MONOXIVENT - SOURCE CAPTURE SYSTEMS 1306 Mill St., Rock Island, IL 61201 877-608-4383 - [email protected] monoxivent.com Mini Arms 2" & 3" Diameter Operation and Maintenance Manual

Transcript of Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning...

Page 1: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

MONOXIVENT - SOURCE CAPTURE SYSTEMS

1306 Mill St., Rock Island, IL 61201877-608-4383 - [email protected]

monoxivent.com

Mini Arms2" & 3" Diameter

Operation and Maintenance Manual

Brady
Rectangle
Brady
Line
Page 2: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

11306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm.

The sole purpose of the Monoxivent mini extractor arm is source capture. This is achieved by loosening the knobs on the three polypropylene joints, positioning the arms for the most effective capture and hand tightening each joint until the arm maintains position. It is recommended that positioning of the arm is best achieved by attending each joint individually starting at the top and working towards the bench top.

Air flow can be regulated by the manual damper located just above the hood after the first adjustment joint.

Exterior cleaning of the extractor arm is an individual preference as this will not hinder the extraction of fumes through the arm.

An interior inspection is recommended periodically, especially if the extractor is not removing fumes as previous. A complete interior inspection can be performed by disassembly of each adjustable joint, removing each section of the arm, and verifying the arm is clean of debris. At this time each O-ring should be inspected for cracks and tears.

Brady
Line
Brady
Line
Page 3: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

21306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

Torx 30

Torx 30

Torx 30

1

Ø 5/16”

Ø 4”Ø 3/16”

1

2L R

Brady
Line
Brady
Line
brady
Line
Page 4: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

31306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

3Torx 30 A

A B C

BC

!D E

4

5

Brady
Line
Brady
Line
brady
Line
brady
Line
Page 5: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

41306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

13a

16

2425

1920

212223

1718

29

30

32

33

35

34

13b

15

2627

14

31

36

1 3

6 9

28

2

4

5

7

3

3

8

12

10

11

3

7

Brady
Line
Brady
Line
brady
Line
brady
Line
brady
Line
brady
Line
brady
Line
Page 6: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

51306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

Extractors with designation MV always include a wall bracket

B

10

Ø C

Ø 3 7/8�

A

10

Ø C

Ø 3 7/8�

A

10

Ø 3 7/8�

3 1

/8�

3 5/8�4 3

/16

�Ø 3/10� (4X)

Ø C

A

B

10

Ø 3 7/8�

3 1

/8�

A

Ø C

Models with two and three jointsCeiling mountDesignation with three joints Size (inches) Weight

Standard PP ESD ATEX A B Total length Ø C lb

MT 1000-50 - ES - 16� 12� 40� 2� 3.3

MT 1300-50 - ES - 22� 18� 51� 2� 3.6

MT 1500-50 - ES - 30� 18� 60� 2� 3.7

MT 1000-75 PP - - 16� 12� 40� 3� 5.0

MT 1300-75 PP - - 22� 18� 51� 3� 5.7

MT 1500-75 PP - - 30� 18� 60� 3� 6.0

MT 2000-75 - - - 40� 26� 80� 3� 7.1

Designation with two joints

MT 650-50 - ES - 12� - 28� 2� 2.2

MT 750-50 - ES - 18� - 34� 2� 2.4

Wall mount

Designation with three joints Size (inches) Weight

Standard PP ESD ATEX A B Total length Ø C lb

MV 1000-50 - ES - 16� 12� 40� 2� 8.0

MV 1300-50 - ES - 22� 18� 51� 2� 8.3

MV 1500-50 - ES - 30� 18� 60� 2� 8.4

MV 1000-75 PP - - 16� 12� 40� 3� 9.7

MV 1300-75 PP - - 22� 18� 51� 3� 10.4

MV 1500-75 PP - - 30� 18� 60� 3� 10.7

MV 2000-75 - - - 40� 26� 80� 3� 11.8

Designation with two joints

MV 650-50 - ES - 12� - 28� 2� 6.9

MV 750-50 - ES - 18� - 34� 2� 7.1

Table mount

Designation with three joints Size (inches) Weight

Standard PP ESD ATEX A B Total length Ø C lb

MB 1000-50 - ES - 16� 12� 40� 2� 3.3

MB 1300-50 - ES - 22� 18� 51� 2� 3.6

MB 1500-50 - ES - 30� 18� 60� 2� 3.7

MB 1000-75 PP - - 16� 12� 40� 3� 5.0

MB 1300-75 PP - - 22� 18� 51� 3� 5.7

MB 1500-75 PP - - 30� 18� 60� 3� 6.0

Designation with two joints

MB 650-50 - ES - 12� - 28� 2� 2.2

MB 750-50 - ES - 18� - 34� 2� 2.4

Extractors with designation MB always include a mounting plate

B

8�

10

8 5/8�

Ø C

A

Ø 3 7/8�

10

8 5/8�

A

Ø 3 7/8�

Ø C

2 1

/2�

8�

2 1

/2�

Model with gas spring MTS*

Air flow cfm

The static pressure drop is measured in accordance with ISO-standard 5167-1. The noise level is measured in accordance with ISO-standard 3741.

Pressure drop diagram models 2, 3 and 4 inch

Sta

tic p

ressure

dro

p (

in/w

g)

0.8

1.6

2.4

3.2

20 40 60 80 100 120 140

49

4237

31

51

48

44

3936

160 180 200 220 240 260 280

34

45

51

2.8

1.2

2.0

0.4

Ø 3� 3 jointsØ 2� 3 joints

Ø 2� 2 joints

Ø 4� 3 joints

56 dB(A)56 dB(A)

52 dB(A)

Brady
Line
Brady
Line
Page 7: Operation and Maintenance Manual · Installation of the Monoxivent extractor arm before beginning operation of ceiling, wall, or bench mount arm. The sole purpose of the Monoxivent

Mini Arms 2 & 3 Inch

61306 Mill St., Rock Island, IL 61201 877-608-4383

[email protected] monoxivent.com

Recommendedmounting height

Designation H inches

MT 1000 64-76

MT 1300 72-84

MT 1500 76-88

MT 2000 84-96

Recommended sidedisplacement in relationto workplace

Designation S inches

MT 1000 12-24

MT 1300 16-28

MT 1500 20-32

MT 2000 28-40

Recommendedmounting height

Designation H inches

MT 1150 68-76

MT 1350 76-84

MT 1650 80-88

MT 1900 88-96

MT 2100 92-100

MT 2650 92-100

Recommended sidedisplacement in relationto workplace

Designation S inches

MT 1150 12-24

MT 1350 16-28

MT 1650 20-32

MT 1900 28-32

MT 2100 28-36

MT 2650 52-36

Typical values for Terfu 4 inch

Range at recommended mountingheight, Terfu 2 - 3 inch

Range at recommendedmounting height, Terfu 4 inch

Size Function Air flow

4 inch 40-80 l/s 85-170 cfm

80 l/s 170 cfm80 l/s 170 cfm

LaboratoryLight industry:-Gluing-Solvents-Brazing-Light welding/TIG

100-125 l/s 210-265 cfm110-125 l/s 235-265 cfm

Size Function Air flow

2 inch Soldering 20 l/s 42 cfmHair dresser 17 l/s 37 cfm

3 inch Laboratory 40 l/s 85 cfmSchools 40 l/s 85 cfm

80 inch.

80 inch.

MT

2000

MT

1500

MT

1300

MT

1000

MT 2000MT 1500MT 1300MT 1000

Maximum radius

MT 2000 MT 1500 MT 1300 T1000

36�

20 40 60

20 40 60

36�

Working radius

H

S

The following mounting heights and sidedisplacements are recommended for optimizationof evacuation.

The following mounting heights and sidedisplacements are recommended for optimizationof evacuation.

100 inch.80

MT

2650

MT

1650

MT

1350

MT

1150

MT

2100

MT

1900

MT 2650MT 2100MT 1900MT 1650MT 1350MT 1150

MT 2650 MT 2100 MT 1900 MT 1650 MT 1350 MT 1150

20 40 60

Working radius

36�

Maximum radius

80 inch.20 40 60

S

*For ceiling mounting theextractor is recommendedto be positioned over thebenchtop, taking otherinterior furnishing inconsideration.

36�

H

*

*

80

20

40

60

100 inch.

80

20

40

60

100 inch.

*For ceiling mounting theextractor is recommendedto be positioned over thebenchtop, taking otherinterior furnishing inconsideration.

Brady
Line
Brady
Line