M494E Biostation IM Perfusion Unit
-
Upload
xicoalexandre -
Category
Documents
-
view
221 -
download
0
Transcript of M494E Biostation IM Perfusion Unit
-
7/23/2019 M494E Biostation IM Perfusion Unit
1/50
Time Lapse Imaging System
BioStation IM
BS-IM Perfusion Unit
Instructions
M 9 E 08.6.NF.1
*M494EN01*
-
7/23/2019 M494E Biostation IM Perfusion Unit
2/50
-
7/23/2019 M494E Biostation IM Perfusion Unit
3/50
i
Introduction
Thank you for purchasing the Nikon products.
This instruction manual is written for the users of the BS-IM Dispense Unit, an accessory for the Time lapse imaging
system BioStation IM.
To ensure correct usage, read this manual carefully before operating the product.
No part of this manual may be reproduced or transmitted in any form without prior written permission from Nikon.
The contents of this manual are subject to change without notice.
Although every effort has been made to ensure the accuracy of this manual, errors or inconsistencies may remain. If
you note any points that are unclear or incorrect, please contact your nearest Nikon representative.
Some of the equipment described in this manual may not be included in the set you have purchased.
If you intend to use any other equipment with this product, read the manual for that equipment too.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may
be impaired.
Microsoft, Windows, and Internet Explorer are registered trademarks of Microsoft Corporation in the U.S. and other
countries. Other product and company names mentioned in this manual are trademarks or registered trademarks of
their respective owners.
-
7/23/2019 M494E Biostation IM Perfusion Unit
4/50
ii
Safety Precautions
To ensure correct and safe operation, read this manual before using the product.
Warning and Caution Symbols Used in This Manual
Although this product is designed to be completely safe during use, incorrect usage or failure to follow the safety
instructions provided may cause personal injury or property damage. To ensure correct usage, read this manual carefully
before using the product. Do not discard this manual and keep it handy for easy reference.
Safety instructions in this manual are marked with the following symbols to highlight their importance. For your safety,
always follow the instructions marked with these symbols.
Symbol Description
WARNINGDisregarding instructions accompanying this symbol may lead to serious injury or death.
CAUTIONDisregarding instructions accompanying this symbol may lead to injury or property
damage.
-
7/23/2019 M494E Biostation IM Perfusion Unit
5/50
Safety Precautions
iii
WARNING
1. Intended application of this product
The BS-IM Dispense Unit is a unit attached inside the BioStation IM (temperature controller area),
allowing the user to inject or to replace test fluids and fluid substances without opening/closing the door
or culture chamber. Do not use this product for other purpose.
The product is classified as an in-vitro diagnostic device.
2. Do not disassemble.
Disassembling may cause malfunction and will lead to the forfeiture of all claims against warranty. Do not
disassemble any part that is not indicated in this manual. If you notice any abnormality, contact your
nearest Nikon representative.
3. Read the instructions carefully.To ensure safety, carefully read this manual and the manuals for other equipment used with this product.
Especially be sure to follow the warnings and cautions indicated in the first of the manual.
4. Note on handing a hazardous specimen
Before handling a biological specimen, its risk must be checked.
When a hazardous specimen is used with this product, consult your safety supervisor or safety standard
of your facility. When a potentially infectious specimen is used, you must wear rubber gloves to avoid
direct touch.
If such specimen is spilled onto this product, the portion must be decontaminated in a safety manner.
Consult your safety supervisor or safety standard of your facility.
CAUTION
1. Disposal of the product
To avoid biohazard risks, dispose of the product as contaminated equipment according to the standard
procedure specified for your facility.
-
7/23/2019 M494E Biostation IM Perfusion Unit
6/50
Safety Precautions
iv
Notes on Handling the Product
1. Handling the Product.
This product is a precision instrument. Handle the product with care to avoid a physical shock or impact.
Cautions on moving the product
Remove the test fluid bottle and disconnect the tubes before carrying the BioStation IM with the Dispense Unit
attached.
Cautions on assembling and installing the product
Take care to avoid pinching your finger or hand.
Scratches or fouling such as fingerprints on optical components (such as lenses or filter) will degrade
microscope images.
Be careful to avoid scratches or direct contact with the objective when assembling the Dispense Unit.
2. Installation location and storage location
This product is an accessory of the BioStation IM. Select an installation location and a storage location as specified
for the BioStation IM. For details, refer to Safety Precautions in the separate manual BioStation IM Instructions
.
-
7/23/2019 M494E Biostation IM Perfusion Unit
7/50
v
ContentsIntroduction ............................................................................................................................................................ i
Safety Precautions ................................................................................................................................................ ii
Chapter 1 Equipment Overview......................................................................................................................... 1
1.1 Components .............................................. .................................................... ................................ 1
1.1.1
Perfusion induction unit (MFC99010) ................................................ ................................. 1
1.1.2
Perfusion medium exchange unit (MFC99015) ..................................................... ............. 2
1.2 Overview of the Perfusion Induction Unit................................... .................................................... 3
1.3 Overview of the Perfusion Medium Exchange Unit............ ........................................................ .... 4
1.3.1 Medium exchange - [degraded culture mediumfresh culture medium].......................... 5
1.3.2 Medium exchange - [culture mediumculture medium containing test fluid] ................... 6
1.3.3 Medium exchange
- [culture mediumculture medium containing test fluid culture medium] ...................7
Chapter 2 Names and Functions of Each Part ................................................................................................. 8
2.1
Parts Names and Functions of the Perfusion Induction Unit............................................. ............. 8
2.2 Parts Names and Functions of the Perfusion Medium Exchange Unit - [degraded culture medium
fresh culture medium] and [culture medium culture medium containing test fluid]................ 9
2.3 Parts Names and Functions of the Perfusion Medium Exchange Unit
- [culture mediumculture medium containing test fluidculture medium] ............................ 10
Chapter 3 Attaching to the BioStation IM ....................................................................................................... 11
3.1 Attaching the Perfusion Induction Unit.......................................................... ............................... 11
3.1.1 Removing the culture chamber from the BioStation IM .................................................... 11
3.1.2 Attaching the dispense needle assembly and bottle holder to the culture chamber ......... 13
3.1.3 Attaching the culture chamber to the BioStation IM..................................................... ..... 14
3.1.4
Installing the bottle and connecting the tubes and fittings ................................................ 15
3.2 Attaching the Perfusion Medium Exchange Unit
(for [degraded culture mediumfresh culture medium] / [culture mediumculture medium
containing test fluid] setup) ................................................... ....................................................... 17
3.2.1 Removing the culture chamber from the BioStation IM .................................................... 17
3.2.2 Attaching the dispense/drain needle assembly and connecting tubes ............................. 19
3.2.3
Attaching the culture chamber to the BioStation IM..................................................... ..... 22
3.2.4 Connecting tubes from outside of the BioStation IM.................................................... ..... 23
3.2.5 Piping arrangement for medium exchange
(for [degraded culture mediumfresh culture medium] /
[culture mediumculture medium containing test fluid] setup) ...................................... 25
3.3 Adjusting the Needle Position......................... ....................................................... ...................... 26
Attaching the Perfusion Medium Exchange Unit ......................................................... ............... 27
(for [culture mediumculture medium containing test fluid culture medium] setup)................. 27
1.
Removing the culture chamber from the BioStation IM .................................................... 27
2. Attaching the dispense/drain needle assembly and connecting tubes ............................. 27
3. Attaching the culture chamber to the BioStation IM..................................................... ..... 32
4. Connecting tubes from outside of the BioStation IM.................................................... ..... 32
5. Piping arrangement for medium exchange
(for [culture medium
culture medium containing test fluid
culture medium] setup)......... 34
Chapter 4 Operating Procedures..................................................................................................................... 35
4.1 Filling the Bottle with Culture Medium................................................ .......................................... 35
-
7/23/2019 M494E Biostation IM Perfusion Unit
8/50
Contents
vi
4.2 Setting and Removing the Specimen........................................... ................................................ 37
4.2.1
Setting the specimen.......... ....................................................... ....................................... 37
4.2.2 Removing the Specimen ................................................... ............................................... 38
4.3 Test Fluid Dispense/Medium Exchange ................................................ ...................................... 39
4.3.1
Dispensing procedure for test fluid, etc. ........................................................ ................... 39
4.3.2 Draining procedure for culture medium, etc...................................................................... 40
4.3.3 Focus adjustment ................................................... ....................................................... ... 40
4.3.4 Use of peristaltic pumps ................................................... ................................................ 40
Chapter 5 Care and Maintenance .................................................................................................................... 41
5.1 Cleaning ............................................ ..................................................... ..................................... 41
5.2 Parts Treatable in Autoclave Sterilization ................................................... ................................. 41
Chapter 6 Specifications.................................................................................................................................. 42
-
7/23/2019 M494E Biostation IM Perfusion Unit
9/50
1
Equipment Overview11
The BS-IM Perfusion Unit is a unit attached inside the BioStation IM (temperature controller area), allowing the user to
inject or to replace test fluids and fluid substances without opening/closing the door or culture chamber.
All the components of the unit are attached to the interior of the temperature controlled area of the BioStation IM. This
configuration eliminates the need for an additional device to warm replacement culture medium, and maintains the
environment of tubes connected to the dish at a controlled temperature.
1.1 Components
This option unit includes perfusion induction unit and a perfusion medium exchange unit.
1.1.1 Perfusion induction unit (MFC99010)
This is the basic unit for the BS-IM Perfusion Unit. Mounting the perfusion induction unit to the BioStation IM culture
chamber allows injection of test fluid to the culture medium without taking the dish out of the culture chamber.
Components in the perfusion induction unit
Dispense needle assembly(BS-P1): x1 Bottle: x1
Tube Clamp (blue): x1 Fitting (female): x1
Silicon tube (1100 mm): x1 Fitting (male) + silicon tube: x1
L-shaped fitting + silicon tube: x2
-
7/23/2019 M494E Biostation IM Perfusion Unit
10/50
Chapter 1 Equipment Overview
2
1.1.2 Perfusion medium exchange unit (MFC99015)
This unit, either partly or entirely, is used in combination with the perfusion induction unit. The perfusion medium
exchange unit alone does not provide any function. The use of this unit allows culture medium replacement without
taking the dish out of the culture chamber.
Components in the perfusion medium exchange unit
Drain needle assembly
(BS-P2): x1
Bottle: x1
Tube clamp
(red, yellow): x2
Fitting (female): x2
Silicon tube (1100 mm): x2
Fitting (male): x1
T-shaped fitting
+ L-shaped fitting
+ silicon tube: x1
Silicon tube (1100 mm): x2
Bottle holder: x1 Check valve: x2 Clamp: x2
Hexagonal socket
head bolt (M3 x 4): x2
L-shaped fitting+ silicon tube: x2
-
7/23/2019 M494E Biostation IM Perfusion Unit
11/50
Chapter 1 Equipment Overview
3
1.2 Overview of the Perfusion Induction Unit
This arrangement allows only test fluid injection to the culture medium.
Operators can observe the sample reactions induced by test fluid injection.
Figure 1.1-1 Overview of the perfusion induction unit
Test fluid bottleCulture chamber
Dish
Objective
Dispense
needle
BioStation IM (slide door)
Pressure applied
(from a syringe or peristaltic
pump) to send test fluid
-
7/23/2019 M494E Biostation IM Perfusion Unit
12/50
Chapter 1 Equipment Overview
4
1.3 Overview of the Perfusion Medium Exchange Unit
This unit allows users to exchange culture medium without taking the dish out of the culture chamber.
This unit and the perfusion induction unit can be configured in two ways, and the two configurations can be used for
three different applications.
Table 1.1-1 Medium exchange combinations
PartQua
ntity
Medium exchange -
[degraded culture medium
fresh culture medium] and
[culture mediumculture
medium containing test fluid]
Medium exchange
[culture mediumculture
medium containing test
fluidculture medium]
Dispense needle assembly x1
Bottle x1
Tube clamp (blue) x1
Fitting (female) x1
L-shaped fitting + silicon tube x2 X
Fitting (male) + silicon tube x1
Perfus
ioninductionunit
Silicon tube (1100 mm) x1
Drain needle assembly x1
Bottle x1 X
Bottle holder x1 X
Tube clamp (red, yellow) x2 (x1)
Fitting (female) x2 (x1)
Fitting (male) x1
Silicon tube (1100 mm) x2 (x1)
Silicon tube (110mm) x2
Hexagonal socket head bolt x2
Check valve x2
Clamp (for check valve) x2
L-shaped fitting + silicon tube x2
Per
fusionmediumexchangeunit
T-shaped fitting + L-shaped fitting+ silicon tube
x1
= used / X = not used / = partially used
-
7/23/2019 M494E Biostation IM Perfusion Unit
13/50
Chapter 1 Equipment Overview
5
Table 1.1-2 Medium exchange applications
Name Application
Medium exchange -
[degraded culture medium
fresh culture medium]
Drain the degraded culture medium launch fresh culture medium
Drain the culture medium degraded over prolonged culturing from the dish, and
launch fresh culture medium.
Medium exchange -
[culture mediumculture
medium containing test fluid]
Drain culture medium launch culture medium containing test fluid
Drain the culture medium in the dish, and launch culture medium containing test
fluid.
Medium exchange -
[culture mediumculture
medium containing test fluid
culture medium]
Drain culture medium launch culture medium containing test fluid drain
culture medium containing test fluid launch culture medium.
Drain the culture medium in the dish, and launch culture medium containing test
fluid. Drain the culture medium containing test fluid in the dish, and launch
culture medium.
1.3.1 Medium exchange - [degraded culture mediumfresh culture medium]
After draining degraded culture medium in the dish, launch fresh culture medium.
Culture medium is exchanged by draining the dish of culture medium degraded over prolonged culturing, and then
injecting fresh culture medium. This enables culturing for an extended period.
Figure 1.1-2 Medium exchange - [degraded culture mediumfresh culture medium]
Culture bottle
Culture chamber
Dish
Objective
Dispense
needle
BioStation IM (slide door)
Pressure applied (from a
syringe or peristaltic pump)
to send fresh culture medium
Pressure applied (from a
syringe or peristaltic pump)
to drain degraded culture
medium
Drain needle
-
7/23/2019 M494E Biostation IM Perfusion Unit
14/50
Chapter 1 Equipment Overview
6
1.3.2 Medium exchange - [culture mediumculture medium containing test fluid]
After draining the culture medium in the dish, launch culture medium containing test fluid.
This is mechanically identical to the medium exchange - [degraded culture medium fresh medium]. Drain the culture
medium in the dish, then launch culture medium containing test fluid. This system allows the user to steadily manage theconcentration of the test substance and the time required to reach the sample.
Figure 1.1-3 Medium exchange - [culture mediumculture medium with test fluid]
Bottle of culture
medium with test fluid
Culture chamber
Dish
Objective
Dispense
needle
BioStation IM (slide door)
Pressure applied (from a
syringe or peristaltic pump)to send culture medium
containing test fluid
Pressure applied (from a
syringe or peristaltic pump)to drain culture medium
Drain needle
-
7/23/2019 M494E Biostation IM Perfusion Unit
15/50
Chapter 1 Equipment Overview
7
1.3.3 Medium exchange -
[culture mediumculture medium containing test fluid culture medium]
After draining the culture medium, launch the culture medium containing a desired test fluid. Drain the culture medium
containing the test fluid to launch the culture medium.Two bottles are provided with this exchange unit: One is applied for the culture medium, the other one for the culture
medium containing the test fluid. After draining the culture medium from the dish, inject the culture medium containing
the test fluid. After observing sample reaction to the test substance, repeat the procedures of draining and injecting the
culture medium until almost any test substance is remained in the dish. In this way, you can verify if the test results are
accurate.
Figure 1.1-4 Medium exchange - [culture mediumculture medium containing test fluid culture medium]
Bottle of culturemedium containingtest fluid
Culture chamber
Dish
Objective
Dispense
needle
BioStation IM (slide door)
Pressure applied (from a
syringe or peristaltic pump)
to send culture medium
containing test fluid
Pressure applied (from a
syringe or peristaltic
pump) to drain culture
medium/ culture medium
containing test fluid
Drain needle
Culture bottle
Check valve
Pressure applied
(from a syringe or
peristaltic pump)
to send culture
medium
Check valve
-
7/23/2019 M494E Biostation IM Perfusion Unit
16/50
8
Names and Functions of Each Part22
This chapter explains the name and function of each part in both units.
2.1 Parts Names and Functions of the Perfusion Induction Unit
This setup is for test fluid dispense only. The unit consists only of the parts used for test fluid dispense.
Figure 2.1-1 Unit configuration
Table 2.1-1 Parts names and functions of the perfusion induction unit
Name Function
Dispense needleassembly
Needle assembly for dispensing test fluid into the dish.The position of the dispense needle can be adjusted axially and vertically. Additionally, the
dispense needle can be set in the dispense position or the standby position. The needle
should be set in the standby position for changing dishes.
Piping parts Silicon tube and couplings.
Bottle 1 bottle for holding the test fluid is installed.
*Culture chamber and dish are not included in this product.
Bottle
Silicon tube (1100 mm)
Fitting (male) + silicon tube
Dispense needle assembly
L-shaped fitting + silicon tube
Tube clamp (blue)
Fitting (female)
-
7/23/2019 M494E Biostation IM Perfusion Unit
17/50
Chapter 2 Names and Functions of Each Part
9
2.2 Parts Names and Functions of the Perfusion Medium Exchange Unit -[degraded culture mediumfresh culture medium] and [culture mediumculture medium containing test fluid]
This setup is for [draining degraded culture medium and injecting fresh culture medium], or [draining culture medium and
injecting culture medium containing a test fluid].
Components of both the perfusion induction unit and the exchange unit are used for this configuration.
Figure 2.1-2 Unit configuration [degraded culture mediumfresh culture medium] and [culture medium culture
medium containing test fluid]
Table 2.1-2 Parts names and functions of the perfusion medium exchange unit [degraded culture mediumfresh culture
medium] and [culture mediumculture medium containing test fluid]
Name Function
Dispense needle
assembly
Needle assembly for dispensing culture medium or culture medium containing test fluid into the dish.
The position of the dispense needle can be adjusted axially and vertically. Additionally, the
dispense needle can be set in the dispense position or the standby position. The needle should
be set in the standby position for changing dishes.
Drain needle
assembly
Needle assembly for draining culture medium or degraded culture medium from the dish.
The position of the drain needle can be adjusted axially and vertically. Additionally, the drain
needle can be set in the drain position or the standby position. Both the dispense and drain
needles should be set in the standby position for changing dishes.
Piping parts Silicon tube and couplings.
Bottle 1 bottle for holding the culture medium or culture medium containing test fluid is installed.
*Culture chamber and dish are not included in this product.
Bottle
Silicon tube (1100 mm)
Fitting (male) + silicon tube
Dispense needle assembly
L-shaped fitting + silicon tube
Tube clamp (blue)
Fitting (female)
Rotatable cover stopper
Fitting (female)Drain needle assembly
Silicon tube (1100 mm)
Fitting (male) Tube clamp (red or yellow)
-
7/23/2019 M494E Biostation IM Perfusion Unit
18/50
Chapter 2 Names and Functions of Each Part
10
2.3 Parts Names and Functions of the Perfusion Medium Exchange Unit - [culturemediumculture medium containing test fluidculture medium]
This setup is for [draining culture mediumdispensing culture medium containing test fluid draining culture medium
containing test fluid
dispensing culture medium].All the components for both the perfusion induction unit and the exchange unit are used for this configuration.
Figure 2.1-3 Unit configuration [culture mediumculture medium containing test fluidculture medium]
Table 2.1-3 Parts names and functions of the perfusion medium exchange unit [culture mediumculture medium containing
test fluidculture medium]
Name Function
Dispense needle
assembly
Needle assembly for dispensing culture medium or culture medium containing test fluid into the
dish.
The position of the dispense needle can be adjusted axially and vertically. Additionally, the
dispense needle can be set in the dispense position or the standby position. Both the dispense
and drain needles should be set in the standby position for changing dishes.
Drain needle
assembly
Needle assembly for draining culture medium or degraded culture medium from the dish.
The position of the drain needle can be adjusted axially and vertically. Additionally, the drain
needle can be set in the drain position or the standby position.
Both the dispense and drain needles should be set in the standby position for changing dishes.
Check valve Prevents the culture medium and culture medium containing test fluid from mixing.
Piping parts Silicon tube and couplings.
Bottle Two bottles for holding the culture medium and culture medium containing test fluid are installed.
*Culture chamber and dish are not included in this product.
Bottle
Silicon tube (1100 mm)
Fitting (male) + silicon tube
Dispense needle assembly
L-shaped fitting +
silicon tube
Tube clamp (blue)
Fitting (female)
Tube clamp (red or yellow)
Fitting (female)
Drain needle assembly
Silicon tube (1100 mm)
Fitting (male)
Silicon tube (110 mm)
Silicon tube (110 mm)
Check valve
Clamp
T-shaped fitting +
L-shaped fitting +
silicon tube
Bottle holder
Tube clamp (red or yellow)
Rotatable cover stopper
-
7/23/2019 M494E Biostation IM Perfusion Unit
19/50
11
Attaching to the BioStation IM33
3.1 Attaching the Perfusion Induction Unit
Necessary tool
The following tool is required for the procedure below.
Hexagonal screwdriver (supplied with BioStation IM)
3.1.1 Removing the culture chamber from the BioStation IM
When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in
advance.
1. Open the slide door on the BioStation IM.
Figure 3.1-1 Opening the slide door
2. Move the diascopic illuminator and duct out of
the way.
Figure 3.1-2 Moving the diascopic illuminator and
duct out of the way
3. Remove the temperature sensor.
Figure 3.1-3 Removing the temperature sensor
Slide door
Diascopic illuminator
Duct
Temperature sensor
-
7/23/2019 M494E Biostation IM Perfusion Unit
20/50
Chapter 3 Attaching to the BioStation IM
12
4. Remove the exhaust tube and wet mixed gas
tube.
Figure 3.1-4 Removing the tubes
5. Remove the culture chamber.
Remove the three (x3) hexagonal socket head bolts using
the hexagonal screwdriver.
Figure 3.1-5 Removing the culture chamber
Exhaust tube
Hexagonal
socket head
bolt
Hexagonal socket head bolt
Wet mixed gas tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
21/50
Chapter 3 Attaching to the BioStation IM
13
3.1.2 Attaching the dispense needle assembly and bottle holder to the culture chamber
Attach the dispense needle assembly and bottle holder to the culture chamber that was removed earlier.
1. Remove the right side plate of the culturechamber.
Remove the two (x2) hexagonal socket head bolts using
the hexagonal screwdriver.
Be careful not to lose the removed plate or hexagonal
socket head bolts.
Figure 3.1-6 Removing the right side plate of
culture chamber
2. Attach the dispense needle assembly to the
culture chamber.
Attach the dispense needle assembly to the culture
chamber using the two (x2) hexagonal socket head bolts.
Figure 3.1-7 Attaching the dispense needle
assembly
CAUTION
Caution in passing the dispense needle through the opening in the culture chamber
When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the
culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.
Hexagonal socket
head bolt
Hexagonal socket
head bolt
Plate
Hexagonal
socket head
bolt
-
7/23/2019 M494E Biostation IM Perfusion Unit
22/50
Chapter 3 Attaching to the BioStation IM
14
3.1.3 Attaching the culture chamber to the BioStation IM
Attach the culture chamber preassembled with the dispense needle assembly to the BioStation IM.
1. Attach the culture chamber.Attach the culture chamber to the BioStation IM using the
three (x3) hexagonal socket head bolts.
Figure 3.1-8 Attaching the culture chamber
2. Connect the exhaust tube and wet mixed gas
tube.
Figure 3.1-9 Connecting the tubes
3. Attach the temperature sensor.
Figure 3.1-10 Attaching the temperature sensor
Hexagonal socket
head bolt
Exhaust tube
Temperature sensor
Hexagonal socket head bolt
Wet mixed
gas tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
23/50
Chapter 3 Attaching to the BioStation IM
15
3.1.4 Installing the bottle and connecting the tubes and fittings
Place the bottle into the bottle holder and connect the tubes and fittings.
1. Place the bottle into the bottle holder.
Figure 3.1-11 Installing the bottle
2. Connect the tubes and fittings.
Connect the tubes and fittings as shown in the figure
below.
Figure 3.1-12 Connecting the tubes and fittings
Bottle
Silicon tube (1100 mm)
Fitting (male) + silicon tube
L-shaped fitting + silicon tube
Fitting (female)
Tube clamp (blue)
-
7/23/2019 M494E Biostation IM Perfusion Unit
24/50
Chapter 3 Attaching to the BioStation IM
16
CAUTION
Caution in connecting the tubes and fittings
Make sure all tubes and fitting couplings are tightened. Test fluid or culture medium may leak from loose tube
connections and cause contamination or damage this instrument.
Pass the silicon tube (1100 mm x1) for dispense through
the points shown in the figure on the right and pull the tube
out of the BioStation IM.
Figure 3.1-13 Routing silicon tubes
CAUTION
Caution in pulling out the silicon tube
When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the
BioStation IM.
Silicon tube
(1100 mm)
for dispense
-
7/23/2019 M494E Biostation IM Perfusion Unit
25/50
Chapter 3 Attaching to the BioStation IM
17
3.2 Attaching the Perfusion Medium Exchange Unit(for [degraded culture mediumfresh culture medium] / [culture mediumculture medium containing test fluid] setup)
This section describes the installation procedure for the perfusion medium exchange setup for [degraded culture medium
fresh culture medium] and [culture medium culture medium containing test fluid].
For installation procedures of the perfusion medium exchange setup for [culture mediumculture medium containing
test fluidculture medium], refer to Section 3.4, Attaching the Perfusion Medium Exchange Unit (for [culture medium
culture medium containing test fluidculture medium]).
Necessary tool
The following tool is required for the procedure below.
Hexagonal screwdriver (supplied with BioStation IM)
3.2.1 Removing the culture chamber from the BioStation IM
When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in
advance.
1. Open the slide door on the BioStation IM.
Figure 3.2-1 Opening the slide door
2. Move the diascopic illuminator and duct out of
the way.
Figure 3.2-2 Moving the diascopic illuminator and
duct out of the way
Slide door
Diascopic illuminator
Duct
-
7/23/2019 M494E Biostation IM Perfusion Unit
26/50
Chapter 3 Attaching to the BioStation IM
18
3. Remove the temperature sensor.
Figure 3.2-3 Removing the temperature sensor
4. Remove the exhaust tube and wet mixed gas
tube.
Figure 3.2-4 Removing the tubes
5. Remove the culture chamber.Remove the three (x3) hexagonal socket head bolts using
the hexagonal screwdriver.
Figure 3.2-5 Removing the culture chamber
Temperature sensor
Exhaust tube
Hexagonal
socket head
bolt
Hexagonal socket head bolt
Wet mixed gas tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
27/50
Chapter 3 Attaching to the BioStation IM
19
3.2.2 Attaching the dispense/drain needle assembly and connecting tubes
Attach the dispense needle assembly and drain needle assembly to the culture chamber that was removed earlier.
1. Remove the right and left side plates of theculture chamber.
Remove the two (x2) hexagonal socket head bolts on each
side using the hexagonal screwdriver.
Be careful not to lose the removed plate or hexagonal
socket head bolts.
Of the plates removed, the left plate is made taller than the
right plate to serve as a stopper for the rotatable cover.
Make sure to return the plates to their correct positions
when reassembling.Figure 3.2-6 Removing the right and left side plates
of culture chamber
2. Attach the dispense needle assembly to the
culture chamber.
Attach the dispense needle assembly to the culture
chamber using the two (x2) hexagonal socket head bolts.
Figure 3.2-7 Attaching the dispense needle
assembly
CAUTION
Caution in passing the dispense needle through the opening in the culture chamber
When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the
culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.
Hexagonal socket head bolt
Plate
Hexagonal socket
head bolt
Hexagonal
socket headbolt
Hexagonal socket head bolt
-
7/23/2019 M494E Biostation IM Perfusion Unit
28/50
Chapter 3 Attaching to the BioStation IM
20
3. Attach the drain needle assembly to the
culture chamber.
Attach the drain needle assembly to the culture chamber
using the two (x2) special hexagonal socket head bolts.
Figure 3.2-8 Attaching the drain needle assembly
CAUTION
Caution in passing the drain needle through the opening in the culture chamber
When passing the drain needle through the culture chamber, be careful not to bump the needle tip against the
culture chamber. This assembly may fail to operate properly, if the drain needle is deformed.
4. Adjust the height of the drain needle.
The height and axial position of both needles are adjustable. Normally, to attach the drain needle, adjust the height
of the drain needle. Apply additional adjustments if necessary.For details on height adjustment procedures for the drain needle, refer to Section 3.3 "Adjusting the Needle
Position".
5. Connect the tubes and fittings to the bottle.
1) Connect the L-shaped fittings and silicon tubes (2 sets)
to the bottle.
2) Connect the male fitting and silicon tube to the
L-shaped fitting and silicon tube on the dispensing side
of the bottle (with a tube inside).
Figure 3.2-9 Connecting tubes to the bottle
CAUTION
Caution in connecting the tubes and fittings
Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tubeconnections and cause contamination or damage this instrument.
Hexagonal socket
head bolt
Hexagonalsocket head bolt
Fitting (male)
+ silicon tube
L-shaped fitting
+ silicon tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
29/50
Chapter 3 Attaching to the BioStation IM
21
6. Place the bottle into the bottle holder and
connect the tubes.
Place the bottle with tubes connected into the bottle holder
and insert the male fitting into the dispense needle.
Figure 3.2-10 Installing and connecting tubes to the
bottle
Fitting (male)
+ silicon tube
Bottle
-
7/23/2019 M494E Biostation IM Perfusion Unit
30/50
Chapter 3 Attaching to the BioStation IM
22
3.2.3 Attaching the culture chamber to the BioStation IM
Attach the culture chamber preassembled with the dispense needle assembly and drain needle assembly to the
BioStation IM.
Figures below illustrate setup for medium exchange - [culture mediumculture medium containing test fluid]. Setupprocedure for this application is the same as the procedure for medium exchange - [culture mediumculture medium
containing test fluidculture medium].
1. Attach the culture chamber.
Attach the culture chamber to the BioStation IM using the
three (x3) hexagonal socket head bolts.
Figure 3.2-11 Attaching the culture chamber
2. Connect the exhaust tube and wet mixed gas
tube.
Figure 3.2-12 Connecting the tubes
Hexagonal socket
head bolt
Exhaust tube
Hexagonal socket
head bolt
Wet mixed
gas tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
31/50
Chapter 3 Attaching to the BioStation IM
23
3. Attach the temperature sensor.
Figure 3.2-13 Attaching the temperature sensor
3.2.4 Connecting tubes from outside of the BioStation IM
Connect the silicon tubes for dispense and drain from outside of the BioStation IM.
1. Insert the silicon tubes for dispense and drain.
Pass the silicon tubes (1100 mm) for dispense (x1) and
drain (x1) through the points shown in the figure on the
right and pull the tubes into the BioStation IM.
Figure 3.2-14 Routing silicon tubes
CAUTION
Caution in pulling out the silicon tube
When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the
BioStation IM.
Temperature
sensor
Silicon tube
(1100 mm)
for dispense
Silicon tube
(1100 mm)
for drain
-
7/23/2019 M494E Biostation IM Perfusion Unit
32/50
Chapter 3 Attaching to the BioStation IM
24
2. Connect each silicon tube (1100 mm) to the BS-IM Dispense Unit.
Test fluid dispense tube
Connect the silicon tube (1100 mm) to the L-shaped fitting and silicon tube on the air pressure side of the bottle
(with no tube inside).
Drain tube
Connect the male fitting to the tip of the silicon tube (1100 mm) and connect the male fitting to the drain needle.
Figure 3.2-15 Connecting the silicon tubes for dispense and drain
CAUTION
Caution in connecting the tubes and fittings
Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tube
connections and cause contamination or damage this instrument.
3. Attach the male fitting and tube clamp to the
silicon tubes for dispense and drain.
Three tube clamps are provided: blue, red and yellow. Besure that each silicon tube is recognized when attaching it.
Figure 3.2-16 Attaching the fitting and the tube
clamp
Silicon tube (1100 mm)
for dispense
Silicon tube (1100 mm) for drain
Fitting (male)
Tube clamp
Fitting (female)
-
7/23/2019 M494E Biostation IM Perfusion Unit
33/50
Chapter 3 Attaching to the BioStation IM
25
3.2.5 Piping arrangement for medium exchange
(for [degraded culture mediumfresh culture medium] / [culture medium culture
medium containing test fluid] setup)
The piping arrangement for medium exchange [degraded culture mediumfresh culture medium] and [culture mediumculture medium containing test fluid] is as shown below.
After connecting the tubes, make sure that the parts and tubes are connected correctly.
Figure 3.2-17 Connecting the tubes and fittings
CAUTION
Caution in connecting tubes to the bottle
Be careful not to connect the wrong tube to the bottle.The inlet with a tube: connect the silicon tube that leads to the dispense needle
The inlet without a tube inside: connect the silicon tube (1100 mm) that leads to outside the BioStation IM.
Silicon tube (1100 mm)
Fitting (male) + silicon tube
L-shaped fitting + silicon tube
Fitting (male)
Silicon tube (1100mm)
Fitting (female)
Fitting (female)
Tube clamp (blue)
Tube clamp (red or yellow)
-
7/23/2019 M494E Biostation IM Perfusion Unit
34/50
Chapter 3 Attaching to the BioStation IM
26
3.3 Adjusting the Needle Position
The needle has adjustments for height and axial positioning.
Normally, only the height of the drain needle needs to be adjusted. The axial position of the drain needle and the
dispense needle position should not require adjustment. Make these adjustments if needed.
Example below describes how to adjust the position for the drain needle, which procedure is also applied to adjust the
position for the dispense needle.
1. Adjust the height of the drain needle.
1) Slightly loosen the hexagonal socket head bolts that
hold the drain needle assembly in place on the left and
right sides.
2) To adjust the clearance between the bottom surface of
the dish and the tip of the drain needle, gently shake
the left side of the mounting piece for the culture
chamber up and down using the hexagonal socket
head bolt on the right side as the pivot.
3) Once adjusted, fix the culture chamber in place by
tightening the left and right hexagonal socket head
bolts.
Figure 3.3-1 Adjusting the height of drain needle
Adjusting the height of the drain needle
The drain needle is made slightly longer than the dispense needle.
When adjusting the height of the drain needle, aim to put the needle tip as close to the bottom of the dish aspossible. If the drain needle tip is off the bottom of the dish, the more the amount of culture medium or test fluid left
undrained in the dish.
2. Adjust the axial position of the drain needle.
1) Loosen the hexagonal socket head bolt (x1) in the top
of the needle assembly using the hexagonal
screwdriver.
2) Slide the cover on the needle in the axial direction to
adjust the position of the tip of the drain needle.
3) Once adjusted, tighten the hexagonal socket head bolt
(x1) in the top of the needle assembly using the
hexagonal screwdriver.Figure 3.3-2 Adjusting the axial position of drain
needle
Hexagonal socket head bolt (pivot side)
Hexagonal socket
head bolt
Hexagonal socket head bolt (adjustment side)
Cover on
needle
assembly
-
7/23/2019 M494E Biostation IM Perfusion Unit
35/50
Chapter 3 Attaching to the BioStation IM
27
Attaching the Perfusion Medium Exchange Unit(for [culture mediumculture medium containing test fluidculture medium] setup)
1. Removing the culture chamber from the BioStation IM
When installing the Perfusion Unit to the culture chamber, remove the culture chamber from the BioStation IM in advance.
Refer to Section 3.2.1, Removing the culture chamber from the BioStation IM for details of the procedure.
2. Attaching the dispense/drain needle assembly and connecting tubes
Attach the dispense needle assembly, drain needle assembly, and bottle holder to the culture chamber that was removed
earlier.
1. Remove the bottle holder from the dispense
needle assembly.Remove the hexagonal socket head bolt using the
hexagonal screwdriver.
Figure 2-1 Removing the bottle holder
2. Combine the bottle holders.
Take the bottle holder removed from the dispense needle
assembly and combine it with the bottle holder included in
the perfusion medium exchange unit using the two (x2)
hexagonal socket head bolts.
Figure 2-2 Combining the bottle holders
3. Attach the clamp for the check valve to the
combined bottle holders.
Secure the clamp by inserting the tab on the clamp into
the slotted mounting hole. With the clamp pressed against
the bottle holder, pull upwards.
Figure 2-3 Attaching the clamps
Hexagonal socket head bolt
Hexagonal sockethead bolt
Clamp
Bottle holder
Hexagonal socket
head bolt
Clamp
-
7/23/2019 M494E Biostation IM Perfusion Unit
36/50
Chapter 3 Attaching to the BioStation IM
28
4. Connect the tubes and fittings to the check
valve.
When connecting, the arrows on the check valves should
be pointed at the T-shaped fitting, L-shaped fitting, andsilicon tube.
Figure 2-4 Connecting the tubes and fittings to
the check valve
CAUTION
Caution in connecting the tubes and fittings to the check valve
Be sure that the arrow on the check valve faces the correct direction. The check valve will not function if it is
connected in the reverse direction.
5. Connect the check valve.
Attach the check valves to each clamp as shown in the
figure on the right, and then pass the tube through the
bottle holders.
Figure 2-5 Connecting the check valves
CAUTION
Caution in connecting the check valves to the clamps
Attach each check valve so that the ribs touch the inside of the clamp. The check valve is not fixed in place
because the recessed surfaces do not touch the inside of the clamp.
6. Attach the assembled bottle holders to the
dispense needle assembly.
Attach the assembled bottle holders to the dispense
needle assembly using the hexagonal socket head bolt
(x1).
Figure 2-6 Attaching the bottle holders
T-shaped fitting + L-shaped fitting + silicon tube
Hexagonal socket head bolt
Check valve
Check valve
Dispense needle
assembly
-
7/23/2019 M494E Biostation IM Perfusion Unit
37/50
Chapter 3 Attaching to the BioStation IM
29
7. Remove the right and left side plates of the
culture chamber.
Remove the two (x2) hexagonal socket head bolts on each
side using the hexagonal screwdriver.
Be careful not to lose the removed plate or hexagonal
socket head bolts.
Of the plates removed, the left plate is made taller than the
right plate to serve as a stopper for the rotatable cover.
Make sure to return the plates to their correct locations
when reassembling.Figure 2-7 Removing the right and left side plates
of culture chamber
8. Attach the dispense needle assembly to the
culture chamber.
Attach the dispense needle assembly to the culture
chamber using the two (x2) hexagonal socket head bolts.
Figure 2-8 Attaching the dispense needle
assembly
CAUTION
Caution in passing the dispense needle through the opening in the culture chamber
When passing the dispense needle through the culture chamber, be careful not to bump the needle tip against the
culture chamber. This assembly may fail to operate properly, if the dispense needle is deformed.
Hexagonal socket
head bolt
Hexagonal socket
head bolt
Hexagonal socket head bolt
Plate
Hexagonal
socket head
bolt
-
7/23/2019 M494E Biostation IM Perfusion Unit
38/50
Chapter 3 Attaching to the BioStation IM
30
9. Attach the drain needle assembly to the
culture chamber.
Attach the drain needle assembly to the culture chamber
using the two (x2) special hexagonal socket head bolts.
Figure 2-9 Attaching the drain needle assembly
CAUTION
Caution in passing the drain needle through the opening in the culture chamber
When passing the drain needle through the culture chamber, be careful not to bump the needle tip against the
culture chamber. This assembly may fail to operate properly, if the drain needle is deformed.
10. Adjust the height of the drain needle.
The height and axial position of both needles are adjustable. Normally, to attach the drain needle, adjust the heightof the drain needle. Apply additional adjustments if necessary.
For details on height adjustment procedures for the drain needle, refer to Section 3.3 Adjusting the Needle
Position.
11. Connect the tube for the dispense needle.
Connect the T-shaped tube assembly to the dispense
needle with the male fitting and silicon tube.
Figure 2-10 Tube connections for the dispense
needle
Hexagonal socket
head bolt
Hexagonal
socket
head bolt
Fitting (male)
+ silicon tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
39/50
Chapter 3 Attaching to the BioStation IM
31
12. Connect the tubes and fittings to each bottle.
Connect the L-shaped fitting and silicon tube to the air
pressure inlet (without a tube) of each bottle.
Figure 2-11 Connecting tubes and fittings to the
bottle
13. Install the bottles.
Place each bottle in the bottle holder.
Figure 2-12 Placing the bottle
14. Connect the tube between the check valve and
bottle.
Connect the silicon tube (110 mm) between the inlet fitting
(with the tube inside) of the test fluid dispense bottle and
the check valve.
Figure 2-13 Placing the bottle and connecting
tubes
L-shaped fitting
+ silicon tube
BottleBottle
Silicon tube
(110 mm)
-
7/23/2019 M494E Biostation IM Perfusion Unit
40/50
Chapter 3 Attaching to the BioStation IM
32
3. Attaching the culture chamber to the BioStation IM
Attach the culture chamber preassembled with the dispense needle assembly and drain needle assembly to the
BioStation IM.
Refer to Section 3.2.3, Attaching the culture chamber to the BioStation IM for details of the procedure.
4. Connecting tubes from outside of the BioStation IM
Connect the silicon tubes for dispense and drain from outside of the BioStation IM.
1. Insert the silicon tubes for dispense and drain.
Pass the silicon tubes (1100 mm) for dispense (x2) and
drain (x1) through the points shown in the figure on the
right and pull the tubes into the BioStation IM.
Figure 4-1 Routing silicon tubes
CAUTION
Caution in pulling out the silicon tube
When pulling the silicon tube (1100 mm) out of the BioStation IM, be careful not to pinch it in the front door of the
BioStation IM.
Silicon tube
(1100 mm)
for dispense
Silicon tube
(1100 mm)
for drain
-
7/23/2019 M494E Biostation IM Perfusion Unit
41/50
Chapter 3 Attaching to the BioStation IM
33
2. Connect each silicon tube (1100 mm) to the BS-IM Dispense Unit.
Test fluid dispense tube
Connect the silicon tube (1100 mm) to the L-shaped fitting and silicon tube on the air pressure side of the bottle
(with no tube inside).
Drain tube
Connect the male fitting to the tip of the silicon tube (1100 mm) and connect the male fitting to the drain needle.
Figure 4-2 Connecting the silicon tubes for dispense and drain
CAUTION
Caution in connecting the tubes and fittings
Make sure all tubes and fittings couplings are tightened. Test fluid or culture medium may leak from loose tube
connections and cause contamination or damage this instrument.
3. Attach the male fitting and tube clamp to the
silicon tubes for dispense and drain.
Three tube clamps are provided: blue, red and yellow. Besure that each silicon tube is recognized when attaching it.
Figure 4-3 Attaching the fitting and the tube clamp
Silicon tube (1100 mm) for dispense
Silicon tube (1100 mm) for drain
Fitting (male)
Tube clamp
Fitting (female)
-
7/23/2019 M494E Biostation IM Perfusion Unit
42/50
Chapter 3 Attaching to the BioStation IM
34
5. Piping arrangement for medium exchange
(for [culture mediumculture medium containing test fluid culture medium] setup)
The piping arrangement for medium exchange [culture mediumculture medium containing test fluid culture
medium] is as shown below.After connecting the tubes, make sure that the parts and tubes are connected correctly.
Figure 5-1 Connecting the tubes and fittings
CAUTION
Caution in connecting tubes to the bottle
Be careful not to connect the wrong tube to the bottle.
The inlet with a tube: connect the silicon tube that leads to the check valve.
The inlet without a tube inside: connect the silicon tube (1100 mm) that leads to outside the BioStation IM.
Silicon tube (1100 mm)
Silicon tube (110 mm)
Connects the bottle and check valve
Fitting (male) + silicon tube
Connects the dispense
needle and T-shaped fitting
Silicon tube (1100 mm)
L-shaped fitting
+ silicon tube
Fitting (male)
Silicon tube
(1100mm)Fitting (female)
Fitting (female)
Tube clamp (blue)
Tube clamp (red or yellow)
Tube clamp (red or yellow)
Check valve
T-shaped fitting
+ L-shaped fitting
+ silicon tube
Silicon tube (110 mm)
Connects the bottle
and check valve
-
7/23/2019 M494E Biostation IM Perfusion Unit
43/50
35
Operating Procedures44
This product does not have any pumps to dispense test fluid or drain culture medium.
Normally, a customer-provided syringe or peristaltic pump is used to dispense test fluid or drain culture medium.
4.1 Filling the Bottle with Culture Medium
1. Open the slide door on the BioStation IM.
2. Remove the bottle to fill with culture medium.
Disconnect and remove the silicon tube from the bottle
and remove the bottle.
Figure 4.1-1 Removing the bottle
3. Remove the lid from the bottle and fill with culture medium.
When removing the lid from the bottle, check if the packing inside the lid is not damaged.Be sure there are no
scratches or cracks or other signs of deterioration.
CAUTION
Packing used in the bottle lid
The bottle lid disassembles into the parts shown in the
figure on the right. The packing inside the lid should be
replaced if any deterioration is found. Make sure it is fit
in the lid after each replacement.
To ensure the bottle is sealed air tight, make sure thepacking is installed properly.
Figure 4.1-2 Packing in the bottle lid
Packing
Bottle lid
Silicon tube
-
7/23/2019 M494E Biostation IM Perfusion Unit
44/50
Chapter 4 Operating Procedures
36
4. Return the bottle to the bottle holder and
reconnect the tubes.
Figure 4.1-3 Installing the bottle
5. Close the slide door on the BioStation IM.
CAUTION
Caution on the temperature of the culture medium/test fluid in the bottle
If the temperature of the culture medium or culture medium containing test fluid filled in the bottle is below
37.0 C, allow the content to reach 37.0 C before starting your experiment by placing the bottle in the BioStation
IM and allowing it to warm up.
-
7/23/2019 M494E Biostation IM Perfusion Unit
45/50
Chapter 4 Operating Procedures
37
4.2 Setting and Removing the Specimen
When setting and removing the specimen (dish), both the dispense needle and drain needle must be in their standby
positions.
4.2.1 Setting the specimen
1. Open the slide door on the BioStation IM and move the diascopic illuminator out of the way.
CAUTION
Caution in moving the needle
When moving the needles to the standby, dispense or drain positions, always hold and turn the needle by the
metal base. Do not hold the silicon tubes, as this may loosen or disconnect the tube and cause it to break.
2. Move the drain needle and dispense needle to
their standby positions.
1) Push the needle slightly toward the culture chamber
until it moves out of the groove.
2) Turn the needle left 90 degrees.
3) Pull the needle toward you.
The needle can be pulled out to the position shown in
the figure below. Place the needle within the guide
groove unless otherwise desired.
Figure 4.2-1
CAUTION
Caution in pulling the needle to the limit.
If you close the slide door on the BioStation with the
needle pulled all the way out of the guide groove, the
door will hit the needle and possibly break it.
Figure 4.2-2 Moving the needle to the standby
position
1) Pull the needle out of the groove.
2) Turn the needle left 90 degrees.
3) Pull the needle toward you.
-
7/23/2019 M494E Biostation IM Perfusion Unit
46/50
Chapter 4 Operating Procedures
38
3. Open the rotatable cover and set the dish.
Set the dish near the center of the culture chamber.
Figure 4.2-3 Setting the dish
4. Move the drain needle and dispense needle to
their drain and dispense positions,respectively.
Return each needle to their original positions by following
the procedures for moving them to the standby position in
reverse.
Figure 4.2-4 Moving each needle to the
drain/dispense positions
5. Close the rotatable cover on the culture
chamber.
Figure 4.2-5 Closing the rotatable cover on the
culture chamber
6. Return the diascopic illuminator to its original position and close the BioStation IM slide
door.
4.2.2 Removing the Specimen
Remove the specimen (dish) by following the setting procedures.
Dish
-
7/23/2019 M494E Biostation IM Perfusion Unit
47/50
Chapter 4 Operating Procedures
39
4.3 Test Fluid Dispense/Medium Exchange
4.3.1 Dispensing procedure for test fluid, etc.
1. Connect a silicon tube to a syringe, and the
bottle containing test fluid or culture medium
mixed with test fluid.
Figure 4.3-1 Connecting to a pump
2. Dispense the required amount of test fluid
from the syringe.
Open the tube clamp when injecting, then close it
afterwards.
Figure 4.3-2 Dispensing test fluid, etc.
3. Record the timing of test fluid dispense in the
software.
- Cell stimulus mark setting button :
For an explanation of each symbol, refer to the separate
manual, BioStation IM Instructions , Chapter 4 Time-Lapse Images in Process
Screen, Section 4.1.8 Time Line and Image Playback
Control.Figure 4.3-3 Recording test fluid dispense
Syringe
Syringe
Fitting (female)
Tube clamp
-
7/23/2019 M494E Biostation IM Perfusion Unit
48/50
Chapter 4 Operating Procedures
40
4.3.2 Draining procedure for culture medium, etc.
1. Connect the silicon tube from the discharge
needle to a syringe.
Figure 4.3-4 Connecting to a pump
2. Draw culture medium into the syringe.
Open the tube clamp when draining, then close it.
Figure 4.3-5 Draining culture medium, etc.
3. Record the timing of draining the culture
medium in the software, if necessary.
- Cell stimulus mark setting button :
For an explanation of each symbol, refer to the separate
manual, BioStation IM Instructions , Chapter 4 Time-Lapse Images in Process
Screen, Section 4.1.8 Time Line and Image Playback
Control.Figure 4.3-6 Recording draining of culture medium
4.3.3 Focus adjustment
After draining culture medium or dispensing test fluid to the dish, focus of the microscope may require adjustment as
focus can be affected by changes in weight of dish content or even by the slightest changes in temperature. Therefore,
verify the focus adjustment after medium exchange or test fluid dispense during time lapse examination. If the image
goes out of focus, readjust the Z coordinate as described in Section 2.4, Observation Point and Focus Adjustment in
the separate manual, BioStation IM Instructions .
4.3.4 Use of peristaltic pumps
To inject the test substance or to drain the culture medium, use a peristaltic pump instead of the syringe, if desired.
Culture medium launching time can be shortened when using a peristaltic pump by increasing the diameter of the silicon
tube with a commercially available conversion adapter.
Syringe
Syringe
Fitting (female)
Tube clamp
-
7/23/2019 M494E Biostation IM Perfusion Unit
49/50
41
Care and Maintenance55
5.1 Cleaning
Clean each component of this product if necessary. Traces of test fluid can harden and clog the needle so it is a goodtime to clean syringes by flushing a cleansing solution through them.
5.2 Parts Treatable in Autoclave Sterilization
All parts used in this product can be sterilized in an autoclave. However, silicon tubes, couplings, and fittings wear with
use and should therefore be replaced periodically.
After treating each part with an autoclave, dry it completely with an apparatus such as dryer (60C).
CAUTION
Caution in autoclaving
The heat applied when using an autoclave to sterilize bottle lids can cause the fitting to become loosened.
Check the fitting in the bottle lid after autoclaving and retighten if loose.
Remove the check valve clamps from the bottle holder before autoclaving.
Unlock the check valve clamps and tube clamps before autoclaving.
-
7/23/2019 M494E Biostation IM Perfusion Unit
50/50
Specifications66
Model BS-P1, BS-P2
Operating condition Temperature: 18 to 28 C (Note that temperature within the culture chamber should be
set at 5 degrees higher than that of the environment)
Transportation
condition (when the
product is packed)
Temperature: -20 to 60 C
Humidity: 90% RH maximum (no condensation)