Hepatic Ultrasonography

download Hepatic Ultrasonography

of 130

Transcript of Hepatic Ultrasonography

  • 7/30/2019 Hepatic Ultrasonography

    1/130

  • 7/30/2019 Hepatic Ultrasonography

    2/130

    Hepatic Elastography UsingUltrasound Waves

    Edited By

    Ioan Sporea and Roxana irli

    Department of Gastroenterology and HepatologyVictor Babe

    University of Medicine and Pharmacy Timioara

    Romania

  • 7/30/2019 Hepatic Ultrasonography

    3/130

    Bentham Science PublishersExecutive Suite Y - 2PO Box 7917, Saif Zone

    Sharjah, [email protected]

    Bentham Science PublishersP.O. Box 446Oak Park, IL 60301-0446

    [email protected]

    Bentham Science PublishersP.O. Box 2941400 AG Bussum

    THE [email protected]

    Please read this license agreement carefully before using this eBook. Your use of this

    eBook/chapter constitutes your agreement to the terms and conditions set forth in this License

    Agreement. Bentham Science Publishers agrees to grant the user of this eBook/chapter, a non-

    exclusive, nontransferable license to download and use this eBook/chapter under the following

    terms and conditions:

    All content available in this eBook is licensed under the terms of the Creative Commons Attribution

    non-commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits

    unrestricted, non-commercial use, distribution and reproduction in any medium, provided that the

    work is properly cited.

    Warranty Disclaimer: The publisher does not guarantee that the information in this publication

    is error-free, or warrants that it will meet the users requirements or that the operation of the

    publication will be uninterrupted or error-free. This publication is provided "as is" without

    warranty of any kind, either express or implied or statutory, including, without limitation,

    implied warranties of merchantability and fitness for a particular purpose. The entire risk as to

    the results and performance of this publication is assumed by the user. In no event will the

    publisher be liable for any damages, including, without limitation, incidental and consequential

    damages and damages for lost data or profits arising out of the use or inability to use the

    publication. The entire liability of the publisher shall be limited to the amount actually paid by

    the user for the eBook or eBook license agreement.

    Limitation of Liability: Under no circumstances shall Bentham Science Publishers, its staff,

    editors and authors, be liable for any special or consequential damages that result from the use

    of, or the inability to use, the materials in this site.

    eBook Product Disclaimer: No responsibility is assumed by Bentham Science Publishers, its

    staff or members of the editorial board for any injury and/or damage to persons or property as a

    matter of products liability, negligence or otherwise, or from any use or operation of any

    methods, products instruction, advertisements or ideas contained in the publication purchased or

    read by the user(s). Any dispute will be governed exclusively by the laws of the U.A.E. and willbe settled exclusively by the competent Court at the city of Dubai, U.A.E.

    You (the user) acknowledge that you have read this Agreement, and agree to be bound by its

    terms and conditions.

  • 7/30/2019 Hepatic Ultrasonography

    4/130

  • 7/30/2019 Hepatic Ultrasonography

    5/130

  • 7/30/2019 Hepatic Ultrasonography

    6/130

    ii

    several editorial boards (i.e.: Ultraschall in der Medizin, Medical

    Ultrasonography, Journal of Gastrointestinal and Liver Diseases). He is author

    and co-author of 195 original papers published in medical journals (42 of them in

    ISI journals and 96 PubMed publications), first author of 12 medical books, co-

    author of 14 medical books, first author of 16 educational medical CDs and DVDs

    (Ultrasound and Endoscopy). He coordinated or participated to numerous research

    projects.

    Special interest in: Contrast enhanced ultrasonography, Elastography, Ultrasound

    in Inflammatory Bowell Disease.

    ROXANA SIRLI

    Roxana Sirli is an Assistant Professor, PhD, in the Department of

    Gastroenterology and Hepatology of the Victor Babe University of Medicine

    and Pharmacy Timioara. She is a senior attendant in Internal Medicine, specialist

    in Gastroenterology, working in the Gastroenterology and Hepatology

    Department of the Victor Babe University of Medicine and Pharmacy

    Timioara. She is a level II specialist in general ultrasonography according to the

    multilevel classification of SRUMB. She is a member of the Board of Directors ofthe Romanian Society of Ultrasound in Medicine and Biology (SRUMB). She is a

    member of the WFUMB (World Federation of Ultrasound in Medicine and

    Biology) Center of Excellence Timioara, also a faculty member of the

  • 7/30/2019 Hepatic Ultrasonography

    7/130

    iii

    Ultrasound Learning Center of UMF Timioara. She participated in several

    courses and scientific sessions in Romania and abroad, mainly in gastroenterology

    and ultrasound. She is author and co-author of 90 original papers published in

    medical journals (32 of them in ISI journals and 58 PubMed publications), co-

    author of 14 medical books, co-author of 6 educational ultrasound CDs and

    DVDs. She participated in numerous research projects.

    ALINA POPESCU

    Alina Popescu is a Lecturer, PhD, in the Department of Gastroenterology and

    Hepatology of the Victor Babe University of Medicine and Pharmacy Timioara.

    She is a senior attendant in Internal Medicine, specialist in Gastroenterology,

    working in the Gastroenterology and Hepatology Department of the Victor Babe

    University of Medicine and Pharmacy Timioara. She is a level II specialist in

    general ultrasonography according to the multilevel classification of the Romanian

    Society for Ultrasound in Medicine and Biology (SRUMB) and she is a member of

    the Board of Directors of SRUMB. She is a member of the WFUMB (World

    Federation of Ultrasound in Medicine and Biology) Center of Excellence Timioara,

    also a faculty member of the Ultrasound Learning Center of UMF Timioara. She isa member of the flying faculty of the International School for Clinical Ultrasound

    ISCUS. She is author and co-author of several original papers published in medical

    journals, medical books and chapters, educational CDs and DVDs.

  • 7/30/2019 Hepatic Ultrasonography

    8/130

  • 7/30/2019 Hepatic Ultrasonography

    9/130

  • 7/30/2019 Hepatic Ultrasonography

    10/130

    vi

    FOREWORD

    Since the introduction of the grey scale B-mode scanners, the liver has been the

    organ with the most extensive and fruitful applications of ultrasonography in the

    abdomen. Starting from the 80s focal liver lesions became detectable even when

    small in size, targeted interventions were made possible with real-time guidance

    even at the bed-side and, slightly later, duplex Doppler ultrasound provided

    functional and not only morphological assessment of the liver vasculature and

    new exciting diagnosis were made possible. It should be acknowledged that the

    introduction of ultrasonography significantly contributed to the recognition of

    hepatology as an independent discipline. In the next 15 years refinements in

    ultrasound equipments were introduced by the industries, but no sustantial change

    in the diagnostic capabilities did really appear. This remained true until the early

    years 2000, which witnessed two revolutionary new ultrasound based techniques.

    One is real-time low acoustic pressure contrast enhanced ultrasound (CEUS),

    introduced into the market in 2002. This technique developed very rapidly and is

    now fully mature and applied in the daily practice worldwide with well

    established guidelines, such as those released by EFSUMB (European Federation

    of Societies for Ultrasound in Medicine and Biology). The second one is

    ultrasound elastography, which was first presented in the medical literature in

    2003. Ultrasound elastography provides a functional assessment of the liver,informing on tissue elasticity and thus on the disease stage. This information is

    obtained with greatest ease, non invasively and very rapidly at the bedside.

    Accordingly, transient elastography has been recently incorporated into

    international guidelines for the management of chronic viral hepatitis. It has also

    applications in other conditions involving the liver, beside chronic hepatitis.

    While contrast enhanced ultrasound underwent technical improvements, but is

    substantially one single modality, elastography is somehow different and various

    modalities are available, requiring different examination techniques and providing

    slightly different clinical information. Most of these modalities have been

    introduced only in the very last few years and their properties are still poorlyknown to clinical ultrasonographers. Therefore, the eBook by Prof. Ioan Sporea

    on liver elastography is very timely presented and greatly desired. In fact the

    ongoing spread of the technical possibility to perform liver elastography must be

  • 7/30/2019 Hepatic Ultrasonography

    11/130

    vii

    paralleled by adequate knowledge of the clinicals information that can be obtained

    by each of the different modalities. Worth to remind that beyond the self standing

    transient elastography equipment, nowadays several ultrasound scanners can be

    implemented with various elastographic techniques, either based on shear wave or

    strain imaging modalities.

    Reading the eBook will be an exciting time, with immediate applicability of the

    information into the daily clinical practice for anyone involved in the management

    of liver disease and the authors are to be commended for their efforts, based on

    long standing clinical and research expertise in this field.

    Fabio Piscaglia, MD PhD,

    University of Bologna

    Italy

    President EFSUMB

  • 7/30/2019 Hepatic Ultrasonography

    12/130

  • 7/30/2019 Hepatic Ultrasonography

    13/130

  • 7/30/2019 Hepatic Ultrasonography

    14/130

  • 7/30/2019 Hepatic Ultrasonography

    15/130

    Hepatic Elastography Using Ultrasound Waves, 2012, 3-24 3

    Ioan Sporea and Roxana irli (Eds)

    All rights reserved- 2012 Bentham Science Publishers

    CHAPTER 1

    Physics and Technical Information

    Ioan Lie*

    Applied Electronics Department, Electronics and Telecommunications Faculty,

    Politehnica University Timioara 2, Vasile Prvan Bv, 300223 Timioara

    Romania

    Abstract: US is defined as acoustic waves with higher frequencies than those that can

    be detected by the human ear, ranging from about 20 kHz to several hundred MHz.

    Medical US typically uses waves ranging from 1 to 15 MHz. A typical US transducer

    employs an array of piezoelectric elements to generate short duration, broadband pulses.The array size determines the imaging systems aperture. The same transducer also

    receives the backscattered signals which are then processed in order to obtain the US

    image of the explored region.Elasticity is the physical property of materials to return to

    their original shape after removing the force that caused the deformation. A

    complementary concept of elasticity is stiffness, which is a measure of the resistance

    opposed by an elastic material to deformation. Quantitative elastography is based on

    shear waves production, tracking and detection. Different elastography methods use

    different techniques for generating and tracking shear waves, but the stiffer the tissue is,

    the higher the shear wave velocity is. Also liver stiffness increases with the severity of

    fibrosis, since scaring tissue is less elastic than the normal liver parenchyma.

    Keywords: Ultrasound waves, elasticity, stiffness, shear waves, liver fibrosis.

    1. ULTRASOUND

    The use of ultrasound (US) in medical practice has found a solid niche among the

    various methods for body imaging. US is defined as acoustic waves with higher

    frequencies than those that can be detected by the human ear, ranging from about

    20 kHz to several hundred MHz [1]. Medical US typically uses only the portion of

    the US spectrum ranging from 1 MHz to 10 MHz, due to the tradeoff between

    frequency and penetration depth. US waves are generated by small acoustic

    transducers, which are electrically driven and typically placed on the skin. The

    *Address correspondence to Ioan Lie: Applied Electronics Department, Electronics andTelecommunications Faculty, Politehnica University Timioara 2, Vasile Prvan Bv, 300223 TimioaraRomania; E-mail: [email protected]

  • 7/30/2019 Hepatic Ultrasonography

    16/130

    4 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    waves propagate into the body tissue, where a portion is reflected from the myriad

    interfaces between tissues with different acoustic properties [1].

    The most commonly used modality in medical US is B-mode imaging, where an

    ultrasound transducer is placed against the skin directly over the region of interest

    (ROI). A typical US transducer employs an array of piezoelectric elements to

    generate short duration, broadband pulses (with a center frequency of about 3-15

    MHz). The array size determines the imaging systems aperture. The same

    transducer also receives the backscattered signals. The transmission signals

    passing to and the received signals passing from the array elements can be

    individually delayed in time, defining a phased array. Phased arrays are used to

    electronically steer and focus the sequence of acoustic pulses through the targetvolume which is known as beam forming. Processing these echo signals routinely

    begins at the individual channel (element) level to produce A-lines (A-mode/ one

    dimensional wave equation of sound energy reflected from the target). The

    general formation of B-mode sequences (Fig. 1) commences with Radio

    Frequency (RF) demodulation or envelope detection storing, resulting A-modes in

    a 2D image matrix, followed by attenuation correction using time gain

    compensation (TGC) or swept and lateral gains, to increase signal amplification

    from increasing depths. Next scan conversion (an 8 bit digitization) allows the B-

    mode to be displayed with a defined resolution (known as a B-scan), and finally

    logarithmic compression is used to adjust the large echo dynamic range (60-100dB). The B-scan sequences captured and analyzed are those processed and

    displayed by the US machine, with a uniform dynamic range intensities ranging

    from 0 to 255 [2].

    Generally, US image analysis is complex, due to the numerous tissue interfaces

    and varying structure of biological tissues causing echogenicity, which is

    described in terms of a speckle formation. A speckle is a structured noise from a

    medium containing many scatterers. Speckle appearance is dependent on the

    bandwidth, frequency and manufacturer of the employed transducer, in addition to

    the geometry and sub-wavelength structure of the tissue. Echographic speckle

    texture of the imaged tissue is mainly due to intensity scattering; implying

    structures are smaller than the sampling volume (a product of spatial pulse length

    and beam cross section). Upon visual inspection, a speckle consists of a relatively

  • 7/30/2019 Hepatic Ultrasonography

    17/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 5high grey level intensity, qualitatively ranging from a hyperechoic (bright) to a

    hypoechoic (dark) domain. Scatter occurs when small imperfections (scatterers) in

    the target cause seemingly random reflections and refractions of the sound wave.

    The textures created do not correspond to the underlying structure, but the

    intensity reflects the local echogenity of the underlying scatterers. Scatterers

    account for a decrease in image quality, causing blurring and decreased intensity

    at impedance boundaries, while within the medium they create speckling. The

    signal statistics depend on the density of scatterers, with a large number of

    randomly located scatterers following a Rayleigh distribution [1].

    Figure 1: The processes used to generate a B-scan. B-scans are composed of a set of axial RF

    signals representing the response magnitude from a pulse generator using a linear array transducer.Since the response magnitude delays exponentially with depth, it is log-amplified prior to

    quantization and display [1].

    Standard medical practice of soft tissue palpation is based on the qualitative

    assessment of stiffness at low frequencies. It is generally known that pathological

    changes are correlated with changes in tissue stiffness. In many cases, despite the

    difference in stiffness, due to the small size of pathological lesions and/or depth to

    which they are located in the body, their detection by palpation is impossible.

    Generally, the lesion may or may not possess echogenic properties detectable with

    US. For example, breast or prostate tumors may be invisible or barely visible in

  • 7/30/2019 Hepatic Ultrasonography

    18/130

    6 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    standard US examination, although they are much more rigid than the tissues they

    are embedded into. In diffuse diseases such as liver cirrhosis, a significant

    increase in tissue stiffness is characteristic, but it may occur normally in a

    conventional US examination. Because tissue echogenity and stiffness are

    generally uncorrelated, it is expected that mapping tissue stiffness or elasticity,

    should provide new information on pathological tissues structure.

    2. PHYSICAL FUNDAMENTALS OF ELASTOGRAPHY

    Elasticity is the physical property of materials to return to their original shape

    after removing the force that caused the deformation. For small deformations,

    most materials show linear elasticity, i.e. a linear dependence between stress

    (force per unit area) and relative deformation (relative change). This dependence

    is known as Hooke's law. A complementary concept of elasticity is stiffness,

    which is a measure of the resistance opposed by an elastic material to

    deformation.

    The elasticity modulus describes mathematical, elastic deformation tendency of an

    object or material. The elasticity modulus of a material is defined as a slope of the

    curve describing the dependence between mechanical stress and deformation,

    considering the elastic deformation region of the curve. As the material is more

    rigid, it will have a higher modulus. Depending on how the mechanical stress is

    applied and how the deformation is measured, several types of elasticity modules

    are defined. The most important are:

    - Young's modulus (E) - this describes the deformation tendency of anobject following a certain axis, if the forces applied along the axis

    have an opposite orientation.

    - Shear modulus (G) - describes an object's tendency to change shapeand keep its volume, when mechanical stress is achieved by opposing

    forces placed in parallel planes.

    - The bulk modulus (K) - describes volumetric elasticity or an objectstendency to deform in all directions, when it supports mechanical

    stress in all directions. It is defined as the ratio between the force per

  • 7/30/2019 Hepatic Ultrasonography

    19/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 7unit volume and the volumetric deformation. Inverse of the bulk

    modulus is compressibility. The bulk modulus can be seen as a three-

    dimensional extension of Young's modulus.

    Poisson's coefficient is often used for the characterization of inhomogeneous

    isotropic media. It is defined as the ratio between transverse contraction per unit

    breadth and longitudinal extension per unit length. Lame's parameters are also

    used in linear elasticity theory. They are a parameterization of elasticity modules

    for homogeneous isotropic environments.

    Lam's first parameter denoted by , expresses the relationship between the bulk

    modulus and the shear modulus. The second parameter of Lam, noted

    (formerly G) is the shear modulus.

    The relationship between the Youngs modulus E, the Poisson coefficient and

    the Lam parameters and , is given by:

    3 2

    2 ( )

    E

    (1)

    The elasticity modulus should not be confused with stiffness. The elasticity

    modulus is a property of the material constituting a certain structure. Stiffness is a

    property of the structure and depends on the material, on its shape and boundary

    condition.

    For biological tissues, consisting mainly of water, compression module (several

    gigaPascals) is much higher than the shear modulus (several kiloPascals) [3]. This

    difference is explained by the fact that the volume change associated with

    compression requires a much greater force than that required for the shear

    deformation, which happens by changing shape at constant volume. The condition

    >> leads to a value of Poisson ratio 0.5, which characterizes the quasi-

    incompressible medium. In these conditions a simple relationship between

    longitudinal and shear modules is established.

  • 7/30/2019 Hepatic Ultrasonography

    20/130

    8 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    E = 3G (2)

    One way of assessing tissue elasticity is based on measuring the propagationvelocity of waves through the tissue. Propagation speed for any type of wave

    depends on the properties of the environment in which they propagate. For

    acoustic waves, the propagation speed depends on the elastic and inertial

    properties. Physical entities associated with these properties are the density ()

    and elasticity modulus. When applying a compressive mechanical stress,

    longitudinal or volumetric waves will propagate through the material, whose

    propagation direction coincides with the mechanical stress direction. Propagation

    velocity of longitudinal waves is given by the following equation:

    L

    KV

    (3a)

    When the material is subjected to shear forces, shear waves will propagate

    through it, which will produce material deformation perpendicular to the forces

    direction. Shear waves propagate at a speed given by the equation:

    S

    GV

    (3b)

    Because the elasticity modules values are significantly different (K = 2.3 GPa

    and G = 0.5-100 kPa) [4], the propagation speeds for the longitudinal waves and

    shear waves are significantly different: VL = 1400-1700 ms-1

    and vs. = 0.5 -10

    ms1.

    The shear modulus in tissue can be deduced from the shear wave velocity, Vs, and

    the mass density, :

    2

    2

    3

    S

    S

    V

    E V

    (4)

    In the hypothesis that soft tissue density is approximately constant (1000 kg/m3),

    the value of elasticity modulus is obtained by measuring the shear wave speed.

  • 7/30/2019 Hepatic Ultrasonography

    21/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 9The relationship (4) is the basis for developing methods for the quantitative

    assessment of elasticity. One of the methods used for measuring shear wave speed

    exploits the big difference between shear wave speed and longitudinal waves

    speed. The shear wave propagation in the region of interest is followed using

    longitudinal ultrasonic beams.

    Qualitative and quantitative description of a medium elasticity can be done in two

    ways:

    - By assessing the relative displacement caused by static or dynamicdeformation, or

    - By measuring the shear waves propagation velocity and indirectdetermination of elasticity modulus.

    Methods in the first category are implemented by qualitative techniques, which

    estimate a deformation rate, which indirectly characterize environmental stiffness.

    Quantitative Evaluation of environmental elasticity can be obtained by measuring

    the shear waves propagation speed and by a simple calculation determining the

    elasticity modulus. Corresponding to these two approachesstrainelastography or

    qualitative elastography and shear wave elastography or quantitative

    elastography were developed [4].

    3. BACKGROUND OF QUALITATIVE (STRAIN) ELASTOGRAPHY

    Consider a system with three springs with the same length without any application

    of force (Fig. 2). Spring constant is defined as the force necessary to stretch (or

    compress) a spring with a one unit length. In the considered system, the springs

    have different spring constants; the spring in the middle has a higher spring

    constant (is stiffer) as compared to the other two springs which have a lower

    spring constant (are softer) than the one in the middle. On application of equal

    forces to the springs, the less rigid spring will yield more displacement ascompared to the rigid one. The rigid spring is mechanically less elastic; thereby

    producing less displacement vis--vis the less rigid spring, which deforms more

    due to the same force [5].

  • 7/30/2019 Hepatic Ultrasonography

    22/130

  • 7/30/2019 Hepatic Ultrasonography

    23/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 11of the pre-compression and post-compression signals are compared by cross correlation. While the

    early windowed segments exhibit virtually no delay, a finite delay (designated del (t)) is detected

    between the later segments [8].

    When an elastic medium is compressed with a constant, axial oriented pressure,

    all points of the environment support a longitudinal deformation, whose main

    component is oriented on the axis of compression. If one or more tissue

    constituent elements have a different stiffness than the others, their deformation

    will be different (lower if the element is stiffer). Longitudinal deformation is

    estimated by analyzing the ultrasonic signals obtained with conventional

    equipment in the following sequence [6]:

    - The region of interest is scanned and the set of appropriate radio-frequency echoes is digitized and stored.

    - A tissue compression force is applied to produce small linear elasticdeformation into the tissue. The ultrasonic transducer or a dedicated

    compressor is used.

    - The region of interest is scanned once again and a new set of echosignals is acquired.

    Pairs of signals corresponding to the same directions of scanning are subdivided

    into small time windows and then compared using cross-correlation techniques.

    The windows are translated in small overlapping steps along the temporal axis of

    the echo line, and the calculation is repeated for all depths. For each direction and

    for each focal point in the direction considered, the differences between U.S.

    wave propagation times are determined in two situations. Since the compressive

    stress amplitude is small, deformation and thus differences in propagation times

    will also be reduced.

    4. THE STRESS EXCITATION METHODS

    Evaluation of tissue elasticity requires its excitation. Excitation methods can beclassified, according to their temporal characteristics, into static methods and

    dynamic methods. Static methods consist of applying a low value compressive

    force, constantly and uniformly distributed. Induced displacements are measured

  • 7/30/2019 Hepatic Ultrasonography

    24/130

  • 7/30/2019 Hepatic Ultrasonography

    25/130

  • 7/30/2019 Hepatic Ultrasonography

    26/130

    14 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    conventional systems. To solve this limitation, elastography dedicated hardware

    architectures have been designed.

    Depending on how shear waves are generated, three types of US elastography

    systems have been implemented.

    6. INDUCTION OF SHEAR WAVES USING AN EXTERNAL ACTUATOR

    TRANSIENT ELASTOGRAPHY

    This method uses an external actuator to produce low-frequency vibrations with

    frequencies in the 50-500 Hz range [9, 10]. The solution used in the "FibroScan"

    commercialized by Echosens, France, combines the actuator and the ultrasonic

    transducer in the same probe [4, 11-15]. Induced shear waves propagate throughthe tissue and produce its elastic deformation. Displacement is reflected in the

    variation of the acquired echo signals. The ultrasonic transducer is used in pulse-

    echo mode to measure displacements induced into the medium by the propagation

    of low frequency shear waves. Both longitudinal and shear waves are generated

    by the same probe and the ultrasonic beam is focused by the actuator axis. The

    assumption of homogeneity and symmetry considerations shows that

    displacement on the transducer axis is purely longitudinal. Diffraction effects

    from the transducer result in a longitudinally polarized shear wave on the axis of

    symmetry. The ultrasonic beam tracks its propagation (Fig. 4) [16].

    By cross-correlating successive lines the tissue deformation is determined. The

    system originally developed is based on single direction data acquisition and

    therefore does not provide a conventional B-mode real time image. Such an image

    is useful to guide the operator in positioning the transducer and choosing the place

    where stiffness is measured.

    Two dimensional representations are obtained when displacements induced by the

    shear wave are measured using cross-correlation of successive high frame rate

    ultrasound lines. From the recorded displacements a strain map is computed. The

    shear wave speed is calculated based on the slope of the wave front visualized on

    the strain map.

  • 7/30/2019 Hepatic Ultrasonography

    27/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 15

    Figure 4: The low frequency shear wave (blue) and the ultrasound beams (red) are generated by

    the same piston-like transducer. Under the assumption of homogeneity, the symmetryconsiderations impose that the displacements on the axis of the transducer be purely longitudinal

    (white arrow).

    7. INDUCTION OF SHEAR WAVES USING ACOUSTIC RADIATION

    FORCE ARFI ELASTOGRAPHY

    Acoustic radiation force is a phenomenon associated with the propagation of

    acoustic waves in attenuating media [17, 18]. Attenuation includes both the

    scattering and absorption of the acoustic wave. Attenuation is a frequency dependent

    phenomenon, and in soft tissues it is dominated by absorption. With increasing

    acoustic frequencies, the tissue does not respond fast enough to the transitions

    between positive and negative pressures, thus its motion becomes out of phase withthe acoustic wave, and energy is deposited into the tissue. This energy results in a

    momentum transfer in the direction of wave propagation and tissue heating. The

    momentum transfer generates a force that causes tissue displacement, the time scale

    of this response being much slower than that of ultrasonic wave propagation. This

    interaction of sound with tissue can be used to derive additional information about

    the tissue, beyond what is normally provided in an ultrasonic image. The magnitude,

    location, spatial extent, and duration of acoustic radiation force can be controlled to

    interrogate the mechanical properties of the tissue.

    The radiation force method causes tissue displacement centered on the focal region.These displacements propagate through the tissue in the form of shear waves and the

    US system is used to monitor the shear waves' propagation. This technique was

    proposed by Sarvazyan [4] and has been adopted by several groups [19, 20].

  • 7/30/2019 Hepatic Ultrasonography

    28/130

    16 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    The Siemens systems, Acuson S2000, implement both the strain and the shear

    wave elastography based on acoustic radiation force [21].

    Principle of Acoustic Radiation Force Impulse

    ARFI imaging involves transmission of an initial ultrasonic pulse at diagnostic

    intensity levels, to obtain a baseline signal for later comparison. A short duration,

    high-intensity acoustic "pushing pulse" is then transmitted by the same transducer,

    followed by a series of diagnostic intensity pulses, which are used to track the

    displacement of the tissue caused by the pushing pulse [17, 22, 23]. The tissue

    response to the radiation force is observed using conventional B-mode imaging

    pulses, and it is possible to display the quantitative shear-wave velocity (Vs; m/s)

    of ARFI displacement. This velocity (m/s) is proportional to the square root oftissue elasticity. Because the shear wave velocity depends on tissue stiffness, it is

    possible to apply ARFI technology to elastography. This technology was named

    Virtual Touch Tissue Quantification by SIEMENS.

    The applications for tissue stiffness assessment using investigative techniques

    based on US provide quite different information as compared to conventional US

    exam. For "Virtual Touch" application software [21], the data acquisition is

    performed in three stages.

    The first step is to obtain a reference B-mode image of the region of interest byconventional US. In the second stage the tissue is disturbed using a short acoustic

    pulse of hundreds of microseconds, which propagates through the tissue. As a

    result of energy transfer from the acoustic pulse to the tissue, it undergoes a

    deformation process dependent on its specific rigidity. Quantitative displacement

    size is tens of microns. Soft tissues, being elastic, will deform more than rigid

    tissue whose elasticity is much lower. The deformation associated with high

    intensity ultrasonic pulse propagation is followed by a process of relaxation after

    which the tissue returns to its original configuration.

    In the final phase, the region is scanned with a normal intensity (diagnostic) USbeam and a new B-mode image is acquired. By comparing it with the reference

    image, displacements occurring in different areas can be calculated. Therefore this

    technique uses different intensity ultrasonic waves to compress tissue and to

  • 7/30/2019 Hepatic Ultrasonography

    29/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 17observe their dynamic behavior due to acoustic radiation force action.

    Commercial systems have implemented acoustic intensity adjustment

    mechanisms, such as power peaks, to be controlled with conventional imaging

    methods. Simultaneously, data processing algorithms allow higher resolution and

    the system hardware has been refined for increased sensitivity to ultrasonic signal

    reception. To determine the delay between two disturbing pulses, ROI size and

    depth are taken into consideration.

    ARFI Elastography Qualitative Approach

    The application software "Virtual Touch Tissue Imaging" made by Siemens [1]

    provides quality map data of relative stiffness of tissue in a ROI (elastogram). The

    information is calculated by the examining of relative displacements ofelementary formations of tissue, arising from the acoustic pulse disturbing action.

    On the elastogram, the elasticity is associated with image brightness. Nestled

    beside a conventional ultrasound B-mode image and an elastogram regions of

    tissue with different borders can be highlighted. This is explained by the fact that

    the mechanisms for determining the contrast in tissue are completely different in

    the two methods.

    By combining lines resulting from successive evaluation mode A, on the

    directions that describe the ROI, the software application synthesizes an image.

    The procedure begins with the line positioned at one end of the ROI (left or right).A signal is obtained which describes, conventionally (mode A), the tissue in that

    direction when it is at rest. Next application of disturbing impulse focused in this

    direction will lead to displacement of tissue. Using conventional ultrasonic beams

    focused on the direction, it acquires signals describing the state of the deformation

    of tissue (Fig. 4). The two signals are compared using cross-correlation algorithm

    and determine differences in tissue position in the relaxed and compressed state,

    along the line considered. Differences calculated for each location relative to the

    maximum, considered as reference, are a measure of tissue elastic properties

    reported to tissue positioned in the location of reference. The process is repeated

    for each line of the ROI, as in a conventional scanning B. Finally the entire ROI

    calculated displacements are converted into an image format (elastogram) which

    shows the relative hardness of the tissue.

  • 7/30/2019 Hepatic Ultrasonography

    30/130

    18 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    Figure 5: Virtual Touch Tissue Imaging utilizes acoustic push pulses (orange) and tracking beams

    (green arrow), sequenced across a user-defined region of interest, to generate an elastogram

    depicting the relative stiffness of tissue from [21].

    ARFI Elastography Quantitative Approach

    ARFI technology allows a quantitative assessment of tissue elasticity based on

    shear wave velocity measurement. An appropriate application is "Virtual Touch

    Tissue Imaging" made by Siemens [21].

    According to the equation (4) shear wave velocity is directly proportional to the

    square modulus of elasticity. Therefore, by measuring the shear wave velocity, we

    obtain a direct characterization of the elastic properties of the tissue. Shear waves

    are generated and propagate perpendicular to the disturbing pulse. Unlike

    longitudinal ultrasonic waves used in conventional investigation, shear waves do

    not interact with the transducer. They are attenuated more than 10,000 times faster

    than conventional waves and therefore require a more sensitive measurement.

    Displacements generated by the shear wave propagation through tissue can be

    detected using ultrasonic beams which scan the ROI. Shear wave velocity arises

    from the determination of the shear wave front position and its correlation with

    the time elapsed between consecutive measurements (Fig. 6).

  • 7/30/2019 Hepatic Ultrasonography

    31/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 19A previously investigated region is identified by locating the ROI on a

    conventional ultrasound image. Then a focused acoustic pulse in this region is

    applied that will induce shear waves that will propagate through the ROI.

    Tracking beams adjacent to the excitation path are sensitive to wavelengths much

    smaller than the wavelength of sound. These are transmitted continuously until the

    detection of the shear wave front. Locating position of peaks at different points in

    time ensure accuracy and reproducibility of measurement results (Fig. 5).

    Figure 6: Virtual Touch Tissue Quantification utilizes an acoustic push pulse (orange) to generate

    shear waves (blue) through a user-placed region of interest. When detection pulses (green arrow)interact with a passing shear wave, they reveal the waves location at a specific time, allowing

    calculation of the shear wave speed. This numerical value is related to the stiffness of the tissue

    within the region of interest from [21].

    8. SHEAR WAVE IMAGING

    Shear wave imaging uses the same principles as the ones presented above. Shear

    waves are generated using a pushing pulse and A-line correlation techniques areused to track them through the tissues. This technique has been developed by a

    group led by Fink [20] and has been implemented commercially (Supersonic

    Imagine, France) [24, 25].

  • 7/30/2019 Hepatic Ultrasonography

    32/130

    20 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    Shear Wave Initiation

    Shear waves induced in the region of interest must be ample enough so that theirpropagation can be detected by focused beams. Initially, single pulses were used

    to generate shear waves. Currently, available commercial systems use several

    pulses, focused at different depths [20]. The cumulative effect of these pulses is

    reflected in the increasing amplitude of shear waves, and in the expansion of the

    region in which they can be tracked. This expands the area that can provide data

    about shear waves and thus about the environment stiffness. Excitation pulses

    form an excitation beam. Rapid change of beam focus depth is equivalent to

    moving high intensity excitation sources through the tissue. If the source moves

    with a higher speed than that of the generated shear wave, it is said that it moves

    with supersonic speed - hence the term supersonic imaging. The shear waves frommultiple sources combine and propagate in the shape of a cone, called a "Mach-

    cone" (Fig. 7).

    Figure 7: Generation of the supersonic shear source: the source is sequentially moved along the

    beam axis, creating two plane- and intense-shear waves [20].

    Shear Wave Detection

    To obtain a quantitative elasticity map of the medium, it is necessary to image the

    propagation of the shear-wave and to measure its velocity. As the shear wavestypically propagate at a few meters per second, a frame rate of several kilohertz is

    needed. This is not possible using conventional US scanners (they typically reach

    a 50-Hz frame rate).

  • 7/30/2019 Hepatic Ultrasonography

    33/130

    Elastography: Physics Hepatic Elastography Using Ultrasound Waves 21So the use of an ultrafast, ultrasonic scanner is needed, able to remotely generate

    the mechanical shear wave, by focusing US at a given location, and image the

    medium during the wave propagation at a very high-frame rate (up to 6000

    images/s) (Fig. 8). The ultrafast frame rate is achieved by reducing the emitting

    mode to a single, plane-wave insonation. This technique allows the acquisition of

    echographic images at a pulse repetition that can reach 6000 Hz.

    Figure 8: Stages necessary to image the propagation of the shear-wave and to measure its velocity[20].

    An ultrafast scanner is used, fully programmable, with a multichannel system made

    of 128 channels, connected to the transducer. All backscattered radio frequency (RF)

    echoes are stored in the memory of each channel and are transferred to a computer

    after acquisition. The beam forming process is done only in the receive mode during

    a post acquisition process. For each elementary transmit-receive sequence, a number

    of parameters can be fixed on each channel independently; to create focalized or flat

    transmits. The delays before and after emission are included, also the pointer

    addresses of transmit and receive signals [20].

    Generation of Radiation Force: To generate the radiation force, the ultrafast

    scanner is used to create an ultrasound-focused beam at a chosen location. The

  • 7/30/2019 Hepatic Ultrasonography

    34/130

    22 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    typical US pulse is made of 400 oscillations at 4.3 MHz. This corresponds to a

    pushing time of 100 s.

    Acquisition Sequence: A first plane-wave insonation is performed to realize a

    reference echographic image of the medium. The pushing sequence is then

    realized by focusing the US beam at a chosen location. Just after the generation of

    the pushing beam, the scanner begins an ultrafast imaging sequence by sending

    plane-wave insonations at a high-frame rate, in order to catch the shear wave

    created by the push.

    Signal Processing: The RF data stored in the scanner memory are transferred to

    the computer. A classical beam forming process then is applied to the data to

    compute the set of echo images. All the images acquired after the push are then

    correlated with the reference echo image using a 1-D correlation algorithm. The

    results are a set of images giving the displacement induced by the shear wave at

    each sample time.

    The final data may be displayed in units of shear wave velocity (m.s-1

    ) or converted

    into units of Youngs modulus (kPa) using the equation (4). Note that the equation

    (4) requires knowledge of the tissue density. Information on how manufacturers

    account for tissue density is not readily available. One possibility is that

    manufacturers simply assume a value for the density, possibly an average value.

    In practice shear wave images demonstrate considerable variability, with values

    affected by the presence of boundaries and by blood vessels [20]. Improved

    understanding of shear waves propagation through biological tissues may result

    in new beam-forming regimes and new signal processing algorithms, which

    improve image quality and reduce image variability.

    CONFLICT OF INTEREST

    The author(s) confirm that this chapter content has no conflict of interest.

    ACKNOWLEDGEMENT

    Declared none.

  • 7/30/2019 Hepatic Ultrasonography

    35/130

  • 7/30/2019 Hepatic Ultrasonography

    36/130

    24 Hepatic Elastography Using Ultrasound Waves Ioan Lie

    [22] Palmeri ML, Frinkley KD, Zhai L, et al. Acoustic radiation force impulse (ARFI) imagingof the gastrointestinal tract. Ultrason Imag 2005; 27: 7588.

    [23]

    Dahl JJ, Pinton GF, Palmeri ML, et al. A parallel tracking method for acoustic radiationforce impulse imaging. IEEE Trans Ultrason Ferroelectr Freq Control2007; 54: 301312.

    [24] 24 Tanter M, Bercoff J, Athanasiou A, et al. Quantitative assessment of breast lesionviscoelasticity: Initial clinical results using supersonic shearimaging. Ultrasound Med Biol

    2008; 34: 13731386.

    [25] Muller M, Gennisson JL, Deffieux T, et al. Quantitative viscoelasticity mapping of humanliver using supersonic shear imaging: preliminary in vivo feasability study. Ultrasound Med

    Biol 2009; 35: 219229.

  • 7/30/2019 Hepatic Ultrasonography

    37/130

    Hepatic Elastography Using Ultrasound Waves, 2012, 25-51 25

    Ioan Sporea and Roxana irli (Eds)

    All rights reserved- 2012 Bentham Science Publishers

    CHAPTER 2

    Transient Elastography (TE)

    Ioan Sporea and Roxana irli*

    Department of Gastroenterology and Hepatology, Victor Babe University of

    Medicine and Pharmacy, 10, Iosif Bulbuca Bv, 300736, Timioara, Romania

    Abstract: Transient Elastography (TE) is the first ultrasound-based method for fibrosis

    assessment, developed by Echosens (France). In order to obtain reliable liver stiffness

    (LS) measurements by means of TE, the manufacturer recommends that at least 10 valid

    shots should be obtained. They should have a success rate (SR: the ratio of valid shots

    to the total number of shots) of at least 60% and an interquartile range (IQR, thedifference between the 75th percentile and the 25th percentile, essentially the range of the

    middle 50% of the data) less than 30% of the median LS value. TE fails if no valid shots

    can be obtained, and is unreliable if fewer than 10 valid shots are obtained. TE failure is

    correlated with the body mass index, increasing in obese patients. Also, unreliable

    results are obtained during aminotransferases flares that can lead to an overestimation of

    fibrosis. The LS upper limit in healthy subjects was estimated to be 5.3 kPa. Several

    meta-analyses assessed LS measurements by TE as a predictor of fibrosis, cut-offs for

    F2 ranging from 7.2-7.6 kPa and for F=4 from 12.5-17.3 kPa, according to the

    etiology of chronic liver disease. Several studies have been published regarding the

    value of TE for predicting the occurrence of cirrhosis complications. The AUROCs for

    predicting clinically significant portal hypertension were 0.945 - 0.99, for cut-off values

    between 13.6 - 21 kPa, while for predicting esophageal bleeding the best cut-offs ranged

    between 50.7 62.7kPa, with AUROCs 0.73-0.75.

    Keywords: Transient elastography, liver stiffness, liver fibrosis, cirrhosis,

    esophageal varices.

    1. TE TECHNIQUE

    Transient Elastography (TE) is an ultrasound-based method, developed by

    Echosens (France), initiating from the principles of Hookes law, which

    characterizes a materials strain response to external stress [1]. A FibroScan

    device is used (Fig. 1), whose ultrasound transducer probe (Fig. 2), mounted on

    *Address correspondence to Roxana irli: Department of Gastroenterology and Hepatology, VictorBabe University of Medicine and Pharmacy, 10, Iosif Bulbuca Bv, 300736, Timioara, Romania;E-mail: [email protected]

  • 7/30/2019 Hepatic Ultrasonography

    38/130

    26 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    the axis of a vibrator, transmits low-frequency vibrations from the right intercostal

    space which creates an elastic shear wave that propagates into the liver. A pulse-

    echo ultrasound acquisition is then used to detect wave propagation velocity,

    which is proportional to tissue stiffness; faster wave progression occurs through

    stiffer material. LS measurement is then performed and measured in kiloPascals

    (kPa) (values between 2.5kPa and 75 kPa are expected).

    Figure 1: The FibroScan device.

    Figure 2: Pediatric (S), standard (M) and obese (XL) FibroScan probes.

    Using TE, liver stiffness measurements (LSMs) are performed in the right liver

    lobe through the intercostal spaces, while the patient lies in a dorsal decubitus

    position with the right arm in maximal abduction. The tip of the transducer is

    covered with coupling gel and placed on the skin between the ribs, aimed at the

  • 7/30/2019 Hepatic Ultrasonography

    39/130

    TE Hepatic Elastography Using Ultrasound Waves 27right liver lobe. The operator, assisted by ultrasound A-mode images provided by

    the system, locates a portion of the liver at least 6 cm thick and free of large

    vascular structures. Once the area of measurement had been located, the operator

    presses the probe button to begin an acquisition. Acquisitions that do not have a

    correct vibration shape or a correct follow-up of the vibration propagation are

    automatically rejected by the software.

    2. PITFALLS OF LS MEASUREMENTS BY MEANS OF TE

    In order to obtain a reliable evaluation by means of TE, the manufacturer

    recommends that at least 10 valid measurements should be obtained. They should

    have a success rate (SR: the ratio of valid shots to the total number of shots) at

    least 60% and an interquartile range (IQR, the difference between the 75 th

    percentile and the 25th

    percentile, essentially the range of the middle 50% of the

    data) less than 30% of the median LSM value.

    Thus, TE is consideredfailedif no valid shots can be obtained, and unreliable if

    fewer than 10 valid shots are obtained, with an IQR greater than 30%, and/or a SR

    less than 60% [2]. In a very large study published by Castera on more than 13,000

    LSMs, the success rate of stiffness evaluation with TE was correlated with the

    body mass index (BMI), decreasing in obese patients (in which it is less than

    80%) [2], but the new probe for obese subjects (the XL probe) has increased the

    percentage of cases with valid results.

    Regarding factors associated with failure, an earlier study performed by Kettaneh

    and et al. [3] on 935 HCV patients, showed that the probability of valid

    measurements (correlated with the histological score) was higher if the operator

    was experienced (with more than 50 FibroScan evaluations performed), if the

    patient was young (OR 0.96/year) and not obese (OR 0.19 if obese). Another

    study by Boursier et al. showed high measurement agreement between novices

    and expert operators, even during the first 10 cases [4], so that a formal session by

    a qualified trainer, followed by practice on 50 cases, should suffice for thetraining of most operators.

    In a prospective study by Foucher et al. [5], the univariant analysis showed that

    failure was associated with: BMI>28 (OR 9.1), diabetes mellitus (OR 2.1), age

  • 7/30/2019 Hepatic Ultrasonography

    40/130

    28 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    >50 years (OR 4.0) and steatohepatitis (OR 3.4). Failure to obtain VM was not

    operator dependent and was not associated with the patients gender, or with the

    aminotransferases level. In the multivariate analysis, the only factor associated

    with failure to obtain VM was BMI>28 (OR 10.0).

    In a study published by our group [6] on 1461 patients, failure to obtain valid

    LSM was observed in 6.9% of the patients. Female gender (OR=1.946), older age

    and higher BMI were significantly associated with failure to obtain valid LSM.

    Also, there are factors that can impair the correlation of LS values by TE with

    liver fibrosis. These factors are: aminotransferases level, liver congestion due to

    heart failure, and extrahepatic cholestasis.

    In a study performed by Coco et al., LS was evaluated considering the

    aminotransferases level, proving that another factor than fibrosis, independently

    associated with LS was ALT for patients with chronic hepatitis [7]. The LS

    dynamics profiles paralleled those of ALT, increasing 1.3 to 3 fold during ALT

    flares. This study also showed that LS remained unchanged in patients with a

    stable biochemical activity. In an Italian study on 12 patients with acute HBV

    hepatitis, repeatedly evaluated by TE and biological tests during a 24 weeks

    follow-up period, Vigano et al. concluded that the initial high values of LS

    mimicking LS cut-off of cirrhosis, likely reflect the liver cell inflammation,

    edema and swelling as they progressively taper down during hepatitis resolution

    [8]. In a study published in 2009, Chan et al. evaluated 161 patients with chronic

    HBV hepatitis and concluded that patients with the same fibrosis staging, but

    higher ALT levels, tend to have higher LSM, and the diagnostic performance for

    low stage fibrosis was most seriously affected when ALT was elevated [9]. All

    three studies confirmed previous results published by Arena and Sagir in 2008

    [10, 11].

    An initial observation of high LS values in a patient with cardiac failure,

    normalized following heart transplantation [12], was confirmed by Millonig et al.in an experimental model on landrace pigs. It showed that the stepwise increase of

    intravenous pressure to 36 cm of water column (3.5 kPa) linearly and reversibly

    increased LS to the upper detection limit of 75 kPa [13]. The experimental data

  • 7/30/2019 Hepatic Ultrasonography

    41/130

    TE Hepatic Elastography Using Ultrasound Waves 29was confirmed in 10 patients with decompensated congestive heart failure, before

    and after recompensation. Initial LS was elevated in all patients, in 8 of them to

    values that suggested liver cirrhosis (median 40.7 kPa). Upon recompensation

    with a median weight loss of 3.0 kg, LS decreased in all 10 patients down to a

    median LS of 17.8 kPa [13].

    The same group of researchers evaluated LS in patients with obstructive jaundice,

    before and after drainage by endoscopic retrograde cholangio-pancreatography.

    After successful biliary drainage, LS decreased by 2.2 to 9.1 kPa, in correlation

    with bilirubin decrease [14]. This observation was confirmed in an animal model

    of bile duct ligation in landrace pigs, where liver stiffness increased from 4.6 kPa

    to 8.8 kPa during 120 minutes of bile duct ligation and decreased to 6.1 kPawithin 30 minutes after decompression [14].

    A significant increase in liver stiffness was observed after food intake for up to 60

    minutes, and the value normalized after 180 minutes. Even if the change was

    modest in most cases (mean change 12 kPa), it determined misclassifications in

    some [15].

    There is conflicting data regarding the influence of steatosis on LS measurements.

    Some studies state that the degree of hepatic steatosis does not appear to affect LS

    [15, 16], while in the study of Lupor et al., the univariant regression analysis

    demonstrated that fibrosis (R2=0.610, p

  • 7/30/2019 Hepatic Ultrasonography

    42/130

    30 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    subjects, in whom VMs were obtained, the mean LS value was 4.81.3 kPa,

    ranging from 2.3 to 8.8 kPa. The mean values of LS in each age group did not

    differ significantly (p=0.5263). (Table 1 and Fig. 3). Also the mean LS in women

    was significantly lower than in men (4.61.2 kPa vs. 5.11.2 kPa, p=0.0082).

    Table 1:Mean liver stiffness values in each age subgroup

    Age group

    (years)

    No. of patients

    with VM

    Mean value of LS

    SD (kPa)

    Minimum (kPa) Maximum (kPa)

    All patients 144 4.81.3 2.3 8.8

    18-29 43 51.3 2.3 8.8

    30-39 24 4.51.2 2.6 7.3

    40-49 17 51.1 3.0 7.1

    50-59 27 4.71.2 2.5 7.7

    60-69 20 51.3 3.2 7.7

    >70 13 4.71.4 3.0 7.1

    Figure 3: Mean LS values according to the age subgroup.

    In a study by Roulot performed on 429 consecutive apparently healthy subjects,

    the mean LS value was 5.491.59 kPa [21], while in a study performed by

    Corpechot et al. [22], a similar mean value (4.8 kPa) was obtained in a group of

    71 healthy subjects. In both studies, LS values were higher in men than in women.

    Overall, the upper limit of normal LS was estimated to be 5.3 kPa [21, 23].

  • 7/30/2019 Hepatic Ultrasonography

    43/130

    TE Hepatic Elastography Using Ultrasound Waves 314. TE IN CHRONIC HEPATOPATHIES

    a) TE in Chronic HCV Hepatitis

    TE assessment of LS was used initially for the evaluation of chronic HCV

    hepatitis. Later, published articles that will be discussed in the following pages,

    proved the methods value in other chronic hepatopathies, such as chronic HBV

    hepatitis, hemochromatosis, primary biliary cirrhosis, human immunodeficiency

    virus (HIV)/HCV co-infection or non-alcoholic steatohepatitis (NASH).

    In HCV viremic patients, if the LS is greater than 6.87.6 kPa (according to the

    results of several studies and meta-analysis) [24-28], there is a great probability of

    finding significant fibrosis on the liver biopsy (F2-F4) and subsequently the

    patient requires antiviral therapy. Probably, in these cases, LB is not required for a

    treatment decision.

    In a multicentre French study coordinated by Beaugrand [29], performed on 494

    HCV patients who were evaluated by means of percutaneous LB (with a

    significant fragment) and valid FibroScan examination, a significant correlation

    was found (p

  • 7/30/2019 Hepatic Ultrasonography

    44/130

    32 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    activity, steatosis or biological activity (ALT) have an important role in the

    assessment of LS by means of FibroScan, as shown in recent studies [7, 17].

    In 324 consecutive patients with chronic HCV hepatitis, evaluated both by TE and

    LB in the same session, the LS values were strongly correlated with fibrosis

    (r=0.759, p

  • 7/30/2019 Hepatic Ultrasonography

    45/130

  • 7/30/2019 Hepatic Ultrasonography

    46/130

    34 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    In a study performed by Ogawa [40] on 68 patients with chronic HBV hepatitis,

    the mean LS values were 3.5 kPa for F0, 6.4 kPa for F1, 9.5 kPa for F2, 11.4 kPa

    for F3, and 15.4 kPa for F4 patients. The values were significantly correlated with

    fibrosis stage (r=0.559, P=0.0093).

    In a prospective study by Marcellin et al., on 202 patients with chronic HBV

    hepatitis, LS was significantly (P

  • 7/30/2019 Hepatic Ultrasonography

    47/130

    TE Hepatic Elastography Using Ultrasound Waves 35A study published in 2011 by Cardoso et al. [43] on 202 HBV patients and 363

    HCV subjects, revealed that TE exhibited comparable accuracies, sensitivities,

    specificities, predictive values and likelihood ratios in HBV and HCV groups.

    Contrary to studies in the Asian population [7-11], AUROC analysis showed no

    influence of ALT levels on the performance of TE in HBV individuals. ALT-

    specific cut-off values did not exhibit significantly higher diagnostic

    performances for predicting fibrosis in HBV patients with elevated ALT.

    In another Asian study, that compared TE performance in HBV vs. HCV patients, the

    conclusion was that discrepancies between LS values and histological fibrosis are due

    to the degree of serum ALT levels, rather than to the cause of hepatitis itself [44].

    The results of these studies, showing a weaker correlation of LS with histological

    fibrosis in HBV than in HCV patients, can be explained in part by the fact that

    high levels of aminotransferases influence the LS values obtained by means of TE

    [7-11]. Thus, LS measurements have to be interpreted in a biochemical context;

    otherwise, there is a risk of overestimating the severity of fibrosis. Also this is

    why LS measurements are not performed in acute hepatitis or during alanine

    aminotransferase (ALT) flares in HBV chronic hepatitis [7, 45].

    In order to minimize the risk of overestimating fibrosis during ALT flares, Chan

    et al. [9] calculated LS cut-off values for various stages of fibrosis considering

    also the aminotransferases levels. In this study, the LS cut-off value for F3 was 9

    kPa in patients with normal ALT and 12 kPa in patients with ALT higher than 5

    times the upper limit of normal. The cut-offs for cirrhosis were 12 kPa in patients

    with normal ALT and 13.4 kPa in those with high ALT.

    The Tsochatzis meta-analysis also assessed the predictive value of LS assessed by

    TE in HBV patients. The pooled cut-off for F2 Metavir was 7 kPa (range 6.97.2,

    lower than in HCV patients), with 0.84 pooled sensitivity and 0.78 pooled specificity

    [32]. In a meta-analysis published by Marcellin, the standardized AUROC of LS

    measurements by TE for F2 Metavir was 0.89 (95% CI 0.83-0.96) [46].

    c) TE in other Chronic Hepatopathies

    Regarding the value of LS measurements by TE in evaluating chronic

    hepatopathies of other etiologies, several studies were performed, in order to

  • 7/30/2019 Hepatic Ultrasonography

    48/130

    36 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    identify significant fibrosis in patients with in HIV-HCV co-infection [47, 48], in

    chronic cholestatic hepatopathies: primary biliary cirrhosis (PBC) and primary

    sclerosing colangitis (PSC) [49] and in NASH [50]. In these studies, the AUROCs

    varied between 0.72 and 0.93, and the cut-off values for F2 ranged between 4

    and 8.7 kPa (Table 3).

    Table 3: Performance of LS for evaluating significant fibrosis in patients with chronic

    hepatopathies other than HCV (PPV Positive Predictive Value; NPV Negative Predictive

    Value)

    Authors De Ledinghen

    et al. [47]

    Vergara et al.

    [48]

    Corpechot et

    al. [49]

    Yoneda et al.

    [50]

    Etiology HCV-HIV HCV-HIV PBC and PSC NAFLD

    No. of patients F 2 44 105 57 33

    Proposed cut-off (kPa) 4.5 7.2 7.3 6.6

    Sensitivity (%) 93.2 88 84 82.7

    Specificity (%) 17.9 66 87 81.3

    NPV (%) 61 75 79 59.1

    PPV (%) 65 88 91 93.5

    AUROC 0.72 0.83 0.92 0.87

    RegardingHCV-HIV coinfection, several studies demonstrated that TE is a useful

    method for fibrosis assessment in patients co-infected with HCV and HIV. In the

    study performed by de Ledinghen et al., LS was significantly correlated to fibrosisstage (Kendall tau-b=0.48; P2, 0.93 (0.85-

    1.00) for F>3 and 0.99 for F4 (cut-offs 7 kPa, 11 kPa and 14 kPa) [51].

    The first study regarding LS by TE in cholestatic hepatitis (primary biliary

    cirrhosis PBC and primary sclerosing colangitis PSC) was published in 2006[49]. In this study, LS was correlated to both fibrosis (Spearman's rho=0.84,

    P

  • 7/30/2019 Hepatic Ultrasonography

    49/130

    TE Hepatic Elastography Using Ultrasound Waves 37curves were 0.92 for F2, 0.95 for F3 and 0.96 for F=4, for the following

    optimal cut-off values 7.3, 9.8, and 17.3 kPa respectively. In another study

    published in 2008 on 80 patients with PBC, LS by TE was significantly correlated

    to the histological fibrosis stage (Kendall coefficient: 0.56; P2 and 0.96 for F=4 [52]. A smaller study in 45 patients

    with PBC showed that the adjusted accuracy of LS by TE for the diagnosis of F2

    was 80%, while for liver cirrhosis it was 95% [53].

    Regarding TE evaluation with nonalcoholic fatty liver disease (NAFLD) and

    nonalcoholic steato-hepatitis (NASH),a positive correlation was found between

    LS values and the histological stage of fibrosis, since even if steatosis may

    attenuate shear waves, it does not modify their speed [54]. LS measurements canbe difficult in patients with NAFLD or NASH, since these conditions are often

    associated with obesity. A first step towards increasing the feasibility of TE in

    these patients was the introduction of the XL probe that increased the number of

    patients that could be evaluated by TE [55-57]. Yoneda et al. evaluated 97

    NAFLD patients by TE and NASH [50]. LS was well correlated with the stage of

    liver fibrosis (Kruskal-Wallis test p

  • 7/30/2019 Hepatic Ultrasonography

    50/130

    38 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    observed in 33 (13.4%) patients. By multivariate analysis, liver biopsy length less

    than 20 mm and F0-2 disease were associated with discordance.

    A new technique, related to TE and performed with a FibroScan device is the

    Controlled Attenuation Parameter (CAP) and it enables steatosis quantification in

    fatty liver. CAP was first validated as an estimate of ultrasonic attenuation at 3.5

    MHz using Field II simulations and tissue-mimicking phantoms. Performance of

    the CAP was then evaluated on 115 patients, taking the histological grade of

    steatosis as reference. CAP was significantly correlated to steatosis (Spearman

    =0.81, p10% and >33% steatosis

    were 0.91 and 0.95 respectively [60].

    Regarding TE evaluation in patients with alcoholic liver disease(ALD), one must

    consider that in most of these patients, inflammation coexists with fibrosis and

    steatosis and it can influence the results of LS measurements, as showed above.

    Higher cut-off values for cirrhosis were reported in patients with ALD, than in

    those with viral hepatitis: 19.5 kPa in the study by Nguyen-Khac et al. [61] and

    22.6 kPa in the Nahon study [62], but the patients included in those studies had

    high ALT levels that were not taken into consideration. In a study by Mueller et

    al. [63], LS was evaluated by TE in a learning cohort of 50 patients with ALD,

    admitted for alcohol detoxification, before and after normalization of serum

    transaminases. LS decreased in almost all patients, within a mean observationinterval of 5.3 days. Of the serum transaminases, the decrease in LS correlated

    best with the decrease in glutamic oxaloacetic transaminase (GOT). No significant

    changes in LS were observed below GOT levels of 100 U/L. In the study cohort

    of 101 patients with histologically confirmed ASH, LS was measured only in

    patients with GOT >100 U/L at the time of LS assessment. In this group, the

    AUROC for cirrhosis detection by FS improved from 0.921 to 0.945 while

    specificity increased from 80% to 90%, at a sensitivity of 96%. A similar AUROC

    was obtained for lower F3 fibrosis stage, if LS measurements were restricted to

    patients with GOT

  • 7/30/2019 Hepatic Ultrasonography

    51/130

    TE Hepatic Elastography Using Ultrasound Waves 395. TE FOR THE DIAGNOSIS OF LIVER CIRRHOSIS

    If the performances of TE for the differentiation of mild from significant fibrosisare only moderate, its real value is for the diagnosis of cirrhosis. Data from 9

    studies were evaluated by Talwalkaret al. [24] showing that TE has 87% pooled

    sensitivity [95% confidence interval (CI): 8490%)] and 91% pooled specificity

    (95% CI: 8992%) for the diagnosis of cirrhosis. In a meta-analysis on 50 studies,

    the mean AUROCs for the diagnosis of significant fibrosis, severe fibrosis, and

    cirrhosis were 0.84, 0.89, and 0.94, respectively [25]. Another meta-analysis from

    2010 [64] evaluated 22 published papers. For a cut-off value of 15.08 kPa, it

    showed a pooled sensitivity of 84.45% (95% CI: 84.2-84.7%) with pooled

    specificity of 94.69% (95% CI: 94.3%-95%). Finally, in a recently published

    meta-analysis which included 40 studies, the summary sensitivity and specificity

    of TE for diagnosing cirrhosis were 0.83 (95% CI: 0.79-0.86) and 0.89 (95% CI:

    0.87-0.91), respectively [32]. The mean optimal cut-off was 154.1 kPa.

    Different cut-off values for the diagnosis of cirrhosis were proposed for different

    etiologies: 12.5 kPa in HCV infection [26]; 13.4 kPa in HBV infection [41]; 10.3

    kPa in NAFLD [59]; 22.4 kPa in ASH [63]; 17.3 kPa in cholestatic chronic

    diseases (primary biliary cirrhosis and primary sclerosing colangitis) [49].

    6. TE FOR THE DIAGNOSIS OF CIRRHOSIS COMPLICATIONS

    The advantage of FibroScan evaluation of liver fibrosis, on other non-invasive

    methods, is that transient elastography can also assess the severity of cirrhosis

    (values up to 75 kPa), as shown in some studies, which proposed cut-off values of

    LS that predict the presence of cirrhosis complications (esophageal varices,

    variceal bleeding, vascular decompensation or hepatocellular carcinoma).

    Esophageal varices and upper digestive hemorrhage are feared complications of

    cirrhosis. The hemorrhage risk depends on the varices size so that primary

    prevention of variceal bleeding should be applied to patients with large EV (grade

    2 or 3) diagnosis established by periodical upper digestive endoscopy (Baveno V

    and AASLD Consensuses) [65, 66]. Such a screening program of periodical

    gastroscopy in all cirrhotics would be very expensive, and repeated endoscopies

    are poorly accepted by the patients. Published studies demonstrated that LS values

  • 7/30/2019 Hepatic Ultrasonography

    52/130

    40 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    10 mmHg AUROC 0.945).

    Robic et al. compared LS measurement by TE to HVPG, as predictors of cirrhosis

    complications. One hundred patients with chronic liver disease were evaluated in

    the same session by TE and HVPG measurements and followed-up for 2 years.

    HVPG and LS measurements showed similar performances for predicting portal

    hypertension: AUROCs 0.830 vs. 0.845. All patients with LS lower than the 21.1

    kPa cut-off value remained free of portal hypertension complications during the 2years follow-up, as compared to 47.5% of those with higher values. The

    performances of LS and HVPG were similar also in the cirrhotic subgroup of

    patients [73].

  • 7/30/2019 Hepatic Ultrasonography

    53/130

    TE Hepatic Elastography Using Ultrasound Waves 41Reibergeret al. performed a study on 122 cirrhotics with EV who were evaluated

    by means of TE and HVPG. There was a better correlation of LS values assessed

    by TE and HVPG in patients with HVPG 12 mmHg than in those with HVPG

    >12 mmHg (r=0.951 vs. r=0.538). Also, the authors observed an improvement in

    the correlation of LS with HVPG under beta-blockers, mainly in hemodynamic

    responders (r=0.864), but not in non-responders (r=0.535), while changes of blood

    pressure, heart rate and LS were similar in responders vs. non-responders. For

    discriminating cirrhotic patients with at least grade 2 EV, from those with grade 1

    EV, for a cut-off value of 47.5 kPa, LS had 80.6% sensitivity and 47.7%

    specificity [74].

    In a review published in 2011, Castera concluded that diagnostic performancesof TE are acceptable for the prediction of clinically significant portal

    hypertension, but far from satisfactory to confidently predict the presence of OV

    in clinical practice and to screen cirrhotic patients without endoscopy [75]. But

    all the studies included in this review evaluated only small numbers of patients

    (ranging from 47 to 211), with contradicting results (cut-off values for significant

    EV ranging from 19.8 to 48 kPa, and AUROCs ranging from 0.73 to 0.87).

    In a study published by our group [76], not available for the Castera review,

    including 1000 consecutive cirrhotic patients, we found out that negative and

    positive predictive values (NPV and PPV) for at least grade 2 EV were 76.2% and71.3%, respectively, for a cut-off value of 31 kPa, chosen to maximize the sum of

    sensitivityand specificity. For >40 kPa criterion, chosen to have a PPV of more

    than 85%, the sensitivity was 77.8%, the specificity 68.3%, with 86% PPV and

    55% NPV (95%CI: 49.6060.23). We also searched for a cut-off value as close as

    possible to a NPV of 90%, and we found out that for 17.1 kPa, the NPV was

    89.3%, with 43.2% PPV, 92.6% sensitivity and 33.5% specificity (AUROC

    0.7807). So, according to our data, at least 8 out of 10 patients with TE values >40

    kPa will have significant portal hypertension, therefore it seems reasonable to

    recommend prophylactic beta-blocker therapy in these patients, without

    endoscopy. Similarly, 5 out of 10 patients with TE values

  • 7/30/2019 Hepatic Ultrasonography

    54/130

    42 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    evaluation, since they have only 1 in 10 risk to present significant EV (NPV

    89.3%).

    In our study group, we also observed that the mean LS value in patients with a

    history of variceal bleeding was significantly higher than in those with no

    bleeding history: 51.921.56 vs. 35.200.91kPa, p

  • 7/30/2019 Hepatic Ultrasonography

    55/130

    TE Hepatic Elastography Using Ultrasound Waves 43predictive value for diagnosis of fibrosis F2, with AUROC 0.90, while for F4 the

    AUROC was 0.98 [79]. Another study that evaluated 95 transplanted HCV

    patients by means of paired liver biopsies and TE, showed that LS changed in

    parallel with grading (r=0.63) and staging (r=0.71), with good sensitivity (86%)

    and specificity (92%) in predicting staging increases [80].

    In a systematic review published in 2010, Cholongitas et al. showed that TE had a

    good discrimination power for significant fibrosis (median AUROC: 0.88, median

    sensitivity 0.86, median NPV 0.90 and median PPV 0.8) [85]. In a recent meta-

    analysis, the pooled data of 5 studies that estimated at least F2 in transplant HCV

    patients were 83% for sensitivity and specificity, 4.95 for the positive likelihood

    ratio, 0.17 for the negative likelihood ratio, and 30.5 for the diagnostic odds ratio.Five studies assessed cirrhosis, and their pooled estimates were 98% for

    sensitivity, 84% for specificity, 7 for the positive likelihood ratio, 0.06 for the

    negative likelihood ratio, and 130 for the diagnostic odds ratio [86].

    As demonstrated above, TE reliably predicts severity of recurrent HCV hepatitis

    following liver transplantation, but its accuracy in non-viral liver graft damage is

    unknown. Rigamontti et al. evaluated 69 transplant recipients (37 hepatitis B/D

    recurrence-free, 20 autoimmune/cholestatic liver disease, 6 alcoholic liver disease

    and 6 mixed) by means of both protocol or on demand liver biopsy and

    concomitant TE. 94% of patients had reliable TE examinations during post-transplant follow-up (median 18 months, range 7-251). Liver biopsy showed graft

    damage in 43% (28) patients. LS values were significantly higher in patients with

    graft damage as compared to the ones without (median 7.8 kPa vs. 5.3 kPa,

    p

  • 7/30/2019 Hepatic Ultrasonography

    56/130

    44 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    One hundred and sixteen consecutive children with various liver diseases were

    evaluated, and only in one TE was not feasible. TE showed the best correlation to

    clinical and biological severity parameters. Also, TE was significantly correlated

    with the Metavir fibrosis score. The AUROCs of TE, FibroTest and APRI for

    predicting cirrhosis were 0.88, 0.73 and 0.73, respectively.

    Nobili et al. evaluated 52 consecutive NASH pediatric patients by means of LB

    and TE [89]. Even if an adult probe was used and most patients were overweight

    and obese, TE proved to be a highly feasible (96% of patients with reliable

    measurements) and highly reproducible (intraclass correlation coefficient 0.961)

    method in children. The AUROCs for prediction of any (>1), significant (>2),

    or advanced fibrosis (>3) were 0.977, 0.992, and 1, for cut-offs

  • 7/30/2019 Hepatic Ultrasonography

    57/130

    TE Hepatic Elastography Using Ultrasound Waves 45could not be obtained [measurements with Success Rate (SR)

  • 7/30/2019 Hepatic Ultrasonography

    58/130

    46 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    REFERENCES

    [1] Beaugrand M. Fibro Scan instructions for use. Journes Francophones de PathologieDigestive. Gastroenterol. Clin Biol 2006; 30: 513- 514.[2] Castera L, Foucher J, Bernard PH, et al. Pitfalls of Liver Stiffness Measurement: A 5-Year

    Prospective Study of 13,369 Examinations. Hepatology 2010; 51: 828-835.

    [3] Kettaneh A, Marcellin P, Douvin C, et al. Features associated with success rate andperformance of FibroScan measurements for the diagnosis of cirrhosis in HCV patients: a

    prospective study of 935 patients. J Hepatol 2007; 46: 628-634.[4] Boursier J, Konate A, Guilluy M, et al. Learning curve and interobserver reproducibility

    evaluation of liver stiffness measurement by transient elastography. Eur. J. Gastroenterol.Hepatol 2008; 20: 693701.

    [5] Foucher J, Castera L, Bernard PH, et al. Prevalence and factors associated with failure ofliver stiffness measurement using FibroScan in a prospective study of 2114 examinations.Eur J Gastroenterol Hepatol 2006; 18: 411-412.

    [6] irli R, Sporea I, Deleanu A, et al. Factors associated with failure of Liver Stiffnessmeasurement using Transient Elastography. Timisoara Med J2009: 49; 34-38.

    [7] Coco B, Oliveri F, Maina AM, et al. Transient elastography; a new surrogate marker ofliver fibrosis influenced by major changes of transaminases. J Viral Hepat 2007; 14: 360-369.

    [8] Vigan M, Massironi S, Lampertico P, et al. Transient elastography assessment of the liverstiffness dynamics during acute hepatitis B. Eur J Gastroenterol Hepatol 2010; 22: 180-184.

    [9] Chan HL, Wong GL, Choi PC, et al. Alanine aminotransferase-based algorithms of liverstiffness measurement by transient elastography (Fibroscan) for liver fibrosis in chronic

    hepatitis B. J Viral Hepatol 2009; 16: 36-44.[10] Sagir A, Erhardt A, Schmitt M, et al. Transient elastography is unreliable for detection of

    cirrhosis in patients with acute liver damage. Hepatology 2008; 47: 592-595.

    [11] Arena U, Vizzutti F, Corti G, et al. Acute viral hepatitis increases liver stiffness valuesmeasured by transient elastography. Hepatology 2008; 47: 380-384.

    [12]

    Millonig G, Friedrich S, Adolf S, et al. Liver stiffness is directly influenced by centralvenous pressure. J Hepatol 2010; 52: 206-210.

    [13] Millonig G, Reimann FM, Friedrich S, et al. Extrahepatic cholestasis increases liverstiffness (FibroScan) irrespective of fibrosis. Hepatology 2008; 48: 1718-1723.

    [14] Mederacke I, Wursthorn K, Kirschner J, et al. Food intake increases liver stiffness inpatients with chronic or resolved hepatitis C virus infection. Liver Int2009; 29: 15001506.

    [15] Arena U, Vizzutti F, Abraldes JG, et al. Reliability of transient elastography for thediagnosis of advanced fibrosis in chronic hepatitis C. Gut 2008; 57: 12881293.

    [16] Wong VW, Vergniol J, Wong GL, et al. Diagnosis of fibrosis and cirrhosis using liverstiffness measurement in nonalcoholic fatty liver disease. Hepatology2010; 51: 454462.

    [17] Lupor M, Badea R, Stefnescu H, et al. Analysis of histopathological changes thatinfluence liver stiffness in chronic hepatitis C. Results from a cohort of 324 patients. J

    Gastrointestin Liver Dis 2008; 17: 155-163.

    [18] Fraquelli M, Rigamonti C, Casazza G, et al. Reproducibility of transient elastography in theevaluation of liver fibrosis in patients with chronic evaluation of liver fibrosis in patients

    with chronic liver disease. Gut2007: 56; 968-973.

    [19] Nobili V, Vizzutti F, Arena U, et al. Accuracy and reproducibility of transient elastographyfor the diagnosis of fibrosis in pediatric nonalcoholic steatohepatitis. Hepatology2008; 48:

    442448.

  • 7/30/2019 Hepatic Ultrasonography

    59/130

    TE Hepatic Elastography Using Ultrasound Waves 47[20] irli R, Sporea I, Tudora A, et al. Transient elastographic evaluation of subjects without

    known hepatic pathology: does age change the liver stiffness? J Gastrointestin Liver Dis

    2009; 18: 57-60.[21] Roulot D, Czernichow S, Le Clsiau H, et al. Liver stiffness values in apparently healthy

    subjects: influence of gender and metabolic syndrome. J Hepatol 2008; 48: 606 -613.

    [22] Corpechot C, El Naggar A, Poupon R. Gender and liver: is the liver stiffness weaker inweaker sex? Hepatology 2006; 44: 513514.

    [23] Kim SU, Choi GH, Han WK, et al. What are true normal liver stiffness values usingFibroScan?: a prospective study in healthy living liver and kidney donors in South Korea.

    Liver Int 2010; 30: 268274.

    [24] Talwalkar JA. Kurtz DM, Schoenleber SJ, et al. Ultrasound-based transient elastographyfor the detection of hepatic fibrosis: systematic review and meta-analysis. Clin

    Gastroenterol Hepatol 2007; 5: 1214-1220.

    [25] Friedrich-Rust M, Ong MF, Martens S, et al. Performance of transient elastography for thestaging of liver fibrosis: a meta-analysis. Gastroenterology 2008; 134: 960-974.

    [26] CasteraL, Vergniol J, Foucher J, et al. Prospective comparison of transient elastography,FibroTest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C.Gastroenterology 2005; 128: 343-350.

    [27] Ziol M, Handra-Luca A, Kettaneh A, et al. Noninvasive assessment of liver fibrosis bymeasurement of stiffness in patients with chronic hepatitis C. Hepatology 2005; 41: 48-54.

    [28] Sporea I, irli R, Deleanu A, et al. Comparison of the liver stiffness measurement bytransient elastography with the liver biopsy. World J Gastroenterol 2008; 14: 6513-6517.

    [29] Beaugrand M, Ziol M, Marcelin P, et al. Liver stiffness cut off values in HCV patients:validation and comparison in an independent population. Hepatology 2006; 44 (Suppl. 1):

    269.[30] irli R, Sporea I, Nicoli D, et al. Are there different values of liver stiffness evaluated by

    means of transient elastography in patients with HBV or HCV chronic hepatitis?Gastroenterology 2011:140 (Suppl 1); S-978.

    [31] Shaheen AA, Wan AF, Myers RP. FibroTest and FibroScan for the prediction of hepatitisC-related fibrosis: a ystematic review of diagnostic test accuracy. Am J Gastroenterol 2007;102: 25892600.

    [32] Tsochatzis EA, Gurusamy KS, Ntaoula S, et al. Elastography for the diagnosis of severityof fibrosis in chronic liver disease: a meta-analysis of diagnostic accuracy. J Hepatol 2011;

    54: 650-659.

    [33] Castera L, Vergniol J, Foucher J, et al. Prospective comparison of transient elastography,FibroTest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C.

    Gastroenterology 2005; 128: 343-350.

    [34] Castera L, Sebastiani G, Le Bail B, et al. Prospective comparison of two algorithmscombining non-invasive methods for staging liver fibrosis in chronic hepatitis C. J Hepatol.

    2010; 52: 191-198.

    [35] Sporea I, irli R, Popescu A, et al. Is it better to use two elastographic methods for liverfibrosis assessment? World J Gastroenterol 2011; 17: 3824-3829.

    [36] Hzode C, Castera L, Roudot-Thoraval F, et al. Liver stiffness diminishes with antiviralresponse in chronic hepatitis C. Aliment Pharmacol Ther 2011; 34: 656-663.

    [37] Andersen ES, Moessner BK, Christensen PB, et al. Lower liver stiffness in patients withsustained virological response 4 years after treatment for chronic hepatitis C. Eur J

    Gastroenterol Hepatol 2011; 23: 41-44.

  • 7/30/2019 Hepatic Ultrasonography

    60/130

    48 Hepatic Elastography Using Ultrasound Waves Sporea andirli

    [38] Ogawa E, Furusyo N, Toyoda K, et al. The longitudinal quantitative assessment bytransient elastography of chronic hepatitis C patients treated with pegylated interferon

    alpha-2b and ribavirin. Antiviral Res 2009; 83: 127-134.[39] Wang JH, Changchien CS, Hung CH, et al. Liver stiffness decrease after effective antiviral

    therapy in patients with chronic hepatitis C: Longitudinal study using FibroScan. J

    Gastroenterol Hepatol 2010; 25: 964-969.

    [40] Ogawa E, Furusyo N, Toyoda K, et al. Transient elastography for patients with chronichepatitis B and C virus infection: Non-invasive, quantitative assessment of liver fibrosis.

    Hepatology Research 2007; 12: 1002-1010.

    [41] Marcellin P, Ziol M, Bedossa P, et al. Non-invasive assessment of liver fibrosis by stiffnessmeasurement in patients with chronic hepatitis B. Liver Int 2009 ; 29: 242-247.

    [42] Sporea I, irli R, Deleanu A, et al. Liver stiffness measurements in patients with HBV vs.HCV chronic hepatitis: a comparative study. World J Gastroenterol 2010 14; 16: 4832-

    4837.

    [43] Cardoso AC, Carvalho-Filho RJ, Stern C, et al. Direct comparison of diagnosticperformance of transient elastography in patients with chronic hepatitis B and chronichepatitis C. Liver Int 2011 Nov 22. doi: 10.1111/j.1478-3231.2011.02660.x. [Epub ahead

    of print].

    [44] Cho HJ, Seo YS, Lee KG, et al. Serum aminotransferase levels instead of etiology affectsthe accuracy of transient elastography in chronic viral hepatitis patients. J Gastroenterol

    Hepatol 2011; 26: 492-500.

    [45] Wong GL, Wong VW, Choi PC, et al. Increased liver stiffness