Principles of Light Microscopy

96
Principles of Light Microscopy Principles of Light Microscopy Peter Evennett Peter Evennett [email protected] [email protected]

Transcript of Principles of Light Microscopy

Page 1: Principles of Light Microscopy

Principles of Light MicroscopyPrinciples of Light MicroscopyPeter EvennettPeter Evennett

[email protected]@microscopical.co.uk

Page 2: Principles of Light Microscopy

True or False ?• Light microscopy is out of date now that we have electron microscopes.

• All graduates are taught how to use the light microscope.

• Because their fittings are standardised, most objectives and eyepieces can beinterchanged and used on any microscope.

• Oil immersion is necessary only for high magnifications.

• It is best to use thick coverglasses because they are stronger.

• It is best to use very thin coverglasses for top-quality work.

• Microscopes are fitted with diaphragms designed to control the intensity ofillumination.

• A good microscope provides a higher magnification than a poorer one.

• An image of the lamp filament should not occur anywhere in the microscope.

• Light microscopy is so much simpler than electron microscopy, that it isunnecessary to attend a course on it.

Page 3: Principles of Light Microscopy
Page 4: Principles of Light Microscopy

Robert Hooke 1635 - 1703Micrographia 1665

Objective lens

LampPowersupply

Condenserlenses

Focusingmechanism

Mechanicalstage

Eyepiece

Page 5: Principles of Light Microscopy

Thin slice of Cork

I could exceedingly plainlyperceive it to be all

perforated and porous,much like a Honey-comb…

…these pores, or cells...

Page 6: Principles of Light Microscopy

Robert Hooke on

Simple and Compound microscopes

“ … ‘tis possible with a single Microscope to

make discoveries much better than with a

double one, because the colours which do

much disturb the clear Vision in double

Microscopes is clearly avoided and prevented in

the single.”

1665

Page 7: Principles of Light Microscopy

Bacteria

Humansperm

Antony van Leeuwenhoek,1632 - 1723

A different approach – using a single lens

50mm

Page 8: Principles of Light Microscopy

LamphouseFocus knobs

Condenser lensStage

Objective lens

Eyepiece

Binocular

Camera tube

Revolving nosepiece

Stand

Page 9: Principles of Light Microscopy

Like paintings, photographs or sculptures in an artgallery, the microscope image is a representationor a likeness, an artefact, and it will be differentfrom the original object.

In order properly to understand microscope images,we need to understand these artefacts, and howthey are produced.

you look at an image of the specimen.

With a microscope you don’t look at thespecimen:

Page 10: Principles of Light Microscopy

The input requirements of the eyedetermine the output requirements of the

microscope:

• The image must be presented atinfinity, so that the image-forming raysare parallel on entering the eye

• Exit pupils must be 3-5mm diameter,to match the pupils of the eyes…

• and separated by about 65mm tomatch the interocular distance.

Page 11: Principles of Light Microscopy

MicroscopyMicroscopy

Microscopes must provide:• Resolution

ability to carry information about fine detail inthe specimen to the image

• Contrastdifferences in the image between featuresand their surroundings

• Magnificationto make the image large enough for the eyeto appreciate the resolved detail

Page 12: Principles of Light Microscopy

The limit to resolution

All optical systemscameras, telescopes, eyeballs, microscopes

have a limit to the smallness of eachimage point – like ‘optical pixels’.

The smaller they are and the more youhave, the better.

Page 13: Principles of Light Microscopy

All optical systems - the eye,cameras, projectors, telescopes,

microscopes - image a point of lightas a disc of light surrounded by

bright and dark rings.

And the diameter of these rings is related to the aperture of the lens.

Because of Diffractionat the image-forming lens:

Page 14: Principles of Light Microscopy

Airy disc

The limit to resolution

The aperture of the lens

The size of the Airy disc depends on:

The wavelength of the imaging radiation

Larger aperture and shorter wavelength

smaller disc better resolution

Image of a pointsource:

The Airy Pattern

This is caused byDiffraction

Original object

Page 15: Principles of Light Microscopy

Image of a point source:the Airy Pattern

Small aperture Larger aperture

Airy discs

with most of its brightness in the centralAiry disc

Page 16: Principles of Light Microscopy

Diffraction in the microscope

• Diffraction occurs whenever light or otherwave motion encounters any kind of obstacle

• The Airy pattern is the result of diffraction atthe objective lens

• Diffraction also occurs at the specimen.• Whether you consider resolution to be limited

by diffraction at the objective lens or thespecimen, the result is the same.

Page 17: Principles of Light Microscopy

Points resolved Not resolved

Two Airy patterns: brightness plots

The Rayleigh Criterion states that two objects can beresolved in an image when they are separated by

distance 0.61λ / n sin α

Page 18: Principles of Light Microscopy

The Rayleigh Criterion for Resolutioncalculated from aperture of objective lens

The peak of one curve falls (approximately) over thecentre of the first dark ring of the other

a bAmplitudesof a and b

Combined amplitudesof a and b

Just detectableby the human eye

radius = λ / 2 n sin α

Page 19: Principles of Light Microscopy

Light passes throughthe specimen

Directly into the eye

And the specimen is coveredby a layer of glass

Transmitted-light, Bright-field

A most unusual way oflooking at things!

Page 20: Principles of Light Microscopy

Light passes throughthe specimen

Directly into the eye

And the specimen is coveredby a layer of glass

Transmitted-light, Bright-field

A most unusual way oflooking at things!

And the microscopeuses light for a job that‘it wasn’t designed for’- looking at things thatare around the same

dimensions as thewavelength of light

Page 21: Principles of Light Microscopy

The job of an ideal lens• To accept as many rays as possible from each point in an object

• To reassemble all the rays from each point at corresponding pointsin the image…

• In such a way that the distance travelled by all the rays from eachobject point to its corresponding image point is the same- so that they all arrive ‘in phase’.

Page 22: Principles of Light Microscopy

A lens has two focal planes

Focal point

Focal point

Rays entering lens parallel...

...are converged on theopposite side of the lens

Rays entering lensthrough a focal point... …emerge parallel on the

opposite side of the lens

as with a projector

as with a camera

Page 23: Principles of Light Microscopy

Ray diagrams - three simple rules

Focal point

1. Rays entering lens parallel to axis...… cross the axis at the focal point on the opposite side of the lens

2. Rays entering lens through focal point... leave the lens parallel to axis

3. Rays passing through centre of lens are undeviated

Focal pointObject

Image

Page 24: Principles of Light Microscopy

What can lenses do??

Camera- forming a reduced-size, real image, close to the lens

Projector- forming an enlarged, real image, distant from the lens

Magnifying glass- not forming a real image; parallel rays to infinity

Lenses can act in a way similar to those of three familiar optical devices:

Page 25: Principles of Light Microscopy

Object in focal plane

f f'

2f 2f'

Camera -portrait

Projector

Camera -landscape

Camera -macro lens

Magnifyingglass

Object at infinity Image in focal plane

Image at infinity

Image on retina

Focal length of lens

Image smaller than object

Image same size as object

Image larger than object

Page 26: Principles of Light Microscopy

SpecimenObjective lens

Primary imageEyepiece

EyeImage on retinaThe objective lens works like a

projector lens and forms thePrimary Image

10mm below the top of theviewing tube

and the eyepiece acts as amagnifying glassand examines the

centre of this image

Ground glass on top of

microscope tube

Page 27: Principles of Light Microscopy

Magnification

Depends on:Image distance

longer higher mag

Objective focal lengthshorter higher mag

Eyepiece focal lengthshorter higher mag

...and in recent microscopes on thefocal length of the tube lens (more later)

Page 28: Principles of Light Microscopy

The objective lens

The most important lens of the microscope

Is the microscopeThe other parts support its function

and adapt the image to the receiving device

Page 29: Principles of Light Microscopy

Objectivelens

The importance of Aperturein the Microscope

Information

collectedInformation

wasted

Consider that every rayleaving the object carries

some information aboutfine detail in the objectSome of these rays

– and some information –will be collected by the objective

and some rays– and some information –

will NOT be collected,and will be wasted

Resolution will therefore dependon the angular aperture

of the objective -the larger the imagingaperture the higher the

resolution

Page 30: Principles of Light Microscopy

Objective lens

Specimen

Medium ofrefractive index

n

NA = n sin α

Numerical NA Aperture

α

Page 31: Principles of Light Microscopy

NumericalAperture

CoverglassMountantSlide

72°

67°39°

NA = 1 x sin 72° = 1 x 0.95

= 0.95

NA = 1.515 x sin 67° = 1.515 x 0.92

= 1.4Refraction makes

angle α in air

representless-oblique raysat the specimen

- which is where itreally matters!

Immersion ObjectiveDry Objective

Oil

Page 32: Principles of Light Microscopy

‘Milky glass’block

Objective lens

Light passing down microscope from objective into glass block

0.3

Magnification 12.5NA 0.3

Magnification 25NA 0.65

0.650.3

Page 33: Principles of Light Microscopy

Magnification 25NA 0.65

0.65 0.65

Magnification 40NA 0.65

Same aperture, different magnifications

Page 34: Principles of Light Microscopy

25 / 0.65

40 / 0.65

Same aperturedifferent magnifications

Same magnificationdifferent apertures

0.65 0.95

40 / 0.95

Page 35: Principles of Light Microscopy

25 / 0.65

40 / 0.65

40 / 0.95

90 / 1.32

1.32

Oil

0.650.95

0.3

12.5 / 0.3

Page 36: Principles of Light Microscopy

Half-aperture angle

Refractive index of medium

Wavelength ofimaging radiation

Minimumresolved distance

Inscription on Ernst Abbe’s memorial

Numerical Aperture

Minimum resolved distance is now commonly expressed asd = 0.61 λ / NA

d

λ

α

n

Why is Numerical Aperture Important?

Page 37: Principles of Light Microscopy

• Resolution depends on NA

• Light transmission of objectivedepends on NA2

• Depth of field of objective is (approximately)inversely proportional to NA2

Why is Numerical Aperture Important?

Page 38: Principles of Light Microscopy

Once you have bought the objective lenses,there is little you can be done to improve

resolution…

…but it can easily be made worse by poorillumination of the specimen

Microscope Illumination

Page 39: Principles of Light Microscopy

• Light up the specimen uniformly

– over a controllable area

• Illuminate the objective aperture uniformly

– over a controllable angle

What are we trying to do whenilluminating a microscopical specimen??

Page 40: Principles of Light Microscopy

Microscope Illumination

Two basic methods of illumination:

Source-focused (or ‘Critical’) Illumination:Light-source imaged on to specimen

Köhler Illumination:Light-source imaged in the aperture of thecondenser

Page 41: Principles of Light Microscopy

Bench lampimaged on

ground glasson stage bycondenser

lens Ground glassImage of

bench lamp

Source-focused Illumination

Page 42: Principles of Light Microscopy

Light sources suitable for source-focusedillumination:

Uniformly-illuminated skyFlame of oil-lampSurface of opal light bulbUniformly-illuminated white paper or ground glass

*note that these are really ‘secondary sources’

Condenser lens acts like a camera lens- throws an image of source on to underside of slide

**

*

Page 43: Principles of Light Microscopy

gave an image of thestink-pipe on theChemistry Building

But looking for aregion of uniformlyilluminated sky inLeeds…

…when themicroscope was set up correctly

Source-focused Illumination

Page 44: Principles of Light Microscopy

Using a normal electric lampgives an image of the writing

on the end of the bulb

…and a modern halogen lamp iseven worse

Source-focused Illumination

Köhler Illuminationsolves this problem

Page 45: Principles of Light Microscopy

SpecimenObjective lens

Primary imageEyepiece

EyeImage on retina

An image of the object

forms the primary imageand this is transferred

to the retina

These are three

conjugate planes - successive images of one

another

… and there are more.

Conjugate planes

Page 46: Principles of Light Microscopy

August Köhler1866 - 1948

publishedA new system of

illumination forphotomicrographic

purposes

(in German) in 1893.

Page 47: Principles of Light Microscopy

The back focal plane of theobjective

Objective lens

Image of objects at ‘infinity’in back focal plane

of objective

Specimen

Ground glassFocal length

(approx)

Back focal planeFocal length

(approx)

Page 48: Principles of Light Microscopy

Condenser

Object

Objective

vvvvv Back focal planeof objective

Imagine a lightsource in the firstfocal plane of the

condenser

Light will passparallel through

the object

And be brought intofocus in the

back focal plane of the objective

Into the objective

Condenser andobjective lenses

vvvvv

How do we light up thespecimen uniformly?

Page 49: Principles of Light Microscopy

Lamp collector

vvvvv

Filament

vvvvv

Condenser

Object

Objective In Köhler Illumination an extra lens, theLamp collector lens

throws an image of the filament into thefirst focal plane of the condenser

Back focal planeof objective

This image of the filament falls also on the

aperture diaphragm of the condenser,the Illuminating aperture diaphragm

The Illuminated field diaphragmfitted just after the lamp collectoris imaged on to the object by the

condenser lens

vvvvv

In this situation the lamp collectorlens appears to be uniformly filled

with light

vvvvv

vvvvv

Object

Illuminatingaperture

diaphragm

Illuminatedfield

diaphragm

To imageHow do we light up the

specimen uniformly?

Page 50: Principles of Light Microscopy

Note that

• Each point in the objectreceives light from

many points on the filament

and that

• Each point of the filamentprovides light to

many points on the object

Condenser

Lamp collector

Object

Objective

vvvvv

vvvvv

Aperture diaphragmof condenser

To image

vvvvvBack focal planeof objective

How do we light up thespecimen uniformly?

Filament

Object

Page 51: Principles of Light Microscopy

Light up the specimen uniformly over a controllable area?

It is unnecessary, and often detrimental, to illuminateparts of the specimen outside the field of view

– some specimens are light-sensitive, and could be damaged

– light can be scattered into field of view from outside this area

– illuminating a large area of specimen produces a largeprimary image, and light can reflect from internal walls ofmicroscope, reducing contrast in the image

Why is it necessary to…

Page 52: Principles of Light Microscopy

vvvvv

Tube

Condenser

Objective

Eyepiecelens

Lens of eye

Lamp collector

Field diaphragmof eyepiece

Aperture diaphragmof objective

Aperture diaphragmof condenser

Object

Primaryimage

Retina

Why control thearea illuminated ?

Large area ofobject illuminated

provides largedisc of light atprimary image

causing reflectionsfrom walls of

microscope andreduction in contrast

Page 53: Principles of Light Microscopy

vvvvv

Condenser

Objective

Eyepiecelens

Lens of eye

Lamp collector

Field diaphragmof eyepiece

Aperture diaphragmof objective

Aperture diaphragmof condenser

Tube

Illuminated FieldDiaphragm

Retina

1. An adjustable diaphragm here

3. And the disc of light at theprimary image is kept off the walls

of the microscope

How do we control thearea illuminated ?

2. Is imaged on to the specimen,so that the area illuminated is

restricted

Page 54: Principles of Light Microscopy

Objectivelens

Information

collected

Consider that every rayleaving the object carries

some information aboutfine detail in the objectSome of these rays

– and some of the information –will be collected by the objective

Illuminate the objective aperture uniformly over a controllable angle?

Why is it necessary to…

and some rays– and some information –

will NOT be collected,and will be wasted.

Informationwasted

Resolution will thus dependon the angular aperture ofthe objective - the larger theaperture the higher theresolution

Page 55: Principles of Light Microscopy

Objectivelens

‘Common sense’ suggeststhat if we expect toreceive light over a

large angle, it isimportant for good

resolution that most ofthe objective apertureshould be illuminated

Why is it necessary to…Illuminate the objective aperture uniformly over a controllable angle?

But why just most ?Why not all ?

Why not a very widecone of light ?

Page 56: Principles of Light Microscopy

Objectivelens

If the illuminatingaperture is too large,light will be scatteredfrom the edges of theobjective lens, thusreducing contrast.

Why is it necessary to…Illuminate the objective aperture uniformly over a controllable angle?

Page 57: Principles of Light Microscopy

Worse

If the illuminating aperture istoo small,

resolution will be reducedand

image quality will beimpaired

though contrast will beincreased.

Illuminate the objective aperture uniformly over a controllable angle?

Why is it necessary to…

Objectivelens

Page 58: Principles of Light Microscopy

ObjectivelensSo for best resolution the

illuminating apertureshould approach theimaging (objective)

aperture

60 to 75% is oftenrecommended

Why is it necessary to…Illuminate the objective aperture uniformly over a controllable angle?

60%

100%

Page 59: Principles of Light Microscopy

vvvvv

Objective

Eyepiece

Lens of eye

Lamp collector

Field diaphragmof eyepiece

Aperture diaphragmof objective

Aperture diaphragmof condenser

Illuminated FieldDiaphragm

Retina

Object

1. Closing thecondenser diaphragm

2. Narrows the angleof rays passing though

the object

Objective

Condenser

vvvvv

vvvvv

View down tube(objective back focal plane)

How do we control the angle ofillumination ?

Page 60: Principles of Light Microscopy

vvvvv

Objective

Eyepiece

Lens of eye

Lamp collector

Field diaphragmof eyepiece

Aperture diaphragmof objective

Aperture diaphragmof condenser

Illuminated FieldDiaphragm

Retina

Object

vvvvv

vvvvvBack focal plane

of objective

Lamp filament

Illuminating ApertureDiaphragm

vvvvv

The aperturediaphragm of the

condenser thus actsas the Illuminating

Aperture Diaphragm

– so called because itis the diaphragm

which regulates theIlluminating Aperture

How do we control the angle ofillumination ?

Page 61: Principles of Light Microscopy

vvvvv

Illuminated FieldDiaphragm

Illuminating ApertureDiaphragm

Condenser

Objective

Eyepiecelens

Lens of eye

Lamp collector

Field diaphragmof eyepiece

Aperture diaphragmof objective

Aperture diaphragmof condenser

Object

Primaryimage

Retina

Back focal planeof objective

Exit pupilof eyepiece

Filament

vvvvv

vvvvv

vvvvv

Field set ofconjugate planes

Aperture set ofconjugate planes

vvvvv

Page 62: Principles of Light Microscopy

Illuminated Field Diaphragm

Condenser lens

Image of IF Diaphragm on specimen

IF Diaphragm also imagedin primary image plane- prevents light fromreflecting from inside oftube

What are the diaphragmsfor?

Page 63: Principles of Light Microscopy

Illuminating aperturediaphragm

Illuminating aperturediaphragm imaged in

back focal planeof objective…

…and also inExit Pupil

of eyepieceWhat are the diaphragmsfor?

Page 64: Principles of Light Microscopy

Condenser lensIlluminating aperture

diaphragm openIlluminating aperture

diaphragm closed

1. Closing thecondenser diaphragm

2. Narrows the angleof rays passing though

the object

Objective

Condenser

View down tube(objective back focal plane)

What are the diaphragmsfor?

Page 65: Principles of Light Microscopy

The Illuminating Aperture Diaphragm

sets the angle of the cone oflight illuminating the specimen,

and is adjusted according toobjective NA

The diaphragms are NOTintended for adjusting thebrightness of the image

The Illuminated Field Diaphragmsets the area of specimenilluminated, and is adjustedaccording to magnification

What are the diaphragmsfor?

Page 66: Principles of Light Microscopy

Illuminating system completely out ofadjustment

Page 67: Principles of Light Microscopy

Image ofIlluminated

FieldDiaphragmnot sharplyfocused onspecimen

Fuzzy edge

Centre ofimage of

diaphragm

Illuminating system completely out ofadjustment

Illuminated Field Diaphragm closed;its image is not centred on field of view

Page 68: Principles of Light Microscopy

Condenser focus adjusted

Image ofIlluminated

FieldDiaphragmnow sharplyfocused onspecimen

Sharp edge

Page 69: Principles of Light Microscopy

Image of Illuminated Field Diaphragmnow centred on field of view

Page 70: Principles of Light Microscopy

Image of Illuminated FieldDiaphragm centred on field of view

How?• moving the condenser lens in x-y• moving the Illuminated Field Diaphragm in x-y• waggling the microscope mirror• moving the entire lamp about on the bench• centring the objective lens

Most commonsystem

Page 71: Principles of Light Microscopy

Illuminated Field Diaphragm openedto illuminate full field of view

Page 72: Principles of Light Microscopy

Back Focal Plane of objective

Lamp notcentred

withcollector

lens

Page 73: Principles of Light Microscopy

Lampcentred

withcollector

lens

Back Focal Plane of objective

Page 74: Principles of Light Microscopy

Diffuser inserted between lamp andcollector lens

Page 75: Principles of Light Microscopy
Page 76: Principles of Light Microscopy

Illuminating Aperture Diaphragmclosed to c75% of objective aperture

75%

100%

Page 77: Principles of Light Microscopy

Illuminating Aperture too large

Image hazy and ‘washed out’

Page 78: Principles of Light Microscopy

Illuminating Aperture correct

Image contrast optimal

Page 79: Principles of Light Microscopy

Illuminating Aperture too small

Image contrast too high; artefacts present

Page 80: Principles of Light Microscopy

Correct Too small

Illuminating Aperture

75%

Page 81: Principles of Light Microscopy

Illuminating Aperture too small

Page 82: Principles of Light Microscopy

Illuminatingaperturetoo small

Illuminatingaperturecorrect

Note this object

This is what it‘should’ look like!

Page 83: Principles of Light Microscopy
Page 84: Principles of Light Microscopy

Illuminating aperture wide open:much larger thanobjective aperture

Gradually close illuminatingaperture diaphragm.

Illuminating aperture now equal toobjective aperture

Removing stray light

Optimum setting:Stray light removed but

still have adequateilluminating aperture

Imagebrightness

Illuminating aperturemuch smaller thanobjective aperture

Reducing illuminating aperture

X

Setting the Illuminatingaperture diaphragm –

a simple way

Poor image!

Page 85: Principles of Light Microscopy

Illuminating aperture too large

Page 86: Principles of Light Microscopy

Illuminating aperture correct

Page 87: Principles of Light Microscopy

Illuminating aperture too small

Page 88: Principles of Light Microscopy

Illuminating aperture much too small – an extreme example

Page 89: Principles of Light Microscopy
Page 90: Principles of Light Microscopy
Page 91: Principles of Light Microscopy
Page 92: Principles of Light Microscopy

Köhler Illumination provides

Control of Area illuminated by theIlluminated Field Diaphragm,

which is adjusted according to magnification.

Control of Angle of illumination by theIlluminating Aperture Diaphragm

(the condenser diaphragm),which is adjusted according to objective aperture.

Page 93: Principles of Light Microscopy

Epi-illumination

Imagine illuminating system is ‘folded over’about the object plane, so that ...

Condenser

Objective

Illuminated FieldDiaphragm

Primaryimage

vvvvv

vvvvv

vvvvv

vvvvv

Lamp collector

Eyepiecelens

Filament

Condenser falls on Objective

Illuminating Aperture Diaphragm falls inBack Focal Plane of Objective

Illuminated Field Diaphragm falls inPrimary Image plane

Lamp collector falls on Eyepiece lens

Page 94: Principles of Light Microscopy

Condenser

Objective

Illuminated FieldDiaphragm

Primaryimage

vvvvv

vvvvv

vvvvv

vvvvv

Lamp collector

Eyepiecelens

Filament

Page 95: Principles of Light Microscopy

vvvvv

vvvvv

Primaryimage

vvvvv

Illuminated Field Diaphragm inserted in planecorresponding to Primary Image Plane

Objective lens actingas its own Condenser

Reflectorinsertedat 45°

Cannot insert Illuminating Aperture Diaphragm here - would also restrict imaging aperture

Illuminating Aperture Diaphragm inserted here.Filament must now be repositioned

Page 96: Principles of Light Microscopy

vvvvv

vvvvv

Illuminated FieldDiaphragm

Primaryimage

Reflectorvvvvv

Objective lens actingalso as Condenser

Cannot insert Illuminating Aperture Diaphragm here - would also restrict imaging aperture

Illuminating ApertureDiaphragm

vvvvv

Filament