EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1 , C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United Kingdom 2 OptoElectronic Engineer, Swansea, Wales, United Kingdom The following potential conflict of interest relationships are germane to my presentation: Financial grant, equipment, consulting fees and travel expenses paid by CyDen Ltd.

Transcript of EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1...

Page 1: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

EVALUATION OFINTENSE PULSED LIGHT DEVICES USING

STANDARDIZED METHODOLOGY

G. Town1, C. Ash2

1RPA2000 Laser Protection Adviser, Haywards Heath, United Kingdom2OptoElectronic Engineer, Swansea, Wales, United Kingdom

 The following potential conflict of interest relationships are germane to my

presentation:Financial grant, equipment, consulting fees

and travel expenses paid by CyDen Ltd.

Page 2: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

• Status of FDA devices used for the material being presented:– NA/Non-Clinical

• Status of off-label use of devices, drugs or other materials that constitute the subject of this presentation:– NA/Non-Clinical

Page 3: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Introduction

• 18 different devices from 15 different manufacturers

and 36 applicators (different cut-off filters) from USA, UK, Israel, Sweden, Switzerland, China and Italy were tested by the authors.

• Data was gathered on-site between patient appointments over a 6-month period in clinics where devices were in daily use.

• Manufacturers’ information was collected from user manuals, company web sites and literature.

Page 4: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Fluence Values

30 IPL applicators were measured at maximum fluence of which 11 were more than 20% below and 8 were more than 10% above fluence levels given on the device display or claimed in user manuals, even where brand-new lamps were

tested.

Page 5: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Fluence Values

9 IPL devices out of 18 had applicators that were outside of the standard formedical Class 4 lasers (> ±20%)

Page 6: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Pulse Duration

Only 14 of 29 pulse duration measurements were within ±20% of the manufacturers stated or system-displayed values.

Study Ref

Stated Pulse Duration (ms)

Measured Pulse Duration (ms)

Study Ref

Stated Pulse Duration (ms)

Measured Pulse Duration (ms)

A 3 x 10 ms = 30 3 x 6 ms = 18 I short:5.5/5.5 (2) 5.5/5.5 (2)

A1 14 14.5 med:3.6/3.6/3.6 (3) 4.0/4.0/4.0 (3)

2 x 2.5 ms = 15 2 x 3 ms = 15.5 long:3.6/3.6/3.6 (3) 4.5/4.5/4.5 (3)

B 5 15-17 J 40 40

C 5 15-17 K 34.8 37

D 50 51 123 121

10-15 no data L 15 black 2.2

E 10-50 10-51 ms 15 blonde 5

F 20-150 missing pulses M 3 3

20-151 missing pulses 5 5

G 10 6 N 35 132

20 6.6 35 132

20 8.7 10 24

H 5.5/5.5/5.5 5.4 6.4 7.0 O 40 42

Q 30/40/50 30/40/50

Page 7: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Electrical Discharge Shape

Only IPLs ‘E’ and ‘G’ exhibited a true single square pulse shape confirming that they used partial discharge capacitor technology although close

pulse-stackingin devices ‘A’ and ‘O’ effectively achieved the same pulse shape and device ‘D’ showed a nearly square pulse shape.

IPL ‘E’

IPL ‘A’

IPL ‘G’ IPL ‘C’

IPL ‘B’IPL ‘Q’

Page 8: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Measurement of UV Content

Shorter wavelengths in the ultraviolet region of the spectrum may burn the patient’s skin or may be hazardous to the operator’s eyes and are therefore removed. Of the 30

applicators tested, 6 IPLs measured more than 1% and two measured more than 2% of unwanted UV output below 400 nm when cut-off filters were set significantly higher.

Page 9: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Average Spectral Output

The spectrum analysis of light emitted from flashlamps contains both discrete line structure and continuum radiation. The continuum radiation is blackbody radiation, characteristic of the temperature of the plasma in the discharge. Of 29 applicators 19 (65.5%) with cut-off filters that were inaccurate by more than 20 nm versus the claimed cut-off value given by

the manufacturer. Only 10 applicators (34.5%) were within 20 nm of the stated cut-off.

Page 10: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Comparing Time-Resolved Spectral Output

Schematic illustration of the difference in the spatial and temporal characteristics of a free discharge and partial discharge pulse to an IPL xenon lamp.

Page 11: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Time-resolved spectrum sampled every 1 ms using an OceanOptics HR2000+ spectrometer and SpectraSuite software to demonstrate the stability and efficiency of spectral output for

free-discharge (IPL ‘C’) vs. square pulse systems (IPL ‘E’)

IPL ‘C’—Free Discharge IPL ‘E’—Square Pulse

Page 12: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘D’ single pulse programs produce a decaying square pulse

Page 13: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘D’ in a multi-pulse program showing more pronounced ‘free-discharge’ characteristics

Page 14: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘G’ —Partial Discharge—square pulse

Page 15: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘N’—Discharge (stated 35 ms pulse duration)

Page 16: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Close “stacking” of sub-pulses produces a square pulse effect unless sub-pulses are spaced too far apart when they simply perform like conventional free-discharge pulses.

IPL ‘A’ —Free Discharge

Page 17: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘L’ showing sub-pulses in typical ‘free discharge’ format

Page 18: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘K’ showing sub-pulses in typical ‘free discharge’ format

Page 19: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

IPL ‘C’ showing a typical ‘free discharge’ of ca 3 ms of energy

Page 20: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

CONCLUSIONS

• Many manufacturers of IPLs make claims that are not substantiated as measured by this study

• Manufacturers operating an ISO 9000 Q/A system (Medical CE-mark) showed greater consistency of stated and actual values for fluence, pulse duration and spectral cut-off filter accuracy over the stated lifetime of the applicator / lamps.

• Partial Discharge (‘square pulse’) SystemsProduce the lowest possible intensity for a given fluence thus minimising discomfort and other side effects such as skin burns (i.e. safer to use)

• Free Discharge (‘rising/falling slope’) SystemsProduce high energy in short pulse durations with spectral shift and wasted energy

Page 21: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

CONCLUSIONS

• A time-resolved ‘Spectral Footprint’ as shown in these measurements is helpful in determining the true pulse duration and spectral pattern of useful energy for optimized clinical efficiency

• This first serious attempt to measure IPL performance characteristics provides criteria that should be carried to national representatives of the competent regulatory body (IEC) to promote an international standard for measuring intense light devices

Page 22: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Acknowledgements

The authors wish to thank the following companies for their contribution in reviewing the fluence test methodology used in this study:

 Cyden Ltd., Swansea, Wales, UK

(www.cyden.co.uk)

Energist Ltd., Swansea, Wales, UK(www.energist-international.com)

Instinctive Technologies Ltd., Bedford, UK(www.instinctiveuk.com)

Lynton Lasers Ltd., Cheshire, UK(www.lynton.co.uk)

Page 23: EVALUATION OF INTENSE PULSED LIGHT DEVICES USING STANDARDIZED METHODOLOGY G. Town 1, C. Ash 2 1 RPA2000 Laser Protection Adviser, Haywards Heath, United.

Measuring key parameters of Intense Pulsed Light (IPL) Devices. The key for the device ID: A = DDD EllipseLight A1 = DDD EllipseFlex B = Active Crystal512 C = Chromogenex Chromolite D = Alma Harmony E = CyDen iPulse F = Freedom (Chinese/UK) G = Unimed NovaLight H = Lumenis Quantum I = Lumenis Aculight J = Greenton Ecolite V K = MBC Plasmalite L = Lynton Lumina M = Energist ULTRA N = Radiancy SkinStation O = Sciton BBL (Profile) P = GP Flash 1 Q = Alma Aria (was Lovely II or SuperiorPro = Harmony with IPL only)