW

! Note!If you select OFF, the only functions shown in this function group are ASSIGN FREQUENCY and OPERATION MODE.

END VALUE FREQUENCY ! Note!

This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to define a full scale frequency for the frequency output.You define the associated measured value of the measuring range in the VALUE F HIGH function described on Page 30.

User input:4-digit fixed-point number: 2...1000 Hz

Factory setting:1000 Hz

Example:• VALUE F HIGH = 1000 kg/h, full scale frequency = 1000 Hz: i.e. a frequency

of 1000 Hz is output at a flow of 1000 kg/h.• VALUE F HIGH = 3600 kg/h, full scale frequency = 1000 Hz: i.e. a frequency

of 1000 Hz is output at a flow of 3600 kg/h.

! Note!In the FREQUENCY operating mode the output signal is symmetrical (on/off ratio = 1:1). At low frequencies the pulse duration is limited to a maximum of 2 seconds, i.e. the on/off ratio is no longer symmetrical.

Page 30: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

30 Endress+Hauser

VALUE F LOW! Note!This function is not available unless the DENSITY, REFERENCE DENSITY or TEMPERATURE option was selected in the ASSIGN FREQUENCY function.

Use this function to assign a value to the start value frequency (0 Hz), (see “Setting the span by means of the f-min. and f-max. value” on Page 30).

Option:5-digit floating-point number (with sign for the TEMPERATURE measured variable)

Factory setting:0.5 [kg/l] or –50 [°C]

VALUE F HIGH Use this function to assign a value to the END VALUE FREQUENCY, (see “Setting the span by means of the f-min. and f-max. value” on Page 30).

Option:5-digit floating-point number (with sign for the MASS FLOW, VOLUME FLOW, CORRECTED VOLUME FLOW and TEMPERATURE measured variables)

Factory setting:Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [°C]

Setting the span by means of the f-min. and f-max. value

The span for the measured variable selected in the ASSIGN FREQUENCY function is specified via the VALUE F LOW and VALUE F HIGH functions.

The span is defined differently, depending on the measured variable selected:

MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW• The VALUE F LOW function is not available; the value for the zero flow

(0 kg/h or 0m3/h) is assigned to the start value frequency.• The flow value for the end value frequency is defined in the VALUE F HIGH

function, (input range –99999 to +99999). The appropriate unit is taken from the UNIT MASS FLOW, UNIT VOLUME FLOW or UNIT CORRECTED VOL-UME FLOW function.

Example (for standard measuring mode):

➀ = Flow value at which a frequency of 0 Hz should be output (preset, cannot be edited).

➁ = Flow value at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

(Continued on next page)

Function description PULSE/FREQUENCY OUTPUT

125

100

Measuring range Measured variable (amount)

Value f minValue f max

Frequency [%]

0 F06

-80x

xxxx

x-05

-xx-

xx-e

n-01

1

Page 31: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

Endress+Hauser 31

Setting the span by means of the f-min. and f-max. value(contd)

DENSITY and REFERENCE DENSITY• The density value for the start value frequency is defined in the

VALUE F LOW function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

• The density value for the end value frequency is defined in the VALUE F HIGH function, (input range 0.0000 to +99999). The appropriate unit is taken from the UNIT DENSITY and UNIT REF. DENSITY function.

Example (for standard measuring mode):

➀ = Density value and reference density at which a frequency of 0 Hz should be output (entry in VALUE F LOW function).

➁ = Density value and reference density at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

TEMPERATURE• The temperature value for the 0/4 mA current is defined in the

VALUE F LOW function, (input range –99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

• The temperature value for the 20 mA current is defined in the VALUE F HIGH function, (input range –99999 to +99999). The appropriate unit is taken from the UNIT TEMPERATURE function.

! Note!Values with different signs cannot be entered for the VALUE F LOW and VALUE F HIGH if SYMMETRY is the option selected in the MEASURING MODE function (see Page 49). The message “INPUT RANGE EXCEEDED” appears on the display.

Example (for standard measuring mode):

➀ = Temperature value at which a frequency of 0 Hz should be output (entry in VALUE F LOW function).

➁ = Temperature value at which the, in the function END VALUE FREQUENCY defined, frequency should be output (entry in VALUE F HIGH function).

Function description PULSE/FREQUENCY OUTPUT

125

100

Measuring range Measured variable (amount)

Value f minValue f max

Frequency [%]

0

➁➁

F06

-80x

xxxx

x-05

-xx-

xx-e

n-01

2

125

100

Measuring range Measured variable (amount)

Value f minValue f max

Frequency [%]

0 F06

-80x

xxxx

x-05

-xx-

xx-e

n-01

3

Page 32: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

32 Endress+Hauser

OUTPUT SIGNAL! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to select the polarity of the frequency signal.

Options:PASSIVE - POSITIVEPASSIVE - NEGATIVE

Factory setting:PASSIVE - POSITIVE

PASSIVE:

! Note!For continuous currents up to 25 mA (Imax = 250 mA / 20 ms)

TIME CONSTANT! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to enter a time constant defining how the frequency output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).

User input:floating-point number: 0.00...100.00 s

Factory setting:0.00 s

Function description PULSE/FREQUENCY OUTPUT

1 5 8 37+

-

=

OpenCollector

U = 30 V DCExternal power supply

max

F-x

xxxx

xx-0

4-xx

-xx-

en-

000

PASSIVE-NEGATIVE

conducting conducting

nonconducting

nonconducting

t t

transistor transistor

PASSIVE-POSITIVE

F-x

xxxx

xx-0

5-xx

-xx-

en-0

02

Page 33: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

Endress+Hauser 33

FAILSAFE MODE! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

For reasons of safety it is advisable to ensure that the frequency output assumes a predefined state in the event of a fault. Use this function to define this state. The setting you select here affects only the frequency output. It has no effect on other outputs and the display (e.g. totalizer).

Options:FALLBACK VALUEOutput is 0 Hz.

FAILSAFE LEVELOutput is the frequency specified in the FAILSAFE VALUE function.

HOLD VALUEMeasured value output is based on the last measured value saved before the fault occurred.

ACTUAL VALUEMeasured value output is based on the current flow measurement. The fault is ignored.

Factory setting:FALLBACK VALUE

FAILSAFE VALUE! Note!This function is not available unless FREQUENCY was selected in the OPERATION MODE function and FAILSAFE LEVEL was selected in the FAILSAFE MODE function.

Use this function to define frequency that the measuring device outputs in the event of an fault.

User input:max. 4-digit number: 0...1250 Hz

Factory setting:1250 Hz

ACTUAL FREQUENCY! Note!This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to view the computed value of the output frequency.

Display shows:0...1250 Hz

Function description PULSE/FREQUENCY OUTPUT

Page 34: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

34 Endress+Hauser

SIMULATION FREQUENCY ! Note!

This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function.

Use this function to activate simulation of the frequency output.

Options:OFFON

Factory setting:OFF

! Note!• The “SIMULATION FREQUENCY OUTPUT” message indicates that

simulation is active.• The measuring device continues to measure while simulation is in progress,

i.e. the currently measured values are output correctly via the other outputs.

" Caution!The setting is not saved if the power supply fails.

VALUE SIMULATION FREQUENCY ! Note!

This function is not available unless FREQUENCY was selected in the OPERATION MODE function and the VALUE SIMULATION FREQUENCY function is active (= ON).

Use this function to define a selectable frequency value (e.g. 500 Hz) to be output at the frequency output. This value is used to test downstream devices and the flowmeter itself.

User input:0...1250 Hz

Factory setting:0 Hz

" Caution!The setting is not saved if the power supply fails.

ASSIGN PULSE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

Use this function to assign a measured variable to the pulse output.

Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

Factory setting:MASS FLOW

! Note!If you select OFF, the only functions shown in this function group are ASSIGN PULSE and OPERATION MODE.

Function description PULSE/FREQUENCY OUTPUT

Page 35: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

Endress+Hauser 35

PULSE VALUE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

Use this function to define the flow at which a pulse is triggered. These pulses can be totaled by an external totalizer, and the total flow quantity since measuring started can be registered in this way.

User input:5-digit floating-point number

Factory setting:Depends on nominal diameter and country,[value] [dm3 ...m3 or US gal...US Mgal] / pulsescorresponding to the factory setting for the pulses value (see Page 58 ff.)

! Note!The appropriate unit is taken from the corresponding UNIT function in the group SYSTEM UNITS (see Page 8).

PULSE WIDTH! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

Use this function to enter the maximum pulse width of the output pulses.

User input:0.5...2000 ms

Factory setting:100 ms

Pulse output is always with the pulse width (B) entered in this function.The intervals (P) between the individual pulses are automatically configured. However, they must at least correspond to the pulse width (B = P).

B = Pulse width entered (the illustration applies to positive pulses)P= Intervals between the individual pulses

! Note!When entering the pulse width, select a value that can still be processed by an external totalizer (e.g. mechanical totalizer, PLC, etc.).

" Caution!If the pulse number or frequency resulting from the pulse value entered, (see function PULSE VALUE on Page 35), and from the current flowis too large to maintain the pulse width selected (interval P is smaller than the pulse width B entered), a system error message (pulse memory) is generated after buffering/balancing time.

Function description PULSE/FREQUENCY OUTPUT

B=P

B

PP

B

B< P

F06-xxxxxxxx-05-xx-xx-xx-012

Page 36: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

10 Group PULSE/FREQUENCY OUTPUT Device Functions PROline Promass 80

36 Endress+Hauser

OUTPUT SIGNAL! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

Use this function to configure the output in such a way that it for example matches an external totalizer. According to the application, the direction of the pulses can be selected here.

Options:PASSIVE - POSITIVEPASSIVE - NEGATIVE

Factory setting:PASSIVE - POSITIVE

PASSIVE:

! Note!For continuous currents up to 25 mA (Imax = 250 mA / 20 ms)

FAILSAFE MODE! Note!This function is not available unless the PULSE setting was selected in the OPERATION MODE function.

For reasons of safety it is advisable to ensure that the pulse output assumes a predefined state in the event of a fault. Use this function to define this state. The setting you select here affects only the pulse output. It has no effect on other outputs and the display (e.g. totalizer).

Options:FALLBACK VALUEOutput is 0 pulse.

HOLD VALUEMeasured value output is based on the last measured value saved before the fault occurred.

ACTUAL VALUEMeasured value output is based on the current flow measurement. The fault is ignored.

Factory setting:FALLBACK VALUE

Function description PULSE/FREQUENCY OUTPUT

1 5 8 37+

-

=

OpenCollector

U = 30 V DCExternal power supply

max

F-x

xxxx

xx-0

4-xx

-xx-

en-

000

B B

PASSIVE-POSITIVEpulses

conducting conducting

nonconducting

nonconducting

t t

transistor transistor

PASSIVE-NEGATIVEpulses

B = Pulse width

F-x

xxxx

xx-0

5-xx

-xx-

en-0

01

Page 37: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 10 Group PULSE/FREQUENCY OUTPUT

Endress+Hauser 37

SIMULATION PULSE ! Note!This function is not available unless the PULSE option was selected in the OPERATING MODE function.

Use this function to activate simulation of the pulse output.

Options: OFFCOUNTDOWNThe pulses specified in the VALUE SIMULATION PULSE function are output.

CONTINUOUSLYPulses are continuously output with the pulse width specified in the PULSE WIDTH function. Simulation is started once the CONTINUOUSLY option is confirmed with the F key.

! Note!Simulation is started by confirming the CONTINUOUSLY option with the F key.The simulation can be switched off again via the SIMULATION PULSE function.

Factory setting:OFF

! Note!• The notice message #631 “SIM. PULSE” indicates that simulation is active.• The on/off ratio is 1:1 for both types of simulation.• The measuring device continues to measure while simulation is in progress,

i.e. the current measured values are output correctly via the other outputs.

" Caution!The setting is not saved if the power supply fails.

VALUE SIMULATION PULSE ! Note!

This function is not available unless the COUNTDOWN option was selected in the SIMULATION PULSE function.

Use this function to specify the number of pulses (e.g. 50) which are output during the simulation. This value is used to test downstream devices and the measuring device itself. The pulses are output with the pulse width specified in the PULSE WIDTH function. The on/off ratio is 1:1.

Simulation is started once the specified value is confirmed with the F key. The display remains at 0 if the specified pulses have been output.

User input:0...10 000

Factory setting:0

! Note!Simulation is started by confirming the simulation value with the F key.The simulation can be switched off again via the SIMULATION PULSE function.

" Caution!The setting is not saved if the power supply fails.

Function description PULSE/FREQUENCY OUTPUT

Page 38: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

11 Group STATUS OUTPUT Device Functions PROline Promass 80

38 Endress+Hauser

11 Group STATUS OUTPUT

Function description STATUS OUTPUT

This group is not available unless the measuring device is fitted with a status output.

ASSIGN STATUS Use this function to assign a switching function to the status output.

Options:OFFON (operation)FAULT MESSAGENOTICE MESSAGEFAULT MESSAGE or NOTICE MESSAGEEMPTY PIPE DETECTION (only if function is active)FLOW DIRECTIONLIMIT MASS FLOWLIMIT VOLUME FLOWLIMIT CORRECTED VOLUME FLOWLIMIT DENSITYLIMIT REFERENCE DENSITYLIMIT TEMPERATURE

Factory setting:FAULT MESSAGE

! Note!• The behaviour of the status output is of the quiescent-current type, in other

words the output is closed (transistor conductive) when normal, error-free measuring is in progress.

• Please read and comply with the information on the switching characteris-tics of the status output (see Page 40, 41).

• If you select OFF, the only function shown in this function group is this func-tion, in other words ASSIGN STATUS.

ON-VALUE! Note!This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT DENSITY, LIMIT REFERENCE DENSITY or FLOW DIRECTION was selected in the ASSIGN STATUS function.

Use this function to assign a value to the switch-on point (status output conductive). The value can be greater or less than the switch-off point.Positive and negative values are permissible.

User input:5-digit floating-point number

Factory setting:0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [°C]

OFF-VALUE! Note!This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT DENSITY, LIMIT REFERENCE DENSITY was selected in the ASSIGN STATUS function.

Use this function to assign a value to the switch-off point (status output not conductive). The value can be greater or less than the switch-on point.Positive and negative values are permissible.

User input:5-digit floating-point number

Factory setting:0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [°C]

Page 39: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 11 Group STATUS OUTPUT

Endress+Hauser 39

TIME CONSTANT! Note!This function is not available unless one of the following was selected in the ASSIGN STATUS function:FLOW DIRECTIONLIMIT MASS FLOWLIMIT VOLUME FLOWLIMIT CORRECTED VOLUME FLOWLIMIT DENSITYLIMIT REFERENCE DENSITYLIMIT TEMPERATURE

Use this function to enter a time constant defining how the status output reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant). The purpose of damping, therefore, is to prevent the status output changing state continuously in response to fluctuations in flow.

User input:5-digit floating-point number: 0.00...100.00 s

Factory setting:0.00 s

ACTUAL STATUS Use this function to check the current status of the status output.

Display shows:NOT CONDUCTIVECONDUCTIVE

SIMULATIONSWITCH POINT

Use this function to activate simulation of the status output.

Options:OFFON

Factory setting:OFF

! Note!• The “SIMULATION STATUS OUTPUT” message indicates that simulation is

active.• The measuring device continues to measure while simulation is in progress,

i.e. the current measured values are output correctly via the other outputs.

" Caution!The setting is not saved if the power supply fails.

VALUE SIMULATION SWITCH POINT ! Note!

This function is not available unless the ON setting was selected in the SIMU-LATION SWITCH POINT function.

Use this function to define the switching response of the status output during the simulation. This value is used to test downstream devices and the flow-meter itself.

User input:NOT CONDUCTIVECONDUCTIVE

Factory setting:NOT CONDUCTIVE

" Caution!The setting is not saved if the power supply fails.

Function description STATUS OUTPUT

Page 40: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

11 Group STATUS OUTPUT Device Functions PROline Promass 80

40 Endress+Hauser

11.1 Information on the response of the status output

GeneralIf you have configured the status output for “LIMIT” or “FLOW DIRECTION”, you can define the requisite switching points in the ON-VALUE and OFF-VALUE functions. When the measured variable in question reaches one of these predefined values, the status output signal switches as shown in the illustrations below.

Status output configured for direction of flowThe value you entered in the ON-VALUE function defines the switching points for the positive and negative directions of flow. If, for example, the switching point you define is = 1 kg/h, the status output is not conductive at –1 kg/h and is conductive at +1 kg/h. Set the switching point to 0 if your process calls for direct switchover (no switching hysteresis). If low flow cut off is used, it is advisable to set hysteresis to a value greater than or equal to the low flow cut off rate.

Switch-off / switch-on point

a = Status output conductiveb = Status output not conductive

Status output configured for limit valueThe status output signal switches as soon as the measured variable falls below or exceeds a defined switching point. Application: Monitoring flow or process-related boundary conditions.

Measured variable

➀ = ON ≤ OFF-VALUE (maximum safety)➁ = ON > OFF-VALUE (minimum safety)➂ = Status output off (not conductive)

F-x

xxxx

xx-0

5-xx

-xx-

xx-0

04

t

➂ ➂

F-x

xxxx

xx-0

5-xx

-xx-

xx-0

05

Page 41: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 11 Group STATUS OUTPUT

Endress+Hauser 41

11.2 Switching action of the status output

Function State Open Collector (Transistor)

ON (operation) System in measuring mode

conductive

System not in measuring mode (power supply failure) not

conductive

Fault message System OK

conductive

(System or process error)Fault ➝ Error response of outputs, inputs and totalizer

not conductive

Notice message System OK

conductive

(System or process error)Fault ➝ Continuation of measuring

not conductive

Fault messageorNotice message

System OK

conductive

(System or process error)Fault ➝ Response to error or Info ➝ Continuation of measuring

not conductive

Empty pipe detection (EPD)

Fluid density above response level, e.g. full measuring tube

conductive

Fluid density below response level, e.g. empty measuring tube

not conductive

Flow direction forward

conductive

reversenot

conductive

Limit value• mass flow• volume flow• corrected vol-

ume flow• density• corrected density• temperature

Limit value not overshot or undershot

conductive

Limit value overshot or undershot not

conductive

Page 42: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

12 Group STATUS INPUT Device Functions PROline Promass 80

42 Endress+Hauser

12 Group STATUS INPUT

Function description STATUS INPUT

This group is not available unless the measuring device is fitted with a status input I/O module.

ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.

Options:OFFRESET TOTALIZER 1POSITIVE ZERO RETURNZEROPOINT ADJUSTMENTRESET TOTALIZER 2RESET ALL TOTALIZERS

Factory setting:OFF

! Note!Positive zero return is active as long as the active level is available at the status input (continuous signal). All other assignments react to a change in level (pulse) at the status input.

ACTIVE LEVEL Use this function to define whether the assigned function (see ASSIGN STATUS INPUT function) is released when the signal level is present (HIGH) or not present (LOW).

Options:HIGHLOW

Factory setting:HIGH

MINIMUM PULSE WIDTH

Use this function to define a minimum width which the input pulse must achieve in order to trigger the defined switching function.

User input:20...100 ms

Factory setting:50 ms

SIMULATION STATUS INPUT

Use this function to activate simulation of the status input, in other words to trigger the function assigned to the status input (see the ASSIGN STATUS INPUT function on Page 38).

Options: OFFON

Factory setting:OFF

! Note!• The “SIMULATION STATUS INPUT” message indicates that simulation is

active.• The measuring device continues to measure while simulation is in progress,

i.e. the current measured values are output correctly via the outputs.

" Caution!The setting is not saved if the power supply fails.

Page 43: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 12 Group STATUS INPUT

Endress+Hauser 43

VALUE SIMULATION STATUS INPUT ! Note!

This function is not available unless the ON setting was selected in the SIMULATION STATUS INPUT function.

Use this function to select the level to be assumed at the status input during the simulation.

Options:HIGHLOW

Factory setting:LOW

" Caution!The setting is not saved if the power supply fails.

Function description STATUS INPUT

Page 44: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

13 Group COMMUNICATION Device Functions PROline Promass 80

44 Endress+Hauser

13 Group COMMUNICATION

Function description COMMUNICATION

TAG NAME Use this function to enter a tag name for the measuring device. You can edit and read this tag name at the local display or via the HART protocol.

User input:max. 8-character text, permissible: A-Z, 0-9, +,–, punctuation marks

Factory setting:“_ _ _ _ _ _ _ _” (without text)

TAG DESCRIPTION Use this function to enter a tag description for the measuring device. You can edit and read this tag description at the local display or via the HART protocol.

User input:max. 16-character text, permissible: A-Z, 0-9, +,–, punctuation marks

Factory setting:“_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _” (without text)

BUS ADDRESS Use this function to define the address for the exchange of data with the HART protocol.

User input:0...15

Factory setting:0

! Note!Addresses 1...15: a constant 4 mA current is applied.

HART PROTOCOL Use this function to display if the HART protocol is active.

Anzeige:OFF = HART protocol not aktiveON = HART protocol aktive

! Note!The HART protocol can be activated with the selection 4-20 mA HART resp. 4-20 mA (25 mA) HART in the CURRENT SPAN function (see Page 23).

MANUFACTURER ID Use this function to view the manufacturer ID in decimal numerical format.

Display shows:Endress+Hauser17 = (≅ 11 hex) for Endress+Hauser

DEVICE ID Use this function to view the device ID in hexadecimal numerical format.

Display shows:50 = (≅ 80 dez) for Promass 80

Page 45: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 14 Group PROCESS PARAMETER

Endress+Hauser 45

14 Group PROCESS PARAMETER

Function description PROCESS PARAMETER

ASSIGN LOW FLOW CUT OFF

Use this function to assign the switching point for the low flow cut off.

Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOW

Factory setting:MASS FLOW

ON VALUE LOW FLOW CUT OFF

Use this function to assign the on value for the low flow cut off. Low flow cut off is active if the setting is a value not equal to 0. The sign of the flow value is highlighted on the display to indicate that low flow cut off is active.

User input:5-digit floating-point number

Factory setting:depends on nominal diameter

! Note!The appropriate unit is taken from the corresponding UNIT function in the group SYSTEM UNITS (see Page 8).

OFF VALUE LOW FLOW CUT OFF

Use this function to enter the switch-off point for low flow cut off.Enter the switch-off point as a positive hysteresis value from the switch-on point.

User input:Integer 0...100%

Factory setting:50%

Example:

Q = Flow [volume/time]t = TimeH = Hysteresisa = ON VALUE LOW FLOW CUT OFF = 200 g/hb = OFF VALUE LOW FLOW CUT OFF = 10%c = Low flow cut off active1 = Low flow cut off is switched on at 200 g/h2 = Low flow cut off is switched off at 220 g/h

1

c c

Q

t

1

2b

a

2

H

F0

6-80

xxxx

xx-0

5-xx

-xx-

xx-0

07

Page 46: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

14 Group PROCESS PARAMETER Device Functions PROline Promass 80

46 Endress+Hauser

EMPTY PIPE DETECTION (EPD)

Use this function to activate the empty pipe detection (EPD). With empty measuring tubes the density of the fluid falls below a specified value (see EPD VALUE LOW function).

Options: OFFON

Factory setting:OFF

" Caution!• Select a correspondingly low EPD VALUE LOW so that the difference to the

effective density of the fluid is sufficiently large enough.This ensures that totally empty measuring tubes and not partially filled ones are detected.

• For gas measurement we strongly recommend to switch off empty pipe detection.

EPD VALUE LOW! Note!This function is not available unless the ON selection was selected in the EMPTY PIPE DETECTION (EPD) function.

Use this function to set an lower threshold for the measured density value, in order to detect possible problems in the process indicated by too low density.

User input:5-digit floating-point number

Factory setting:0.2000 g/cc

EPD VALUE HIGH! Note!This function is not available unless the ON selection was selected in the EMPTY PIPE DETECTION (EPD) function.

Use this function to set an upper threshold for the measured density value.

User input:5-digit floating-point number

Factory setting:6.0000 g/cc

EPD RESPONSE TIME Use this function to enter the time span for which the criteria for an empty pipe have to be satisfied without interruption before a notice message or fault mes-sage is generated.

User input:fixed-point number: 1.0...60.0 s

Factory setting:1.0 s

Function description PROCESS PARAMETER

Page 47: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 14 Group PROCESS PARAMETER

Endress+Hauser 47

FIXED REFERENCE DENSITY

In this function, a fixed value for the reference density can be entered, with which the corrected volume flow or corrected volume is calculated.

User input:5-digit floating-point number

Factory setting:1 kg/Nl

ZERO POINT ADJUST This function enables a zero point adjustment to be automatically carried out. The new zero point determined by the measuring system is adopted by the ZERO POINT function (see Page 52).

Options:CANCELSTART

Factory setting:CANCEL

" Caution!Before carrying out the calibration, please refer to BA 057D/06/en “Promass 80 Operating Instructions” where a detailed description of the zero point adjust-ment is given.

! Note!• Programming is locked during zero point adjustment and the display shows:

“ZERO ADJUST RUNNING”.• If the zero point adjustment is not possible, e.g. with a flow velocity

> 0.1 m/s, or has been cancelled, then the alarm message “ZERO ADJUST NOT POSSIBLE” is shown on the display.

• If the Promass 80 measuring electronics are fitted with a status input, then the zero point can also be activated by using this input.

• After Zero point adjustment is completed, the new zero point can be called up with theF key. If the F key is pressed again, you return to the ZERO POINT ADJUST function.

DENSITY SET VALUE In this function, enter the density set value of the particular fluid for which you want to carry out a field density adjustment.

User input:5 digit floating-point number, incl. units (corresponding to 0.1...5.9999 kg/l)

! Note!• The preset density entered here should not vary from the actual fluid density

by a more than ±10%.• The appropriate unit is taken from the corresponding UNIT function in the

group SYSTEM UNITS (see Page 8).

MEASURE FLUID In this function the actual density of the fluid is measured for the density adjustment.

Options:CANCELSTART

Function description PROCESS PARAMETER

Page 48: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

14 Group PROCESS PARAMETER Device Functions PROline Promass 80

48 Endress+Hauser

DENSITY ADJUST With this function a density adjustment can be carried out on site. The density set value will thus be recalculated and stored. This ensures that the values dependent on density calculations (e.g. volume flow) are as accurate as possi-ble.

" Caution!Before carrying out a density adjustment, please refer to BA 057D/06/en “Promass 80 Operating Instructions” where a detailed description of the density adjustment is given.

! Note!The density adjustment can be executed if:• The sensor does not accurately measure the density which the operator

expects based on laboratory trials.• The characteristics of the fluid are outside the measuring points set at the

factory or reference conditions under which the flowmeter has been calibrated.

• The plant is used solely for measuring a fluid whose density is to be determined very accurately under constant conditions.

User input:CANCELDENSITY ADJUST

Factory setting:CANCEL

RESTORE ORIGINAL With this function the original density coefficient determined at the factory are restored.

Options:NOYES

Factory setting:NO

PRESSURE MODE Use this function to configure an automatic pressure correction. In this way, the effect of a pressure deviation between the calibration and process pressures on the measured error for mass flow is compensated for, (see also Operating Instructions PROline promass 80, BA 057D/06/en, Accuracy Chapter).

Options:OFFFIX (A fixed process pressure for pressure correction is specified)

Factory setting:OFF

PRESSURE! Note!This function is not available unless FIX was selected in the PRESSURE MODE function.

Use this function to enter the value for the process pressure which should be used during pressure correction.

User input:7-digit floating-point number

Factory setting:0 bar g

! Note!The appropriate unit is taken from the function group SYSTEM UNITS (see Page 8).

Function description PROCESS PARAMETER

Page 49: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 15 Group SYSTEM PARAMETER

Endress+Hauser 49

15 Group SYSTEM PARAMETER

Function description SYSTEM PARAMETER

INSTALLATION DIRECTION SENSOR

Use this function to reverse the sign of the measured variable, if necessary.

Options:NORMAL (flow as indicated by the arrow)INVERSE (flow opposite to direction indicated by the arrow)

Factory setting:NORMAL

! Note!Ascertain the actual direction of fluid flow with reference to the direction indicated by the arrow on the sensor (nameplate).

MEASURING MODE Use this function to define the measuring mode for all outputs.

Options:STANDARDSYMMETRY

Factory setting:STANDARD

The responses of the individual outputs in each of the measuring modes are described in detail below:

Current and frequency outputSTANDARDThe output signals of the current and frequency output are proportional to the measured variable. The flow components outside the scaled measuring range (between VALUE 0_4 mA or VALUE F LOW ➀ and the VALUE 20 mA or VALUE F HIGH ➁) are not taken into account for signal output, but a message “CURRENT RANGE AT FULL SCALE VALUE” or “FREQUENCY RANGE AT FULL SCALE VALUE” is issued.

Example for current output:

SYMMETRYThe output signals of the current and frequency output are independent of the direction of flow (absolute amount of the measured variable). The “VALUE 20 mA” or “VALUE F HIGH” ➂ (e.g. backflow) corresponds to the mirrored VALUE 20 mA or VALUE F HIGH ➁ (e.g. flow).

Example for current output:

Q

20

4

0➀ ➁

mA

F-x

xxxx

xx-0

5-xx

-xx-

xx-0

03

Q

20

0➀ ➁➂

mA

4 F-x

xxxx

xx-0

5-xx

-xx-

xx-0

07

Page 50: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

15 Group SYSTEM PARAMETER Device Functions PROline Promass 80

50 Endress+Hauser

MEASURING MODE(Continuation)

Pulse outputSTANDARDOnly positive flow components are totaled. Negative components are not taken into account.

SYMMETRYPositive and negative flow components are taken into account.

! Note!The direction of flow can be output via the configurable status output.

Status output

! Note!Only if in the ASSIGN STATUS function the LIMIT option is selected.

STANDARDThe status output signal switches at the defined switching points.

SYMMETRYThe status output signal switches at the defined switching points, irrespective of the sign. In other words, if you define a switching point with a positive sign the status output signal switches as soon as the value is reached in the nega-tive direction (negative sign) (see illustration).

Example for the SYMMETRY measuring mode:

Switch-on point: Q = 4Switch-off point: Q = 10➀ = Status output switched on (conductive)➁ = Status output switched off (non-conductive)

POSITIVE ZERO RETURN

Use this function to interrupt evaluation of measured variables. This is neces-sary when a piping system is being cleaned, for example. This setting acts on all functions and outputs of the measuring device.

Options:OFFON (signal output is set to zero flow value, temperature and density are output normally)

Factory setting:OFF

Function description SYSTEM PARAMETER

0

- 4

- 10

4

10

Q

t

➀ ➀ ➀➁ ➁

F-x

xxxx

xx-0

5-xx

-xx-

xx-0

05

Page 51: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 15 Group SYSTEM PARAMETER

Endress+Hauser 51

DENSITY DAMPING The density filter allows the sensitivity of the density measuring signal to be lowered with respect to variations in the density of the fluid, e.g. with inhomo-geneous liquids.

User input:max. 5-digit number, including unit: 0.00...100.00 s

Factory setting:0.00 s

! Note!The damping acts on all functions and outputs of the measuring device.

FLOW DAMPING Using the interference blanking (= time constant for exponential filter) the sensitivity of the flow measurement signal can be reduced with respect to transient flows and interference peaks; e.g. with fluid containing solids or gas bubbles, etc.

User input:0...100 s

Factory setting:0 s

! Note!The damping acts on all functions and outputs of the measuring device.

Function description SYSTEM PARAMETER

Page 52: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

16 Group SENSOR DATA Device Functions PROline Promass 80

52 Endress+Hauser

16 Group SENSOR DATA

Function description SENSOR DATA

All sensor data, including calibration factor, zero point, nominal diameter, etc. are set at the factory. All the sensor's parameter settings are saved on the S-DAT™ memory chip.

" Caution!Under normal circumstances you should not change these parameter settings, because changes affect numerous functions of the entire measuring facility in general, and the accuracy of the measuring system in particular. Consequently, most of the functions described below can be accessed only by entering a special service code, which is not the same as your private code number.

Contact the E+H service organization if you have any questions about these functions.

K-FACTOR This function shows the current calibration factor for the sensor.

Factory setting:depends on nominal diameter and calibration

! Note!If the service code is used to call this function, this value can be edited.

ZERO POINT This function shows the current zero-point correction value for the sensor. The zero-point correction factor is calculated and set at the factory.

User input:max. 5-digit number: –99999...+99999

Factory setting:depends on calibration

NOMINAL DIAMETER This function shows the nominal diameter for the sensor. The nominal diameter depends on the size of the sensor and is set at the factory.

Factory setting:depends on the size of the sensor

! Note!If the service code is used to call this function, this value can be edited.

TEMPERATURE COEFFICIENT KM

This function shows the temperatur coefficient KM.

TEMPERATURE COEFFICIENT KM 2

This function shows the temperatur coefficient KM 2.

TEMPERATURE COEFFICIENT KT

This function shows the temperatur coefficient KT.

CALIBRATION COEFFICIENT KD 1

This function shows the calibration coefficient KD 1.

CALIBRATION COEFFICIENT KD 2

This function shows the calibration coefficient KD 2.

Page 53: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 16 Group SENSOR DATA

Endress+Hauser 53

DENSITY COEFFICIENT C 0

This function shows the actual density coefficient C 0.

" Caution!A density adjustment can alter the calibration value of this coefficient.

DENSITY COEFFICIENT C 1

This function shows the actual density coefficient C 1.

" Caution!A density adjustment can alter the calibration value of this coefficient.

DENSITY COEFFICIENT C 2

This function shows the actual density coefficient C 2.

" Caution!A density adjustment can alter the calibration value of this coefficient.

DENSITY COEFFICIENT C 3

This function shows the actual density coefficient C 3.

" Caution!A density adjustment can alter the calibration value of this coefficient.

DENSITY COEFFICIENT C 4

This function shows the actual density coefficient C 4.

" Caution!A density adjustment can alter the calibration value of this coefficient.

DENSITY COEFFICIENT C 5

This function shows the actual density coefficient C 5.

" Caution!A density adjustment can alter the calibration value of this coefficient.

MINIMAL TEMPERATURE MEASURED

Display of the lowest fluid temperature measured.

MAXIMAL TEMPERATURE MEASURED

Display of the highest fluid temperature measured.

MINIMAL TEMPERATURE CARRIER TUBE

Display of the lowest carrier tube temperature measured.

MAXIMAL TEMPERATURE CARRIER TUBE

Display of the highest carrier tube temperature measured.

Function description SENSOR DATA

Page 54: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

17 Group SUPERVISION Device Functions PROline Promass 80

54 Endress+Hauser

17 Group SUPERVISION

Function description SUPERVISION

ACTUAL SYSTEM CONDITION

Use this function to check the current system status.

Display shows:“SYSTEM OK” or the fault / notice message with the highest priority.

PREVIOUS SYSTEM CONDITION

Use this function to view the fifteen most recent fault and notice messages since measuring last started.

Display shows:The 15 most recent fault or notice messages.

ASSIGN SYSTEM ERROR

Use this function to view all system messages and the associated error categories (fault message or notice message). If you select a single system fault you can change its error category.

Display shows:List of system errors

! Note!• Press the F key twice to call the ERROR CATEGORY function.• Use the OS key combination or select CANCEL in the system error list to

exit the function.

ERROR CATEGORY Use this function to define whether a system fault triggers a notice message or a fault message. If you select “FAULT MESSAGES”, all outputs respond to a fault in accordance with their defined error response patterns.

Options:NOTICE MESSAGES (display only)FAULT MESSAGES (outputs and display)

! Note!• Press the F key twice to call the ASSIGN SYSTEM ERROR function.• Use the OS key combination to exit the function.

ASSIGN PROCESS ERROR

Use this function to view all process errors and the associated error categories (fault message or notice message). If you select a single process error you can change its error category.

Display shows:List of process errors

! Note!• Press the F key twice to call the ERROR CATEGORY function.• Use the OS key combination or select CANCEL in the process error list to

exit the function.

Page 55: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 17 Group SUPERVISION

Endress+Hauser 55

ERROR CATEGORY Use this function to define whether a process error triggers a notice message or a fault message. If you select “FAULT MESSAGES”, all outputs respond to a fault in accordance with their defined error response patterns.

Options: NOTICE MESSAGES (display only)FAULT MESSAGES (outputs and display)

! Note!• Press the F key twice to call the ASSIGN PROCESS ERROR function.• Use the OS key combination to exit the function.

ALARM DELAY Use this function to define a time span for which the criteria for an error have to be satisfied without interruption before an error or notice message is generated.

Depending on the setting and the type of fault, this suppression acts on:• Displa• Current output• Frequency output• Status output

User input:0...100 s (in steps of one second)

Factory setting:0 s

" Caution!If this function is activated fault and notice messages are delayed by the time corresponding to the setting before being forwarded to the higher-order controller (process controller, etc.). It is therefore imperative to check in advance in order to make sure whether a delay of this nature could affect the safety requirements of the process. If fault and notice messages cannot be suppressed, a value of 0 seconds must be entered here.

SYSTEM RESET Use this function to perform a reset of the measuring system.

Options:NORESTART SYSTEM (restart without interrupting line supply)

Factory setting:NO

OPERATION HOURS The hours of operation of the device appear on the display.

Display:Depends on the number of hours of operation elapsed:Hours of operation < 10 hours → display format = 0:00:00 (hr:min:sec)Hours of operation 10...10,000 hours → display format = 0000:00 (hr:min)Hours of operation > 10,000 hours → display format = 000000 (hr)

Function description SUPERVISION

Page 56: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

18 Group SIMULATION SYSTEM Device Functions PROline Promass 80

56 Endress+Hauser

18 Group SIMULATION SYSTEM

Function description SIMULATION SYSTEM

SIMULATION FAILSAFE MODE

Use this function to set all inputs, outputs and totalizer to their defined fault response modes, in order to check whether they respond correctly. During this time, the words “SIMULATION FAILSAFE MODE” appear on the display.

Options:OFFON

Factory setting:OFF

SIMULATION MEASURAND

Use this function to set all inputs, outputs and totalizer to their defined flow-response modes, in order to check whether they respond correctly. During this time, the words “SIMULATION MEASURAND” appear on the display.

Options:OFFMASS FLOWVOLUME FLOWCORRECTED VOLUME FLOWDENSITYREFERENCE DENSITYTEMPERATURE

Factory setting:OFF

" Caution!• The measuring device cannot be used for measuring while this simulation is

in progress.• The setting is not saved if the power supply fails.

VALUE SIMULATION MEASURAND ! Note!

This function is not available unless the SIMULATION MEASURAND function is active.

Use this function to define a selectable value (e.g. 12 kg/s). This value is used to test downstream devices and the flowmeter itself.

User input:5-digit floating-point number

Factory setting:0 kg/h (MASS FLOW)0 m3/h (VOLUME FLOW)0 Nm3/h (CORRECTED VOLUME FLOW)0 kg/l (DENSITY)0 kg/NI (REFERENCE DENSITY)0 °C (TEMPERATURE)

" Caution!The setting is not saved if the power supply fails.

Page 57: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 19 Group SENSOR VERSION

Endress+Hauser 57

19 Group SENSOR VERSION

20 Group AMPLIFIER VERSION

Function description SENSOR VERSION

SERIAL NUMBER Use this function to view the serial number of the sensor.

SENSOR TYPE Use this function to view the sensor type (e.g. Promass F).

SOFTWARE REVISION NUMBER S-DAT

Use this function to view the software revision number of the S-DAT.

Function description AMPLIFIER VERSION

SOFTWARE REVISION NUMBER AMPLIFIER

Use this function to view the software revision number of the amplifier.

LANGUAGE GROUP Use this function to view the language group.

The following language groups can be ordered: WEST EU / USA, EAST EU / SCAND., ASIA.

Display:available language group

! Note!• The language options of the available language group are displayed in the

LANGUAGE function.• You can change the language group via the configuration software

FieldTool. Please do not hesitate to contact your E+H sales office if you have any questions.

I/O MODUL TYPE Use this function to view the I/O type (input/output type).

SOFTWAREREVISION NUMBER I/O

Use this function to view the software revision number of the I/O module.

Page 58: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

21 Factory settings Device Functions PROline Promass 80

58 Endress+Hauser

21 Factory settings

21.1 SI units (not for USA and Canada)

Low flow cut off, full scale value, pulse value

Language

Nominal diameter

Low flow cut off Full scale value Pulse value

[mm] (approx. v = 0.04 m/s) (approx. v = 2 m/s) (approx. 2 pulse/s at 2 m/s)

1 0.08 kg/h 4 kg/h 0.001 kg/p

2 0.40 kg/h 20 kg/h 0.010 kg/p

4 1.80 kg/h 90 kg/h 0.010 kg/p

8 8.00 kg/h 400 kg/h 0.100 kg/p

15 26.00 kg/h 1300 kg/h 0.100 kg/p

15 FB 72.00 kg/h 3600 kg/h 1.000 kg/p

25 72.00 kg/h 3600 kg/h 1.000 kg/p

25 FB 180.00 kg/h 9000 kg/h 1.000 kg/p

40 180.00 kg/h 9000 kg/h 1.000 kg/p

40 FB 300.00 kg/h 15000 kg/h 10.000 kg/p

50 300.00 kg/h 15000 kg/h 10.000 kg/p

80 720.00 kg/h 36000 kg/h 10.000 kg/p

100 1200.00 kg/h 60000 kg/h 10.000 kg/p

150 2600.00 kg/h 130000 kg/h 100.000 kg/p

* DN 15, 25, 40 “FB” = Full bore versions Promass I

Country Language

Australia English

Austria Deutsch

Belgium English

Czech Republic Czech

Denmark English

England English

Finland Suomi

France Francais

Germany Deutsch

Hong Kong English

Hungary English

India English

Indonesia Bahasa Indonesia

Instruments International English

Italy Italiano

Japan Japanese

Malaysia English

Netherlands Nederlands

Norway Norsk

Poland Polish

Portugal Portuguese

Russia Russian

Singapore English

South Africa English

Spain Espanol

Sweden Svenska

Switzerland Deutsch

Thailand English

Page 59: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 21 Factory settings

Endress+Hauser 59

Density, length, temperature, reference density

21.2 US units (only for USA and Canada)

Low flow cut off, full scale value, pulse value

Language, density, length, temperature, reference density

Unit

Density kg/lLength mmTemperature °C

Reference density kg/Nl

Nominal diameter

Low flow cut off Full scale value Pulse value

[mm] (approx. v = 0.04 m/s) (approx. v = 2 m/s) (approx. 2 pulse/s at 2 m/s)

1 0.003 lb/min 0.15 lb/min 0.002 lb/p

2 0.015 lb/min 0.75 lb/min 0.020 lb/p

4 0.066 lb/min 3.30 lb/min 0.020 lb/p

8 0.300 lb/min 15.00 lb/min 0.200 lb/p

15 1.000 lb/min 50.00 lb/min 0.200 lb/p

15 FB 2.600 lb/min 130.00 lb/min 2.000 lb/p

25 2.600 lb/min 130.00 lb/min 2.000 lb/p

25 FB 6.600 lb/min 330.00 lb/min 2.000 lb/p

40 6.600 lb/min 330.00 lb/min 2.000 lb/p

40 FB 11.000 lb/min 550.00 lb/min 20.000 lb/p

50 11.000 lb/min 550.00 lb/min 20.000 lb/p

80 26.000 lb/min 1300.00 lb/min 20.000 lb/p

100 44.000 lb/min 2200.00 lb/min 20.000 lb/p

150 95.000 lb/min 4800.00 lb/min 200.000 lb/p

* DN 15, 25, 40 “FB” = Full bore versions Promass I

UnitLanguage English

Density g/ccLength INCHTemperature °F

Reference density g/Scc

Page 60: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

21 Factory settings Device Functions PROline Promass 80

60 Endress+Hauser

Page 61: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 22 Index of key words

Endress+Hauser 61

22 Index of key words

AAccess code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Active Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Actual

Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39System condition . . . . . . . . . . . . . . . . . . . . . . . . 54

AdjustDensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Zero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Assign

Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Display line 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Display line 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . 45Process error . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38System error . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

BBus address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

CCalibration coefficient

KD 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KD 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Carrier tube temperatureMaximal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Minimal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

CodeAccess code counter . . . . . . . . . . . . . . . . . . . . . 16

CoefficientCalibration

KD 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KD 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

DensityC 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

TemperatureKM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KM 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Contrast LDC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Current

OutputCurrent span . . . . . . . . . . . . . . . . . . . . . . . . 24

Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 24Current Output

Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 28Value 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

DDamping

Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

Define private code . . . . . . . . . . . . . . . . . . . . . . . . . . 16Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

CoefficientC 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53C 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Set value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Density adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Density damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Device ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Display

Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

EEmpty pipe detection

EPD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46High value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Low value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

End value frequency . . . . . . . . . . . . . . . . . . . . . . . . . 29Error category

Process error . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55System error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

FFactory setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58Failsafe mode

Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Simulation system . . . . . . . . . . . . . . . . . . . . . . . . 56Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Failsafe valueFrequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Flow damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Frequency

Page 62: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

22 Index of key words Device Functions PROline Promass 80

62 Endress+Hauser

Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29End value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Failsafe value . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Value F high . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Value F low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 34

Funktion matrixGraphical illustration . . . . . . . . . . . . . . . . . . . . . . . 6Layout and use . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

GGroup

Amplifier Version . . . . . . . . . . . . . . . . . . . . . . . . . 57Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 44Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Measuring values . . . . . . . . . . . . . . . . . . . . . . . . . 7Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Process parameter . . . . . . . . . . . . . . . . . . . . . . . 45Pulse/Frequency output . . . . . . . . . . . . . . . . . . . 29Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Sensor data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Sensor Version . . . . . . . . . . . . . . . . . . . . . . . . . . 57Simulation system . . . . . . . . . . . . . . . . . . . . . . . . 56Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Status/Relais output . . . . . . . . . . . . . . . . . . . . . . 38Supervision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54System parameter . . . . . . . . . . . . . . . . . . . . . . . . 49System units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

HHART protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

II/O Modul type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Installation direction sensor . . . . . . . . . . . . . . . . . . . 49

KK-Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

LLanguage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15LCD Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Low flow cut off

Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Off-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45On-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

MManufacturer ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Mass flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Maximal

Carrier tube temperature . . . . . . . . . . . . . . . . . . 53Temperature measured . . . . . . . . . . . . . . . . . . . 53

Measured fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Minimal

Carrier tube temperature . . . . . . . . . . . . . . . . . . 53Temperature measured . . . . . . . . . . . . . . . . . . . 53

Minimum pulse width . . . . . . . . . . . . . . . . . . . . . . . 42

NNominal Diameter . . . . . . . . . . . . . . . . . . . . . . . . . . 52

OOff-value

Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . 45Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

On-valueLow flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . 45Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

OperationMode Pulse-/Freq.-output . . . . . . . . . . . . . . . . . 29

Output signalFrequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

OverflowTotalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

PPositive zero return . . . . . . . . . . . . . . . . . . . . . . . . . 50Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Pressure mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Previous system condition . . . . . . . . . . . . . . . . . . . 54Process Error

Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Error category . . . . . . . . . . . . . . . . . . . . . . . . . . 55

PulseAssign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Failsafe value . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

QQuick Setup

Commission . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Commission (Diagram) . . . . . . . . . . . . . . . . . . . 14

RReset

System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Restore original . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

SSensor type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Serial number sensor . . . . . . . . . . . . . . . . . . . . . . . 57Simulation

Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Page 63: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Device Functions PROline Promass 80 22 Index of key words

Endress+Hauser 63

Switch point . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Software revision number

Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57S-DAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

StatusAccess . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Status inputActive level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Minimum pulse width . . . . . . . . . . . . . . . . . . . . . 42Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 43

Status outputActual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Direction of flow . . . . . . . . . . . . . . . . . . . . . . . . . 40General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Limit value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Off-Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38On-Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Switching action . . . . . . . . . . . . . . . . . . . . . . . . . 41Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

SumTotalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Switch pointSimulation Status output . . . . . . . . . . . . . . . . . . . 39Value Simulation Status output . . . . . . . . . . . . . . 39

SystemCondition

Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Previous . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55System Error

Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Error category . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

TTag

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Temperature coefficient

KM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KM 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52KT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Temperature measuredMaximal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Minimal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Test Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Time constantCurrent output . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

TotalizerAssign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Overflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Sum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

UUnit

Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Mass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Mass flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

VValue

Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Value F high . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Value F low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Value Simulation

Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Switch point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Value 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Version

Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

ZZero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Zero point adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Numerics100% Value

Line 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Line 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Page 64: PROline promass 80 Valid as of software version: Coriolis ... · Device Functions PROline Promass 80 Table of Contents Endress+Hauser 3 Table of Contents 1 Function matrix PROline

Europe

Austria – Wien❑ Endress+Hauser Ges.m.b.H.Tel. (01) 88 05 60, Fax (01) 88 05 63 35

Belarus – MinskBelorgsintezTel. (017) 2 50 84 73, Fax (017) 2 50 85 83

Belgium / Luxembourg – Bruxelles❑ Endress+Hauser S.A. / N.V.Tel. (02) 2 48 06 00, Fax (02) 2 48 05 53

Bulgaria – SofiaIntertech-Automation Ltd.Tel. (02) 9 62 71 52, Fax (02) 9 62 14 71

Croatia – Zagreb❑ Endress+Hauser GmbH+Co.Tel. (01) 6 63 77 85, Fax (01) 6 63 78 23

Cyprus – NicosiaI+G Electrical Services Co. Ltd.Tel. (02) 48 47 88, Fax (02) 48 46 90

Czech Republic – Praha❑ Endress+Hauser Czech s.r.o.Tel. (02) 66 78 42 00, Fax (026) 66 78 41 79

Denmark – Søborg❑ Endress+Hauser A/STel. (70) 13 11 32, Fax (70) 13 21 33

Estonia – TartuElvi-Aqua OÜTel. (7) 30 27 32, Fax (7) 30 27 31

Finland – Helsinki❑ Metso Endress+Hauser OyTel. (204) 8 31 60, Fax (204) 8 31 61

France – Huningue❑ Endress+Hauser S.A.Tel. (389) 69 67 68, Fax (389) 69 48 02

Germany – Weil am Rhein❑ Endress+Hauser Messtechnik GmbH+Co. KGTel. (07621) 9 75 01, Fax (07621) 97 55 55

Great Britain – Manchester❑ Endress+Hauser Ltd.Tel. (0161) 2 86 50 00, Fax (0161) 9 98 18 41

Greece – AthensI & G Building Services Automation S.A.Tel. (01) 9 24 15 00, Fax (01) 9 22 17 14

Hungary – Budapest❑ Endress+Hauser MagyarországTel. (01) 4 12 04 21, Fax (01) 4 12 04 24

Iceland – ReykjavikSindra-Stál hfTel. 5 75 00 00, Fax 5 75 00 10

Ireland – Clane / County Kildare❑ Flomeaco Endress+Hauser Ltd.Tel. (045) 86 86 15, Fax (045) 86 81 82

Italy – Cernusco s/N, Milano❑ Endress+Hauser S.p.A.Tel. (02) 92 19 21, Fax (02) 92 19 23 62

Latvia – RigaElekoms Ltd.Tel. (07) 33 64 44, Fax (07) 33 64 48

Lithuania – KaunasUAB Agava Ltd.Tel. (03) 7 20 24 10, Fax (03) 7 20 74 14

Macedonia – BeogradMeris d.o.o.Tel. (11) 44 42 96 6, Fax (11) 30 85 77 8

Moldavia – ChisinauS.C. Techno Test SRLTel. (02) 22 61 60, Fax (02) 22 83 13

Netherlands – Naarden❑ Endress+Hauser B.V.Tel. (035) 6 95 86 11, Fax (035) 6 95 88 25

Norway – Lierskogen❑ Endress+Hauser A/STel. 32 85 98 50, Fax 32 85 98 51

Poland – Wroclaw❑ Endress+Hauser Polska Sp. z o.o.Tel. (071) 7 80 37 00, Fax (071) 7 80 37 60

Portugal – Cacem❑ Endress+Hauser Lda.Tel. (21) 4 26 72 90, Fax (21) 4 26 72 99

Romania – BucharestRomconseng S.R.L.Tel. (021) 41 12 50 1, Fax (021) 41 01 63 4

Russia – Moscow❑ Endress+Hauser GmbH+CoTel. (095) 78 32 85 0, Fax (095) 78 32 85 5

Slovak Republic – BratislavaTranscom Technik s.r.o.Tel. (2) 44 88 86 90, Fax (2) 44 88 71 12

Slovenia – Ljubljana❑ Endress+Hauser (Slovenija) D.O.O.Tel. (01) 5 19 22 17, Fax (01) 5 19 22 98

Spain – Sant Just Desvern❑ Endress+Hauser S.A.Tel. (93) 4 80 33 66, Fax (93) 4 73 38 39

Sweden – Sollentuna❑ Endress+Hauser ABTel. (08) 55 51 16 00, Fax (08) 55 51 16 55

Switzerland – Reinach/BL 1❑ Endress+Hauser Metso AGTel. (061) 7 15 75 75, Fax (061) 7 11 16 50

Turkey – Levent/IstanbulIntek Endüstriyel Ölcü ve Kontrol SistemleriTel. (0212) 2 75 13 55, Fax (0212) 2 66 27 75

Ukraine – KievPhotonika GmbHTel. (44) 2 68 81 02, Fax (44) 2 69 07 05

Yugoslavia Republic – BeogradMeris d.o.o.Tel. (11) 4 44 29 66, Fax (11) 3 08 57 78

Africa

Algeria – AnnabaSymes Systemes et MesuresTel. (38) 88 30 03, Fax (38) 88 30 02

Egypt – Heliopolis/CairoAnasia Egypt For Trading (S.A.E.)Tel. (02) 2 68 41 59, Fax (02) 2 68 41 69

Morocco – CasablancaOussama S.A.Tel. (02) 22 24 13 38, Fax (02) 2 40 26 57

Rep. South Africa – Sandton❑ Endress+Hauser (Pty.) Ltd.Tel. (011) 2 62 80 00, Fax (011) 2 62 80 62

Tunisia – TunisCMR Controle, Maintenance et RegulationTel. (07) 17 93 07 7, Fax (07) 17 88 59 5

America

Argentina – Buenos Aires❑ Endress+Hauser Argentina S.A.Tel. (11) 45 22 79 70, Fax (11) 45 22 79 09

Brazil – Sao Paulo❑ Samson Endress+Hauser Ltda.Tel. (011) 50 33 43 33, Fax (011) 50 31 30 67

Canada – Burlington, Ontario❑ Endress+Hauser Canada Ltd.Tel. (905) 68 19 29 2, Fax (905) 68 19 44 4

Chile – Santiago de Chile❑ Endress+Hauser (Chile) Ltd.Tel. (02) 3 21 30 09, Fax (02) 3 21 30 25

Colombia – Bogota D.C.Colsein Ltda.Tel. (01) 2 36 76 59, Fax (01) 6 10 78 68

Costa Rica – San JoseEuro-Tec S.A.Tel. 2 20 28 08, Fax 2 96 15 42

Ecuador – QuitoInsetec Cia. Ltda.Tel. (02) 2 26 91 48, Fax (02) 2 46 18 33

El Salvador – San SalvadorAutomatizacion y Control Industrial de El Salvador, S.A. de C.V.Tel. 2 60 24 24, Fax 2 60 56 77

Guatemala – Ciudad de GuatemalaAutomatizacion y Control Industrial, S.A.Tel. (03) 34 59 85, Fax (03) 32 74 31

Honduras – San Pedro Sula, CortesAutomatizacion y Control Industrial de Honduras, S.A. de C.V.Tel. 5 57 91 36, Fax 5 57 91 39

Mexico – México, D.F❑ Endress+Hauser (México), S.A. de C.V.Tel. (5) 5 55 68 24 07, Fax (5) 5 55 68 74 59

Nicaragua – ManaguaAutomatización y Control Industrial de Nicaragua, S.A.Tel. 2 22 61 90, Fax 2 28 70 24

Peru – MirafloresCorsusa InternationalTel. (1) 44 41 20 0, Fax (1) 44 43 66 4

USA – Greenwood, Indiana❑ Endress+Hauser Inc.Tel. (317) 5 35 71 38, Fax (317) 5 35 84 98

USA – Norcross, Atlanta❑ Endress+Hauser Systems & Gauging Inc.Tel. (770) 4 47 92 02, Fax (770) 4 47 57 67

Venezuela – CaracasControval C.A.Tel. (212) 9 44 09 66, Fax (212) 9 44 45 54

Asia

Azerbaijan – BakuModcon Systems - BakuTel. (12) 92 98 59, Fax (12) 99 13 72

Brunei – Negara Brunei DarussalamAmerican International Industries (B) Sdn. Bhd.Tel. (3) 22 37 37, Fax (3) 22 54 58

Cambodia – Khan Daun Penh, Phom PenhComin Khmere Co. Ltd.Tel. (23) 42 60 56, Fax (23) 42 66 22

China – Shanghai❑ Endress+Hauser (Shanghai) Instrumentation Co. Ltd.Tel. (021) 54 90 23 00, Fax (021) 54 90 23 03

China – Beijing❑ Endress+Hauser (Beijing) Instrumentation Co. Ltd.Tel. (010) 65 88 24 68, Fax (010) 65 88 17 25

Hong Kong – Tsimshatsui / Kowloon❑ Endress+Hauser (H.K.) Ltd.Tel. 8 52 25 28 31 20, Fax 8 52 28 65 41 71

India – Mumbai❑ Endress+Hauser (India) Pvt. Ltd.Tel. (022) 56 93 83 33, Fax (022) 56 93 88 330

Indonesia – JakartaPT Grama BazitaTel. (21) 7 95 50 83, Fax (21) 7 97 50 89

Iran – TehranPatsa IndustryTel. (021) 8 72 68 69, Fax (021) 8 71 96 66

Israel – NetanyaInstrumetrics Industrial Control Ltd.Tel. (09) 8 35 70 90, Fax (09) 8 35 06 19

Japan – Tokyo❑ Sakura Endress Co. Ltd.Tel. (0422) 54 06 11, Fax (0422) 55 02 75

Jordan – AmmanA.P. Parpas Engineering S.A.Tel. (06) 5 53 92 83, Fax (06) 5 53 92 05

Kazakhstan – AlmatyBEI ElectroTel. (72) 30 00 28, Fax (72) 50 71 30

Korea, South – Seoul❑ Endress+Hauser (Korea) Co. Ltd.Tel. (02) 26 58 72 00, Fax (02) 26 59 28 38

Kuwait – SafatUnited Technical Services Est. For General TradingTel. 2 41 12 63, Fax 2 41 15 93

Lebanon – Jbeil Main EntryNetwork EngineeringTel. (3) 94 40 80, Fax (9) 54 80 38

Malaysia – Shah Alam, Selangor Darul Ehsan❑ Endress+Hauser (M) Sdn. Bhd.Tel. (03) 78 46 48 48, Fax (03) 78 46 88 00

Pakistan – KarachiSpeedy AutomationTel. (021) 7 72 29 53, Fax (021) 7 73 68 84

Philippines – Pasig City, Metro Manila❑ Endress+Hauser (Phillipines) Inc.Tel. (2) 6 38 18 71, Fax (2) 6 38 80 42

Saudi Arabia – JeddahAnasia Trading Est.Tel. (02) 6 53 36 61, Fax (02) 6 53 35 04

Singapore – Singapore❑ Endress+Hauser (S.E.A.) Pte. Ltd.Tel. (65) 66 82 22, Fax (65) 66 68 48

Sultanate of Oman – RuwiMustafa & Sultan Sience & Industry Co. L.L.C.Tel. 63 60 00, Fax 60 70 66

Taiwan – TaipeiKingjarl CorporationTel. (02) 27 18 39 38, Fax (02) 27 13 41 90

Thailand – Bangkok 10210❑ Endress+Hauser (Thailand) Ltd.Tel. (2) 9 96 78 11-20, Fax (2) 9 96 78 10

United Arab Emirates – DubaiDescon Trading L.L.C.Tel. (04) 2 65 36 51, Fax (04) 2 65 32 64

Uzbekistan – TashkentIm Mexatronika-TesTel. (71) 1 91 77 07, Fax (71) 1 91 76 94

Vietnam – Ho Chi Minh CityTan Viet Bao Co. Ltd.Tel. (08) 8 33 52 25, Fax (08) 8 33 52 27

Australia + New Zealand

Australia – North Ryde NSW 2113❑ Endress+Hauser Australia Pty. Ltd.Tel. (02) 88 77 70 00, Fax (02) 88 77 70 99

New Zealand – AucklandEMC Industrial Group Ltd.Tel. (09) 4 15 51 10, Fax (09) 4 15 51 15

All other countries❑ Endress+Hauser GmbH+Co. KGInstruments InternationalWeil am Rhein, GermanyTel. (07621) 9 75 02, Fax (07621) 97 53 45

❑ Members of the Endress+Hauser group

http://www.endress.com

05.03

BA058D/06/en/10.0350098512FM+SGML 6.0