Endoscope Reprocessing

download Endoscope Reprocessing

of 34

Transcript of Endoscope Reprocessing

  • 8/14/2019 Endoscope Reprocessing

    1/34

  • 8/14/2019 Endoscope Reprocessing

    2/34

    Endoscopy Reprocessing

    THETA Education Day

  • 8/14/2019 Endoscope Reprocessing

    3/34

    Endoscope Reprocessing

    Device Classification

    Manual Cleaning

    Personal Protective Equipment

    Biofilms within GI Endoscopy

    Reprocessing Room Standards

    Endoscope Reprocessing Protocols

    Leakage Testing

    Endoscope Handling

  • 8/14/2019 Endoscope Reprocessing

    4/34

    Endoscope Reprocessing

    Review of Terms

    Biofilmrefers to a complex community of microorganisms thatform a matrix of extracellular material composed ofexopolysaccharides (EPS)

    Minimum Effective concentration (MEC)refers to the lowestconcentration of active ingredient necessary to meet the label claimof a reusable high-level disinfectant / sterilantchemical teststrips should be used to determine whether and effectiveconcentration of the active ingredient is present despite repeated

    use

    Reuse-liferefers to a statement by the manufacturer indicatingthe maximum number of days of a reusable high-level disinfectant/ sterilant might be effective

  • 8/14/2019 Endoscope Reprocessing

    5/34

    Endoscope Reprocessing

  • 8/14/2019 Endoscope Reprocessing

    6/34

    Endoscope Reprocessing

    Two classifications of Endoscopes

    Critical Endoscope:

    Endoscopes used in the examination of critical spaces such as jointsand sterile cavities. Many of these endoscopes are rigid with nolumen. Examples of critical endoscopes are arthroscopes and

    laproscopes

    Semicritical Endoscope

    Fiberoptic or video endoscopes used in the examination of the hollowviscera. These endoscopes generally invade only semicritical spaces,although some of their components might enter tissue or other critical

    spaces ie. Forceps. Examples of semi-critical endoscopes arecolonoscopes, gastroscopes, duodenoscopes, sigmoidoscopes,laryngoscopes, nasopharangeal endoscopes and enteroscopes.

    Opininons differ regarding the reprocessing requirements ofbronchoscopes; a minimum of high level disinfection is required

  • 8/14/2019 Endoscope Reprocessing

    7/34

    Medical Device Classification

    Spaulding Classification

    Determines the degree of disinfection or sterilization required forvarious medical devices.

    Critical: A device that enters normally sterile tissue, including thevascular system. These devices should be sterilized

    1)Semicritical: A device that comes into contact with intact mucous

    membranes and does not ordinarily penetrate sterile tissue. Thesedevices are cleaned followed by High Level Disinfection

    2)Noncritical: Devices that do not ordinarily touch the patient ortouch only intact skin. These devices may be cleaned by low-leveldisinfection

  • 8/14/2019 Endoscope Reprocessing

    8/34

    Manual Cleaning

    Manual Cleaning

    Thorough and meticulous manual cleaning of all instruments mustprecede exposure to any high-level disinfectant or sterilant. Thisprocess significantly reduces the organic and microbial challenge tothe high-level disinfectant or sterilant. An item that has not beencleaned cannot be assuredly disinfected or sterilized.

    Refer to endoscope manufacturers guidelines for design featuresunique to a particular instrument

    Personal Protective Equipment

    Should be used when reprocessing endoscopes. Gowns, gloves andprotective eyewear are recommended when handling any high-leveldisinfectant / sterilant

  • 8/14/2019 Endoscope Reprocessing

    9/34

    Personal Protective Equipment

    Personal Protective Equipment

    Gownsimpervious to fluid, long sleeves that fit snugly aroundthe wrist, and wrap to cover as much of the body as possible.Dispose of or launder gowns if they become wet or are exposed

    to contaminated materialGlovesinspect for tears or holes before use. Gloves should be

    long enough to extend up the arm to protect the forearm orclothing from splashes or seepage. To prevent cross-contamination, change gloves and wash hands whenever moving

    from a dirty to clean task or environmentEye and / or face protection are necessarycontact lenses are

    not sufficient eye protection. A face shield is recommended. Donot use high filtration masks since they may actually trapvapours.

  • 8/14/2019 Endoscope Reprocessing

    10/34

    Biofilms

    Biofilms within GI Endoscopy

    A biofilmis a structured community of microorganismsencapsulatedwithin a self-developed polymeric matrix and adherent to a living orinert surface.If the colonists are not immediately separated from the surface, theycan anchor themselves more permanently using cell adhesionBiofilms within endoscopy may form within endoscopes if propermanual pre-cleaning and CSGNA guidelines are not followed forendoscope reprocessing

    Staphylococcus

    aureus biofilm

    Exopolysaccharide (EPS)

    Staphylococcus aureus

    http://en.wikipedia.org/wiki/Microorganismhttp://en.wikipedia.org/wiki/File:Staphylococcus_aureus_biofilm_01.jpghttp://en.wikipedia.org/wiki/Microorganism
  • 8/14/2019 Endoscope Reprocessing

    11/34

    Reprocessing Room Standards

    The process and products used for cleaning, disinfection and

    or/sterilization of endoscopes must be compatible with theequipment being used

    Each health care setting in which endoscopic procedures areperformed should have written detailed procedures for thecleaning and handling of endoscopes

    Reprocessing of contaminated patient equipment should bedone in an area designated and dedicated for this function

    This room should be separate from where endoscopicprocedures are performed

    Ventilation must be capable of removing toxic vapours

    generated by, or emitted from, cleaning or disinfectant agentsthe vapour concentration of the chemical disinfectant beingused should not exceed allowable limits (eg. 0.05 ppm forglutaraldehyde)

    Minimum of 10-12 air exchanges per hour in the reprocessing

    area

  • 8/14/2019 Endoscope Reprocessing

    12/34

    Reprocessing Room Standards

    Tap water and / or water that has been filtered by passage througha 0.2micron filter or water of equivalent quality should be availablein the reprocessing area

    Manual cleaning should include a medical grade, low-foaming,neutral pH enzymatic formulated for endoscopes that contain

    enzymes to digest all components of bioburden; including, blood, fat,carbohydrate, uric acid, starch

    Accessories

    Accessories which are classified as critical devices (e.g. biopsy

    foceps) require sterilization. Critical items labeled for single-useshould not be reprocessed and / or reused.

  • 8/14/2019 Endoscope Reprocessing

    13/34

    Transportation and Handling of

    Contaminated Endoscopes

    Covered containers with easily cleanable surfaces should beused for handling and transporting soiled endoscopes

    Soild endoscopes should be transported by direct routes where

    cleaning will be performed

    Containers used to transport soiled endoscopes should becleaned after each use

  • 8/14/2019 Endoscope Reprocessing

    14/34

    Endoscope Reprocessing Protocols

    Basic steps to clean and perform high-leveldisinfection of gastrointestinal endoscopes

    1) Pre-cleaning

    2) Leakage testing3) Cleaning

    4) Rinsing

    5) Disinfection

    6) Rinsing

    7) Drying

    8) Storage

  • 8/14/2019 Endoscope Reprocessing

    15/34

    Endoscope Reprocessing Protocols

    A. Pre-cleaning

    1) Immediately after removal of the insertion tube from thepatient and prior to disconnecting the endoscope from thepower source

    Prepare for bedside cleaning:

    PPE Container with enzymatic Sponge or lint-free cloth Air and water channel cleaning adapters per manufacturers

    instruction

    Protective video cap

    2) Wipe the insertion tube with the wet cloth or sponge soakedin the freshly prepared enzymatic solution. Note that thecloth / sponge should be disposed of between cases

  • 8/14/2019 Endoscope Reprocessing

    16/34

    Endoscope Reprocessing Protocols

    A. Pre-cleaning

    3) Place the distal end of the endoscope into the enzymatic solution.Suction the solution through the biopsy / suction channel, alternatesuctioning enzymatic solution and air several times until thesolution is visibly cleanFinish by suctioning air

    Note:Alternate suctioning of fluid and air is more effective thansuctioning fluid alone in the removal of debris from lumens immediate flushing of the biopsy/suction and air/water channelsprecludes drying of debris on lumen surfaces

  • 8/14/2019 Endoscope Reprocessing

    17/34

    Endoscope Reprocessing Protocols

    A. Pre-cleaning

    4) Flush or blow out air and water channels in accordance withthe endoscope manufacturers instructions

    5) Flush the auxiliary water channel

    6) Detach the endoscope from the light source and suction pump7) Attach protective video cap if using a video endoscope

    8) Transport the Endoscope to the reprocessing area in anenclosed container

    Note: Containers, sinks, and basins should be large enough thatthe endoscope will not be damaged by being coiled too tightly

  • 8/14/2019 Endoscope Reprocessing

    18/34

    Endoscope Reprocessing Protocols

    B. Cleaning the Endoscope in the Reprocessing Area

    Prepare the following:

    PPE Leakage testing equipment Channel cleaning adapters

    Large basin of endoscope detergent solution Channel cleaning brushes Sponge or lint-free cloth

    C. Leak Testing

    Leak Testing detects damage to the interior or exterior of theendoscope

    The leak test is done before immersion of the endoscope inreprocessing solutions to minimize damage to parts of theendoscope not designed for fluid exposure.

  • 8/14/2019 Endoscope Reprocessing

    19/34

    Endoscope Reprocessing Protocols

    C. Leak Testing

    1. Manual Leak Testing

    Remove suction valves, air water valves, and biopsy valves

    Attached the leak tester and pressurize the scope beforesubmerging it in water

    With the pressurized insertion tube completely submerged,flex the distal portion of the scope in all directions, observingfor bubbles

    Submerge the entire endoscope and, observing the control

    head of the scope, depress the freeze and release buttons.

    Check the insertion tube and distal bending section as well asthe universal cord for bubbles coming from the interior of thescope

  • 8/14/2019 Endoscope Reprocessing

    20/34

    Endoscope Reprocessing Protocols

    C. Leak Testing

    1. Computerized leak testing Remove suction valves, air water valves, and biopsy valves Attach the leak tester to the computer unit Input data including scope ID and user Move knobs and depress the freeze and release buttons when

    indicated Reprocess when test is complete

    Follow the endoscope manufacturersinstructions if a leak or high humidity

    is detected or if the endoscope appearsdamaged

  • 8/14/2019 Endoscope Reprocessing

    21/34

    Existing Manual Leak Test Methods

    Two Methods: Wet Test & Dry Test Both methods should take ~3 minutes

    of uninterrupted technician time if done properly

    Human Error Factors:

    Takes skill, commitment and dedication

    Rushing

    Leak detection problems

    Lack of consistency and/or training

    Manual Equipment Error Factors:

    No automated detection

    100% reliance on visual leak observation

    Lack of procedure control

    No record keeping

    Endoscope Reprocessing Overview

  • 8/14/2019 Endoscope Reprocessing

    22/34

    Fluid Invasion

    Pathogen inpatient #1

    enters

    scope via

    leak

    Cross-patient

    infection

    breakout

    e.g. Pseudomonas,Salmonella, HIV,

    Hepatitis B and C,

    Staph, E. coli, C. diff

    Leak

    harbors

    pathogen

    from

    disinfection

    Pathogen

    dislodged in

    subsequent

    procedures

    into other

    patients

    *APICGuideline for

    infection

    prevention and

    control in

    flexible

    endoscopy

    Volume 28,

    number 2, p.145

  • 8/14/2019 Endoscope Reprocessing

    23/34

    Endoscope Reprocessing Study Results

    1%of endoscopy procedures result in the endoscope

    developing a leak

    65%of leaks are currently detected (35% undetected) and

    undetected leaks lead to fluid invasion

    60%of endoscope repair costs are fluid invasion-related

    .2%of endoscopy procedures result in fluid invasion

    without a leak present (usually due to poor handling

    such as cleaning cap left off or poorly sealed)

    10-15%of patient-ready endoscopes possess a leak

    Endoscope Reprocessing Overview

  • 8/14/2019 Endoscope Reprocessing

    24/34

    Endoscope Reprocessing Protocols

    D. Cleaning Solutions

    Composition of soil found on endoscopes includes, proteins, fats,carbohydrates and the various chemical salts that exist in bloodand other body fluids.

    Ideally, a cleaning solution should have a broad spectrum of

    effectiveness against these various contaminants and not harmthe device being cleaned

    i) Enzymatic cleaning solutions use surfactants to breakdown anddigest bioburden. They are specifically selected to have anegligible effect on surface tension while still suspending soilparticles. This feature provides easy rinsibility.

  • 8/14/2019 Endoscope Reprocessing

    25/34

    Endoscope Reprocessing Protocols

    E. Cleaning

    Manual cleaning of endoscopes is necessary immediatelyafter removing the endoscope from the patient and priorto automated or manual disinfection

    Performed as the first and most important step in removing themicrobial burden from an endoscope

    Retained debris may inactivate or interfere with the capability ofthe active ingredient of the chemical solution to effectively killand /or inactivate microorganisms

  • 8/14/2019 Endoscope Reprocessing

    26/34

    Endoscope Reprocessing Protocols

    E. Cleaning

    1. Fill a sink with freshly prepared solution of water and a medicalgrade, low-foaming, neutral pH detergent formulated forendoscopesEnzymatic detergent must be discarded after eachuse.

    2. Dilute and use according to the detergent manufacturersinstructions.

    3. Immerse the endoscope

    4. Wash all the debris from the exterior of the endoscope bybrushing and wiping the instrument while submerged in thedetergent solution. Note that the instrument should be leftunder water during the cleaning process to prevent splashing ofcontaminated fluid and aerosolization of bioburden

  • 8/14/2019 Endoscope Reprocessing

    27/34

    Endoscope Reprocessing Protocols

    E. Cleaning

    5. Use a small, soft brush to clean all removable parts, including insideand under the suction valve, air/water valve, and biopsy port cover andopenings.

    6. Brush all accessible endoscope channels including the body, insertiontube and the umbilicus of the endoscopebrushes used for cleaninglumens should be of an appropriate size, inspected before and after useand discarded or cleaned, high-level disinfected and dried following use

    7. After each passage, rinse the brush in enzymatic solution, removing anyvisible debris before retracting and reinserting it

    8. Continue brushing until there is no debris visible on the brush9. Cleanining items should be disposable or thoroughly cleaned and

    minimum of high-level disinfected between cases.

  • 8/14/2019 Endoscope Reprocessing

    28/34

    Endoscope Reprocessing ProtocolsE. Cleaning

    10. Attach the endoscope cleaning adapters for suction, biopsy, air, andwater channelsNote:Automated pumps are available for this step that eliminate

    the manual flush

    11. Attach the manufacturers cleaning adapters for special endoscope channels(dual channel, elevator channel, auxilliary channel)

    i. To achieve adequate flow through all lumens, various adapters or channelrestrictors may be required

    ii. The elevator channel of a duodenoscope is a small lumen, this channelrequires manual reprocessing using a 2-5mm syringe

    12. Flush all channels with the detergent solution to remove debris

    13. Soak the endoscope and its internal channels for the period of time specifiedby the label

    14. If immediate reprocessing is not possible the endoscope shouldbe leak-tested, flushed, brushed, and allowed to soak in aenzymatic solution until it can be thoroughly reprocessed

  • 8/14/2019 Endoscope Reprocessing

    29/34

    Endoscope Reprocessing Protocols

    F. Rinse After Cleaning

    Thoroughly rinse the endoscope and all removable parts withclean water to remove

    Purge water from all channels using forced air. Dry the exteriorof the endoscope with a soft, lint-free cloth to prevent dilution of

    the liquid chemical germicide used in subsequent stepsG. High Level Disinfection

    Recognized as the standard of reprocessing for endoscopes by:

    CSGNA / SGNA

    CPSOCollege of Surgeons of Ontario ASGEAmerican Society for Gastrointestinal Endoscopy ACG - American College of Gastroenterology AGA - American Gastroenterological Association APIC - Association for Professionals in Infection Control and

    Epidemiology

    CDCCenters for Disease Control and Prevention

  • 8/14/2019 Endoscope Reprocessing

    30/34

    Automated Endoscope

    Reprocessing

    Ensure that the endoscope and endoscope components arecompatible with the automated endoscope reprocessor (AER)

    Follow the OEM instructions for use in the AER

    AER provide a method by which a permanent record ofendoscope use and reprocessing can be monitored andvalidated

    Some AER have a system capable of tracking endoscopes andpatients. For each procedure the patients name and recordnumber, the date and time of procedure, type of procedure,the endoscopist and the serial number of the endoscope arerecorded and stored to assist in outbreak investigation.

  • 8/14/2019 Endoscope Reprocessing

    31/34

    Endoscope Reprocessing ProtocolsF. DRYING

    Purge all channels with air until dry

    Flush all channels, including accessory channels, with alcoholuntil the alcohol until the alcohol can be seen exiting theopposite end of each channel

    i. 70% isopropyl alcohol is used to assist in drying the interiorchannel surfaces

    ii. Use alcohol that has been properly stored in a closed containerbetween usesalcohol, when exposed to air, rapidly evaporates,and if below the recommended percentage level, cannot berelied upon to assist in the drying process

    iii. Alcohol flushes should be used even when sterile water is usedfor rinsing

    Purge all channels with air. Alcohol mixes with the remaining water onthe channel surfaces and acts to encourage evaporation of the residualwater as air flows through the channel

  • 8/14/2019 Endoscope Reprocessing

    32/34

    Endoscope Reprocessing Protocols

    F. DRYING Purge all channels with air. Alcohol mixes with the remaining

    water on the channel surfaces and acts to encourageevaporation of the residual water as air flows through thechannel

    Remove all channel adapters

    Dry the exterior of the endoscope with a soft, clean lint-freetowel

    Thoroughly rinse and dry all removable parts. Do not attachremovable parts to the endoscope during storage as this lowersthe risk of trapping liquid inside the instrument

  • 8/14/2019 Endoscope Reprocessing

    33/34

    Endoscope Reprocessing ProtocolsF. DRYING

    Drying the endoscope after every reprocessing cycle, bothbetween patient procedures and before storage is a requisitepractice crucial to the prevention of bacterial transmission

    Drying is as important to the prevention of diseasetransmission as cleaning and high level disinfection

    G. STORAGE

    Hang the endoscope vertically, with the distal tip hanging freely in aclean, well-ventilated dust-free area

    A storage area with good ventilation will encourage continued air drying

    of the surfaces, and prevent undue moisture build-up, thusdiscouraging any microbial contamination

    Caps, valves and other detachable components should be removedduring storage and reassembled before use

    Colonoscopes have a minimum shelf life of 7 days, if stored dry

  • 8/14/2019 Endoscope Reprocessing

    34/34

    Endoscope Reprocessing Protocols

    Thank you...

    Questions?