ECT Supplement
-
Upload
arya-thekul -
Category
Documents
-
view
231 -
download
1
description
Transcript of ECT Supplement
Supplementary Material Emission Computed
Tomography
Thanks to those that post interesting material on the
internet. This supplement is a collection from several.
Emission Tomography
f(x,y,z)f(x,y,z)
ProjectionsProjections
ReconstructiReconstructionon
f(x,y,z)f(x,y,z)
SlicesSlices
projection
projection
SPECTSingle Photon Emission Computed Tomography
collimator
only one gamma photon is detected per decay
NaI(Tl) crystal
Rotating scintillator camera
detector
detector
PETPositron Emission Tomography
What do we want to detect in PET?– 2 photons of 511 keV in coincidence, coming in a
straight line from the same annihilation
e-
e+
unstable nucleus emits positron positron annihilates with electron
two 511 photons are emitted
simultaneously in opposite directions
TRUE coincidence
Types of Coincidence True coincidence is the simultaneous detection of the
two emissions resulting from a single decay event. Scatter coincidence is when one or both photons from
a single event are scattered and both are detected. Random coincidence is the simultaneous detection of
emission from more than one decay event.
Coincidences: True Scatter Random
PET radiation detection PET scanner
– Typical configuration: whole-body (patient port ~60 cm; axial FOV~15 cm) scintillator crystals coupled to photomultiplier tubes (PMTs) cylindrical geometry ~24-32 rings of detector crystals hundreds of crystals/ring several millions of Lines Of Response (LORs)
(only a few are shown)
True True
Other configurations for special-purpose applications:
- brain imaging
- animal PET
- mammography, other
PET CT
PET/CT
CT PET CT+PET
General Electric Medical Systems
PET data acquisition Organization of data
– True counts in LORs are accumulated– In some cases, groups of nearby LORs are
grouped into one average LOR (“mashing”)– LORs are organized into projections
etc…
PET data acquisition 2D and 3D acquisition modes
septa
2D mode (= with septa)
3D mode(= no septa)
In the 3D mode there are no septa: photons from a larger number of incident angles are accepted, increasing the sensitivity.
Note that despite the name, the 2D mode provides three-dimensional reconstructed images (a collection of transaxial, sagittal and transaxial slices), just like the 3D mode!
PET data acquisition 2D mode vs. 3D mode
2D mode (= with septa)
3D mode(= no septa)
True
detected
True
not detected (septa block
photons)
PET data acquisition Organization of data
– In 3D, there are extra LORs relative to 2D
3D mode same planes as 2D oblique planes
...
...
+
+
PET evolution: spatial resolutionHuman brain
Animal PET~1998
Monkey brain
Image credits: CTI PET Systems
Image credits: Crump Institute, UCLA
J. Fessler, 2002
“Part of the goal is to bring order to this alphabet soup.”
PET image reconstruction
Projection
Object
P( 2 ,r)
∫=),(
),(),(rline
dlyxfrPφ
φ
Radon Transform
1
2
P(1 ,r)
f(x,y)
r
r
PET image reconstruction
Sinogram
r
Projection
angle
Projection bin
Object
PET image reconstruction
SinogramObject
r
PET image reconstruction
SinogramObject
r
PET image reconstruction
SinogramObject
r
PET image reconstruction
SinogramObject
r
PET image reconstruction
r
SinogramObject
Sinogram
Other representations can be used instead of the sinogram (linogram, planogram)
SPECT: 360º (1 photon)
PET: 180º (2 opposite photons)
PET image reconstruction 2D Reconstruction
– Each parallel slice is reconstructed independently (a 2D sinogram originates a 2D slice)
– Slices are stacked to form a 3D volume f(x,y,z)
2D reconstruction
Plane 5
Slice 5etc
etc
2D reconstruction
Plane 4
Slice 4
2D reconstruction
Plane 3
Slice 3
2D reconstruction
Plane 2
Slice 2
2D reconstruction
Plane 1
Slice 1
2D Reconstruction2D Reconstruction
PET image reconstruction Projection and Backprojection
Projection Backprojection
2D Reconstruction2D Reconstruction
4 projections 16 projections 128 projections
Backprojection
FilteredBackprojection
Object
PET image reconstruction2D Reconstruction2D Reconstruction
Noise In PET Images Noise in PET images is dominated by the counting
statistics of the coincidence events detected. Noise can be reduced at the cost of image resolution
by using an apodizing window on ramp filter in image reconstruction (FBP algorithm).
105 106 107 counts
Unapodized ramp filter
Hanning window, 4mm
Hanning window, 8mm
Scatteredcoincidences component
Attenuation
Random coincidencescomponent
Detector efficiency
effects Truecoincidencescomponent
PET image reconstruction Data corrections are necessary
– the measured projections are not the same as the projections assumed during image reconstruction
Object(uniformcylinder)
projectionmeasured
projectionassumed
integral of the activity along the line or tube
of response
Analytical methods Advantages
– Fast– Simple– Predictable, linear behaviour
Disadvantages– Not very flexible– Image formation process is not modelled image
properties are sub-optimal (noise, resolution)
Iterative methods
Advantages– Can accurately model the image formation process (use
with non-standard geometries, e.g. not all angles measured, gaps)
– Allow use of constraints and a priori information (non-negativity, boundaries)
– Corrections can be included in the reconstruction process (attenuation, scatter, etc)
Disadvantages– Slow– Less predictive behaviour (noise? convergence?)
PET Image reconstruction Iterative methods
Current Current estimateestimate
Measured Measured projectionprojection
CompareCompare
(e.g. (e.g. –– or or // ) )
Error Error projectionprojection
projectionprojectionEstimated Estimated projectionprojection
image spaceimage space projection spaceprojection space
backprojectionbackprojectionErrorErrorimageimage
UpdateUpdate
Iteration 1
PET Image reconstruction Iterative methods
Current Current estimateestimate
Measured Measured projectionprojection
CompareCompare
(e.g. (e.g. -- or or // ) )
Error Error projectionprojection
projectionprojectionEstimated Estimated projectionprojection
image spaceimage space projection spaceprojection space
backprojectionbackprojectionErrorErrorimageimage
UpdateUpdate
Estimated Estimated projectionprojectionEstimated Estimated projectionprojection
Iteration 2
PET Image reconstruction Iterative methods
Current Current estimateestimate
Measured Measured projectionprojection
CompareCompare
(e.g. (e.g. -- or or // ) )
Error Error projectionprojection
projectionprojection
image spaceimage space projection spaceprojection space
backprojectionbackprojectionErrorErrorimageimage
UpdateUpdate
Estimated Estimated projectionprojectionEstimated Estimated projectionprojectionEstimated Estimated projectionprojectionEstimated Estimated projectionprojection
Iteration N
Algorithm comparison 600 000 counts (including scatter)
original
3DRP Hanning
3D OSEM + filt.6 subsets5 iter.Gauss .5cm
FORE + OSEM
6 subsets2 iter.
3D OSEM
6 subsets2 iter.
Image credits: Kris ThielemansMRC CU, London (now IRSL – www.irsl.org)
Reconstruction of a slice from projectionsexample = myocardial perfusion, left ventricle, long axis
courtesy of Dr. K. Kouris
Iterative reconstruction methodsconventional iterative algebraic methods
algebraic reconstruction technique (ART) simultaneous iterative reconstruction technique (SIRT)
iterative least-squares technique (ILST)
iterative statistical reconstruction methods (with and without using a priori information)
gradient and conjugate gradient (CG) algorithms maximum likelihood expectation maximization (MLEM)
ordered-subsets expectation maximization (OSEM) maximum a posteriori (MAP) algorithms
algorithm (a recipe)
(1) make the first arbitrary estimate of the slice (homogeneous image),
(2) project the estimated slice into projections analogous to those measured by the camera (important: in this step, physical corrections can be introduced - for attenuation, scatter, and depth-dependent collimator resolution),
(3) compare the projections of the estimate with measured projections (subtract or divide the corresponding projections in order to obtain correction factors - in the form of differences or quotients),
(4) stop or continue: if the correction factors are approaching zero, if they do not change in subsequent iterations, or if the maximum number of iterations was achieved, then finish; otherwise
(5) apply corrections to the estimate (add the differences to individual pixels or multiply pixel values by correction quotients) - thus make the new estimate of the slice,
(6) go to step (2).
Iterative reconstruction - multiplicative corrections
Iterative reconstruction - differences between individual iterations
Iterative reconstruction - multiplicative corrections
Filtered back-projection very fast direct inversion of the
projection formula
corrections for scatter, non-uniform attenuation and other physical factors are difficult
it needs a lot of filtering - trade-off between blurring and noise
quantitative imaging difficult
Iterative reconstruction
discreteness of data included in the model
it is easy to model and handle projection noise, especially when the counts are low
it is easy to model the imaging physics such as geometry, non-uniform attenuation, scatter, etc.
quantitative imaging possible
amplification of noise long calculation time
References:Groch MW, Erwin WD. SPECT in the year 2000: basic principles. J Nucl Med Techol 2000; 28:233-244, http://www.snm.org.
Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med Technol 2001; 20:9-15, http://www.snm.org.
Bruyant PP. Analytic and iterative reconstruction algorithms in SPECT. J Nucl Med 2002; 43:1343-1358, http://www.snm.org.
Zeng GL. Image reconstruction - a tutorial. Computerized Med Imaging and Graphics 2001; 25(2):97-103, http://www.elsevier.com/locate/compmedimag.
Vandenberghe S et al. Iterative reconstruction algorithms in nuclear medicine. Computerized Med Imaging and Graphics 2001; 25(2):105-111, http://www.elsevier.com/locate/compmedimag.
References:Patterson HE, Hutton BF (eds.). Distance Assisted Training Programme for Nuclear Medicine Technologists. IAEA, Vienna, 2003, http://www.iaea.org.
Busemann-Sokole E. IAEA Quality Control Atlas for Scintillation Camera Systems. IAEA, Vienna, 2003, ISBN 92-0-101303-5, http://www.iaea.org/worldatom/books, http://www.iaea.org/Publications.
Steves AM. Review of nuclear medicine technology. Society of Nuclear Medicine Inc., Reston, 1996, ISBN 0-032004-45-8, http://www.snm.org.
Steves AM. Preparation for examinations in nuclear medicine technology. Society of Nuclear Medicine Inc., Reston, 1997, ISBN 0-932004-49-0, http://www.snm.org.
Graham LS (ed.). Nuclear medicine self study program II: Instrumentation. Society of Nuclear Medicine Inc., Reston, 1996, ISBN 0-932004-44-X, http://www.snm.org.
Saha GB. Physics and radiobiology of nuclear medicine. Springer-Verlag, New York, 1993, ISBN 3-540-94036-7.