A brief history of OCTA brief history of OCT
Nawat Watanachai
Sir Charles Gairdner Hospital
October 2007
Nawat Watanachai
Sir Charles Gairdner Hospital
October 2007
![Page 2: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/2.jpg)
![Page 3: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/3.jpg)
•Huang, Hee, Fujimoto, Puliafito 1991 Huang, Hee, Fujimoto, Puliafito 1991 •1st in vivo retinal images at the MIT in 19931st in vivo retinal images at the MIT in 1993•Commercially available in 1996Commercially available in 1996
•Optical B-scans (Cross-sections)Optical B-scans (Cross-sections)•Resolution ~ 10 micronsResolution ~ 10 microns
OCT : Optical Coherence TomographyOCT : Optical Coherence Tomography
1st OCT image of the retinaResolution: 17 μmdepth ~1.5 mm into the tissue
![Page 4: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/4.jpg)
BASIC OCTBASIC OCT
![Page 5: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/5.jpg)
Clinical ApplicationsClinical Applications Visualization of vitreoretinal interface Cross-sectional visualization retinal pathology Nerve fiber layer analysis for glaucoma Optic nerve head analysis for glaucoma Other jobs eg. Anterior segment
Visualization of vitreoretinal interface Cross-sectional visualization retinal pathology Nerve fiber layer analysis for glaucoma Optic nerve head analysis for glaucoma Other jobs eg. Anterior segment
![Page 6: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/6.jpg)
Does OCT make a difference?Does OCT make a difference?
Clinical Diagnosis/ Decision Making
Medical Record/ Documentation
Patient Education/ Satisfaction
Clinical Diagnosis/ Decision Making
Medical Record/ Documentation
Patient Education/ Satisfaction
![Page 7: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/7.jpg)
Medical ProofMedical Proof OCT provides objective evidence for treatment
decisions, both to treat or not to treat, and of response to treatment
Examples: cystoid macular oedema retinal vein occlusion diabetic maculopathy Others eg fleck diseases
OCT provides objective evidence for treatment decisions, both to treat or not to treat, and of response to treatment
Examples: cystoid macular oedema retinal vein occlusion diabetic maculopathy Others eg fleck diseases
CNV and AMDMacular holePMF
![Page 8: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/8.jpg)
Time domain OCTTime domain OCT
Current clinical OCT devices utilize time domain technology ---> limit in signal acquisition time
700-900 nm* Z Dimension: Samples tissue with 1024 data points
over 2mm depth
X-Y Dimension: Takes a sample every 5-60 microns apart, 128-512 scans
Current clinical OCT devices utilize time domain technology ---> limit in signal acquisition time
700-900 nm* Z Dimension: Samples tissue with 1024 data points
over 2mm depth
X-Y Dimension: Takes a sample every 5-60 microns apart, 128-512 scans
![Page 9: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/9.jpg)
Time domain OCT : problemsTime domain OCT : problems 1 detector, Each A-scan requires the movement of
several mechanical parts
Excellent image quality…if device and patient are not moving
1 detector, Each A-scan requires the movement of several mechanical parts
Excellent image quality…if device and patient are not moving
![Page 10: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/10.jpg)
Time domain OCT : problemsTime domain OCT : problems
~400 A-scan per sec take time to catch the image Can give poor image resolution
No image registration --> not easily reproduce Error in both quantitative and qualitative
measurement in some situations
~400 A-scan per sec take time to catch the image Can give poor image resolution
No image registration --> not easily reproduce Error in both quantitative and qualitative
measurement in some situations
![Page 11: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/11.jpg)
So, what are we looking for in the newer OCTs?So, what are we looking for in the newer OCTs?So, what are we looking for in the newer OCTs?So, what are we looking for in the newer OCTs?
• Higher resolutionHigher resolution• Higher imaging speedHigher imaging speed• Image registrationImage registration• Quantitative information Quantitative information
extractionextraction
• Higher resolutionHigher resolution• Higher imaging speedHigher imaging speed• Image registrationImage registration• Quantitative information Quantitative information
extractionextraction
![Page 12: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/12.jpg)
Newer OCTsNewer OCTs
1. Spectral domain OCT -- available
2. Other OCTs
2.1 Ultrahigh resolution OCT
2.2 Adaptive optic OCT
2.3 Hybrid machines eg OCT/SLO, HRA/OCT
3. OCT for anterior segment
1. Spectral domain OCT -- available
2. Other OCTs
2.1 Ultrahigh resolution OCT
2.2 Adaptive optic OCT
2.3 Hybrid machines eg OCT/SLO, HRA/OCT
3. OCT for anterior segment
![Page 13: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/13.jpg)
1. Spectral domain OCT1. Spectral domain OCT
![Page 14: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/14.jpg)
1. Spectral domain OCT1. Spectral domain OCT High speed high resolution OCT Frequency swept light source at around 850-1040 nm
Improved axial image resolutions (<6 microns axial resolution appears possible)
High speed high resolution OCT Frequency swept light source at around 850-1040 nm
Improved axial image resolutions (<6 microns axial resolution appears possible)
![Page 15: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/15.jpg)
1. Spectral domain OCT1. Spectral domain OCT
The output is measured with the use of a spectrometer (may be > 2,000 detector elements), doesn’t need moving part during scanning 16,000-75,000 scan/sec (TD-OCT ~400)
The output is measured with the use of a spectrometer (may be > 2,000 detector elements), doesn’t need moving part during scanning 16,000-75,000 scan/sec (TD-OCT ~400)
![Page 16: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/16.jpg)
1. Spectral domain OCT1. Spectral domain OCT
Real time display and data streaming capabilities enable video-rate imaging at more than 30fps
Real time display and data streaming capabilities enable video-rate imaging at more than 30fps
QuickTime™ and aCinepak decompressor
are needed to see this picture.
![Page 17: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/17.jpg)
1. Spectral domain OCT1. Spectral domain OCT Can produce 3-D projection image which can be aligned with the actual
fundus image to provide pixel-to-pixel registration 3-D imaging produces layers of information
Isolation of retinal layers makes image analysis possible in a broad fashion
3-D thickness map
Can produce 3-D projection image which can be aligned with the actual fundus image to provide pixel-to-pixel registration
3-D imaging produces layers of information Isolation of retinal layers makes image analysis possible in a broad
fashion 3-D thickness map
QuickTime™ and aMicrosoft Video 1 decompressorare needed to see this picture.
![Page 18: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/18.jpg)
1. Spectral domain OCT1. Spectral domain OCT
Pixel-to-pixel registration Precise scan location Image reproducibility
Some machines can generate fundus imaging and register the spatial location of each OCT section image
Pixel-to-pixel registration Precise scan location Image reproducibility
Some machines can generate fundus imaging and register the spatial location of each OCT section image
QuickTime™ and aCinepak decompressor
are needed to see this picture.
![Page 19: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/19.jpg)
1. Spectral domain OCT1. Spectral domain OCT
Pros Higher resolution/ 3D imaging Enhanced imaging speed Pinpoint registration
Cons Signal strength and depth resolution is dependent on the
path difference between the retina and the reference mirror --> the greater the distance (myopic eye), the weaker the signal and the lower the resolution
Yes………it’s price!
Pros Higher resolution/ 3D imaging Enhanced imaging speed Pinpoint registration
Cons Signal strength and depth resolution is dependent on the
path difference between the retina and the reference mirror --> the greater the distance (myopic eye), the weaker the signal and the lower the resolution
Yes………it’s price!
![Page 20: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/20.jpg)
2. Other OCTs2. Other OCTs
2.1 Ultrahigh resolution OCT
2.2 Adaptive optic OCT
2.3 Hybrid machines eg OCT/SLO, HRA/OCT
2.1 Ultrahigh resolution OCT
2.2 Adaptive optic OCT
2.3 Hybrid machines eg OCT/SLO, HRA/OCT
![Page 21: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/21.jpg)
2.1 Ultra-high Resolution OCT2.1 Ultra-high Resolution OCT
First described by Drexler and Fujimoto in 2001 Broad bandwidth, 150nm femtosecond Ti-sapphire
laser source (TD-OCT 10-25 nm) 1-3 microns axial resolution
First described by Drexler and Fujimoto in 2001 Broad bandwidth, 150nm femtosecond Ti-sapphire
laser source (TD-OCT 10-25 nm) 1-3 microns axial resolution
QuickTime™ and a decompressor
are needed to see this picture.
![Page 22: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/22.jpg)
2.1 Ultra-high Resolution OCT2.1 Ultra-high Resolution OCT
Video-rate with up to 25-50 B-scans/sec3,000 axial pixels and 600 transverse pixels (OCT-3 1,024 and 512 pixels)
Video-rate with up to 25-50 B-scans/sec3,000 axial pixels and 600 transverse pixels (OCT-3 1,024 and 512 pixels)
QuickTime™ and aCinepak decompressor
are needed to see this picture.
![Page 23: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/23.jpg)
2.1 Ultra-high Resolution OCT2.1 Ultra-high Resolution OCT
![Page 24: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/24.jpg)
2.1 Ultra-high Resolution OCT2.1 Ultra-high Resolution OCT Pros
Enhanced anatomic detail
Enhanced visualization of the subretinal CNV
Pros Enhanced anatomic
detail Enhanced
visualization of the subretinal CNV QuickTime™ and a
Cinepak decompressorare needed to see this picture.
![Page 25: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/25.jpg)
2.1 Ultra-high Resolution OCT2.1 Ultra-high Resolution OCT
Cons Slow acquisition time of 4-5 sec (OCT-3 can do
the job in 1.3 sec) Require significant technical support for day-to-
day operation Limited ability to localize scans in relation to
precise fundus landmarks Alignment of serial scans dependent on fixation is
inaccurate
Cons Slow acquisition time of 4-5 sec (OCT-3 can do
the job in 1.3 sec) Require significant technical support for day-to-
day operation Limited ability to localize scans in relation to
precise fundus landmarks Alignment of serial scans dependent on fixation is
inaccurate
![Page 26: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/26.jpg)
2.2 Adaptive optic OCT2.2 Adaptive optic OCT
Use electromagnetic deformable mirror (adaptive optic) to improve the spatial resolution Correct chromatic aberration of the eye to get
better light pathway
Use electromagnetic deformable mirror (adaptive optic) to improve the spatial resolution Correct chromatic aberration of the eye to get
better light pathway
![Page 27: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/27.jpg)
2.2 Adaptive Optic OCT2.2 Adaptive Optic OCT
Pixel resolution of 3µm x 3µm Very fine image : able to Imaging the retinal cells
Pixel resolution of 3µm x 3µm Very fine image : able to Imaging the retinal cells
![Page 28: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/28.jpg)
2.2 Adaptive optic OCT2.2 Adaptive optic OCT 3D video Voxel resolution of 3µm x 3µm x 3 µm
3D video Voxel resolution of 3µm x 3µm x 3 µm
QuickTime™ and aCinepak decompressor
are needed to see this picture.
![Page 29: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/29.jpg)
2.3 OCT/SLO : multimodal imaging2.3 OCT/SLO : multimodal imaging
Introduced by Podoleanu and Jackson in 1998
Employs OCT scan simultaneous with confocal SLO images SLO = surface detail OCT = internal detail
Introduced by Podoleanu and Jackson in 1998
Employs OCT scan simultaneous with confocal SLO images SLO = surface detail OCT = internal detail
![Page 30: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/30.jpg)
2.3 OCT/SLO : multimodal imaging2.3 OCT/SLO : multimodal imaging
Single illumination source with parallel detector systems ensures pixel-to-pixel correlation between views
Overlay capability permits view of internal anatomy beneath surface landmarks
Single illumination source with parallel detector systems ensures pixel-to-pixel correlation between views
Overlay capability permits view of internal anatomy beneath surface landmarks
QuickTime™ and aCinepak decompressor
are needed to see this picture.
![Page 31: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/31.jpg)
2.3 OCT/SLO : multimodal imaging2.3 OCT/SLO : multimodal imaging
Coronal OCT scans capture details often lost in fixation-driven B scanning
SLO channel facilitates integration of functional testing such as angiography, microperimetry, and mfERG
Coronal OCT scans capture details often lost in fixation-driven B scanning
SLO channel facilitates integration of functional testing such as angiography, microperimetry, and mfERG
![Page 32: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/32.jpg)
2.3 OCT/SLO2.3 OCT/SLO
TD-OCT
OCT/SLO
![Page 33: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/33.jpg)
Matching C-Scan Images Provide Linear Reference
![Page 34: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/34.jpg)
2.3 OCT/HRA2.3 OCT/HRA
OCT + laser angiography Image registration --> precisely define the scan
location -->good for repeat measurement Record fundus image in the same time
OCT + laser angiography Image registration --> precisely define the scan
location -->good for repeat measurement Record fundus image in the same time
![Page 35: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/35.jpg)
2.3 other multimodaling machines2.3 other multimodaling machines
Ultra-high Resolution OCT/SLO Combined Spectral Domain/ Time Domain High-
Resolution OCT/SLO
Ultra-high Resolution OCT/SLO Combined Spectral Domain/ Time Domain High-
Resolution OCT/SLO
![Page 36: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/36.jpg)
Will you own this machine?Will you own this machine?
Keeps in mind OCT has become gold standard for some diseases
eg. PMF, MH Non invasive
limitation OCT does NOT provide dynamic information (FA
does) Depend on operator techniques Degraded in the presence of media opacity Computer can do wrong things
Keeps in mind OCT has become gold standard for some diseases
eg. PMF, MH Non invasive
limitation OCT does NOT provide dynamic information (FA
does) Depend on operator techniques Degraded in the presence of media opacity Computer can do wrong things
![Page 37: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/37.jpg)
Which OCT is ‘the one’ for you?Which OCT is ‘the one’ for you?devices B-scan/ 3D Axial
resolutionScanning
speedNon-OCT imaging
(TD-OCT) Yes/no 10 400 Near IR
Heidelberg Spectralis HRA/OCT
Yes/yes 7 40,000 SLO, ICG, autoF
Optopol Copernicus 2
Yes/yes 6 25,000 Near IR
Optovue RTVue-1001
Yes/yes 5 26,000 Near IR
OTI OCT/SLO1 Yes/yes 5 28,000 SLO, ICG, microperimetry
Topcon 3D-OCT-10002
Yes/yes 6 18,000 Near IR/ color
Carl Zeiss Meditec Cirrus1
Yes/yes 5 27,000 LSLO
![Page 38: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/38.jpg)
Which OCT is ‘the one’ for you?Which OCT is ‘the one’ for you?
Time domain OCT (~60,000+ AUD) Spectral domain OCT (>100,000 AUD)
ResearchersRetinal specialists who can use this
images to help with decision making in selected difficult cases
Time domain OCT (~60,000+ AUD) Spectral domain OCT (>100,000 AUD)
ResearchersRetinal specialists who can use this
images to help with decision making in selected difficult cases
![Page 39: NW2007 Ophthalmic OCT](https://reader036.fdocuments.net/reader036/viewer/2022062417/54b4d11e4a7959bd488b480d/html5/thumbnails/39.jpg)
Thank youThank you