ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

16
George Mavunganidze BMLS (Hons) (UZ), MBA (ESAMI) ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ANALYSERS: ZIMBABWE EXPERIENCE 35 th SANBTC 2019– SUN CITY , SOUTH AFRICA 6 AUGUST 2019

Transcript of ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Page 1: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

George MavunganidzeBMLS (Hons) (UZ), MBA (ESAMI)

ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING

ANALYSERS: ZIMBABWE EXPERIENCE35th SANBTC 2019– SUN CITY ,

SOUTH AFRICA

6 AUGUST 2019

Page 2: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

1. Introduction & Background2. Aims & Objectives3. Study Design & Methods4. Results5. Conclusion 6. Recommendations7. Acknowledgements

Presentation Format

Page 3: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Introduction

• RBC Immunohematology a key component for blood safety through donations testing.

• ABO & RhD (Forward, Reverse); Ab Screening• AfSBT Accreditation standards requirement:

Section B, Technical requirements 4.2.4: “Serum or plasma from donors shall be tested for unexpected red cell antibodies,….plasma from units with clinically significant Abs shall not be used for transfusion.” High titre allo-antibodies.

Page 4: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

ZIMBABWE MAP

Page 5: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Background

• NBSZ automated its donations blood grouping in 2011, (Tango Optimo, Bio-Rad).

• Replacing the Conventional Test Tube (CTT) method which remained in use for confirmatory blood group testing.

• The need for a robust/ higher throughput automated blood grouping system. Two systems evaluated

• The JOURNEY – Kigali, AfSBT(2016), Cairo (2016), Tender (2017), Eval (Apr 1018) Go Live (April 2019)

Page 6: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Selection of the Blood Group Testing Systems

Page 7: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Aims and Objectives

To evaluate the performance and efficiency of two fully automated immuno-hematological systems, and assess their role in donor blood group and antibody screening at NBSZ, with the intention of selecting one for routine use.

Page 8: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Study Design and Methods

• Two automated analysers (Qwalys 3 – erythro-magnetic technology) and (NEO IRIS Solid phase red cell adherence) were compared with the Tango Optimo – Gold Standard.

• Comparing: ABO & RhD typing and Ab Screening• Operational performance was assessed on

throughput, loading options, reagent shelf life and instrument errors.

Page 9: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Sample TestingSample Testing NEO IRIS QWALYS 3

ABO Blood Grouping 6012 6118

RhD Blood Grouping 6012 6118

Weak D Testing 198 22

Irregular Ab Screening 6167 6082

Irregular Ab Screening 12 N/A

High ABO Antibody Titre 3453 N/A

Page 10: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Results -SummaryTable 1: NEO Iris and Qwalys3 Results

Tango Optimo/CTT

NEO Iris Qwalys3

ABO Rhesus Weak D ABO Rhesus Weak D

Concordance 99.5% 99.25% 99% 93.7% 93.3% *

Discordance 0 0.03% 0.5% 0 0.4% *Uninterpretable 0.5% 0.72% 0.5% 6.3% 6.3% *

Total 100% 100% 100% 100% 100% *

Page 11: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Results,…Ctd• The NEO IRIS showed a concordance in results for

5982/6012 (99.5%) ABO and 5967/6012 (99.25%) RhD. There were 64/6012 (1.1%) samples with initially un-interpretable results.

• The Qwalys 3 showed a concordance in results for 5734/6118 (93.7%) ABO and 5708/6118 (93.3%) RhD. 384/6118 (6.3%) were un-interpretable

• 198 samples were tested for weak D on NEO IRIS whilst only 22 were tested on Qwalys 3 due to reagent outages/ expiries machine downtime.

Page 12: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

User Experience Comparison SummaryTechnology/ Analyser NEO Iris Qwalys3 loaded samples at a time 176 144

STAT Function Present Not Present

Throughput Less time for ABO Rh and antibody screen (135.1 min)

More time for ABO Rh and antibody screen (153.4 min)

Antibody Identification Yes No

ABO Antibody Titration Yes No

Test Run Flexibility Dynamic work flow Unable to change work flow pattern within run

Assay order flexibility and reflex testing

Ability to perform multiple tests on each sampleCustomize Reflex tests: Weak D & antibody ID

Only module testing possible: -ABO/Rh and Ab Screen module-Weak-D module-Rhesus phenotyping module

Liquid waste container Can be directly connected to the sink or drain

Emptied only possible by user

Page 13: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Conclusion

• Both analysers were easy to operate and are complete walk away with alarms systems for errors.

• Based on: high concordance of ABO/RhD results, lesser un-interpretable results encountered, higher throughput & TAT and more flexibility in assay ordering and adjustment of workflow, the NEO IRIS was selected for routine use at NBSZ.

Page 14: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Recommendations

1. IMMUCOR’S high throughput, fully automated NEO IRIS can be used for routine, ABO, RhD & Ab Screening for maximum productivity

2. Onsite evaluation is the way to go before procurement especially for high cost Laboratory equipment.

Page 15: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...

Acknowledgements

1. National Blood Service Zimbabwe Management and Staff

2. IMMUCOR & HAEMOTEC (Qwalys3)

3. SANBTC 2019 for the opportunity to share our experience

Page 16: ONSITE EVALUATION OF TWO AUTOMATED BLOOD GROUPING ...