2
Principles of Color Measurement 460 480 500 520 540 560 580 600 620 Kelvin ( K ) 10000 6000 4000 3000 2500 3000 2000 1500 100 % 50 % 0 % Spectral Locus Colors at maximum saturation. Dominant Wavelength Intersection point on the spectral locus by a line originating at white and then projected through a given color. Purity (Excitation Purity) The ratio of the distance of white to a stimulus with respect to the distance of white to the dominant wavelength on the spectral locus (expressed as a percentage). Planckian Locus Color of light emitted by a black body at a given temperature. Used to calculate Correlated Color Temperature (CCT) of white light sources in Kelvin (K). 0.0 400 500 /nm 600 700 0.5 1.0 1.5 2.0 CIE Color Matching Functions MacAdam ellipses indistinguishable to the human eye (ellipses shown here at ten times their actual size). Color gamut illustrates the RGB limits reproducible by digital displays. Calculating CIE color coordinates of a stimulus: For a spectral power distribution (SPD), given as , the XYZ tristimulus values are computed as: 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 CIE 1931 CIE 1931 sRGB DCI-P3 Rec.2020 CIE 1976 0.1 0.2 0.3 0.4 0.5 0.6 0 0.1 0.2 0.3 0.6 0.4 0.5 sRGB DCI-P3 Rec.2020 Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically evaluate light, color, and surface quality. Calibrated to www.RadiantVisionSystems.com Copyright ©2019 Radiant Vision Systems LLC. All rights reserved. 2019/03/21 Anatomy of the CIE Chromaticity Diagram CIE 1931 CIE 1976 v’ u’ y x 0.1 0.2 0.3 0.4 0.5 0.6 0 0.1 0.2 0.3 0.6 0.4 0.5 v’ u’ CIE 1976 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 y x Example Stimulus 490

Transcript of Principles of Color Measurement - Radiant Vision Systems...Radiant Vision Systems engineers imaging...

Page 1: Principles of Color Measurement - Radiant Vision Systems...Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically

Principles of Color Measurement

460

480

500

520

540

560

580

600

620

Kelvin ( K )

10000

6000

4000

3000

25003000

2000

1500

100%50%0%

Spectral Locus Colors at maximum saturation.

Dominant Wavelength Intersection point on the spectral locus by a line originating at white and then projected through a given color.

Purity (Excitation Purity) The ratio of the distance of white to a stimulus with respect to the distance of white to the dominant wavelength on the spectral locus (expressed as a percentage).

Planckian Locus Color of light emitted by a black body at a given temperature. Used to calculate Correlated Color Temperature (CCT) of white light sources in Kelvin (K).

0.0400 500

𝜆𝜆/nm600 700

0.5

1.0

1.5

2.0 𝒙𝒙𝒙𝒙𝜆𝜆𝒙𝒚𝒚𝒙𝒙𝜆𝜆𝒙𝒛𝒛𝒙𝒙𝜆𝜆𝒙

CIE Color Matching FunctionsMacAdam ellipses define areas of color on the CIE chromaticity diagram that are indistinguishable to the human eye (ellipses shown here at ten times their actual size).

Color gamut illustrates the RGB limits reproducible by digital displays.Calculating CIE color coordinates of a stimulus:

For a spectral power distribution (SPD), given as 𝑃𝑃𝒙𝜆𝜆𝒙, the XYZ tristimulus values are computed as:

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

CIE 1931

CIE 1931

sRGB

DCI-P3

Rec.2020

CIE 1976

0.1

0.2

0.3

0.4

0.5

0.6

0 0.1 0.2 0.3 0.60.4 0.5

sRG

BDCI-P

3Rec

.202

0

Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically evaluate light, color, and surface quality. Calibrated to simulate human visual perception, Radiant cameras perform scientific measurement of displays and illuminated components to ensure quality that accurately reflects human experience.

www.RadiantVisionSystems.com Copyright ©2019 Radiant Vision Systems LLC. All rights reserved. 2019/03/21

Anatomy of the CIE Chromaticity Diagram

𝑋𝑋𝑋𝑋𝑃𝑃𝒙𝜆𝜆𝒙𝒙𝒙𝒙𝒙𝜆𝜆𝒙𝑋𝑋𝜆𝜆𝑌𝑌𝑋𝑋𝑃𝑃𝒙𝜆𝜆𝒙𝒚𝒚𝒙𝒙𝜆𝜆𝒙𝑋𝑋𝜆𝜆𝑍𝑍𝑋𝑋𝑃𝑃𝒙𝜆𝜆𝒙𝒛𝒛𝒙𝒙𝜆𝜆𝒙𝑋𝑋𝜆𝜆

Cx𝑋X/𝒙𝒙X+Y+Z𝒙𝒙

Cy𝑋Y/𝒙𝒙X+Y+Z𝒙𝒙

u′𝑋4X/𝒙𝒙X+15Y+3Z𝒙𝒙

v′𝑋9Y/𝒙𝒙X+15Y+3Z𝒙𝒙

CIE 1931 CIE 1976

v’

u’

y

x

0.1

0.2

0.3

0.4

0.5

0.6

0 0.1 0.2 0.3 0.60.4 0.5

v’

u’

CIE 1976

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

y

x

ExampleStimulus

490

Principles of Color Measurement

460

480

500

520

540

560

580

600

620

Kelvin ( K )

10000

6000

4000

3000

25003000

2000

1500

100%50%0%

Spectral Locus Colors at maximum saturation.

Dominant Wavelength Intersection point on the spectral locus by a line originating at white and then projected through a given color.

Purity (Excitation Purity) The ratio of the distance of white to a stimulus with respect to the distance of white to the dominant wavelength on the spectral locus (expressed as a percentage).

Planckian Locus Color of light emitted by a black body at a given temperature. Used to calculate Correlated Color Temperature (CCT) of white light sources in Kelvin (K).

0.0400 500

/nm600 700

0.5

1.0

1.5

2.0

CIE Color Matching FunctionsMacAdam ellipses indistinguishable to the human eye (ellipses shown here at ten times their actual size).

Color gamut illustrates the RGB limits reproducible by digital displays.Calculating CIE color coordinates of a stimulus:

For a spectral power distribution (SPD), given as , the XYZ tristimulus values are computed as:

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

CIE 1931

CIE 1931

sRGB

DCI-P3

Rec.2020

CIE 1976

0.1

0.2

0.3

0.4

0.5

0.6

0 0.1 0.2 0.3 0.60.4 0.5

sRG

BDCI-P

3Rec

.202

0

Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically evaluate light, color, and surface quality. Calibrated to

www.RadiantVisionSystems.com Copyright ©2019 Radiant Vision Systems LLC. All rights reserved. 2019/03/21

Anatomy of the CIE Chromaticity Diagram

Cx=X/((X+Y+Z))

Cy=Y/((X+Y+Z))

CIE 1931 CIE 1976

v’

u’

y

x

0.1

0.2

0.3

0.4

0.5

0.6

0 0.1 0.2 0.3 0.60.4 0.5

v’

u’

CIE 1976

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

y

x

ExampleStimulus

490

Page 2: Principles of Color Measurement - Radiant Vision Systems...Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically

Radiant Vision Systems engineers imaging colorimeters and photometers, application software, and goniometric systems to critically evaluate light, color, and surface quality. Calibrated to simulate human visual perception, Radiant cameras perform scientific measurement of displays and illuminated components to ensure quality that accurately reflects human experience.

www.RadiantVisionSystems.com Copyright ©2019 Radiant Vision Systems LLC. All rights reserved. 2019/03/21

The Language of Light

Human Photopic Response

The human eye is not equally sensitive to wavelengths of light across the visible electromagnetic spectrum. Sensitivity of the human photopic response is given by the CIE 1931 luminosity function for a standard photopic observer (aligns with the 𝒚𝒚 color matching curve), peaking at around 555 nanometers. Thus, green wavelengths are typically brightest to the eye.

Spectral Power Distribution Every light source is defined by its unique spectral power distribution (SPD), which is the radiant power (Watts) emitted by the light source at each wavelength in the visible electromagnetic spectrum.

Units of Measurement: Photometry

Photometry vs. Radiometry

Luminance(candela per meter squared)

Luminous Intensity(candela)

Illuminance(lumens per meter squared)

Luminous Flux(lumens)

Lumen (lm)lm = cd * sr

Luminous Flux

Candela (cd)cd = lm/sr

Luminous Intensity

Lux (lm/m2)

Illuminance

cd/m2

1 cd/m2 = 1 nit

Luminance

Total output of a lightsource in all directions

Luminous flux emittedper unit solid angle

Light incident on asurface per unit area

Directional lightemitted or reflectedback from a surfaceper unit solid angle

AB

PhotometricHuman Visual Perception

RadiometricAll Radiation

LUMINOUS FLUXlumens (lm)

1 Im = 1 cd * 1 sr

RADIANT FLUXWatts (W)

LUMINOUS INTENSITYcandela (cd)1 cd = 1 lm/sr

RADIANT INTENSITYW/sr

ILLUMINANCE

LUMINANCE

IRRADIANCEW/m2

RADIANCEW/sr * m2cd/m2

1 cd/m2 = 1 nitfoot-lambert (fL)

1 fL = 1/π * cd/ft2

lux1 lx = 1 lm/m2

foot-candle (fc)1 fc = 1 lm/ft2

Total light output

Light froma direction

Light incidenton a surface

Brightness

Wavelength 𝜆𝜆 (nm)

Wavelength 𝜆𝜆 (nm)cd = candela sr = steradian lm = lumen W = Watt fc = foot-candle fL = foot-lambert

390

400

410

420

430

440

450

460

470

480

500

510

520

530

540

550

560

570

580

590

600

610

620

630

640

650

670

680

690

700

SpectralSensitivity of

the Human Eye

D65 (Sunlight)

Illuminant A

LCD Blue

LED Red

HeNe Laser

390 430 470 510 550 590 630 670 710