FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform...

17
FTIR Spectroscopy (Fourier Transform Infrared) Infrared spectroscopy is an important technique in organic chemistry. It is an easy way to identify the presence of certain functional groups in a molecule. Also, one can use the unique collection of absorption bands to confirm the identity of a pure compound or to detect the presence of specific impurities. This page will step you through the principles of operation of the FTIR Spectrometer, and how it differs from classical continuous-wave (CW) instruments. To skip directly to spectral interpretation, Continuous Wave spectrometers Before delving into FT spectrometry, let's review the principles of a classical spectrometer. If you have used an optical or UV spectrometer, the principles are identical:

Transcript of FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform...

Page 1: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

FTIR Spectroscopy (Fourier Transform Infrared)

Infrared spectroscopy is an important technique in organic chemistry.  It is an easy way to identify the presence of certain functional groups in a molecule.  Also, one can use the unique collection of absorption bands to confirm the identity of a pure compound or to detect the presence of specific impurities.

This page will step you through the principles of operation of the FTIR Spectrometer, and how it differs from classical continuous-wave (CW) instruments.  To skip directly to spectral interpretation,

Continuous Wave spectrometers

Before delving into FT spectrometry, let's review the principles of a classical spectrometer.  If you have used an optical or UV spectrometer, the principles are identical:  

A source generates light across the spectrum of interest. A monochromater (in IR this can be either a salt prism or a

grating with finely spaced etched lines) separates the source radiation into its different wavelengths.

A slit selects the collection of wavelengths that shine through the sample at any given time.

Page 2: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

In double beam operation, a beam splitter separates the incident beam in two; half goes to the sample, and half to a reference.

The sample absorbs light according to its chemical properties. A detector collects the radiation that passes through the

sample, and in double-beam operation, compares its energy to that going through the reference.

The detector puts out an electrical signal, which is normally sent directly to an analog recorder.  A link between the monochromater and the recorder allows you to record energy as a function of frequency or wavelength, depending on how the recorder is calibrated.

Although very accurate instruments can be designed on these principles, there are several important limitations. 

First, the monochromater/slit limits the amount of signal one can get at a particular resolution.  To improve resolution, you must narrow the slit and decrease sensitivity. 

Second, there is no easy way to run multiple scans to build up signal-to-noise ratios.

Finally, the instrument must be repetitively calibrated, because the analog connection between the monochromater position and the recording device is subject to misalignment and wear.

Page 3: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

FT instrumentation

Now let's look at an FT instrument.

We still have a source, a sample and a detector, but everything else is different.  Now, we send all the source energy through an interferometer and onto the sample.  In every scan, all source radiation gets to the sample!  The interferometer is a fundamentally different piece of equipment than a monochromater.  The light passes through a beamsplitter, which sends the light in two directions at right angles.  One beam goes to a stationary mirror then back to the beamsplitter.  The other goes to a moving mirror.  The motion of the mirror makes the total path length variable versus that taken by the stationary-mirror beam.  When the two meet up again at the beamsplitter, they recombine, but the difference in path lengths creates constructive and destructuive interference:  an interferogram:

The recombined beam passes through the sample.  The sample absorbs all the different wavelengths characteristic of its spectrum, and this subtracts specific wavelengths from the interferogram.  The detector now reports variation in energy versus time for all wavelengths simultaneously.  A laser beam is superimposed to provide a reference for the instrument operation.

Energy versus time is an odd way to record a spectrum, until you recognize the relationship between time and frequency:  they are reciprocals!  A mathematical function called a Fourier tranform allows us to convert an intensity-vs.-time spectrum into an intensity-vs.-frequency spectrum.

Page 4: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

The Fourier transform:

  A(r) and X(k) are the frequency domain and time domain points, respectively, for a spectrum of N points.    

An interferogram 

A spectrum

Now, you have to do the Fourier transform for every point in the interferogram.  You may like to do exponential functions by hand, but I (and most organic chemists) are far too lazy for that.  Fortunately, even a slow computer can efficiently perform this operation.  The output of the detector is digitized, and a small computer program will do the transform in a matter of seconds (or less).  All modern FT instruments are computer-interfaced.

Page 5: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

There are several advantages to this design:

All of the source energy gets to the sample, improving the inherent signal-to-noise ratio.

Resolution is limited by the design of the interferometer.  The longer the path of the moving mirror, the higher the resolution.  Even the least expensive FT instrument provides better resolution that all but the best CW instruments were capable of.

The digitization and computer interface allows multiple scans to be collected, also dramatically improving the signal-to-noise ratio.

Most of the computer programs today allow further mathematical refinement of the data:  you can subtract a reference spectrum, correct the baseline, edit spurious peaks or otherwise correct for sample limitations.

The one minor drawback is that the FT instrument is inherently a single-beam instrument; it cannot use the "channel ratio" trick used in CW operation.  One result is that IR-active atmospheric components (CO2, H2O) will appear in the spectrum.  Usually, a "Background" spectrum is run, then automatically subtracted from every spectrum.  The spectrum below is such a background scan.

 

Page 6: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

You can see CO2 as the storng doublet at around 2400 cm-1, and water as the "spiky" peaks in the 3800 and 1600 cm-1.  The "bell curve" shape of the spectrum reflects the output spectrum of the source:  strong in the middle, but falling off at the ends.  

FTIR Spectroscopy (Fourier Transform Infrared) A word about units.

Most spectra using electromagnetic radiation are presented with wavelength as the X-axis.  Originally, IR spectra were presented in units of micrometers.  Unfortunately, a linear axis in micrometers compresses the region of the spectrum 10-15 mm) that usually has the largest number of peaks.

One could rectify this by presenting the spectrum on a linear scale vs. frequency (Hz), but the magnitude is unwieldy (10 mm = 3 x 1013 Hz).  A different measure, the wave number, is given the unit cm-1. 

Spectral Interpretation  

Once you collect a spectrum, the real work begins. Spectra of organic compounds have two general areas:  

4000-1500 cm-1 1500-400 cm-1

The Functional Group Region 

Peaks in this region are characteristic of specific kinds of bonds, and therefore can be used to identify whether a specific functional group is present.

The Fingerprint Region   

Peaks in this region arise from complex deformations of the molecule.  They may be characteristic of molecular symmetry, or combination

Page 7: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

bands arising from multiple bonds deforming simultaneously.     

     The two regions of the spectrum overlap to a degree.  (In fact, one always finds overlap between different regions of any spectrum; these designations are "guideposts" to help you orient yourself.)  For example, carbon-chlorine bonds appear at around 800 cm-1, and C-O single bonds appear at around 1200-1300 cm-1.  Also, benzene rings show "overtones" in the 1500-1700 cm-1 region, even though these arise from complex ring deformations.       The normal way to approach interpretation of an IR spectrum is to examine the functional group region to determine which groups might be present, then to note any unusually strong bands or particularly prominent patterns in the fingerprint region.  Finally, if you think you have identified the compound (usually you need additional information) you can compare the spectrum with a reference.  Matching the fingerprint region is a very rigorous test.

 

Some important IR-active functional groups, and examples of spectra.    

Group Region Examples of spectra. (Try to find the characteristic peaks.)

Page 8: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

C-H

3000-3100 cm-1 (sp2)  2800-3000 cm-1 (sp3)

C=O

1600-1800 cm-1  Acids: 1650-1700  Esters: 1740-1750  Aldehydes: 1720-1750  Ketones: 1720-1750  Amides:1650-1715

O-H  (alcohol)

3300-3600 cm-1  Monomeric forms: sharp. 

H-bonding leads to broadening.

Page 9: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

O-H  (acids)

2400-3000 cm-1  Very broad, medium intensity 

C C�   C N�

2200-2100 cm-1  Usually weak; maybe not visible  in internal alkynes.  Nitriles are quite strong.

C-O

1200-1300 cm-1  Often difficult to assign,  depending on fingerprint region.

Page 10: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

N-H

3400 cm-1  Usually sharper than O-H.

A final word about symmetry.

Molecular vibrations give rise to IR bands only if they cause a change in the dipole moment of the molecule.  (This comes out of the quantum mechanics of molecular absorption of energy, so we aren't concerned too much with why, yet.)  If a stretch does not change the dipole moment, there won't be any IR band.  This is why O2 and N2 in the atmosphere don't show any IR bands.  CO2, however, has a stretch where one O moves in and the other moves out:

 

Thus we see this band at 2400 cm-1.

Page 11: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented

Use of the Nicolet IR Spectrometers

 

This tutorial will help you with sample preparation, instrument setup and use of OMNIC software.  Please follow the menu below based on which experiment you are doing.   You should have read the background on instrument design and spectral interpretation before using this area.

 

Sample preparation:

Thin film on salt plates (Viscous liquids, as in the Synthesis)

Solution cell (Volatile liquids, as in the Hacac/Fe(acac)3 measurement)

KBr pellet (solids, as in the Fe(acac)3

Page 13: FTIR Spectroscopy - University of Babylon  · Web viewFTIR Spectroscopy (Fourier Transform Infrared) A word about units. Most spectra using electromagnetic radiation are presented