Your Most Valuable QA and Dosimetry Tools
PROFILER™ 2, IC PROFILER™ & SRS PROFILER™
PROFILER Family
Every clinic should have a PROFILER. The extended PROFILER family delivers fast and powerful beam QA at thousands of clinics, as well as Linac manufacturing facilities and Linac service groups. PROFILER provides a wealth of information about beam performance in real-time, and takes only minutes to setup. It’s that simple, and that powerful.
THE INDUSTRY STANDARD
1 I www.lightages.com2 | SUN NUCLEAR CORPORATION // sunnuclear.com
Routine QAUse instead of a water tank for high resolution routine QA of many parameters
CommissioningReal time measurements of modulated fields such as electronic wedges; export to planning system
Trouble ShootingHighly flexible real-time measurements allows for trouble shooting of beam and linac anomalies, and verification testing after Linac service and parts replacement
Real-Time Machine QA, Commissioning and Trouble Shooting The PROFILER family of 2D arrays share a common mission: real-time precision measurements for a wide variety of routine QA, commissioning1, and troubleshooting applications. Each PROFILER model is designed for maximum flexibility within a specific application, although each is useful across many applications.
SUN NUCLEAR CORPORATION // sunnuclear.com | 3
1 PROFILER 2 and IC PROFILER
PROFILER Family
PROFILER Software• Multiple real time parameter calculations
• Profile comparison , subtraction, overlay
• All relevant international and OEM calculation protocols
• Beam tuning and data plotting interface
• Real-time movie mode measurement and playback
• Import water tank data and compare to PROFILER files
• Developed in cooperation with market leading Linac manufacturers
• Report generation with output to PDF®
• Analysis parameters
- Flatness
- Symmetry
- Field size
- Light-radiation field coincidence
- Penumbra
- Dose rate
- Beam center
- Output
- Beam stabilization time (movie mode)
- Electron energy verification (with accessory wedge)
- Photon wedge angle calculation
- Pulse counting
- Dose/pulse
- Pulse/second
1 I www.lightages.com4 | SUN NUCLEAR CORPORATION // sunnuclear.com
Features & Benefits• Full-field real-time measurements
- Flatness, symmetry, field size, beam center, penumbra width, light radiation field
- Identify startup/time dependent anomalies
- Beam constancy, steering, diagnostics, collimator and rotational sag QA
- Electron energy verification
- Photon energy verification – IC PROFILER
• Quick and easy setup
- Portable and lightweight
- Single power/data cable
- No pre-irradiation required
- No warm-up required
• Measure at isocenter for any gantry angle
- IMF (PROFILER 2, SRS PROFILER) and GMF (IC PROFILER)
• Low signal to noise (0.15%)
• User calibrated every 1-2 years with 15 minute patented calibration method
- No need to send back to manufacturer
• Used by Varian, Elekta and Accuray
• Supports FFF beams
• MRI compatibility – IC PROFILER MR
PROFILER™ 2A 22 x 32 cm array using SunPoint Diode Detectors on X and Y axes for maximum signal and no dose volume averaging
Beam Start-upMeasure beam start-up stability with time or selected beam-on stabilization: Dose rate, dose/pulse, pulse/sec, point symmetry, and more!
Beam Start-upMeasure beam start-up stability with time or selected beam-on stabilization: Dose rate, dose/pulse, pulse/sec, point symmetry, and more!
SUN NUCLEAR CORPORATION // sunnuclear.com | 5
22 cm
32 c
m
PROFILER 2
Detector Type: SunPoint™ Diode Detectors
Detector Quantity: 139
Detector Width: (mm) 0.8
Field Size: (cm) 22 x 32
Detector Axes: X , Y
Detector Spacing: (mm) 4
Specifications• SunPoint® Diode Detectors have the highest sensitivity,
reliability and lowest available volume of any dosimetric measurement system, ensuring accurate beam measurement
• 1.0 cm of integrated water equivalent buildup
• IMF™ holds PROFILER 2 at isocenter for rotational measurements at any gantry angle
- Add up to 10 cm buildup
1 I www.lightages.com6 | SUN NUCLEAR CORPORATION // sunnuclear.com
IC PROFILER™ A 32 x 32 cm array using uniquely shaped ionization chambers on X, Y, and Diagonal axes for closest dosimetric match to water tanks
IC PROFILER
Detector Type: Ion Chambers
Detector Quantity: 251
Detector Width: (mm) 2.9
Field Size: (cm) 32 x 32
Detector Axes: X, Y, diagonals
Detector Spacing: (mm) 5
Specifications• Unique curved ion chambers maximize signal and
minimizes volume averaging
• IC PROFILER accuracy is within 0.5% of a water tank
• Full field measurements of X, Y, Diagonals
- 45 cm x 45 cm diagonals
- 40 cm x 40 cm X and Y field measurement at 75 cm SDD
• GMF™ holds IC PROFILER at isocenter for rotational measurements at any gantry angle
- Add up to 1.5 cm (Varian) or 3.0 cm (Elekta) buildup
Beam SteeringEasy viewing of real-time beam changes of selected parameters
AccessoriesElectron Energy Quad Wedge Plate
• Used for precise electron energy measurements independent of wedge positioning
• Aluminum-based design
• Suitable for analysis of beam energies, ranging from 4 MeV to 22 MeV
Photon Energy Quad Wedge Plate
• Used for precise photon energy measurements independent of wedge positioning
• Copper-based design
• Suitable for analysis of beam energies, ranging from 6 MV to 18 MVranging from 6 MV to 18 MV
32 c
m
32 cm
SUN NUCLEAR CORPORATION // sunnuclear.com | 7
SRS PROFILER™
A 12 x 14 cm array using SunPoint Diode Detectors in offset X, Y, and Diagonal axes for maximum SRS penumbra definition
SRS PROFILER
Detector Type: SunPoint™ Diode Detectors
Detector Quantity: 125
Detector Width: (mm) 0.8
Largest Field Size: (cm) 14.4 cm diameter
Detector Axes: X , Y , diagonals
Detector Spacing: (mm) 4
Specifications • SunPoint® Diode Detectors have the highest sensitivity,
reliability and lowest available volume of any dosimetric measurement system, ensuring accurate beam measurement
• Full field measurements of X, Y, Diagonals
• Attaches to the Accuray® CyberKnife® birdcage
- Measure at any node position
• On Accuray Recommended Equipment List
12 cm
14 c
m
Corporate Headquarters3275 Suntree BoulevardMelbourne, FL 32940 USA
+1 321 259 6862
sunnuclear.com
PROFILER_Solutions_D111215©2015 Sun Nuclear Corporation. All Rights Reserved. All data used is best available at time of publication. Data is subject to change without notice.