Host Interface Document Version 1.0 Sysmex XE-2100 Automated Hematology Analyzer Part No. 011600600

Transcript of Host manual sysmex-xe2100

Page 1: Host manual sysmex-xe2100

Host Interface DocumentVersion 1.0

Sysmex XE-2100 Automated Hematology AnalyzerPart No. 011600600

Page 2: Host manual sysmex-xe2100

WarrantyRoche Diagnostics makes no warranties for the Sysmex XE-2100 System Host Interface Manualbeyond those set forth in the operator’s manuals for the Sysmex XE-2100 Analyzer. RocheDiagnostics disclaims all other warranties, express or implied, including but not limited to theimplied warranties of merchantability and fitness for a particular purpose.

In no event shall Roche Diagnostics be liable for the incidental or consequential damages arisingfrom the use of the interface.

© 2000, Roche Diagnostics and Sysmex Corporation. All rights reserved.

The contents of this manual, including all graphics and photographs are the property of SysmexCorporation. Information in this document is subject to change without notice. RocheDiagnostic and Sysmex Corporation shall not be liable for technical or editorial errors oromissions contained herein.

No part of this document may be reproduced or transmitted in any form or by any means,electronic or mechanical, for any purpose, without the express written permission of RocheDiagnostics and Sysmex Corporation.

Sysmex is a registered trademark of Sysmex Corporation.

This manual was created by the Sysmex Corporation and the Roche Diagnostics TechnicalPublications Department. Questions/comments regarding the content of this manual can bedirected to your local Roche Diagnostics representative, or to:

Roche Diagnostics CorporationP.O. Box 504579115 Hague RoadIndianapolis, IN USA46250-0457Attention: Technical Publications Department

US Order Number: 011600600

Page 3: Host manual sysmex-xe2100

Revised Manual PagesRevised pages for this manual are provided by Roche Diagnostics when necessary. No part ofthis publication may be reproduced in any form or by any means without prior writtenpermission.

Publication Date Pages AffectedReference No.

Version 1.0 July 2000 entire document

Page 4: Host manual sysmex-xe2100

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

APPENDIX A TECHNICAL INFORMATION

1. OUTPUT FORMAT TO HOST COMPUTER ........................................ A-11.1 Hardware ........................................................................................ A-11.2 Software.......................................................................................... A-2

2. COMMUNICATION SPECIFICATIONS OF ANALYSISINFORMATION WITH HOST COMPUTER ............................. A-13

2.1 Outline.......................................................................................... A-132.2 Hardware ...................................................................................... A-142.3 Software........................................................................................ A-15

3. ID BAR CODE SPECIFICATIONS ...................................................... A-22

Page 5: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-1

1. OUTPUT FORMAT TO HOST COMPUTER

The bit serial voltage type, which conforms to the RS-232C interface, is used for host computeroutput of XE-2100. The serial interface port for the connection with the host computer is on therear panel of the IPU.The specifications conform to Standard for Bit Serial Interface of Japan Clinical Instrumentautomation Society.

1.1 Hardware

1. Connector

• Connect to the host computer output connector on the rear panel of the IPU.• Use a 9-pin, D-SUB female connector.• Pitch of connector fixing screws adopts the size in inch.

2. Connector Signal

Pin No. Signal Name Signal Direction12 Receive Data (RxD) To XE-2100 from HOST3 Transmit Data (TxD) From XE-2100 to HOST4 Data Terminal Ready (DTR) From XE-2100 to HOST5 Signal Ground (SG)6 Data Set Ready (DSR) To XE-2100 from HOST7 Request to Send (RTS) From XE-2100 to HOST8 Clear to Send (CTS) To XE-2100 from HOST9

Table A-1: Connector Signal

3. Communication Format

The format is start-stop a synchronization half duplex (asterisk indicates setting at shipment fromfactory), and adjustment can be made as follows.

Baud Rate 600, 1200, *2400, 4800, 9600, 14400 (BPS)Code *7-bit, 8-bitStop Bit 1-bit, *2-bitParity Bit None, *Even, OddClass Class A, *Class BInterval 0, 1, *2, 3, 5, 7, 10, 15 (seconds)

Table A-2: Communication Format Setting

Page 6: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-2 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

4. Signal Level

Signal level of the RS-232C conforms to EIA RS-232C V.24.

Level Data Signal Control Signal+3V or Higher Logic "0," Start Bit ON-3V or Lower Logic "1," Stop Bit OFF

Table A-3: Signal Level

1.2 Software

1. Communication Format

1) CodeASCII codes are used for all the output to the computer.

2) Text Format

Order of Transmission

STX

ETX

Figure A-1: Text Format

"STX" is sent at the beginning, and "ETX" is sent at the end of data.

3) Transmission ProcedureTwo classes are prepared and can be selected depending on the system status.(Class B is selected at the shipment from factory.)• Class A

One-way data transmission which requires no response from the computer.• Class B

Two-way communication which requires response (ACK, NAK) from the computer.

Page 7: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-3

Sends data (1 text)

In case of ACK, sends next text. In case of NAK, resends (up to 3 times)

Sends ACK (06H) if there is no error, or NAK (15H) if there is an error

[Computer][XE-2100]

Figure A-2: Transmission Procedure

4) Transmission ErrorIf an error is detected, the transmission is interrupted and the error message is displayed onthe IPU. The operator must recover the error. Cases of errors are as follows.• Control signal "DSR" is OFF.

• After the data transmission, there is no response from the computer in 15 seconds. (ClassB only)

• "NAK" is sent back after transmitting three times. (Class B only)

5) Transmission TimingSelection between the transmission for every analysis cycle and the batch transmission ofstored data is possible by the IPU settings. The data transmission interval can be set (fromzero second) also.

6) Transmission IntervalThe transmission interval of sample data can be set on the IPU. The interval is a time periodafter the ACK/NAK response until the next data transmission in the case of class B.

2. Data Format

The output to host computer consists of sample data and QC data, and they differ in length andcontent of the text. (They are classified by the sample class codes.)

• Output of analysis results is the text data only, and the text class I is always "D."

• The text distinction II is usually "1." But when the text length is 255 bytes or more, the textis divided into two or more for transmission, and the text block order is expressed by thenumeral of the text distinction II. (ETB code is not used.)

• Sample distinction code is "U" in the case of sample data, and is "C" in the case of QC data.

• The QC data is output in the QC format. This data can be output by the cursor rangedesignation in the QC menu of the IPU.

NOTE: • XE-2100 corresponds to two types of host format. As a default, the format Ais installed. If you wish to change the format from A to B, contact your Sysmexservice representative.

Page 8: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-4 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

3. Analysis Data Format

1) Output OrderThe order of transmission is from the top parameter to the bottom. The data sent is the mostsignificant digit first. Zero-suppression is not carried out.

2) Decimal PointDecimal point is not sent. Therefore, it is necessary to add decimal point specified for eachitem at the host computer.

3) DateThe order of Year/Month/Day is fixed. Zero-suppression is not carried out.

4) Rack No.This is the number assigned to a sample rack, and consists of 6-digit number. Zero-suppression is not carried out.

5) Tube PositionThis indicates the analysis position of aimed sample in a sample rack, and consists ofnumber from 1 to 10. Zero-suppression is not carried out.

6) Sequence No.This indicates the sequence number of the sample analyzed on the same day, and consists of10-digit number. Zero-suppression is not carried out.

7) Sample ID NumberThis number consists of 15 digits in numeral or alphabet which may include a hyphen "-"(2HD) between numerals as needed. A hyphen “-” have to be included in 15 digits. Zerosuppression is not carried out.

8) Construction and Flag of Numerical ValueThe numerical value is constructed as follows. Zero suppression is not carried out.For the RESERVED parameters, "0" is output.

Details of Flag0: Normal1: Analysis data is greater than the preset Upper

Patient Mark Limit.2: Analysis data is less than the preset Lower

Mark Limit.3: Out of linearity limit.4: Analysis data is less reliable

9) Abnormal Value DataWhen the value data is displayed with "----," the data is output in the form of "*0000."However, in case that analysis for that data is not ordered, it is reported as " " (all spaces).

Least Significant Digit

Most Significant Digit

X X X X F

FlagData

Page 9: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-5

10) Instrument NameThis is the information to identify each instrument, and consists of alpha-numeric (capital). Itcan be used when more than one XE-2100 is connected to one host computer.It can be set up in the IPU, and the one for 16-character of the head is outputted. The space isput at the back when the number of the characters is under 16-character.

11) Analysis InformationThis is to display the analysis status of a sample.0: Normal analysis1: Abnormal analysis

12) Sample Judgment InformationThis is to display the judgment information of aimed sample whether re-analysis isnecessary.0: Negative1: Positive only2: Error only3: Positive+ErrorQ: QC data

13) Order InformationThis indicates the analysis order information for each analysis parameter."0": Not analyze"1": Analyze

14) Sample Information (Flag)

PLT SUSPECTPLT ABNORMALRBC SUSPECTRBC ABNORMALWBC SUSPECTWBC ABNORMAL

The existence of the IP message of WBC, RBC or PLT is indicated.0: None1: Existing

15) Sample ID InformationThis displays the implication of the sample number.4: ID bar code reader readable sample number2: ID bar code reader not readable sample number0: Other

Page 10: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-6 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

16) Analysis ModeThis is to show the one of following.1: Manual analysis2: Sampler analysis3: Closed analysis4: Capillary analysis

17) Patient IDThis number consists of maximum 16 digits in numeral or alphabet. The space is put at theback when the number of the characters is under 16-character.

18) Positive (Diff)When a blood cell differentiation value is abnormal, output “1,” if not, output “0.”

19) Positive (Morph)When a blood cell morphology is abnormal, output “1,” if not, output “0.”

20) Positive (Count)When a blood cell count value is abnormal, output “1,” if not, output “0.”

21) Error (Func)When an analysis error other than the ID bar code read error has been occurred, output “1,”if not, output “0.”

22) Error (Result)When an analysis error which one of "cannot absorb blood," "insufficient amount of blood,"or "sample innate error" has been occurred, output “1,” if not, output “0.”

23) The Unit InformationWhen the Netherlands SI unit is being used, output “1,” if others, output “0.”

24) WBC InformationWhen WBC value is corrected by the NRBC value, output “1,” if others, output “0.”

25) PLT InformationWhen optical PLT is adopted as a PLT value, output “1,” if others, output “0.”

26) Instrument IDA manufacturer has an instrument ID composed of fixed 22 characters with the informationto distinguish an instrument.

Page 11: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-7

Analysis Data Format 1

Parameter No. of Chars. ExampleSTX 1 (02H)Text Distinction Code 1 1 "D"Text Distinction Code 2 1 "1"Sample Distinction Code 1 "U"Instrument Name 16 XX~XXSequence No. 10 XX~XXRESERVED 3 XXXSample ID No. 15 XX~XXYear 4 XXXXMonth 2 XXDay 2 XXHour 2 XXMinute 2 XXRESERVED 2 XXRack No. 6 XXXXXXTube Position 2 XXSample ID Information 1 XAnalysis Mode 1 XPatient ID 16 XX~XXAnalysis Information 1 XSample Judgment Information 1 XPositive (Diff) 1 XPositive (Morph) 1 XPositive (Count) 1 XError (Func) 1 XError (Result) 1 XOrder Information 1 XSample Information (INTERPRETATION) 6 XXXXXXThe Unit Information 1 XWBC Information 1 XPLT Information 1 XRESERVED 63 “00~00”RESERVED (Instrument ID: Fixed) 22 “XE-2100XXX---X”ETX 1 (03H)Total 191

Table A-4: Analysis Data Format 1

Page 12: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-8 Sysmex XE-2100 IPU Operator’s Manual -- November 1999 N.A.

Analysis Data Format 2

Parameter No. of Chars. ExampleSTX 1 (02H)Text Distinction Code 1 1 "D"Text Distinction Code 2 1 "2"Sample Distinction Code 1 "U"Instrument Name 16 XX~XXSequence No. 10 XX~XXRESERVED 3 XXXSample ID No. 15 XX~XXWBC (x103/µL) 6 XXX.XXFRBC (x106/µL) 5 XX.XXFHGB (g/dL) 5 XXX.XFHCT (%) 5 XXX.XFMCV (fL) 5 XXX.XFMCH (pg) 5 XXX.XFMCHC (g/dL) 5 XXX.XFPLT (x103/µL) 5 XXXXFLYMPH% (%) 5 XXX.XFMONO% (%) 5 XXX.XFNEUT% (%) 5 XXX.XFEO% (%) 5 XXX.XFBASO% (%) 5 XXX.XFLYMPH# (x103/µL) 6 XXX.XXFMONO# (x103/µL) 6 XXX.XXFNEUT# (x103/µL) 6 XXX.XXFEO# (x103/µL) 6 XXX.XXFBASO# (x103/µL) 6 XXX.XXFRDW-CV (%) 5 XXX.XFRDW-SD (fL) 5 XXX.XFRESERVED 5 XXXXXMPV (fL) 5 XXX.XFRESERVED 5 XXXXXRET% (%%) 5 XXX.XFRET# (x104/µL) 5 XX.XXFIRF (%) 5 XXX.XFRESERVED 5 XXXXXRESERVED 5 XXXXXRESERVED 5 XXXXXRESERVED 5 XXXXXNRBC% (/100WBC) 6 XXXX.XFNRBC# (x103/µL) 6 XXX.XXFRESERVED 16 “00~00”RESERVED (Instrument ID: Fixed) 22 “XE-2100XXX---X”ETX 1 (03H)Total 255

Table A-5: Analysis Data Format 2

Page 13: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-9

4. QC Data Format

The QC data format is used to output to XMcontrol file, and X/L-J control files.The QC numbers indicate the following QC files.

QC Data Format 1

Parameter No. of Chars. ExampleSTX 1 (02H)Text Distinction Code 1 1 "D"Text Distinction Code 2 1 "1"Sample Distinction Code 1 "C"Quality Control No. 1 XYear 2 XXMonth 2 XXDay 2 XXHour 2 XXMinute 2 XXInstrument Name 16 XX~XXRBC (x106/µL) 4 XX.XXHGB (g/dL) 4 XXX.XHCT (%) 4 XXX.XMCV (fL) 4 XXX.XMCH (pg) 4 XXX.XMCHC (g/dL) 4 XXX.XRDW-CW (%) 4 XXX.XRDW-SD (fL) 4 XXX.XPLT (x103/µL) 4 XXXXPDW (fL) 4 XXX.XMPV (fL) 4 XXX.XP-LCR (%) 4 XXX.XPCT (%) 4 XXX.XWBC (x103/µL) 5 XXX.XXNEUT% (%) 4 XXX.XLYMPH% (%) 4 XXX.XMONO% (%) 4 XXX.XEO% (%) 4 XXX.XBASO% (%) 4 XXX.X

(To continue to next page)

Table A-6-1: QC Data Format 1

Page 14: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-10 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

(Continued from previous page)

NEUT# (x103/µL) 5 XXX.XXLYMPH# (x103/µL) 5 XXX.XXMONO# (x103/µL) 5 XXX.XXEO# (x103/µL) 5 XXX.XXBASO# (x103/µL) 5 XXX.XXNRBC% (/100WBC) 5 XX.XXRET# (x104/µL) 4 XX.XXRET% (%%) 4 XX.XXHFR (%) 4 XX.XXMFR (%) 4 XX.XXLFR (%) 4 XX.XXIRF (%) 4 XX.XXNRBC# (x103/µL) 5 XXX.XXRESERVED 65 “00~00”Closed/Manual 1 XRESERVED (Instrument ID) 22 “XE-2100XXX---X”ETX 1 (03H)Total 256

* A decimal point is not contained in the output data.

Table A-6-2: QC Data Format 1

QC No. Closed/Manual Corresponding File QC No. Closed/Manual Corresponding File1 0 No. 1 1 1 No. 212 0 No. 2 2 1 No. 223 0 No. 3 3 1 No. 234 0 No. 4 4 1 No. 245 0 No. 5 5 1 No. 256 0 No. 6 6 1 No. 267 0 No. 7 7 1 No. 278 0 No. 8 8 1 No. 289 0 No. 9 9 1 No. 29A 0 No. 10 A 1 No. 30B 0 No. 11 B 1 No. 31C 0 No. 12 C 1 No. 32D 0 No. 13 D 1 No. 33E 0 No. 14 E 1 No. 34F 0 No. 15 F 1 No. 35a 0 No. 16 a 1 No. 36b 0 No. 17 b 1 No. 37c 0 No. 18 c 1 No. 38d 0 No. 19 d 1 No. 39e 0 No. 20 e 1 No. 40M 0 XbarM M 1 None

(Note: In the Closed/Manual column, "0" indicates "Manual," and "1" indicates "Closed.")

Table A-7: File Corresponding to QC No.

Page 15: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-11

QC Data Format 2

Parameter No. of Chars. ExampleSTX 1 (02H)Text Distinction Code 1 1 "D"Text Distinction Code 2 1 "2"Sample Distinction Code 1 "C"Quality Control No. 1 XYear 2 XXMonth 2 XXDay 2 XXHour 2 XXMinute 2 XXInstrument Name 16 XX~XXWBC/BASO-X (CH) 4 XX.XXWBC/BASO-Y (CH) 4 XX.XXDIFF-X (CH) 4 XX.XXDIFF-Y (CH) 4 XX.XXRESERVED 5 "00000"NRBC-X (CH) 4 XXX.XNRBC-Y (CH) 4 XXX.XIMI# 4 XXXXIMI-DC (x10-1/fL) 4 XXXXIMI-RF (x10-1/fL) 4 XXXXRBC-O (x106/µL) 4 XX.XXPLT-O (x103/µL) 4 XXXXRBC-X (CH) 4 XXX.XRBC-Y (CH) 4 XXX.Xd-RBC (x10-1/%) 4 XXX.Xd-PLT (x10-1/%) 4 XXXXDw/X (x10-1/%) 4 XXX.XDw/Y (x10-1/%) 4 XXX.XRESERVED 128 “00~00”RESERVED RE(Instrument ID) 22 “XE-2100XXX---X”ETX 1 (03H)Total 255

*A decimal point is not contained in the output data.

Table A-8: QC Data Format 2

Page 16: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-12 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

QC No. Closed/Manual Corresponding File QC No. Closed/Manual Corresponding File1 0 No. 1 1 1 No. 212 0 No. 2 2 1 No. 223 0 No. 3 3 1 No. 234 0 No. 4 4 1 No. 245 0 No. 5 5 1 No. 256 0 No. 6 6 1 No. 267 0 No. 7 7 1 No. 278 0 No. 8 8 1 No. 289 0 No. 9 9 1 No. 29A 0 No. 10 A 1 No. 30B 0 No. 11 B 1 No. 31C 0 No. 12 C 1 No. 32D 0 No. 13 D 1 No. 33E 0 No. 14 E 1 No. 34F 0 No. 15 F 1 No. 35a 0 No. 16 a 1 No. 36b 0 No. 17 b 1 No. 37c 0 No. 18 c 1 No. 38d 0 No. 19 d 1 No. 39e 0 No. 20 e 1 No. 40M 0 XbarM M 1 None

(Note: In the Closed/Manual column, "0" indicates "Manual," and "1" indicates "Closed.")

Table A-9: File Corresponding to QC No.

Page 17: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-13

2. COMMUNICATION SPECIFICATIONS OF ANALYSISINFORMATION WITH HOST COMPUTER

2.1 Outline

XE-2100 has a function to load analysis information (analysis order information and patientinformation) from the host computer, and perform the analysis automatically according to theinformation.

1. Analysis Information Inquiry Method

XE-2100 has the following two inquiry methods.

(1) Real-time inquiry is performed just before an analysis, using the sample ID number, asa key word, read from the bar code label attached to a sample tube.

(2) Batch inquiry is performed before an analysis by specifying a rack number. The keywords in this case are the rack number and the tube position.

2. Sample Information Inquiry SettingKey: Sample ID Rack No./Tube Pos.Real-time Inquiry (Manual Mode): [Sample ID]Real-time Inquiry (Sampler Mode): [Key]

Page 18: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-14 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

2.2 Hardware

The bit serial voltage type, which conforms to the RS-232C interface, is used for host computeroutput of XE-2100. The serial interface port for the connection with the host computer is on therear panel of the IPU.

1. Connector• Connect to the host computer output connector on the rear panel of the IPU.• Use a 9-pin, D-SUB female connector.• Pitch of connector fixing screws adopts the size in inch.

2. Connector Signal

Pin No. Signal Name Signal Direction12 Receive Data (RxD) To XE-2100 from HOST3 Transmit Data (TxD) From XE-2100 to HOST4 Data Terminal Ready (DTR) From XE-2100 to HOST5 Signal Ground (SG)6 Data Set Ready (DSR) To XE-2100 from HOST7 Request to Send (RTS) From XE-2100 to HOST8 Clear to Send (CTS) To XE-2100 from HOST9

Table A-10: Connector Signal

3. Communication FormatThe format is start-stop a synchronization half duplex (asterisk indicates setting at shipment fromfactory), and adjustment can be made as follows.

Baud Rate 600, 1200, *2400, 4800, 9600, 14400 (BPS)Code *7-bit, 8-bitStop Bit 1-bit, *2-bitParity Bit None, *Even, OddClass Class A, *Class BInterval 0, 1, *2, 3, 5, 7, 10, 15 (seconds)

Table A-11: Communication Format Setting

4. Signal Level

Level Data Signal Control Signal+3V or Higher Logic "0," Start Bit ON-3V or Lower Logic "1," Stop Bit OFF

Table A-12: Signal Level

Page 19: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-15

2.3 Software

1. Communication Format

1) CodeASCII codes are used for all the output to the computer.

2) Text Format"STX" is sent at the beginning, and "ETX" is sent at the end of data.STX: 02H, ETX: 03H

Order of Transmission

STX

ETX

Figure A-3: Text Format

3) Transmission ProcedureThe transmission method is class B (fixed). Make sure to set it class B. If the class A is set,correct communication is not possible. The procedure to make inquiry on analysisinformation from XE-2100 to the host computer is as follows:• Analysis information inquiry text is sent from XE-2100.• The host computer returns NAK if an error occurs with the data reception, and returns

ACK when there is no error, and then send the analysis information corresponding to theinquiry.

• In the case of NAK, XE-2100 resends the analysis information inquiry text. In the case ofACK, after receiving the analysis information text, XE-2100 sends ACK when there is noerror with the received data, or sends NAK if there is an error.

• The host computer closes the communication for one set of analysis information when thereturned response is ACK, or resends the analysis parameter text if the response is NAK.

Sends NAK if there is an error.or ACK when there is no error.Sends sample information text for the inquired samples.(Resends if NAK.)(Resends up to 3 times.)

ACK: 06H, NAK: 15H

Sends sample information inquiry text.

(Resends if NAK.)(Resends up to 3 times.)

Sends NAK if there is an error.or ACK when there is no error.

[XE-2100] [Computer]

Figure A-4: Transmission Procedure

Page 20: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-16 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

4) Transmission ErrorIf a transmission error is detected, the transmission is interrupted and the error message isdisplayed on the IPU. The operator must recover the error. Cases of errors are as follows.• Control signal "DSR" is OFF.• After the data transmission, there is no response from the computer within 15 seconds.• "NAK" is sent back after transmitting three times.

5) Inquiry Timing• Real-time inquiry by sample ID number

At the time when a sample ID number is read from the bar code attached to the tube, theinquiry is performed. If XE-2100 already has the analysis information, the inquiry is notperformed. After sending out the analysis information inquiry text, wait while receivingACK and receiving the analysis information text, then blood is absorbed after that.

• Batch inquiry by rack number and tube positionAt the time when a rack number is specified by the IPU program of XE-2100, the inquiry ismade for the tube positions from 1 through 10 for the specified rack. There is not time limitto receive the analysis information text after sending the analysis information inquiry text.

6) Transmission IntervalThe transmission interval from receiving analysis information text from the host computerand sending out ACK until the next sending of analysis information inquiry text, can be setby the IPU program of XE-2100. This transmission interval is common with that of analysisdata.

7) Note on Analysis Data Automatic OutputWhen the analysis data is set to be output automatically by XE-2100, the communication ofanalysis information and the output of analysis data may be performed alternately. In thiscase when two different types of information are handled simultaneously, the response of thehost computer may become slower. Therefore, set the instrument to make the batch output,not the automatic output to avoid slow response of the host computer.

2. Text Format

There are text formats for "Analysis information inquiry" and "Analysis information" regardingthe communication of analysis information between XE-2100 and the host computer. And theseare identified by Text Distinction Code I.• The Text Distinction Code I of the "Analysis information inquiry" sent from XE-2100 to the

host computer is always "R."• The Text Distinction Code I of the "Analysis information" sent from the host computer to XE-

2100 is always "S."

Page 21: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-17

3. Analysis Information Inquiry Format

Parameter No. of Chars.STX 1 (02H)Text Distinction Code I 1 "R"Inquiry Mode 1 (X)RESERVED 3 (XXX)Inquiry Sample ID No. 15 (XX~XX)RESERVED 2 (XX)Rack No. 6 (XXXXXX)Tube Position 2 (XX)RESERVED 31 (XX~XX)ETX 1 (03H)Total 63

Table A-13: Analysis Information Inquiry Format

1) Output OrderThe upper items are sent first, and upper digits of data are sent first. Zero-suppression is notcarried out.

2) Inquiry ModeThe mode of inquiry is indicated."1": Real-time inquiry by sample ID number as the key word."2": Batch inquiry by rack No. and tube position as the key words.

3) Inquiry Sample ID NumberThis parameter becomes effective with the real-time inquiry by sample ID number as the keyword. It consists of 15-digit number, but may include hyphen "-" (2DH) between numeralsdepending on the usage.The hyphen "-" is included in 15 digits.

4) Rack No.This parameter becomes effective with the batch inquiry by rack No. and tube position as thekey words. This is the number assigned to a sample rack. It contains of 6-digit number.

5) Tube PositionThis parameter becomes effective with the batch inquiry by rack No. and tube position as thekey words. It consists of number from 1 to 10 for an analysis position on a sample rack.

Page 22: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-18 Sysmex XE-2100 IPU Operator’s Manual -- November 1999 N.A.

4. Analysis Information Format 1

Parameter No. of Chars.STX 1 (02H)Text Distinction Code 1 1 "S"Text Distinction Code 2 1 “1”Information Status 1 (X)Date Ordered 8 (XXXXXXXX)RESERVED 3 (XX~XX)Sample ID No. 15 (XX~XX)RESERVED 2 (XX)Rack No. 6 (XXXX)Tube Position 2 (XX)Inquiry Mode 1 (X)Patient ID No. 16 (XX~XX)Patient Name 40 (XX~XX)Sex 1 (X)Birthday 8 (XXXXXXXX)Doctor 20 (XX~XX)Ward 20 (XX~XX)Sample Comments 40 (XX~XX)RESERVED 18 (XX~XX)WBC 1 (X)RBC 1 (X)HGB 1 (X)HCT 1 (X)MCV 1 (X)MCH 1 (X)MCHC 1 (X)PLT 1 (X)LYMPH% 1 (X)MONO% 1 (X)NEUT% 1 (X)EO% 1 (X)BASO% 1 (X)LYMPH# 1 (X)MONO# 1 (X)NEUT# 1 (X)EO# 1 (X)BASO# 1 (X)RDW-CV 1 (X)RDW-SD 1 (X)RESERVED 1 "0"MPV 1 (X)RESERVED 1 "0"

(To continue to next page)

Table A-14-1: Analysis Information Format 1

Page 23: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- November 1999 N.A. A-19

(Continued from previous page)

RESERVED 2 "00"RET% 1 (X)RET# 1 (X)IRF 1 (X)RESERVED 1 "0"RESERVED 1 "0"RESERVED 1 "0"RESERVED 1 “0”RESERVED 1 "0"NRBC% 1 (X)NRBC# 1 (X)RESERVED 15 (XX~XX)EXT 1 (03H)Total 255

Table A-14-2: Analysis Information Format 1

5. Analysis Information Format 2

Parameter No. of Chars.STX 1 (02H)Text Distinction Code 1 1 "S"Text Distinction Code 2 1 “2”Information Status 1 (X)Date Ordered 8 (XXXXXXXX)RESERVED 3 (XX~XX)Sample ID No. 15 (XX~XX)RESERVED 2 (XX)Rack No. 6 (XXXX)Tube Position 2 (XX)Inquiry Mode 1 (X)Patient ID No. 16 (XX~XX)Patient Comments 100 (XX~XX)RESERVED 97 (XX~XX)EXT 1 (03H)Total 255

Table A-15: Analysis Information Format 2

1) Output OrderThe upper items are sent first, and upper digits of data are sent first. Zero-suppression is notcarried out.

Page 24: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-20 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

2) Information StatusThis parameter indicates if the inquired analysis information is registered. If the requiredsample is not registered, make sure to return "0" (Not registered) in the analysis informationtext."0": Not registered"1": Registered"2": Quality control

3) Date OrderedThis parameter indicates the requested date of analysis of the inquired sample."YYYYMMDD"

YYYY: Year, MM: Month, DD: Day

4) Sample ID NumberIn the case of real-time inquiry by sample ID number as the key word, this number becomesthe same with that in the inquiry text. In the case of batch inquiry by rack No. and tubeposition as the key words, the sample ID number corresponding to the specified rack No. andtube position will be assigned.It consists of 15-digit number or alphabet, but may include hyphen "-" (2DH) betweennumerals depending on the usage. The hyphen "-" is included in 15 digits.

5) Rack No.In the case of batch inquiry by rack No. and tube position as the key words, this numberbecomes the same with that in the inquiry text. In the case of real-time inquiry by sample IDnumber as the key word, no particular setting is used.This number is assigned to a sample rack, and consists of 6 digit number.

6) Tube PositionIn the case of batch inquiry by rack No. and tube position as the key words, this numberbecomes the same with that in the inquiry text. In the case of real-time inquiry by sample IDnumber as the key word, no particular setting is used.This is the analysis position of the inquired sample in the sample rack, and consists ofnumber from 1 to 10.

7) Inquiry ModeThe mode of inquiry is indicated."1": Real-time inquiry by sample ID number as the key word."2": Batch inquiry by rack No. and tube position as the key words.

8) Patient ID No.This parameter is the patient ID for the inquired sample, and is unique to a patient.It consists of 16-digit number or alphabet, but may include hyphen "-" (2DH) betweennumerals depending on the usage. The hyphen "-" is included in 16 digits.

9) Patient NameThis is the patient name for the inquired sample. Last name and first name (20 charactersmax. for each) can be entered in alphabet characters.

Page 25: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-21

10) SexThis is the sex of the patient."1": Male"2": Female

11) BirthdayThis is the birthday of the patient."YYYYMMDD"

YYYY: Year, MM: Month, DD: Day

12) DoctorThis is the name of the doctor in charge, and consists of up to 20 alphabets.

13) WardThis is the ward (medical section) in which the patient is staying, and consists of up to 20alphabets.

14) Sample CommentsThis is the comments for the inquired sample, and consists of up to 40 alphabets.

15) Patient CommentsThis is the comments of the patient for the inquired sample, and consists of up to 100alphabets.

16) Order InformationThis indicates the analysis order information for each analysis parameter."0": Not analyze"1": Analyze

Page 26: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-22 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

3. ID BAR CODE SPECIFICATIONS

By affixing the bar code label on the sample tube, a sample ID number can be automaticallyread. Information read from ID number can be also corrected by processing stored data. In usinga bar code, make sure it meets the bar code label specifications applicable to XE-2100 ID barcode reader.The specifications of the bar code label is described in this section.

1. Acceptable Bar Codes

The types of bar codes acceptable to the instrument and the check digit(s) are listed below.

1) Sample ID number

Type of Bar Code Check Digit No. of DigitsITF Not Used Max. 15 digits (Sample ID No.)

Modulus 10 Max. 15 digits (Sample ID No.)+ 1 digit (Check digit) = 16 digits MaxNW-7(*) Not Used Max. 15 digits (Sample ID No.)

Modulus 11W. Modulus 11 Max.15 digits (Sample ID No.)+ 1 digit (Check digit) = 16 digits MaxModulus 16

CODE 39 Not Used Max. 15 digits (Sample ID No.)Modulus 43 Max.15 digits (Sample ID No.)+ 1 digit (Check digit) = 16 digits Max

JAN-13 Modulus 10 12 digits (Sample ID No.) + 1 digit (Check digit) = 13 digitsJAN-8 Modulus 10 7 digits (Sample ID No.) + 1 digit (Check digit) = 8 digitsCODE 128 Modulus 103 Max.15 digits (Sample ID No.)+ 1 digit (Check digit) = 16 digits Max

Table A-16: Sample ID No. Bar Code

NOTE: • Do not use the bar code of Rack ID No. as that of Sample ID No.• For CODE 128, do not use the function characters.• *: As the Start/Stop code for NW-7, use one of the characters "A," "B," "C,"

"a," "b," or "c."

2) Rack ID No.

Type of Bar Code Check Digit No. of DigitsNW-7 Modulus 16 6 digits (Rack No.) + 1 digit (Check Digit) = 7 digitsCODE 39 Modulus 43 6 digits (Rack No.) + 1 digit (Check Digit) = 7 digits

Table A-17: Rack ID No. Bar Code

NOTE: • As the Start/Stop code, use either "D" or "d."

Page 27: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-23

3) Quality Control

Type of Bar Code Check Digit No. of DigitsCODE 128 (*1) Modulus 103 3 digits (Fixed character string "QC-") + 8 digits (Lot No.)

+ 1 digit (Check Digit) = 12 digitsNW-7 (*2) Not Used 4 digits - 13 digits (File No.) (*3)

Table A-18: Quality Control Bar Code

NOTE: • *1: The bar code of CODE 128 for quality control is a special code used forthe control blood of Sysmex.

• *2: As the Start/Stop code for NW-7, use either "C" or "c."• *3: The numerals that can be used are 1 to 9 and all the digits must be

identical.

2. Dimension of Bar Code Elements

Narrow Element 190 µ mWide Element 1.2 mmNarrow Element Gap between characters Wide Element

3. Narrow/Wide Ratio

For each character, the wide element to narrow element ratio must comply with the following:Narrow (MAX) : Wide (MIN) = 1 : 2.2 or moreNarrow (MIN) : Narrow (MAX) = 1 : 1.3 or lessWide (MIN) : Wide (MAX) = 1 : 1.4 or less

4. PCS (Print Contrast Signal)

Reflectivity at the space

Reflectivity at the space - Reflectivity at the black inked barPCS =

The measuring method conforms to JIS (Japanese Industrial Standards) X 0501, "5.3 OpticalCharacteristic of Bar Code Symbols."Standard: PCS value ≥ 0.45

5. Reflection Characteristics of the Label Surface

It is possible that a laminated label cannot be read.

Page 28: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-24 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

6. Irregularity and Roughness of Printing

When a bar element is magnified, the following may be observed.

MAX

MIN

Bar element

Figure A-5: Roughness of Printing

When the variation coefficient (S) in the width of a bar is defined:

MAX

MAX - MINS = X 100%

Then the variation coefficient (S) must be less than or equal to 20%.

7. Dimensions of Bar Code Label

Bar Code of Effective Length Space Space

Bar Height

Figure A-6: Dimensions of Bar Code Label

Space: 2.5 mm or more (Normally, at least 5 mm or both right and left.)Bar Code Effective Length: 48 mm or less (Optimum: 40 mm or less)Bar Height: 20 mm or more (Rack label height: 6 mm or more)

Page 29: Host manual sysmex-xe2100

TECHNICAL INFORMATION

Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999 A-25

8. Check Digit

To improve the reliability of an ID No. read, check digit(s) can be added.Taking the Sample ID No. of "258416" as an example, let us explain how to calculate the checkdigit for modulus 11 and weighted modulus 11.

1) Modulus 11

(1) Each digit is weighted. The weight corresponding to each digit is as follow.

Digit 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1Weight 6 5 4 3 2 1 10 9 8 7 6 5 4 3 2

(2) Add up the multiplied results as given below:S = 14 + 30 + 40 + 16 + 3 + 12 = 115

(3) When the S is divided by 11, calculate the remainder and obtain the complement of theremainder. This complement will be the check digit.115/11= 10 with remainder 511 - 5 = 6, thus the check digit is 6.However, all English symbols except the numerals of 0 - 9 are regarded as 0 in makingcalculation. Also, when S is divisible by 11 with remainder 0 and when calculation ofthe check digit results in 10, zero is entered as the check digit.

2) Weighted Modulus 11Weighted modulus 11 has two sets of weight. When the check digit is computed to 10 as aresult of applying the first weight set, the second weight set is applied. The result shouldalways be one of the 0 to 9 values. Calculation method is entirely the same as modulus 11except for difference in weighting.

(1) Weighing Each Digit

Weight: W12 W11 W10 W9 W8 W7 W6 W5 W4 W3 W2 W1First Set: 6 3 5 9 10 7 8 4 5 3 6 2Second Set: 5 8 6 2 10 4 3 7 6 8 5 9

2 5 8 4 1 6X X X X X X8 4 5 3 6 2

Weight 16 20 40 12 6 12

(2) Add up the multiplied results as given below:S = 16 + 20 + 40 + 12 + 6 + 12 = 106

Page 30: Host manual sysmex-xe2100

TECHNICAL INFORMATION

A-26 Sysmex XE-2100 IPU Operator’s Manual -- Revised November 1999

(3) When the S is divided by 11, calculate the remainder and obtain the complement of theremainder. This complement will be the check digit.106/11= 9 with remainder 711 - 7 = 4, thus the check digit is 4.However, all English symbols except the numerals of 0 - 9 are regarded as 0 in makingcalculation. Also, when S is divisible by 11 with remainder 0 and when calculation ofthe check digit results in 0, zero is entered as the check digit.

NOTE: • For Weighted Modulus 11, weight for the 13th, 14th and 15th digits areassumed 0.