Environmental scanning electron microscopy a technique that offers unique possibilities for exposure studies Martin Ebert and Stephan Weinbruch

Transcript of Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ......

Page 1: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Environmental scanning electron microscopya technique that offers unique possibilities for exposure studies

Martin Ebert and

Stephan Weinbruch

Page 2: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

CONTENTS

1. Individual Particle Analysis by SEM

2. Basics of Environmental Scanning Electron Microscopy (ESEM)

3. ESEM applications in exposure studies

Page 3: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 4: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 5: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 6: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 7: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Cabability of Scanning Electron Microscopy

-Excellent resolution (~ 1nm)

-Excellent depth of field

-Chemical element identification by EDX

But all samples in the SEM are exposed to high vacuum

⇒ no volatile or nonconductive samples can be investigated(without sample preparation)

Page 8: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Why do we need high vacuum in a SEM ?

Page 9: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 10: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 11: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 12: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

3. Basics of Environmental Scanning Electron Microscopy (ESEM)

Page 13: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 14: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 15: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 16: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 17: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Gaseous Secondary Electron Detector (GSE)

• gas molecules are used foramplification

p contrast

best image quality~4 Torr

• more gas molecules cause more scattering

p resolution

Page 18: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 19: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 20: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

- 30 0 30 60 1000 1500

temperature [°C]

Cooling stage

Heating stage 1

Heating stage 2

Temperature working range in ESEM

Page 21: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Druck u. Temperatur-Arbeitsbereich im ESEM

Der mögliche Druck und Temperatur-Arbeitsbereich im Esem ermöglicht die Untersuchung von Wasser sowohl im flüssigen als auch im festen Zustand.

Druck in Torr

Page 22: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und
Page 23: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

4. ESEM applications in exposure studies

4.1. Characterization of volatile and astable aerosol components

At pressures of 1- 10 mbar most of the volatile components of theaerosol can be studied in the ESEM (sulfates, nitrates, organics, liquid water). Secondary electron picture

of ammonium nitrate particlesin the ESEM

Page 24: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

fresh soot in the ESEM same particleexposed to high vacuum

Characterization of volatile aerosol components

Page 25: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

a b

Particles deposited on gelantine covered substrates can directlybe ínvestigated in the ESEM

Investigation of pollen and spores

Page 26: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

Investigation of pollen and spores

a b c

Spores bursting and degassing under electron bombardement

Page 27: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

4.2. Investigation of water containing samplesand interactions with water

soot with solved inclusions undried or even living biogenicmaterial (e.g. dermal tissue)

mites

Page 28: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

increasingrelative humidity

Activation of unsoluble particlesat high relative humidities

RH = 90%

RH = 99%

RH = 100%

soot

water

Page 29: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

drop formation recrystallization

RH = 80% RH = 85% RH = 60%

Deliquescence and efflorescence

Deliquescence and efflorescence of a sodium sulfate particle

Page 30: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

60 65 70 75 80 85

60

65

70

75

80

85Na2SO4

(NH4)2SO4

NaCl

NH4NO3

DR

H [%

] at 2

5°C

, thi

s st

udy

DRH [%] at 25°C, references*

DRH at 25°C determined in the ESEM, compared with the values obtained by other techniques

*Ebert et al., 2002

Page 31: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

NaCl

Deliquescence und efflorescence of sodium chloride

20 30 40 50 60 70 80 90 100

1.0

1.5

2.0

2.5

crystallisationJoutsensaari et al.

crystallisation ESEM

dropformation

RH inc., ESEM, DP > 100 nm, T = 5° RH dec., Jout. et al., DP = 100 nm T = 23°C RH dec., ESEM, DP > 100 nm, T = 5° RH inc., Hämeri et al., DP = 50 nm, T = 25°C RH inc., Jout. et al., DP = 100 nm, T = 23°C RH dec., Hämeri et al., DP = 50 nm, T = 25°C

grow

th fa

ctor

relative humitidy [%]

Page 32: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

RH 72 % RH 95 %

Deliquescence behavior of Ni-containing particles

b) Partial deliquescence

Page 33: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

4.3. Ice in the ESEM

Page 34: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

aerosol particles and ice nucleation in the atmosphere

Heterogeneous ice nucleation (induced by special aerosol particles):

was observed already at -5°C !!!

Homogeneous ice nucleation of a supercooled cloud droplet:

does not start until -38°C !!!

aerosol particles

? CCNicenuclei

Page 35: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

sample chamber of the ESEM with cooling stage

samplepeltiercoolingstage

watercooling

SE detector

gaseous SE detector

pole shoe

CCD camera

Page 36: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

mica ice

T = - 5°C

Heterogeneous ice nucleation in the ESEM

ice

mica

Page 37: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

4.4. additional devices: Micro manipulator

scratching the surface

electrical conducting

micromanipulator in ESEM

drop off or pick up

Page 38: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

micro injector (drop off or pick up solution)

Page 39: Environmental scanning electron microscopy - IAG · Environmental scanning electron microscopy ... values obtained by other techniques *Ebert et al., 2002. NaCl Deliquescence und

CONTEXTAdditionally to the capabilities of conventional SEM, ESEM enables

electronmicroscopic analysis at pressures up to 50 Torr (~6700 Pa) and temperatures between –30 to 1500°C.

No charging effects ⇒ no coating of non-conductive samples.

Analysis of non conductive, wet, astable and living material becomessuitable. This enables a more complete picture of the ambient aerosols.

Observation of in-situ processes (SE/BSE/EDX).

The scanning electron microscope can be used as a reactionchamber, customized for different requirements (cryo-SEM, micro-manipulator, microinjector, residual gas analysis).

Hygroscopic behaviour (e.g. deliquescnece, efflorescence, activationice forming prperties..) of particles can be studied in situ.