Endoscope Reprocessing
Transcript of Endoscope Reprocessing
-
8/14/2019 Endoscope Reprocessing
1/34
-
8/14/2019 Endoscope Reprocessing
2/34
Endoscopy Reprocessing
THETA Education Day
-
8/14/2019 Endoscope Reprocessing
3/34
Endoscope Reprocessing
Device Classification
Manual Cleaning
Personal Protective Equipment
Biofilms within GI Endoscopy
Reprocessing Room Standards
Endoscope Reprocessing Protocols
Leakage Testing
Endoscope Handling
-
8/14/2019 Endoscope Reprocessing
4/34
Endoscope Reprocessing
Review of Terms
Biofilmrefers to a complex community of microorganisms thatform a matrix of extracellular material composed ofexopolysaccharides (EPS)
Minimum Effective concentration (MEC)refers to the lowestconcentration of active ingredient necessary to meet the label claimof a reusable high-level disinfectant / sterilantchemical teststrips should be used to determine whether and effectiveconcentration of the active ingredient is present despite repeated
use
Reuse-liferefers to a statement by the manufacturer indicatingthe maximum number of days of a reusable high-level disinfectant/ sterilant might be effective
-
8/14/2019 Endoscope Reprocessing
5/34
Endoscope Reprocessing
-
8/14/2019 Endoscope Reprocessing
6/34
Endoscope Reprocessing
Two classifications of Endoscopes
Critical Endoscope:
Endoscopes used in the examination of critical spaces such as jointsand sterile cavities. Many of these endoscopes are rigid with nolumen. Examples of critical endoscopes are arthroscopes and
laproscopes
Semicritical Endoscope
Fiberoptic or video endoscopes used in the examination of the hollowviscera. These endoscopes generally invade only semicritical spaces,although some of their components might enter tissue or other critical
spaces ie. Forceps. Examples of semi-critical endoscopes arecolonoscopes, gastroscopes, duodenoscopes, sigmoidoscopes,laryngoscopes, nasopharangeal endoscopes and enteroscopes.
Opininons differ regarding the reprocessing requirements ofbronchoscopes; a minimum of high level disinfection is required
-
8/14/2019 Endoscope Reprocessing
7/34
Medical Device Classification
Spaulding Classification
Determines the degree of disinfection or sterilization required forvarious medical devices.
Critical: A device that enters normally sterile tissue, including thevascular system. These devices should be sterilized
1)Semicritical: A device that comes into contact with intact mucous
membranes and does not ordinarily penetrate sterile tissue. Thesedevices are cleaned followed by High Level Disinfection
2)Noncritical: Devices that do not ordinarily touch the patient ortouch only intact skin. These devices may be cleaned by low-leveldisinfection
-
8/14/2019 Endoscope Reprocessing
8/34
Manual Cleaning
Manual Cleaning
Thorough and meticulous manual cleaning of all instruments mustprecede exposure to any high-level disinfectant or sterilant. Thisprocess significantly reduces the organic and microbial challenge tothe high-level disinfectant or sterilant. An item that has not beencleaned cannot be assuredly disinfected or sterilized.
Refer to endoscope manufacturers guidelines for design featuresunique to a particular instrument
Personal Protective Equipment
Should be used when reprocessing endoscopes. Gowns, gloves andprotective eyewear are recommended when handling any high-leveldisinfectant / sterilant
-
8/14/2019 Endoscope Reprocessing
9/34
Personal Protective Equipment
Personal Protective Equipment
Gownsimpervious to fluid, long sleeves that fit snugly aroundthe wrist, and wrap to cover as much of the body as possible.Dispose of or launder gowns if they become wet or are exposed
to contaminated materialGlovesinspect for tears or holes before use. Gloves should be
long enough to extend up the arm to protect the forearm orclothing from splashes or seepage. To prevent cross-contamination, change gloves and wash hands whenever moving
from a dirty to clean task or environmentEye and / or face protection are necessarycontact lenses are
not sufficient eye protection. A face shield is recommended. Donot use high filtration masks since they may actually trapvapours.
-
8/14/2019 Endoscope Reprocessing
10/34
Biofilms
Biofilms within GI Endoscopy
A biofilmis a structured community of microorganismsencapsulatedwithin a self-developed polymeric matrix and adherent to a living orinert surface.If the colonists are not immediately separated from the surface, theycan anchor themselves more permanently using cell adhesionBiofilms within endoscopy may form within endoscopes if propermanual pre-cleaning and CSGNA guidelines are not followed forendoscope reprocessing
Staphylococcus
aureus biofilm
Exopolysaccharide (EPS)
Staphylococcus aureus
http://en.wikipedia.org/wiki/Microorganismhttp://en.wikipedia.org/wiki/File:Staphylococcus_aureus_biofilm_01.jpghttp://en.wikipedia.org/wiki/Microorganism -
8/14/2019 Endoscope Reprocessing
11/34
Reprocessing Room Standards
The process and products used for cleaning, disinfection and
or/sterilization of endoscopes must be compatible with theequipment being used
Each health care setting in which endoscopic procedures areperformed should have written detailed procedures for thecleaning and handling of endoscopes
Reprocessing of contaminated patient equipment should bedone in an area designated and dedicated for this function
This room should be separate from where endoscopicprocedures are performed
Ventilation must be capable of removing toxic vapours
generated by, or emitted from, cleaning or disinfectant agentsthe vapour concentration of the chemical disinfectant beingused should not exceed allowable limits (eg. 0.05 ppm forglutaraldehyde)
Minimum of 10-12 air exchanges per hour in the reprocessing
area
-
8/14/2019 Endoscope Reprocessing
12/34
Reprocessing Room Standards
Tap water and / or water that has been filtered by passage througha 0.2micron filter or water of equivalent quality should be availablein the reprocessing area
Manual cleaning should include a medical grade, low-foaming,neutral pH enzymatic formulated for endoscopes that contain
enzymes to digest all components of bioburden; including, blood, fat,carbohydrate, uric acid, starch
Accessories
Accessories which are classified as critical devices (e.g. biopsy
foceps) require sterilization. Critical items labeled for single-useshould not be reprocessed and / or reused.
-
8/14/2019 Endoscope Reprocessing
13/34
Transportation and Handling of
Contaminated Endoscopes
Covered containers with easily cleanable surfaces should beused for handling and transporting soiled endoscopes
Soild endoscopes should be transported by direct routes where
cleaning will be performed
Containers used to transport soiled endoscopes should becleaned after each use
-
8/14/2019 Endoscope Reprocessing
14/34
Endoscope Reprocessing Protocols
Basic steps to clean and perform high-leveldisinfection of gastrointestinal endoscopes
1) Pre-cleaning
2) Leakage testing3) Cleaning
4) Rinsing
5) Disinfection
6) Rinsing
7) Drying
8) Storage
-
8/14/2019 Endoscope Reprocessing
15/34
Endoscope Reprocessing Protocols
A. Pre-cleaning
1) Immediately after removal of the insertion tube from thepatient and prior to disconnecting the endoscope from thepower source
Prepare for bedside cleaning:
PPE Container with enzymatic Sponge or lint-free cloth Air and water channel cleaning adapters per manufacturers
instruction
Protective video cap
2) Wipe the insertion tube with the wet cloth or sponge soakedin the freshly prepared enzymatic solution. Note that thecloth / sponge should be disposed of between cases
-
8/14/2019 Endoscope Reprocessing
16/34
Endoscope Reprocessing Protocols
A. Pre-cleaning
3) Place the distal end of the endoscope into the enzymatic solution.Suction the solution through the biopsy / suction channel, alternatesuctioning enzymatic solution and air several times until thesolution is visibly cleanFinish by suctioning air
Note:Alternate suctioning of fluid and air is more effective thansuctioning fluid alone in the removal of debris from lumens immediate flushing of the biopsy/suction and air/water channelsprecludes drying of debris on lumen surfaces
-
8/14/2019 Endoscope Reprocessing
17/34
Endoscope Reprocessing Protocols
A. Pre-cleaning
4) Flush or blow out air and water channels in accordance withthe endoscope manufacturers instructions
5) Flush the auxiliary water channel
6) Detach the endoscope from the light source and suction pump7) Attach protective video cap if using a video endoscope
8) Transport the Endoscope to the reprocessing area in anenclosed container
Note: Containers, sinks, and basins should be large enough thatthe endoscope will not be damaged by being coiled too tightly
-
8/14/2019 Endoscope Reprocessing
18/34
Endoscope Reprocessing Protocols
B. Cleaning the Endoscope in the Reprocessing Area
Prepare the following:
PPE Leakage testing equipment Channel cleaning adapters
Large basin of endoscope detergent solution Channel cleaning brushes Sponge or lint-free cloth
C. Leak Testing
Leak Testing detects damage to the interior or exterior of theendoscope
The leak test is done before immersion of the endoscope inreprocessing solutions to minimize damage to parts of theendoscope not designed for fluid exposure.
-
8/14/2019 Endoscope Reprocessing
19/34
Endoscope Reprocessing Protocols
C. Leak Testing
1. Manual Leak Testing
Remove suction valves, air water valves, and biopsy valves
Attached the leak tester and pressurize the scope beforesubmerging it in water
With the pressurized insertion tube completely submerged,flex the distal portion of the scope in all directions, observingfor bubbles
Submerge the entire endoscope and, observing the control
head of the scope, depress the freeze and release buttons.
Check the insertion tube and distal bending section as well asthe universal cord for bubbles coming from the interior of thescope
-
8/14/2019 Endoscope Reprocessing
20/34
Endoscope Reprocessing Protocols
C. Leak Testing
1. Computerized leak testing Remove suction valves, air water valves, and biopsy valves Attach the leak tester to the computer unit Input data including scope ID and user Move knobs and depress the freeze and release buttons when
indicated Reprocess when test is complete
Follow the endoscope manufacturersinstructions if a leak or high humidity
is detected or if the endoscope appearsdamaged
-
8/14/2019 Endoscope Reprocessing
21/34
Existing Manual Leak Test Methods
Two Methods: Wet Test & Dry Test Both methods should take ~3 minutes
of uninterrupted technician time if done properly
Human Error Factors:
Takes skill, commitment and dedication
Rushing
Leak detection problems
Lack of consistency and/or training
Manual Equipment Error Factors:
No automated detection
100% reliance on visual leak observation
Lack of procedure control
No record keeping
Endoscope Reprocessing Overview
-
8/14/2019 Endoscope Reprocessing
22/34
Fluid Invasion
Pathogen inpatient #1
enters
scope via
leak
Cross-patient
infection
breakout
e.g. Pseudomonas,Salmonella, HIV,
Hepatitis B and C,
Staph, E. coli, C. diff
Leak
harbors
pathogen
from
disinfection
Pathogen
dislodged in
subsequent
procedures
into other
patients
*APICGuideline for
infection
prevention and
control in
flexible
endoscopy
Volume 28,
number 2, p.145
-
8/14/2019 Endoscope Reprocessing
23/34
Endoscope Reprocessing Study Results
1%of endoscopy procedures result in the endoscope
developing a leak
65%of leaks are currently detected (35% undetected) and
undetected leaks lead to fluid invasion
60%of endoscope repair costs are fluid invasion-related
.2%of endoscopy procedures result in fluid invasion
without a leak present (usually due to poor handling
such as cleaning cap left off or poorly sealed)
10-15%of patient-ready endoscopes possess a leak
Endoscope Reprocessing Overview
-
8/14/2019 Endoscope Reprocessing
24/34
Endoscope Reprocessing Protocols
D. Cleaning Solutions
Composition of soil found on endoscopes includes, proteins, fats,carbohydrates and the various chemical salts that exist in bloodand other body fluids.
Ideally, a cleaning solution should have a broad spectrum of
effectiveness against these various contaminants and not harmthe device being cleaned
i) Enzymatic cleaning solutions use surfactants to breakdown anddigest bioburden. They are specifically selected to have anegligible effect on surface tension while still suspending soilparticles. This feature provides easy rinsibility.
-
8/14/2019 Endoscope Reprocessing
25/34
Endoscope Reprocessing Protocols
E. Cleaning
Manual cleaning of endoscopes is necessary immediatelyafter removing the endoscope from the patient and priorto automated or manual disinfection
Performed as the first and most important step in removing themicrobial burden from an endoscope
Retained debris may inactivate or interfere with the capability ofthe active ingredient of the chemical solution to effectively killand /or inactivate microorganisms
-
8/14/2019 Endoscope Reprocessing
26/34
Endoscope Reprocessing Protocols
E. Cleaning
1. Fill a sink with freshly prepared solution of water and a medicalgrade, low-foaming, neutral pH detergent formulated forendoscopesEnzymatic detergent must be discarded after eachuse.
2. Dilute and use according to the detergent manufacturersinstructions.
3. Immerse the endoscope
4. Wash all the debris from the exterior of the endoscope bybrushing and wiping the instrument while submerged in thedetergent solution. Note that the instrument should be leftunder water during the cleaning process to prevent splashing ofcontaminated fluid and aerosolization of bioburden
-
8/14/2019 Endoscope Reprocessing
27/34
Endoscope Reprocessing Protocols
E. Cleaning
5. Use a small, soft brush to clean all removable parts, including insideand under the suction valve, air/water valve, and biopsy port cover andopenings.
6. Brush all accessible endoscope channels including the body, insertiontube and the umbilicus of the endoscopebrushes used for cleaninglumens should be of an appropriate size, inspected before and after useand discarded or cleaned, high-level disinfected and dried following use
7. After each passage, rinse the brush in enzymatic solution, removing anyvisible debris before retracting and reinserting it
8. Continue brushing until there is no debris visible on the brush9. Cleanining items should be disposable or thoroughly cleaned and
minimum of high-level disinfected between cases.
-
8/14/2019 Endoscope Reprocessing
28/34
Endoscope Reprocessing ProtocolsE. Cleaning
10. Attach the endoscope cleaning adapters for suction, biopsy, air, andwater channelsNote:Automated pumps are available for this step that eliminate
the manual flush
11. Attach the manufacturers cleaning adapters for special endoscope channels(dual channel, elevator channel, auxilliary channel)
i. To achieve adequate flow through all lumens, various adapters or channelrestrictors may be required
ii. The elevator channel of a duodenoscope is a small lumen, this channelrequires manual reprocessing using a 2-5mm syringe
12. Flush all channels with the detergent solution to remove debris
13. Soak the endoscope and its internal channels for the period of time specifiedby the label
14. If immediate reprocessing is not possible the endoscope shouldbe leak-tested, flushed, brushed, and allowed to soak in aenzymatic solution until it can be thoroughly reprocessed
-
8/14/2019 Endoscope Reprocessing
29/34
Endoscope Reprocessing Protocols
F. Rinse After Cleaning
Thoroughly rinse the endoscope and all removable parts withclean water to remove
Purge water from all channels using forced air. Dry the exteriorof the endoscope with a soft, lint-free cloth to prevent dilution of
the liquid chemical germicide used in subsequent stepsG. High Level Disinfection
Recognized as the standard of reprocessing for endoscopes by:
CSGNA / SGNA
CPSOCollege of Surgeons of Ontario ASGEAmerican Society for Gastrointestinal Endoscopy ACG - American College of Gastroenterology AGA - American Gastroenterological Association APIC - Association for Professionals in Infection Control and
Epidemiology
CDCCenters for Disease Control and Prevention
-
8/14/2019 Endoscope Reprocessing
30/34
Automated Endoscope
Reprocessing
Ensure that the endoscope and endoscope components arecompatible with the automated endoscope reprocessor (AER)
Follow the OEM instructions for use in the AER
AER provide a method by which a permanent record ofendoscope use and reprocessing can be monitored andvalidated
Some AER have a system capable of tracking endoscopes andpatients. For each procedure the patients name and recordnumber, the date and time of procedure, type of procedure,the endoscopist and the serial number of the endoscope arerecorded and stored to assist in outbreak investigation.
-
8/14/2019 Endoscope Reprocessing
31/34
Endoscope Reprocessing ProtocolsF. DRYING
Purge all channels with air until dry
Flush all channels, including accessory channels, with alcoholuntil the alcohol until the alcohol can be seen exiting theopposite end of each channel
i. 70% isopropyl alcohol is used to assist in drying the interiorchannel surfaces
ii. Use alcohol that has been properly stored in a closed containerbetween usesalcohol, when exposed to air, rapidly evaporates,and if below the recommended percentage level, cannot berelied upon to assist in the drying process
iii. Alcohol flushes should be used even when sterile water is usedfor rinsing
Purge all channels with air. Alcohol mixes with the remaining water onthe channel surfaces and acts to encourage evaporation of the residualwater as air flows through the channel
-
8/14/2019 Endoscope Reprocessing
32/34
Endoscope Reprocessing Protocols
F. DRYING Purge all channels with air. Alcohol mixes with the remaining
water on the channel surfaces and acts to encourageevaporation of the residual water as air flows through thechannel
Remove all channel adapters
Dry the exterior of the endoscope with a soft, clean lint-freetowel
Thoroughly rinse and dry all removable parts. Do not attachremovable parts to the endoscope during storage as this lowersthe risk of trapping liquid inside the instrument
-
8/14/2019 Endoscope Reprocessing
33/34
Endoscope Reprocessing ProtocolsF. DRYING
Drying the endoscope after every reprocessing cycle, bothbetween patient procedures and before storage is a requisitepractice crucial to the prevention of bacterial transmission
Drying is as important to the prevention of diseasetransmission as cleaning and high level disinfection
G. STORAGE
Hang the endoscope vertically, with the distal tip hanging freely in aclean, well-ventilated dust-free area
A storage area with good ventilation will encourage continued air drying
of the surfaces, and prevent undue moisture build-up, thusdiscouraging any microbial contamination
Caps, valves and other detachable components should be removedduring storage and reassembled before use
Colonoscopes have a minimum shelf life of 7 days, if stored dry
-
8/14/2019 Endoscope Reprocessing
34/34
Endoscope Reprocessing Protocols
Thank you...
Questions?