M494E Biostation IM Perfusion Unit

download M494E Biostation IM Perfusion Unit

of 50

Transcript of M494E Biostation IM Perfusion Unit

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    1/50

    Time Lapse Imaging System

    BioStation IM

    BS-IM Perfusion Unit

    Instructions

    M 9 E 08.6.NF.1

    *M494EN01*

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    2/50

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    3/50

    i

    Introduction

    Thank you for purchasing the Nikon products.

    This instruction manual is written for the users of the BS-IM Dispense Unit, an accessory for the Time lapse imaging

    system BioStation IM.

    To ensure correct usage, read this manual carefully before operating the product.

    No part of this manual may be reproduced or transmitted in any form without prior written permission from Nikon.

    The contents of this manual are subject to change without notice.

    Although every effort has been made to ensure the accuracy of this manual, errors or inconsistencies may remain. If

    you note any points that are unclear or incorrect, please contact your nearest Nikon representative.

    Some of the equipment described in this manual may not be included in the set you have purchased.

    If you intend to use any other equipment with this product, read the manual for that equipment too.

    If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may

    be impaired.

    Microsoft, Windows, and Internet Explorer are registered trademarks of Microsoft Corporation in the U.S. and other

    countries. Other product and company names mentioned in this manual are trademarks or registered trademarks of

    their respective owners.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    4/50

    ii

    Safety Precautions

    To ensure correct and safe operation, read this manual before using the product.

    Warning and Caution Symbols Used in This Manual

    Although this product is designed to be completely safe during use, incorrect usage or failure to follow the safety

    instructions provided may cause personal injury or property damage. To ensure correct usage, read this manual carefully

    before using the product. Do not discard this manual and keep it handy for easy reference.

    Safety instructions in this manual are marked with the following symbols to highlight their importance. For your safety,

    always follow the instructions marked with these symbols.

    Symbol Description

    WARNINGDisregarding instructions accompanying this symbol may lead to serious injury or death.

    CAUTIONDisregarding instructions accompanying this symbol may lead to injury or property

    damage.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    5/50

    Safety Precautions

    iii

    WARNING

    1. Intended application of this product

    The BS-IM Dispense Unit is a unit attached inside the BioStation IM (temperature controller area),

    allowing the user to inject or to replace test fluids and fluid substances without opening/closing the door

    or culture chamber. Do not use this product for other purpose.

    The product is classified as an in-vitro diagnostic device.

    2. Do not disassemble.

    Disassembling may cause malfunction and will lead to the forfeiture of all claims against warranty. Do not

    disassemble any part that is not indicated in this manual. If you notice any abnormality, contact your

    nearest Nikon representative.

    3. Read the instructions carefully.To ensure safety, carefully read this manual and the manuals for other equipment used with this product.

    Especially be sure to follow the warnings and cautions indicated in the first of the manual.

    4. Note on handing a hazardous specimen

    Before handling a biological specimen, its risk must be checked.

    When a hazardous specimen is used with this product, consult your safety supervisor or safety standard

    of your facility. When a potentially infectious specimen is used, you must wear rubber gloves to avoid

    direct touch.

    If such specimen is spilled onto this product, the portion must be decontaminated in a safety manner.

    Consult your safety supervisor or safety standard of your facility.

    CAUTION

    1. Disposal of the product

    To avoid biohazard risks, dispose of the product as contaminated equipment according to the standard

    procedure specified for your facility.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    6/50

    Safety Precautions

    iv

    Notes on Handling the Product

    1. Handling the Product.

    This product is a precision instrument. Handle the product with care to avoid a physical shock or impact.

    Cautions on moving the product

    Remove the test fluid bottle and disconnect the tubes before carrying the BioStation IM with the Dispense Unit

    attached.

    Cautions on assembling and installing the product

    Take care to avoid pinching your finger or hand.

    Scratches or fouling such as fingerprints on optical components (such as lenses or filter) will degrade

    microscope images.

    Be careful to avoid scratches or direct contact with the objective when assembling the Dispense Unit.

    2. Installation location and storage location

    This product is an accessory of the BioStation IM. Select an installation location and a storage location as specified

    for the BioStation IM. For details, refer to Safety Precautions in the separate manual BioStation IM Instructions

    .

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    7/50

    v

    ContentsIntroduction ............................................................................................................................................................ i

    Safety Precautions ................................................................................................................................................ ii

    Chapter 1 Equipment Overview......................................................................................................................... 1

    1.1 Components .............................................. .................................................... ................................ 1

    1.1.1

    Perfusion induction unit (MFC99010) ................................................ ................................. 1

    1.1.2

    Perfusion medium exchange unit (MFC99015) ..................................................... ............. 2

    1.2 Overview of the Perfusion Induction Unit................................... .................................................... 3

    1.3 Overview of the Perfusion Medium Exchange Unit............ ........................................................ .... 4

    1.3.1 Medium exchange - [degraded culture mediumfresh culture medium].......................... 5

    1.3.2 Medium exchange - [culture mediumculture medium containing test fluid] ................... 6

    1.3.3 Medium exchange

    - [culture mediumculture medium containing test fluid culture medium] ...................7

    Chapter 2 Names and Functions of Each Part ................................................................................................. 8

    2.1

    Parts Names and Functions of the Perfusion Induction Unit............................................. ............. 8

    2.2 Parts Names and Functions of the Perfusion Medium Exchange Unit - [degraded culture medium

    fresh culture medium] and [culture medium culture medium containing test fluid]................ 9

    2.3 Parts Names and Functions of the Perfusion Medium Exchange Unit

    - [culture mediumculture medium containing test fluidculture medium] ............................ 10

    Chapter 3 Attaching to the BioStation IM ....................................................................................................... 11

    3.1 Attaching the Perfusion Induction Unit.......................................................... ............................... 11

    3.1.1 Removing the culture chamber from the BioStation IM .................................................... 11

    3.1.2 Attaching the dispense needle assembly and bottle holder to the culture chamber ......... 13

    3.1.3 Attaching the culture chamber to the BioStation IM..................................................... ..... 14

    3.1.4

    Installing the bottle and connecting the tubes and fittings ................................................ 15

    3.2 Attaching the Perfusion Medium Exchange Unit

    (for [degraded culture mediumfresh culture medium] / [culture mediumculture medium

    containing test fluid] setup) ................................................... ....................................................... 17

    3.2.1 Removing the culture chamber from the BioStation IM .................................................... 17

    3.2.2 Attaching the dispense/drain needle assembly and connecting tubes ............................. 19

    3.2.3

    Attaching the culture chamber to the BioStation IM..................................................... ..... 22

    3.2.4 Connecting tubes from outside of the BioStation IM.................................................... ..... 23

    3.2.5 Piping arrangement for medium exchange

    (for [degraded culture mediumfresh culture medium] /

    [culture mediumculture medium containing test fluid] setup) ...................................... 25

    3.3 Adjusting the Needle Position......................... ....................................................... ...................... 26

    Attaching the Perfusion Medium Exchange Unit ......................................................... ............... 27

    (for [culture mediumculture medium containing test fluid culture medium] setup)................. 27

    1.

    Removing the culture chamber from the BioStation IM .................................................... 27

    2. Attaching the dispense/drain needle assembly and connecting tubes ............................. 27

    3. Attaching the culture chamber to the BioStation IM..................................................... ..... 32

    4. Connecting tubes from outside of the BioStation IM.................................................... ..... 32

    5. Piping arrangement for medium exchange

    (for [culture medium

    culture medium containing test fluid

    culture medium] setup)......... 34

    Chapter 4 Operating Procedures..................................................................................................................... 35

    4.1 Filling the Bottle with Culture Medium................................................ .......................................... 35

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    8/50

    Contents

    vi

    4.2 Setting and Removing the Specimen........................................... ................................................ 37

    4.2.1

    Setting the specimen.......... ....................................................... ....................................... 37

    4.2.2 Removing the Specimen ................................................... ............................................... 38

    4.3 Test Fluid Dispense/Medium Exchange ................................................ ...................................... 39

    4.3.1

    Dispensing procedure for test fluid, etc. ........................................................ ................... 39

    4.3.2 Draining procedure for culture medium, etc...................................................................... 40

    4.3.3 Focus adjustment ................................................... ....................................................... ... 40

    4.3.4 Use of peristaltic pumps ................................................... ................................................ 40

    Chapter 5 Care and Maintenance .................................................................................................................... 41

    5.1 Cleaning ............................................ ..................................................... ..................................... 41

    5.2 Parts Treatable in Autoclave Sterilization ................................................... ................................. 41

    Chapter 6 Specifications.................................................................................................................................. 42

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    9/50

    1

    Equipment Overview11

    The BS-IM Perfusion Unit is a unit attached inside the BioStation IM (temperature controller area), allowing the user to

    inject or to replace test fluids and fluid substances without opening/closing the door or culture chamber.

    All the components of the unit are attached to the interior of the temperature controlled area of the BioStation IM. This

    configuration eliminates the need for an additional device to warm replacement culture medium, and maintains the

    environment of tubes connected to the dish at a controlled temperature.

    1.1 Components

    This option unit includes perfusion induction unit and a perfusion medium exchange unit.

    1.1.1 Perfusion induction unit (MFC99010)

    This is the basic unit for the BS-IM Perfusion Unit. Mounting the perfusion induction unit to the BioStation IM culture

    chamber allows injection of test fluid to the culture medium without taking the dish out of the culture chamber.

    Components in the perfusion induction unit

    Dispense needle assembly(BS-P1): x1 Bottle: x1

    Tube Clamp (blue): x1 Fitting (female): x1

    Silicon tube (1100 mm): x1 Fitting (male) + silicon tube: x1

    L-shaped fitting + silicon tube: x2

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    10/50

    Chapter 1 Equipment Overview

    2

    1.1.2 Perfusion medium exchange unit (MFC99015)

    This unit, either partly or entirely, is used in combination with the perfusion induction unit. The perfusion medium

    exchange unit alone does not provide any function. The use of this unit allows culture medium replacement without

    taking the dish out of the culture chamber.

    Components in the perfusion medium exchange unit

    Drain needle assembly

    (BS-P2): x1

    Bottle: x1

    Tube clamp

    (red, yellow): x2

    Fitting (female): x2

    Silicon tube (1100 mm): x2

    Fitting (male): x1

    T-shaped fitting

    + L-shaped fitting

    + silicon tube: x1

    Silicon tube (1100 mm): x2

    Bottle holder: x1 Check valve: x2 Clamp: x2

    Hexagonal socket

    head bolt (M3 x 4): x2

    L-shaped fitting+ silicon tube: x2

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    11/50

    Chapter 1 Equipment Overview

    3

    1.2 Overview of the Perfusion Induction Unit

    This arrangement allows only test fluid injection to the culture medium.

    Operators can observe the sample reactions induced by test fluid injection.

    Figure 1.1-1 Overview of the perfusion induction unit

    Test fluid bottleCulture chamber

    Dish

    Objective

    Dispense

    needle

    BioStation IM (slide door)

    Pressure applied

    (from a syringe or peristaltic

    pump) to send test fluid

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    12/50

    Chapter 1 Equipment Overview

    4

    1.3 Overview of the Perfusion Medium Exchange Unit

    This unit allows users to exchange culture medium without taking the dish out of the culture chamber.

    This unit and the perfusion induction unit can be configured in two ways, and the two configurations can be used for

    three different applications.

    Table 1.1-1 Medium exchange combinations

    PartQua

    ntity

    Medium exchange -

    [degraded culture medium

    fresh culture medium] and

    [culture mediumculture

    medium containing test fluid]

    Medium exchange

    [culture mediumculture

    medium containing test

    fluidculture medium]

    Dispense needle assembly x1

    Bottle x1

    Tube clamp (blue) x1

    Fitting (female) x1

    L-shaped fitting + silicon tube x2 X

    Fitting (male) + silicon tube x1

    Perfus

    ioninductionunit

    Silicon tube (1100 mm) x1

    Drain needle assembly x1

    Bottle x1 X

    Bottle holder x1 X

    Tube clamp (red, yellow) x2 (x1)

    Fitting (female) x2 (x1)

    Fitting (male) x1

    Silicon tube (1100 mm) x2 (x1)

    Silicon tube (110mm) x2

    Hexagonal socket head bolt x2

    Check valve x2

    Clamp (for check valve) x2

    L-shaped fitting + silicon tube x2

    Per

    fusionmediumexchangeunit

    T-shaped fitting + L-shaped fitting+ silicon tube

    x1

    = used / X = not used / = partially used

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    13/50

    Chapter 1 Equipment Overview

    5

    Table 1.1-2 Medium exchange applications

    Name Application

    Medium exchange -

    [degraded culture medium

    fresh culture medium]

    Drain the degraded culture medium launch fresh culture medium

    Drain the culture medium degraded over prolonged culturing from the dish, and

    launch fresh culture medium.

    Medium exchange -

    [culture mediumculture

    medium containing test fluid]

    Drain culture medium launch culture medium containing test fluid

    Drain the culture medium in the dish, and launch culture medium containing test

    fluid.

    Medium exchange -

    [culture mediumculture

    medium containing test fluid

    culture medium]

    Drain culture medium launch culture medium containing test fluid drain

    culture medium containing test fluid launch culture medium.

    Drain the culture medium in the dish, and launch culture medium containing test

    fluid. Drain the culture medium containing test fluid in the dish, and launch

    culture medium.

    1.3.1 Medium exchange - [degraded culture mediumfresh culture medium]

    After draining degraded culture medium in the dish, launch fresh culture medium.

    Culture medium is exchanged by draining the dish of culture medium degraded over prolonged culturing, and then

    injecting fresh culture medium. This enables culturing for an extended period.

    Figure 1.1-2 Medium exchange - [degraded culture mediumfresh culture medium]

    Culture bottle

    Culture chamber

    Dish

    Objective

    Dispense

    needle

    BioStation IM (slide door)

    Pressure applied (from a

    syringe or peristaltic pump)

    to send fresh culture medium

    Pressure applied (from a

    syringe or peristaltic pump)

    to drain degraded culture

    medium

    Drain needle

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    14/50

    Chapter 1 Equipment Overview

    6

    1.3.2 Medium exchange - [culture mediumculture medium containing test fluid]

    After draining the culture medium in the dish, launch culture medium containing test fluid.

    This is mechanically identical to the medium exchange - [degraded culture medium fresh medium]. Drain the culture

    medium in the dish, then launch culture medium containing test fluid. This system allows the user to steadily manage theconcentration of the test substance and the time required to reach the sample.

    Figure 1.1-3 Medium exchange - [culture mediumculture medium with test fluid]

    Bottle of culture

    medium with test fluid

    Culture chamber

    Dish

    Objective

    Dispense

    needle

    BioStation IM (slide door)

    Pressure applied (from a

    syringe or peristaltic pump)to send culture medium

    containing test fluid

    Pressure applied (from a

    syringe or peristaltic pump)to drain culture medium

    Drain needle

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    15/50

    Chapter 1 Equipment Overview

    7

    1.3.3 Medium exchange -

    [culture mediumculture medium containing test fluid culture medium]

    After draining the culture medium, launch the culture medium containing a desired test fluid. Drain the culture medium

    containing the test fluid to launch the culture medium.Two bottles are provided with this exchange unit: One is applied for the culture medium, the other one for the culture

    medium containing the test fluid. After draining the culture medium from the dish, inject the culture medium containing

    the test fluid. After observing sample reaction to the test substance, repeat the procedures of draining and injecting the

    culture medium until almost any test substance is remained in the dish. In this way, you can verify if the test results are

    accurate.

    Figure 1.1-4 Medium exchange - [culture mediumculture medium containing test fluid culture medium]

    Bottle of culturemedium containingtest fluid

    Culture chamber

    Dish

    Objective

    Dispense

    needle

    BioStation IM (slide door)

    Pressure applied (from a

    syringe or peristaltic pump)

    to send culture medium

    containing test fluid

    Pressure applied (from a

    syringe or peristaltic

    pump) to drain culture

    medium/ culture medium

    containing test fluid

    Drain needle

    Culture bottle

    Check valve

    Pressure applied

    (from a syringe or

    peristaltic pump)

    to send culture

    medium

    Check valve

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    16/50

    8

    Names and Functions of Each Part22

    This chapter explains the name and function of each part in both units.

    2.1 Parts Names and Functions of the Perfusion Induction Unit

    This setup is for test fluid dispense only. The unit consists only of the parts used for test fluid dispense.

    Figure 2.1-1 Unit configuration

    Table 2.1-1 Parts names and functions of the perfusion induction unit

    Name Function

    Dispense needleassembly

    Needle assembly for dispensing test fluid into the dish.The position of the dispense needle can be adjusted axially and vertically. Additionally, the

    dispense needle can be set in the dispense position or the standby position. The needle

    should be set in the standby position for changing dishes.

    Piping parts Silicon tube and couplings.

    Bottle 1 bottle for holding the test fluid is installed.

    *Culture chamber and dish are not included in this product.

    Bottle

    Silicon tube (1100 mm)

    Fitting (male) + silicon tube

    Dispense needle assembly

    L-shaped fitting + silicon tube

    Tube clamp (blue)

    Fitting (female)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    17/50

    Chapter 2 Names and Functions of Each Part

    9

    2.2 Parts Names and Functions of the Perfusion Medium Exchange Unit -[degraded culture mediumfresh culture medium] and [culture mediumculture medium containing test fluid]

    This setup is for [draining degraded culture medium and injecting fresh culture medium], or [draining culture medium and

    injecting culture medium containing a test fluid].

    Components of both the perfusion induction unit and the exchange unit are used for this configuration.

    Figure 2.1-2 Unit configuration [degraded culture mediumfresh culture medium] and [culture medium culture

    medium containing test fluid]

    Table 2.1-2 Parts names and functions of the perfusion medium exchange unit [degraded culture mediumfresh culture

    medium] and [culture mediumculture medium containing test fluid]

    Name Function

    Dispense needle

    assembly

    Needle assembly for dispensing culture medium or culture medium containing test fluid into the dish.

    The position of the dispense needle can be adjusted axially and vertically. Additionally, the

    dispense needle can be set in the dispense position or the standby position. The needle should

    be set in the standby position for changing dishes.

    Drain needle

    assembly

    Needle assembly for draining culture medium or degraded culture medium from the dish.

    The position of the drain needle can be adjusted axially and vertically. Additionally, the drain

    needle can be set in the drain position or the standby position. Both the dispense and drain

    needles should be set in the standby position for changing dishes.

    Piping parts Silicon tube and couplings.

    Bottle 1 bottle for holding the culture medium or culture medium containing test fluid is installed.

    *Culture chamber and dish are not included in this product.

    Bottle

    Silicon tube (1100 mm)

    Fitting (male) + silicon tube

    Dispense needle assembly

    L-shaped fitting + silicon tube

    Tube clamp (blue)

    Fitting (female)

    Rotatable cover stopper

    Fitting (female)Drain needle assembly

    Silicon tube (1100 mm)

    Fitting (male) Tube clamp (red or yellow)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    18/50

    Chapter 2 Names and Functions of Each Part

    10

    2.3 Parts Names and Functions of the Perfusion Medium Exchange Unit - [culturemediumculture medium containing test fluidculture medium]

    This setup is for [draining culture mediumdispensing culture medium containing test fluid draining culture medium

    containing test fluid

    dispensing culture medium].All the components for both the perfusion induction unit and the exchange unit are used for this configuration.

    Figure 2.1-3 Unit configuration [culture mediumculture medium containing test fluidculture medium]

    Table 2.1-3 Parts names and functions of the perfusion medium exchange unit [culture mediumculture medium containing

    test fluidculture medium]

    Name Function

    Dispense needle

    assembly

    Needle assembly for dispensing culture medium or culture medium containing test fluid into the

    dish.

    The position of the dispense needle can be adjusted axially and vertically. Additionally, the

    dispense needle can be set in the dispense position or the standby position. Both the dispense

    and drain needles should be set in the standby position for changing dishes.

    Drain needle

    assembly

    Needle assembly for draining culture medium or degraded culture medium from the dish.

    The position of the drain needle can be adjusted axially and vertically. Additionally, the drain

    needle can be set in the drain position or the standby position.

    Both the dispense and drain needles should be set in the standby position for changing dishes.

    Check valve Prevents the culture medium and culture medium containing test fluid from mixing.

    Piping parts Silicon tube and couplings.

    Bottle Two bottles for holding the culture medium and culture medium containing test fluid are installed.

    *Culture chamber and dish are not included in this product.

    Bottle

    Silicon tube (1100 mm)

    Fitting (male) + silicon tube

    Dispense needle assembly

    L-shaped fitting +

    silicon tube

    Tube clamp (blue)

    Fitting (female)

    Tube clamp (red or yellow)

    Fitting (female)

    Drain needle assembly

    Silicon tube (1100 mm)

    Fitting (male)

    Silicon tube (110 mm)

    Silicon tube (110 mm)

    Check valve

    Clamp

    T-shaped fitting +

    L-shaped fitting +

    silicon tube

    Bottle holder

    Tube clamp (red or yellow)

    Rotatable cover stopper

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    19/50

    11

    Attaching to the BioStation IM33

    3.1 Attaching the Perfusion Induction Unit

    Necessary tool

    The following tool is required for the procedure below.

    Hexagonal screwdriver (supplied with BioStation IM)

    3.1.1 Removing the culture chamber from the BioStation IM

    When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in

    advance.

    1. Open the slide door on the BioStation IM.

    Figure 3.1-1 Opening the slide door

    2. Move the diascopic illuminator and duct out of

    the way.

    Figure 3.1-2 Moving the diascopic illuminator and

    duct out of the way

    3. Remove the temperature sensor.

    Figure 3.1-3 Removing the temperature sensor

    Slide door

    Diascopic illuminator

    Duct

    Temperature sensor

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    20/50

    Chapter 3 Attaching to the BioStation IM

    12

    4. Remove the exhaust tube and wet mixed gas

    tube.

    Figure 3.1-4 Removing the tubes

    5. Remove the culture chamber.

    Remove the three (x3) hexagonal socket head bolts using

    the hexagonal screwdriver.

    Figure 3.1-5 Removing the culture chamber

    Exhaust tube

    Hexagonal

    socket head

    bolt

    Hexagonal socket head bolt

    Wet mixed gas tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    21/50

    Chapter 3 Attaching to the BioStation IM

    13

    3.1.2 Attaching the dispense needle assembly and bottle holder to the culture chamber

    Attach the dispense needle assembly and bottle holder to the culture chamber that was removed earlier.

    1. Remove the right side plate of the culturechamber.

    Remove the two (x2) hexagonal socket head bolts using

    the hexagonal screwdriver.

    Be careful not to lose the removed plate or hexagonal

    socket head bolts.

    Figure 3.1-6 Removing the right side plate of

    culture chamber

    2. Attach the dispense needle assembly to the

    culture chamber.

    Attach the dispense needle assembly to the culture

    chamber using the two (x2) hexagonal socket head bolts.

    Figure 3.1-7 Attaching the dispense needle

    assembly

    CAUTION

    Caution in passing the dispense needle through the opening in the culture chamber

    When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the

    culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.

    Hexagonal socket

    head bolt

    Hexagonal socket

    head bolt

    Plate

    Hexagonal

    socket head

    bolt

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    22/50

    Chapter 3 Attaching to the BioStation IM

    14

    3.1.3 Attaching the culture chamber to the BioStation IM

    Attach the culture chamber preassembled with the dispense needle assembly to the BioStation IM.

    1. Attach the culture chamber.Attach the culture chamber to the BioStation IM using the

    three (x3) hexagonal socket head bolts.

    Figure 3.1-8 Attaching the culture chamber

    2. Connect the exhaust tube and wet mixed gas

    tube.

    Figure 3.1-9 Connecting the tubes

    3. Attach the temperature sensor.

    Figure 3.1-10 Attaching the temperature sensor

    Hexagonal socket

    head bolt

    Exhaust tube

    Temperature sensor

    Hexagonal socket head bolt

    Wet mixed

    gas tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    23/50

    Chapter 3 Attaching to the BioStation IM

    15

    3.1.4 Installing the bottle and connecting the tubes and fittings

    Place the bottle into the bottle holder and connect the tubes and fittings.

    1. Place the bottle into the bottle holder.

    Figure 3.1-11 Installing the bottle

    2. Connect the tubes and fittings.

    Connect the tubes and fittings as shown in the figure

    below.

    Figure 3.1-12 Connecting the tubes and fittings

    Bottle

    Silicon tube (1100 mm)

    Fitting (male) + silicon tube

    L-shaped fitting + silicon tube

    Fitting (female)

    Tube clamp (blue)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    24/50

    Chapter 3 Attaching to the BioStation IM

    16

    CAUTION

    Caution in connecting the tubes and fittings

    Make sure all tubes and fitting couplings are tightened. Test fluid or culture medium may leak from loose tube

    connections and cause contamination or damage this instrument.

    Pass the silicon tube (1100 mm x1) for dispense through

    the points shown in the figure on the right and pull the tube

    out of the BioStation IM.

    Figure 3.1-13 Routing silicon tubes

    CAUTION

    Caution in pulling out the silicon tube

    When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the

    BioStation IM.

    Silicon tube

    (1100 mm)

    for dispense

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    25/50

    Chapter 3 Attaching to the BioStation IM

    17

    3.2 Attaching the Perfusion Medium Exchange Unit(for [degraded culture mediumfresh culture medium] / [culture mediumculture medium containing test fluid] setup)

    This section describes the installation procedure for the perfusion medium exchange setup for [degraded culture medium

    fresh culture medium] and [culture medium culture medium containing test fluid].

    For installation procedures of the perfusion medium exchange setup for [culture mediumculture medium containing

    test fluidculture medium], refer to Section 3.4, Attaching the Perfusion Medium Exchange Unit (for [culture medium

    culture medium containing test fluidculture medium]).

    Necessary tool

    The following tool is required for the procedure below.

    Hexagonal screwdriver (supplied with BioStation IM)

    3.2.1 Removing the culture chamber from the BioStation IM

    When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in

    advance.

    1. Open the slide door on the BioStation IM.

    Figure 3.2-1 Opening the slide door

    2. Move the diascopic illuminator and duct out of

    the way.

    Figure 3.2-2 Moving the diascopic illuminator and

    duct out of the way

    Slide door

    Diascopic illuminator

    Duct

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    26/50

    Chapter 3 Attaching to the BioStation IM

    18

    3. Remove the temperature sensor.

    Figure 3.2-3 Removing the temperature sensor

    4. Remove the exhaust tube and wet mixed gas

    tube.

    Figure 3.2-4 Removing the tubes

    5. Remove the culture chamber.Remove the three (x3) hexagonal socket head bolts using

    the hexagonal screwdriver.

    Figure 3.2-5 Removing the culture chamber

    Temperature sensor

    Exhaust tube

    Hexagonal

    socket head

    bolt

    Hexagonal socket head bolt

    Wet mixed gas tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    27/50

    Chapter 3 Attaching to the BioStation IM

    19

    3.2.2 Attaching the dispense/drain needle assembly and connecting tubes

    Attach the dispense needle assembly and drain needle assembly to the culture chamber that was removed earlier.

    1. Remove the right and left side plates of theculture chamber.

    Remove the two (x2) hexagonal socket head bolts on each

    side using the hexagonal screwdriver.

    Be careful not to lose the removed plate or hexagonal

    socket head bolts.

    Of the plates removed, the left plate is made taller than the

    right plate to serve as a stopper for the rotatable cover.

    Make sure to return the plates to their correct positions

    when reassembling.Figure 3.2-6 Removing the right and left side plates

    of culture chamber

    2. Attach the dispense needle assembly to the

    culture chamber.

    Attach the dispense needle assembly to the culture

    chamber using the two (x2) hexagonal socket head bolts.

    Figure 3.2-7 Attaching the dispense needle

    assembly

    CAUTION

    Caution in passing the dispense needle through the opening in the culture chamber

    When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the

    culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.

    Hexagonal socket head bolt

    Plate

    Hexagonal socket

    head bolt

    Hexagonal

    socket headbolt

    Hexagonal socket head bolt

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    28/50

    Chapter 3 Attaching to the BioStation IM

    20

    3. Attach the drain needle assembly to the

    culture chamber.

    Attach the drain needle assembly to the culture chamber

    using the two (x2) special hexagonal socket head bolts.

    Figure 3.2-8 Attaching the drain needle assembly

    CAUTION

    Caution in passing the drain needle through the opening in the culture chamber

    When passing the drain needle through the culture chamber, be careful not to bump the needle tip against the

    culture chamber. This assembly may fail to operate properly, if the drain needle is deformed.

    4. Adjust the height of the drain needle.

    The height and axial position of both needles are adjustable. Normally, to attach the drain needle, adjust the height

    of the drain needle. Apply additional adjustments if necessary.For details on height adjustment procedures for the drain needle, refer to Section 3.3 "Adjusting the Needle

    Position".

    5. Connect the tubes and fittings to the bottle.

    1) Connect the L-shaped fittings and silicon tubes (2 sets)

    to the bottle.

    2) Connect the male fitting and silicon tube to the

    L-shaped fitting and silicon tube on the dispensing side

    of the bottle (with a tube inside).

    Figure 3.2-9 Connecting tubes to the bottle

    CAUTION

    Caution in connecting the tubes and fittings

    Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tubeconnections and cause contamination or damage this instrument.

    Hexagonal socket

    head bolt

    Hexagonalsocket head bolt

    Fitting (male)

    + silicon tube

    L-shaped fitting

    + silicon tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    29/50

    Chapter 3 Attaching to the BioStation IM

    21

    6. Place the bottle into the bottle holder and

    connect the tubes.

    Place the bottle with tubes connected into the bottle holder

    and insert the male fitting into the dispense needle.

    Figure 3.2-10 Installing and connecting tubes to the

    bottle

    Fitting (male)

    + silicon tube

    Bottle

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    30/50

    Chapter 3 Attaching to the BioStation IM

    22

    3.2.3 Attaching the culture chamber to the BioStation IM

    Attach the culture chamber preassembled with the dispense needle assembly and drain needle assembly to the

    BioStation IM.

    Figures below illustrate setup for medium exchange - [culture mediumculture medium containing test fluid]. Setupprocedure for this application is the same as the procedure for medium exchange - [culture mediumculture medium

    containing test fluidculture medium].

    1. Attach the culture chamber.

    Attach the culture chamber to the BioStation IM using the

    three (x3) hexagonal socket head bolts.

    Figure 3.2-11 Attaching the culture chamber

    2. Connect the exhaust tube and wet mixed gas

    tube.

    Figure 3.2-12 Connecting the tubes

    Hexagonal socket

    head bolt

    Exhaust tube

    Hexagonal socket

    head bolt

    Wet mixed

    gas tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    31/50

    Chapter 3 Attaching to the BioStation IM

    23

    3. Attach the temperature sensor.

    Figure 3.2-13 Attaching the temperature sensor

    3.2.4 Connecting tubes from outside of the BioStation IM

    Connect the silicon tubes for dispense and drain from outside of the BioStation IM.

    1. Insert the silicon tubes for dispense and drain.

    Pass the silicon tubes (1100 mm) for dispense (x1) and

    drain (x1) through the points shown in the figure on the

    right and pull the tubes into the BioStation IM.

    Figure 3.2-14 Routing silicon tubes

    CAUTION

    Caution in pulling out the silicon tube

    When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the

    BioStation IM.

    Temperature

    sensor

    Silicon tube

    (1100 mm)

    for dispense

    Silicon tube

    (1100 mm)

    for drain

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    32/50

    Chapter 3 Attaching to the BioStation IM

    24

    2. Connect each silicon tube (1100 mm) to the BS-IM Dispense Unit.

    Test fluid dispense tube

    Connect the silicon tube (1100 mm) to the L-shaped fitting and silicon tube on the air pressure side of the bottle

    (with no tube inside).

    Drain tube

    Connect the male fitting to the tip of the silicon tube (1100 mm) and connect the male fitting to the drain needle.

    Figure 3.2-15 Connecting the silicon tubes for dispense and drain

    CAUTION

    Caution in connecting the tubes and fittings

    Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tube

    connections and cause contamination or damage this instrument.

    3. Attach the male fitting and tube clamp to the

    silicon tubes for dispense and drain.

    Three tube clamps are provided: blue, red and yellow. Besure that each silicon tube is recognized when attaching it.

    Figure 3.2-16 Attaching the fitting and the tube

    clamp

    Silicon tube (1100 mm)

    for dispense

    Silicon tube (1100 mm) for drain

    Fitting (male)

    Tube clamp

    Fitting (female)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    33/50

    Chapter 3 Attaching to the BioStation IM

    25

    3.2.5 Piping arrangement for medium exchange

    (for [degraded culture mediumfresh culture medium] / [culture medium culture

    medium containing test fluid] setup)

    The piping arrangement for medium exchange [degraded culture mediumfresh culture medium] and [culture mediumculture medium containing test fluid] is as shown below.

    After connecting the tubes, make sure that the parts and tubes are connected correctly.

    Figure 3.2-17 Connecting the tubes and fittings

    CAUTION

    Caution in connecting tubes to the bottle

    Be careful not to connect the wrong tube to the bottle.The inlet with a tube: connect the silicon tube that leads to the dispense needle

    The inlet without a tube inside: connect the silicon tube (1100 mm) that leads to outside the BioStation IM.

    Silicon tube (1100 mm)

    Fitting (male) + silicon tube

    L-shaped fitting + silicon tube

    Fitting (male)

    Silicon tube (1100mm)

    Fitting (female)

    Fitting (female)

    Tube clamp (blue)

    Tube clamp (red or yellow)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    34/50

    Chapter 3 Attaching to the BioStation IM

    26

    3.3 Adjusting the Needle Position

    The needle has adjustments for height and axial positioning.

    Normally, only the height of the drain needle needs to be adjusted. The axial position of the drain needle and the

    dispense needle position should not require adjustment. Make these adjustments if needed.

    Example below describes how to adjust the position for the drain needle, which procedure is also applied to adjust the

    position for the dispense needle.

    1. Adjust the height of the drain needle.

    1) Slightly loosen the hexagonal socket head bolts that

    hold the drain needle assembly in place on the left and

    right sides.

    2) To adjust the clearance between the bottom surface of

    the dish and the tip of the drain needle, gently shake

    the left side of the mounting piece for the culture

    chamber up and down using the hexagonal socket

    head bolt on the right side as the pivot.

    3) Once adjusted, fix the culture chamber in place by

    tightening the left and right hexagonal socket head

    bolts.

    Figure 3.3-1 Adjusting the height of drain needle

    Adjusting the height of the drain needle

    The drain needle is made slightly longer than the dispense needle.

    When adjusting the height of the drain needle, aim to put the needle tip as close to the bottom of the dish aspossible. If the drain needle tip is off the bottom of the dish, the more the amount of culture medium or test fluid left

    undrained in the dish.

    2. Adjust the axial position of the drain needle.

    1) Loosen the hexagonal socket head bolt (x1) in the top

    of the needle assembly using the hexagonal

    screwdriver.

    2) Slide the cover on the needle in the axial direction to

    adjust the position of the tip of the drain needle.

    3) Once adjusted, tighten the hexagonal socket head bolt

    (x1) in the top of the needle assembly using the

    hexagonal screwdriver.Figure 3.3-2 Adjusting the axial position of drain

    needle

    Hexagonal socket head bolt (pivot side)

    Hexagonal socket

    head bolt

    Hexagonal socket head bolt (adjustment side)

    Cover on

    needle

    assembly

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    35/50

    Chapter 3 Attaching to the BioStation IM

    27

    Attaching the Perfusion Medium Exchange Unit(for [culture mediumculture medium containing test fluidculture medium] setup)

    1. Removing the culture chamber from the BioStation IM

    When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in advance.

    Refer to Section 3.2.1, Removing the culture chamber from the BioStation IM for details of the procedure.

    2. Attaching the dispense/drain needle assembly and connecting tubes

    Attach the dispense needle assembly, drain needle assembly, and bottle holder to the culture chamber that was removed

    earlier.

    1. Remove the bottle holder from the dispense

    needle assembly.Remove the hexagonal socket head bolt using the

    hexagonal screwdriver.

    Figure 2-1 Removing the bottle holder

    2. Combine the bottle holders.

    Take the bottle holder removed from the dispense needle

    assembly and combine it with the bottle holder included in

    the perfusion medium exchange unit using the two (x2)

    hexagonal socket head bolts.

    Figure 2-2 Combining the bottle holders

    3. Attach the clamp for the check valve to the

    combined bottle holders.

    Secure the clamp by inserting the tab on the clamp into

    the slotted mounting hole. With the clamp pressed against

    the bottle holder, pull upwards.

    Figure 2-3 Attaching the clamps

    Hexagonal socket head bolt

    Hexagonal sockethead bolt

    Clamp

    Bottle holder

    Hexagonal socket

    head bolt

    Clamp

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    36/50

    Chapter 3 Attaching to the BioStation IM

    28

    4. Connect the tubes and fittings to the check

    valve.

    When connecting, the arrows on the check valves should

    be pointed at the T-shaped fitting, L-shaped fitting, andsilicon tube.

    Figure 2-4 Connecting the tubes and fittings to

    the check valve

    CAUTION

    Caution in connecting the tubes and fittings to the check valve

    Be sure that the arrow on the check valve faces the correct direction. The check valve will not function if it is

    connected in the reverse direction.

    5. Connect the check valve.

    Attach the check valves to each clamp as shown in the

    figure on the right, and then pass the tube through the

    bottle holders.

    Figure 2-5 Connecting the check valves

    CAUTION

    Caution in connecting the check valves to the clamps

    Attach each check valve so that the ribs touch the inside of the clamp. The check valve is not fixed in place

    because the recessed surfaces do not touch the inside of the clamp.

    6. Attach the assembled bottle holders to the

    dispense needle assembly.

    Attach the assembled bottle holders to the dispense

    needle assembly using the hexagonal socket head bolt

    (x1).

    Figure 2-6 Attaching the bottle holders

    T-shaped fitting + L-shaped fitting + silicon tube

    Hexagonal socket head bolt

    Check valve

    Check valve

    Dispense needle

    assembly

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    37/50

    Chapter 3 Attaching to the BioStation IM

    29

    7. Remove the right and left side plates of the

    culture chamber.

    Remove the two (x2) hexagonal socket head bolts on each

    side using the hexagonal screwdriver.

    Be careful not to lose the removed plate or hexagonal

    socket head bolts.

    Of the plates removed, the left plate is made taller than the

    right plate to serve as a stopper for the rotatable cover.

    Make sure to return the plates to their correct locations

    when reassembling.Figure 2-7 Removing the right and left side plates

    of culture chamber

    8. Attach the dispense needle assembly to the

    culture chamber.

    Attach the dispense needle assembly to the culture

    chamber using the two (x2) hexagonal socket head bolts.

    Figure 2-8 Attaching the dispense needle

    assembly

    CAUTION

    Caution in passing the dispense needle through the opening in the culture chamber

    When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the

    culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.

    Hexagonal socket

    head bolt

    Hexagonal socket

    head bolt

    Hexagonal socket head bolt

    Plate

    Hexagonal

    socket head

    bolt

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    38/50

    Chapter 3 Attaching to the BioStation IM

    30

    9. Attach the drain needle assembly to the

    culture chamber.

    Attach the drain needle assembly to the culture chamber

    using the two (x2) special hexagonal socket head bolts.

    Figure 2-9 Attaching the drain needle assembly

    CAUTION

    Caution in passing the drain needle through the opening in the culture chamber

    When passing the drain needle through the culture chamber, be careful not to bump the needle tip against the

    culture chamber. This assembly may fail to operate properly, if the drain needle is deformed.

    10. Adjust the height of the drain needle.

    The height and axial position of both needles are adjustable. Normally, to attach the drain needle, adjust the heightof the drain needle. Apply additional adjustments if necessary.

    For details on height adjustment procedures for the drain needle, refer to Section 3.3 Adjusting the Needle

    Position.

    11. Connect the tube for the dispense needle.

    Connect the T-shaped tube assembly to the dispense

    needle with the male fitting and silicon tube.

    Figure 2-10 Tube connections for the dispense

    needle

    Hexagonal socket

    head bolt

    Hexagonal

    socket

    head bolt

    Fitting (male)

    + silicon tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    39/50

    Chapter 3 Attaching to the BioStation IM

    31

    12. Connect the tubes and fittings to each bottle.

    Connect the L-shaped fitting and silicon tube to the air

    pressure inlet (without a tube) of each bottle.

    Figure 2-11 Connecting tubes and fittings to the

    bottle

    13. Install the bottles.

    Place each bottle in the bottle holder.

    Figure 2-12 Placing the bottle

    14. Connect the tube between the check valve and

    bottle.

    Connect the silicon tube (110 mm) between the inlet fitting

    (with the tube inside) of the test fluid dispense bottle and

    the check valve.

    Figure 2-13 Placing the bottle and connecting

    tubes

    L-shaped fitting

    + silicon tube

    BottleBottle

    Silicon tube

    (110 mm)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    40/50

    Chapter 3 Attaching to the BioStation IM

    32

    3. Attaching the culture chamber to the BioStation IM

    Attach the culture chamber preassembled with the dispense needle assembly and drain needle assembly to the

    BioStation IM.

    Refer to Section 3.2.3, Attaching the culture chamber to the BioStation IM for details of the procedure.

    4. Connecting tubes from outside of the BioStation IM

    Connect the silicon tubes for dispense and drain from outside of the BioStation IM.

    1. Insert the silicon tubes for dispense and drain.

    Pass the silicon tubes (1100 mm) for dispense (x2) and

    drain (x1) through the points shown in the figure on the

    right and pull the tubes into the BioStation IM.

    Figure 4-1 Routing silicon tubes

    CAUTION

    Caution in pulling out the silicon tube

    When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the

    BioStation IM.

    Silicon tube

    (1100 mm)

    for dispense

    Silicon tube

    (1100 mm)

    for drain

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    41/50

    Chapter 3 Attaching to the BioStation IM

    33

    2. Connect each silicon tube (1100 mm) to the BS-IM Dispense Unit.

    Test fluid dispense tube

    Connect the silicon tube (1100 mm) to the L-shaped fitting and silicon tube on the air pressure side of the bottle

    (with no tube inside).

    Drain tube

    Connect the male fitting to the tip of the silicon tube (1100 mm) and connect the male fitting to the drain needle.

    Figure 4-2 Connecting the silicon tubes for dispense and drain

    CAUTION

    Caution in connecting the tubes and fittings

    Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tube

    connections and cause contamination or damage this instrument.

    3. Attach the male fitting and tube clamp to the

    silicon tubes for dispense and drain.

    Three tube clamps are provided: blue, red and yellow. Besure that each silicon tube is recognized when attaching it.

    Figure 4-3 Attaching the fitting and the tube clamp

    Silicon tube (1100 mm) for dispense

    Silicon tube (1100 mm) for drain

    Fitting (male)

    Tube clamp

    Fitting (female)

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    42/50

    Chapter 3 Attaching to the BioStation IM

    34

    5. Piping arrangement for medium exchange

    (for [culture mediumculture medium containing test fluid culture medium] setup)

    The piping arrangement for medium exchange [culture mediumculture medium containing test fluid culture

    medium] is as shown below.After connecting the tubes, make sure that the parts and tubes are connected correctly.

    Figure 5-1 Connecting the tubes and fittings

    CAUTION

    Caution in connecting tubes to the bottle

    Be careful not to connect the wrong tube to the bottle.

    The inlet with a tube: connect the silicon tube that leads to the check valve.

    The inlet without a tube inside: connect the silicon tube (1100 mm) that leads to outside the BioStation IM.

    Silicon tube (1100 mm)

    Silicon tube (110 mm)

    Connects the bottle and check valve

    Fitting (male) + silicon tube

    Connects the dispense

    needle and T-shaped fitting

    Silicon tube (1100 mm)

    L-shaped fitting

    + silicon tube

    Fitting (male)

    Silicon tube

    (1100mm)Fitting (female)

    Fitting (female)

    Tube clamp (blue)

    Tube clamp (red or yellow)

    Tube clamp (red or yellow)

    Check valve

    T-shaped fitting

    + L-shaped fitting

    + silicon tube

    Silicon tube (110 mm)

    Connects the bottle

    and check valve

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    43/50

    35

    Operating Procedures44

    This product does not have any pumps to dispense test fluid or drain culture medium.

    Normally, a customer-provided syringe or peristaltic pump is used to dispense test fluid or drain culture medium.

    4.1 Filling the Bottle with Culture Medium

    1. Open the slide door on the BioStation IM.

    2. Remove the bottle to fill with culture medium.

    Disconnect and remove the silicon tube from the bottle

    and remove the bottle.

    Figure 4.1-1 Removing the bottle

    3. Remove the lid from the bottle and fill with culture medium.

    When removing the lid from the bottle, check if the packing inside the lid is not damaged.Be sure there are no

    scratches or cracks or other signs of deterioration.

    CAUTION

    Packing used in the bottle lid

    The bottle lid disassembles into the parts shown in the

    figure on the right. The packing inside the lid should be

    replaced if any deterioration is found. Make sure it is fit

    in the lid after each replacement.

    To ensure the bottle is sealed air tight, make sure thepacking is installed properly.

    Figure 4.1-2 Packing in the bottle lid

    Packing

    Bottle lid

    Silicon tube

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    44/50

    Chapter 4 Operating Procedures

    36

    4. Return the bottle to the bottle holder and

    reconnect the tubes.

    Figure 4.1-3 Installing the bottle

    5. Close the slide door on the BioStation IM.

    CAUTION

    Caution on the temperature of the culture medium/test fluid in the bottle

    If the temperature of the culture medium or culture medium containing test fluid filled in the bottle is below

    37.0 C, allow the content to reach 37.0 C before starting your experiment by placing the bottle in the BioStation

    IM and allowing it to warm up.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    45/50

    Chapter 4 Operating Procedures

    37

    4.2 Setting and Removing the Specimen

    When setting and removing the specimen (dish), both the dispense needle and drain needle must be in their standby

    positions.

    4.2.1 Setting the specimen

    1. Open the slide door on the BioStation IM and move the diascopic illuminator out of the way.

    CAUTION

    Caution in moving the needle

    When moving the needles to the standby, dispense or drain positions, always hold and turn the needle by the

    metal base. Do not hold the silicon tubes, as this may loosen or disconnect the tube and cause it to break.

    2. Move the drain needle and dispense needle to

    their standby positions.

    1) Push the needle slightly toward the culture chamber

    until it moves out of the groove.

    2) Turn the needle left 90 degrees.

    3) Pull the needle toward you.

    The needle can be pulled out to the position shown in

    the figure below. Place the needle within the guide

    groove unless otherwise desired.

    Figure 4.2-1

    CAUTION

    Caution in pulling the needle to the limit.

    If you close the slide door on the BioStation with the

    needle pulled all the way out of the guide groove, the

    door will hit the needle and possibly break it.

    Figure 4.2-2 Moving the needle to the standby

    position

    1) Pull the needle out of the groove.

    2) Turn the needle left 90 degrees.

    3) Pull the needle toward you.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    46/50

    Chapter 4 Operating Procedures

    38

    3. Open the rotatable cover and set the dish.

    Set the dish near the center of the culture chamber.

    Figure 4.2-3 Setting the dish

    4. Move the drain needle and dispense needle to

    their drain and dispense positions,respectively.

    Return each needle to their original positions by following

    the procedures for moving them to the standby position in

    reverse.

    Figure 4.2-4 Moving each needle to the

    drain/dispense positions

    5. Close the rotatable cover on the culture

    chamber.

    Figure 4.2-5 Closing the rotatable cover on the

    culture chamber

    6. Return the diascopic illuminator to its original position and close the BioStation IM slide

    door.

    4.2.2 Removing the Specimen

    Remove the specimen (dish) by following the setting procedures.

    Dish

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    47/50

    Chapter 4 Operating Procedures

    39

    4.3 Test Fluid Dispense/Medium Exchange

    4.3.1 Dispensing procedure for test fluid, etc.

    1. Connect a silicon tube to a syringe, and the

    bottle containing test fluid or culture medium

    mixed with test fluid.

    Figure 4.3-1 Connecting to a pump

    2. Dispense the required amount of test fluid

    from the syringe.

    Open the tube clamp when injecting, then close it

    afterwards.

    Figure 4.3-2 Dispensing test fluid, etc.

    3. Record the timing of test fluid dispense in the

    software.

    - Cell stimulus mark setting button :

    For an explanation of each symbol, refer to the separate

    manual, BioStation IM Instructions , Chapter 4 Time-Lapse Images in Process

    Screen, Section 4.1.8 Time Line and Image Playback

    Control.Figure 4.3-3 Recording test fluid dispense

    Syringe

    Syringe

    Fitting (female)

    Tube clamp

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    48/50

    Chapter 4 Operating Procedures

    40

    4.3.2 Draining procedure for culture medium, etc.

    1. Connect the silicon tube from the discharge

    needle to a syringe.

    Figure 4.3-4 Connecting to a pump

    2. Draw culture medium into the syringe.

    Open the tube clamp when draining, then close it.

    Figure 4.3-5 Draining culture medium, etc.

    3. Record the timing of draining the culture

    medium in the software, if necessary.

    - Cell stimulus mark setting button :

    For an explanation of each symbol, refer to the separate

    manual, BioStation IM Instructions , Chapter 4 Time-Lapse Images in Process

    Screen, Section 4.1.8 Time Line and Image Playback

    Control.Figure 4.3-6 Recording draining of culture medium

    4.3.3 Focus adjustment

    After draining culture medium or dispensing test fluid to the dish, focus of the microscope may require adjustment as

    focus can be affected by changes in weight of dish content or even by the slightest changes in temperature. Therefore,

    verify the focus adjustment after medium exchange or test fluid dispense during time lapse examination. If the image

    goes out of focus, readjust the Z coordinate as described in Section 2.4, Observation Point and Focus Adjustment in

    the separate manual, BioStation IM Instructions .

    4.3.4 Use of peristaltic pumps

    To inject the test substance or to drain the culture medium, use a peristaltic pump instead of the syringe, if desired.

    Culture medium launching time can be shortened when using a peristaltic pump by increasing the diameter of the silicon

    tube with a commercially available conversion adapter.

    Syringe

    Syringe

    Fitting (female)

    Tube clamp

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    49/50

    41

    Care and Maintenance55

    5.1 Cleaning

    Clean each component of this product if necessary. Traces of test fluid can harden and clog the needle so it is a goodtime to clean syringes by flushing a cleansing solution through them.

    5.2 Parts Treatable in Autoclave Sterilization

    All parts used in this product can be sterilized in an autoclave. However, silicon tubes, couplings, and fittings wear with

    use and should therefore be replaced periodically.

    After treating each part with an autoclave, dry it completely with an apparatus such as dryer (60C).

    CAUTION

    Caution in autoclaving

    The heat applied when using an autoclave to sterilize bottle lids can cause the fitting to become loosened.

    Check the fitting in the bottle lid after autoclaving and retighten if loose.

    Remove the check valve clamps from the bottle holder before autoclaving.

    Unlock the check valve clamps and tube clamps before autoclaving.

  • 7/23/2019 M494E Biostation IM Perfusion Unit

    50/50

    Specifications66

    Model BS-P1, BS-P2

    Operating condition Temperature: 18 to 28 C (Note that temperature within the culture chamber should be

    set at 5 degrees higher than that of the environment)

    Transportation

    condition (when the

    product is packed)

    Temperature: -20 to 60 C

    Humidity: 90% RH maximum (no condensation)