Jan Visser Detector R&D at Nikhef Gaseous detectors & semiconductor detectors

Transcript of Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower...

Page 1: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Jan Visser

Detector R&D at Nikhef

Gaseous detectors & semiconductor detectors

Page 2: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Outline

• Nikhef

• Medipix chips

• Gaseous detectors

• Semiconductor detectors

• Future developments

14 February, 2012 2 Jan Visser

Page 3: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Detector R&D at

• Experiment specific

– Atlas, LHC-b, KM3Net, etc.

• Technology driven

– Gaseous detectors post-processed

on ASICs

– Tilable semiconductor detectors

• Focus on

– Chip design

– Read-out electronics

– Mechanics

– Cooling (CO2)

– Assembly of modules and systems

– Characterising sensors, chips, systems

14 February, 2012 3

LHC-Atlas Forward Silicon Detector

Jan Visser

Page 4: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Facilities

• Clean room

– Probe station; wafer probing

– Keithley; sensor characterisation (IV, CV)

– SEM; fault analysis (dice, polish, Au coating)

– Wire-bonder; shear and pull tester

• Lab

– X-ray tube (60kV, pulsed)

• Future

– X-ray tube (100kV, continuous, tuneable)

– Laser for semiconductor tests (inject charge)

– Cosmic-ray set-up

14 February, 2012 4 Jan Visser

Page 5: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Test and assembly lab infrastructure

Scanning Electron Microscope

Probe station Wire bonder

14 February, 2012 5 Jan Visser

Shear / pull testing

Page 6: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Gas vs Semiconductors

• Gaseous detectors

– Low material budget

– Vectors instead of points

• Semiconductors

– Higher stopping power

– Proven technology

– Strips, pixels, 3D,...

• Issues

– Reliability

– Radiation hardness

– Cooling

14 February, 2012 6 Jan Visser

Page 7: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix

14 February, 2012 7 Jan Visser

Courtesy of PANalytical

Page 8: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix based hybrid detectors • Photon counting principle (E > 4keV = 1000 e-) • Large dynamic range • Pixel size: 55 x 55 μm2 (256 x 256 pixels) • Active area per chip is ~ 14 x 14 mm2 • I/O periphery (wire bonded)

8

Pixelated 300 μm silicon sensor (256 x 256 pixels)

Medipix2 readout ASIC

Detector bias voltage

14 mm

14 February, 2012 Jan Visser

Page 9: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix chips

• Medipix1 (1µm)

– First attempt in late 1990s to take HEP results to other applications

• Medipix2 (0.25µm)

– Started in 1999 and still going strong!

– First commercial application at PANalytical in x-ray diffraction field

• Timepix (0.25µm)

– Update of MPX2 for timing purposes in 2006

• Medipix3 (0.13µm)

– New generation, solving charge sharing issues

– Under test in 2010; a lot of progress, some issues to solve/fix

14 February, 2012 9 Jan Visser

Page 10: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2 operation principles

Jan Visser 14 February, 2012 10

Page 11: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Timepix operation principles

Jan Visser 14 February, 2012 11

Counting mode

• Count number of events over threshold

Energy information

• Count while signal is over threshold

Arrival time

• Time from hit till close of shutter

Page 12: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2

Jan Visser

Threshold equalisation mask Masked pixels

Setting up the Medipix2 chip for data taking:

• Equalising pixels; i.e. same reaction to noise level

• Switching off dead pixels: up to few hundred

14 February, 2012 12

Page 13: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2

Jan Visser

Noise measurement 55Fe source

Taking data with the Medipix2 chip with a silicon sensor

90Sr source

14 February, 2012 13

Page 14: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2

Jan Visser

Noise measurement Taking data with the Medipix2

chip with a silicon sensor:

- Noise hits can be suppressed

- Cosmic rays will always occur

Algorithms to remove the

tracks of cosmic rays in certain

imaging applications

14 February, 2012 14

Page 15: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2 versus Medipix3 Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge between neighbouring pixels and assigning the sum to the pixel with the highest energy deposition.

• Cluster of 4 pixels

(110 x 110 μm2)

• Charge Summing

• Wider distribution due

to noise summing.

Jan Visser 15 14 February, 2012

Page 16: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Chip design at

• Gossipo

– Development of fast timing circuits

for read-out of gaseous detectors

• Kenniswerkers

– 2FTE from Bruco (chip design-house in Borne)

– Development of high-speed read-out

• From 1Gb/s to 10Gb/s

• Test chip planned for summer 2010

– Asynchronous read-out chips

• Read out only those pixels that were hit,

instead of all of them!

• Compress data sample

14 February, 2012 16 Jan Visser

X-ray diffraction pattern

Page 17: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Gaseous Detectors

14 February, 2012 17 Jan Visser

Page 18: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Gas chamber

Anode

Drift volume

Grid

Amplification

Protection layer

CMOS chip

14 February, 2012 18 Jan Visser

1.2mm

Field strengths

– Drift region: 300 – 3kV/cm

– Amplification: 400 – 600V/50µm (80 – 120kV/cm)

Gas detectors on Timepix

Page 19: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Gas detectors on Timepix

14 February, 2012 19 Jan Visser

Page 20: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 20

Gossip

• Gaseous detector – Start with Timepix chip

– Build pillars (SU8) to support grid

– Place aluminium grid on top

– Assemble onto PCB

– Build into system providing anode

Jan Visser

Page 21: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

GEM grid

• Instead of pillars with grid on top, make holes in full layer to support grid.

• To improve the gain the holes in the metal sheet were given a smaller diameter than the holes in the SU8

14 February, 2012 21 Jan Visser

Page 22: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Discharges

• Unavoidable in gaseous detectors

• ASICs can’t withstand these discharges

• Protection required

14 February, 2012 22 Jan Visser

Page 23: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Discharge protection

Protection layer made of Si3N4

• only 7 µm thick

• high resistivity bulk material

14 February, 2012 23 Jan Visser

5 layers of Si3N4 Damage through pinholes

Page 24: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Twin grid

14 February, 2012 24 Jan Visser

•Twin grid to have additional tuning parameter • Two lower gains

• One high gain and one low gain

• Aim to reduce chance of discharges and possible damage

Page 25: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 25

Applications

• Xenon – Position sensitive Xe TPC

– WIMP-nucleon cross-section

– Direct Dark Matter Search

Jan Visser

• Polapix – Detect polarised x-rays

– e.g. from black holes

– Future satellite mission

• High energy physics – Tracking

Page 26: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Photosensitive GridPix

InGrid with CsI on Al anode UV light 200-400 nm UV well absorbed by fingerprint

InGrid (Al) coated with 200 nm CsI

transfer of primary e−

photon

mesh cathode

14 February, 2012 26Jan Visser

Univ. Twente and Weizmann institute

Page 27: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 27

Gaseous detectors status & future

Jan Visser

Status

• Tracks can be constructed really well

• Resolution in agreement with simulations

• Protection layer big step forward

Future

• Construction

• Investigate different materials for more robustness

• Make larger area for ILC-TPC

• Tests

• Tests at Nikhef (source and cosmics)

• Test beam at DESY and CERN (GeV particles)

Page 28: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 28 Jan Visser

Semiconductor Detectors

Page 29: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

• Hybrid medipix pixel detectors advantages – Photon counting

– No electronic noise (> 4keV)

– High sensitivity, large dynamic range and low contrast detectability; • Possible dose reduction

– Better than many detectors in use in many fields

– High speed imaging and readout

• Issues – Radiation hardness (Synchrotron applications)

– Small! (1.4cm x 1.4cm)

– Periphery on one side

• Solutions – Current design (0.13µm) very radiation tolerant

– Make them bigger!

14 February, 2012 29 Jan Visser

Medipix2 Hybrid detector

Page 30: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Technologies:

• Edgeless sensor technology (Canberra) • Through Silicon Vias and Ball-Grid-Array (IMEC) • Ceramic printed circuit board (Nikhef) • 1 Gbit/s ethernet data transmission (Nikhef) • CO2-cooling (Nikhef) • Product development (PANalytical)

Large Area Detectors

30

RELAXD: high Resolution Large area X-ray detector

14 February, 2012 Jan Visser

Page 31: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

1. Reduce the dead area around the sensor edgeless sensors. 2. Decrease the Medipix periphery design & dicing. 3. Through-Silicon-Vias replacement of wire bonds.

Tiling

TSV BGA

(Ball-Grid-Array)

(Through-Silicon-Via)

31 14 February, 2012 Jan Visser

Page 32: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Reduction of edge induced currents

Suppression: Guard Ring vs. Stop Ring

14 February, 2012 32 Jan Visser

Page 33: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

TecPlot – visualisation

No stop ring 50 µm wide stop ring

Hole current density as a function of the surface recombination at the edge

(Changing parameter: electron and hole recombination velocity)

14 February, 2012 33 Jan Visser

Page 34: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Traditional blade dicing

Dicing methods Stealth dicing (laser)

Deep Reactive Ion Etching

14 February, 2012 34 Jan Visser

Page 35: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Dicing techniques results

DRIE dicing

Standard

dicing

Stealth dicing

14 February, 2012 35 Jan Visser

Page 36: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Proof-of-concept

• 51 edgeless samples

• 40 % of them showed an acceptable leakage current density in the active area

• 45 % of the sensors with p+ SR showed leakage current reduction due to bias on the SR

Effect of biasing the Stop Ring Edgeless vs. Conventional

14 February, 2012 36 Jan Visser

Page 37: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

50µm

Bonding device wafer face down

on temporary carrier

Thinning down to 50-100um

Annular silicon DRIE

stopping on oxide

Dielectric polymer fill of annular

trenches

Dielectric opening

at the top surface

DRIE/wet etching of center Si

block + RIE of oxide

Seed layer deposition and Cu

electroplating

Through-Silicon-Vias

35µm

min 60µm pitch

14 February, 2012 37 Jan Visser

Page 38: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Printed circuit boards Board requirements

• Through-silicon-via chips are thinned

• Thermal expansion coefficients play a role!

• Ceramic close to silicon in thermal behaviour

• Looking into stablcor multi layer boards

• Vacuum applications for low energy ion or electron detection

• Ceramic is vacuum compatible

• Liquid crystal polymer coated FR4 works well

14 February, 2012 38 Jan Visser

Page 39: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 39

Hybrid detector challenges Status

• Edgeless sensors under development

– Need to position with 10µm precision

– Dicing accuracy crucial (both sensor and ASIC)

– Design sensor with four chips underneath with identical pixels

• Hybridisation

– Periphery obstacle to full tiling possibility

– Pitch adjustment between sensor and ASIC still an issue

Jan Visser

Page 40: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Hybrid detectors future

Separate functionality

• Sensors – Precisely diced edgeless

– Uniform pixel matrix

• Hybridisation – Get rid of periphery

– Separate analogue and digital

(select optimal technology)

– Thinned ASICs

– Through-wafer-vias

– Ball-grid-array

– Wafer to wafer bonding

14 February, 2012 40 Jan Visser

analogue

sensor

ceramic

digital

Gb ethernet

Page 41: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medical x-ray application

14 February, 2012 41

Interventional x-ray imaging

• large area

• little gaps

• efficient detection of

~100keV x-rays

Jan Visser

Hidralon

• Collaboration with

Philips Healthcare in Best

• Development of CdTe

detectors with minimal edge

Page 42: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Sensor materials • Silicon

• Relatively light material

• Abundant in good quality

• Works well for

• Tracking purposes

• Low energetic X-rays < 20keV

• “Transparent” for high energetic X-rays 20-120keV

• Alternatives

• Cadmium Zinc Telluride

• Gallium Arsenide

• Challenges

• Good material scarce and difficult to make

• Small wafers

• Nasty materials; brittle and poisonous

14 February, 2012 42 Jan Visser

Page 43: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 43 Jan Visser

0

0.2

0.4

0.6

0.8

1

1.2

10 15 20 30 40 50 60 80 100 150 200

Ab

so

rpti

on

eff

icie

ncy

X-ray photon energy (keV)

Si

CdZnTe

GaAs

Sensor materials

Page 44: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

MARS: Medipix All Resolution System

Phase Contrast Angiography on mouse: GREEN: Calcium – bone RED: Iodine contrast in blood vessels PURPLE: Barium in the lungs

Spectroscopic mode: up to 8 thresholds (7 energy windows)

Images emulated with multiple exposures of mouse by Medipix2

14 February, 2012 44/41 Jan Visser

Page 45: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix projects summary • Medipix3

– Test new features

– Read-out 1Gb/s (10gb/s)

• Timepix2

– Next generation with improved timing

– Data driven read-out

• Photon detection

– Photocathode + MCP’s + TPX

• Ion or electron detection

– MCP’s + MPX or TPX @ Amolf, Leiden, TUT

14 February, 2012 45 Jan Visser

Advance with Medipix in LEEM (Leiden)

Page 46: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Gaseous detector summary

• Gridpix

– Post-processing structures on TPX chips

– Signal formation studies

– Position resolution

– Simulations

• Gridpix projects

– Gossip for Atlas tracking

– TPC for vertexing at ILC

– Xenon; GridPix as a single electron detector in bi-phase Xenon filled

WIMP search experiments

– Polapix; measuring polarisation of x-rays in space

14 February, 2012 46 Jan Visser

Page 47: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Semiconductors summary

• Relaxd

– Edgeless silicon sensor

– Through-silicon-vias

– 1Gb/s read-out

• Hidralon

– Edgeless CdZnTe sensors

– Simulations

– 10Gb/s read-out

14 February, 2012 47 Jan Visser

Page 48: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Thank you for your attention

14 February, 2012 Jan Visser 48

Page 49: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 Jan Visser 49

Page 50: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

3D view of discharge current

14 February, 2012 50 Jan Visser

Page 51: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 51

Edgeless sensors

• Characterisation – Example of dicing techniques:

– Standard with blade results in chipping: stay mm’s from edge

– Deep Reactive Ion Etching (DRIE); possible 50-100 micron from edge

– Stealth Dicing works well for MPX chips, try on sensors

DRIE dicing Standard dicing

Jan Visser

Page 52: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Mouse pelvis and surrounding soft tissue structures demonstrates extremely high contrast of images taken by Medipix2 Even structure of overlaying hair is seen through at the side of mouse’s body

Jan Visser 52

Thanks to the Czech Technical University in Prague

14 February, 2012

Medipix2 x-ray radiography

Page 53: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Computed Tomography reconstruction

Thanks to the University of Canterbury, Christchurch, NZ

GREEN: Calcium - bone BROWN: Iodine contrast in the stomach YELLOW: Intestine

Jan Visser 53 14 February, 2012

Page 54: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

5-10µm Cu

100µm Si wafer

500µm

400µm (80% of the ball size)

5µm of polymer

5µm of BCB

3-5µm of Ni, 500nmAu

Ball-Grid-Array

14 February, 2012 54 Jan Visser

Page 55: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

XFEL Experiments

• Coherent imaging of “small things” (structure)

• Photon Correlation Spectroscopy (time evolution)

• Pump-Probe diffraction (non-equilibrium states)

• Small Quantum systems

• Warm dense matter (“plasmas”)

• Unexpected effects

Physics for AMOLF, Rijnhuizen, Bio-medical, Pharma,

… But Detector R&D compatible with HEP-ILC !!

14 February, 2012 55 Jan Visser

Page 56: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

HPAD: a pixel detector for XFEL

- 1Mpixel: 1024´1024 pixel

- 4 Quadrants to allow beam hole

- 32 modules for 256´128 pixel

- 8 ASIC's/module: 64´64pixel

-pixel: 200µm´200µm

Module:

size: ~ 51.2´25.6mm2

pixels: 32K

XFEL Current first dates for: 1st beam injector ~ end. 2012

1st beam Linac ~ end. 2013

1st SASE(1) at 0.2nm ~ end. 2014

14 February, 2012 56 Jan Visser

Page 57: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Module mechanics

Each interface has to fit behind active area: no depth limitation

Separate PCB constructions for low-EMI analogue and high dense digital

Heat per module: 40W

Principle quite similar to our RELAXD module!

14 February, 2012 57 Jan Visser

Page 58: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

14 February, 2012 58

Hidralon

• General – New generation of CMOS imagers;

e.g. automotive and entertainment industry

– Large European project (Catrene)

• Nikhef

– Collaboration with

• Philips Healthcare in Best

– Development of Cadmium-Telluride detectors

with minimal edge

– Aim is to efficiently detect X-rays with an energy

between 30 and 120keV in a tiled area, ultimately 20x30cm2

– First CdTe has arrived, mount on MPX chips for testing in 2010

CdT detectors

Jan Visser

Page 59: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Philips Healthcare: human diagnostics

14 February, 2012 59

Interventional x-ray imaging

requires large area with little

gaps and efficient detection

of ~100keV x-rays

Jan Visser

Page 60: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Characteristics

• Direct X-ray conversion in semiconductor sensor • Photon counting within given energy region (Eγ > 4 keV) • High maximum count rates • High sensitivity, large dynamic range and low contrast detectability; • potential dose reduction • Noise-free • No sensitivity to dark currents • High speed imaging and readout

Larger area coverage is needed. Enlargement of the chip size, however, leads to a degradation of the production yield. The solution is TILING. 3-side buttable limits the Medipix2 tiling possibilities to 2*N array configuration.

Limitation:

Characteristics:

60 14 February, 2012 Jan Visser

Page 61: Detector R&D at Nikhef · Charge sharing: The spectrum of a single pixel is distorted towards lower energies. Medipix3 readout architecture: Mitigating effect by summing the charge

Medipix2 Hybrid detector

• Direct X-ray conversion in semiconductor sensor • Photon counting within given energy region (Eγ > 4 keV)

• No sensitivity to dark currents • Noise-free

• High maximum count rates • High sensitivity, large dynamic range and low contrast detectability;

• Potential dose reduction • High speed imaging and readout

Characteristics:

Jan Visser 14 February, 2012 61