Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting...

12
Technical Information TI 207C/07/en No. 51501593 Smartec S is a compact measuring system for inductive conductivity measurement, specially designed for measurements in highly conductive liquids. The various process connections of the Smartec S CLD 132 allow an individual adaption of the transmitter to different mounting situations. Areas of application Alkali/acid concentration control Phase separation of product/water and product/product mixtures in pipe systems Monitoring and control of bottle cleaning systems Product monitoring in breweries, dairies and the beverage industry CIP system control. Benefits at a glance Transmitter housing made of stainless steel (IP 67) Sensor made of highly resistant synthetic material (PEEK) Sterilisable sensor (140 °C) Insensitive to polarisation and soiling Various mounting possibilities Sensor versions for all industrial process connections Compact transmitter Simple installation Easy operation due to logical menu structure Large two-line display allows simultaneous display of measured value and temperature Direct calibration access via CAL key Safe operation due to overvoltage protection according to EN 61000-4-5 Alarm contact for error signalling Communication interface HART ® or PROFIBUS Standard version extendable by: Measuring range remote switching (MRS) Temperature coefficient determination Alarm contact switchable as limit contactor. Endress Hauser The Power of Know How Conductivity Measurement smartec S CLD 132 Transmitter for measurement of conductivity and concentration with inductive sensor Quality made by Endress+Hauser ISO 9001

Transcript of Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting...

Page 1: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

TechnicalInformationTI 207C/07/enNo. 51501593

Smartec S is a compact measuringsystem for inductive conductivitymeasurement, specially designed formeasurements in highly conductiveliquids. The various process connectionsof the Smartec S CLD 132 allow anindividual adaption of the transmitter todifferent mounting situations.

Areas of application• Alkali/acid concentration control• Phase separation of product/water

and product/product mixtures in pipesystems

• Monitoring and control of bottlecleaning systems

• Product monitoring in breweries,dairies and the beverage industry

• CIP system control.

Benefits at a glance• Transmitter housing made of stainless

steel (IP 67)• Sensor made of highly resistant

synthetic material (PEEK)• Sterilisable sensor (140 °C)• Insensitive to polarisation and soiling• Various mounting possibilities• Sensor versions for all industrial

process connections• Compact transmitter• Simple installation• Easy operation due to logical menu

structure• Large two-line display allows

simultaneous display of measuredvalue and temperature

• Direct calibration access via CAL key• Safe operation due to overvoltage

protection according to EN 61000-4-5• Alarm contact for error signalling• Communication interface HART® or

PROFIBUS• Standard version extendable by:

– Measuring range remote switching(MRS)

– Temperature coefficientdetermination

– Alarm contact switchable as limitcontactor.

Endress HauserThe Power of Know How

Conductivity Measurementsmartec S CLD 132

Transmitter for measurement of conductivityand concentration with inductive sensor

Quality made byEndress+Hauser

ISO 9001

Page 2: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Measuring system A complete measuring systemcomprises:• the transmitter Smartec S CLD 132• conductivity sensor CLS 52 with

integrated temperature sensor andfixed cable or

• the compact instrument Smartec SCLD 132 with integrated conductivitysensor CLS 52

Optional for the separate version:

• extension cable CLK 5• junction box VBM• Mounting kit for pipe mounting.

Basic version andadditional functions

ENDRESS+HAUSERSMARTEC SENDRESS+HAUSERSMARTEC S

ALARM

CAL

E

+

Smartec S CLD 132

Conductivity sensorCLS 52

Smartec S CLD 132with integrated

conductivity sensor

LD132SYS.CDR

Completemeasuring systems withSmartec S CLD 132

top:Separate version andconductivity sensorCLS 52

bottom:Compact version withintegrated conductivitysensor CLS 52

Fuctions of the basic version Options

MEASUREMENT

CALIBRATION of installation factorCALIBRATION of cell constantCALIBRATION of residual coupling

Read instrument DATA

Linear CURRENT OUTPUTCURRENT OUTPUT simulation

CHECK of measuring system by PCS alarm(live check)

TEMPERATURE COMPENSATION selectable(1 free table)

CONCENTRATION table selectable(4 defined tables, 1 free table)

RELAY as alarm contact

SERVICE functions

2nd current output for temperature

MEASURING RANGE REMOTE SWITCHING:

REMOTE SWITCHING of max. 4 measuring ranges

TEMPERATURE COMPENSATION selectable(4 free tables)

CONCENTRATION table selectable(4 defined tables, 4 free tables)

RELAY configurable as alarm or limit contactor

TEMPERATURE COEFFICIENTS can be determined

2

Page 3: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Measurement principle In inductive conductivity measurement,a transmitting coil generates a magneticalternating field that induces an electricvoltage in a liquid. The ions present inthe liquid enable a current flow whichincreases with increasing ionconcentrations.

The current in the liquid generatesa magnetic alternating field in thereceiving coil. The resulting current inthe receiving coil is measured and usedto determine the conductivity value. Theconductivity serves as a measure of ionconcentration.

This measurement principle has thefollowing advantages:• No electrodes, therefore no

polarisation• Error-free measurement in media with

a tendency to sedimentate• Complete galvanic isolation of

measurement from medium.

Measuring technology according tothe purity law

The sensor, injection-moulded for highlychemically, mechanically and thermallyresistant PEEK (polyether ether ketone),does not have joints or crevices and istherefore biologically safe.

The temperature sensor Pt 100 is indirect thermal contact with the mediumvia the stainless steel thermalconductivity socket. This assuresextremely fast temperature response(t90 < 5 s).

The use of special components andmaterials makes the sensor suitable forcontinuous exposure to temperaturesof +125 °C and brief exposures(max. 30 min) to +140 °C for sterilization.

ENDRESS+HAUSERSMARTEC SENDRESS+HAUSERSMARTEC S

ALARM

CAL

E

+

LD132FCT.CDR

1

3

4

6

2

5

7

8

9

Measurement andfunction principleof Smartec S

LS51-T12.EPS

Detail view of theconical measuringchannel

LS52-03.EPS

Sensor CLS 52 withtemperature sensorPt 100 with flow lines

3

Page 4: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Dimensions

LD132-01.CDR

207

242

(180

)

142

35.5

∅ 49

34

Dimensions in parentheses denote thetotal length of the compact device withsensor Varivent, APV or 2 “ Perlick.

Dimensions ofSmartec S CLD 132,compact version

LD132-02.CDR

142

225

175

109

170

95

Dimensions ofSmartec S CLD 132,separate transmitterwith post mountingaccessories

225

4

Page 5: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Mounting

160

225

LD132-10.CDR

∅ 7

Wall mounting of theseparate transmitterSmartec S CLD 132

LD132-11.CDR

Back plateincluded in delivery

Post mounting kit(accessories)

LD132-13.CDR

∅ max. 60

Post mounting of theseparate transmitterSmartec S CLD 132

5

Page 6: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Mounting

The minimum pipe socket diameterdepends on the sensor used.

The pipe socket and the wall distanceof the built-in sensor determine theinstallation factor.

Installation factor The electrical conductivity of the liquidprimarily depends on the ionconcentration. However, installation andsensor geometry must also be takeninto account for measurement. The cellconstant completely describes thesensor geometry.

If the distance from the wall is sufficient(a > 15 mm), then it is not necessary toconsider the installation factor (f = 1.00).If the distance from the wall is smaller,then the installation factor increases inthe case of electrically insulating pipes(f > 1) and decreases in the case ofelectrically conductive pipes (f < 1).

LD132-08.CDR

Horizontal flow Vertical flow

b

b

a

Flowdirection

Flow direction

a = wall distanceb = immersion depth

min. DN 40

min. DN 40

Mounting of thecompact version ofSmartec S CLD 132in flow pipes

DIAGR.CDR

2

1

1.40

1.20

1.00

0.80

Inst

alla

tion

fact

orf

0 5 10 15 20 25Wall distance a in mm

Installation factor fis dependent ondistance a from pipe wall1 Electrically

conductive pipe2 Insulating pipe

Sensor Diameter pipe socket Installation factor

Varivent, 2 “ Perlick , APV ≥ DN 40 f > 1

Dairy fitting, 2" SMS,G 1 ½ thread, 2" Clamp

≥ DN 50 f > 1

no influence of diameter ≥ DN 65 f = 1

6

Page 7: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Process connections

LD132-06.CDR

Dairy fitting DN 50(DIN 11851)

2" Clamp connection(ISO 2852)

G 1 ½ internal thread

83

60 AF

83 102

2 “ ClampG 1½

Varivent connectionDN 40 ... DN 125

2" Perlick connection

∅ 84 ∅ 90

46 33

∅ 92

83

∅ 84

2 “ SMS

APV connectionDN 40 ... DN 100

∅ 100

46

Process connectionsof Smartec S CLD 132,compact version

LD132-07.CDR

Dairy fitting DN 50(DIN 11851)

G 1 ½ internal thread 2" Clamp connection(ISO 2582)

60 AF

2 “ Clamp

83

G 1½

102

∅ 84

8383

Varivent connectionDN 40 ... DN 125

2" Perlick connection

46 33

∅ 89∅ 84

2 “ SMS

∅ 92

∅ 100

APV connectionDN 40 ... DN 100

46

Process connectionsof conductivity sensorsCLS 52

7

Page 8: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Electrical connection

Pressure andtemperature ranges

15 V85

86

10-50 V

81

82

10-50 V

93

94

41

42

43

Cd

mA31

32

mA33

3411

12

13

S

84

83

15

16

∼–∼–

Temperaturesensor Pt 100

Power supply

LD13212E.CDR

Auxiliary voltageoutput

Digital input 1(HOLD,MRS 3+4*)

Signal output 1:conductivity

Alarm(contact status:no current)

Digital input 2(MRS 1+2*)

Sensor

optionalSignal output 2:temperature(optional)

PE

* MRS = measuring range remote switching (optional)Electrical connectionof Smartec S CLD 132

60 80 100 120 14020 40125

0

10

0

–10–5

20

30

40

5055

55

35

60

70

[° C]

[° C]

brief exposurefor sterilization

(< 30 min)

T-DIAGR.CDRMedium temperature

Compact version

Sensor CLS 52 with separate version

Am

bien

ttem

pera

ture

Permissible pressureand temperature rangesof Smartec S CLD 132

60 80 100 120 14020 4012590

0

42

–5

8

12

16

20

[bar]

[° C]

Med

ium

pres

sure

Medium temperature

LS52-DTB.CDR

brief exposurefor sterilization

(< 30 min)

Permissible pressureand temperatureranges of the sensorCLS 52

8

Page 9: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Operation Everything at a glanceThe display simultaneously shows thecurrent measured value and thetemperature – the essential processdata. Brief informational texts in theconfiguration menu provide assistancewith parameter configuration.

Intelligent and simpleAll instrument control functions arearranged in a logical menu structure.The individual parameters can only bemodified by following the access codeentry. The current position within themenu structure is displayed (as shownin figure below – O213).

ENDRESS+HAUSERSMARTEC S

ALARM

CAL

E

+

2000mS/cm

O213

20 mA

DISPLAY1.CDRUser interface:Display and keys

SETUP 1A

SETUP 2B

ALARMF

CHECKC

RELAYSR

ALPHA TABLET

CONCENTRATIONK

SERVICES

INTERFACEI

TEMP. COEFF.D

MEAS. RANGEREMOTE SWITCHING

M

CURRENT OUTPUTO

Selection of measurement unit, entry of measured value damping

Adjustment of current output assignment, curve and simulation for verification

Selection of contact type; adjustment of switching behaviour;adjustment of error current

Adjustment of live-check alarm

Configuration of contact as limit contactor or alarm contact

Definition of table for temperature compensation (up to 10 value pairs)α

Selection and editing of concentration curves(4 defined, up to 4 free curves)

Selection of language version; configuration of HOLD function;entry of release codes

Entry of device address for communication interface HART or PROFIBUS

Determination of temperature coefficients

Selection of digital inputs and parameter sets for remote switching ofmeasuring range

Temperature measurement and temperature compensation

CALIBRATIONC

Calibration of cell constant; calibration of residual coupling;entry of installation factor; adaptive calibration

Current temperature display can be suppressed by pressing the PLUS key

Display of active errors (sequence according to priority)

MEAS. VALUE

TEMPERATURE

Display

LD132BEE.CDR

Overview of Smartec SCLD 132 menu.This overview coversall options that can beinstalled.

9

Page 10: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Technical data

General specifications Manufacturer Endress+Hauser

Product designation Smartec S CLD 132

Physical data / design Dimensions of separate transmitter with mounting rail(L × W × D)

225 × 142 × 109 mm

Weight approx. 2.5 kg

Dimensions of compact deviceversions MV1, CS1, GE1, SMS (L × W × D)

225 × 142 × 242 mm

Dimensions of compact deviceversions VA1, AP1, PER (L × W × D)

225 × 142 × 180 mm

Weight with sensor CLS 52 approx. 3 kg

Measured value display LC display, two-line, five and nine digits for text,current status display

Materials Housing stainless steel SS 304 S

Front window polycarbonate

Sensor CLS 52 PEEK

Input Measured quantities conductivity, concentration, temperature

Minimum distance for 0 / 4 ... 20 mA signal

For conductivity measurement meas. value 0 ... 19.99 µS/cm: 2 µS/cmmeas. value 20 ... 199.9 µS/cm: 20 µS/cmmeas. value 200 ... 1999 µS/cm: 200 µS/cmmeas. value 2 ... 19.99 mS/cm: 2 mS/cmmeas. value 20 ... 200 mS/cm: 20 mS/cmmeas. value 200 ... 2000 mS/cm: 200 mS/cm

For concentration measurement no minimum distance

Inductive conductivity measurement

Measuring range 10 µS ... 2000 mS/cm

Cell constant k = 5.9 cm–1

Maximum cable length 55 m (CLK 5)

Measuring frequency 2 kHz

Temperature measurement

Temperature sensor Pt 100, class A acc. to IEC 60751

Measuring range –10 ... +150 °C

Temperature offset range ±5.0 °C

Temperature compensation

Compensation types none (α = 0), linear, table, NaCl

Range –10 ... +150 °C

Minimum distance 1 K

Reference temperature 25 °C

Digital inputs 1 and 2

Voltage 10 ... 50 V

Current consumption max. 10 mA at 50 V

Output Conductivity signal output (not with PROFIBUS)

Current range 0 / 4 ... 20 mA, galvanically isolated; error current 2.4 / 22 mA

Load max. 500 ΩMaximum resolution 700 digits/mA

Transmission range adjustable

Separation voltage max. 350 Veff / 500 V DC

Overvoltage (lightning) protection acc. to EN 61000-4-5:1995

Temperature signal output (optional, not with PROFIBUS)

Current range 0 / 4 ... 20 mA, galvanically isolated

Load max. 500 ΩMaximum resolution 700 digits/mA

Transmission range adjustable, ∆ 10 ... ∆ 100 % of measuring range

Separation voltage max. 350 Veff / 500 V DC

Overvoltage protection acc. to EN 61000-4-5:1995

10

Page 11: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Technical dataAuxiliary voltage output

Output voltage 15 V ± 0.6 V

Output current max. 10 mA

Contact outputs

Switching current with ohmic load (cos ϕ = 1) max. 2 A

Switching current with inductive load (cos ϕ =0.4) max. 2 A

Switching voltage max. 250 V AC, 30 V DC

Switching power with ohmic load (cos ϕ = 1) max. 1250 VA AC, 150 W DC

Switching power with inductive load (cos ϕ = 0.4) max. 500 VA AC, 90 W DC

Limit contactor

Pickup / dropout delay 0 ... 2000 s

Alarm

Function (switchable) steady contact / fleeting contact

Alarm delay 0 ... 2000 s (min)

Accuracy Conductivity measurement

Deviation of indication1 max. 0.5 % of measured ± 4 digits

Reproducibility max. 0.2 % of measured ± 2 digits

Measurement deviation1, conductivity signal output 0.75 % of current output range

Temperature measurement

Resolution 0.1 °C

Deviation of indication1 max. 0.6 % of measuring range

Measurement deviation1, temperature signal output max. 0.75 % of current output range

Power requirements Supply voltage 100 / 115 / 230 V AC +10 / –15 %, 48 ... 62 Hz24 V AC/DC +20 / –15 %

Power consumption max. 7.5 VA

Fuse protection fine-wire fuse, medium time-lag, 250 V / 3.15 A

Operating data compact version Operating temperature max. 55 °C at 55 °C ambient temperaturemax. 140 °C (30 min) at 35 °C ambient temperature

Operating pressure max. 16 bar (90 °C)

Ambient conditions Storage and transport temperature –25 ... +70 °C

Relative humidity (nominal operating conditions) 5 ... 95 %, non-condensing

Ingress protection of housing IP 67

Electromagnetic compatibility interference emission and interference immunityacc. to EN 61326-1:1997

Vibrational resistanceacc. to IEC 770

Mounting place pipe

Vibration frequency 10 ... 60 Hz

Peak amplitude 0.21 mm

Impact test Impact strength of display window 9 J

Sensor data CLS 52 Conductivity measuring range 10 µS ... 2000 mS/cm

Cell constant k = 5,9 cm-1

Temperature sensor Pt 100, class A acc. to IEC 60751

Temperature measuring range Pt 100 –5 ... +140 °C

Temperature response time t90 < 5 s

Measuring value deviation ±10 µS/cm + 0.5% of meas. value at –5 ... 100 °C±30 µS/cm + 0.5% of meas. value at > 100 °C

max. cable length 55 m

Material in contact with medium PEEK, stainless steel SS 316 S, Chemraz

Ambient temperature –10 ... +70 °C

max. operating temperature 140 °C (max. 30 min)

max. operating pressure max. 16 bar (90 °C)

Ingress protection IP 67

Supplementary documentation Technical Information CLS 52 order no.: 50086110

1acc. to IEC 60746-1, for nominal operating conditions Subject to modifications.

11

Page 12: Endress Hausershlevel.co.kr/ehprduct/pdf/CLD132.pdfMounting 160 225 LD132-10.CDR ∅7 Wall mounting of the separate transmitter Smartec S CLD 132 LD132-11.CDR Back plate included in

Accessories

Product structure

Endress+Hauser GmbH+Co.- Instruments International -

P.O. Box 2222D-79574 Weil am Rhein

Tel. (07621) 975-02Fax (07621) 975-345

E-Mail: [email protected]

TI 207C/07/en/12.99Printed in Germany / DT / CV5 (51501593)

Endress HauserThe Power of Know How

VersionP Compact versionW Separate transmitter, 5 m cable lengthX Separate transmitter, 10 m cable length

Process connectionMV1 Dairy fitting DN 50 (acc. to DIN 11851)CS1 2" Clamp connection (acc. to ISO 2852)GE1 G 1 ½ internal threadVA1 Varivent connection from DN 40 upwardAP1 APV connection from DN 40 upwardSMS 2" SMS connectionPER 2" Perlick connection

Cable entry1 Cable gland Pg 13.53 Cable gland M 20 × 1.55 Conduit adapter NPT ½"

Power supply0 230 V AC1 115 V AC5 100 V AC8 24 V AC / DC

Current output / communicationAA Current output conductivityAB Current output conductivity and temperatureHA Current output conductivity, HARTHB Current output conductivity and temperature, HARTPE PROFIBUS-PAPF PROFIBUS-PA, M 12 plugPP PROFIBUS-DP

Additional equipment1 Basic version2 Measuring range remote switching /

extended functionality

CLD 132- complete order code

Conductivity and resistance transmitter Smartec S CLD 132

Mounting accessories

Type Features Order number

Postmounting kit

Kit for mounting of field housing on horizontalor vertical pipes (∅ max. 60 mm)Material: stainless steel (SS 304 S)

50062121

Cable extension

Type Features Order number

CLK 5 Extension cable for inductive sensorsup to 55 m

50085473

VBM Junction box for inductive measuring cables 50003987