INDEPENDENT MONITORING CONSULTANTS(M) SDN BHD

30
Schedule SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA MS ISO/IEC 17025 Issue date: 2 November 2017 Valid until: 15 October 2020 NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017) Page: 1 of 30 Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation LABORATORY LOCATION: (PERMANENT LABORATORY) METCAL TECHNOLOGIES (M) SDN. BHD. NO. 36, CANGKAT BUKIT BELAH 11920 BAYAN LEPAS PULAU PINANG MALAYSIA FIELDS OF CALIBRATION: DIMENSIONAL, MASS, FORCE, TORQUE, TEMPERATURE, PRESSURE, VOLUME, ELECRTRICAL, TIME AND FREQUENCY This laboratory has demonstrated its technical competence to operate in accordance with MS ISO/IEC 17025:2005 (ISO/IEC 17025:2005). This laboratory’s fulfillment of the requirements of ISO/IEC 17025 means the laboratory meets both the technical competence requirements and management system requirements that are necessary for it to consistently deliver technically valid test results and calibrations. The management system requirements in ISO/IEC 17025 are written in language relevant to laboratory operations and operate generally in accordance with the principles of ISO 9001 (see Joint ISO-ILAC-IAF Communiqué dated April 2017). * The expanded uncertainties are based on an estimated confidence probability of approximately 95% and have a coverage factor of k=2 unless stated otherwise. SCOPE OF CALIBRATION: DIMENSIONAL Instrument Calibrated/ Measurement Parameter Range Calibration and Measurement Capability Expressed as an Uncertainty(±)* Remarks Feeler Gauge 0.01 mm to 3 mm 0.78 μm Direct measurement with micrometer based on JIS B 7524:1992 External Micrometer 0 mm to 25 mm 25 mm to 100 mm 100 mm to 250 mm 250 mm to 500 mm 0.44 μm 0.82 μm 2.3 μm 4.7 μm Comparison with gauge block based on JIS B 7520:1981 Vernier, Dial & Digital Caliper 0 mm to 600 mm 6 μm Comparison with caliper checker / gauge block based on JIS B 7507:1993 Dial Gauge / Digital Indicator 0 mm to 50 mm 1.5 μm Comparison with gauge tester based on JIS B 7503:2011 Dial Test Indicator 0 mm to 1.6 mm 1.5 μm Comparison with gauge tester based on JIS B 7533-1990

Transcript of INDEPENDENT MONITORING CONSULTANTS(M) SDN BHD

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 1 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

LABORATORY LOCATION: (PERMANENT LABORATORY)

METCAL TECHNOLOGIES (M) SDN. BHD. NO. 36, CANGKAT BUKIT BELAH 11920 BAYAN LEPAS PULAU PINANG MALAYSIA

FIELDS OF CALIBRATION: DIMENSIONAL, MASS, FORCE, TORQUE, TEMPERATURE, PRESSURE, VOLUME, ELECRTRICAL, TIME AND FREQUENCY

This laboratory has demonstrated its technical competence to operate in accordance with MS ISO/IEC 17025:2005 (ISO/IEC 17025:2005). This laboratory’s fulfillment of the requirements of ISO/IEC 17025 means the laboratory meets both the technical competence requirements and management system requirements that are necessary for it to consistently deliver technically valid test results and calibrations. The management system requirements in ISO/IEC 17025 are written in language relevant to laboratory operations and operate generally in accordance with the principles of ISO 9001 (see Joint ISO-ILAC-IAF Communiqué dated April 2017). * The expanded uncertainties are based on an estimated confidence probability of approximately 95% and have a coverage factor of k=2 unless stated otherwise.

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Feeler Gauge 0.01 mm to 3 mm

0.78 µm Direct measurement with micrometer based on JIS B 7524:1992

External Micrometer 0 mm to 25 mm 25 mm to 100 mm 100 mm to 250 mm 250 mm to 500 mm

0.44 µm 0.82 µm 2.3 µm 4.7 µm

Comparison with gauge block based on JIS B 7520:1981

Vernier, Dial & Digital

Caliper

0 mm to 600 mm 6 µm Comparison with caliper checker / gauge block based on JIS B 7507:1993

Dial Gauge / Digital

Indicator

0 mm to 50 mm 1.5 µm Comparison with gauge tester based on JIS B 7503:2011

Dial Test Indicator 0 mm to 1.6 mm

1.5 µm Comparison with gauge tester based on JIS B 7533-1990

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 2 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Vernier, Dial & Digital

Height Gauge

0 mm to 100 mm 100 mm to 600 mm

(0.2 + 4.5 L) µm (-0.3 + 9.5 L) µm

‘L’ in metre

Comparison with caliper checker / gauge block based on JIS B 7517:1993

Gauge Block

0.5 mm to 100 mm

(0.11 + 3L) µm

L in m

Comparison with master gauge block based on ISO 3650 / JIS B 7506:1997

Dial & Digital Thickness Gauge

0 to 20 mm 2.5 µm Comparison with gauge block based on JIS B 7503:1997

Plain Pin Gauge (diameter only)

0.10 mm to 25 mm

0.26 µm

Direct measurement with micrometer based on JIS B 7420:1997

Depth Micrometer/Gauge

0 mm to 300 mm 6 µm Comparison with gauge block based on JIS B 7544:1994

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien 4. Chew Siou Lian

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 3 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Height Master

Up to 300 mm 2.3 µm Comparison with

gauge block based on AS 3779:1990

Riser Block

Up to 300 mm 2.3 µm Comparison with

gauge block based on AS 3779:1990

Electrical Comparator (Mu-checker)

± 5 µm ± 15 µm ± 50 µm ± 150 µm ± 500 µm ± 1500 µm

0.10 µm 0.16 µm 0.31 µm 1.2 µm 2.9 µm 12 µm

Comparison with gauge block based on JIS B 7536:1982

Parallel Thread Plug Gauge

Major diameter 1 mm to 25 mm

Pitch diameter

0.25 mm to 6 mm

Pitch 0.25mm to 3 mm

0.62 µm

1.7 µm

3.4 µm

Three wire method based on JIS B 0261:1992

Signatories:

1. Eunice Wong Phaik Gaik (Except for Scope Parallel Thread Plug Gauge)

2. Ching Kar Chien 3. Chew Siou Lian

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 4 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Plain Ring Gauge 20 mm to 100 mm 0.4 µm + 12 L µm ‘L’ in metre

Length Comparator JIS B 7420

Glass Scale

0 mm to 200 mm

6.2 µm Measuring

System and Reference Scale on JIS B 7541:2001

Steel Ruler 0 mm to 300 mm 0.068 mm Measuring

Projector JIS B 7516:2005

Angle Protractor

0 Degree to 60 Degree 0.13 Degree Sine Bar, Slip Gauges, Comparator VDI/WDE/DGQ 2618

Signatories:

1. Chew Siou Lian

2. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 5 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: DIMENSIONAL SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Profile Projector (Measuring Accuracy of Individual Axis)

x, y-axis: 0 to 300 mm

3 µm Comparison with standard scale based on JIS B 7184:1999

Measuring Scope (Measuring Accuracy of Individual Axis)

x, y-axis: 0 to 300 mm

2 µm Comparison with standard scale based on JIS 7184:1999

Surface Plate

600 x 600 mm 1200 x 1200mm 1800 x 1800mm

1.3 µm 1.5 µm 1.7 µm

Planekator, micro-indicator based on JIS B 7513:1992

Vernier, Dial & Digital Height Gauge / Linear Height

0 mm to 600 mm (1.3 + 4.7L) µm ‘L’ in metre

Comparison with gauge blocks, height master based on JIS B 7517:1993

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien 4. Chew Siou Lian

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 6 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: MASS

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Standard Weights

1 mg 2 mg 5 mg 10 mg 20 mg 50 mg 100 mg 200 mg 500 mg 1 g 2 g 5 g 10 g 20 g 50 g 100 g 200 g 500 g 1 kg 2 kg 5 kg 10 kg 20 kg 50 kg

0.009 mg 0.009 mg 0.009 mg 0.009 mg 0.010 mg 0.010 mg 0.010 mg 0.011 mg 0.011 mg 0.012 mg 0.013 mg 0.015 mg 0.018 mg 0.018 mg 0.046 mg 0.12 mg 0.14 mg 0.89 mg 0.96 mg 1.5 mg 4.7 mg 0.012 g

0.018 g 9.3 g

Direct Comparison using the ABBA or AB1 …BnA weighing sequence.

Signatories:

1. Ching Kar Chien

2. Chew Siou Lian

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 7 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: MASS SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Weighing Balance (Analytical Balance, Scale, Electronic type, platform or pallet)

Up to 50 g Up to 250 g Up to 5000 g Up to 10 kg Up to 20 kg Up to 150 kg Up to 300 kg Up to 500 kg Up to 1000 kg Up to 3000 kg

0.039 mg 0.27 mg 0.019 g 0.65 g 2.6 g

0.049 kg 0.067 kg 0.12 kg 0.45 kg 1.6 kg

Standard weight based on ASTM E898-88:2013

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien 4. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 8 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: FORCE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Durometer Hardness

Spring force for Type A, B & O Type C, D & DO

0.65 durometer point 0.65 durometer point

Dead weights method

Signatory:

1. Ching Kar Chien

SCOPE OF CALIBRATION: FORCE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Force Gauge / Push-Pull Gauge

0 N to 10 N 10 N to 50 N 50 N to 500 N

500 N to 1000 N

± 0.0011 N ± 0.011 N ± 0.080 N ± 0.80 N

Dead weights method

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien 4. Chew Chii Tsong

SCOPE OF CALIBRATION: FORCE SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Hardness Testing Machine (HRC)

30 to 60 HRC

0.25 HRC Standard hardness block

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 9 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: PRESSURE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Pressure Indicating Devices (Pressure Gauge, Vacuum Gauge, Manometer, Magnehelic) Manometer Vacuum Pneumatic/Hydraulic Hydraulic

-5.8 to 5.8 psi

-12 to 0 psi

0 to 300 psi

300 to 10000 psi

0.00078 psi

0.022 psi

0.049 psi

4.8 psi

Comparison with Reference Pressure Calibrator

Signatories:

1. Ching Kar Chien 2. Chew Siou Lian 3. Chew Chii Tsong

SCOPE OF CALIBRATION: PRESSURE SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Pressure Indicating Devices (Pressure Gauge, Vacuum Gauge, Manometer, Magnehelic) Manometer Vacuum Pneumatic/Hydraulic Hydraulic

-5.8 to 5.8 psi

-12 to 0 psi

0 to 300 psi

300 to 10000 psi

0.00078 psi

0.022 psi

0.049 psi

4.8 psi

Comparison with Reference Pressure Calibrator

Signatories:

1. Ching Kar Chien 2. Chew Siou Lian 3. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 10 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TORQUE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Torque Measurement Devices

0 N.m to 1 N.m Above 1 N.m to 10 N.m Above 10 N.m to 40 N.m

0.0012 N.m 0.020 N.m 0.042 N.m

Calibration using deadweights and known radius of torque wheel.

Signatories:

1. Kenny Ching Ah Hoo

2. Eunice Wong Phaik Gaik 3. Ching Kar Chien

SCOPE OF CALIBRATION: TORQUE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Torque Tools (Torque Wrench, Torque Driver, Fixed or Adjustable)

Up to 1 N.m Above 1 N.m to 10 N.m

Above 10 N.m to 40 N.m Above 40 N.m to 150 N.m Above 150 N.m to 600 N.m

Above 600 N.m to 1000 N.m Above 1000 N.m to 1500 N.m

0.0016 N.m 0.016 N.m 0.18 N.m 0.48 N.m 1.4 N.m 1.8 N.m 2.7 N.m

Based on ISO 6789:2003

Signatories:

1. Ching Kar Chien

2. Chew Siou Lian 3. Chew Chii Tsong

SCOPE OF CALIBRATION: TORQUE SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Torque Tools (Torque Wrench, Torque Driver Fixed or Adjustable)

0 N.m to 1 N.m Above 1 N.m to 5 N.m

0.0055 N.m 0.048 N.m

Based on ISO 6789:2017

Signatories:

1. Ching Kar Chien

2. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 11 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement Capability

Expressed as an Uncertainty(±)*

Remarks

Temperature Sensor -70 oC to -20 oC -20 oC to 0 oC 0 oC to 100 oC

100 oC to 300 oC 300 oC to 500 oC

0.30 oC 0.041 oC 0.022 oC 0.100 oC 0.510 oC

Comparison with PT 100 in Liquid Bath and Temperature Block Calibrator

Surface Sensor

0 oC to 50 oC 50 oC to 200 oC 200 oC to 300 oC

0.50 oC 0.55 oC 0.71 oC

Comparison with reference surface sensor on electronic hotplate / equalizing block of dry bath

Temperature Block Calibrator

-20 oC to 300 oC 300 oC to 600 oC 600 oC to 700 oC

0.061 oC 0.12 oC 1.0 oC

Comparison with PT 100 / Thermocouple Type – K EURAMET cg-13 Version 3.0

Radiation / Infrared Thermometer

0 oC to 50 oC 50 oC to 100 oC 100 oC to 140 oC

0.52 oC 1.3 oC 1.4 oC

Comparison with PT 100 in blackbody source ASTM E 1256:2007

Signatories:

1. Ching Kar Chien 2. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 12 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Themohygro Devices A) Relative Humidity Temperature B) Relative Humidity (Fixed Point) Temperature

20 %rh to 80 %rh 80 %rh to 90 %rh

0 °C to 60 oC

11 %rh 33 %rh 75 %rh 97 %rh

25 oC

1.8 %rh 2.2 %rh

0.10 oC

1.1 %rh 1.3 %rh 1.4 %rh 1.7 %rh

0.31 oC

Comparison with Reference Standard in Temperature/Humidity Chamber Fixed Point Calibration in Salt Chamber

Liquid – In – Glass Thermometer (Total Immersion)

0 oC to 200 oC 0.10 oC Comparison with PT100 in a stirred liquid bath

Signatories:

1. Ching Kar Chien 2. Chew Siou Lian 3. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 13 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Temperature

Indicator

K

J

T

E

N

B

R

S

Pt 100

-200 to 0 oC

0 to 1370 oC

-210 to 0 oC

0 to 1200 oC

-200 to 0 oC

0 to 400 oC

-200 to 0 oC

0 to 1000 oC

-200 to 0 oC

0 to 1300 oC

250 to 600 oC

600 to 1820 oC

-50 to 0 oC

0 to 1760 oC

-50 to 0 oC

0 to 1760 oC

-200 to 850 oC

0.14 oC

0.12 oC

0.12 oC

0.10 oC

0.12 oC

0.10 oC

0.10 oC

0.10 oC

0.16 oC

0.10 oC

0.50 oC

0.15 oC

0.35 oC

0.14 oC

0.33 oC

0.16 oC

0.065 oC

Electrical

simulation using

multi-product

calibrator based

on ITS-90 Table

Signatories:

1. Ching Kar Chien 2. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 14 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Temperature

Simulator

K

J

T

E

N

B

R

S

Pt 100

-200 to 0 oC

0 to 1370 oC

-210 to 0 oC

0 to 1200 oC

-200 to 0 oC

0 to 400 oC

-200 to 0 oC

0 to 1000 oC

-200 to 0 oC

0 to 1300 oC

250 to 600 oC

600 to 1820 oC

-50 to 0 oC

0 to 1760 oC

-50 to 0 oC

0 to 1760 oC

-200 to 850 oC

0.19 oC

0.13 oC

0.16 oC

0.11 oC

0.18 oC

0.10 oC

0.13 oC

0.10 oC

0.25 oC

0.11 oC

0.93 oC

0.23 oC

0.64 oC

0.21 oC

0.60 oC

0.24 oC

0.059 oC

Measurement

using multimeter

based on ITS-90

Table

Signatories:

1. Ching Kar Chien 2. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 15 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Temperature Simulator

K J

T

E

N

B

R

S

Pt 100

-100 ˚C to 1300 ˚C

-100 ˚C to 1200 ˚C

-100 ˚C to 400 ˚C

-100 ˚C to 950 ˚C

-200 ˚C to 1300 ˚C

600 to 1800 ˚C

0 to 1700 ˚C

0 to 1700 ˚C

-100 to 800 ˚C

0.19 ˚C

0.16 ˚C

0.18 ˚C

0.13 ˚C

0.25 ˚C

0.93 ˚C

0.64 ˚C

0.60 ˚C

0.059 ˚C

Measurement using Temperature Calibrator

Temperature Indicator

K J

T

E

N

B

R

S

Pt 100

-100 to 1300 ˚C

-100 to 1200 ˚C

-100 to 400 ˚C

-100 to 950 ˚C

-200 ˚C to 1300 ˚C

600 to 1800 ˚C

0 to 1700 ˚C

0 to 1700 ˚C

-100 to 800 ˚C

0.14 ˚C

0.12 ˚C

0.12 ˚C

0.10 ˚C

0.16 ˚C

0.50 ˚C

0.35 ˚C

0.33 ˚C

0.065 ˚C

Electrical simulation using Temperature Calibrator based on EURAMET/cg-11/v.01

Signatories:

1. Ching Kar Chien 2. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 16 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TEMPERATURE SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Temperature Enclosure

-25 oC to 500 oC

0.58 oC

Using temperature sensor / logger based on AS 2853-1986

Humidity Enclosure

10 % RH to 98 % RH

10 oC to 95 oC

2.5 % RH

0.58 oC

Using Dry bulb & Wet bulb sensor / logger based on JTM K01-1991

Temperature Sensor

-10 oC to 0 oC 0 oC to 100 oC

100 oC to 300 oC 300 oC to 500 oC

0.12 oC 0.10 oC 0.12 oC 0.51 oC

Comparison with PT 100 in Liquid Bath and Temperature Block Calibrator

Signatories:

1. Ching Kar Chien 2. Chew Siou Lian 3. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 17 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: VOLUME

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Volumetric Apparatus (To Deliver) Piston-Operated Pipette, Type A

1 to 100 µl Above 100 to 200 µl

Above 200 to 1000 µl Above 1000 to 2000 µl Above 2000 to 5000 µl

0.17 µl 0.24 µl 3.1 µl 6.5 µl 7.2 µl

Gravimetric Method

One-Mark Pipette

1 to 10 ml Above 10 to 100 ml

Above 100 to 200 ml Above 200 to 500 ml

Above 500 to 1000 ml

0.0066 ml 0.019 ml 0.029 ml 0.069 ml 0.15 ml

Volumetric Apparatus (To Contain) Measuring Cylinder Flask

1 to 10 ml Above 10 to 100 ml

Above 100 to 200 ml Above 200 to 500 ml

Above 500 to 1000 ml

0.041 ml 0.24 ml 0.35 ml 0.58 ml 1.6 ml

Signatories:

1. Chew Siou Lian 2. Ching Kar Chien

SCOPE OF CALIBRATION: VOLUME

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

pH Meter

4.01 pH 7.01 pH 10.01 pH

0.03 pH 0.03 pH 0.06 pH

Standard Buffer Solutions

Refractometer / Sucrose Mass Fraction / Refractive Index

0 % to 50 % Brix

(1.333 nD ~ 1.420 nD)

± 0.32 % Brix

Sucrose Powder with analytical balance

Signatories:

1. Chew Siou Lian 2. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 18 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Measuring Instrument

DC Voltage

0 V to 320 mV 0.32 V to 3.2 V

3.2 V to 32 V 32 V to 320 V

320 V to 1050 V

5.7 μV/V + 1.1 μV 14 μV/V - 1.4 μV 13 μV/V + 2.4 μV 12 μV/V + 33 μV 16 μV/V - 1.5 mV

Generation using calibrator model Wavetek 9100

AC Voltage

0 V to 1050 V

See Matrix A

DC Current

0 A to 320 µA

0.32 mA to 3.2 mA 3.2 mA to 32 mA 32 mA to 320 mA 0.32 A to 3.2 A 3.2 A to 20 A

390 μA/A + 5.2 mA 5.9 μA/A + 0.13 mA 34 μA/A + 0.038 mA 40 μA/A - 0.14 mA

0.83 mA/A - 0.26 mA 0.58 mA/A + 0.55 mA

AC Current

0 A to 20 A

See Matrix B

Resistance

0 Ω to 40 Ω

40 Ω to 400 Ω 400 Ω to 4 kΩ 4 kΩ to 40 kΩ

40 kΩ to 400 kΩ 400 kΩ to 4 MΩ 4 MΩ to 40 MΩ

40 MΩ to 400 MΩ

0.055 mΩ/Ω + 0.18 mΩ 0.017 mΩ/Ω + 1.7 mΩ 0.024 mΩ/Ω - 0.96 mΩ 0.019 mΩ/Ω + 19 mΩ 0.036 mΩ/Ω – 0.67 Ω

0.14 mΩ/Ω - 42 Ω 0.020 mΩ/Ω + 0.43 kΩ

0.25 mΩ/Ω - 9.2 kΩ

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 19 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

Matrix A AC Voltage Measurement

Range

Frequency

Hz kHz

0 to 10 10 to 50 0.05 to 1 1 to 30 30 to 50 50 to 100

0 mV to 320 mV 0.0007 +

0.12 0.0027 +

0.084 0.0024 + 0.084 0.00028 +

0.21 0.00093 + 0.21 0.00028 + 0.47

320 mV to 3.2 V 0.34 + 0.006 0.24 + 0.0093 0.2 + 0.0092 0.66 +

0.00010 1.6 – 0.29 1.5 + 0.00010

3.2 V to 32 V 0.59 – 0.79 0.48 – 0.77 0.48 – 0.77 1.2 – 1.6 0.62 + 2.7 2.2 – 2.4

32 V to 320 V 0.34 + 7.2 0.35 + 3.4 0.35 + 3.4 0.11 +32

320 V to 1050 V 0.63 - 84 0.63 - 84 0.98 - 195

The expanded uncertainties given in above table are expressed in mV/V + mV

Matrix B AC Current Measurement

Range

Frequency (Hz )

Frequency ( kHz )

0 to 10 0.01 to 3 3 to 10 10 to 30

0 to 320 µA 0.11 mA/A + 285 mA

0.096 mA/A + 285 mA

0.068 mA/A + 285 mA

0.089 mA/A + 285 mA

320 µA to 3.2 mA 0.16 mA/A + 0.27 μA

0.27 mA/A + 0.23 μA

0.22 mA/A + 0.24 μA

0.22 mA/A + 0.25 μA

3.2 mA to 32 mA 0.39 mA/A - 0.49 μA

0.40 mA/A - 0.18 μA

0.40 mA/A - 0.35 μA

0.40 mA/A - 0.35 μA

32 mA to 320 mA 0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.32 A to 3.2 A 0.73 mA/A - 0.053 mA

0.69 mA/A + 0.067 mA

0.69 mA/A + 0.067 mA

-

3.2 A to 20 A - 0.35 mA/A - -

Signatories:

1. Kenny Ching Ah Hoo 2. Gabriel Ching Kar Ee 3. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 20 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Measuring Instruments Frequency

0.5 Hz to 320 Hz 320 Hz to 3.2 kHz 3.2 kHz to 32 kHz 32 kHz to 3.2 MHz 3.2 MHz to 10 MHz

29 ppm 29 ppm 29 ppm 29 ppm 29 ppm

Generation using Calibrator model Wavetek 9100

Capacitance

0.5 to 4 nF

4 nF to 40 nF 40 nF to 400 nF 400 nF to 4 µF 4 µF to 40 µF

40 µF to 400 µF 400 µF to 4 mF 4 mF to 40 mF

4.9 mF/F 3.7 mF/F 2.4 mF/F 2.7 mF/F 3.7 mF/F 3.9 mF/F 3.9 mF/F 5.1 mF/F

Resistance

10 mΩ to 90 mΩ

100 mΩ to 900 mΩ 1 Ω to 9 Ω

10 Ω to 90 Ω 100 Ω to 900 Ω

1 kΩ to 9 kΩ 10 kΩ to 90 kΩ

100 kΩ to 900 kΩ 1 MΩ to 9 MΩ

10 MΩ to 90 MΩ 100 MΩ to 199.99 MΩ

8.7 mΩ/Ω 1.2 mΩ/Ω 0.24 mΩ/Ω 0.17 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω

Direct measurement with TIME Decade Resistance Box

Clamp Meter DC Current - 50 Turn Coil

(+/- polarities) 1.6 A to 16 A 16 A to 160 A 160 A to 525 A 525 A to 1000 A

0.37 mA/A 0.37 mA/A 0.12 mA/A 0.58 mA/A

Generating using Calibrator model Wavetek 9100 with Current Coil

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 21 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Clamp Meter AC Current - 50 Turn Coil

1.6 A to 16 A 10 Hz to 100 Hz 100 Hz to 440 Hz

16 A to 160 A

10 Hz to 100 Hz 100 Hz to 440 Hz

160 A to 1000 A 10 Hz to 100 Hz 100 Hz to 440 Hz

1.3 mA/A 3.2 mA/A

1.4 mA/A 3.4 mA/A

1.3 mA/A 3.2 mA/A

Generating using Calibrator model Wavetek 9100 with Current Coil

Generating Instruments DC Voltage

0 mV to 100 mV 100 mV to 1 V

1 V to 10 V 10 V to 100 V

100 V to 1000 V

36 nV/V + 1.7 µV 7.7 µV/V + 0.85 µV 6.9 µV/V + 1.6 µV 8.3 µV/V – 12 µV 9.6 µV/V – 140 µV

Measurement using KeySight 3458A Digital Multimeter

AC Voltage

0 to 1000 V

See Matrix C

DC Current

0 µA to 100 µA 100 µA to 1 mA 1 mA to 10 mA

10 mA to 100 mA 100 mA to 1 A

46 mA/A + 17 nA - 43 µA/A + 67 nA 4.9 µA/A + 19 nA 5.6 µA/A + 12 nA

0.12 mA/A - 12 µA

1 A to 3 A

1.8 mA/A - 1.7 mA

Measurement using Agilent 34401A Multimeter

AC Current

0 to 3 A

See Matrix D

Resistance

0 Ω to 10 Ω

10 Ω to 100 Ω 100 kΩ to 1 kΩ 1 kΩ to 10 kΩ

10 kΩ to 100 kΩ 100 kΩ to 1 MΩ 1 MΩ to 10 MΩ

10 MΩ to 100 MΩ

42 µΩ/Ω + 0.32 mΩ 15 µΩ/Ω + 0.18 mΩ 13 µΩ/Ω + 0.44 mΩ 13 µΩ/Ω - 0.15 mΩ 15 µΩ/Ω - 21 mΩ 26 µΩ/Ω - 1.1 Ω 73 µΩ/Ω - 49 Ω

360 µΩ/Ω - 2.9 kΩ

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 22 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Generating Instruments Frequency Capacitance

1 Hz to 40 Hz 40 Hz to 100 Hz

100 Hz to 100 kHz 100 kHz to 10 MHz

1 µF to 10 µF

10 µF to 100 µF

566 ppm 114 ppm 114 ppm 127 ppm

40 µF/F + 150 nF 65 mF/F – 0.50 µF

Measuring using SANWA CD 800a Multimeter

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 23 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

AC Voltage: Matrix C

Frequecny Voltage

Hz kHz

0 to 40 0.04 to 1 1 to 20 20 to 50 50 to 100

0 to 10 mV 2.9 mV/V + 19 μV

0.88 mV/V + 3.7 μV

0.55 mV/V + 7.4 μV

0.31 mV/V + 15 μV

1.1 mV/V + 15 μV

10 mV to 100 mV - 0.23 mV/V + 52 μV

0.12 mV/V + 12 μV

0.11 mV/V + 12 μV

0.80 mV/V + 9.6 μV

0.71 mV/V + 18 μV

100 mV to 1 V 109 μV/V + 18 μV

114 μV/V + 12 μV

102 μV/V + 12 μV

184 μV/V + 71 μV

185 μV/V + 70 μV

1 V to 10 V 76 μV/V + 50 μV

77 μV/V + 49 μV

69 μV/V + 45 μV

188 μV/V + 67 μV

190 μV/V + 66 μV

10 V to 100 V 79 μV/V + 19 μV

80 μV/V + 19 μV

82 μV/V - 81 μV

245 μV/V - 500 μV

245 μV/V - 488 μV

100 V to 1000 V 68 μV/V + 1.2 mV

68 μV/V + 1.2 mV

- - -

AC Current: Matrix D Frequency Voltage

Hz kHz Measuring Instrument 0 to 50 0.05 to 1 1 to 5 5 to 30

0 A to 1 mA 0.26 mA/A + 0.38 μA

0.17 mA/A + 70 nA

0.21 mA/A + 59 nA

0.92 mA/A + 59 nA

KeySight 3458A

1 mA to 10 mA 0.19 mA/A + 0.45 μA

0.23 mA/A + 9.9 nA

0.23 mA/A + 45 nA

0.15 mA/A + 0.84 nA

KeySight 3458A

10 mA to 100 mA 0.21 mA/A + 0.21 μA

0.21 mA/A + 0.17 μA

0.21 mA/A + 0.17 μA

0.21 mA/A + 0.17 μA

KeySight 3458A

100 mA to 1 A 0.46 mA/A - 25 μA

0.45 mA/A - 25 μA

0.45 mA/A - 25 μA

- KeySight 3458A

1 A to 3 A 1.6 mA/A - 1.1 mA

1.6 mA/A - 1.1 mA

1.6 mA/A - 1.1 mA

- Agilent 34401A

Signatories:

1. Gabriel Ching Kar Ee 2. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 24 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Measuring Instruments DC Voltage

0 mV to 320 mV 0.32 V to 3.2 V 3.2 V to 32 V 32 V to 320 V

320 V to 1050 V

5.7 μV/V + 1.1 μV 14 μV/V - 1.4 μV 13 μV/V + 2.4 μV 12 μV/V + 33 μV 16 μV/V - 1.5 mV

Generation using Calibrator model Wavetek 9100

AC Voltage

0 V to 1050 V

See Matrix A

DC Current

0 µA to 320 µA

0.32 mA to 3.2 mA 3.2 mA to 32 mA 32 mA to 320 mA 0.32 A to 3.2 A 3.2 A to 20 A

390 μA/A + 5.2 mA 5.9 μA/A + 0.13 mA 34 μA/A + 0.038 mA 40 μA/A - 0.14 mA

0.83 mA/A - 0.26 mA 0.58 mA/A + 0.55 mA

AC Current

0 A to 20 A

See Matrix B

Resistance

0 Ω to 40 Ω

40 Ω to 400 Ω 400 Ω to 4 kΩ 4 kΩ to 40 kΩ

40 kΩ to 400 kΩ 400 kΩ to 4 MΩ 4 MΩ to 40 MΩ

40 MΩ to 400 MΩ

0.055 mΩ/Ω + 0.18 mΩ 0.017 mΩ/Ω + 1.7 mΩ 0.024 mΩ/Ω - 0.96 mΩ 0.019 mΩ/Ω + 19 mΩ 0.036 mΩ/Ω – 0.67 Ω

0.14 mΩ/Ω - 42 Ω 0.020 mΩ/Ω + 0.43 kΩ

0.25 mΩ/Ω - 9.2 kΩ

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 25 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL SITE: CATEGORY I Matrix A AC Voltage Measurement

Range

Frequency

Hz kHz

0 to 10 10 to 50 0.05 to 1 1 to 30 30 to 50 50 to 100

0 mV to 320 mV 0.0007 +

0.12 0.0027 +

0.084 0.0024 + 0.084 0.00028 +

0.21 0.00093 + 0.21 0.00028 + 0.47

320 mV to 3.2 V 0.34 + 0.006 0.24 + 0.0093 0.2 + 0.0092 0.66 +

0.00010 1.6 – 0.29 1.5 + 0.00010

3.2 V to 32 V 0.59 – 0.79 0.48 – 0.77 0.48 – 0.77 1.2 – 1.6 0.62 + 2.7 2.2 – 2.4

32 V to 320 V 0.34 + 7.2 0.35 + 3.4 0.35 + 3.4 0.11 +32

320 V to 1050 V 0.63 - 84 0.63 - 84 0.98 - 195

The expanded uncertainties given in above table are expressed in mV/V + mV

Matrix B AC Current Measurement

Range

Frequency (Hz )

Frequency ( kHz )

0 to 10 0.01to 3 3 to 10 10 to 30

0 to 320 µA 0.11 mA/A + 285 mA

0.096 mA/A + 285 mA

0.068 mA/A + 285 mA

0.089 mA/A + 285 mA

320 µA to 3.2 mA 0.16 mA/A + 0.27 μA

0.27 mA/A + 0.23 μA

0.22 mA/A + 0.24 μA

0.22 mA/A + 0.25 μA

3.2 mA to 32 mA 0.39 mA/A - 0.49 μA

0.40 mA/A - 0.18 μA

0.40 mA/A - 0.35 μA

0.40 mA/A - 0.35 μA

32 mA to 320 mA 0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.95 mA/A - 18 μA

0.32 A to 3.2 A 0.73 mA/A - 0.053 mA

0.69 mA/A + 0.067 mA

0.69 mA/A + 0.067 mA

-

3.2 A to 20 A - 0.35 mA/A - -

The expanded uncertainties given in above table are expressed in mA/A

Signatories:

1. Kenny Ching Ah Hoo 2. Gabriel Ching Kar Ee 3. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 26 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Measuring Instruments Frequency

0.5 Hz to 320 Hz 320 Hz to 3.2 kHz 3.2 kHz to 32 kHz 32 kHz to 3.2 MHz 3.2 MHz to 10 MHz

29 ppm 29 ppm 29 ppm 29 ppm 29 ppm

Generation using Calibrator model Wavetek 9100

Capacitance

0.5 nF to 4 nF 4 nF to 40 nF

40 nF to 400 nF 400 nF to 4 µF 4 µF to 40 µF

40 µF to 400 µF 400 µF to 4 mF 4 mF to 40 mF

4.9 mF/F 3.7 mF/F 2.4 mF/F 2.7 mF/F 3.7 mF/F 3.9 mF/F 3.9 mF/F 5.1 mF/F

Resistance

10 mΩ to 90 mΩ

100 mΩ to 900 mΩ 1 Ω to 9 Ω

10 Ω to 90 Ω 100 Ω to 900 Ω

1 kΩ to 9 kΩ 10 kΩ to 90 kΩ

100 kΩ to 900 kΩ 1 MΩ to 9 MΩ

10 MΩ to 90 MΩ 100 MΩ to 199.99 MΩ

8.7 mΩ/Ω 1.2 mΩ/Ω 0.24 mΩ/Ω 0.17 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω 0.16 mΩ/Ω

Direct measurement with TIME Decade Resistance Box

Clamp Meter DC Current - 50 Turn Coil

(+/- polarities) 1.6 to 16 A 16 to 160 A 160 to 525 A 525 to 1000 A

0.37 mA/A 0.37 mA/A 0.12 mA/A 0.58 mA/A

Generating using Calibrator model Wavetek 9100 with Current Coil

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 27 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Clamp Meter AC Current - 50 Turn Coil

1.6 to 16 A 10 to 100 Hz 100 to 440 Hz

16 to 160 A 10 to 100 Hz 100 to 440 Hz

160 to 1000 A 10 to 100 Hz 100 to 440 Hz

1.3 mA/A 3.2 mA/A

1.4 mA/A 3.4 mA/A

1.3 mA/A 3.2 mA/A

Generating using Calibrator model Wavetek 9100 with Current Coil

Generating Instruments DC Voltage

0 to 100 mV 100 mV to 1 V

1 V to 10 V 10 V to 100 V

100 V to 1000 V

3.2 nV/V + 1.7 μV 7.7 μV/V + 0.85 μV 6.9 μV/V + 1.6 μV 8.3 μV/V - 12 μV 9.6 μV/V - 140 μV

Measurement Keysight 3458A

Digital Multimeter

AC Voltage 0 to 1000 V

See Matrix C

DC Current

0 to 10 mA

100 μA to 1 mA 1 mA to 10 mA

10 mA to 100 mA 100 mA to 1 A

1 A to 3 A

46 mA/A + 17 nA

- 43 μA /A + 67 nA 4.9 μA /A + 19 nA 5.6 μA/A + 12 nA

0.12 mA/A - 12 μA 18 mA/A – 17 mA

AC Current 0 to 3 A See Matrix D

Resistance

0 to 10 Ω

10 Ω to 100 Ω 100 Ω to 1 kΩ 1 kΩ to 10 kΩ

10 kΩ to 100 kΩ 100 kΩ to 1 MΩ 1 MΩ to 10 MΩ

10 MΩ to 100 MΩ

42 nΩ/Ω + 0.32 mΩ 15 μΩ/Ω + 0.18 mΩ 13 μΩ/Ω + 0.44 mΩ 13 μΩ/Ω - 0.15 mΩ 15 μΩ/Ω - 21 mΩ 26 μΩ/Ω - 1.1 Ω 73 μΩ/Ω - 49 Ω

360 μΩ/Ω - 2.9 kΩ

Frequency 1 Hz to 40 Hz 40 Hz to 100 Hz

100 Hz to 100 kHz 100 kHz to 10 MHz

566 ppm 114 ppm 114 ppm 127 ppm

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 28 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: ELECTRICAL SITE: CATEGORY I Generating Instruments

AC Voltage:

Matrix C

Frequecny Voltage

Hz kHz

0 to 40 0.04 to 1 1 to 20 20 to 50 50 to 100

0 to 10 mV 2.9 mV/V + 19 μV

0.88 mV/V + 3.7 μV

0.55 mV/V + 7.4 μV

0.31 mV/V + 15 μV

1.1 mV/V + 15 μV

10 mV to 100 mV - 0.23 mV/V + 52 μV

0.12 mV/V + 12 μV

0.11 mV/V + 12 μV

0.80 mV/V + 9.6 μV

0.71 mV/V + 18 μV

100 mV to 1 V 109 μV/V + 18 μV

114 μV/V + 12 μV

102 μV/V + 12 μV

184 μV/V + 71 μV

185 μV/V + 70 μV

1 V to 10 V 76 μV/V + 50 μV

77 μV/V + 49 μV

69 μV/V + 45 μV

188 μV/V + 67 μV

190 μV/V + 66 μV

10 V to 100 V 79 μV/V + 19 μV

80 μV/V + 19 μV

82 μV/V - 81 μV

245 μV/V - 500 μV

245 μV/V - 488 μV

100 V to 1000 V 68 μV/V + 1.2 mV

68 μV/V + 1.2 mV

- - -

The expanded uncertainties given in above table are expressed in mV/V

AC Current: Matrix D Frequency Voltage

Hz kHz Measuring Instrument 0 to 50 0.05 to 1 1 to 5 5 to 30

0 A to 1 mA 0.26 mA/A + 0.38 μA

0.17 mA/A + 70 nA

0.21 mA/A + 59 nA

0.92 mA/A + 59 nA

KeySight 3458A

1 mA to 10 mA 0.19 mA/A + 0.45 μA

0.23 mA/A + 9.9 nA

0.23 mA/A + 45 nA

0.15 mA/A + 0.84 nA

KeySight 3458A

10 mA to 100 mA 0.21 mA/A + 0.21 μA

0.21 mA/A + 0.17 μA

0.21 mA/A + 0.17 μA

0.21 mA/A + 0.17 μA

KeySight 3458A

100 mA to 1 A 0.46 mA/A - 25 μA

0.45 mA/A - 25 μA

0.45 mA/A - 25 μA

- KeySight 3458A

1 A to 3 A 1.6 mA/A - 1.1 mA

1.6 mA/A - 1.1 mA

1.6 mA/A - 1.1 mA

- Agilent 34401A

The expanded uncertainties given in above table are expressed in mA/A

Signatories:

1. Gabriel Ching Kar Ee 2. Ching Kar Chien

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 29 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TIME AND FREQUENCY

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Stopwatch and Timer Tachometer (Non-contact)

1 sec to 60 minutes

0 to 2000 rpm

2000 to 5000 rpm

0.063 sec

2.6 rpm

7.1 rpm

Calibrated by using comparison method according to NIST stopwatch and timer calibration. Comparison with reference

tachometer.

Signatories:

1. Gabriel Ching Kar Ee 2. Ching Kar Chien 3. Chew Chii Tsong

Schedule

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA

MS ISO/IEC 17025

Issue date: 2 November 2017 Valid until: 15 October 2020

NO: SAMM 256 (Issue 2, 2 November 2017 replacement of SAMM 256 dated 28 September 2017)

Page: 30 of 30

Scan t

his

QR

Co

de

or

vis

it w

ww

.jsm

.go

v.m

y/c

ab

-dir

ecto

ries f

or

the c

urr

ent

sco

pe o

f a

ccre

ditatio

n

SCOPE OF CALIBRATION: TIME AND FREQUENCY SITE: CATEGORY I

Instrument Calibrated/

Measurement Parameter

Range

Calibration and Measurement

Capability Expressed as an Uncertainty(±)*

Remarks

Stopwatch and Timer Tachometer (Non-contact)

0 to 60 sec

1 minute to 5 minute

5 minute to 60 minute

0 to 2000 rpm

2000 to 5000 rpm

0.12 sec

0.12 sec

0.34 sec

2.6 rpm

7.1 rpm

Calibrated by using comparison method according to NIST stopwatch and timer calibration. Comparison with reference tachometer.

Signatories:

1. Gabriel Ching Kar Ee 2. Ching Kar Chien 3. Chew Chii Tsong